If you read nothing else in this
manual please read this
Quick Reference Guide.
PLEASE READ and COMPREHEND ALL
WARNINGS, CAUTIONS and ADVISEMENTS
CONTAINED WITHIN THIS MANUAL.
Failure to comply is beyond the responsibility of
the Myron L Company.
WARNING: ALL MONITOR/CONTROLLERS ARE
FACTORY SET TO OPERATE ON 115 VAC. BEFORE
APPLYING POWER ENSURE THE INPUT POWER
“115/230 VAC” SELECTION IS CORRECT FOR YOUR
REQUIREMENTS. FAILURE TO DO SO IS BEYOND
THE RESPONSIBILITY OF THE MYRON L COMPANY.
See section II.E.2. and figure II.E.1.
WARNING: ENSURE POWER IS OFF WHILE
INSTALLING ELECTRICAL EQUIPMENT. IF
MONITOR/ CONTROLLER IS INSTALLED, ENSURE
THE POWER IS OFF BEFORE SERVICING. FAILURE
TO DO SO COULD CAUSE DAMAGE TO THE
INSTRUMENT, AND COULD BE HARMFUL OR FATAL
TO PERSONNEL. ONLY QUALIFIED PERSONNEL
SHOULD INSTALL OR SERVICE ELECTRICAL
EQUIPMENT.
WARNING: THE DISPLAY WILL BE IRREPARABLY
DAMAGED IF THE DISPLAY HARNESS IS INSTALLED
UPSIDE-DOWN OR MISALIGNED. THE HARNESS
MUST BE INSTALLED AS SHOWN IN FIGURE II.E.5.
CAUTIONS:
Before installation, ensure you have the correct
model (with options), AND it is ranged for your
application. See sections I.A., I.B. & I.G.
Do you have the correct sensor? See section I.E.
Mounting requirements. What is needed? See
section II.B.
The following will give the installer and user a quick
overview. See the sections listed for details.
REMOVING FRONT PANEL
NOTE: When opening instrument, remove front cover with care;
a ribbon cable connects the front panel and main board.
1. Ensure power is OFF.
2. Remove the two (2) screws on the front panel.
3. Carefully wiggle the front panel and pull gently toward
you. Do not pull more than about 8 inches/20CM or you
could damage the wiring harness.
REASSEMBLY
1.Carefully reinstall the front panel, bottom first.
Ensure no wires have been pinched between enclosure
and front panel.
2.Reinstall the two (2) screws and tighten.
3.To operate, turn power ON.
INTRODUCTION - Section I.
This section covers the specifications of your new
AQUASWITCH including sensor information.
INSTALLATION - Section II.
This section covers how to install your new AQUASWITCH;
mechanically and electrically.
OPTIONS & ACCESSORIES - Section III.
This section covers the specifications, installation, set up, and
operation of each option.
QUICK LOCATOR
-4A/4AO MODULE (4-20mA), see section III.A.
-PA/PAO (Piezo Alarm), see section III.B.
-TP/TPO (Temperature Module, see section III.C.
RA (Remote Alarm), see section III.D.
OPERATING PROCEDURES - Section IV.
This section covers a brief description of the models and their
features; how they work, and how to set them up for your
particular use.
QUICK SET POINT CONVERSION (SPC) /
REVERSING SET POINT - See Section IV.B.1.
Resistivity AQUASWITCH II‘s are configured to trigger the
alarm relay as the resistivity reading decreases.
Conductivity/TDS AQUASWITCH II’s are configured to trigger
the alarm relay as the conductivity/TDS reading increases.
To reverse:
1. Locate the jumper block for the alarm to be configured.
See figure V.A.1.
2.Remove and rotate the jumpers 1/4 turn and reinstall
them on their posts.
QUICK CHECK-OUT PROCEDURE -
See Section IV.C.
It is assumed that the AQUASWITCH power is ON, that it is
connected to an appropriate Sensor, and that the Sensor is
immersed in water within the range that the AQUASWITCH II
will be required to control; and the front panel is removed.
1. Make a note of the reading on the display.
2. While pressing the Calibration/Full Scale Test Switch (FS
SW), verify that the front panel display is indicating a full
scale reading. If not, see Calibration, section V.C.
3. Press and hold the "SET POINT" switch on the front
panel. Using a tweaker or a small screwdriver, adjust the
Set Point trimmer adjustment screw on the circuit board
to sweep the display from zero to full scale. (A digital
display may be blank at the full scale end. This is
normal.) Listen for the alarm relay to click on and off as
the alarm set point moves past the water reading.
4. Adjust the alarm to the desired set point value. Release
the "SET POINT" switch.
Continued
QUICK SET POINT ADJUSTMENT -
See Section IV.B.2.
The set point setting is based upon the user's particular water
purity specifications or requirements.
1. While pressing the "SET POINT" switch, turn the Set
Point #1 adjustment screw (see figure V.A.1) until the
desired set point value is indicated on the display.
HYSTERESIS (DEAD BAND) ADJUSTMENT -
See Section IV.B.3.
ELECTRONIC CALIBRATION (CIRCUIT ONLY) -
See Section V.D.1.
Full Scale Calibration V.D.1.a.
1. Press and hold the Full Scale Test switch. The display
should indicate Full Scale for the particular range
selected, i.e. 0-20 MΩ should indicate 20. If not, set to
Full Scale with the CALibration control.
2. Turn power OFF.
3. Re-install front panel as described in “REASSEMBLY”.
4. To operate, turn power ON.
PRIMARY COMPONENT IDENTIFICATION -
Section V.A.
Review the figure below to familiarize yourself with the different
circuit boards and component locations. The diagram has the
second alarm/control module option installed.
“PURGE CYCLE” CALIBRATION PROCEDURE
Section V.B.
“ALARM RESET” MODE CHANGE PROCEDURE
Section V.C.
QUICK CALIBRATION - Section V.D.
WARNING: When performing calibration procedures,
the technician must take extreme care to avoid
contacting the circuitry other than the CALibration
control. Failure to do so could result in damage to
the equipment, property and/or personal injury.
The following assumes the front panel has been
removed and the power is ON.
AQUASWITCH II
AQUASWITCH II MAIN MONITOR/CONTROLLER BOARD
BK WT RD GN NU R- R+
RANGE
MODULE
3S
}
UP
-11
0-20mΩ
751 756
752 757
753 758
754 759
MYRON L
COMPANY
TRANSFORMER
115/
230
FUSE*
CALIBRATION
CONTROL
CAL
PWR C GD NC NO CM
HYS1SP1FS SW
SPC
INC
DIS
DEC
REMOVE TO
INSTALL
SECOND
RELAY
AQUASWITCH II DIGITAL DISPLAY
DISPLAY
DISPLAY CONNECTOR
DISPLAY
HARNESS
DISPLAY
HARNESS
CONNECTOR
DISPLAY
CALIBRATION
CONTROL
4-20
CONNECTOR
E
N
C
L
O
S
U
R
E
F
R
O
N
T
P
RED
A
N
TO MONITOR/
CONTROLLER
E
CIRCUIT BOARD
DISPLAY
CONNECTOR
LLOW
SET POINT
SWITCH
GREEN
10VDC Calibration - See Section V.D.1.b.
USING STANDARD SOLUTIONS - See Section V.D.2.
The BEST method of verifying and recalibrating your
conductivity/TDS AQUASWITCH II is with NIST traceable
Standard Solution (available from the Myron L Company).
Because it includes the sensor, the entire system is recalibrated.
NOTE: Since standard solution calibrations are NOT practicable
with resistivity models, another means of verification or
calibration of resistivity models is to use the transfer standard
method, using a hand-held or portable instrument capable of
resistivity measurements, such as the Myron L Ultrameter™. See
section V.C.4 for description.
Thank you for selecting one of the Myron L Company’s newest
AQUASWITCH Monitor/controllers. The new AQUASWITCH™
series is based on input from ‘you’ - our customers, time proven
designs, and many years of instrumentation experience.
Since 1957, the Myron L Company has been providing customers
with quality products at an affordable price by designing and
producing products that are Accurate, Reliable, Simple to use.
Quality you have come to rely and depend on.
As you read through this operation manual you will see the new
AQUASWITCH is truly designed to be user friendly with simple
to install options, and easy re-rangeability as conditions or
applications change. This manual is actually more complex than
the AQUASWITCH, but must be to address all the variables.
A. SCOPE
This operation manual provides the user with the necessary
information to install, operate and maintain the
Myron L Company's latest AQUASWITCH Resistivity &
Conductivity/TDS Controllers.
Section I. Descriptions, Applications, Specifications.
Section II. Installation; mounting, wiring and set up.
Section III. Options and Accessory installation procedures.
Section IV. Operating procedures.
Section V. Identifies their primary components and provides the
user with easy-to-use calibration and preventive care
procedures.
Section VI. Options & Accessories List.
Section VII. Replacement Components.
Section VIII. Warranty information.
Section IX. Glossary, definitions.
Section X. Notes.
Section XI. Addendum.
1. FUNCTIONAL DESCRIPTIONS
Both models have water & corrosion resistant IP64/NEMA 3
housings suitable for panel, bench or surface mounting.
The AQUASWITCH I is packaged within a compact 6.0" x 4.8"
(152mm x 122mm) housing. The AQUASWITCH II is packaged
within a 6.0" x 10.8" (152mm x 274mm) housing.
The AQUASWITCH I does not monitor water quality itself, but
should be used with any reliable resistivity or conductivity/TDS
controller such as the Myron L Company’s 750 Series II
Resistivity or Conductivity/TDS Monitor/controllers.
The AQUASWITCH II is an integrated unit requiring no other
Monitor/controller.
NOTE: All specifications for the AQUASWITCH I also apply to
AQUASWITCH II unless otherwise indicated. Bright green/red
LEDs indicate HIGH/LOW set point readings. All AQUASWITCHII’s feature a heavy-duty 10 amp output relay, operating on either
increasing or decreasing readings. May use the optional 4-20mA
output for PLC or SCADA operations.
2. APPLICATIONS
Conductivity/TDS
Reverse Osmosis
Process Control
Seawater Desalinization
Food Processing
Plating
Power Plants
Laboratories
Printing
Are just a few of the applications
Resistivity
Deionization (DI) and Distillation Ultrapure Water
Treatment Systems
Electronics
Pharmaceutical
Laboratories
Food Processing
Plating
Power Plants
Are just a few of the applications
B. SPECIFICATIONS
1. GENERAL ASI & ASII
INPUT POWER:
115/230 VAC ±15%, 50/60 Hz (User selectable)
overvoltage category II
100 mA Maximum Current
Double Insulated (with circuit board ground for OEM operation)
Fuse - 100mA Slow Blow (T.10A) for both 115 & 230 VAC (V~)
User replaceable
Humidity - 20-90% non-condensing
Max. Altitude -
40,000 ft/12,000 meters non-operating
10,000 ft/3000 meters operating
Pollution degree 2
24 VAC or 24 VDC Option available on Special Order
Overvoltage category II (24 VAC)
250 mA Maximum Current
Double Insulated (with circuit board ground for OEM operation)
Fuse - 250mA Slow Blow (T.25A) User replaceable
AQUASWITCHI:
Provides automatic switching from exhausted to fresh Dl banks;
push-button override for immediate bank change-over; adjustable
purge timer for “FAILSAFE” operation; front panel LED displays
provide continual status of both Dl banks; and an alarm is
automatically activated as each bank is depleted.
AQUASWITCH II:
AQUASWITCH II combines the operating features of the
AQUASWITCH I with the additional features of a
Monitor/controller. These include: a 3 1/2 digit Liquid Crystal
Display and a front panel accessible “SET POINT CHECK” switch.
(Set point setting is internal to discourage unauthorized
adjustments).
AMBIENT TEMPERATURE RANGE
32°F ( 0°C) to 131°F (55°C)
OUTPUT POWER:
24 VAC (Standard if ordered with Myron L valves)
115 VAC (Standard if ordered without Myron L valves)
230 VAC (Optional)
MAXIMUM LOAD PER VALVE OUTPUT:
200 milliamps;
(5 watts for a 24 V valve)
(25 watts for a 115 V valve)
(50 watts for a 230 V valve)
4
AMBIENT TEMPERATURE RANGE:
32°F (0°C) to 140°F (60°C)
HOUSING CONSTRUCTION
Fully gasketed heavy-duty ABS for splashproof and corrosion
-PC115 VAC Powercord (8 ft. with USA plug and strain
relief) - NOT for use with 230 VAC.
-24VA24 VAC isolated power supply (special order)
-24VD24 VDC isolated power supply (special order)
-PTSPanel mounted Full Scale Test Switch ((ASII special order)
D.ACCESSORIES (ordered separately)
4AO4-20 mA Isolated output
PAO70 db Piezo Electric Alarm ONLY (ASII)
35BLO 3 1/2 digit backlit LCD (ASII)
PCO115 VAC Powercord (8 ft. with USA plug and strain
0-10 VDC @ 5mA max. (linear); standard on all models
OUTPUT IMPEDANCE
100Ω ±5%
SENSOR INPUT
Resistivity — 1 CS10 sensor
Conductivity/TDS — 1 CS51LC sensor up to 0-10
1 CS51 sensor above 0-10
CONTROL FUNCTION
Models ASIIC & ASIIR:
Single set point alarm/control continuously adjustable
5
E. SENSORS
All sensors feature integral thermistor style temperature sensors
to ensure accurate, rapid and reliable automatic temperature
compensation.
1. CONDUCTIVITY/TDS
AQUASWITCHII Conductivity/TDS Monitor/controllers use the
CS51LC or CS51 Series sensor depending on range. The 1.0 cell
constant CS51 model is recommended for ranges of 0-10 through
100 µS/ppm. Its compact size allows mounting in a standard 3/4"
tee. The sturdy polypropylene bushing is modular for easy,
inexpensive replacement. Other models available by special
order.
CS51LC sensor has a cell constant of 0.1 and is used for
conductivity/TDS values of 0-1 to 0-5 µS/ppm.
2. RESISTIVITY
AQUASWITCHII Resistivity Monitor/controllers use the CS10
sensor. It has a cell constant of 0.05 and is used for ultra pure
water applications. Other models available by special order.
Special order High Temperature, High Pressure sensors; and Low
Cost, Low Temperature, Low Pressure sensors are available.
For detailed descriptions of these and other sensors, see Sensor
Selection Guide and specific sensor data sheets available from
your local distributor, the Myron L Company, or on line at
www.myronl.com.
3. SENSOR SPECIFICATIONS
STANDARD MODELS — 316 Stainless SteelCONSTANT
CS10:For ALL Resistivity ranges0.05
CS51 LC: For ranges 0-5 µS/ppm & below 0.1
CS51:For ranges between 0-10 & 0- 100 µS/ppm1.0
See Range Selection Guide, Page 8.
SPECIAL ORDER MODELS — 316 Stainless Steel
CS40Resistivity/Conductivity/TDS
Valve Insertable 100PSI@150°C
(1.0, 0.1 & 0.05)
CS40HTAbove High Temperature Model 250PSI@205°C
CS41Cond/TDS High Temperature Model 100PSI@150°C
CS41HTCond/TDS High Temperature Model 250PSI@205°C
CS50Conductivity/TDS 100PSI@95°C (1.0 & 0.1)
CSALow Cost Conductivity/TDS - NO Temperature
Compensation 75PSI@60°C (1.0)
CSATCLow Cost Conductivity/TDS - with Temperature
Compensation 75PSI@60°C (1.0)
4. SENSOR OPTIONS
-TTitanium - in place of Stainless Steel
-2525’ Shielded Cable
-100100’ Shielded Cable
-PV 1/2" PVDF fitting for CS10 & CS51 ONLY
{replaces polypropylene}
-HPSS 1/2" 316 STAINLESS STEEL fitting
for CS10 & CS51 ONLY (replaces
polypropylene).
-STFSanitec Fitting for CS10, CS51 & CS51LC. 1/2”
thru 4”. State size, i.e. STF-1/2.
Special Order Options
-WTVWet-Tap Valve for CS40 (0.1 & 1.0 only)
-JBJunction Box - Class I, Group D, Div. S,
Explosion proof, Weather proof, Aluminum. For
CS40 ONLY.
TEMPERATURE COMPENSATION
Automatic to 25°C, between 32-212°F (0-100°C)
except high temperature models - up to 205°C.
PRESSURE/TEMPERATURE LIMITS
CS10 & CS51 - 100 psi (689.6 kPa) at 212°F (100°C)
For higher limits, see specifications below.
