6. 2oz. bottle of Alkalinity Standard Solution Model ALK-1002OZ
7. 2oz. bottle of Hardness Standard Solution Model HARD-2002OZ
8. 2oz. bottle of pH 4 buffer Model PH42OZ
9. 2oz. bottle of pH 7 buffer Model PH72OZ
10. 2oz. bottle of pH 10 buffer Model PH102OZ
11. 2oz. bottle of pH/ORP Sensor Storage Solution Model SS2OZ
12. 2oz. bottle of Alkalinity Reagent A1 Model H2SO4-12OZ
13. 2oz. bottle of Hardness Reagent HB, Model HBUFF2OZ
14. 2oz. bottle of Hardness Reagent H4, Model EDTA-B-LC2OZ
15. 2oz. bottle of Hardness Reagent H5, Model EDTA-B-HC2OZ
16. 1oz. bottle of ORP ISA Sensor Conditioner Model ORPCOND1OZ
17. 2oz. bottle of KCl-7000 Model KCL-70002OZ
18. 2oz. bottle of 442-3000 Model 442-30002OZ
19. Titration T-plunger Model TPLUNGER
i
Page 4
ii
Page 5
I. INTRODUCTION
Thank you for selecting the feature-packed Ultrameter III™, one of the
Myron L® Company’s latest in an increasing line of instruments utilizing
advanced microprocessor-based circuitry and SMT manufacturing
processes. This circuitry makes the instrument extremely accurate,
reliable and very easy to use.
The Ultrameter III has been designed to include titration measurements
for Alkalinity, Hardness and an LSI Calculator for water balance analysis.
The Ultrameter III also features Myron L® Company's exclusive Free
Chlorine Equivalent (FC
TM
E
) function for making ORP-based free chlorine
measurements, as well as optional Bluetooth® wireless data transfer.
Other features include a clock with time and date, memory of up to 100
locations with time and date stamp, the ability of the user to adjust the
timeout “Auto OFF”, and enhanced performance. See Features and
Specications on pages 2-3.
The most exciting feature is data logging with the ability to wirelessly
download the memory or stored test data with its corresponding time, date
and unit name (requires -BD option or bluDockTM accessory). This feature
allows the user to create spreadsheets and graphs with ease, and quickly
and accurately manipulate data more effectively. The bluDock's Guardian
TM
2 TM
application will operate on Windows® 7 or later/macOS® 10.13 or later with
Bluetooth® capability. The data may be exported to spreadsheet formats
such as Microsoft Excel® (.xls and .xlsx) or universal formatting (.csv). The
user can also save data in a secure, encrypted format (.mlcx).
Please Note: Although the Myron L® Company has performed extensive
testing, we cannot guarantee compatibility of all applications and
formats. We suggest testing your application and format for compatibility
before relying on it.
For your convenience, a brief set of instructions is provided on the
bottom side of your Ultrameter III. A Quick Start Guide with abbreviated
instructions is also included with the instrument as a quick reference.
Special note ... Conductivity, resistivity, and TDS require mathematical
correction to 25°C values (ref. Temperature Compensation, pg. 56). On
the left of the Ultrameter III’s liquid crystal display is shown an indicator of
the salt solution characteristic used to model temperature compensation
of conductivity and its TDS conversion. The indicator may be KCl,
NaCl, 442™ or User. Selection affects the temperature correction of
conductivity and the calculation of TDS from compensated conductivity
(ref. Conductivity Conversion to Total Dissolved Solids (TDS), pg. 59).
The selection can affect the reported conductivity of hot or cold solutions
and will change the reported TDS of a solution. Using KCl for conductivity,
NaCl for resistivity, and 442™ (Natural Water characteristic) for TDS is
consistent with present industry practice for standardization. This is how
your instrument, as shipped from the factory, is set to operate. For use in
seawater desalination for example, both the conductivity and TDS may
easily be changed to NaCl.
1
Page 6
II. FEATURES and SPECIFICATIONS
A. Features
•ORP-based FCE Free Chlorine measurement; displays as ppm
concentration
•Alkalinity and Hardness Conductometric Titrations
•Langelier Saturation Index (LSI) Calculator
•Superior resolution 4 digit LCD displays full 9999 µS/ppm
•Cond/TDS Accuracy of ±1% of READING / ±0.1% at calibration point
•All electrodes are internal for maximum protection
•Improved 4-electrode sensor technology
•Waterproof to 1 meter/3 feet
•Autoranging conductivity/TDS/resistivity
•Prompts for easy pH calibration
•Factory calibrations stored in microprocessor
•3 conductivity/TDS solution conversions preprogrammed into
microprocessor
•User mode feature allows:
Programming your own Cond/TDS conversion factor
Programming your own temperature compensation factor
Disabling temperature compensation
•Real Time Clock with Time and Date
•Data Logging with TIME and DATE in memory
•Memory stores 100 readings
•User adjustable timeout “Auto OFF”
•Bluetooth® wireless download capability with optional bluDock™
accessory package
B. General Specications
Display 4 Digit LCD
Dimensions (LxWxH) 196 x 68 x 64 mm/
7.7 x 2.7 x 2.5 in.
Weight 352 g/12.4 oz.
Case Material VALOX*
Cond/Res/TDS Cell Material VALOX*
Cond/TDS Electrodes (4) 316 Stainless Steel
Cond/Res/TDS Cell Capacity 5 ml/0.17 oz.
pH/ORP Sensor Well Capacity 1,2 ml/0.04 oz.
Power 9V Alkaline Battery
Battery Life >100 Hours/5000 Readings
Operating/Storage Temperature 0-55°C/32-131°F
Protection Ratings IP67/NEMA 6 (waterproof to
1 meter/3 feet)
EMI/EMC Ratings EN61326-1: 2006 + Annex A: 2008
(hand-held devices)
(Conformité Européenne)
IEC 61000-4-2: 2001 and,
IEC 61000-4-3: 2002
* ™ SABIC Innovative Plastics IP BV
Additional information is available on our website:
CISPR 11: 2003
www.myronl.com
MADE IN USA
2
Page 7
C. Specication Chart
*
*
*
Auto
Temperature Compensation
0-71ºC
32-160ºF
0-71ºC
32-160ºF
0-71ºC
32-160ºF
0-71ºC
32-160ºF
0-71ºC
32-160ºF
ParametersRangesResolution
0.01(<100µS)
0.1(<1000µS)
1(<10mS)
0.01(<100mS)
0.1(<200mS)
0.01(<100ppm)
0.1(<1000ppm)
1(<10ppt)
0.01(<100ppt)
0.1(<200ppt)
0.01(<100KΩ)
0.1(<1000KΩ)
0.1(>1MΩ)
0.01ppm
0.1(<1000 ppm)
1(<1710 ppm)
0.1(<100 grains)
0.1ºC/ºF±0.1ºC
350≤ORPmV<725
)
725≤ORPmV<825
0-9999µS/cm
10-200mS/cm
in 5 autoranges
0-9999ppm
10-200ppt
in 5 autoranges
0.00-9.99ppm**
and 0.0≤pH<9.9
and 0.0≤pH<8.9
0-1710ppm
(0-100 grains)
0-71ºC
32-160ºF
Conductivity
TDS
Resistivity10KΩ-30MΩ
pH0-14pH0.01pH±0.01pH
ORP±999mV1mV±1mV
Free Chlorine
Equivalent
TM
E
(FC
Alkalinity Titration10-800ppm0.1ppm
Hardness Titration
LSI Calculator-10 to +100.01
Temperature
*EM Susceptibility: When Measuring 10.0 pH Reference Solution
±0.2 pH in the presence of RF elds ≥ 3 V/m.
±0.37 pH in the presence of RF elds at 300 MHz (±30MHz)
**If either ORP or pH is outside the specied limits, the instrument will display "-Or-".
Accu-
racy
±1% of
reading
±1% of
reading
±1% of
reading
±0.3ppm
<1.00ppm*
±0.2ppm
≥1.00ppm
Adjustable
Temperature Compensation
0-9.99%/ºC
0-9.99%/ºC
0-9.99%/ºC
Cond/TDS
Ratios
Prepro-
grammed
KCl, NaCl,
442™
Adjust-
able
Cond/
TDS
Ratio
Factor
0.20-
7.99
WARNING: These products can expose you to chemicals including
Di(2-ethylhexyl)phthalate (DEHP), which is known to the State of
California to cause cancer and birth defects or other reproductive harm.
For more information go to www.P65Warnings.ca.gov
D. Warranty/Service
The Myron L® Ultrameter III, excluding the pH/ORP sensor, has a Two
(2) Year Limited Warranty. The pH/ORP sensor has a Six (6) Month
Limited Warranty for materials and workmanship. If an instrument fails
to operate properly, see Troubleshooting Chart, pgs. 52-53. The battery,
pH/ORP sensor and cell extender are user-replaceable. For other
service, return the instrument prepaid to the Myron L® Company.
If, in the opinion of the factory, failure was due to materials or
workmanship, repair or replacement will be made without charge. A
reasonable service charge will be made for diagnosis or repairs due to
normal wear, abuse or tampering. This warranty is limited to the repair or
replacement of the Ultrameter III only. The Myron L® Company assumes
no other responsibility or liability.
Because of our commitment to product improvement, the substance and
style of this manual may change. When changes are made, the updated
manual is posted for download in PDF format from the Myron L® Website:
A. Operation
NOTE: The cell extender does not interfere with normal operation.
Using the instrument is simple:
• Individual or multiple parameter readings may be obtained by
lling individual sensors or entire cell cup area.
• Rinse the conductivity cell and/or pH/ORP sensor well with test
solution 3 times and rell. Temperature and/or measurement
extremes will require additional rinses for maximum accuracy.
• Press the desired measurement key to start measurement.
• For titrations, user intuitive display prompts guide you through
the addition of reagents and measurements.
• Pressing any parameter key again in measurements restarts
the 15 second “Auto off” timer. Pressing any parameter key
again during titrations will cancel the titration in progress.
• Note the value displayed and/or press the MS key to store the
reading (ref. Memory Storage, pg. 36). It’s that simple!
B. Characteristics of the Keys
• Though your Ultrameter III has a variety of sophisticated
options, it is designed to provide quick, easy, accurate
measurements by simply pressing a key.
• All functions are performed one key at a time.
• THERE IS NO “OFF” KEY. If the instrument remains inactive
for 15 seconds after you press any parameter key (user
adjustable up to 75 seconds), the instrument turns itself off.
The instrument turns itself off after 60 seconds of inactivity in
CAL mode, 3 minutes of inactivity in each titration screen and
60 seconds of inactivity in each LSI calculator value screen.
• Rarely is it necessary to press and hold a key (as in Procedure
to Select a Solution, pg. 23; Conductivity or TDS Calibration,
pg. 28; and ORP/FCE Mode Selection, pg. 20).
C. Operation of the Keys (See Instrument Illustration, pg. ii)
1. Measurement Keys in General
Any of the 8 measurement keys turns on the instrument in the mode
selected. The mode is shown at the bottom of the display for COND,
RES, TDS, ORP and pH. The ORP mode is indicated by the units
displayed, mV for ORP or ppm for free chlorine. Measurement units
appear at the right. Pressing any measurement key puts the unit in that
measurement mode even if you are in a calibration sequence and also
serves to cancel a change (ref. Leaving Calibration, pg. 27).
8
Page 13
2. COND, RES and TDS Keys
These 3 keys are used with solution in the Conductivity Cell.
Precautions:
• While lling cell cup, ensure no air bubbles cling on the cell wall.
• If the proper solution is not selected (KCl, NaCl, 442™ or
User), refer to Why Solution Selection is Available, pg. 23 and
Procedure to Select a Solution, pg. 23.
a. COND Key
Solution to be tested is introduced into the conductivity cell and a press
of displays conductivity with units on the right. On the left is
shown the solution type selected for conductivity.
b. RES Key
A press of
is shown solution type selected for resistivity (ref. Solution Selection,
pg. 23). The range of display of resistivity is limited to between 10
kilohms (KΩ) and 30 megohms (MΩ). A solution outside that range will
only show [- - - -] in the display.
c. TDS Key
A press of
This is a display of the concentration of ionized material calculated from
compensated conductivity using the characteristics of a known material.
On the left is shown solution type selected for TDS (ref. Solution
Selection, pg. 23).
displays resistivity with units on the right. On the left
displays Total Dissolved Solids with units on the right.
3. Alkalinity, Hardness and LSI Keys
The ALK and HARD keys enter the titration measurement functions. The
LSI key also allows you to access the LSI Calculator.
