Myron L PoolPro PS9 Quick Start Manual

PoolPro PS9 Quick Start Guide
Handheld Conductivity, Mineral / Salt Concentration, TDS, ORP, Hardness, Alkalinity, LSI, Free Chlorine, pH, and Temperature Meter
PS9QSI REV: 07-16
This Quick Start Guide is for the Myron L® Company Pool Pro Model PS9. Download the full PS9 Operation Manual (P/N PS9OM) from www.myronl.com to get more detailed instructions for taking measurements (including Alkalinity Hardness and LSI calculation), storing or recalling measurements, changing instrument settings, maintenance procedures, calibration procedures, Troubleshooting Guide and FACTORY CAL reset.
QUICK REFERENCE INSTRUCTIONS
Press and release any one of these buttons to turn on the PS9 and begin taking measurements.
Press to calculate LSI.
Press and release the CAL/MCLR button to calibrate (CAL) the active measurement parameter or press and hold to clear displayed memory location (MCLR).
Press and release to store (MS) a measurement or move upward on the list of stored measurements (Location 1 to 100) while in memory recall mode.
Press and release to view the most recent stored measurements or move downward on the list of stored measurements (Location 100 to 1). Also used to enter
Settings Mode.
PS9 FEATURES
1. TEMPERATURE SENSOR – Houses thermistor for making temperature measurements.
2. CONDUCTIVITY CELL – Contains electrodes that generate a flux field in defined area for all measurements except
pH, ORP, and FCE.
3. PH SENSOR (user replaceable) – Measures test sample’s hydrogen ion concentration. The reference junction is
located under the glass bulb.
4. pH SENSOR PROTECTIVE CAP – Snaps in place to protect pH Sensor when not in use and seals in Storage Solution.
5. LCD DISPLAY – Displays measurements, units of measure, active measurement mode, current solution mode setting, low battery warning (LOBATT), and other information (see below).
6. MEASUREMENT KEYS – Press and release any one of these to turn on the PS9 and begin taking measurements.
7. FUNCTION KEYS – Press and release to store the current measurement (MS), recall a stored measurement (MR). Also used to move between memory locations and to select various mode settings.
8. CAL/MEMORY CLEAR KEY – Use to calibrate (CAL) the active measurement parameter, enter or clear the
contents of the currently displayed memory location (MCLR).
9. BLUDOCK SYMBOL – Indicates that this PoolPro PS9 is equipped with Myron L® Company’s optional wireless transceiver for wireless data download.
10. CELL EXTENDER – Holds sample and reagents during Alkalinity and Hardness measurements.
11. PLUNGER – Used to create proper cell cup volume of liquid during Alkalinity and Hardness measurements.
12. CELL EXTENDER CAP – Seals cell extender while mixing sample and reagents during Alkalinity and Hardness
measurements.
LCD DISPLAY
1. MAIN VALUE DISPLAY – Displays the current measurement value and settings menu information. During Alkalinity & Hardness titrations, prompts appear here at
various stages of the process.
2. UNITS OF MEASURE ICONS – Appropriate icon appears to show the units of measure for the current measurement.
3. TEMPERATURE UNITS OF MEASURE – Displays °C (Celsius) or °F (Fahrenheit) based on setting selected for
Temperature readings.
4. MEASUREMENT ICONS – Appropriate icon appears to show what type of measurement is being made.
5. BUFFER ICON – Appears when the PS9 is in pH Calibration mode to indicate that the instrument is expecting a pH
buffer solution.
6. TEMPERATURE DISPLAY– Displays the temperature of the sample solution, memory location number (1 to 100) or buffer during pH calibration.
7. SOLUTION MODE ICONS – Appropriate icon appears to indicate the current solution temperature compensation mode setting for Conductivity, Min/Salt and TDS measurements.
8. LOBATT ICON – Appears when the PS9 ’s battery requires replacement.
9. CAL ICON – Appears when the PS9 is in Calibration mode.
10. MEMORY ICON – Appears to indicate that the values and icons being displayed are for a measurement stored in the PS9 ’s memory (NOT a live measurement).