*CS51LC conductivity sensor (0. 1 constant) required for these ranges.
Conductivity/TDS AQUASWITCH II Controllers require a sensor with a cell constant of 1.0. The Model CS51 is the
most commonly selected sensor due to its ease of use and low cost.
Resistivity AQUASWITCH II Controllers require a sensor with a cell constant of 0.05. The Model CS10 is the most
common selected due to its ease of use and low cost.
NOTE: UNLESS ppm/NaCl is specified with order, TDS/ppm type conductivity instruments will be calibrated to the
Myron L "442™" Natural Water standard. For NaCl ppm/ppt add the letter "N" after the number, i.e. -114N.
8
II.INSTALLATION
A.GENERAL
This section provides the recommended procedures for properly
installing the AQUASWITCH I Controller, or AQUASWITCHII Monitor/controller, Valve(s), and Sensor. Figure II.A.1 gives
an overview of the entire system. Alternate configurations on
page 39.
rotated or mounted upside down so that the cutouts are
on the opposite side.
4.AQUASWITCH II
For best results, position your Monitor/controller and
sensor as close as possible to the point(s) being
controlled. The AQUASWITCH II Resistivity &
Conductivity/TDS Monitor/controllers are not designed to
operate with a sensor cable length that exceeds 100' (30
meters).
AQUASWITCH II
SOURCE
WATER
MONITOR/CONTROLLER
MEGOHMS - CM
AQUASWITCH II
BANK A
DIDIDI
DIDIDI
BANK B
BANK A
PILOT VALVE
NC
NC
BANK B
PILOT VALVE
BLEED
PILOT VALVE
(NC)
WASTE WATER
750II
WASTE
WATER
AQUASWITCH I/II SYSTEMS DIAGRAM
(SEE PAGE 39 FOR ALTERNATIVE CONFIGURATIONS)
AQUASWITCH I
AQUASWITCH I
RESISTIVITY
SENSOR
PROCESS
WATER
OPTIONAL
PROCESS
PILOT VALVE
(NO)
VAC
ALARM
RELAY
Figure II.A.1.
CAUTION - READ FOLLOWING CAREFULLY
!
WARNING: THE MYRON L COMPANY RECOMMENDS
THAT ALL MOUNTING AND ELECTRICAL
INSTALLATIONS BE PERFORMED BY QUALIFIED
PERSONNEL ONLY. FAILURE TO DO SO COULD
CAUSE DAMAGE TO INSTRUMENT, AND COULD BE
HARMFUL OR FATAL TO PERSONNEL.
B.MECHANICAL INSTALLATION
All AQUASWITCH electronics are packaged inside
drip/weather-proof housings. The physical dimensions of the
housing is suitable for panel, bench or surface mounting.
There are four basic guidelines to consider when selecting a
mounting location:
1.Select a site that limits the AQUASWITCH‘s exposure
to excessive moisture and corrosive fumes.
2.If at all possible, mount the AQUASWITCH at eye level
for viewing convenience.
3.AQUASWITCH I
If needed, the AQUASWITCH I enclosure may be
1. SURFACE MOUNTING WITH SMP
NOTE: A Surface Mounting Plate (AQUASWITCH I - #SMP50)
or (AQUASWITCH II - #SMP60) may be required when access
to the back side of the mounting site is impractical or if the
AQUASWITCH must be mounted on a solid wall. Both SMP’s
come with the proper hardware to mount the AQUASWITCH to
the SMP, however, the installer must provide the four (4)
additional screws/bolts to mount the SMP to the wall or fixture.
Their type and size is to be determined by the user.
1. Select your mounting location. Mark and drill the four (4)
required mounting holes. For hole locations, use the SMP
as a template.
2. Drill the corner holes in the SMP according to the size of
the screws or bolts selected.
3. Attach and securely fasten the SMP to the Monitor using
the 1/4" X 20 X 3/8" screws provided.
4. Mount the SMP to the prepared site using the selected
screws or bolts.
2. SURFACE MOUNTING WITHOUT SMP
NOTE: Surface mounting will require two (2) 1/4 " X 20 screws of
a length equal to the thickness of the mounting site plus 3/8"
9
1. Select mounting site location. Mark and drill the required
mounting holes. For hole drilling locations, see figure
II.B.1.
2. Insert the 1/4" X 20 screws into the holes from the side
opposite the mounting site.
3. Hold the Monitor in place while starting and tightening the
mounting screws.
AQUASWITCH I
4.80
(122)
Face Plate Gasket
Panel Gasket
0.88 DIA
(22)
1.13 DIA
(29)
3. PANEL MOUNTING
A panel mounting fastening kit is provided with all
AQUASWITCH’s. Panel mounting will require the use of the
fastening kit's two (2) 4-40 mounting screws/nuts or two (2) #4 x
1/2" sheet metal screws. See figures II.B.1 & II.B.2. for panel
cutout dimensions.
1. Select your mounting location. Mark the appropriate
panel cutout and complete the necessary panel cut.
0.31 DIA, X2
(8)
0.53
(14)
0.13 RAD MAX, X4
(3)
3.96
(101)
6.00
(152)
AQUASWITCH I
6-32x3/8" x4
Face Plate Gasket
Panel Gasket
0.63 THK
3.89
(16)
(99)
AQUASWITCH II
10.81
(275)
0.50 DIA
(13)
0.5
(13)
Figure II.B.2
3.78
(96)
6.00
(152)
0.60 DIA
(15)
4.94
(126)
5.00
SURFACE
MOUNT
2.84
(72)
(127)
PANEL
CUTOUT
1.05
(27)
3.10
(79)
0.40
(10)
4.17
(106)
NOT TO SCALE
DIMENSIONS IN INCHES
(MILLIMETERS)
SURFACE AND PANEL MOUNTING DIAGRAMS
9.85
0.63 THK
(1.6)
5.00
(127)
0.50
(13)
0.75
(19)
5.00
(127)
3.75
(95)
3.77
(96)
0.31 DIA 2PL
(8)
SURFACE
MOUNT
(250)
0.113
(3)
2.54
(64)
0.11
(3)
2.00
(51)
AQUASWITCH II
0.50 DIA
(13)
9.85
(250)
Figure II.B.3
0.75
(19)
01.03 DIA 2 PL
(29)
0.68 DIA 2 PL
(22)
0.60 DIA
(15)
3.90
(99)
5.10
(129)
PANEL
CUTOUT
NOT TO SCALE
DIMENSIONS IN
INCHES
(MILLIMETERS)
DOUBLE INSULATED
SURFACE AND PANEL MOUNTING DIAGRAMS
9.92
(252)
10.19
(259)
0.12 RAD MAX 4PL
(3)
0.13 2PL
(3)
2.57
(65)
0.13
(3)
10
2. Carefully unfasten and separate the AQUASWITCH's
PILOT
VALVE
Figure II.D.2
!
front panel from its enclosure.
3. Disconnect all panel cable(s)/wires from the
AQUASWITCH's Control board.
4. Slide the enclosure through the panel cutout
until its flange contacts the panel.
5. Insert mounting screws through the flange
mounting holes and tightly secure.
6. Reconnect all panel cable(s)/wires and reinstall the front
panel.
C. SOLENOID VALVE INSTALLATION
There are many types of solenoid valves. The user must decide
which type is best suited for the specific application. The
following is an example installation of one type of valve,
1. Insert the pilot valve unit so that the systems’ INFLOW
source and OUTFLOW process piping connecting ends
are inserted into the valve as shown in figure II.C.1.
2. Repeat step until all appropriate pilot valve units have
been installed.
NOTE: The number of pilot valves used will be based upon the
user’s desired application.
24VAC
INTERCONNECTING
MANUAL OVERRIDE
ADJUSTABLE
FLOW CONTROL
OPERATES
NORMALLY
CLOSED
WIRES
OPERATES
NORMALLY
OPEN
SECURING NUT
FLANGE
CABLE
SS
WASHER
"T" FITTING
PLASTIC
WASHER
O-RING
THREADED FITTING
3/4" FNPT
OUT
IN
INSERTION MODE ASSEMBLY
Figure II.D.1.
2. IMMERSION OR DIP SENSOR ASSEMBLY
Use approved sealant, i.e. Teflon tape as required.
1. Verify that the Sensor’s Fitting assembly is properly
assembled as shown in figure II.D.2.
2. Insert and pull the Sensor’s cable through the extension
tube and then tightly attach extension tube to Sensor
assembly as shown in figure II.D.2.
PLASTIC
WASHER
O-RING
THREADED FITTING
3/4" FNPT
FLANGE
INFLOWOUTFLOW
OPTIONAL INFLOW
SOLENOID VALVE MOUNTING DIAGRAM
D.SENSOR INSERTION/IMMERSION MOUNTING
The SENSOR’s mounting orientation must provide a continuous
and adequate circulation flow to prevent the trapping of air
bubbles within the Sensor’s electrode area (CS10 shown in
figures II.D.1 & II.D.2). Failure to do so will result in conditions
that will prevent the Sensor from functioning properly.
1. INSERTION MODE (in-line installation)
Use approved sealant, i.e. Teflon tape as required.
1.Verify that the Sensor’s Fitting assembly is properly
assembled as shown in figure II.D.1.
2.Insert the Sensor Fitting assembly into the "T" fitting with
electrode aligned as shown in figure II.D.1. and tightly
SENSOR TIP
COUPLING
3/4" NPT
CABLE
EXTENSION TUBE
IMMERSION OR DIP SENSOR ASSEMBLY
E.ELECTRICAL INSTALLATION
The AQUASWITCH I and AQUASWITCH II require the user
to follow separate electrical installation procedures. In addition,
based on the user’s solenoid voltage specifications, the user will
be required to connect Solenoid Power.
NOTE: All cable watertight restraints are user supplied.
WARNING !
11
A device to disconnect the AQUASWITCH from the
power supply is required. It is recommended that
this switch or circuit breaker be labeled as the
disconnection device for the AQUASWITCH.
1. AQUASWITCH I MAIN INPUT POWER
WARNING: All AQUASWITCH I’s are factory set for
115 VAC. Before starting, ensure the input power
“115/230” selection is correct for your requirements.
Failure to do so is beyond the responsibility of the
Myron L Company. See section II.E.3. and figure
II.E.1 and II.E.2.
NOTE: Some models may have either a 24 VAC or a
24 VDC input power requirement - check labels
carefully.
1. Verify that the main AC power source is turned "OFF" or
disconnected.
2. Using a standard slot screwdriver remove the two (2)
screws on the front panel.
3. Carefully wiggle the front panel to loosen the gasket and
pull gently toward you. Do not pull more than about 8
inches/20CM or you could damage the wiring harness.
4. Turn the front panel around so that the back side is
facing you and set aside for now.
5. Using the enclosure cutouts, install the proper wire and
watertight cable restraint (not provided) to comply with
local electrical codes.
6. Neatly connect wires to the AQUASWITCH I's
connectors*, as shown in figures II.E.1.
*CAUTION: The input power connectors require only a small
screwdriver, or a pen to push on the release levers†. The release
levers may be broken or damaged if not pushed straight toward
the circuit board. DO NOT push the release levers sideways.
† Early versions have screw terminals.
AQUASWITCH™
INPUT FROM
CONTROLLER
RELAY
ALARM
RELAY
{
SOLENOID
VALVE
OUTPUT
POWER
POWER BOARD
LOGIC
BOARD
CONNECTION
TRANSFORMER
NCNOCOM
GRDCAC
{
BK B
BK A
BLD
NO
COM
NC
BANK B
BANK A
BLEED
COM
ASI ELECTRICAL CONNECTION DIAGRAM
115/230
JUMPERS
LINE / +DC
(BLK)
NEUTRAL / -DC
(WHT)
WHT
BLK
COM
LINE
NEUTRAL
OR
115/230
SWITCH
FUSE*
}
FUSE
1/8 Amp
AGC 1/8
MAIN
INPUT
POWER
}
SOLENOID
INPUT
POWER
Figure II.E.1
2. AQUASWITCH II MAIN INPUT POWER
WARNING: All AQUASWITCH II’s are factory set for
115 VAC. Before starting, ensure the input power
“115/230” selection is correct for your requirements.
Failure to do so is beyond the responsibility of the
Myron L Company. See section II.E.3. and figure
II.E.2.
MONITOR/CONTROLLER
MAIN CIRCUIT BOARD
MAIN
INPUT
POWER
FROM AS
POWER
BOARD
{
115/230
SWITCH
NEU-WHT
LINE-BLK
L N
FUSE
Figure II.E.2
(+)
(-)
NEU
GRN
RED
WHT
BLK
0-10VDC
}
OUTPUT
}
CONTROL
RELAY
AQUASWITCH™
*115/230VAC
FUSE
100mA
(T.10A) 5X20
Slow Blow/
Time Delay
*24VAC
FUSE
250mA
(T.25A) 5X20
SENSOR
ELECTRICAL CONNECTION DIAGRAM
Slow Blow/
Time Delay
GRN
ORN
WHT
ALARM
RELAY
AQUASWITCH II™
POWER BOARD
LOGIC
BOARD
CONNECTION
NO
COM
NC
{
{
BANK B
BANK A
BLEED
COM
SOLENOID
VALVE
OUTPUT
POWER
TRANSFORMER
NCNOCOM
BK B
BK A
GRDCAC
LINE / +DC
(BLK)
NEUTRAL / -DC
(WHT)
WHT
BLD
COM
115/230
JUMPERS
BLK
LINE
NEUTRAL
OR
115/230
SWITCH
FUSE*
}
SOLENOID
INPUT
}
POWER
FUSE
1/8 Amp
AGC 1/8
MAIN
INPUT
POWER
12
The following procedures are to be used to install the
AQUASWITCH II to 115 VAC main power source. For the
procedures to install the optional 230 VAC main power source, the
user must first complete the installation procedures in Section
II.E.3. Failure to do so could result in damage to equipment and/or
property.
NOTE: Both early and later models function exactly the same.
II.E.3.a. Early Model Power Boards with
Buss Wire JUMPERS
This requires extreme care as circuit board may be damaged if
improperly attempted.
1. Verify that the main power source is turned "OFF" or
disconnected.
2. Using a standard slot screwdriver remove the two (2)
screws on the front panel.
3. Carefully wiggle the front panel and pull gently toward
you. Do not pull more than about 8 inches/20CM or you
could damage the wiring harness.
4. Turn the front panel around so that the back side is
facing you and set aside for now.
5. Using the enclosure cutouts, install the proper wire and
watertight cable restraint (not provided) to comply with
local electrical codes.
6. Neatly connect wires to the AQUASWITCH II's
connectors*, as shown in figures II.E.2.
*CAUTION: The input power connectors require only a small
screwdriver, or a pen to push on the release levers†. The release
levers may be broken or damaged if not pushed straight toward
the circuit board. DO NOT push the release levers sideways.
† Early versions have screw terminals.
3. 115/230 VAC CONVERSION
Before turning power ON to the AQUASWITCH ensure the
proper input voltage has been selected. For the AQUASWITCHI, see figures II.E.1 & II.E.3. This is required in TWO locations on
the AQUASWITCH II, see figures II.E.2 & II.E.3. Failure to do
so will blow the fuse(s). It could, under some conditions, cause
injury and damage the instrument voiding the warranty.
On the early model AQUASWITCH Power Board the conversion
must be accomplished by desoldering and resoldering a buss wire
jumper as shown in figure II.E.3, (to desolder and resolder the
JUMPERS, the circuit board (CB) must be removed from the
enclosure), or may be accomplished as shown in figure II.E.4.
Later models have switches, see section II.E.3.b and figure
II.E.2.
1.Locate the jumpers below the transformer.
2.Cut jumpers in location as shown in figure II.E.4.
3.Carefully, bend leads toward each other (center) as
shown in figure II.E.4.
4.Solder the ends of the leads together.
Cut and carefully bend leads toward center
115 VAC
Solder ends together
230 VAC
JUMPERS LOCATED UNDER TRANSFORMER
EARLY MODEL POWER BOARD
115/230 VAC JUMPER CONVERSION
Figure II.E.4
II.E.3.b. Later Model Power Boards with SWITCH
1.Locate switch next to the fuse holder.
2.Using a screwdriver, turn switch to required voltage.
II.E.3.c. AQUASWITCH II Main Control Board
ALL AQUASWITCH II Main Control Boards have a switch. See
figure II.E.2.