NOTE: All titration measurements require the installation of the cell
extender.
a. Alkalinity Key
A press of
b. Hardness Key
A press of
c. LSI Key
A press of
used to measure LSI by pulling the most recent alkalinity, hardness, pH
and temperature values or for hypothetical water balance adjustments.
enters the Alkalinity titration function.
enters the unit in the Hardness titration function.
enters the unit in LSI calculator mode, which can be
9
Page 14
4. pH and ORP/FC Keys
E
Measurements are made on solution held in the pH/ORP sensor well
(ref. pH and ORP/FCE, pg. 62). The protective cap is removed and
the sensor well is lled and rinsed with the sample enough times to
completely replace the storage solution.
After use, the pH/ORP sensor well must be relled with Myron L®
Storage Solution, and the protective cap reinstalled securely (ref.
Maintenance of the pH/ORP Sensor, pg. 11 and Cleaning Sensors, 2.
pH/ORP, pg. 50).
a. pH Key
A press of
displays pH readings. No units are displayed on the right.
b. ORP/Fr Chl Key
In ORP mode, a press of
displays Oxidation-Reduction Potential
/REDOX reading in millivolts; "mV" is displayed. When the FCE mode
is activated, a press of
reading in "ppm" alternating with the FCE predictive ORP reading in "mV".
displays the Free Chlorine Equivalent
5. CAL/MCLR Key
While measuring conductivity, TDS, or pH, a press of
allows
you to enter the calibration mode. Once in CAL mode, a press of this
key accepts the new value. If no more calibration options follow, the
instrument returns to measuring (ref. Leaving Calibration, pg. 27).
If is held down for about 3 seconds, CAL mode is not entered,
but “SEL” appears to allow Solution Selection (ref. pg. 23) with the UP or DOWN keys. As in calibration, the CAL key is now an “accept” key.
While measuring ORP or Free Chlorine, holding CAL down for about
3 seconds allows ORP/FCE mode selection (ref. pg. 20).
For titrations and LSI Calculator, the CAL key steps you through the
procedure and accepts values for nal calculations.
Once in CAL mode, a press of this key accepts the new value.
While reviewing stored records, the MCLR side of the key is active to
allow clearing records (ref. Clearing a Record/Memory Clear, pg. 37).
6. UP or DOWN Keys
While measuring in any parameter, the or
keys activate
the Memory Store and Memory Recall functions. A single press steps
the display and holding either key scrolls the value rapidly.
10
Page 15
While in calibration or calculator mode, the keys step or scroll the
displayed value up or down.
While in Memory Recall, the keys scroll the display up and down
through the stack of records (ref. Memory Recall, pg. 37).
IV. AFTER USING THE ULTRAMETER
A. Maintenance of the Conductivity Cell
Rinse out the cell cup with clean water. Do not scrub the cell. For
oily lms, squirt in a foaming non-abrasive cleaner and rinse (ref.
Cleaning Sensors, pg. 50). Even if a very active chemical discolors
the electrodes, this does not affect the accuracy; leave it alone.
B. Maintenance of the pH/ORP Sensor
The sensor well must be kept wet. Before replacing the rubber cap,
rinse and ll the sensor well with Myron L® pH Sensor Storage Solution.
Although not ideal, if Myron L® Storage Solution is not available, you
can temporarily use a strong KCl solution, a pH 4 Buffer Solution, or a
saturated solution of table salt and tap water until pH Storage Solution is
obtained. NEVER use distilled water.
If all ORP solutions are reporting approximately the same value,
clean ONLY the platinum ORP electrode with an MLC ORP
conditioner-soaked cotton swab, being careful not to touch the
swab to the glass bulb of the pH sensor.
V. SPECIFIC RECOMMENDED MEASURING PROCEDURES
A. Parameter Methods
For Conductivity and Total Dissolved Solids measurements, if the
proper solution is not selected (KCl, NaCl, 442TM or User), see Solution
Selection, pg. 23.
III
NOTE: After sampling high concentration solutions or temperature
extremes, more rinsing may be required. When sampling low
conductivity solutions, be sure the pH cap is well seated so that no
solution washes into the conductivity cell from around the pH cap.
B. Titration Methods
For Alkalinity, Hardness titrations, assemble the required materials
before you begin: a pipette with 100µL tips, the required reagents, the
cell extender, and the T-plunger.
11
Page 16
NOTE: The accuracy of titration measurements
Plunger Button
is affected by your technique. Be careful when
removing the cap of the cell extender to add
reagents. Flicking the cap or popping it off can
Plunger Rod
cause solution to spill out of the cell extender.
Always use a new pipette tip when changing
solutions to avoid contamination.
CE
1. Pipette Instructions
100µl
Always use a clean tip when changing
solutions! To install a tip, press the wide end of
Shaft Coupling
100 µL
the tip on the end of the pipette until it is secure.
Avoid contaminating the pipette tips. To remove
a tip, simply pull it off the end of the pipette.
Disposable Tip
Shaft Coupling
100µl
100 µL
Pipette Assembly
CE
Shaft Bottom
Disposable Tip
Fully Assembled Pipette
NOTE: Do NOT twist the cap off as this may unscrew the shaft bottom
from the pipette.
The pipette has 3 positions: REST, FIRST STOP, and SECOND STOP.
Practice pushing in and letting up on the pipette to feel the different
positions.
Rest Position: The
plunger is all the way
U P.
First Stop: The plunger
is depressed until
resistance is felt.
a. To DRAW a
sample: Grasp the pipette by the shaft
with your hand. Rest your thumb on top
of the plunger button. Use your thumb
to depress the top of the pipette to the
FIRST STOP. Insert the tip end down into
sample until only the tip is submerged,
about 2-3mm / 1/8". Slowly release the
top and let it return to the REST position
being careful to keep the tip submerged.
b. To
DISPENSE a sample: With the pipette
in the REST position, place the tip end
12
Second Stop: The
plunger is depressed all
the way DOWN.
Do not allow the pipette tip to touch
the cell wall or the sample.
Page 17
over the conductivity cell being careful
not to touch the tip to the existing
solution. Depress the top of the pipette
to the SECOND STOP (all the way
down) being careful to keep the tip over
the cell. Release the plunger button and
let it return to the REST position.
2. T-Plunger Instructions
To PLUNGE the cell:
Clean and dry the T-plunger. With the
cell extender installed and solution in
the cell, insert the tip of the T-plunger
in the cell extender until the arms of the
T-plunger are ush against the rim of the
cell extender. Solution will overow the
cell.
Keeping the arms of the T-plunger ush
against the rim of the cell extender, rotate
the T-plunger from side to side, allowing
the arms of the T-plunger to slide along
the rim of the cell extender.
Remove the T-plunger from the cell by
lifting it straight up out of the cell, being
careful to keep it centered over the mouth
of the cell extender. When the T-plunger
is completely out of solution, tap the
T-plunger on the inside rim of the cell
extender to dispel any clinging solution
back into the cell.
Insert the T-plunger all the way
down into the cell extender then
rotate it from side to side.
Hold the T-plunger over the cell
then tap rmly so that solution is
dispelled back into the cell cup.
Remove the T-plunger completely from
the cell area.
3. Mixing Solution in Cell
a. Agitate
When the display says to agitate (“AGit”):
Grasp the instrument on both sides of
the keypad (keypad facing up) with your
hands so you don’t accidentally drop or
release it. Agitate the solution in the cell
by swinging the cell cup area of the 9P up
“AGit”: When “AGit” displays, swing
the cell end of the 9P up and down
ve or more times.
and down ve or more times.
b. Hold
When the display says to hold (“HOLd”):
Hold the instrument steady in the
horizontal position while the instrument
records a measurement.
“HOLd”: When “HOLd” displays, hold
the 9P still with bottom of case parallel
to the ground.
13
Page 18
C. Measuring Conductivity & Total Dissolved Solids (TDS)
1. Rinse cell cup 3 times with sample to be measured. (This
conditions the temperature compensation network and prepares
the cell.)
2. Rell cell cup with sample.
3. Press or
4. Take reading. A display of [- - - -] indicates an overrange
condition.
D. Measuring Resistivity
Resistivity is for low conductivity solutions. In a cell cup the value may
drift from trace contaminants or absorption from atmospheric gasses, so
measuring a owing sample is recommended.
1. Ensure pH protective cap is secure to avoid contamination.
2. Hold instrument at 30° angle (cup sloping downward).
3. Let sample ow continuously into conductivity cell with no aeration.
4. Press
NOTE: If reading is lower than 10 kilohms, display will be dashes:
[ - - - - ]. Measure solution Conductivity instead.
E. Measuring Alkalinity
Prepare the materials required for titration: Reagent: A1, cell extender,
T-plunger, and 100µL pipette with a clean tip installed.
NOTE: If you do not complete each titration point within the 3-minute
timeout period, the unit exits the Alkalinity function. Start the titration
process over.
key; use best reading.
.
If you make a gross error in your titration method, e.g., you forget to add
sample to the cell, “Err” will display. Start the titration over.
1. Ensure pH protective cap is secure to avoid contamination.
2. Install the cell extender by pushing the base of the cell extender
into the conductivity cell until it is fully seated.
3. Rinse the conductivity cell cup and cell extender 3 times with
solution to be measured.
4.Rell cell cup and cell extender with sample solution.
5. Insert T-plunger into open cell extender until the “T” rests ush
on the lip of the cell extender. This will cause uid beyond what
is required for titration to overow the cell extender.
14
Page 19
6. Rotate the T-plunger from side to side, allowing the arms of the
T-plunger to slide along the rim of the cell extender.
7. Carefully remove the T-plunger so as not to ick or spill remaining
sample. Tap the T-plunger on the inside rim of the cell extender
to remove sample that may be stuck to the T-plunger.
8. Press . “AL” alternating with “tdS” briey displays along
with the “PPM” value of the sample. “PrES CAL” will display
when the reading stabilizes.
9. Press CAL to accept and advance to the rst titration prompt.
“Add” alternating with “A1” displays.
10. Using the pipette, add 100µL of Reagent: A1, close the cell
extender cap securely, Press CAL to continue.
11. “AGit t1” ashes on the screen. Agitate to mix the sample. When
agitation is done, tap the cell extender cap to dispel any solution
clinging to the cap back into the cell. Press CAL to continue.
(The number after "t" indicates which titration point you are on.
The number "1" here indicates this is the rst titration point.
Subsequent titration points are sequenced numerically: 2, 3, 4,
etc.)
12. “HOLd” displays. Keep holding the meter steady until “Add”
alternating with “A1” displays. Open the cell extender cap carefully
to avoid spilling, and use the pipette to add 100µL of Reagent A1.
Close the cell extender cap securely. Press CAL to continue.
13. Repeat Steps 11 and 12 until the meter reports the alkalinity
result.
14. Note the value for your records or press MS to store the value.
The unit will automatically power off.
15
Page 20
F. Measuring Hardness
1. Hardness Unit Selection
The 9P offers the ability to set the hardness unit preference to either
“PPM” CaCO3 or grains of hardness. To change hardness units, press
HARD, then press and hold CAL down until“HArd”and“SEL”are
displayed. Press UP or DOWN to toggle between “PPM” and grains (no
units are displayed for grains). Press CAL to accept. This also sets the
hardness unit preference for the LSI Calculator simultaneously.
3. Hardness Titration Procedure
Prepare the materials required for titration: Reagent: HB, Reagent: H4,
or Reagent: H5, cell extender, T-plunger, 100µL pipette with a clean tip
installed, and one extra clean tip.
NOTE: If you do not complete each titration point within the 3-minute
timeout period, the unit exits the Hardness function. Start the titration
process over.
1. Ensure pH protective cap is secure to avoid contamination.
2. Install the cell extender by pushing the base of the cell extender
into the conductivity cell until it is fully seated.
3. Rinse the conductivity cell cup and cell extender 3 times with
solution to be measured.
4. Rell cell cup and cell extender with sample solution.
5. Insert T-plunger into open cell extender until the “T” rests ush
on the lip of the cell extender. This will cause uid beyond what
is required for titration to overow the cell extender.
6. Rotate the T-plunger from side to side, allowing the arms of the
T-plunger to slide along the rim of the cell extender.
16
Page 21
7. Carefully remove the T-plunger so as not to ick or spill remaining
sample. Tap the T-plunger on the inside rim of the cell extender
to remove sample that may be stuck to the T-plunger.
8. Press . “CA” alternating with “tdS” displays along with
“PPM” sample value. “PrES CAL” displays when the reading stabilizes.
9. Press CAL. "HC" or "LC" will display; use the UP or DOWN
keys to switch to HC for the higher range (0-1710 ppm) or
LC for the lower range (0-200ppm). Press CAL to accept and
continue.
10. “Add” alternating with “buFF” displays. Using the pipette, add
100µL of Reagent HB to the sample, close the cell extender cap
securely. Change the pipette tip for adding EDTA in following
steps. Press CAL to continue.
11. “AGit” ashes on the screen. Agitate to mix the sample. When
agitation is done, tap the cell extender cap to dispel any solution
clinging to the cap back into the cell. Press CAL to continue.
