Myron L 512M10 User Manual

SINGLE RANGE DS METERS
Model 512M5 shown
Range “S” Button (Internal Standard)
Cell Cup
Black Button (µmhos or ppm)
Upper Electrode
Meter Zero
Adjust
Pointer
OPERATING INSTRUCTIONS
Us er ’s Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10
DESCRIPTION
This manual describes your Myron L DS meter, tells you how to use it, and how to keep it working accurately for many years.
Your model 512 single range DS Meter is a compact instrument which operates on the principal of electrical conductivity. It will quickly determine the conductivity or ppm/Total Dissolved Solids of almost any solution. Models 512M2, 512M3, 512M4, 512M5 and 512M10 have dials calibrated in micromhos (µmhos), which is equivalent to the metric microsiemens (µS). Models 512T2, 512T3, 512T4, 512T5 and 512T10 convert the conductivity directly into parts per million (ppm) of Total Dissolved Solids.
All are 3.4” x 4.5” x 4.0” (85 x 129 x 126mm) and weigh less than one pound (0.45kg). DS Meters are completely self-contained. The built-in cell is automatically temperature compensated from 50º to 160ºF (10º to 71ºC). They are powered by a 9 volt battery good for at least 2000 tests, or one year shelf life.
512T 512M
MODEL RANGE MODEL RANGE
512T2 0-50 ppm 512M2 0-50 µmhos
512T3 0-500 ppm 512M3 0-5000 µmho
512T4 0-2500 ppm 512M4 0-2500 µmho
512T5 0-5000 ppm 512M5 0-5000 µmho
512T10 0-10,000 ppm 512M10 0-10,000 µmho
The pictures in this manual show the major operating parts of your Myron L DS Meter. Handle your DS Meter and identify these parts to become familiar with it.
By using an RE-10 Range Extender (see ACCESSORIES), the maximum range of most meters may be increased ten times.
USING YOUR METER CALIBRATION
Battery Strap
Calibration Control
Lower Electrode (inside cell)
1. Rinse the cell cup three times with the sample you want to test. (For very hot or very cold samples see TEMPERATURE COMPENSATION below.)
NEVER FILL THE CELL BY DIPPING THE METER INTO WATER!
2. Fill the cell with another sample to at least 1/4” (6mm) above the upper electrode.
3. Press the black button.
4. Read the dial value indicated by the pointer to determine parts per million (ppm) Total Dissolved Solids on the 512T models or the conductivity in micromhos (µmhos) on the 512M models.
5. If the pointer goes off the scale to the right, try RANGE DOUBLING (see FIELD CALIBRATION) or an RE-10 Range Extender (see ACCESSORIES).
NOTE:
When you are nished with the meter, RINSE THE CELL CUP with clean water, preferably distilled or deionized.
TEMPERATURE COMPENSATION
For very hot or very cold solutions, let the three rinse samples each remain in the cell for several seconds. Then immediately ll the cell with the sample you want to test (step 2 above). This allows the automatic temperature compensation feature time to work properly.
CAUTION
DO NOT use with samples hotter than 160ºF (71ºC).
The readings WILL NOT be accurate.
DO NOT splash solvents such as lacquer thinner, acetone, benzene or chlorinated solvents on the plastic case.
DO NOT x or modify the meter. That will void your warranty. See SERVICE for details or consult Myron L Company.
DO NOT DIP THE INSTRUMENT INTO WATER. If water does get inside the instrument, see MAINTENANCE for instructions on drying it.
STANDARD SOLUTIONS: A Standard Solution has a known conductivity and ppm. Your meter was calibrated at the factory using a Standard Solution. You can keep your meter accurate by using the same Standard Solution. To pick the right Standard Solution for your meter see ACCESSORIES.
CHECKING CALIBRATION
1. Test a sample of appropriate Standard Solution.
CAUTION: Throw the Standard Solution away as you use it. Don’t put the used samples back in the bottle.
2. If the DS Meter does not indicate the same value as is on the Standard Solution bottle’s label, rst clean the cell. For directions on how to do this see CELL inside. Rinse the cell thoroughly and test the Standard Solution again. If the DS Meter still does not indicate the correct value, recalibrate it as described below.
TO RECALIBRATE THE METER
1. Remove the bottom cover using ngernails or a small screwdriver to loosen the front or rear edge. Identify the Calibration Control (see photo below left) so you can
nd it by touch while calibrating.
2. Test another sample of the Standard Solution (be careful to not to get any solution inside the meter).
3. Adjust the Calibration Control until the meter indicates the value that is on the Standard Solution label.
4. Press the red “S” button. Compare the meter reading to the INTERNAL STANDARD value on the meter’s bottom label. If they are noticeably different, mark the new value in place of the old one.
NOTE:
A 512T (ppm) Meter may be calibrated to read either 442™ or NaCl parts per million, but not micromhos. A 512M (conductivity) Meter may only be set to read micromhos.
INTERNAL STANDARD: An INTERNAL STANDARD value for each meter is on the label on the bottom cover of each meter. Use it between normal calibrations as a eld check of your meter’s accuracy. To verify your meter’s calibration:
1. Press the red “S” button.
2. If the reading matches the INTERNAL STANDARD value on the bottom label, your meter is in calibration. If not, see FIELD CALIBRATION.
NOTE:
The feature described above is intended as a quick eld calibration check, or for using the range doubling technique. It is not a replacement for calibration with Standard Solution.
Loading...
+ 2 hidden pages