115 VAC
230 VAC
JUMPERS LOCATED UNDER TRANSFORMER
EARLY MODEL POWER BOARD
115/230 VAC JUMPER CONVERSION
Figure II.E.3
1.Locate switch next to the fuse holder.
2.Using a screwdriver, turn switch to required voltage.
4. CONNECTING THE SENSOR CABLE
This section provides the procedure to properly wiring the sensor.
1.Place the SENSOR interface cable and user supplied
watertight cable restraint into the enclosure's
appropriate access hole.
2.Install the SENSOR cable wire to comply with local
electrical codes.
3.Follow the color code as marked. See figure II.E.2.
*CAUTION: The sensor connectors require only a small
screwdriver or a pen to push on the release levers. The release
levers may be broken or damaged if not pushed straight toward
the circuit board. DO NOT push the release levers sideways.
13
a. USP 25 MODIFICATION
(No Temperature Compensation)
This simple modification will allow your AQUASWITCH II
Monitor/controller to meet the USP 25 requirements by defeating
the normal temperature compensation circuit thus giving
“uncompensated” readings as required.
SENSOR LEADS
REMOVE THESE
THREE LEADS
Specifications:
As required to meet USP25.
Installation
Briefly For Resistivity models, two resistors are installed in place of the
sensor “temperature” sensing leads.
For Conductivity/TDS models, a resistor is installed in place of the
sensor “temperature” sensing leads.
The extra sensor leads are either cut off or the ends are wrapped
in tape to prevent shorting.
CAUTION - READ FOLLOWING CAREFULLY
!
WARNING: BEFORE STARTING, IF THE
AQUASWITCH II IS INSTALLED, ENSURE THE
POWER IS OFF. FAILURE TO DO SO COULD CAUSE
DAMAGE TO THE INSTRUMENT, AND COULD BE
HARMFUL OR FATAL TO PERSONNEL. ONLY
QUALIFIED PERSONNEL SHOULD INSTALL OR
SERVICE ELECTRICAL EQUIPMENT.
Requirements:
For Resistivity; one 100kΩ 1% resistor, and one 5.49KΩ 1%
resistor, user supplied or may be ordered from the Myron L
Company.
For Conductivity/TDS; one 10kΩ 1% resistor, user supplied or
may be ordered from the Myron L Company.
NOTE: When opening instrument, remove front cover with care;
a ribbon cable connects the front panel and main board. If the
front panel has all ready been removed from the enclosure skip to
#4.
BK WT RD GN NU R- R+
3S
CAL
Resistivity Main CB Assembly
Figure II.E.5
SENSOR LEADS
BK WT RD GN NU R- R+
3S
INSTALL 5.49K
RESISTOR HERE
Resistivity Main CB Assembly
Figure II.E.6
FS SW
INSTALL 100K
RESISTOR HERE
FS SW
CAL
1.Using a standard slot screwdriver remove the two (2)
screws on the front panel.
2.Carefully wiggle the front panel and pull gently toward
you. Do not pull more than about 8 inches/20CM or you
could damage the wiring harness.
3.Turn the front panel around so that the back side is
facing you and set aside.
4.For Resistivity Monitor/controllers;
a.If sensor is installed, locate and remove the BLACK
(BK), RED (RD), and the NEUTRAL (NU) leads
from MAIN Circuit Board, as shown in figure
II.E.5.
b.Cut off or tape BLACK (BK), RED (RD), and the
NEUTRAL (NU) leads from sensor.
c.Install 100kΩ resistor at BLACK (BK) and GREEN
(GN) connector locations, as shown in figure
II.E.6.
d.Install 5.49kΩ resistor at RED (RD) and the
NEUTRAL (NU) connector locations, as shown in
figure II.E.6.
For Conductivity/TDS Monitor/controllers;
a.If sensor is installed, locate and remove the RED
(RD) and the GREEN (GN) leads from MAIN
Circuit Board, as shown in figure II.E.7.
b.Cut off or tape RED (RD) and the GREEN (GN)
leads from sensor.
c.Install 10kΩ resistor at RED (RD) and the GREEN
(GN) connector locations, as shown in figure
II.E.8.
5.Carefully reinstall the front panel, bottom first. Ensure no
wires have been pinched between enclosure and front
panel.
6.Reinstall the two (2) screws and tighten.
7.To operate, turn power ON.
NOTE: Recalibration will require both the solution and sensor be
at 25°C for maximum accuracy.
14
SENSOR LEADS
BK WT RD GN NU R- R+
3S
}
UP
REMOVE THESE TWO LEADS
0-10VDC
OUTPUT
CAL
FULL SCALE
PUSH TO TEST
HYS1SP1FS SW
SPC
INC
DIS
-11
0-20mΩ
DEC
Conductivity/TDS Main CB Assembly
Figure II.E.7
SENSOR LEADSINSTALL 10K
RESISTOR HERE
BK WT RD GN NU R- R+
3S
FS SW
CAL
Conductivity/TDS Main CB Assembly
Figure II.E.8
5. SOLID STATE OUTPUT
24 VDC Unregulated 30mA max. The following instructions are
assuming the Monitor/controller enclosure is already open.
a. Piezo Electric Alarm - PA/PAO (option)
For additional information, see Piezo Alarm under Options in
section III.I.
1.If not already installed, peel tape backing from PIEZO
and press into place per figure III.I.3.
2.Attach connector to main control circuit board per figure
III.I.4.
NOTE: If remotely mounted; cut wires and splice as necessary,
use comparable wire. Piezo requires 1/4” (6.35mm) hole in user
panel.
b. Remote Alarm - RA™ (option)
For additional information,see RA Instructions under options in
section III.J.
1. Run user supplied #22, 2 conductor speaker type wire
from Monitor/controller to RA location as necessary.
Additional wire may be ordered, part #RAW-200, see
Options & Accessories.
2. Open the RA by removing the four screws.
3. Locate and remove the 8” 2 conductor wire attached to
RA.
4. At the controller, connect the extension wires to the 8”
2 conductor wire with the wire nuts provided — Black to
Positive (+) and White to Negative (-). Be sure to first
pass the wire through the user supplied waterproof strain
relief in the enclosure.
5. Plug the reddish brown female connector into the male
connector on the controller CB marked either RA or PA
(see inside case label for location). It will only go on the
connector one way.
6. At the RA, connect the wires to the connector — Black
to Positive (+) and White to Negative (-).
7. To test, simply turn on the controller and adjust controller
set point until the alarm/piezo sounds off. If controller is
not yet connected to a functioning sensor, on
conductivity/TDS controllers it will be necessary to press
and hold the Full Scale test switch.
The black button on the front of the RA will mute the
piezo alarm for approximately three minutes or until you
improve the water quality (readjust controller set point).
The piezo alarm will continue to sound off every three
minutes until the user has improved the alarm condition
inside the controller. If three minutes muting is fine for
your application, skip to step 9.
8. If three minutes is too long or too short, adjust time delay
control inside RA until desired mute time is achieved
adjustable from approximately 6 seconds to 10 minutes).
9. Replace the bottom of the RA, and secure RA to the
surface you have selected for its installation.
NOTE: If the RA does not sound off;
1. Check the polarity of the extension wire connections.
2. Be sure the controller is actually switching (relay will
click).
c.Connect to your own alarm or indicator
Use the following as guidelines.
Connector is a standard 2 wire Methode* style
connector. Connector with 8” wires, part #RAH, is
available from the Myron L Company.
Ensure your requirements do not exceed the 24 VDC
Unregulated 30mA maximum.
Ensure the polarity is correct (RED is positive), see
figure V.A.1.
Attach wires to RA.
Attach connector to controller connector (RA) per figure
V.A.1.
*Methode is registered trademark of Methode Electronics, Inc.
REASSEMBLY
1.Carefully reinstall the front panel, bottom first.
Ensure no wires have been pinched between
enclosure and front panel.
2.Reinstall the two (2) screws and tighten.
3.To operate, turn power ON.
15
6. SOLENOID VALVE CONNECTIONS
The solenoid valves require electrical power to operate. This
power must be supplied by the user. See figure II.E.10 for the
suggested valve wiring. This section provides the procedures for
making the appropriate electrical connections.
a. External Power Connection
Valves powered by an external USER supplied source, i.e.
115/230 or 24 VAC.
1. Place the USER supplied power cable and watertight
cable restraint into the enclosure’s appropriate access
hole.
2. Connect the cable wires to the Power Board terminal
block as shown in figure II.E.1 or II.E.2.
For 24 VAC applications, the Myron L Company offers a 115 VAC
to 24 VAC transformer, Model #VR. Other voltages must be usersupplied.
AQUASWITCH POWER BOARD
CUSTOMER POWERED
SOLENOID OUTPUT
(STANDARD)
OPTIONAL
SELF-POWERED
SOLENOID OUTPUT
(ADD BUSS WIRE
JUMPERS AS SHOWN)
SOLENOID VALVE POWER JUMPER
DIAGRAM
FOR EARLY MODELS
For jumper location, see Figure II.E.11.
VALVE
A
VALVE
B
BLEED
VALVE
OPTIONAL
PROCESS
VALVE
TO
COM
TO
BLEED
TO
BANK B
SUGGESTED VALVE WIRING
DIAGRAM
Figure II.E.10
TO
BANK A
Figure II.E.9
b. Valve Wiring
Suggested method of wiring solenoid valves. See figure II.E.10.
1. Place the Bank A Valve solenoid interface cable and
watertight cable restraint into the enclosure’s
appropriate access hole.
2. Neatly connect the cable wires to Power Board terminal
block as shown in figure II.E.1 or II.E.2.
3. Repeat steps 1 and 2 to install the Bank B Valve solenoid
interface cable.
4. Repeat steps 1 and 2 to install the Bleed Valve solenoid
interface cable.
5. (Optional) Repeat steps 1 and 2 to install the Process
valve solenoid interface cable.
REASSEMBLY
1.Carefully reinstall the front panel, bottom first.
Ensure no wires have been pinched between
enclosure and front panel.
2.Reinstall the two (2) screws and tighten.
3.To operate, turn power ON.
INPUT FROM
CONTROLLER
RELAY
ALARM
RELAY
SOLENOID
VALVE
OUTPUT
POWER
AQUASWITCH™
POWER BOARD
LOGIC
BOARD
CONNECTION
NCNOCOM
{
NO
COM
NC
TRANSFORMER
GRDCAC
BK B
BK A
BLD
115/230
JUMPERS
LINE / +DC
(BLK)
NEUTRAL / -DC
(WHT)
WHT
BLK
COM
LINE
NEUTRAL
FUSE
1/8 Amp
AGC 1/8
{
BANK B
BANK A
BLEED
COM
EARLY MODEL AQUASWITCH
ELECTRICAL CONNECTION DIAGRAM
Figure II.E.11
SOLENOID
VALVE
POWER
JUMPERS
}
SOLENOID
INPUT
}
POWER
MAIN
INPUT
POWER
16
7. ALARM/CONTROL RELAY CONNECTIONS
The Myron L Company AQUASWITCH is equipped with a “Dry
Contact” relay which is designed to energize/de-energize when
the set point is crossed. (See section IV.B.2) for set point
adjustment procedure) The relay energizes on increasing or
decreasing readings as set by the user, see section IV.B.1.
When energized (above set point), the Common (COM) will
disconnect from the Normally Closed (NC) contact and connect to
the Normally Open (NO) contact. Devices may be operated using
either the Normally Open contact or Normally Closed contact; or
both relay contacts may be used to control two devices of the
same voltage.
AQUASWITCH
CONTROLLER
CIRCUIT BOARD
RELAY
NC
COM
NO
115VAC
POWER
SOURCE
115 to 24VAC
TRANSFORMER
VR
COM
NO
WARNING: CONNECTING BOTH POWER SOURCE
LEADS TO THE RELAY TERMINAL BLOCK
CONNECTERS WILL DAMAGE THE CIRCUIT BOARD
AND MAY CAUSE PERSONAL INJURY.
1. Place the user supplied Alarm relay interface cable and
watertight cable restraint into the enclosure's
appropriate access hole.
2. Neatly connect the relay interface cable wires to the
AQUASWITCH power board terminal block connectors,
see figures II.E.1 or II.E.2, and II.E.12 or II.E.13.
CAUTION: The connectors require only a small screwdriver or a
pen to push on the release levers. The release levers may be
broken or damaged if not pushed straight toward the circuit board.
DO NOT push the release levers sideways. Early models have
screw terminals.
NOTE: See Section IV.C for the “ALARM RESET” jumper
adjustments.
The easiest method of connecting the relay is shown in figures
II.E.1 or II.E.2 and II.E.12. These show how the dry contact relay
can use incoming power to activate a controlled device (alarm,
solenoid valve, etc.) of 10 amps or less.
For 24 VAC applications, the Myron L Company offers a 115 VAC
to 24 VAC transformer, Model #VR, see figure II.E.13. Other
voltages must be user-supplied.
ALARM RELAY
TERMINAL BLOCK
ALARM
Figure II.E.13
8. CONNECTING DISPLAY HARNESS TO DISPLAY
If the installation required the removal of the display harness from
the display, the following procedure will ensure it is reinstalled
without damaging the display.
WARNING: THE DISPLAY WILL BE IRREPARABLY
DAMAGED IF THE HARNESS IS INSTALLED UPSIDEDOWN OR MISALIGNED. THE HARNESS MUST BE
INSTALLED AS SHOWN IN FIGURE II.E.14.
1. Grasp connector and align wires DOWN on display or with
the small edge of the display as shown in figure II.E.14.
2. Press connector onto display pins. Ensure pins are
aligned or they may become bent. Wiggle connector
slightly “end to end” if necessary.
DISPLAY
DISPLAY CONNECTION
AQUASWITCH
CONTROLLER
CIRCUIT BOARD
RELAY
ALARM RELAY
TERMINAL BLOCK
NC
COM
COM
NO
NO
Figure II.E.12
POWER
SOURCE
DISPLAY
HARNESS
PANEL MOUNTED DISPLAY
REAR VIEW
Figure II.E.14
ALARM
REASSEMBLY
1.Carefully reinstall the front panel, bottom first.
Ensure no wires have been pinched between
enclosure and front panel.
2.Reinstall the two (2) screws and tighten.
3.To operate, turn power ON.
17
F. 0-10VDC RECORDER OUTPUT
The 0-10VDC output is designed to give the user the capability of
sending a signal to a remote meter, recorder, PLC or SCADA
system.
1. CONNECTION
1.Place the user supplied interface cable and watertight
cable restraint into the enclosure's appropriate access
hole. Skip for OEM.
2.Connect the Recorder's plus (+) and minus (-) terminal
wires to the Recorder output connectors. (See figure
II.E.1.)
3.Refer to Section V.D.1.b for the procedures to calibrate
the 0-10 VDC voltage output.
+
0-10V Output
-
A
B
Figure II.F.1
+
Recording
Device
-
2. VOLTAGE DIVIDER
A voltage divider gives the user the ability to scale or tailor the
output to a particular need or requirement due to the input of
another device, i.e. the output of the Main CB is 0-10V while the
input requirement of a particular recording device is 0-5V.
a. INSTALLATION
Briefly—
Two resistors are installed across the 0-10V output.
The output is recalibrated to required voltage.
Requirements
Select two (2) resistors as listed;
For 0-5V Output both "A & B" are 2K Resistors.
For 0-1V Output "A" is a 9K resistor and "B" is a 1K resistor.
CAUTION - READ FOLLOWING CAREFULLY
!
WARNING: BEFORE STARTING, IF THE
AQUASWITCH IS INSTALLED, ENSURE THE POWER
IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE
TO THE INSTRUMENT, AND COULD BE HARMFUL OR
FATAL TO PERSONNEL. ONLY QUALIFIED
PERSONNEL SHOULD INSTALL ELECTRICAL
EQUIPMENT.
Physical
If the front panel has all ready been removed from the enclosure
skip to #3.
1. Using a standard slot screwdriver remove the two (2)
screws on the front panel.
2. Carefully wiggle the front panel and pull gently toward
you. Do not pull more than about 8 inches/20CM or you
could damage the wiring harness.
3. Turn the front panel around so that the back side is
facing you and set aside for now.
4. Solder two selected resistors together as shown in figure
II.F.1.
5. Attach leads to recording device as shown in figure
II.F.1.
6. Attach resistors to 0-10V Output as shown in figure
II.F.2.
Ensure resistors and leads DO NOT short to each other
or to any part of the CB assembly.
7. Recalibration is required, see Calibration Procedures,
section V.D.