12. “HOLd” displays. Keep holding the meter steady until “Add”
alternating with “EdtA” displays. Open the cell extender cap
carefully to avoid spilling, and use the pipette to add 100µL of
EDTA-B-HC or EDTA-B-LC, as indicated on the display. Close
the cell extender cap securely. Press CAL to continue.
13. “AGit t1” ashes on the screen. Agitate to mix the sample. When
agitation is done, tap the cell extender cap to dispel any solution
clinging to the cap back into the cell. Press CAL to continue.
(The number after “t” indicates which titration point you are on.
The number “1” here indicates this is the rst titration point.
Subsequent titration points are sequenced numerically: 2, 3, 4,
etc.)
14. Repeat Steps 12 and 13 until the meter reports the hardness result.
15. Note the hardness value for your records or press MS to store
the value. The unit will automatically power off.
NOTE: If there is a gross error in your titration method, e.g., you forgot
to add sample to the cell, “rEAd Err” will alternate with “rEPt tit”.
Repeat the titration from Step 1 to obtain a reading. If the reading is
over range, the display indicates “Or”. If the reading is under range, the
display indicates “0.00”.
17
Page 22
G. LSI Calculator Functions
The Ultrameter III features an LSI Calculator that can be used to perform
actual LSI or for hypothetical water balance calculations.
The LSI Calculator computes a Langelier Saturation Index value using
measured, default, or user-adjusted values for alkalinity, hardness, pH
and temperature.
If you have stored alkalinity, hardness or pH and temperature values,
the calculator will automatically display the most recent stored value. All
other values will display as default. You can adjust any or all of the values
displayed to determine the effect of the change(s) on the LSI value.
1. Measuring LSI
To compute the actual saturation index of a solution:
Before you enter the calculator function, you must measure and store
values for all water balance variables used by the calculator to compute
saturation index.
1. Perform an alkalinity titration of the sample solution (ref.
Measuring Alkalinity, pg. 14). Press MS to store the reading in
memory.
2. Perform a hardness titration of the sample solution (ref.
Measuring Hardness, pg. 16). Press MS to store the reading in
memory.
3. Measure the pH of the sample solution (ref. Measuring pH, pg.
19) Press MS to store the reading in memory.
4. Press
5. The last stored alkalinity value is displayed.
6. Press CAL to accept value and advance to the hardness value
screen. The last stored hardness value is displayed.
7. Press CAL to accept value and advance to the pH value screen.
The last stored pH value is displayed.
8. Press CAL to accept value and advance to the temperature
value screen. The last stored temperature value (taken from
the last stored hardness or alkalinity titration) is displayed.
9. Press CAL to accept and calculate LSI. The saturation index
value will display. Press MS to store the reading. The unit will
automatically power off after the period of inactivity dened in
the AUTO OFF setting (ref. AUTO OFF, pg. 43).
18
.
Page 23
If you want to modify any of the input values and recalculate LSI based
on those changes, press CAL again and repeat Steps 5-9 using the UP and DOWN keys to change the values. If you want to change hardness
units (ppm/grains), you must do so in the hardness parameter. See
Hardness Unit Selection, pg. 16.
2. Hypothetical LSI Calculations
To compute saturation index using hypothetical alkalinity, hardness, pH
or temperature values:
1. Press
2. Either the last stored value or the default value of “120” is
displayed.
3. Press the UP or DOWN keys to adjust the alkalinity value or
leave as displayed.
4. Press CAL to accept and advance to the hardness value
screen. Either the last stored value or the default value of “166”
is displayed.
5. Press the UP or DOWN keys to adjust the hardness value or
leave as displayed.
6. Press CAL to accept hardness value and advance to the pH
value screen. Either the last stored value or the default value of
“7.20” is displayed.
7. Press UP or DOWN to adjust the pH value or leave as displayed.
8. Press CAL to accept pH value and advance to the temperature
value screen. Either the last stored value (taken from the last
stored hardness or alkalinity titration) or the default value of
“25 ºC” is displayed.
9. Press UP or DOWN to adjust the temperature value or leave as
displayed.
10. Press CAL to accept and calculate LSI. A saturation index value
will display.
If you want to modify any of the input values and recalculate LSI based
on those changes, press CAL again and repeat Steps 3-10. If you want
to change hardness units, you must do so in the hardness parameter.
See Hardness Unit Selection, pg. 16.
.
H. Measuring pH
1. Remove protective cap by rotating while grasping and pulling up.
2. Rinse pH/ORP sensor well and conductivity cell cup 3 times
with sample to be measured. Shake out each sample to remove
any residual liquid.
19
Page 24
3. Rell both sensor well and cell cup with sample.
4. Press
5. Note value displayed.
.
6. IMPORTANT: The sensor well must be kept wet. Before
replacing the rubber cap, rinse and ll the sensor well with
Myron L® pH Sensor Storage Solution.
I. Measuring ORP
The Ultrameter III features the ability to measure the activity of
oxidizing or reducing chemicals in solution as ORP mV. The instrument
also includes an innovative Free Chlorine Equivalent (FCE) feature
(Measuring Free Chlorine Using FCE, pg. 21) that uses ORP and pH to
measure free available chlorine (FAC) concentration in ppm. ORP mV
and ppm of free available chlorine (FAC) are the two most commonly
used sanitizer units of measure in water quality management.
1. ORP / FCE Mode Selection
The Ultrameter III allows the user to choose between measuring
oxidizing sanitizers using either ORP mV or as parts per million (ppm)
of equivalent free chlorine. Use ORP to directly measure the oxidizing
power of all sanitizers like ozone, bromine, peracetic acid or chlorine. Use
FCE to measure the strength of oxidizing sanitizers as ppm of equivalent
free chlorine. To select between ORP and Free Chlorine Equivalent modes:
1. Press
.
2. Press and hold
for approximately 3 seconds.
The current preference for ORP units of measure is displayed. Factory
setting for this preference is ORP mV.
3. Press the or
keys to toggle between mV (standard
ORP mode) and FCE ppm. The setting chosen is displayed.
PPM
20
Page 25
4. Press any parameter key to exit ORP unit preference selection
or let the unit time out. ORP unit preference will be saved.
2. Measuring ORP
1. Ensure the 9P is in ORP mode (ref. ORP/FCE Mode Selection,
pg. 20).
2. Remove protective cap by rotating while grasping and pulling up.
3. Rinse sensor well and cell cup 3 times with sample to be
measured. Shake out each sample to remove any residual liquid.
4. Rell both sensor well and cell cup with sample.
5. Press .
6. Take reading.
7. Press MS to store reading in memory, if desired.
IMPORTANT: The sensor well must be kept wet. Before
replacing the rubber cap, rinse and ll the sensor well with
Myron L® pH Sensor Storage Solution.
J. Measuring Free Chlorine Using FC
E
The FCE function can be used to measure owing solution, bodies of
water and discrete samples. The measurement techniques are unique
to each type of sample. For best results, use the FCE Flow Method or
the FCE Immersion Method. For discrete samples, follow the FCE Flow
method, but instead of a running stream, rinse the conductivity cell cup
and pH sensor well three times then ll with the sample to be tested.
Read through Section 4: FCE Best Practices before you begin.
1. Prepare for FCE Measurement
1. For ease of measurement, set the instrument’s Auto oFF feature
to 75 sec (ref. AUTO OFF, pg. 43).
2. Ensure the FCE mode has been activated (ref. ORP/FCE Mode
Selection, pg. 20).
3. Remove protective cap from the pH/ORP sensor by rotating
while grasping and pulling up.
2. FCE Flow Method
1. Empty the pH/ORP sensor well of all storage solution.
2. Hold the 9P at a 30º angle (cup sloping downward).
3. Thoroughly ush the pH sensor well and conductivity cell cup
with a steady stream of the solution you intend to measure by
allowing the solution to ow into and out of the sensor well and
cell cup for at least 10 seconds.
21
Page 26
4. Let sample ow continuously into conductivity cell cup with no
aeration.
5. Allow both the pH sensor well and conductivity cell cup to remain
lled with sample.
6. Press
predicted nal ORP value and a free chlorine equivalent
concentration in ppm. Both readings will change rapidly at rst.
7. Wait for the readings to stabilize. When the mV and ppm values
are unchanging for 5 consecutive readings, the FCE reading
has reached a stable level. This may take 1 to 2 minutes.
NOTE: If the reading takes more than 1 minute to stabilize,
press the
from disturbing the measurement process. Annunciators will
alert you when either the pH or ORP of the nal FCE are Out of
Range (“-Or-”).
8. Press MS to store reading in memory if desired.
3. FCE Immersion Method
NOTE: Use this method for pools, spas and other large standing bodies
of water.
1. Hold instrument beneath the surface of the water to avoid
surface effects on the water’s chemistry.
2. Swirl the instrument around for at least 10 seconds to thoroughly
rinse the conductivity cell cup and pH sensor well.
. The instrument will begin alternating between a
after 1 minute to prevent the Auto oFF feature
3. Continue holding the instrument under the surface while taking
the reading.
4. Press .
5. The instrument will begin alternating between a predicted nal
ORP value and a free chlorine equivalent concentration in ppm.
Both readings will change rapidly at rst.
6. Wait for the readings to stabilize. When the mV and ppm values
are unchanging for 5 consecutive readings, the FCE reading
has reached a stable level. This may take 1 to 2 minutes.
NOTE: If the reading takes longer than 1 minute to stabilize,
press after 1 minute to prevent the Auto-OFF feature
from
22
Page 27
disturbing the measurement process. Annunciators will alert you
when either the pH or ORP of the nal FCE are Out of Range
(“-Or-”).
7. Press MS to store reading in memory if desired.
4. FCE Best Practices
For best results it is recommended that you:
1. Take at least 3 consecutive FCE measurements and record the
readings.
2. Discard measurements that are signicantly different from the
others.
3. Calculate the average of the remaining measurements. Use
this value.
IMPORTANT: The sensor well must be kept wet. Before
replacing the rubber cap, rinse and ll the sensor well with
Myron L® pH Sensor Storage Solution.
VI. SOLUTION SELECTION
A. Why Solution Selection is Available
Conductivity, resistivity, and TDS require temperature correction to
25°C values (ref. Standardized to 25°C, pg. 56). Selection determines
the temperature correction of conductivity and calculation of TDS from
compensated conductivity (ref. Cond. Conversion to TDS, pg. 59).
B. The 4 Solution Types
On the left side of the display is the salt solution characteristic used
to model temperature compensation of conductivity and its TDS
conversion. Generally, using KCl for conductivity, NaCl for resistivity, and
442TM (Natural Water characteristic) for TDS will reect present industry
practice for standardization. This is how your instrument is shipped from
the factory (ref. Solution Characteristics, pg. 59).
The User selection allows a custom value to be entered for the
temperature compensation of conductivity and also the conversion ratio
if measuring TDS.
C. Calibration of Each Solution Type
There is a separate calibration for each of the 4 solution types. Note that
calibration of a 442TM solution does not affect the calibration of a NaCl
solution. For example: Calibration (ref. Conductivity or TDS Calibration,
pg. 28) is performed separately for each type of solution one wishes to
measure (ref. Conductivity/TDS Standard Solutions, pg. 54).
D. Procedure to Select a Solution
NOTE: Check display to see if solution displayed (KCl, NaCl, 442TM or
User) is already the type desired. If not:
23
Page 28
1. Press
which you wish to change the solution type.
,
or to select the parameter on
2. Press and hold
appear (see Figure 1). For demonstration purposes, all 4
solution types are shown simultaneously.
3. Use the
(ref. Solution Characteristics, pg. 59). The selected solution
type will be displayed: KCl, NaCl, 442TM or User.
4. Press to accept new solution type.
E. Application of User Solution Type
1. User Programmable Temperature Compensation
This feature allows you to change your Ultrameter III’s temperature
compensation factor to another factor between 0-9.99%/°C (ref.
Temperature Compensation, pg. 56). This feature does not apply to pH
or ORP.
or key to select type of solution desired
key for 3 seconds to make “SEL”
(Tempco)
a. As in Procedure to Select a Solution, above, select User mode.
b. With User mode now selected, press
adjust a temperature compensation from .00%/°C to 9.99%/°C,
by pressing or . See example in Figure 2, pg. 25.
24
. You may now
Page 29
c. Press twice to skip calibration adjustment and accept
the new tempco (3 times if in TDS mode). You are now ready to
measure samples with your new temperature compensation factor.
2. Disabling Temperature Compensation
a. Select User mode (ref. Procedure to Select a Solution, pg. 23).
b. With “User” selected, press . If the display does not
show .00%/°C, hold long enough to bring the tempco to
.00%/°C (see Figure 3).
c. Press
compensation is now disabled (=0) for measurements in User
mode.
