Nanoentek EVE Plus User Manual

Automated cell counter
User Manual
Copyright NanoEnTek, Inc. All rights reserved.
All the materials in this user manual are protected by Korean and international copyright laws. They cannot be reproduced, translated, published or distributed without the permission of the copyright owner.
EVE™ PLUS User Manual
Website: www.nanoentek.com E-mail: sales@nanoentek.com
NanoEnTek, Inc.
851-14, Seohae-ro, Paltan-myeon, Hwaseong-si, Gyeonggi-do, 18531, Korea Tel. +82-2-6220-7940 Fax. +82-2-6220-7999
NanoEnTek America, Inc.
240 Bear Hill Road, Suite 101, Waltham, MA 02451, USA Tel. +1-781-472-2558 Fax. +1-781-790-5649
The information contained in this document is subject to change without notice. Firmware and software changes and updates may change without prior consent or notication.
Copyright © 2019 by NanoEnTek Inc.
All rights reserved. Published in Korea.
Documentation: NESMU-EVE2-001E
Revision history: V.0.0 JUN 2019
Table of contents
Introduction
Product contents
Product overview
Features and benets
Product description
Front view
Side view
Rear view
Installation
Install EVETM PLUS
Activation code
Setting
Setting menu
Mode
Parameter
Calibration
Date
Operation
Recommend actions
Cell counting (Auto focus)
Cell counting (Manual focus)
Maintenance and cleaning
2
3
4
Troubleshooting
Warranty
Product specications
Ordering information
5
5
6
7
8
9
9
9
12
14
15
16
19
Safety precautions
Safety symbols
Technical support
30
31
33
34
35
36
38
39
DATA
Check the image
Edit the histogram
Data export (Report & image)
Data export (CSV le)
Data export (Mail)
24
26
27
28
29
Introduction
Product contents
EVE™ PLUS is shipped with the following components. Please check that all items listed below were shipped, receiving the instru­ment. If any items are missing or damaged, contact your local distributor or e-mail sales@nanoentek.com.
Automated cell counter
1 EA
Cat. No. EVE-MC2 Cat. No. EVS-050
Power cord
Europe/Korea, USA/Japan, Australia,
UK, or China
1 SET
Cell counting slides
with 1.5 mL of Trypan blue (0.4%)
1 BOX (50 slides/box)
Adaptor
1 SET
User manual
1 EA
User
manual
2
Quick manual
1 EA
Quick
manual
Introduction
Product overview
EVE™ PLUS Automated cell counter uses state-of-the-art optics and image analysis for automatic cell counting. EVE™ PLUS is a benchtop counter de­signed to measure cell count and viability (live, dead, and total cells) accu­rately and precisely, using the standard trypan blue technique.
Using the same amount of sample that you currently use with the hemo­cytometer, EVE™ PLUS takes less than 1 second per sample for a typical cell count with manual focus option and is compatible with a wide variety of eukaryotic cells and provides information on cell size.
EVE™ PLUS is supplied with disposable EVE™ Cell counting slides that con­tain two enclosed slides to hold the sample to allow you to measure two dif­ferent samples or perform replicates of the same sample. The cell count­ing occurs in the central location on the counting slide and the volume counted is 0.4 µL, the same as counting four (1 mm × 1 mm) squares in a standard hemocytometer.
EVE™ PLUS, Automated Cell Counter 3
Introduction
Features and benets
User-friendly, compact design for simple, fast, automated cell count and
viability measurements within 1 second.
It provides data on cell size and is compatible with various types of eu-
karyotic cells, regardless of whether the cell size is small or large.
Measures cell concentrations ranging from 1 × 10
cells with sizes ranging from 5 µm to 60 µm.
Provides the clumpy cell counting function to get more accurate results.
Uses disposable cell counting slides that you can eliminate washing steps
and cross contamination between samples with.
Up to 500 test results are automatically saved in the DATA tab.
Presents comprehensive data with graphical reports and as a .CSV
(comma separated value) le for sample comparisons.
4
to 2 × 107 cells/mL and
4
Product description
Front view
Side view
Item Description
LCD touch screen Located in the front of the instrument contains buttons
Slide port The slide port is used to insert the EVE™ Cell counting
Item Description
Image adjustment
(focus) knob
USB port
Power button
for all the functions needed and displays data from the cell count.
slide containing the sample with trypan blue stain into the counter for analysis.
The image adjustment (focus) knob is used to adjust the image quality to obtain better contrast between live (bright centers) and dead (dark blue centers) cells by manually. This is important to obtain accurate cell counts and viability measurements.
The USB port allows you to transfer and save the cell count data and image to your computer for record keeping and printing purposes. Any other standard USB drive can be used for data transfer.
To turn the power on and o, press and hold for 3 seconds. The unlighted status indicates that the instrument is o; the red status light indicates that the instrument is on.