NEW
OUTPUT
AS
SENSOR LEADS
-+
BK WT RD GN NU R- R+
3S
SELECTED
0-10VDC
OUTPUT
FS SW
CAL
Main CB Assembly
Figure II.F.2
Reassembly
1. Carefully reinstall the front panel, bottom first, ensure no
wires have been pinched.
2. Reinstall the two (2) screws and tighten.
3. To operate, turn power ON.
18
G. RE-RANGE YOUR AQUASWITCH II
!
(Range Module Installation)
1. DESCRIPTION
The AQUASWITCH II Monitor/controllers have been designed
for easy field re-rangeability. The Range Module consists of a 16
pin Header that plugs into a 16 pin socket.
For available ranges, see Range Selection Guide I.G. When
making large range changes in conductivity/TDS, i.e. 1 ppm to 50
ppm, a different sensor may also be required as noted in the
Range Selection Guide. Order Range Module by adding the prefix
“RM” to the range number as in examples below.
Resistivity Range Modules—
RMXX i.e. = RM11is a 0-20MΩ
Conductivity/TDS Range Modules —
RMXXXi.e. = RM111is a 0-50µS
2. INSTALLATION
Briefly The new Range Module simply replaces the Range Module
presently installed, see figure II.G.1.
Type label may be required if changing from Microsiemens to
PPM.
The Full Scale reading is recalibrated.
POINTER
CAUTION - READ FOLLOWING CAREFULLY
WARNING: BEFORE STARTING, IF THE
AQUASWITCH IS INSTALLED, ENSURE THE POWER
IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE
TO THE INSTRUMENT, AND COULD BE HARMFUL OR
FATAL TO PERSONNEL. ONLY QUALIFIED
PERSONNEL SHOULD INSTALL OR SERVICE
ELECTRICAL EQUIPMENT.
Physical
NOTE: When opening instrument, remove front cover with care;
a ribbon cable connects the front panel and main board. If the
front panel has already been removed from the enclosure skip to
#4.
1. Using a standard slot screwdriver remove the two (2)
screws on the front panel.
2. Carefully wiggle the front panel and pull gently toward
you. Do not pull more than about 8 inches/20CM or you
could damage the wiring harness.
3. Turn the front panel around so that the back side is
facing you and set aside.
4. Locate and remove existing Range Module from MAIN
Circuit Board, as shown in figure II.G.1. It is not easy to
remove, it was designed to stay in place under adverse
conditions.
5. With the pointer up, carefully align the new Range Module
to the socket on the MAIN Circuit Board as shown in
figure II.G.1.
6. Press firmly into place.
7. Recalibrate, see CALIBRATION PROCEDURES, section
V.D.
0-10VDC
OUTPUT
BK WT RD GN NU R- R+
3S
}
UP
-11
0-20mΩ
751 756
752 757
753 758
754 759
MYRON L
COMPANY
FSSW
CAL
RANGE
MODULE
FULL SCALE
PUSH TO TEST
HYS1SP1
SPC
INC
DIS
DEC
REMOVE TO
INSTALL
SECOND
RELAY
TRANSFORMER
115/
230
FUSE*
PWR C GD NC NO CM
CHS
MAIN MONITOR/CONTROLLER CIRCUIT BOARD
Figure II.G.1
PA RA
GND
19
III.OPTIONS &
ACCESSORIES
A.4A/4AO MODULE (4-20mA OPTION)
-4A4-20mA Self/Remote-powered Isolated output module
ordered with Monitor/controller.
The 4-20mA option gives the AQUASWITCH II
Monitor/controller the ability to send a signal very long distances
with minimal interferences and signal degradation. The output is
an Isolated 4-20mA signal that corresponds to the full scale range
of the Monitor/controller into which it is installed. This output is
easily configured to be either self-powered or remote-powered as
required for your particular application.
NOTE: The maximum impedance of the user's current input
instrument should not exceed 600 ohms.
without any degradation of performance.
There are two modes in which current loop transmitters operate;
Self-Powered and Remote-Powered.
Self-Powered — the transmitter provides the power to drive
the 4 to 20 mA current. See figure III.A.1.
Remote-powered — the receiving instrument provides the
power to drive the 4 to 20 mA current. See figure III.A.2.
Specifications
Self-Powered and Remote-Powered
Drive Impedance — 0 to 600Ω
Common Mode Maximum — 120VAC @ 60 Hz
Isolation — 100pf max. to Model 750II circuit common
100pf max. to input power line
Calibration
Two multi-turn pots — Factory Set.
4mA = Zero (0)
20mA = Full Scale
+
-
SO PO PISO PO PI
+
POWER IN
POWER OUT
(-)
(+) SIGNAL OUT
N
(+)
Self - Powered
Figure III.A.1
(+)
+
-
+
POWER IN
POWER OUT
(-)
(+) SIGNAL OUT
N
Remote - Powered
Figure III.A.2
4A (4-20mA) Wiring Options
Since the output is an isolated current loop, it is the ideal choice
for applications requiring; a control signal to be run very long
distances, systems requiring a 4-20mA input or in instances
where isolation is necessary.
As the output is isolated, and a current, it is useful for long
distance interface, especially where wiring resistances may be
high, and/or the ground potentials may differ between the sensor
input ground and the current receiving instruments ground.
The 4-20mA output will not be degraded in accuracy even when
the ground differences are as much as 120VAC @ 60Hz.
Interface wire resistance of 350Ω will not degrade the accuracy of
the output when interfaced to a typical 250Ω input impedance of a
transmitter current input device.
The output is capable of driving a minimum of 600Ω worse case,
therefore, will satisfy virtually all modern interface requirements.
Current input devices usually have an input impedance of 250Ω,
however, some older designs can have as high as 500Ω or as low
as 10Ω. This “-4A” option will drive any impedance from 0 to 600Ω
CURRENT INPUT INSTRUMENT
CURRENT INPUT INSTRUMENT
+ 35 VDC
MAXIMUM
Calibration is NOT required. However, if you feel you must verify
or recalibrate, see Recalibration below.
2. INSTALLATION
Briefly The 4-20 Module replaces the plastic display retainer plate
attached to the front panel.
The 4-20 Module harness is attached to the main circuit board as
marked ‘4-20’. See figure III.A.4.
The 4-20 output is wired as required - Self-powered or Remotepowered. See figures III.B.1 & III.A.2.
CAUTION - READ FOLLOWING CAREFULLY
!
WARNING: BEFORE STARTING, IF THE
AQUASWITCH IS INSTALLED, ENSURE THE POWER
IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE
TO THE INSTRUMENT, AND COULD BE HARMFUL OR
FATAL TO PERSONNEL. ONLY QUALIFIED
PERSONNEL SHOULD INSTALL ELECTRICAL
EQUIPMENT.
20
Physical
If the front panel has all ready been removed from the enclosure
skip to #3.
1. Using a standard slot screwdriver remove the two (2)
screws on the front panel.
2. Carefully wiggle the front panel and pull gently toward
you. Do not pull more than about 8 inches/20CM or you
could damage the wiring harness.
3. Turn the front panel around so that the back side is
facing you.
4A/4AO - 4-20mA MODULE
RECONNECT
DISPLAY
CONNECTOR
HERE
DISPLAY
have already connected the other end of the wires as
required.
a.Place the remote interface cable and user supplied
watertight cable restraint into the enclosure's
appropriate access hole.
b.Neatly connect the signal cable wires to the
Monitor's appropriate connectors as shown in figure
III.A.3.
3.To test, turn power ON.
4.Press the Full Scale Test switch and monitor the output
at your remote site, or with a DVM set to DC milliamps.
DISPLAY IS ON BACK SIDE OF 4-20 MODULE
420
TO CIRCUIT BOARD
4-20
(+)
+
-
SO PO PI
+
POWER IN
POWER OUT
(-)
(+) SIGNAL OUT
Figure III.A.3.
4. Using a standard slot screwdriver remove the four (4)
screws holding the plastic display retainer plate to the
front panel. When the screws have been removed, the
plastic display retainer plate and the display will be free
from the front panel.
5. Set front panel down or carefully allow to hang from the
harness. Do not drop as the harness connector will pull
out allowing the front panel to fall.
6. While holding the display and the plastic display retainer
plate, carefully remove the display harness connector.
Do not drop the display. Remove and discard the plastic
display retainer plate.
7. While still holding the display in the palm of your hand,
set the 4-20 Module over it with the display pins
protruding through the center opening as shown in figure
III.A.3.
8. Reconnect the harness to the display as shown in figure
III.A.3.
9. While holding the front panel, align the display to the
opening while at the same time align the 4-20 Module
mounting holes to the front panel.
10. Reinstall the four (4) screws and tighten.
Electrical
1.Connect the 4-20 Module (five) 5 wire harness to the
main circuit board at the location next to the display
harness marked “4-20” as shown in figure III.A.4.
2.Connect the signal and power wires as required, as
shown in figures III.A.1. & III.A.2. This assumes you
Attach the DVM to the output connectors per your
requirements, i.e. self-powered or remote-powered, see
figures III.A.1. & III.A. 2. If the 4A module is connected
properly it will indicate 20mA.
5.Turn power OFF.
6.Carefully reinstall the front panel, bottom first, ensure no
wires have been pinched between enclosure and front
panel.
7.Reinstall the two (2) screws and tighten.
8.To operate, turn power ON.
3. RECALIBRATION
The 4-20 Module was calibrated at the factory, however, if you
wish to check the calibration the following procedure will help you
to accomplish this task. Exercise caution while performing this
procedure.
Requirements; a DVM set to DC milliamps, a tweaker or small
standard slot screwdriver.
This procedure assumes the front panel is removed.
1.If sensor is connected, disconnect sensor wires from
sensor terminal block.
2.Attach the DVM to the output connectors per your
requirements, i.e. self-powered or remote-powered, see
figure III.A.5.
3.Turn power ON, with the front panel meter/display at
ZERO, the DVM should indicate 4 milliamps.
21
BK WT RD GN NU R- R+
3S
}
UP
-11
0-20mΩ
0-10VDC
CAL
FS SW
FULL SCALE
TEST SWITCH
HYS1SP1
SPC
INC
DIS
DEC
4.If not, adjust the CAL control marked “4mA” until the
DVM indicates 4mA, see figure III.A.5.
5.Press the Full Scale Test Switch, the DVM should
indicate 20 milliamps.
6.If not, adjust the CAL control marked “20mA” until the
DVM indicates 20mA. See figure III.A.5.
7. Calibration is complete
8.Turn power OFF.
9.Reconnect sensor wires to sensor terminal block as
shown in figure III.A.4.
10. Carefully reinstall the front panel, bottom first, ensure no
wires have been pinched between enclosure and front
panel.
11. Reinstall the two (2) screws and tighten.
12. To operate, turn power ON.
751 756
752 757
753 758
754 759
MYRON L
COMPANY
TRANSFORMER
115/
230
FUSE*
PWR C GD NC NO CM
4-20
REMOVE TO
INSTALL
SECOND
RELAY
CHS
MONITOR/CONTROLLER CIRCUIT
BOARD ASSEMBLY
Figure III.A.4.
(+)
+
-
SO PO PISO PO PI
+
POWER IN
POWER OUT
(-)
(+) SIGNAL OUT
Self - Powered
4-20
CONNECTOR
CONNECT 4AO
HARNESS HERE
PA RA
GND
N
DV
(+)
+
-
+
POWER IN
POWER OUT
(-)
(+) SIGNAL OUT
N
+
-
Remote
Power
Supply
Remote - Powered
DV
Figure III.A.5.
22
4. CONVERTING A CURRENT
TO A VOLTAGE
Current measuring devices actually measure voltage, but have an
internal resistor as shown in figures III.A.1 & III.A.2 and are
scaled to display in current. If you have a voltage input
instrument and you wish to utilize the current from the 4A (420mA) Module, the following will help you to make this conversion.
To convert a 4-20 mA current output to operate or drive a voltage
input device it is necessary to install a LOAD resistor across the
input terminals as shown in figures III.A.6 & III.A.7. The value of
the resistor is chosen to match the input voltage range, i.e. 0-10
Volts requires a 500Ω resistor and will produce a 2 to 10 input
voltage. This floating zero is useful to indicate a broken 4-20 input
wire when the indication is zero volts.
For an input voltage range of:
10.0 Volts the resistor value is 500 ohmsindicates2.0 V @ 4 mA.
For other input ranges, divide the input voltage range by 0.02, the
answer will be in ohms.
NOTE: The tolerance of the load resistor directly affects the
accuracy of the resulting voltage, i.e. 5% resistor = 5% error.
VOLTAGE INPUT INSTRUMENT
VOLTAGE INPUT INSTRUMENT
+ 35 VDC
MAXIMUM
23
B. PIEZO ELECTRIC ALARM
-PA Piezo Electric Alarm ordered with AQUASWITCH II.
PAO Piezo Electric Alarm ordered separately for
AQUASWITCH II.
1. DESCRIPTION
The PA/PAO is an electronic sound device capable of emitting a
80dB or more @ 30cm, high pitched squeal. See figure III.B.1.
Specifications
Oscillating Frequency — 3.0≠0.5KHz
Operating Voltage — 24VDC Nom. (1.5-30VDC Max.)
Sound Pressure Level (Min) 30cm/12VDC — 80dB
Current consumption (Max) @ 12VDC — 12mA
Tone — Constant
Operating Temperature — -20 - +60°C
Size — 24 x 9.5 mm
Double Backed Tape
Harness ~9"
CAUTION - READ FOLLOWING CAREFULLY
!
WARNING: BEFORE STARTING, IF THE
AQUASWITCH II IS INSTALLED, ENSURE THE
POWER IS OFF. FAILURE TO DO SO COULD CAUSE
DAMAGE TO THE INSTRUMENT, AND COULD BE
HARMFUL OR FATAL TO PERSONNEL. ONLY
QUALIFIED PERSONNEL SHOULD INSTALL OR
SERVICE ELECTRICAL EQUIPMENT.
Physical
NOTE: Remote mounting will require a .25” (6.35mm) hole, and
extending the harness. Use #22 gauge speaker wire. Observe
polarity.
If the front panel has all ready been removed from the enclosure
skip to #4.
1. Remove rubber plug as shown in figure III.B.2.
2. Using a standard slot screwdriver remove the two (2)
screws on the front panel.
Piezo Alarm
Figure III.B.1
2. INSTALLATION
Briefly A rubber plug is removed from the front panel, figure III.B.2.
The PA/PAO Piezo Alarm attaches to the front panel with the tape
supplied, figure III.B.3.
The wire harness plugs into a methode connector on the Main
Monitor/controller circuit board, figure III.B.4.
BANK A
IN USE GOOD POOR
BANK B
MEGOHMS - CM
SET POINT
HIGHLOW
BANK SWITCH ALARM RESET
Methode Connector
3. Carefully wiggle the front panel and pull gently toward
you. Do not pull more than about 8 inches/20CM or you
could damage the wiring harness.
4. Turn the front panel around so that the back side is
facing you.
5. Peal off tape backing from PA, and install as shown in
figure III.B.3.
6. Set front panel aside.
7. Connect wire harness to Main Monitor/controller circuit
board as shown in figure III.B.4.
Remove Rubber Plug
WATER QUALITY
BLEED
MYRON L
COMPANY
AQUASWITCH II
AQUASWITCH II FRONT PANEL
Figure III.B.2
24
Test
1. Turn power ON. Depending on configuration, the Piezo
will sound off when —.
NOTE: If the sensor is connected, the solution value, set point
value and Set Point Conversion (SPC) jumpers may affect the
test.
a.For a Resistivity AQUASWITCH II, as shipped
from the Myron L Company, the alarm will sound
off when the power is applied and stop when the Full
Scale Test Switch is pressed.
b.For a Conductivity/TDS AQUASWITCH II, as
shipped from the Myron L Company, the alarm will
sound off when the Full Scale Test Switch is
pressed
c.If the Set Point Conversion (SPC) jumpers have
been reversed, the opposite will be true in the
above descriptions.
2. Turn power OFF.
REASSEMBLY
Before reassembly, ensure set point control is set to desired trip
point.
1. Carefully reinstall the front panel, bottom first. Ensure no
wires have been pinched between enclosure and front
panel.