3. User Programmable Conductivity to TDS Ratio
This feature allows you to select a custom conductivity to TDS conversion
ratio within the range of 0.20-7.99 for User mode measurements.
To determine the conversion ratio for a custom solution of known TDS
ppm value, measure the solution conductivity at 25ºC with the Ultrameter III and divide the ppm value by the µS value. For example, a solution of
known 75 ppm TDS and measured 100 µS conductivity at 25ºC would
have a conversion ratio of 75/100 or 0.75. Enter the new conversion
ratio as follows:
twice (3 times if in TDS mode). Temperature
25
Page 30
a. While in User mode, press .
b. Press twice (to skip over tempco adjustment) and
“RATIO” will appear (see Figure 4).
c. Adjust with
displayed.
d. Press twice (to skip over calibration adjustment) to
accept new conversion ratio. You are now ready to measure
samples with the new conductivity/TDS ratio.
VII. CALIBRATION
A. Calibration Intervals
Generally, calibration is recommended about once per month with
Conductivity or TDS solutions. Calibration with pH solutions should
be checked twice a month. Calibration of ORP is not necessary (ref.
CALIBRATION INTERVALS TABLE under Calibration Steps, pg. 26).
B. Rules for Calibration of the Ultrameter
1. Calibration Steps
The following table lists measurement functions and their corresponding
calibration requirements:
or
key until new conversion ratio is
III
26
Page 31
CALIBRATION INTERVALS TABLE:
Function
KCl, NaCl or 442
TM
User
CondGain OnlyTempco, then Gain
ResDone in ConductivityDone in Conductivity
TDSGain OnlyTempco, Ratio, then Gain
pH7, acid and/or base
ORPZero set with pH 7
automatically
AlkalinityGain Only
HardnessGain Only
a. Starting Calibration
For Alkalinity and Hardness, calibration is begun by pressing
when the titration is complete and a value displays.
For Conductivity, TDS or pH, calibration is begun by pressing
while measuring Conductivity, TDS or pH. Measuring continues, but
the “CAL” icon is on, indicating calibration is now changeable. The
reading is changed with the
known value.
and
keys to match the
The calibration for each of the 4 solution types may be performed in
either conductivity or TDS mode.
Once in CAL mode, the
new calibration.
For pH, at each point, pressing
key becomes an “ACCEPT” key for the
accepts the new calibration
value and steps you to the next adjustment.
To bypass a calibration, simply press to accept the present
value as-is.
b. Leaving Calibration
Pressing accepts calibration and exits CAL mode if there are no
more steps. Calibration is complete when the “CAL” icon goes out.
Pressing any measurement key during calibration cancels changes
not yet accepted and exits calibration mode. Leaving pH after the 2nd
buffer results in the same gain being entered in place of the 3rd buffer.
27
Page 32
2. Calibration Limits
There are calibration limits for all functions for which calibration is
required. The 9P will not allow you to calibrate beyond its calibration
limits. Attempts to calibrate too far up or down from the ideal “FAC” value
stored in the unit by the factory will cause the displayed value to be
replaced with “FAC”. If you accept it (press the CAL key), you will have
the original default factory calibration for this measurement. The need to
calibrate so far out that “FAC” appears indicates a procedural problem,
incorrect standard solution, a very dirty cell cup or an aging pH/ORP
sensor (ref. Troubleshooting Chart, pg. 52).
C. Calibration Procedures
1. Conductivity or TDS Calibration
a. Rinse conductivity cell 3 times with proper standard (KCl, NaCl,
or 442TM) (ref. Cond/TDS Standard Solutions, pg. 54). For user
calibration see User Calibration Conductivity/TDS below.
b. Rell conductivity cell with same standard. KCl-7000 shown.
c. Press
appear on the display (see Figure 5).
d. Press
standard’s value (7032 > 7000) or hold a key down to scroll
rapidly through the reading.
e. Press
sequence for this particular solution type. If another solution
type is also to be measured, change solution type now and
repeat this procedure.
2. User Calibration Conductivity/TDS
Instrument must be in User mode, see Solution Selection, pg. 23.
a. Rinse conductivity cell three times with YOUR standard.
or , then press
or to step the displayed value toward the
once to conrm new value and end the calibration
; “CAL” icon will
b. Rell conductivity cell with same standard.
28
Page 33
c. Press or , then press twice in COND/
three times in TDS. The “CAL” icon will appear on the display.
d. Press
standard’s value or hold a key down to scroll rapidly through
the reading.
e. Press once to conrm new value and end the calibration
sequence for this particular solution type.
3. Resistivity Calibration
Resistivity is the reciprocal of Conductivity. To calibrate resistivity,
calibrate conductivity for the solution type you wish to measure (ref.
Conductivity or TDS Calibration, pg. 28).
4. Reloading Factory Calibration (Cond or TDS)
If calibration is suspect or known to be incorrect, and no standard solution
is available, the calibration value can be replaced with the original factory
value for that solution. This “FAC” value is the same for all Ultrameter IIIs,
and returns you to a known state without solution in the cell. The “FAC”
internal electronics calibration (which bypasses the electrodes and cell)
is not intended to replace calibration with conductivity/TDS standard
solutions. If another solution type requires resetting, change solution
type and repeat this procedure.
a. Press
b. Press
twice if in Conductivity, and three times if in TDS to skip over
tempco and ratio adjustments.)
or to step the displayed value toward the
or .
. (If in “User” solution mode — press CAL key
c. Press key until “FAC” appears and release.
d. Press to accept the factory calibration setting.
5. Alkalinity Calibration
Prepare the materials required for titration: Alkalinity Standard Solution
100PPM, Reagent: A1, cell extender, T-plunger, and 100µL pipette with
a clean tip installed.
NOTE: If you do not complete each titration point within the 3-minute
timeout period, the unit exits the Alkalinity function. Start the titration
process over.
29
Page 34
If you make a gross error in your titration method, e.g., you forget to add
sample to the cell, “Err” will display. Start the titration over.
1. Ensure pH protective cap is secure to avoid contamination.
2. Install the cell extender by pushing the base of the cell extender
into the conductivity cell until it is fully seated.
3. Rinse the conductivity cell cup and cell extender 3 times with
Alkalinity Standard 100PPM.
4. Rell cell cup and cell extender with Alkalinity Standard 100PPM.
5. Insert T-plunger into open cell extender until the “T” rests ush
on the lip of the cell extender. This will cause uid beyond what
is required for titration to overow the cell extender.
6. Rotate the T-plunger from side to side, allowing the arms of the
T-plunger to slide along the rim of the cell extender.
7. Carefully remove the T-plunger so as not to ick or spill remaining
sample. Tap the T-plunger on the inside rim of the cell extender
to remove sample that may be stuck to the T-plunger.
8. Press . “AL” alternating with “tdS” briey displays along
with the “PPM” value of the sample. “PrES CAL” will display
when the reading stabilizes.
9. Press CAL to accept and advance to the rst titration prompt.
“Add” alternating with “A1” displays.
10. Using the pipette, add 100µL of Reagent: A1, close the cell
extender cap securely, Press CAL to continue.
11. “AGit t1” ashes on the screen. Agitate to mix the sample. When
agitation is done, tap the cell extender cap to dispel any solution
clinging to the cap back into the cell. Press CAL to continue.
30
Page 35
(The number after "t" indicates which titration point you are
on. The number "1" here indicates this is the rst titration point.
Subsequent titration points are sequenced numerically: 2, 3, 4,
etc.).
12. “HOLd” displays. Keep holding the meter steady until “Add”
alternating with “A1” displays. Open the cell extender cap
carefully to avoid spilling, and use the pipette to add 100µL of
Reagent A1. Close the cell extender cap securely. Press CAL to
continue.
13. Repeat Steps 11 and 12 until the meter reports the alkalinity
result.
14. Press CAL. The “CAL” icon will display.
15. Adjust the displayed value to "100" using the UP or DOWN key.
NOTE: If you scroll beyond ±10% of the reading "FAC" will display. If
you press the CAL key while "FAC" is displayed, the instrument will
return Alkalinity to factory calibration (erasing any previous wet calibration adjustments).
16. Press CAL to accept. The unit will automatically power off.
6. Hardness Calibration
Prepare the materials required for titration: Hardness Standard Solution
200PPM, Reagent: HB, Reagent: H4, or Reagent: H5, cell extender,
T-plunger, 100µL pipette with a clean tip installed, and one extra clean
tip.
NOTE: If you do not complete each titration point within the 3-minute
timeout period, the unit exits the Hardness function. Start the titration
process over.
1. Ensure pH protective cap is secure to avoid contamination.
2. Install the cell extender by pushing the base of the cell extender
into the conductivity cell until it is fully seated.
3. Rinse the conductivity cell cup and cell extender 3 times with
Hardness Standard 200PPM.
4. Rell cell cup and cell extender with Hardness Standard 200PPM.
5. Insert T-plunger into open cell extender until the “T” rests ush
on the lip of the cell extender. This will cause uid beyond what
is required for titration to overow the cell extender.
31
Page 36
6. Rotate the T-plunger from side to side, allowing the arms of the
T-plunger to slide along the rim of the cell extender.
7. Carefully remove the T-plunger so as not to ick or spill remaining
sample. Tap the T-plunger on the inside rim of the cell extender
to remove sample that may be stuck to the T-plunger.
8. Press . “CA” alternating with “tdS” displays along with
“PPM” sample value. “PrES CAL” displays when the reading stabilizes.
9. Press CAL. "HC" or "LC" will display; use the UP or DOWN keys
to switch to HC for the higher range (0-1710 ppm) or LC for the
lower range (0-200ppm). Press CAL to accept and continue.
10. “Add” alternating with “buFF” displays. Using the pipette, add
100µL of Reagent: HB to the sample, close the cell extender cap
securely. Change the pipette tip for adding EDTA in following steps.
Press CAL to continue.
11. “AGit” ashes on the screen. Agitate to mix the sample. When
agitation is done, tap the cell extender cap to dispel any solution
clinging to the cap back into the cell. Press CAL to continue.
12. “HOLd” displays. Keep holding the meter steady until “Add”
alternating with “EdtA” displays. Open the cell extender cap
carefully to avoid spilling, and use the pipette to add 100µL of
EDTA-B-HC or EDTA-B-LC, as indicated on the display. Close
the cell extender cap securely. Press CAL to continue.
13. “AGit t1” ashes on the screen. Agitate to mix the sample.
When agitation is done, tap the cell extender cap to dispel any
solution clinging to the cap back into the cell. Press CAL to continue. (The number after “t” indicates which titration point you are on.
The number “1” here indicates this is the rst titration point.
Subsequent titration points are sequenced numerically: 2, 3, 4, etc.)
14. Repeat Steps 12 and 13 until the meter reports the hardness result.
15. Press CAL. The “CAL” icon will display.
16. Adjust the displayed value to "200" using the UP or DOWN key.
NOTE: If you scroll beyond ±10% of the reading "FAC" will display. If
you press the CAL key while "FAC" is displayed, the instrument will
return Hardness to factory calibration (erasing any previous wet calibration
adjustments).
32
Page 37
17. Press CAL to accept.
The unit will automatically power off.
7. pH Calibration
IMPORTANT: Always “zero” your Ultrameter III with a pH 7 buffer
solution before adjusting the gain with acid or base buffers, i.e., 4 and/
or 10, etc.
a. pH Zero Calibration
1. Rinse sensor well and cell cup 3 times with 7 buffer solution.
2. Rell both sensor well and cell cup with 7 buffer solution.
3. Press to verify the pH calibration. If the display shows
7.00, skip the pH Zero Calibration and proceed to section b.
pH Gain Calibration.
4. Press to enter calibration mode. The “CAL”, “BUFFER”
and “7” annunciators will appear (see Figure 6). Displayed
value will be the uncalibrated sensor.
NOTE: If a wrong buffer is added (outside of 6-8 pH), “7”and “BUFFER”
will ash, and the Ultrameter III will not adjust. The uncalibrated pH value
displayed in Step 4 will assist in determining the accuracy of the pH sensor.
If the pH reading is above 8 with pH 7 buffer solution, the sensor well needs
additional rinsing or the pH sensor is defective and needs to be replaced.
5. Press or
NOTE: Attempted calibration of >1 pH point from factory calibration will
cause “FAC” to appear. This indicates the need for sensor replacement
(ref. Troubleshooting pg. 52) or fresh buffer solution. The “FAC” internal
electronic calibration is not intended to replace calibration with pH
buffers. It assumes an ideal pH sensor Each “FAC” indicates a factory
setting for that calibration step (i.e., 7, acid, base).
until the display reads “7.00”.