OPERATING INSTRUCTIONS
DEFAULT SETTINGS:
Temperature Units: °C (Temperature is displayed in Degrees Celsius)
Solution Mode: Conductivity – KCl
MEASUREMENT SETUP: Before you take a reading, make sure the PS9 is
clean, calibrated, and if measuring COND or TDS, that the desired solution mode has been selected (See Section II of Operating Instructions, below). The sample solution must also be within the specified range.
I. MEASUREMENT
1. Rinse and empty the appropriate sensor (conductivity or pH / ORP) 3 times with sample to be measured.
2. Fill with test sample.
3. Press appropriate Measurement Key:
COND to measure CONDUCTIVITY 0-9999 MICROSIEMENS (µS/cm), 10-200 (mS)/cm
TDS: Measures TOTAL DISSOLVED SOLIDS 0-9999 parts per million (ppm), 10-200 (ppt)
MIN / SALT: Same as TDS but displays in 0 - 200 parts per thousand (ppt). displays as PPT of NaCl
ORP: Measures Oxidation Reduction Potential (REDOX) ± 999 mV
pH: Measures 0-14 pH
Press one of these to preform an Alkalinity or Hardness measurement.
Press to calculate LSI. Uses stored or hypothetical Alkalinity and Hardness data.
For complete instructions on performing these measurements, download the full PoolPro PS9 Operation Manual and watch our YouTube videos (Search for: Ultrameter III 9P and 9PTK).
PoolPro PS9 Quick Start Guide
Handheld Conductivity, Mineral / Salt Concentration, TDS, ORP, Hardness, Alkalinity, LSI, Free Chlorine, pH, and Temperature Meter
MYRON L® COMPANY
2450 Impala Drive u Carlsbad, CA 92010-7226 u Phone: +1-760-438-2021
E-Mail: Customer Service – info@myronl.co m u Technical Support – techquestions@myro nl.com
Website: www.myronl.com PS9QSI Rev:07-16
NOTES:
When finished making Conductivity, TDS, MIN/SALT ALK or HARD measurements, rinse the cell cup with clean water (preferably DI, RO, or distilled).
When finished measuring pH, ORP or FCE, rinse sensor well with clean water (preferably DI, RO, or Distilled), refill it with Myron L Storage Solution then reinstall protective cap.
II. Solution Mode Selection
1. Press , or , to select which parameter is having its solution
type changed.
2. Press and hold about 3 seconds or until "SEL" appears on the display.
3. Use the or key to select type of solution desired. The
possible solution types are KCI, NaCl or 442™(a Myron L® Company natural water).
4. Press to accept new solution type.
CALIBRATION
CALIBRATION SOLUTIONS
Mode
Standard Solution or Buffer
Conductivity KCl
KCl: 7000 µS
TDS 442
442: 3000 PPM
Displays as 2027 ppm in TDS NaCl or MIN/SALT mode
TDS NACL & MIN/SALT
pH and FCE
4.0, 7.0 and 10.0 buffer
Alkalinity
100 ppm Alkalinity Standard Solution
Hardness
200 ppm Hardness Standard Solution
FCE ORP calibration is electronic and based on pH Cal Results
I. CONDUCTIVITY, MIN /SALT OR TDS CALIBRATION
1. Rinse cell cup 3 times with proper standard solution.
2. Refill cell cup with fresh, standard solution.
3. Press , or , then press , "CAL" will appear.
4. Press or until display agrees with standard solution.
5. Press to accept value.
NOTE: A deviation of more than 10% from standard solution value will
produce a "FAC" on the display. Press to accept factory calibration, or clean the cell and recalibrate.
ALKALINITY and HARDNESS CALIBRATION
For complete instructions on performing these measurements, download the full PoolPro PS9 Operation Manual.
III. pH/ CALIBRATION
1. Rinse sensor well 3 times with 7.0 buffer solution.
2. Refill sensor well with 7.0 buffer solution.
3. Press , then , "CAL", "7" and "BUFFER" will appear on the
display.
4. Press or until display agrees with buffer value.
5. Press once to accept the value.
6. Rinse 3 times with either ACID (Acd) (pH 1- 6) or BASE (bAS) (pH 8-14)
buffer solution.