2. Reinstall the two (2) screws and tighten.
3. To operate, turn power ON.
Piezo
Alarm
FULL SCALE
PUSH TO TEST
BK WT RD GN NU R- R+
3S
}
UP
-11
0-20mΩ
751 756
752 757
753 758
754 759
CAL
MYRON L
COMPANY
INC
DIS
HYS1SP1FS SW
SPC
DEC
REMOVE TO
INSTALL
SECOND
RELAY
AQUASWITCH II
FRONT PANEL
Rear View
Figure III.B.3
TRANSFORMER
115/
230
FUSE*
PWR C GD NC NO CM
MAIN MONITOR/CONTROLLER
CIRCUIT BOARD
PA RA
PA Connection
CHS
(may use either
GND
connector)
Figure III.B.4
25
C.TEMPERATURE MODULE
-TP Temperature Module ordered with Monitor/controller.
TPOTemperature Module Kit ordered separately.
Requires; CSXX-TP sensor, Special Order, i.e. CS10-TP
or CS40-TP, see sensor data sheet.
.
1. DESCRIPTION
The Temperature Module (TP) gives the Monitor/controller more
flexibility for the user by being in the same package.
The Temperature Module is driven by the main display output and
is very simple to install.
The Temperature Module and Temperature Sensor SYSTEM
utilize a unique 3 wire technique by which errors are greatly
reduced.
The Temperature Module has its own 0-5VDC output.
TPO kit comes with all items necessary to install and operate: TP
Module; front panel harness with switch, bezel, cap and two orings (006 & 008); and TEMPERATURE label (# LTEMP).
NOTE: Requires the CSXX-TP sensor temperature wires be
routed directly to the TP/TPO CB.
Specifications
±200°C (±200°F Special Order Only)
Accuracy: ±0.2 w/ICE±0.5 w/out ICE
Display Resolution = 0.1
Output:0-5 VDC Connections - 2
Sensor Input:Cond/TDS, RES sensor with SPECIAL Platinum
RTD added or 1000Ω RTD temperature sensor.
Connections - 3 (3 wire for increased
accuracy over long distances)
Calibration
Zero and Full Scale, FS (0 & 5 VDC): Factory SetSimple pots
Display (CAL): Full Scale, FS (199.9) Factory SetSimple pot
Operates on safe ±5 VDC supplied by the Main CB.
2. INSTALLATION
Briefly The Temperature Module replaces the plastic display retainer
plate attached to the front panel.
The display harness plugs into the Temperature Module instead of
the display.
The -TP sensor leads are connected to the Temperature Module.
See figure III.C.1.
The display switch is installed in the lower front panel as shown in
figure III.C.4.
A “TEMPERATURE” label is added to lower front panel next to the
Display Select switch as shown in figure III.C.4.
CAUTION - READ FOLLOWING CAREFULLY
!
WARNING: BEFORE STARTING, IF AQUASWITCH II
IS INSTALLED, ENSURE THE POWER IS OFF.
FAILURE TO DO SO COULD CAUSE DAMAGE TO THE
INSTRUMENT, AND COULD BE HARMFUL OR FATAL
TO PERSONNEL. ONLY QUALIFIED PERSONNEL
SHOULD INSTALL OR SERVICE ELECTRICAL
EQUIPMENT.
Physical
If the front panel has all ready been removed from the enclosure
skip to 3.
1. Using a standard slot screwdriver remove the two (2)
screws on the front panel.
2. Carefully wiggle the front panel and pull gently toward
you. Do not pull more than about 8 inches/20CM or you
could damage the wiring harness.
3. Turn the front panel around so that the back side is
facing you.
4. Using a standard slot screwdriver remove the four (4)
screws holding the plastic display retainer plate to the
front panel. When the screws have been removed, the
plastic display retainer plate and the display will be
free from the front panel.
5. Set front panel down or carefully allow to hang from the
harness. Do not drop as the harness connector will pull
out allowing the front panel to fall.
6. While holding the display and the plastic display retainer
plate, carefully remove the display harness connector.
Do not drop the display.
7. Remove and discard the plastic display retainer plate.
8. Insert display connector pins into the Temperature
Module female connector. See figure III.C.1.
RECONNECT
DISPLAY
HARNESS
HERE
Figure III.C.1
HYS2
TP/TPO - TEMPERATURE MODULE
DISPLAY
DISPLAY CONNECTION
FSZERODISSP2
SPC
INC
INC
DEC
DEC
TEMP
CONTROL
HARNESS
DISPLAY
SELECT
+
0-5VDC
-
OR
BLU
BRN
BR BL OR - +
JUMPER
(+)
Orange
Blue
Brown
DISPLAY IS ON THE
BACK SIDE OF THE
TEMPERATURE
MODULE
0-5VDC
OUTPUT(-)
TEMPERATURE
SENSOR INPUT
26
9. While holding the front panel, align the display to the
opening and at the same time, align the Temperature
Module mounting holes to the front panel.
10. Reconnect display harness with leads down as shown in
figures III.C.1 & III.C.2.
11. Reinstall the four (4) screws and tighten.
Display Select Switch installation
1. Using a small sharp knife or 1/4” (6.35mm) drill, carefully
cut open the lower center hole of the front panel, when
viewed from the back. See figure III.C.2.
2. Install push-button display switch into this hole and
tighten bezel. See figures III.C.2 & III.C.4.
3. Install push-button/cap on switch.
FSZERODISSP2HYS2
SPC
INC
INC
DEC
DEC
TEMP
CONTROL
HARNESS
DISPLAY
DISPLAY CONNECTION
DISPLAY
SELECT
+
0-5VDC
-
BLU
BRN
TEMP
BR BL OR - +
SENSOR
(+)
0-5VDC
OUTPUT
(-)
TEMPERATURE
SENSOR INPUT
WARNING: There are two (2) o-rings installed on the switch, one
(1) on the threaded shank and the other is under the push button.
Both of these must be re-installed to maintain IP64/NEMA 3
ratings. See figure III.C.3.
4. Place “TEMPERATURE” label ABOVE the switch as
shown in figure III.C.4.
Electrical
1. Connect the display switch harness to the Temperature
Module as shown in figure III.C.1.
2. Connect the Temperature Sensor leads to the
Temperature Module as labeled in figure III.C.1.
CAUTION: The input connectors require only a small
screwdriver or a pen to push on the release levers. The release
levers may be broken or damaged if not pushed straight toward the
CB. DO NOT push the release levers sideways. Follow the color
code as labeled.
3. Connect 0-5 VDC output, if desired.
4. To test, turn power ON.
5. Press “TEMPERATURE” front panel switch, display will
show the temperature of the sensor.
6. Turn power OFF.
7. Continue or reinstall the front panel and tightly secure
both retaining screws, see REASSEMBLY below.
AQUASWITCH II FRONT PANEL
Figure III.C.2
Rear View
MEGOHMS - CM
SET POINT
TEMPERATURE
Switch Location
HIGHLOW
TEMPERATURE
Label Location
WATER QUALITY
BANK A
IN USE GOOD POOR
BANK B
BANK SWITCH ALARM RESET
SWITCH
O-RING
FRONT PANEL
BEZEL
PUSH BUTTON
O-RING
SWITCH and O-RING ASSEMBLY
Figure III.C.3
BLEED
MYRON L
COMPANY
TEMPERATURE
Switch Location
Figure III.C.4
TEMPERATURE
AQUASWITCH II
27
3. RECALIBRATION
Electronic Calibration Only. For System Calibration, see below.
The Temperature Module was calibrated at the factory, however,
if you wish to check the calibration the following procedure will
help you to accomplish this task. Exercise caution while
performing this procedure.
Requirements:
TP Calibration Module (TPC) or
Two .1% precision resistors -
ZERO - 1000.0Ω (0°C = 0.0V)
SPAN - 1758.56Ω (200°C = 5.0V)
One (1) jumper
A DVM set to DC Volts, a tweaker or small standard slot
screwdriver.
a. TPC “Calibration” Module Procedure
This procedure assumes the front panel is removed.
1. Ensure power is OFF.
2. Remove the sensor leads from the Temperature Module.
3. Remove the 0-5VDC leads from the Temperature Module.
4. Press TP Calibration Module (TPC) firmly into sensor
input connectors as shown in figure III.C.5.
5. Attach the DVM to the 0-5VDC output on the TPC
module. See figure III.C.5.
6. Turn power ON. The DVM should indicate 0.0 volts.
7. If not, adjust the calibration control marked “ZERO” (see
figure III.C.1 for location) until the DVM indicates 0.0
volts as shown in figure III.C.5.
8. Press switch on the TPC. See figure III.C.6.
9. The DVM should indicate 5.0 volts. See figure III.C.6.
10. If not, while pressing the TPC Module switch, adjust the
calibration control marked “FS” (see figure III.C.1 for
location) until the DVM indicates 5.0 volts as shown in
figure III.C.6.
11. Press the front panel TEMPERATURE select switch and
the TPC Module switch, the display should indicate
199.9 as shown in figure III.C.7.
Figure III.C.5
Figure III.C.6
TPC Module
PRESS BUTTON FOR
199.9ºC/5.0VDC
MYRON LCOMPANY
BLUE
BROWN
0-5VDC +
OUTPUT -
ORANGE
Press TPC Module into
Press TPC Module into
TPO connector as shown.
TPO connector as shown.
-
5.0
+
DV
Connect DVM leads as shown.
BR BL OR - +
199.9
FRONT PANEL DISPLAY
Figure III.C.7
12. If not adjust the calibration control marked “DIS” (see
figure III.C.1 for location) until the reading is 199.9 as
shown in figure III.C.7.
13. Calibration is complete.
14. Turn power OFF.
15. Remove TPC Module by pressing on each Phoenix
connector release lever*.
16. Reconnect sensor leads as labeled.
17. Reconnect 0-5 VDC output leads as labeled.
18. Continue or reinstall the front panel and tightly secure
both retaining screws, see REASSEMBLY below.
TPC Module
PRESS BUTTON FOR
199.9ºC/5.0VDC
MYRON LCOMPANY
BLUE
BROWN
BR BL OR - +
0-5VDC +
OUTPUT -
ORANGE
Press TPC Module into
Press TPC Module into
TPO connector as shown.
TPO connector as shown.
-
0.0
+
DV
Connect DVM leads as shown.
*CAUTION: The sensor input and 0-10 VDC output connectors
require only a small screwdriver or a pen to push on the release
levers. The release levers may be broken or damaged if not
pushed straight toward the CB. DO NOT push the release levers
sideways.
b. Precision Resistor Calibration Procedure
This procedure assumes the front panel is removed.
1. Ensure power is OFF.
2. Remove the sensor leads from the Temperature Module.
3. Install ZERO resistor (1000.0Ω) across sensor terminals
BL and OR. See figure III.C.8.
4. Install JUMPER between BR & BL as shown in figures
III.C.8 & III.C.9.
5. Attach the DVM to the 0-5 VDC output. See figures
III.C.8 & III.C.9.
6. Turn power ON. The DVM should indicate 0.0 volts.
28
7. If not, adjust the calibration control marked “ZERO” (see
figure III.C.1 for location).
8. Install SPAN resistor (1758.56Ω) across sensor
terminals BL and OR. See figure III.C.9.
9. The DVM should indicate 5.0 volts.
10. If not, adjust the calibration control marked “FS” (see
figure III.C.1 for location) until the DVM indicates 5.0
volts as shown in figure III.C.9.
11. Press the front panel TEMPERATURE select switch, the
display should indicate 199.9 as shown in figure
III.C.10.
12. If not adjust the calibration control marked “DIS” (see
figure III.C.1 for location) until the reading is 199.9.
13. Calibration is complete.
14. Turn power OFF.
15. Remove DVM from 0-5 VDC.
16. Remove resistor and jumper by pressing on each
Phoenix connector release lever*.
17. Reconnect sensor leads as labeled.
18. Reconnect 0-5 VDC output leads as labeled.
19. Continue or reinstall the front panel and tightly secure
both retaining screws, see REASSEMBLY below.
*CAUTION: The sensor input and 0-5 VDC output connectors
require only a small screwdriver or a pen to push on the release
levers. The release levers may be broken or damaged if not
pushed straight toward the CB. DO NOT push the release levers
sideways.
1000.0Ω RESISTER
JUMPER
-
0.0
+
DV
Connect DVM leads as shown.
c. System Calibration
By following these steps the complete temperature system,
module and sensor, may be calibrated to better than ±0.2°
centigrade accuracy. This procedure is similar to the electronic
calibration except the sensor is attached and is allowed to
equilibrate in “ICE” water before adjusting the ZERO calibration
control.
NOTE: One of the above electronic calibration procedures, TPC
Module or precision resistors, must be performed BEFORE the
system calibration is performed. This is required to preset the
span between zero and full scale.
1. Ice must be crushed in water to form a very thick slurry.
A slurry is that condition where the water to ice ratio is
such that only sufficient water is present to allow easy
stirring. At this point, the temperature of the water will
be 0.000°C.
2. Immerse sensor and cable approximately 2” (50mm) into
slurry.
3. Allow to equilibrate. Slurry must be constantly stirred.
4. Continue stirring until no further change in temperature is
observed on the display.
5. Adjust TP Module ZERO calibration control to 000.0.
6. Continue or reinstall the front panel and tightly secure
both retaining screws, see REASSEMBLY below.
REASSEMBLY
1. Carefully reinstall the front panel, bottom first. Ensure no
wires have been pinched between enclosure and front
panel.
2. Reinstall the two (2) screws and tighten.
3. To operate, turn power ON.
JUMPER
BR BL OR - +
Figure III.C.8
1758.56Ω RESISTER
Connect DVM leads as shown.
BR BL OR - +
Figure III.C.9
199.9
FRONT PANEL DISPLAY
Figure III.C.10
-
5.0
+
DV
29
D. REMOTE ALARM - RA™
RAOrdered as an accessory, includes 8” harness with
connector, and two (2) wire nuts.
1. DESCRIPTION
This remotely mounted AUDIBLE and VISUAL alarm connects to
any Myron L Company AQUASWITCH II Monitor/controller, or
brand “X” controller with dry contacts. When activated by the
controller the Remote Alarm will provide both an audible and a
visual alert at a location other than at the controller.
A mute button will silence the piezo alarm for up to 10 minutes
while the LED remains illuminated. After the preset time the piezo
alarm will again sound, this will repeat until the water quality is
corrected. Thus allowing servicing of the system under control,
while still acting as a reminder if the problem has not been
corrected.
The Remote Alarm - RA™ is an inexpensive way to alert personnel
of a trouble situation. For example; the Monitor/controller may be
located with an RO system while the service technicians are on
another floor or even in another building - great for hospitals. The
AUDIBLE alarm may be silenced, but stays in alarm (RED LED is
still illuminated) until the trouble is corrected. The timer may be
set from 15 seconds up to 10 minutes, thereby, giving personnel
the time to correct the problem while not being able to ignore the
trouble because of the reoccurring AUDIBLE and VISUAL alerts.
The Remote Alarm - RA connector is labeled 24 VDC INPUT and
has a 8” 2 conductor wires attached — Black is Positive (+) and
White is Negative (-).
When extending the wires you may use any two color wire you
wish just remember the polarity — Black is Positive (+) and White
is Negative (-). Wire lengths of 500 feet/152 meters have been
tested and have no adverse effect on performance although a
length of 500 feet/152 meters would be unusually long.
Ensure the unit is completely wired, tested and adjusted before
installing RA to mounting surface as the tape will not remove from
mounting surface without damaging the adhesive.
For OEM models skip steps referencing enclosure or front panel.
5
4
3
2
CONTROLLER
+ -10
1
30
15
6
7
8
9
MINSEC
Specifications
Audible Alert — Piezo Electric
Oscillating Frequency — 3.0±0.5KHz
Operating Voltage (750II) — 24 VDC Nom. (1.5-30VDC Max.)
Sound Pressure Level (Min) 30cm/12VDC — 80dB
Current consumption (Max) @ 12 VDC — 12mA
Tone — Constant
Operating Temperature — -20 - +60°C
Size — 24 x 9.5 mm
Visual Alert — Bright RED LED
Time Delay — 15 seconds to 10 minutes (USER adjustable)
Power — 24 VDC (supplied from Monitor/controller)
Case Material — ABS plastic
Dimensions — 4.75L x 2.56W x 1.56H (121L x 65W x 40H)
Mounting — Double Backed Tape (supplied)
Operational Distance — Wire lengths of 500 feet (152 meters)
have been tested with no adverse effects
Additional Wire available, order RAW-200. — 200 feet (61 meters)
2. INSTALLATION
Briefly —
Only 2 wires to connect from the controller to the RA.
Set the time.
Mount on the wall or on the bench.
CAUTION - READ FOLLOWING CAREFULLY
!