33
Page 38
You may press
Figure 8
CAL
to accept the preset factory value, or you may
reduce your variation from factory setting by pressing
or
6. Press to accept the new value. The pH Zero Calibration
is now complete. You may continue with pH Gain Calibration or
exit by pressing any measurement key.
b. pH Gain Calibration
IMPORTANT: Always calibrate or verify your Ultrameter III with a pH 7
buffer solution before adjusting the gain with acid or base buffers, i.e.,
4 and/or 10, etc. Either acid or base solution can be used for the 2nd
point “Gain” calibration and then the opposite for the 3rd point. The
display will verify that a buffer is in the sensor well by displaying either
“Acd” or “bAS”.
1. The pH calibration mode is initiated by either completion of the
pH Zero Calibration, or verifying 7 buffer and pressing the
key twice while in pH measurement mode.
2. At this point the “CAL”, “BUFFER” and “Acd” or “bAS”
annunciators will be displayed (see Figures 7 and 8).
.
BUFFER
pH
NOTE: If the “Acd” and “bAS” indicators are blinking, the unit is indicating an
error and needs either an acid or base solution present in the sensor well.
3. Rinse sensor well 3 times with acid or base buffer solution.
4. Rell sensor well again with same buffer solution.
5. Press
or
until display agrees with buffer value.
6. Press to accept 2nd point of calibration. Now the
display indicates the next type of buffer to be used.
34
Page 39
Single point Gain Calibration is complete. You may continue for the 3rd
point of calibration (2nd Gain) or exit by pressing any measurement key.
Exiting causes the value accepted for the buffer to be used for both acid
and base measurements.
To continue with 3rd point calibration, use basic buffer if acidic buffer
was used in the 2nd point, or vice-versa. Again, match the display to the
known buffer value as in Step 2 and continue with the following steps:
7. Repeat Steps 3 through 6 using opposite buffer solution.
8. Press
completes the calibration procedure. Fill sensor well with
Myron L® Storage Solution and replace protective cap.
8. ORP Calibration
ORP electrodes rarely give false readings without problems in the
reference electrode. For this reason, and because calibration solutions
for ORP are highly reactive and potentially hazardous, your Ultrameter III
has an electronic ORP calibration. This causes the zero point on the
reference electrode to be set whenever pH 7 calibration is done.
9. Temperature Calibration
Temperature calibration is not necessary in the Ultrameter III.
VIII. CALIBRATION INTERVALS
There is no simple answer as to how often one should calibrate an
instrument. The Ultrameter III is designed to not require frequent
recalibration. The most common sources of error were eliminated in
the design, and there are no mechanical adjustments. Still, to ensure
specied accuracy, any instrument must be checked against chemical
standards occasionally.
A. Suggested Intervals
On the average, we expect calibration need only be checked monthly
for the Conductivity, RES or TDS functions. The pH function should be
checked every 2 weeks to ensure accuracy. Measuring some solutions
will require more frequent intervals.
B. Calibration Tracking Records
To minimize your calibration effort, keep records. If adjustments you
are making are minimal for your application, you can check less often.
Changes in conductivity calibration should be recorded in percent.
Changes in pH calibration are best recorded in pH units.
Calibration is purposely limited in the Ultrameter III to ±10% for the conductivity
cell, as any change beyond that indicates damage, not drift. Likewise,
calibration changes are limited to ±1 pH unit, as any change beyond that
indicates the end of the sensor’s lifetime and replacement is recommended.
to accept 3rd point of calibration, which
35
Page 40
C. Conductivity, RES, TDS, Practices to Maintain Calibration
1. Clean oily lms or organic material from the cell electrodes
with foaming cleaner or mild acid. Do not scrub inside the cell.
2. Calibrate with solutions close to the measurements you make.
Readings are compensated for temperature based on the
type of solution. If you choose to measure tap water with a
KCl compensation, which is often done (ref. An Example of 2
different solution selections and the resulting compensation,
pg. 57), and you calibrate with 442TM solution because it is
handy, the further away from 25°C you are, the more error
you have. Your records of calibration changes will reect
temperature changes more than the instrument’s accuracy.
3. Rinse out the cell with pure water after taking measurements.
Allowing slow dissolving crystals to form in the cell
contaminates future samples.
4. For maximum accuracy, keep the pH sensor cap on tight so
that no uid washes into the conductivity cell.
D. pH and ORP/Free Chlorine Equivalent Practices to
Maintain Calibration
1. The sensor well must be kept wet. Before replacing the rubber
cap, rinse and ll the sensor well with Myron L® pH Sensor
Storage Solution. Although not ideal, if Myron L® Storage
Solution is not available, you can temporarily use a strong
KCl solution, a pH 4 Buffer Solution, or a saturated solution of
table salt and tap water until pH Storage Solution is obtained.
NEVER use distilled water.
2. Rinse away caustic solutions immediately after use.
ORP calibration solutions are caustic, and ±5% is considered very
accurate. By using the pH zero setting (0 mV = 7 pH) for ORP and
precision electronics for detection, the Ultrameter III delivers better
accuracy without calibration than a simpler instrument could using
calibration solutions.
IX. MEMORY
This feature allows up to 100 readings with their temperatures to be
stored simultaneously for later recall. At the same time, the TIME and
DATE are also recorded. To download the memory to a computer, ref.
bluDock™ WIRELESS DATA TRANSFER INSTRUCTIONS, pg. 47.
A. Memory Storage
1. While displaying a measurement, press
displayed value.
36
to record the
Page 41
2. “MEMORY” will appear and the temperature display will
be momentarily replaced by a number (1-100) showing the
position of the record. Figure 9 shows a reading of 1806 µS
stored in memory record #4.
B. Memory Recall
1. Press any measurement key.
2. Press ; “MEMORY” will appear, and the display will show
the last record stored.
3. Press
(the temperature display alternates between temperature
recorded and location number).
4. Press
5. Press any measurement key to leave memory recall or allow to
automatically turn off.
C. Clearing a Record/Memory Clear
After recalling a certain record location, press and HOLD
clear that memory. This space will be the place for the next memory
record, unless you scroll to another empty position before ending the
recall sequence. The next memory stored will go into the next highest
available memory location.
Example: You have locations 1-7 lled and wish to clear the
conductivity reading stored in record location #3 and replace it with a
pH reading.
1. Press
2. Press and HOLD
3. Fill pH/ORP sensor well with sample.
4. Press
reading in location #3.
or
to display time and date stamp.
and scroll to location #3.
to measure sample and press to store
to scroll to the record location desired
to
to clear old record #3.
37
Page 42
5. The next memory stored will go into location #8.
6. To clear all records: After pressing
ALL” will be displayed (see Figure 10).
7.
X. TIME and DATE
The Time and Date may easily be changed as you travel.
A. Setting TIME
Time is always displayed in 24 hour time.
Example shown in Figure 11, 16:05 equals 4:05 PM.
1. Press .
Press
.
All records will be cleared.
, scroll down. “CLr
2. Press
stored readings, “PC OFF”, and “CLr ALL” to time, e.g.,
“16:05”).
3. Press
time (see Figure 11).
4. Press
38
until the time is displayed (scrolling through
to initiate. “CAL” will be displayed along with the
or to change the time.
Page 43
5. Press
to accept the change (new time).
B. Setting DATE
Example shown in Figure 12 is in US format, i.e., mo/dy/yr.
Figure 12
NOTE: The default format is US. Date format may be changed (ref.
Date Format “US and International (Int)”, pg. 40).
1. Press .
2. Press repeatedly until the date is displayed (scrolling
through stored readings, “PC OFF”, “CLr ALL” and time to the
date, e.g., “01.23 12” (Figure 12), January 23, 2012).
3. Press to initiate. “CAL” will be displayed along with the
YEAR (see Figure 13).
CAL
4. Press
5. Press
6. Press
7. Press
pg. 40).
Figure 13
or
to change the YEAR.
to accept the change (new year).
or
to change the month.
to accept the change (new month), (see Figure 14,
39
Page 44
CAL
Figure 14
8. Press the or
to change the day.
9. Press
to accept the change (new day) (see Figure 15).
CAL
Figure 15
C. DATE FORMAT “US & International (Int)”
1. Press .
2. Press repeatedly until the format is displayed (scrolling
through stored readings, “PC OFF”, “CLr ALL”, time and date
to date format).
3. Press
to change. Display will now indicate other format
(see Figures 16 & 17).
40
Page 45
4. Press any measurement key or allow to automatically turn off.
XI. TEMPERATURE FORMAT “Centigrade & Fahrenheit”
1. Press .
2. Press
3. Press
format location. The display will show a “C” or “F”
(see Figures 18 and 19).
4. Press
5. Press
readings.
NOTE: Tempco will still be shown in %/°C.
to display the stored memory records.
repeatedly until you pass the “US” or “Int” date
to switch units.
to accept unit preference for all temperature
XII. TOTAL RETURN to FACTORY SETTINGS “FAC SEL”
There may come a time when it would be desirable to quickly reset all
the recorded calibration values in the instrument back to the factory
settings. This might be to ensure all calibrations are set to a known
value, or to give the instrument to someone else free of adjustments or
recorded data for a particular application.
NOTE: All stored data will be lost.
1. Press
2. Press to display the stored memory records.
.
41
Page 46
3. Press
C-F locations. The display will show a “FAC SEL” (see Figure 20).
4. Press to accept the resetting. Display will return to
Conductivity.
XIII. CELL CHECK
The cell check veries the cleanliness of the conductivity/TDS/
resistivity sensor. In normal use the cell may become dirty or coated
and require cleaning. If the display is showing “.00” when the cell cup
is dry, the sensor is probably clean. However, when testing high purity
water in resistivity (“RES”) mode improved accuracy may be desired.
No matter what a manufacturer claims, a sensor can and will become
contaminated or coated and, therefore, require cleaning. A true 4-wire
sensor, as in the Ultrameter III, helps to mitigate contamination, but
NO SENSOR IS 100% IMMUNE.
repeatedly until you pass the “CLr ALL” and the
1. Press .
2. Press
3. Press
The display will show a “CELL ch” (see Figure 21).
42
to display the stored memory records.
repeatedly until you pass the FAC SEL location.
Page 47
4. Press
displayed (see Figure 22). If cell is dirty, “CELL cLn” will be
displayed (see Figure 23) (ref. Cleaning Sensors, pg. 50).
XIV. AUTO OFF
Auto off allows the user to adjust the time the instrument is ON (up to
75 seconds) after each press of a key. Default time is 15 seconds with
60 seconds in “CAL” (calibration) mode.
to test. If cell is clean, “Good” will momentarily be
1. Press
2. Press to display the stored memory records.
3. Press
The display will show “Auto oFF” (see Figure 24).
4. Press
SEC” or current Auto oFF value (see Figure 25, pg. 44).
.
repeatedly until you pass the “CELL ch” location.
to initiate. “CAL” will be displayed along with “15
43
Page 48
5. Press
Figure 26). Maximum time of 75 seconds is shown.
or
to change the amount of time (see
6. Press
XV. USER Mode CALIBRATION LINC™ FUNCTION
The Linc™ function allows easy calibration when in User mode and the
user does not have a user standard solution to calibrate the instrument.
This function will ensure more repeatable and accurate measurements
than many other calibration methods. It is recommended that this
function be used to provide the highest degree of condence when the
Ultrameter III is used in User mode. When Linc is used, the User mode
is linked to another standard, i.e., if User and KCl are linked, a KCI
standard solution is used to calibrate the instrument. It is that simple.
A. Calibration of Ultrameter
1. Press or
2. Calibrate the unit using a Standard Solution (ref.
CALIBRATION, pg. 26).
3. Place the Ultrameter III in User mode (ref. SOLUTION
SELECTION, pg. 23).
4. Verify/Set the calibration linc. (See below – Setting “User”
Mode Calibration “Linc”.)
44
to accept the change (new time).
for use in User mode
III
key.
Page 49
B. Setting User mode Calibration “Linc”
The Linc function sets or “links” the calibration gain factor of a Standard
Solution to the User solution mode. Once set, the “Linc” will stay intact
with future calibrations unless the Linc has been canceled. For more
information on canceling the User mode Calibration Linc refer to the
section C. Canceling User mode Calibration “Linc”, pg. 46.
Follow the steps below to set either the KCl, NaCl or 442TM calibration
factor to the User solution mode.
1. Press measurement key desired to be “Linked”, i.e., ,
or .
2. Place the Ultrameter III in User mode (ref. SOLUTION
SELECTION, pg. 23, for selecting the User mode).
3. Press
Figure 27).
4. Press
“User” Icon (see Figure 28).
arrow key until the menu “Linc” appears (see
key. The instrument will display “SEL” and the
Any additional display of KCl, NaCl or 442TM icons indicates
a “Linc” between the User solution and the other solution
displayed.
45
Page 50
5. Press or keys to select a Standard Solution to be
linked to the User mode calibration constant. In Figure 29, pg.
46, the display indicates that “User” is linked to “KCl”.
