ULTRAPEN™ PT5/PTBT5 SENSOR REPLACEMENT INSTRUCTIONS
n
Acc ura cy • Rel ia bil ity • Si mpl ic ity
NOTE: Please read these instructions thoroughly and carefully — Damage caused by improper installation or mishandling will void
the warranties of the instrument and the new sensor.
CAUTION: Moisture trapped in instrument will cause damage to the electronic circuitry. Careful adherence to the following
procedure will signicantly reduce the potential for failure due to moisture contamination or damaged connectors.
CAUTION: AVOID TOUCHING AND/OR CONTAMINATING THE ULTRAPEN ELECTRONICS OR SENSOR CONNECTOR/
CIRCUITRY AT ALL TIMES!
For clarity, this procedure is broken down into three sections. Follow each section carefully to avoid damage. Refer to images as
necessary. All images are of model PT5 for example.
• Always replace the sensor dry and ll with electrolyte after installation.
• Only remove/replace the sensor in a CLEAN and DRY environment.
I. Preparation of ULTRAPEN
1. Remove the protective/hydration cap from your ULTRAPEN by wiggling it side to side while gently pulling it off the
ULTRAPEN.
2. Remove the membrane cap assembly by turning it in a counter-clockwise direction.
3. Empty the electrolyte well of ALL electrolyte solution.
4. Rinse the sensor end and surrounding area with clean water (preferably DI, RO, or distilled) and dry thoroughly.
5. Ensure hands are clean and dry. Moisture will damage the electronics.
6. Unscrew the battery cap and remove the battery tray. (There is a pressure relief tube running from the battery
compartment to the sensor compartment, which allows pressure equalization in the sensor area.)
7. Grasp the sensor body and gently pull the sensor out of the ULTRAPEN.
8. Carefully clean off any residual salt or liquid around the opening of the ULTRAPEN’s sensor compartment using a
towel. Caution: Do NOT contaminate the inside of the ULTRAPEN.
®
Water Qu al ity Ins tr umentat io

II. Preparing the New Sensor
NOTE: The O-rings of the sensor are pre-greased for ease of insertion. Carefully remove any packaging from the shaft before
installation and handle the sensor by the protective/hydration cap to avoid migration of the grease to the membrane.
1. Holding the new sensor, remove the warning label from around the protective/hydration cap.
2. Remove the Connector Cap from the connector end of the sensor and inspect to ensure 3 O-rings are on the
sensor body.
III. Installing the New Sensor
1. Align the Large and Small alignment tabs on the sensor with the Large and Small grooves on the ULTRAPEN. (It
will only go in one way.) WARNING: Damage WILL occur to the connectors inside your ULTRAPEN if you attempt
to install the sensor incorrectly.
2. While maintaining alignment, gently apply pressure to insert the sensor into the ULTRAPEN. You may need to
apply a slight twisting motion to start each O-ring, ensuring all 3 O-rings on sensor slide evenly into ULTRAPEN.
CAUTION: A damaged O-ring will allow solution to leak into the instrument. Press sensor STRAIGHT into the
ULTRAPEN until it is securely in place.
3. Replace the battery tray by aligning the groove in the side of the enclosure with the groove in the battery tray, then
sliding the tray into the case until it is fully seated. The groove in the case corresponds to a guide bump inside the
enclosure that guides the tray back into the correct position. Screw the battery cap back on.
Your instrument is now ready for calibration. See Operation Manual for calibration instructions.
2450 Impala Drive
Carlsbad, CA 92010-7226 USA
Tel: +1-760-438-2021
Fax: +1-800-869-7668 / +1-760-931-9189
www.myronl.com
Printed in U.S.A.
Built On Trust. Founded in 1957, the Myron L® Company
is one of the world’s leading manufacturers of water quality
instruments. Because of our commitment to product
improvement, changes in design and specications are
possible. You have our assurance any changes will be guided
by our product philosophy: accuracy, reliability, and simplicity.
© Myron L® Company 2018
®
RPT5I 10-18