WARNING: BEFORE STARTING, IF THE
AQUASWITCH IS INSTALLED, ENSURE THE POWER
IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE
TO THE INSTRUMENT, AND COULD BE HARMFUL OR
FATAL TO PERSONNEL. ONLY QUALIFIED
PERSONNEL SHOULD INSTALL ELECTRICAL
EQUIPMENT.
RA CB ASSEMBLY
Figure III.D.1
1. Run #22, 2 conductor speaker type wire, NOT supplied,
from Monitor/controller to RA location as necessary. Wire
may be ordered from the Myron L Company, part #RAW-
200.
2. Open the RA by removing the four screws.
3. Locate and remove the 8” 2 conductor wire with the
reddish brown connector attached to the RA. See figure
III.D.1.
4. At the RA, connect the extension wires to the connector
on the RA circuit board — Black to Positive (+) and White
to Negative (-) as shown in figure III.D.1.
5. Using a standard slot screwdriver remove the two (2)
screws on the Myron L Monitor/controller front panel.
6. Carefully wiggle the front panel to loosen the gasket and
pull gently toward you. Do not pull more than about 8
inches/20CM or you could damage the wiring harness.
7. Connect the extension wires to the 8” 2 conductor wire
with the wire nuts provided — Black to Positive (+) and
White to Negative (-). Be sure to first pass the wire
through the user supplied waterproof strain relief in the
enclosure.
8. Plug the reddish brown connector into the male
connector on the controller circuit board marked RA (see
inside case label or figure III.D.2 for location). It will only
go on the connector one way.
9. To test, simply turn ON the controller and adjust
controller set point until the alarm sounds off*. If
controller is not yet connected to water, on
conductivity/TDS controllers it will be necessary to press
and hold the Full Scale Test Switch. See figure III.D.2.
The black button on the front of the RA will mute the piezo
30
alarm for approximately three minutes or until you
improve the water quality (readjust controller set point).
If three minutes muting is fine for your application, skip
to #10.
10. If three minutes is too long or too short, adjust control
inside RA until desired mute time is achieved (adjustable
from approximately 15 seconds to 10 minutes). see
figure III.D.1.
11. Replace the bottom of the RA, and secure to surface you
have selected for its installation.
12. Carefully reinstall the controller front panel, bottom first.
Ensure no wires have been pinched between enclosure
and front panel.
13. Reinstall the two (2) screws and tighten.
14. To operate, turn power ON.
*NOTE: If the RA does not sound off;
1. Check the polarity of the extension wire connections.
2. Be sure the controller is actually switching (relay will
click).
FULL SCALE
PUSH TO TEST
BK WT RD GN NU R- R+
3S
HYS1SP1FS SW
CAL
}
UP
INC
DIS
-11
0-20mΩ
751 756
752 757
753 758
754 759
MYRON L
COMPANY
TRANSFORMER
115/
230
FUSE*
PWR C GD NC NO CM
MAIN MONITOR/CONTROLLER
CIRCUIT BOARD
Figure III.D.2
SPC
DEC
REMOVE TO
INSTALL
SECOND
RELAY
CHS
RA
Connector
PA RA
GND
31
IV. OPERATING PROCEDURES
The front panel illustrations, switch and indicator operational
descriptions have been provided to assist the user in identifying
and operating the AQUASWITCH I & AQUASWITCH II
Monitor/controllers.
Section IV.A provides the operational descriptions for each of the
AQUASWITCH I and AQUASWITCH II switch and indicator
controls.
Section IV.B provides the user with recommended Setup
procedures
Section IV.C provides the user with recommended checkout
procedures.
A.SWITCH AND INDICATOR CONTROLS
1. AQUASWITCH I
BANK A/B “IN USE” Indicators
The amber “IN USE” LED indicator light is ON only when its
respective Bank’s solenoid valve is energized.
BANK A
BANK B
WATER QUALITY
IN USE GOOD POOR
BANK SWITCHALARM RESET
MYRON L
COMPANY
AQUASWITCH I
BLEED
NOTE: The units have been designed so that only one (1) Bank
(either A or B) can be in use at any time.
Water Quality “GOOD” Status Indicators
The green “GOOD” status LED light indicators are ON when their
respective Bank’s water quality is acceptable or ABOVE
Controller Set Point - Resistivity (the reverse is true for
Conductivity - BELOW).
Water Quality “POOR” Status Indicators
The red “POOR” status LED light indicators are ON only when
their respective Bank’s water quality is unacceptable or BELOW
Controller Set Point - Resistivity (the reverse is true for
Conductivity - ABOVE).
“BLEED” Indicator Light
The amber “BLEED” LED indicator light is ON only when the
AQUASWITCH is in its Purge Cycle (Bleed/Process Water
valve(s) are activated).
“BANK SWITCH” Push Button
When the “BANK SWITCH” is pressed, the AQUASWITCH
automatically switches Banks, i.e. turns OFF the currently
selected Bank and turns ON the alternate Bank, i.e. Bank A to
Bank B or visa versa.
“ALARM RESET” Push Button
When the-”ALARM RESET” switch is pressed, the
AQUASWITCH will automatically turn OFF the Alarm.
2. AQUASWITCH II
The AQUASWITCH II incorporates all of the features listed
above in the AQUASWITCH I plus those listed below.
3 1/2 Digit LCD
The 3 1/2 Digit LCD provides a continuous readout of the water
being controlled. A 3 1/2 digit “backlit” LCD is available as an
option.
AQUASWITCH I FRONT PANEL
Figure IV.A.1
The red LED indicator light is ON only when the resistance of the
water is BELOW the Controller’s internal Set Point.
NOTE: The reverse is true for Conductivity.
“SET POINT CHECK” Switch
When the “SET POINT CHECK” switch is pressed, the internal Set
Point reading is immediately displayed on the LCD.
3. AQUASWITCH II OPTIONS
Piezo Alarm
Audible alarm sounds off automatically when the set point is
reached. Figure IV.A.2. shows the location of this option. See
section III.B.
Backlit Display
A 3 1/2 digit “backlit” LCD is available.
Temperature Module adds the ability to momentarily display the
solution temperature by pressing a switch mounted on the from
panel as shown in figure IV.A.2. see section III.C for details.
WATER QUALITY
PIEZO ALARM
MEGOHMS - CM
SET POINT
HIGHLOW
TEMPERATURE
AQUASWITCH II
BANK A
IN USE GOOD POOR
BANK B
BANK SWITCHALARM RESET
BLEED
MYRON L
COMPANY
“HIGH/LOW” Set Point Indicators
The green LED indicator light is ON only when the resistance of
the water is ABOVE the Controller’s internal Set Point.
TEMPERATURE
OPTION
AQUASWITCH II FRONT PANEL
Figure IV.A.2
32
CAUTION - READ FOLLOWING CAREFULLY
!
B.SETUP PROCEDURES
These Setup procedures cover (1) converting the alarm circuit to
trigger on a decreasing (conductivity) or increasing (Resistivity)
reading, (2) setting the alarm circuit set point, and (3) adjusting
the hysteresis.
The following require that the front panel be removed. While doing
so be careful not to strain the cable(s).
NOTE: A small screwdriver or tweaker will be required for several
of the following operations.
1. SET POINT CONVERSION (SPC) /
REVERSING SET POINT
The alarm/control circuit on all Resistivity AQUASWITCH II
Monitor/controllers are configured to trigger the alarm relay as the
Resistivity reading decreases.
The alarm/control circuit on all Conductivity/TDS AQUASWITCHII Monitor/controllers are configured to trigger the alarm relay as
the conductivity/TDS reading increases.
If the user's application requires it, the alarm circuit may be easily
reconfigured to trigger the alarm relay as the conductivity (or
ppm) reading decreases or increases for Resistivity. Refer to
figure V.A.1. for the locations of the jumpers referred to in this
section.
NOTE: These instructions describe the general procedures for
converting the AQUASWITCH II Monitor/controller without
reference to jumper numbers or orientation.
2. SET POINT ADJUSTMENT
The set point setting is based upon the user's particular water
purity specifications or requirements.
1. Being careful not to strain the cable, unfasten and
remove the Monitor's front panel.
2. While depressing the "SET POINT" switch, turn
the Set Point #1 adjustment screw (See figure V.A.1)
until the desired set point value is indicated on the
display.
3. If Hysteresis adjustment is required see below. If not,
reinstall the front panel and tightly secure both retaining
screws.
3. HYSTERESIS (DEAD BAND) ADJUSTMENT
The hysteresis or dead band is approximately ±3% of the set point
at full scale as it leaves the factory. Under normal (most)
conditions it will not be necessary to adjust.
However, if you desire to make an adjustment please keep the
following in mind.
The adjustment is very simple and is based on set point location.
If the set point* is in the upper 75-100% of the scale, the
hysteresis control pot should be turned fully to the right.
If the set point is in the lower portion of the scale, i.e. 5-25% of
scale, the control pot may be turned fully to the left.
If you are operating in the center, 25-75% of scale, the the control
pot may be adjusted in the center.
Or the hysteresis or dead band may be adjusted to tighten the
control of a particular process.
CAUTION: adjusting the hysteresis too tight may cause the
alarm to fluctuate (on-off) due to flow, chemical mixing or bubbles
causing the relay to chatter. This condition is to be avoided, it
could damage your valves, pumps, etc. and will eventually
damage the relay.
1.Ensure power is OFF.
2. Locate the jumper block for the alarm to be configured.
3. Make a note of the current orientation of the jumpers.
4. Remove both jumpers. This is easily done by hand. Take
care not to crush the jumpers if using pliers.
5. Rotate the jumpers 1/4 turn and reinstall them on their
posts as shown in figure IV.B.1.
INCREASINGDECREASING
CONDUCTIVITY/TDS
DECREASINGINCREASING
RESISTIVITY
JUMPER CONVERSION
Figure IV.B.1
The following is assuming the front panel is already removed and
the set points have been set, if not see above.
1. Locate the hysteresis control adjustment located next to
the Set Point #1 (SP1) adjustment - it is a single turn pot.
See figure V.A.1.
2. Adjust as described above or as desired.
C. OPERATIONAL CHECKOUT PROCEDURES
The following checkout procedures are used to verify that the
AQUASWITCH I or AQUASWITCH II is operating properly.
To complete the checkout procedure, it is assumed that the
system is fully installed and the AQUASWITCH I is connected
to a reliable Monitor/controller or an AQUASWITCH II is
installed. It is also assumed that two fresh Dl banks or RO system
are on line with the system and the Monitor/controller is indicating
water quality in the desired range.
During this test procedure, it will be necessary to simulate GOOD
and POOR water quality. This is done by setting the Set Point to
BELOW or ABOVE the Monitor/controller’s actual water reading.
When the Set Point is set BELOW (counterclockwise), the monitor
will simulate “GOOD” water quality. “POOR” water quality is
simulated by setting the Set Point ABOVE (clockwise) the actual
water quality.
1. AQUASWITCH II (ONLY)
For AQUASWITCH II users, the term “Controller” refers to the
ASII Monitor/controller Circuit Board functions.
33
1.Using the “SET POINT CHECK” switch, verify that the
display indicates the desired Set Point.
NOTE: If the display does not indicate the correct Set Point
setting, reference Section IV.B.2 and figure V.A.1 for the
procedures to readjust the AQUASWITCH II’s internal Set
Point.
2. Observe reading of water being monitored.
3. With front panel removed, press “SET POINT CHECK”
switch and observe the reading on the display. With a
small screwdriver, adjust the Set Point trimmer (See
figure V.A.1) to above the reading observed in step
2. Release the “SET POINT CHECK” switch and observe
that the red “POOR” light is ON.
4.Press the “SET POINT CHECK” switch and adjust screw
adjustment until the Set Point is below the reading
observed in step 2.
5.Release the “SET POINT CHECK” switch and observe
that the green “ABOVE” light is ON. Also note the audible
click of the relay as the lights change.
6.Readjust the Set Point adjustment to the desired Set
Point setting and replace front panel. This completes the
AQUASWITCH II Monitor/controller Board checkout.
2. AQUASWITCH I and AQUASWITCH II
For AQUASWITCH II users, the term “Controller” refers to the
ASII Monitor/controller Circuit Board functions.
1.Turn the Purge Cycle trimmer adjustment screw fully
clockwise to obtain an approximate one minute cycle.
(See figure V.A.1)
2.Adjust Controller “Set Point” to simulate “GOOD” water
quality (for the AQUASWITCH II, turn the Set Point
trimmer adjustment screw counterclockwise (see
Section IV.B.2 and See figure V.A.1 )) and verify the
following sequence of operation.
3.Push “BANK SWITCH” several times to verify the
following Controller operations.
a.Verify that the “IN USE” indicator lights switch ON
and OFF between Banks A and B each time “BANK
SWITCH” is pressed.
b.Verify that both Banks A/B “GOOD” water quality
lights remain ON.
c.Verify that the “BLEED” light remains OFF.
d.Leave BANK A ON for next step. (If required, push
the “ALARM RESET” switch to reset alarm.)
valve should close and the alarm should be
activated. The system is now entirely shut down.
NOTE: If an external alarm is connected, it will be activated when
the AQUASWITCH switches Banks. (Press the “ALARM
RESET” switch to turn OFF alarm.)
If an alarm is not connected, an alarm condition can be monitored
by connecting an ohm meter (DVM) across Power Board Alarm
Terminals COM and NO. The meter should indicate a CLOSED
CIRCUIT when in the alarm condition.
5.Adjust the Controller Set Point to simulate “GOOD” water
quality. (For the AQUASWITCH II, turn Set Point trimmer
adjustment screw counterclockwise and verify the
sequence of operation.)
6.Press the “BANK SWITCH.” The BANK A “IN USE” and
“GOOD” water quality indicator lights should turn ON.
Press the “BANK SWITCH” again and BANK B “IN USE”
and “GOOD” water quality indicator lights should turn
ON.
7.Complete the following steps to return the
AQUASWITCH to its normal operating status.
a.Using the “BANK SWITCH,” select the desired Bank “IN
USE” status.
b.Adjust the Controller “Set Point” to the desired minimum
water quality. (Refer to Section IV.B.2 and figure V.A.1. )
c.Set Purge Cycle to the desired time delay. (Refer to
Section V.B)
REASSEMBLY
1. Carefully reinstall the front panel, bottom first. Ensure no
wires have been pinched between enclosure and front
panel.
2.Reinstall the two (2) screws and tighten.
3.To operate, turn power ON.
NOTE: If the Logic Module Alarm Reset jumper has been
installed to provide the user with the optional Manual Bank Switch
alarm condition, the alarm will be activated each time the “BANK
SWITCH” is pressed. Otherwise, the alarm will be activated only
when the Purge Cycle time setting has expired. (Refer to Section
V.C)
4.Adjust Controller Set Point to simulate “POOR” water
quality. (For the AQUASWITCH II, turn Set Point
trimmer adjustment screw clockwise, and verify the
following sequence of operation.)
a.BANK A water quality indicates “POOR.”
b.Yellow “BLEED” light turns ON and water is purged
from system. (This should continue for about 1
minute.)
c.After 1 minute, BANK A will turn OFF and BANK B
will turn ON and indicate “POOR” water quality.
d.After approximately 1 minute, BANK B should turn
OFF, the “BLEED” light should turn OFF, the Bleed
34
V.COMPONENT IDENTIFICATION,
CALIBRATION AND PREVENTIVE CARE
This section provides detailed illustrations to assist the user in
identifying the AQUASWITCH I and AQUASWITCH II
primary components and cable connector designations.
Section A. Provides the user with the Primary Component
Identification.
Section B. Provides the user with the Purge Cycle Calibration
Procedure.
Section C. Provides the user with the Alarm Reset Mode
Calibration Procedure.
Section D. Provides the user with the AQUASWITCH II
Calibration Procedures.
Section E. Provides the user with the Myron L Company
recommended Preventive Care Procedures.
Section F. Provides the user with Alternative Valve
Configurations.
AQUASWITCH II
AQUASWITCH II MAIN MONITOR/CONTROLLER BOARD
BK WT RD GN NU R- R+
RANGE
MODULE
3S
}
UP
-11
0-20mΩ
751 756
752 757
753 758
754 759
MYRON L
COMPANY
TRANSFORMER
115/
230
CALIBRATION
CONTROL
CAL
FUSE*
PWR C GD NC NO CM
INC
DIS
HYS1SP1FS SW
SPC
DEC
REMOVE TO
INSTALL
SECOND RELAY
DISPLAY
HARNESS
CONNECTOR
DISPLAY
CALIBRATION
CONTROL
4-20
CONNECTOR
E
N
C
L
O
S
U
R
E
A. PRIMARY COMPONENT IDENTIFICATION
As identified in Section IV, the AQUASWITCH I and
AQUASWITCH II switch and indicator components are
mounted directly to the front panel.