If none of the Solution Selection icons are displayed (i.e., KCl,
NaCl or 442TM), nothing has been linked to User mode.
6. Press
measurement keys will exit without changing the setting. User
mode “Linc” is now complete. The User mode will now use
the calibration gain constant used for the calibration of the
Standard Solution as outlined above.
C. Canceling User mode Calibration “Linc”
The Ultrameter III must be in User linked mode in order to cancel the
“Linc” (ref. Solution Selection, pg. 23).
1. Press “Linked” measurement key ,
Two solution icons will be shown in the left side of display —
“User” and another, e.g., “KCl”.
2. Press
3. Press
the “User” Icon.
4. Press
displayed.
key to accept the setting. Pressing any of the
.
or
key until the menu “Linc” appears (see Figure 27).
key; the instrument will display both “SEL” and
key until “User” is the only solution icon being
5. Press
6. The User mode calibration “Linc” has now been canceled.
46
key.
Page 51
NOTES:
1. To maintain repeatability, use the same standard solutions for
future calibrations.
2. Calibration of the Ultrameter III Gain Factor for User mode is
not available when the calibration linc has been established.
The other calibration functions (i.e., Temperature Compensation
%/C settings and TDS Ratio settings) are still intact. To perform
a calibration of the User mode as described in User Calibration
Conductivity/TDS, pg. 28, the User mode Linc should be canceled.
See above, “Canceling User mode Calibration “Linc””.
3. Once a “Linc” has been established for User mode, the “Linc”
will apply to all measurement modes using User solution
selection (i.e., TDS/User, Cond/User or Res/User).
XVI. bluDock™ WIRELESS DATA TRANSFER INSTRUCTIONS
NOTE: The Bluetooth® word mark and logos are registered trademarks
owned by the Bluetooth® SIG, Inc. and any use of such marks by
Mryon L® Company is under license.
Requires a bluDock enabled instrument. If your 9P does not include
the bluDock option ("-BD") you may purchase the Myron L® bluDock
accessory package, Model # BLUDOCK.
Package includes Ultrameter III hardware modication that allows the
unit to communicate wirelessly with a personal computer congured for
wireless device communication. Package includes the bluDock software,
Guardian
TM
2
, that will operate on Windows®* 7 or later/macOS®** 10.13 or
later based computer systems.
A. Software Installation
1. Download Guardian2 application from theMyron L® website: http://www.myronl.com/.
Select the Downloads tab and click on Guardian2 Applications.
2. Download and install the Guardian2 application for your
operating system. You will also nd the Quick Start Guide
and Operation Manual for the Guardian2 Application on this
webpage. Additional drivers may be required. See our website
for the latest information.
B. Hardware Setup
For a computer without Bluetooth capability:
A Bluetooth dongle may be purchased from your local electronics
store. Plug in the dongle and install per manufacturer’s instructions.
* Windows® is a registered trademark of Microsoft Corporation.
** macOS® is a registered trademark of Apple, Inc.
47
Page 52
For computers with Bluetooth capability/Bluetooth dongle installed:
First time use of the bluDock:
1. Press any parameter button to turn the Ultrameter III on.
2. Put the Ultrameter III in “PC On” mode by pressing the
key until “PC OFF” appears (see Figure 30).
3. Then press the key. “PC On” will be displayed (see
Figure 31).
NOTE: “PC Ini” may momentarily be displayed while initializing (see
Figure 32).
4. The bluDock device will appear as MyronL-XXXX in your
computer's Bluetooth® devices list. While pairing with the
bluDock device it will ask for the pin, THE BLUDOCK DEVICE
PIN IS 1234.
5. After pairing, note the bluDock number for your device,
MyronL-XXXX.
(For complete instructions see the Guardian2 Operation
Manual for your computer operating system.)
NOTE: The unit will automatically power down after 3 minutes. If the
unit powers down during pairing, repeat Steps 1-3 above and continue.
48
Page 53
C. Memory Stack Download
1. With the Ultrameter III in “PC On” mode, open the software
application.
2. Verify that the device name selected matches the bluDock
number for your device. (MyronL-XXXX).
3. In the application, click on the data download button. A data
transfer bar will appear while the data is being downloaded.
Once downloaded, the data may be manipulated, or printed within
the Myron L® application, or the data may be exported to spreadsheet
formats such as Microsoft Excel® (.xls and .xlsx) or universal formatting
(.csv). The user can also save data in a secure, encrypted format (.mlcx).
Please Note: Although the Myron L® Company has performed extensive
testing, we cannot guarantee compatibility with all computers/operating
systems.
Additional features such as assigning a name to the unit, setting time
and date and erasing data are available. Visit our website for the latest
instructions: http://www.myronl.com
4.Turn off Ultrameter III PC download mode by selecting any
measurement function. Failure to do so will reduce battery life.
XVII. CARE and MAINTENANCE
•Ultrameter IIIs should be rinsed with clean water after use.
•Solvents should be avoided.
•Do not drop, throw or otherwise strike the instrument. Shock
damage may cause instrument failure.
A. Temperature Extremes
Solutions in excess of 71°C/160°F should not be placed in the cell
cup area; this may cause damage. The pH sensor may fracture if the
Ultrameter III temperature is allowed to go below 0°C/32°F. Care should
be exercised not to exceed rated operating temperature.
Leaving the Ultrameter III in a vehicle or storage shed on a hot day
can easily subject the instrument to over 66°C/150°F. This will void the
warranty.
B. Battery Replacement
Dry Instrument THOROUGHLY. Remove the four (4) bottom screws.
Open instrument carefully. Carefully detach battery from circuit board.
Replace with 9 Volt alkaline battery. Replace bottom, ensuring the
sealing gasket is installed in the groove of the top half of case. Re-install
screws, tighten evenly and securely.
49
Page 54
NOTE: Because of nonvolatile EEPROM circuitry, all data stored in
memory and all calibration settings are protected even during power
loss or battery replacement. However, loss of time and date may occur if
battery is removed for longer than 3 minutes (180 seconds).
C. pH/ORP Sensor Replacement
Order model RPR. When ordering, be sure to include the model and serial
number of your instrument to ensure receipt of the proper type. Complete
installation instructions are provided with each replacement sensor.
D. Cleaning Sensors
1. Conductivity/TDS/Resistivity
The conductivity cell cup should be kept as clean as possible. Flushing
with clean water following use will prevent buildup on electrodes.
However, if very dirty samples — particularly scaling types — are
allowed to dry in the cell cup, a lm will form. This lm reduces accuracy.
When there are visible lms of oil, dirt, or scale in the cell cup or on the
electrodes, use isopropyl alcohol or a foaming non-abrasive household
cleaner. Rinse out the cleaner and your Ultrameter III is again ready for
accurate measurements.
2. pH/ORP
The unique pH/ORP sensor in your Ultrameter III is a nonrellable
combination type that features a porous liquid junction. It should not be
allowed to dry out. However, if this occurs, the sensor may sometimes
be rejuvenated by rst cleaning the sensor well with Isopropyl alcohol
or a liquid spray cleaner such as Windex™ or Fantastic™ and rinsing
well. Do not scrub or wipe the pH/ORP sensor.
50
Page 55
Then use one of the following methods:
1. Pour a HOT salt solution ~60°C/140°F — a potassium
chloride (KCI) solution such as Myron L® pH/ORP Sensor
Storage Solution is preferable, but HOT tap water with table
salt (NaCl) will work ne — in the sensor well and allow to
cool. Retest.
or
2. Pour DI water in the sensor well and allow to stand for no more
than 4 hours (longer can deplete the reference solution and
damage the glass bulb). Retest.
If neither method is successful, the sensor must be replaced.
“Drifting” can be caused by a lm on the pH sensor bulb and/or
reference. Use isopropyl alcohol (IPA) or spray a liquid cleaner such
as Windex™ or Fantastic™ into the sensor well to clean it. The sensor
bulb is very thin and delicate. Do not scrub or wipe the pH/ORP sensor.
Leaving high pH (alkaline) solutions in contact with the pH sensor for
long periods of time is harmful and will cause damage. Rinse such
liquids from the pH/ORP sensor well and rell it with Myron L® Storage
Solution to extend the useful life of the sensor.
Samples containing chlorine, sulfur, or ammonia can “poison” any pH
electrode. If it is necessary to measure the pH of any such sample,
thoroughly rinse the sensor well with clean water immediately after
taking the measurement. Any sample element that reduces (adds an
electron to) silver, such as cyanide, will attack the reference electrode.
Replacement sensors are available only from the Myron L® Company
or its authorized distributors (ref. Replacement pH/ORP Sensor, pg.
56).
It is possible to desensitize the ORP sensor when measuring
high ORP solutions. To check the ORP sensor function, take
an ORP reading of Myron L® pH/ORP sensor storage solution.
If the reading is outside the range of 350-400 mV, clean ONLY
the platinum ORP electrode with Myron L® ORP Conditioner
solution-soaked cotton swab, being careful not to touch the
swab to the glass bulb of the pH sensor.
No response to pH changesSensor bulb is cracked or an
Will not adjust down to pH 7pH/ORP sensor has lost KCl.Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 50) and recalibrate. If no
Battery weak or not connected.Check connections or replace battery. Ref. Battery Replacement, pg. 49.
pg. 34.
2. Cross-contamination from residual pH
buffers or samples in sensor well.
3. Calibration with expired pH buffers.
electromechanical short caused by an
internal crack.
pH readings drift or respond slowly to
changes in buffers/samples or “FAC” is
displayed repeatedly
Unstable Conductivity/TDS/ Resistivity readings
Unable to calibrate Conductivity/TDSFilm or deposits on electrodes.Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 50.
Resistivity readings much lower than
expected
Alkalinity or Hardness titration with
calibration solution value not as
stated on bottle
Low ORP Reading
Slow or no response to ORP changes
FCE responds very slowly or returns an
atypically high Predictive ORP value
1. Temporary condition due to memory” of
solution in pH sensor well for long periods.
2. Bulb dirty or dried out.
3. Reference junction clogged or coated.
1. Dirty electrodes.
2. Actual resistance is changing due to
atmospheric contamination.
1. Contamination from previous sample or
from pH sensor well.
2. Carbon dioxide in test sample.
1. Operation error.
2. Dirty Cell Cup.
3. Dirty pipette.
4. Faulty pipette.
5. Contaminated calibration solutions.
ORP platinum electrode is dirty.
1. Dirty platinum electrode (see above).
2. ORP sensor memory/battery effect.
Some ORP sensors exhibit a residual
charge when measuring LOW Free
Chlorine concentrations soon after
measuring a HIGH Free Chlorine
Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 50) and recalibrate. If no
improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 56).
1. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 50.
2. Minimize test sample exposure to air by taking a owing sample.
Ref. Measuring Resistivity, pg. 14.
1. Rinse cell cup more thoroughly before measurement. Ensure pH cap is snugly in
place.
2. See Measuring Resistivity, pg. 14.
1. Repeat titration.
2. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 50.
3. Clean pipette per manufacturer’s instructions.
4. Use a new pipette.
5. Use fresh solutions. Be sure to use a new tip for each type of reagent used.
Check the ORP sensor function. Take an ORP reading of Myron L® pH/ORP
Sensor Storage Solution (ref. pH Sensor Storage Solution, pg. 55). If the reading is
outside the range of 350-400 mV, clean ONLY the platinum ORP
electrode with Myron L® ORP Conditioner solution-soaked cotton swab (ref. ORP Sensor Conditioner Solution, pg. 55), being careful not to touch the swab to the glass bulb
of the pH sensor.
1. Rinse the pH/ORP sensor well briey with a small amount of ORP Sensor
Conditioner Solution. DO NOT leave the conditioning solution in the sensor well
for more than 10 seconds.
2. Rinse the pH/ORP sensor 3 times with Sensor Storage Solution.
3. Fill the sensor well with Sensor Storage Solution and let rest for 5 minutes.
53
Page 58
XIX. ACCESSORIES
NOTE: SDSs are available on the Myron L® website for all solutions:
http://www.myronl.com/main/Material_Safety_DS_DL.htm
A. Conductivity/TDS Standard Solutions
Your Ultrameter III has been factory calibrated with the appropriate
Myron L® Company NIST traceable KCl, NaCl, and our own 442™
standard solutions. Most Myron L® conductivity standard solution bottles
show three values referenced at 25°C: Conductivity in microsiemens/
micromhos, the ppm/TDS equivalents (based on our 442 Natural
Water™) and NaCl standards. All standards are within ±1.0% of
reference solutions. Available in 2 oz., quarts/liters, and gallon/~3.8 liter
bottles.
1. Potassium Chloride (KCl)
The concentrations of these reference solutions are calculated from
data in the International Critical Tables, Vol. 6. The 7000 µS is the
recommended standard. Order KCL-7000.