7. Refill and repeat steps 4 & 5.
8. Rinse 3 times with opposite buffer solution, refill and repeat steps
4 & 5 or press to exit.
9. When calibration is complete, refill sensor well with Myron L Storage Solution or a high KCI solution and reinstall cap.
IV. ORP/ FCE CALIBRATION
ORP electrodes rarely give false readings without problems in the reference
electrode.
For this reason, and because calibration solutions for ORP are highly
reactive and potentially hazardous, your PoolPro has an electronic ORP calibration.
This causes the zero point on the reference electrode to be set whenever
pH 7 calibration is done.
MAINTENANCE
I. BATTERY REPLACEMENT (LOBATT)
1. Clean and dry Instrument THOROUGHLY.
WARNING: ONLY open PS9 in a clean, dry environment!
2. Remove the four (4) bottom screws.
3. Open instrument CAREFULLY.
4. Carefully detach battery from circuit board.
5. Replace with 9-volt alkaline battery.
6. Replace bottom, ensuring the sealing gasket is installed in the groove of
the top half of case.
7. Re-install the 4 screws, tightening them evenly and securely.
II. CAUTIONS
Solutions in excess of 71°C/160°F should not be placed in the cell cup area;
this may cause damage.
The pH sensor may fracture if the PS9 temperature is allowed to go below
0°C/32°F.
The Conductivity cell cup should be kept as clean as possible. Flushing with
clean water following use will prevent buildup on electrodes. If very dirty samples (particularly scaling types) are allowed to dry in the cell cup, a film will form reducing accuracy.
The pH sensor in your PS9 should not be allowed to dry out.
Whenever the pH sensor is not in use make sure it is filled with
Myron L® Company Storage Solution and that the protective cap is in place.
III. CLEANING SENSORS
A. Conductivity, TDS, MIN/SALT, ALK, HARD, and LSI
When there are visible films of oil, dirt, or scale in the cell cup or on the electrodes or if readings are not as expected, use isopropyl alcohol or a foaming non-abrasive household cleaner.
Rinse out the cleaner, and your PS9 is ready for accurate calibration then measurements.
B. pH: If the pH sensor dries out or becomes dirty, it may be cleaned and/or
reconditioned. For instructions on preforming these operations, download the full PS9 Operation Manual from the Myron L® Company website.
IV. pH Sensor Replacement
Order model RPR. Be sure to include the model and serial number of your instrument to ensure receipt of the proper type. NOTE: Complete installation instructions are provided with each replacement sensor.
SPECIFICATIONS
Spec
pH
CONDUCTIVITY
TDS & MIN / SALT
ORP
ALKALINITY
HARDNESS
LSI
FCE
TEMP
Ranges
0-14 pH
0 - 9999 µS/cm; 10 - 200 mS
In 5 autoranges
0 - 9999 ppm; 10 - 200 ppt
In 5 autoranges
± 999
mV
10-800ppm
0-1710 ppm
(0-100 grains)
-10 to +10
0.00 – 9.99 ppm
ORP = 350mV - 725 mV, <9.9 pH
ORP = 725 mV - 825 mV, <8.9 pH
0°C – 71°C
32°F – 160°F
Resolution
0.01 pH
0.01 < 100 µS
0.1 (<1000µS)
1.0 (<10 mS
0.01 < 100 mS)
0.1 mS ( 100 & < 200 mS)
0.01 < 100 ppm
0.1 (<1000 ppm)
1.0 (<10 ppt
0.01 < 100 ppt)
0.1 (≥ 100 & < 200 ppt)
±.1 mV
0.1(<100 ppm) 1(<800 ppm)
0.1(<100 ppm)
1(<1710 ppm
0.1
0.01 ppm
0.1 °C / °F
Accuracy
± 0.01 pH
±1% of reading
±.1 mV
±. 0.3 ppm <1.0 ppm ±. 0.2 ppm 1.0 ppm
±0.1 °C / °F
COND/TDS
Ratios
Programmed: KCl, NaCl or 442™ Adjustable 0.20 -07.99
Temp Co
Auto: 0-71°C, 32-160°F Adjustable (COND & TDS) 0 – 9.99% / °C
!
Loading...