ULTRAPEN™ PT2 – Quick Start Guide
pH and Temperature Pen
PT2QSI 06-18
3-POINT CALIBRATION: Myron L
®
Company recommends performing a 3-point calibration of the PT2 using pH 7, 4, and 10 Buffers twice a month
or any time measurements are not as expected.
For maximum accuracy, fill 2 clean containers with each pH buffer. Arrange them in such a way that you can clearly remember which is the rinse
solution and which is the calibration buffer. If you don’t have enough buffer, you can use 1 container of each buffer for calibration and 1 container of
clean water for all rinsing. Always rinse the pH sensor between buffer solutions. Ensure the pH sensor is clean and free of debris.
1. Thoroughly rinse the pen by submerging the sensor in pH 7 buffer rinse solution and swirling it around.
2. Push and release the push button to turn the PT2 ON.
3. Push and hold the push button to go into menu mode.
4. Release the button when “CAL” displays. The display will indicate “CAL” and the LED will flash rapidly.!
5. While the LED flashes rapidly, dip the pen in pH 7 buffer calibration solution so that the sensor is completely submerged.
6. While the LED flashes slowly, the pH calibration point will display along with “CAL”. Swirl the pen around to remove bubbles, keeping the sensor
submerged.
7. If the pH 7 calibration is successful, the display will indicate “SAVEd”, then “PUSHCONT” will be displayed.
8. Push and release the push button to continue. The LED will begin flashing rapidly.
9. Repeat steps 5 through 8 with pH 4 and 10 buffer calibration solutions.
10. After the 3rd calibration point is successfully saved, the display will indicate “SAVEd” and power OFF.
11. Verify calibration by retesting the calibration solution.
• If “Error” is displayed, check to make sure you are using a proper pH buffer solution. If the solution is correct, clean the sensor. For sensor
cleaning instructions, download the full PT2 Operation Manual from the Myron L® Company website for troubleshooting instructions.
1 AND 2 POINT CALIBRATION: To perform a one or two point calibration, follow the above procedure, but let the PT2 time out after the last
point you wish to calibrate.
FACTORY CALIBRATION: Factory Calibration (FAC CAL) resets the PT2 to its factory settings. For instructions on how to perform a FAC CAL,
download the full PT2 Operation Manual from the Myron L® Company website.
MAINTENANCE
I. BATTERY REPLACEMENT
The PT2 display has an indicator that depicts the pen battery's charge level. When the charge level indicator only shows one bar, IMMEDIATELY
replace the battery with a new N type battery.
1. In a CLEAN DRY place unscrew the battery cap in a counter-clockwise motion.
2. Slide the cap and battery housing out of the pen.
3 Remove the depleted battery from its housing.
4 Insert a new battery into the battery housing oriented with the negative end
touching the spring.
5. Align the groove along the battery housing with the guide bump inside the pen
case and slide the battery housing in.
6. Screw the pen cap back on in a clockwise direction. Do not over tighten.
II. ROUTINE MAINTENANCE
1. ALWAYS rinse the pH sensor with clean water after each use.
2. ALWAYS replace the protective/hydration cap half filled with pH Sensor
Storage Solution to prevent the sensor from drying out after each use.
3. If the sensor becomes dirty, clean the sensor surface with an isopropyl soaked cotton swab. Then rinse thoroughly with clean water.
4. Do not drop, throw or otherwise strike the pen. This voids the warranty.
5. Do not store the pen in a location where the ambient temperatures exceed its specified Operating/Storage Temperature limits.
III. PH SENSOR REPLACEMENT (Follow the instructions that come with your Replacement Sensor.) Only remove or replace the pH sensor in a
CLEAN and DRY environment!
1. To remove the pH sensor: Remove the protective/hydration cap; make sure the PT2 (including the pH sensor) is clean and dry. Loosen the battery
tray (to allow pressure equalization).
2. Firmly grasp the pH sensor body and slowly pull the pH sensor out.
3. To install a new pH sensor: Line up the alignment tabs on the pH sensor with the alignment slots on the PT2 unit. Gently push the pH sensor into
position, then close the battery cap.