2. 442 Natural Water™
442 Natural Water Standard Solutions are based on the following salt
proportions: 40% sodium sulfate, 40% sodium bicarbonate, and 20%
sodium chloride, which represent the three predominant components
(anions) in freshwater. This salt ratio has conductivity characteristics
approximating fresh natural waters and was developed by the
Myron L® Company over ve decades ago. It is used around the world
for measuring both conductivity and TDS in drinking water, ground
water, lakes, streams, etc. 3000 ppm is the recommended standard.
Order 442-3000.
3. Sodium Chloride (NaCl)
This is especially useful in seawater mix applications, as sodium chloride
is the major salt component. Most Myron L® standard solution labels
show the ppm NaCl equivalent to the conductivity and to ppm 442
values. The 14.0 mS is the recommended standard. Order NACL-14.0.
B. Titration Standard Solutions
Myron L® titration standard solution bottles show both the actual chemical
in solution and the ppm calcium carbonate (CaCO3) equivalent. All
standards are within ±1.0% of reference solutions. Available in 2 oz.,
and quart/liter bottles.
1. Alkalinity Standard Solution
ALK(NaHCO3)-100 is a solution of sodium bicarbonate (NaHCO3)
equivalent to 100 ppm calcium carbonate that is used for the alkalinity
calibration: Order ALK-100.
54
TM
Page 59
2. Hardness Standard Solution
HARD(CaCl2)-200 solution is a solution of calcium chloride (CaCl2)
equivalent to 200 ppm calcium carbonate that is used for the hardness
calibration: Order HARD-200.
C. Titration Reagent Solutions
Titration reagents are required for all conductometric titrations - alkalinity
and hardness. Reagents are available in 2 oz. and quarts/liters.
1. Reagent: A1
Reagent: A1 is a ≤1%WV solution of sulfuric acid (H2SO4-1) required for
alkalinity titration: Order H2SO4-1.
2. Reagent: HB
Reagent: HB is a ≤15% WV solution of pH buffer required for hardness
titration: Order HBUFF.
3. Reagent: H4
Reagent: H4 is a low concentration ≤5%WV solution of
Ethylenediaminetetraacetic acid (EDTA) required for hardness titration:
Order EDTA-B-LC.
4. Reagent: H5
Reagent: H5 is a high concentration ≤10%WV solution of
Ethylenediaminetetraacetic acid (EDTA) required for hardness titration:
Order EDTA-B-HC.
D. pH Buffer Solutions
pH buffers are available in pH values of 4, 7 and 10. Myron L® Company
buffer solutions are traceable to NIST certied pH references and are
color-coded for instant identication. They are also mold inhibited and
accurate to within ±0.01 pH units @ 25°C. Order 4, 7 or 10 Buffer.
Available in 2 oz., quarts/liters, and gallon/~3.8 liter bottles.
E. pH Sensor Storage Solution
Myron L® pH Sensor Storage Solution prolongs the life of the pH sensor.
Available in 2 oz., quarts/liters, and gallon/~3.8 liter bottles. Order SS.
F. ORP Sensor Conditioner Solution
Myron L® ORP Conditioner Solution removes contaminants and
conditions the ORP electrode. Available in 1 oz. Order ORPCOND1OZ.
G. Soft Protective Carry Cases
Padded Nylon carrying case features a belt clip for hands-free mobility.
Two colors to choose from;
Blue - Model #: UCC
Desert Tan - Model #: UCCDT
55
Page 60
H. Hard Protective Carry Cases
Large, foam-lined case without UMIII 9P includes the 100µL pipette,
12 disposable pipette tips, 3 buffers (pH 4, 7, and 10), pH/ORP sensor
storage solution, ORP Conditioner solution, reagents A1, HB, H4, H5,
and standard solutions, KCI-7000, 442-3000, ALK-100 and HARD-200.
All bottles are 2 oz/59 ml (except for ORP Conditioner solution which is 1
oz.). Model: TK9B
Small case (for instrument only, no solutions) - Model #: UPP
I. Replacement pH/ORP Sensor
pH/ORP sensor is gel lled and features a unique porous liquid junction.
It is user-replaceable and comes with easy to follow instructions.
Model #: RPR
J. Pipette Kit with Replacement Tips
3 color coded M’Pet pipettes with 48 replacement tips. This allows you
to have a dedicated pipette for each titration reagent. Colored dots on
each pipette match the colored dots on reagent solution bottles to help
you avoid confusion and contamination. Model #: PROKITB
K. Replacement M’Pet Pipette
1 replacement 100µL xed volume pipette. Model #: FVMP-100
L. M’Pet Pipette Replacement Tips
Bag of 48 pipette tips. Model #: PTIP100-48
M. bluDock Wireless Data Transfer Accessory Package
TM
This accessory allows the operator to download the Ultrameter III
memory stack to a spreadsheet on a computer. The package includes
bluDock modied circuit board to be installed inside the unit, software
installation and operating instructions. Model #: BLUDOCK
XX. TEMPERATURE COMPENSATION (Tempco)
of Aqueous Solutions
Electrical conductivity indicates solution concentration and ionization
of the dissolved material. Since temperature greatly affects ionization,
conductivity measurements are temperature dependent and are normally
corrected to read what they would be at 25°C.
A. Standardized to 25°C
Conductivity is measured with great accuracy in the Ultrameter III using a
method that ignores ll level, electrolysis, electrode characteristics, etc.,
and features a microprocessor to perform temperature compensation. In
simpler instruments, conductivity values are usually assigned an average
correction similar to that of KCl solutions for correction to 25°C. The
correction to an equivalent KCl solution is a standard set by chemists
that standardizes the measurements and allows calibration with precise
KCl solutions. In the Ultrameter III, this correction can be set to other
solutions or tailored for special measurements or applications.
56
Page 61
B. Tempco Variation
Most conductivity instruments use an approximation of the temperature
characteristics of solutions, perhaps even assuming a constant value.
The value for KCl is often quoted simply as 2%/°C. In fact, KCl tempco
varies with concentration and temperature in a non-linear fashion. Other
solutions have more variation still. The Ultrameter III uses corrections
that change with concentration and temperature instead of single
average values. See Chart 1.
C. An Example of 2 different solution selections and the
resulting compensation
How much error results from treating natural water as if it were KCl at
15°C?
A tap water solution should be compensated as 442TM with a tempco of
1.68 %/°C, where the KCl value used would be 1.90 %/°C.
Suppose a measurement at 15°C/59°F is 900 microsiemens of true
uncompensated conductivity.
Using a 442TM correction of 10 (degrees below 25) x 1.68% indicates the
solution is reading 16.8% low. For correction, dividing by (.832) yields
1082 microsiemens as a compensated reading.
57
Page 62
A KCl correction of 10 (degrees below 25) x 1.9% indicates the solution
is reading 19% low. Dividing by (.81) yields 1111 microsiemens for a
compensated reading. The difference is 29 out of 1082 = 2.7%.
D. A Chart of Comparative Error
In the range of 1000 µS, the error using KCl on a solution that should
be compensated as NaCl or as 442TM, is illustrated in the graph below.
Users wanting to measure natural water based solutions to 1% would
have to alter the internal compensation to the more suitable preloaded
“442TM” values or stay close to 25°C. Users who have standardized
to KCl- based compensation may want to stick with it, regardless of
increasing error as you get further from 25°C. The Ultrameter III will
provide the repeatability and convertibility of data necessary for relative
values for process control.
E. Other Solutions
A salt solution like seawater or liquid fertilizer acts like NaCl. An internal
correction for NaCl can be selected for greatest accuracy with such
solutions. Many solutions are not at all similar to KCl, NaCl or 442TM. A
sugar solution, or a silicate, or a calcium salt at a high or low temperature
may require a “User” value unique for the application to provide readings
close to the true compensated conductivity.
58
Page 63
Clearly, the solution characteristics should be chosen to truly represent
the actual water under test for rated accuracy of ±1%. Many industrial
applications have historically used relative measurements seeking a
number to indicate a certain setpoint or minimum concentration or trend.
The Ultrameter III gives the user the capability to collect data in “KCl
conductivity units” to compare to older published data, in terms of NaCl
or 442TM, or as appropriate. The Ultrameter III can be used to reconcile
data taken with other compensation assumptions, especially with its
ability to allow custom characteristics through the User mode.
XXI. CONDUCTIVITY CONVERSION to TOTAL DISSOLVED
SOLIDS (TDS)
Electrical conductivity indicates solution concentration and ionization
of the dissolved material. Since temperature greatly affects ionization,
conductivity measurements are temperature dependent and are
normally corrected to read what they would be at 25°C (ref. Temperature
Compensation, pg. 56).
A. How it’s Done
Once the effect of temperature is removed, the compensated conductivity
is a function of the concentration (TDS). Temperature compensation of
the conductivity of a solution is performed automatically by the internal
processor with data derived from chemical tables. Any dissolved salt at
a known temperature has a known ratio of conductivity to concentration.
Tables of conversion ratios referenced to 25°C have been published by
chemists for decades.
B. Solution Characteristics
Real world applications have to measure a wide range of materials and
mixtures of electrolyte solutions. To address this problem, industrial users
commonly use the characteristics of a standard material as a model for
their solution, such as KCl, which is favored by chemists for its stability.
Users dealing with seawater, etc., use NaCl as the model for their
concentration calculations. Users dealing with freshwater work with
mixtures including sulfates, carbonates and chlorides, the three
predominant components (anions) in freshwater that the
Myron L® Company calls “natural water”. These are modeled in a
mixture called “442™” which the Myron L® Company markets for use
as a calibration standard, as it does standard KCl and NaCl solutions.
The Ultrameter III contains algorithms for these 3 most commonly
referenced compounds. The solution type in use is displayed on the
left. Besides KCl, NaCl, and 442TM, there is the “User” choice. The
benet of “User” is that one may enter the temperature compensation
and TDS ratio by hand, greatly increasing accuracy of readings for a
specic solution. That value remains a constant for all measurements
and should be reset for different dilutions or temperatures.
59
Page 64
C. When does it make a lot of difference?
First, the accuracy of temperature compensation to 25°C determines the
accuracy of any TDS conversion. Assume we have industrial process
water to be pretreated by RO. Assume it is 45°C and reads 1500 µS
uncompensated.
1. If NaCl compensation is used, an instrument would report 1035
µS compensated, which corresponds to 510 ppm NaCl.
2. If 442TM compensation is used, an instrument would report
1024 µS compensated, which corresponds to 713 ppm 442TM.
The difference in values is 40%.
In spite of such large error, some users will continue to take data in
the NaCl mode because their previous data gathering and process
monitoring was done with an older NaCl referenced device.
Selecting the correct Solution Type on the Ultrameter III will allow the
user to attain true TDS readings that correspond to evaporated weight.
If none of the 3 standard solutions apply, the User mode must be used.
Temperature Compensation (Tempco) and TDS Derivation below, details
the User mode.
XXII. TEMPERATURE COMPENSATION (Tempco) and TDS
DERIVATION
The Ultrameter III contains internal algorithms for characteristics of the
3 most commonly referenced compounds. The solution type in use is
displayed on the left. Besides KCl, NaCl, and 442TM, there is the User
choice. The benet of User mode is that one may enter the tempco and
TDS conversion values of a unique solution via the keypad.
A. Conductivity Characteristics
When taking conductivity measurements, the Solution Selection
determines the characteristic assumed as the instrument reports what
a measured conductivity would be if it were at 25°C. The characteristic
is represented by the tempco, expressed in %/°C. If a solution of 100
µS at 25°C increases to 122 µS at 35°C, then a 22% increase has
occurred over this change of 10°C. The solution is then said to have a
tempco of 2.2 %/°C.
Tempco always varies among solutions because it is dependent on their
individual ionization activity, temperature and concentration. This is why
the Ultrameter III features mathematically generated models for known
salt characteristics that also vary with concentration and temperature.
60
Page 65
B. Finding the Tempco of an Unknown Solution
One may need to measure compensated conductivity of some solution
unlike any of the 3 standard salts. In order to enter a custom xed tempco
for a limited measurement range, enter a specic value through the User
function. The tempco can be determined by 2 different methods:
1. Heat or cool a sample of the solution to 25°C, and measure its
conductivity. Heat or cool the solution to a typical temperature
where it is normally measured. After selecting User function,
set the tempco to 0 %/°C as in Disabling Temperature
Compensation, pg. 25 (No compensation). Measure the new
conductivity and the new temperature. Divide the % decrease
or increase by the 25°C value. Divide that difference by the
temperature difference.
2. Heat or cool a sample of the solution to 25°C, and measure
its conductivity. Change the temperature to a typical measuring
temperature. Set the tempco to an expected value as in User
Programmable Temperature Compensation, pg. 25. See if the
compensated value is the same as the 25°C value. If not, raise
or lower the tempco and measure again until the 25°C value is
read.