The AQUASWITCH I Power Board is contained within and
mounted to the back of the enclosure.
The AQUASWITCH II Main Monitor/controller Circuit Board and
Power Board are mounted to the back of the enclosure. Some
models with options have an additional circuit board mounted
behind the front panel, i.e. Model ASIIR-4A (4-20 Module).
Review figure V.A.1 below to familiarize yourself with the different
circuit board and component locations.
AQUASWITCH I
AQUASWITCH POWER BOARD
AQUASWITCH™ POWER
ALARM OUTPUT
CONNECTORS
BOARD
LOGIC
BOARD
CONNECTION
RELAY
NO
COM
TRANSFORMER
NCNOCOM
GRDCPWR
BK B
BK A
BLD
COM
NC
WHT
BK B
115/230
JUMPERS
OR
115/230
SWITCH
FUSE*
FUSE
1/8 Amp
AGC 1/8
*115/230VAC
FUSE
100mA
(T.10A) 5X20
Slow Blow/
Time Delay
*24 VAC/24 VDC
FUSE
250mA
(T.25A) 5X20
Slow Blow/
Time Delay
F
R
O
N
T
P
A
N
E
L
AQUASWITCH II DIGITAL DISPLAY
DISPLAY
DISPLAY CONNECTOR
DISPLAY
HARNESS
RED
TO MONITOR/
CONTROLLER
CIRCUIT BOARD
DISPLAY
CONNECTOR
LOW
AQUASWITCH LOGIC BOARD
SET POINT
SWITCH
GREEN
PRIMARY COMPONENT IDENTIFICATION
Figure V.A.1
STANDARD
ALARM
RESET
ENABLE
JUMPER
CABLE
CONNECTS TO
POWER BOARD
PURGE CYCLE
CONTROL
35
B.“PURGE CYCLE” CALIBRATION
PROCEDURE
The Purge Cycle control adjustment (see figure V.A.1) determines
the time span that the Bleed and Process solenoid valves are
energized during either a Bank A or B Purge Cycle.
Turning the adjustment screw fully clockwise equals a minimum
Purge Cycle of approximately one (1) minute. Turning the
adjustment screw fully counterclockwise equals a maximum
Purge Cycle of nine (9) to thirteen (13) minutes.
AQUASWITCH LOGIC BOARD
SET FULLY
COUNTERCLOCKWISE
FOR 9-13 MINUTE
PURGE CYCLE
113
SET FULLY
CLOCKWISE
FOR ONE (1) MINUTE
PURGE CYCLE
1.Being careful not to excessively strain the cable(s),
unfasten the enclosure’s front panel.
2.Turn the Logic Board adjustment to the desired Purge
Cycle setting. See figures V.A.1 and V.B.1.
C.“ALARM RESET” MODE CHANGE PROCEDURE
The AQUASWITCH I and AQUASWITCH II are shipped with
an Alarm condition that occurs whenever the “IN USE” Banks are
switched automatically.
This section provides the user with the procedures for activating
an Alarm condition every time the “BANK SWITCH” is manually
pressed.
NOTE: The Manual Bank Switch alarm condition is in addition to
the standard Automatic Bank Switch alarm condition.
To activate the Manual Bank Switch alarm condition, the user
must reinstall the Logic Module Alarm Reset jumper as shown in
figure V.C.1.
1.Being careful not to excessively strain the cable(s),
unfasten the enclosure’s front panel.
2.Remove the Logic Board Jumper and reinstall as shown
in figure V.C.1.
*115/230VAC
FUSE
100mA
(+)
}
(-)
NEU
GRN
RED
WHT
BLK
MONITOR/CONTROLLER
MAIN CIRCUIT BOARD
0-10VDC
OUTPUT
SENSOR
}
(T.10A) 5X20
Slow Blow/
Time Delay
*24VAC
FUSE
250mA
(T.25A) 5X20
Slow Blow/
Time Delay
GRN
ORN
WHT
"PURGE CYCLE" ADJUSTMENT CONTROL
For alarm jumper location, see figure V.A.1
AQUASWITCH LOGIC BOARD
AUTO
RESET
"ALARM RESET" JUMPER DIAGRAM
For alarm jumper location, see figure V.A.1
AQUASWITCH™
POWER BOARD
LOGIC
BOARD
CONNECTION
NCNOCOM
Figure V.B.1
Figure V.C.1
TRANSFORMER
GRDCPWR
LINE / +DC
(BLK)
NEUTRAL / -DC
(WHT)
AUTO/MANUAL
RESET
115/230
JUMPERS
OR
115/230
SWITCH
FUSE*
}
FUSE
1/8 Amp
AGC 1/8
MAIN
INPUT
POWER
MAIN AC
INPUT
POWER
FROM AS
{
115/230
SWITCH
Figure V.A.2
NEU-WHT
LINE-BLK
FUSE
L N
CONTROL
RELAY
NO
COM
NC
ALARM
RELAY
AQUASWITCH II™
ELECTRICAL CONNECTION DIAGRAM
{
SOLENOID
VALVE
OUTPUT
POWER
{
BANK B
BANK A
BLEED
COM
BK B
BK A
BLD
WHT
COM
BLK
LINE
NEUTRAL
SOLENOID
INPUT
}
POWER
36
CAUTION - READ FOLLOWING CAREFULLY
!
D.CALIBRATION PROCEDURES
AQUASWITCH I does not require calibration.
All Myron L AQUASWITCH II Monitor/controllers are factory
calibrated prior to shipping and are ready to install without further
calibration. Calibration should be checked occasionally with the
internal Calibration/Full Scale Test switch (FS SW) to ensure
continued accuracy. The following procedures are provided in the
event that recalibration becomes necessary. The only equipment
required are a tweaker or small screwdriver, and standard
solution.
Refer to Figure V.A.1. to locate the components described in this
section.
NOTE: When opening instrument, remove front cover with care;
a ribbon cable connects the front panel and main board.
When performing calibration procedures, the
technician must take extreme care to avoid
contacting the circuitry other than the CALibration
control. Failure to do so could result in damage to
the equipment, property and/or personal injury.
1. ELECTRONIC CALIBRATION
(CIRCUIT ONLY)
This method is faster than the following method using standard
solution, but it does not include the sensor. Therefore, it should
be used only in applications where the chance of sensor
contamination or damage are slight.
NOTE: The DISplay calibration control has been set and
sealed prior to leaving the factory, and under normal conditions
will not require readjustment. This allows the customer to use the
panel meter/display to easily set the electronic Full Scale value. If
the DISplay calibration control has been adjusted it will be
necessary to, first, verify or adjust the Full Scale using a DVM as
described in “Full Scale Calibration” below, followed by adjusting
the DISplay calibration control to Full Scale on panel
meter/display.
a. Full Scale Calibration
1. Ensure power is OFF.
2. Using a standard slot screwdriver remove the two (2)
screws on the front panel.
3. Carefully wiggle the front panel to loosen the gasket and
pull gently toward you. Do not pull more than about 8
inches/20CM or you could damage the wiring harness.
4. Turn the front panel around so that the back side is
facing you and set aside.
5. Turn power ON.
6. Press and hold the Full Scale Test switch. The display
should indicate Full Scale for the particular range
selected, i.e. 0-50 ppm should indicate 50. If not, set to
Full Scale with the CALibration control.
NOTE: When setting digital models with a full scale of 2000
(1999), it is advisable to adjust Full Scale to 1990, otherwise an
overrange condition may occur.
7. Turn power OFF.
8. Re-install front panel as described below in
“REASSEMBLY” or continue.
9. To operate, turn power ON.
b. 0-10VDC Recorder Output Calibration
For output voltages other than 10VDC, substitute as required, i.e.
5.00VDC or 2.00VDC per Section II.F.2.
Requires a DVM set to DC Volts.
1. Ensure power is OFF.
2. Attach DVM to RECORDER output connection.
3. Turn power ON.
4. Press and hold the Full Scale Test switch (FS SW).
5. The DVM should indicate 10.00VDC.
6. If not, adjust CALibration control until DVM indicates
10.00VDC.
7. Verify the panel meter/display, it should indicate Full
Scale reading as shown on Range Module.
8. If not, adjust DISplay control until a full scale reading is
displayed. When setting digital models with a full scale of
2000 (1999), it is advisable to adjust Full Scale to 1990,
otherwise an overrange condition may occur.
9. Turn power OFF.
10. Re-install front panel as described below in
“REASSEMBLY” or continue.
11. To operate, turn power ON.
2. CALIBRATION USING
STANDARD SOLUTION
The best method of recalibrating your conductivity/TDS
AQUASWITCH II Monitor/controllers is with NIST traceable
Standard Solution (available from the Myron L Company).
Because it includes the sensor, the entire system is recalibrated.
NOTE: Since standard solution calibrations are NOT practicable
with Resistivity models, another means of verification or
calibration of resistivity models is to use the transfer standard
method, using a hand-held or portable instrument capable of
resistivity measurements, i.e. the Myron L Ultrameter™. See
section V.D.4 description.
The following procedure describes the easiest method for
standard solution calibration of your Conductivity/TDS
Monitor/controller.
1. Using a standard solution which is 60-90% of full scale
of the instrument, i.e. for 0-20 µS range use KCl-18,
rinse thoroughly and fill a clean glass beaker with the
standard solution.
2.Place sensor in the beaker of standard solution. The
level of standard solution must be high enough to cover
at least 1/2" above cross hole.
3.Slowly shake the sensor to remove air bubbles from
inside the sensor bore hole.
4. Allow 5-10 minutes for temperature to equilibrate. For the
quickest and the best results, both the sensor and
solution should be at the same temperature.
5.Turn power ON.
6.Read the display of the instrument. The display should
match the value and units of measure located on the
bottle of standard solution. If the reading is different,
adjust CALibration control on the main circuit board until
the reading matches the solution value. This will require
removal of the front panel. For removal, see section
V.D.1.a, 1-4.
7.After adjustment, turn power OFF.
8.Re-install front panel as described below in
“REASSEMBLY”.
9.To operate, turn power ON.
37
3. SENSOR SUBSTITUTE CALIBRATION
NIST traceable Sensor Substitutes are commonly use to verify
and calibrate Resistivity Monitor/controllers. Normally they are
not needed due to the “built-in “ electronic calibration or “Full
Scale Test”. However, your requirements may be such that a
crosscheck or verification is required. Sensor Substitutes are
available from the Myron L Company, see accompanying chart,
figure V.D.1, for part number.
8. Display reading should be full scale of range. If not,
adjust CALibration control to read full scale, i.e. 0-50 KΩ
range should indicate 50, 0-10.00 MΩ = 10.00, and 0-
20.00 MΩ = 20.00 at full scale.
9. After adjustment, turn power OFF.
10. Re-install front panel as described below in
“REASSEMBLY”.
11. To operate, turn power ON.
If the proper Resistivity Sensor Substitute is not readily available
and you can not wait for one to be delivered, one may be
constructed using the equivalent resister values listed on the
accompanying chart, figure V.D.1 and schematic, figure V.D.2.
NOTE: If you have previously performed a system calibration
with either a NIST Standard Solution, or using the transfer
standard method, using this procedure will make that calibration
invalid. You must decide which is more important, a system
calibration, or an electronic calibration.
1. Ensure power is OFF.
2. Using a standard slot screwdriver remove the two (2)
screws on the front panel.
3. Carefully wiggle the front panel and pull gently toward
you. Do not pull more than about 8 inches/20CM or you
could damage the wiring harness.
4. Turn the front panel around so that the back side is
facing you and set aside.
5. Locate and remove the sensor leads from the sensor
connector as shown in figure V.A.3.
6. Install Sensor Substitute with label toward transformer
as shown in figure V.D.3.
7. Turn power ON.
RESISTIVITY MONITOR/CONTROLLER
RANGEPART NUMBERRESISTER VALUE - X*
CS-11
20 MEG
RESISTIVITY
SENSOR
SUBSTITUTE
NIST TRACEABLE
INSTALL WITH
LABEL TOWARD
TRANSFORMER
MYRON L
COMPANY 5-01
SENSOR
SUBSTITUTE
LABEL
TRANSFORMER
SENSOR SUBSTITUTE
LABEL TOWARD
TRANSFORMER
(+)
0-10VDC
}
OUTPUT
(-)
NEU
GRN
RED
WHT
BLK
SENSOR
}
LINE-BLK/+DC
NEU-WHT/-DC
GND-GRN
NC
NO
COM
}
MAIN
INPUT
}
POWER
ALARM
CONTROL
RELAY
0-20 MCS-111 M
0-10 MCS-1250 K
0-5 MCS-1320 K
0-2 MCS-14100 K
0-1 MCS-155 K
0-500 KCS-1624.9 K
0-200 KCS-1710 K *1%
Figure V.D.1
5.49K 1%
100K 1%X
BK WT RD GN NU R- R+
RESISTIVITY
BK WT RD GN NU R- R+
CONDUCTIVITY/TDS
NC
10K 1%X
Figure V.D.2
L N
115/230
SWITCH
ELECTRICAL CONNECT DIAGRAM
FUSE
CHASSIS GROUND for
OEM INSTALLATIONS ONLY
Figure V.D.3
4. TRANSFER STANDARD METHOD
For maximum accuracy of any AQUASWITCH II
Monitor/controllers, the transfer standard method should be
utilized. Instead of removing the entire Monitor/controller and
sensor, and either returning it to the manufacture or sending it to
a third party laboratory for recertification, the transfer standard
allows quick recertification and return to service — less down
time. While being the most accurate method it is also very easy to
perform, and may be used to calibrate ANY manufacture
Monitor/controller — resistivity or conductivity/TDS. This method
still has the benefit of third party verification, if so desired.
1.A high quality hand-held instrument, one capable of
accurate conductivity/TDS and/or “resistivity” readings
such as the Myron L Ultrameter™, is calibrated using a
standard solution, or if so desired, sent to a third party
laboratory for calibration & certification. Preferably the
standard solution should be as close as practicable to
38
the area of the AQUASWITCH II Monitor/controllers operation,
i.e. for a conductivity AQUASWITCH II
Monitor/controllers that is operating between 15 and
20µS, the recommended Myron L standard solution is
KCl-18.
NOTE: For resistivity values, the Myron L Company can supply
NaCl standards as low as ~25µS on special order. Extremely low
values are available from other sources. Calibrated at 25µS the
Ultrameter is capable of very accurate resistivity measurements.
2.Once the Ultrameter is calibrated, the process water is
measured. (For resistivity measurements, follow the
recommended procedure described in the Ultrameter
Operation Manual, section IV.B, Measuring Resistivity.)
3.Finally, the AQUASWITCH II Monitor/controllers is
calibrated/adjusted to match the value measured by the
Ultrameter by adjusting the Main CALibration control
shown in figure V.A.1.
4.Turn power OFF.
5.Re-install front panel as described below in
“REASSEMBLY”.
REASSEMBLY
1.Carefully reinstall the front panel, bottom first.
Ensure no wires have been pinched between
enclosure and front panel.
2.Reinstall the two (2) screws and tighten.
3.To operate, turn power ON.
E.PREVENTIVE CARE
The Myron L Company recommends that the following Preventive
Care procedures be observed.
1. Try to prevent exposure to excessive heat and moisture.
2. The AQUASWITCH II Monitor/controllers's main AC
power source should be protected against excessive
voltage "spikes."
3. Take care not to damage the Monitor during handling.
F.ALTERNATIVE VALVING CONFIGURATIONS &
OPERATION for CONTINUOUS
PROCESS WATER
The following methods will provide an uninterrupted water supply
to process if a fresh Dl bank is NOT in place when the other Dl
bank exhausts. Refer to figures II.A.1, II.E.1 & II.E.2.
OPTION 1
Requirements:
Continuous process. No interruptions allowed.
Connections:
“A” output (BANK A) to “A” valve AND “B” valve.
“B” output not connected.
Bleed not connected.
Omit bleed valve.
Configure “A” valve Normally Closed & “B” valve Normally Open.
OPTION 2
Requirements:
Continuous process. Holding tank between Dl beds and process.
Purge ok but no prolonged shutdown.
Connections:
Bleed output (BLEED) connected to bleed valve.