Slide port
LCD touch screen
Power button
USB port
Image adjustment knob
EVE™ PLUS, Automated Cell Counter 5
Product description
Rear view
The front view showing various parts of the ADAM-MC2 is shown below:
Control buttons Description
USB port The USB port allows you to transfer and save the cell count
data and image to your computer for record keeping and
printing purposes. Any standard USB drive can be inserted into
the USB port for data transfer.
Power inlet
Connect the counter to an electrical outlet using the supplied
power cord and the appropriate plug, based on the electrical
outlet conguration in your country.
USB Port
Power inlet
6
Installation
Install EVE™ PLUS
1. After unpacking the instrument, place the instrument on a at and dry sur­face.
2. Connect adaptor and power cord, then plug the power cord to EVETM PLUS.
3. Plug the power cord into the electrical outlet. Be sure to use only the power cord supplied with your instrument. Powering the instrument with an unapproved power cord may damage the instrument.
4. When you are ready to use, start the EVE™ PLUS by pressing and hold 3 seconds the Power button.
5. When the instrument is turned on, the startup screen is displayed. Here you can proceed immediately to cell counting, set up the instrument for cell types, or adjust the parameter for cell counting.
EVE™ PLUS, Automated Cell Counter 7
Installation
Activation code
To use the EVE™ PLUS for the rst time, you need the activation code, which is in the EVE™ slide box enclosed with the instrument. Entering the pro­vided code into the pop up window allows normal cell counting.
This code has a limit on the number of uses, and it is possible to use the counting function of 200 times per code.
If the limit is exceeded, this popup window will reappear. In this case, you must enter a new code from the new box.
* This code is intended to provide accurate calculation results using correct consumables.
8
Setting
Setting menu
Mode
Parameter
Press SETTING from the startup screen to display settings. The setting menu allows you to set up the following:
• MODE to operate the instrument for cell counting (choose single cell) or clumy cell counting (choose CLUMP CELL)
PARAMETER (see below and next page for details)
CALIBRATION to calibrate the instrument image background level (page 12)
UPDATE to install new rmware versions as they become available
DATE to set up date and time (page 14)
WIFI to connect to the internet to send results via mail
Select count mode for cell count­ing. Please choose counting mode (SINGLE CELL or CLUMP CELL).
1. Press PARAMETER from the setting screen to display parameters screen.
Parameter function allows you to change the image analysis algorithm for specic or mixed cell types, and the specic parameters must be deter­mined empirically.
EVE™ PLUS, Automated Cell Counter 9
Setting
Parameter
2. The parameters are described below:
[Sensitivity]
Sensitivity (refers to the contrast of the objects against the background). Adjusting the sensitivity higher makes instrument more sensitive to ob­jects; useful for cells that do not stain well with trypan blue while adjust­ing the sensitivity lower makes the instrument less sensitive and is useful if there is a lot of background.
[Minimum cell size]
Minimum cell size is used to determine the low range of cell size to include in the measurement. The algorithm rst identies all objects, and calcu­lates the average size (e.g., 15 µm). From the percent of average size set­ting, the algorithm calculates the smallest object size to include in the nal measurement (e.g., 70% of 15 is 10.5 µm; 15–10.5 = 4.5 µm; 4.5 µm would be the smallest particle included in the count). Adjusting the number up, increases inclusiveness thereby decreasing the lower cell size range (e.g., 50% of 15 is 7.5 µm; 15–7.5 = 7.5 µm).
Maximum cell size is used to determine the high range of cell size to in­clude in the measurement. The algorithm rst identies all objects, then calculates the average size (e.g., 15 µm). From the percent of average size setting, the algorithm calculates the largest object size to include in the nal measurement (e.g., 200%; 200% of 15 µm = 30 µm; 30 µm is the largest cell size included in the measurement.
Circularity is used to determine the objects to include in the measurement based on roundness. Increasing the value from 80% requires objects to be more round for inclusion in the measurement. Decreasing the value from 80% allows objects to be less round. Adjusting this may be useful if the cell type is not particularly circular or perhaps oddly shaped.
4Ch
After modifying any parameters, press APPLY button to make the changes.
To restore default parameters, press DEFAULT button.
3. Press NEW button to create a new protocol.
10
Setting
Parameter
4. Type protocol name and user name in the appropriate elds, and press SAVE button again.
5. Once a protocol is saved, it is available for use at any time. Press LOAD button.
6. The protocol appears in the protocol menu. Use the up and down arrows to nd your saved protocol. To use the protocol, select one of protocol and press APPLY button.
» Press DELETE button to delete protocol.
Press EDIT button to edit protocol.
Press CLOSE button to exit the screen.
EVE™ PLUS, Automated Cell Counter 11