C. Finding the TDS Ratio of an Unknown Solution
Once the effect of temperature is removed, the compensated
conductivity is a function of the concentration (TDS). There is a ratio
of TDS to compensated conductivity for any solution, which varies
with concentration. The ratio is set during calibration in User mode
as in User Programmable Conductivity to TDS Ratio, pg. 25. A truly
unknown solution has to have its TDS determined by evaporation
and weighing. Then the solution whose TDS is now known can be
measured for conductivity and the ratio calculated. Next time the same
solution is to be measured, the ratio is known.
61
Page 66
XXIII. pH and ORP/FC
E
A. pH
1. pH as an Indicator
pH is the measurement of Acidity or Alkalinity of an aqueous solution. It
is also stated as the Hydrogen Ion activity of a solution. pH measures
the effective, not the total, acidity of a solution.
A 4% solution of acetic acid (pH 4, vinegar) can be quite palatable, but
a 4% solution of sulfuric acid (pH 0) is a violent poison. pH provides the
needed quantitative information by expressing the degree of activity of
an acid or base.
In a solution of one known component, pH will indicate concentration
indirectly. However, very dilute solutions may be very slow reading,
just because the very few ions take time to accumulate.
2. pH Units
The acidity or alkalinity of a solution is a measurement of the relative
availabilities of hydrogen (H+) and hydroxide (OH-) ions. An increase in
(H+) ions increases acidity, while an increase in (OH-) ions increases
alkalinity. The total concentration of ions is xed as a characteristic
of water, and balance would be 10
-
7
mol/liter (H+) and (OH-) ions in a
neutral solution (where pH sensors give 0 voltage).
pH is dened as the negative logarithm of hydrogen ion concentration.
Where (H+) concentration falls below 10-7, solutions are less acidic
than neutral, and therefore are alkaline. A concentration of 10-9 mol/
liter of (H+) would have 100 times less (H+) ions than (OH-) ions and be
called an alkaline solution of pH 9.
3. The pH Sensor
The active part of the pH sensor is a thin glass surface that is selectively
receptive to hydrogen ions. Available hydrogen ions in a solution will
accumulate on this surface and a charge will build up across the glass
interface. The voltage can be measured with a very high impedance
voltmeter circuit; the dilemma is to connect the voltmeter to solution on
each side.
The glass surface encloses a captured solution of potassium chloride
holding an electrode of silver wire coated with silver chloride. This is
the most inert connection possible from a metal to an electrolyte. It can
still produce an offset voltage, but using the same materials to connect
to the solution on the other side of the membrane causes the 2 equal
offsets to cancel.
62
Page 67
The problem is, on the other side of the membrane is an unknown test
solution, not potassium chloride. The outside electrode, also called the
Reference Junction, is of the same construction with a porous plug
in place of a glass barrier to allow the junction uid to contact the
test solution without signicant migration of liquids through the plug
material. Figure 33 shows a typical 2 component pair. Migration does
occur, and this limits the lifetime of a pH junction from depletion of
solution inside the reference junction or from contamination. The
junction may be damaged if dried out because insoluble crystals may
form in a layer, obstructing contact with test solutions.
4. The Myron L Integral pH Sensor
®
The sensor in the Ultrameter III (see Figure 34) is a single construction
in an easily replaceable package. The sensor body holds an oversize
solution supply for long life. The reference junction “wick” is porous to
provide a very stable, low permeable interface, and is located under
the glass pH sensing electrode. This construction combines all the
best features of any pH sensor known.
63
Page 68
5. Sources of Error
The basics are presented in pH and ORP/FCE, pg. 62.
a. Reference Junction
The most common sensor problem will be a clogged junction because
a sensor was allowed to dry out. The symptom is a drift in the “zero”
setting at 7 pH. This is why the Ultrameter III 9P does not allow more
than 1 pH unit of offset during calibration. At that point the junction is
unreliable.
b. Sensitivity Problems
Sensitivity is the receptiveness of the glass surface. A lm on the surface
can diminish sensitivity and cause a long response time.
c. Temperature Compensation
pH sensor glass changes its sensitivity slightly with temperature, so the
further from pH 7 one is, the more effect will be seen. A pH of 11 at 40°C
would be off by 0.2 units. The Ultrameter III 9P senses the sensor well
temperature and compensates the reading.
B. ORP/Oxidation-Reduction Potential/REDOX
1. ORP as an Indicator
ORP is the measurement of the ratio of oxidizing activity to reducing
activity in a solution. It is the potential of a solution to give up electrons
(oxidize other things) or gain electrons (reduce).
Like acidity and alkalinity, the increase of one is at the expense of the
other, so a single voltage is called the Oxidation-Reduction Potential,
with a positive voltage showing, a solution wants to steal electrons
(oxidizing agent). For instance, chlorinated water will show a positive
ORP value.
2. ORP Units
ORP is measured in millivolts, with no correction for solution temperature.
Like pH, it is not a measurement of concentration directly, but of activity
level. In a solution of only one active component, ORP indicates
concentration. Also, as with pH, a very dilute solution will take time to
accumulate a readable charge.
3. ORP Sensors
An ORP sensor uses a small platinum surface to accumulate charge
without reacting chemically. That charge is measured relative to the
solution, so the solution “ground” voltage comes from a reference
junction - same as the pH sensor uses.
64
Page 69
4. The Myron L ORP Sensor
®
Figure 34, pg. 63, shows the platinum button in a glass sleeve. The
same reference is used for both the pH and the ORP sensors. Both
pH and ORP will indicate 0 for a neutral solution. Calibration at zero
compensates for error in the reference junction.
A zero calibration solution for ORP is not practical, so the Ultrameter III
uses the offset value determined during calibration to 7 in pH calibration
(pH 7 = 0 mV). Sensitivity of the ORP surface is xed, so there is no gain
adjustment either.
5. Sources of Error
The basics are presented in pH and ORP/FCE, pg. 62, because sources
of error are much the same as for pH. The junction side is the same,
and though the platinum surface will not break like the glass pH surface,
its protective glass sleeve can be broken. A surface lm will slow
the response time and diminish sensitivity. It can be cleaned off with
detergent or acid, as with the pH glass.
C. Free Chlorine Equivalent
1. Free Chlorine Equivalent as an Indicator
Chlorine, which kills bacteria by way of its power as an oxidizing agent,
is the most popular germicide used in water treatment. Chlorine is not
only used as a primary disinfectant, but also to establish a sufcient
residual level of Free Available Chlorine (FAC) for ongoing disinfection.
FAC is the chlorine that remains after a certain amount is consumed by
killing bacteria or reacting with other organic (ammonia, fecal matter) or
inorganic (metals, dissolved CO2, Carbonates, etc) chemicals in solution.
Measuring the amount of residual free chlorine in treated water is a well
accepted method for determining its effectiveness in microbial control.
The Myron L® Company FCE method for measuring residual disinfecting
power is based on ORP, the specic chemical attribute of chlorine (and
other oxidizing germicides) that kills bacteria and microbes.
2. Free Chlorine Equivalent Units
The 9P is the rst handheld device to detect free chlorine directly, by
measuring ORP. The ORP value is converted to a concentration reading
(ppm) using a conversion table developed by Myron L® Company
through a series of experiments that precisely controlled chlorine levels
and excluded interferants.
Other test methods typically rely on the user visually or digitally
interpreting a color change resulting from an added reagent-dye. The
reagent used radically alters the samples pH and converts the various
chlorine species present into a single, easily measured species. This
ignores the effect of changing pH on free chlorine effectiveness and
65
Page 70
disregards the fact that some chlorine species are better or worse
sanitizers than others.
The Myron L® Company 9P avoids these pitfalls. The chemistry of the
test sample is left unchanged from the source water. It accounts for the
effect of pH on chlorine effectiveness by including pH in its calculation.
For these reasons, the Ultrameter III’s FCE feature provides the best
reading-to-reading picture of the rise and fall in sanitizing effectivity of
free available chlorine.
The 9P also avoids a common undesirable characteristic of other ORPbased methods by including a unique Predictive ORP value in its FCE
calculation. This feature, based on a proprietary model for ORP sensor
behavior, calculates a nal stabilized ORP value in 1 to 2 minutes rather
than the 10 to 15 minutes or more that is typically required for an ORP
measurement.
XXIV. ALKALINITY, HARDNESS AND LSI FUNCTIONS
A. Alkalinity and Hardness Titrations
The 9P uses a conductometric method of titration. This means that
instead of measuring a color or pH change, it measures the change
in conductivity when reagent is added. The endpoint is determined
based on the conductivity changes and then the alkalinity/hardness is
calculated based on the total volume of titrant consumed. For hardness
titration, the sample conductivity must be ≤10,000 uS/cm in KCl mode.
It may not work with strong buffered samples or samples with extreme
pH (e.g. <2 pH). The titration is based on EDTA complexation reactions.
Metal ions from the alkaline earth metal group, such as Ba2+ and Sr2+
will be measured as hardness. Interferences including aluminum, iron,
nickel, cobalt, manganese, etc. may be corrected by inhibitors.
B. Langelier Saturation Index
The LSI calculator function uses a calcium carbonate saturation index
algorithm developed by Dr. Wilfred Langelier in 1936:
SI = PH + TF + CF + AF – 12.1
Where:
PH = pH value
TF = 0.0117 x Temperature value – 0.4116
CF = 0.4341 x ln(Hardness value) – 0.3926
AF = 0.4341 x ln(Alkalinity value) – 0.0074
The following is a general industry guideline for interpreting LSI values
66
Page 71
• An index between -0.5 and +0.5 indicates balanced water
• An index of more than +0.5 indicates scale-forming water
• An index below -0.5 indicates corrosive water
You can use the calculator to adjust any of the water balance
parameter variables used in the calculation to analyze the effect of the
change on water balance.
The LSI calculator function generates an accurate saturation index
value that indicates the state of the system at the time the calculation
is made.
C. Hardness Units
NOTE: The 9P MUST be in the HARDNESS FUNCTION for hardness
unit preference selection.
The hardness unit preference is set simultaneously for both functions
that capture a hardness value. That means when you set the unit
preference in Hardness Function, it is set for the LSI Calculator
function at the same time.
The LSI Calculator allows you to calculate with either ppm or grains as
hardness units.
The hardness unit conversion is based on the following equivalency:
17.1 ppm (mg/L) = 1 grain
XXV. SOFTWARE VERSION
Contact the Myron L® Company to see if a software upgrade is available.
1. Press any parameter key.
2. Press key until three numbers are displayed as shown
in Figure 35.
3. Press any parameter key; instrument will time out in 15
seconds. lving a mathematical problem.
67
Page 72
XXVI. GLOSSARY
Anions Negatively charged ions.
See Solution Characteristics, pg. 59.
AlgorithmA procedure for solving a mathematical problem.See Temperature Compensation (Tempco) and
TDS Derivation, pg. 61.
FAC Free Available Chlorine. The amount of chlorine that remains active in solution and is available for ongoing
disinfection. See Free Chlorine as an Indicator, pg. 65.
FCE FCE directly measures ORP, the germ killing
TM
property of chlorine and other oxidizing germicides.
It displays both the ORP reading (in mVDC) as well
as an equivalent free chlorine concentration (in
familiar ppm). For more information see FCE :
TM
Groundbreaking Measurement of Free Chlorine
Disinfecting Power in a Hand-Held Instrument on
the Myron L® Company website.
LogarithmAn arithmetic function. The inverse of an exponential
function. See pH Units, pg. 62.
ORPOxidation-Reduction Potential or REDOX, See ORP/ Oxidation-Reduction Potential/REDOX, pg. 64.
REDOX An abbreviation for Reduction-Oxidation reactions.
Reaction This is the basic electrochemical process by which
chlorine destroys microbes by grabbing electrons from
the microbe’s proteins, denaturing the protein and
killing the organism. ORP directly measures the
strength of a solutions’ REDOX potential and,
therefore, sanitizing strength.
TDSTotal Dissolved Solids or the Total Conductive Ions in a solution. See Conductivity Conversion to
Total Dissolved Solids (TDS), pg. 59.
TempcoTemperature CompensationSee Temperature Compensation of Aqueous
Solutions, pg. 56.
UserA mode of operation that allows the instrument user(operator) to set a tempco and/or a TDS factor for
their specic solution type. See Temperature Compensation of Aqueous Solutions, pg. 56
and Temperature Compensation (Tempco) and
TDS Derivation, pg. 59.
For details on specic areas of interest refer to the Table of Contents, pg. 5.
68
Page 73
69
Page 74
MYRON L® COMPANY
2450 Impala Drive
Carlsbad, CA 92010-7226
USA
Tel: +1-760-438-2021
Fax: +1-760-931-9189
E-Mail: info@myronl.com
techquestions@myronl.com
www.myronl.com
Made In USA
70
9POM 24JUL19
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.