Bleed configured as Normally Closed.
Other connections same as Option 1.
OPTION 3
Requirements:
Continuous process. Purge required only at system startup.
Connections:
Same as Option 2.
Add manual valve in line with Bleed Valve.
NOTE: Daily, weekly or monthly maintenance schedules are
based upon the frequency of use and the severity of the Monitor's
environment and operating conditions.
4.Repeat the AQUASWITCH II Monitor/controllers's
Check-Out procedures to verify satisfactory operation
and/or isolate possible troubleshooting symptoms.
5. Check all cable connections to ensure that they are free
of moisture and contamination.
6. Inspect and replace damaged component boards and
cable assemblies.
7. Periodically remove, clean and inspect the Sensor
(ASII).
39
VI. OPTIONS & ACCESSORIES
A.OPTIONS ORDERED WITH MONITOR/CONTROLLER
PART #DESCRIPTION
B.OPTIONS & ACCESSORIES ORDERED SEPARATELY
PART #DESCRIPTION
4AO4-20mA Self-Powered Isolated Output Module (ASII only)
PAOPiezo Electric Alarm Only (ASII only)
35BLO3 1/2 Digit Backlit Liquid Crystal Display (LCD) (ASII only)
TPOTemperature Module 0-200°C, Requires -TP Sensor (ASII only)
TPCCalibration Module (Temperature Module)
PCOPowercord 115 VAC (8 ft. with USA plug and Strain Relief) - NOT for 230 VAC use
SMP50Surface Mounting Plate for AQUASWITCH I with two (2) 1/4-20 x 3/8 screws
SMP60Surface Mounting Plate for AQUASWITCH II with two (2) 1/4-20 x 3/8 screws
RARemote Alarm - RA™ (ASII only)
RAW200Wire, 200 ft/60 meters, Remote alarm
VRPowersupply, 24 VAC, 20VA. (115 VAC, Indoor use Only)
024-124 VAC 1” Solenoid Valve
CS-1120 Megohm Resistivity Calibration Module (Sensor Substitute) NIST Traceable
CS-142 Megohm Resistivity Calibration Module (Sensor Substitute) NIST Traceable
CS-1720 Kilohm Resistivity Calibration Module (Sensor Substitute) NIST Traceable
CC or CRNIST Certificate without Sensor C = Conductivity/TDS, R = Resistivity
CCSNIST Certificate with Sensor in Standard Solution
40
C.STANDARD SOLUTIONS & BUFFERS
1. CONDUCTIVITY/TDS
STANDARD SOLUTIONS
Your Conductivity/TDS Monitor/controller has been factory
certified to meet certain specifications based on the appropriate
Myron L Company NIST traceable standard solutions. These
same solutions are available to you.
DESCRIPTION
The Myron L Company manufactures three basic types of
conductivity/TDS standard solutions — KCl, NaCl and 442™. See
below.
Most Myron L standard solution bottles show three values
referenced at 25°C:
Conductivity in microsiemens/micromhos, while the ppm/TDS
equivalents are based on our 442 Natural Water™, and NaCl
standards. All standards are within ±1.0% of reference solutions.
These same buffers may be used for calibration and
recertification of your Myron L or other handheld instrument.
a. Potassium Chloride (KCl)
The concentrations of these reference solutions are calculated
from data in the International Critical Tables, Vol. 6.
b. 442 Natural Water™
442 Natural Water Standard Solutions are based on the following
salt proportions: 40% sodium sulfate, 40% sodium bicarbonate,
and 20% sodium chloride which represent the three predominant
components “anions” in freshwater. This salt ratio has
conductivity characteristics approximating fresh natural waters
and was developed by the Myron L Company over three decades
ago. It is used around the world for measuring both conductivity
and TDS in drinking water, ground water, lakes, streams, etc.
c. Sodium Chloride (NaCl)
This is especially useful in sea water mix applications, as sodium
chloride is its major salt component. Most Myron L standard
solution labels show the ppm NaCl equivalent to the conductivity
and to ppm 442 values.
b. pH Sensor Storage Solution
Myron L Storage Solution prolongs the life of the pH sensor. It is
available in quarts and gallons. Order SSQ or SSG.
ALL standard solutions and buffers:
$16.00US per quart/liter
$58.00US per gallon
Prices as of July 1, 2001. Subject to change—call for latest
prices.
Order example: 442-3000QT
KCL-7000GAL
C.1.a. Potassium Chloride Standards
KCl-18
KCl-180
KCl-1800
KCl-18,000
KCl-70
KCl-700
KCl-7000
KCl-70,000
C.1.b. 442 Natural Water Standards
442-15
442-150
442-1500
442-15,000
442-30
442-300
442-3000
442-30,000
442-500
442-1000
2.pH and ORP BUFFER SOLUTIONS
For your pH and/or ORP Monitor/controller. Your pH and ORP
Monitor/controller requires the same care and calibration as your
Conductivity/TDS Monitor/controller. With the following Myron L
Company NIST traceable buffer solutions it is possible for you to
maintain the accuracy required in your process.
NOTE: pH and/or ORP Monitor/controllers ORP ZERO is
calibrated using 7pH buffer. SPAN may be calibrated using any
high quality ORP mV solution.
These same buffers may be used for calibration and
recertification of your Myron L or other handheld instrument.
a. pH Buffer Solutions
pH buffers are available in pH values of 4, 7 and 10. Myron L
Company buffer solutions are traceable to NIST certified pH
references and are color-coded for instant identification. They are
also mold inhibited and accurate to within ±0.01 pH units @ 25°C.
Order 4, 7 or 10 Buffer.
C.1.c. Sodium Chloride Standards
NaCl-12.5
NaCl-13.4
NaCl-14.0
C.2.a. pH Buffer Solutions
pH4.0
pH7.0
pH10.0
C.2.b. pH/ORP Storage Solution
SSQ - Quart/liter
SSG - Gallon
Customer Standard Solutions available by special order @
$90.00 per Gallon. Contact us with your request.
41
VII. REPLACEMENT PARTS
PRICES AS OF July 1, 2001
PRICES SUBJECT TO CHANGE — contact the Myron L Company for the latest prices.
Prices are in US dollars.
PART #CIRCUIT BOARD ASSEMBLIESPRICE
XXXAQUASWITCH I & AQUASWITCH II Power Board $XXX.00
XXXAQUASWITCH I & AQUASWITCH II Logic Board $XXX.00
753IIRBASIIR RESISTIVITY Monitor/controller Circuit Board $301.00
758IICBASIIC CONDUCTIVITY Monitor/controller Circuit Board $290.00
(Must select RANGE MODULE from RANGE SELECTION GUIDE, see page 8) NO ADDITIONAL CHARGE
24VA24 VAC Isolated Power, add to above prices$18.00
24VD24 VDC Isolated Power, add to above prices$45.00
If VOLTAGE is not specified 115/230 VAC will be supplied.
RANGE MODULES SELECT from RANGE SELECTION GUIDE, see page 8
RMXXX Range Module Only$16.00
RMXXXDWith Overlay Labels$18.00
DIGITAL DISPLAYS
353 1/2 Digit without Bezel $70.00
35BL3 1/2 Digit Backlit without Bezel$130.00
HARNESS (STANDARD 9 in./229 mm)
XXHarness, Power to Logic Board$XX.00
DHHarness, ASIIR or ASIIC Monitor/controller$73.00
ENCLOSURE
ASIECAQUASWITCH I Rear Housing (BOX)$36.00
ASIIECAQUASWITCH II Rear Housing (BOX)$XX.00
FRONT PANELwith UPPER and LOWER LABELS ONLY
ASIFPAQUASWITCH I Front Panel $60.00
ASIIFPAQUASWITCH II Front Panel $XX.00
OPTIONS & ACCESSORIES
4AMO4-20mA Self-Powered Isolated Output Module $108.00
4AHHarness, 4A (4-20mA) $32.00
TPMOTemperature Module 0-200°C $116.00
TPAMOAbove Temperature Module with Adjustable Hysteresis (SPECIAL ORDER)$118.00
TDHHarness, TP and TPA (Temperature Module)$44.00
TPCCalibration Module (Temperature Module)$119.00
F100Fuse, 115/230 VAC Monitor/controller & Power Board (late), 100mA (T.10A) 5 X 20 Slow Blow/Time Delay $2.00
F250Fuse, 24 VAC Monitor/controller & Power Board (late), 250mA (T.25A) 5 X 20 Slow Blow/Time Delay $2.00
F18Fuse, 115/230 VAC, 1/8 amp, AGC 1/8, Power Board (early - with open fuse)$1.12
024-124 VAC 1” Solenoid Valve$65.00
42
VIII. LIMITED WARRANTY
All Myron L Company AQUASWITCH I Controllers & AQUASWITCH II Monitor/controllers and sensors have a Two Year
Limited Warranty. If any AQUASWITCH or sensor fails to function normally, return the faulty unit to the factory prepaid. If, in
the opinion of the factory, failure was due to materials or workmanship, repair or replacement will be made without charge.
A reasonable service charge will be made for diagnosis or repairs due to normal wear, abuse or tampering. Warranty is
limited to the repair or replacement of Monitor/controller or sensor only. The Myron L Company assumes no other
responsibility or liability.
MYRON L COMPANY
2450 Impala Drive
Carlsbad, CA 92010-7226
USA
Tel: 760-438-2021
Fax: 760-931-9189
www.myronl.com
Custom Monitors/controllers available, contact us with your special needs. Price and delivery upon request.
TERMS & CONDITIONS (additional information available upon request)
ALL SPECIAL ORDER ITEMS ARE NON-RETURNABLE AND NON-REFUNDABLE.
Minimum order $25.00. All prices are US dollars and are F.O.B Carlsbad, CA. USA.
Terms: 1% 10 days, net 30 days upon approved credit.
Credit Cards Accepted: Visa, MasterCard and American Express
Because of our policy of continuous product improvement, the Myron L Company
reserves the right to make changes in design, specifications, and prices without notice.
43
IX.GLOSSARY
442™An Internationally recognized “natural water”
standard developed by the Myron L Company in
1964.
AlgorithmA procedure for solving a mathematical
problem. See Addendum, Temperature
Compensation and TDS Derivation.
Anions Negatively charged ions
DVMDigital Volt Meter
HysteresisDead Band — related to alarm set
point. The amount of delay or overlap
between change from high to low or low
to high.
LCDLiquid Crystal Display
KClPotassium Chloride — used as a
standard for many applications.
Normally, micromhos or microsiemens
are the units of measure.
systems controlled using equipment without
temperature compensation. Requires a
separate temperature measuring device
(temperature measurements may be part of
Cond/TDS, Resistivity or pH device but must
not correct measurements for temperature
errors).
NaClSodium Chloride — used as a
standard for sea water and other
applications.
NISTNational Institute Standards & Testing
PLCProgrammable Logic Controller
PPM Parts Per Million — common units of measure
for TDS.
PPTParts Per Thousand — common units of
measure for TDS.
SCADASupervisory Control And Data Acquisition
Tempco (TC) Temperature Compensation — See Addendum,
Temperature Compensation.
TDSTotal Dissolved Solids or the Total Conductive
Ions in a solution — normally displayed as ppm
or ppt. See Addendum, Conductivity
Conversion to TDS.
TweakerA screwdriver specially designed for adjusting
electrical calibration controls.
µM or µSMicromho or microsiemen — common unit of
measure for conductivity.
Meg or ohm Megohm — common unit of measurement for
resistivity.
USP25United States Pharmaceutical regulation,
revision number 25. In part, requires
Conductivity/TDS, resistivity or pH
measurements be taken, recorded and/or
44
X. NOTES
45
XI.ADDENDUM
A. CONDUCTIVITY, TDS, RESISTIVITY and
TEMPERATURE RELATIONSHIPS
1.TEMPERATURE COMPENSATION
(Tempco) of Aqueous Solutions
Electrical conductivity indicates solution concentration and
ionization of the dissolved material. Since temperature greatly
affects ionization, conductivity measurements are temperature
dependent and are normally corrected to read what they would be
at 25°C.
a. Standardized to 25°C
Conductivity or Resistivity are accurately measured in the
AQUASWITCH II by a method that ignores electrolysis,
electrode characteristics, etc., and uses a unique circuit to
perform temperature compensation. In simpler instruments,
conductivity values are usually assigned an average correction
similar to KCl solutions for correction to 25°C. The correction to an
equivalent KCl solution is a standard set by chemists. It
standardizes the measurements and allows calibration with
precise KCl solutions recognized for stability.
b. Tempco Variation
Most conductivity instruments use an approximation of the
temperature characteristics of solutions, perhaps even assuming
a constant value. The value for KCl is often quoted simply as
2%/°C. In fact, KCl tempco varies with concentration and
temperature in a non-linear fashion. Other solutions have more
variation still. The AQUASWITCH II uses corrections that
change with concentration and temperature instead of single
average values. See Chart 1.
2.500%
2.400%
a. How it’s Done
Once the effect of temperature is removed, the compensated
conductivity is a function of the concentration (TDS).
Temperature compensation of the conductivity of a solution is
performed automatically by the electronic circuit, using data
derived from chemical tables. Any dissolved salt at a known
temperature has a known ratio of conductivity to concentration.
Tables of conversion ratios referenced to 25°C have been
published by chemists for decades.
b. Solution Characteristics
Real world applications have to measure a wide range of materials
and mixtures of electrolyte solutions. To solve this problem,
industrial users commonly use the characteristics of a standard
material as a model for their solution, like the KCl favored by
chemists for its stability.
Users dealing with sea water, etc., commonly use NaCl as the
model for their concentration calculations. Users dealing with
freshwater work with mixtures including sulfates, carbonates and
chlorides, the three predominant components (anions) in
freshwater that Myron L Company calls “natural water”. These are
modeled in a mixture called “442™” which the Myron L Company
developed and markets for use as a calibration standard, as it
does standard KCl and NaCl solutions.
c. When does it make a lot of difference?
First, the accuracy of temperature compensation to 25°C
determines the accuracy of any TDS conversion. Assume we
have industrial process water to be pretreated by RO. Assume it
is 45°C and reads 1500 µS uncompensated.
1. If NaCl compensation is used, an instrument would report 1035
µS compensated, which corresponds to 510 ppm NaCl.
2. If 442 compensation is used, an instrument would report 1024
µS compensated, which corresponds to 713 ppm 442.
2.300%
2.200%
2.100%
2.000%
1.900%
1.800%
1.700%
1.600%
1.500%
The AQUASWITCH II will provide the repeatability of data
needed for relative values for process control.
2.CONDUCTIVITY CONVERSION to TOTAL
Electrical conductivity indicates solution concentration and
ionization of the dissolved material. Since temperature greatly
affects ionization, conductivity measurements are temperature
dependent and are normally corrected to read what they would be
at 25°C (ref. Temperature Compensation).
% / °C
KCl % / °C
Temperature
0 5 10 15 20 25 30 35 40 45 50 55 60
Chart 1
DISSOLVED SOLIDS (TDS)
The difference in values is 40%.
In spite of such large error, some users will continue to take data
in the NaCl mode because their previous data gathering and
process monitoring was done with an older NaCl referenced
device.
Those who want true TDS readings that will correspond to
evaporated weight will select the correct Solution Type.
3.TEMPERATURE COMPENSATION
(Tempco) and TDS DERIVATION
When making conductivity/resistivity measurements, the Solution
Selection determines the characteristic assumed as the
instrument reports what a measured conductivity or resistivity
would be if it were at 25°C. The characteristic is represented by
the tempco, expressed in %/°C. If a solution of 100 µS at 25°C
increases to 122 µS at 35°C, then a 22% increase has happened
over this change of 10°C. The solution is said to have a tempco of
2.2 %/°C.
Another solution would have a different tempco because of its
ionization activity. And, that tempco may be a little different at a
different concentration or temperature.
NOTE: Resistivity TEMPCO’s change very drastically above
10MΩ.
46
XI.ADDENDUM Continued
4. RESISTIVITY
Resistivity is the inverse of conductivity (the reverse is also
true), i.e. 18.2 MΩ = 0.0549 µM or µS, and 0.2 µM/µS = 5 MΩ.
47
Water Quality Instrumentation
Accuracy • Reliability • Simplicity
MYRON L COMPANY
2450 Impala Drive
Carlsbad, CA 92010-7226
USA
Tel: 760-438-2021
Fax: 760-931-9189
www.myronl.com
Made in USA
ACCURACY • RELIABILITY • SIMPLICITY
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.