DESCRIPTION
This manual describes your Myron L pDS meter , tells you how to
use it, and how to keep it working accurately for many years.
Myron L pDS Meters are compact, light-weight, self-contained conductivity/pH meters. They contain a built-in conductivity cell with a replaceable pH sensor. Models EP11/pH and M6/pH have dials calibrated in micromhos (equivalent to the metric microsiemens) and in pH units. They will quickly determine the conductivity and pH of almost any solution.
Both are 3.4" x 4.5" x 4.0" (85 x 129 x 126mm) and weigh less than one pound (0.45 kg). pDS Meters are completely self-contained. The built-in conductivity cell is automatically temperature compensated from 50° to 160°F (10° to 71°C). They are powered by two 9 volt transistor batteries good for at least 2000 tests, or one year shelf life.
MODEL CONDUCTIVITY RANGES pH RANGE
EP11/pH 0-10, 0-100, 0-1000,
0-10,000 micromhos (µmhos) 2-12 pH
M6/pH 0-5000 micromhos (µmhos) 2-12 pH ACCURACY: (Both Models)
Conductivity: ±2% of full scale
pH: ±0.2 pH units
REPEATABILITY: (Both Models) ±1%
By using an RE-10 Range Extender (seeACCESSORIES), the maximum conductivity range of each meter may be increased ten times.
The pictures in this manual show the major operating parts of your Myron L DS Meter. Handle your DS Meter and identify these parts to become familiar with it.
IMPORTANT: The pH sensor in your pDS Meter has a limited life much like a battery. Sensor longevity is af fected both by the types of solution tested and care given your meter. pH sensors are user replaceable. See MAINTENANCE for additional information.