MULTI MEASURING INSTRUMENTS CO.,LTD. MIS-2D, MIS-4D, MIS-3D Instruction Manual

DIGITAL INSULATION
INSULATION RESISTANCE TESTER
INSTRUCTION MANUAL
Thank you very much for selecting our digital insulation resistance tester.
This model is complex instrument and employs a very reliable mechanical/ electronic design.
Before use the instrument, read this instruction manual completely and familiarize yourself thoroughly with all functions.
Keep this instruction manual carefully to take out whenever you need.
MULTI MEASURING INSTRUMENTS CO., LTD
Akihabara Murai Bldg. 7F, 1-26 Kanda Sakuma-cho, Chiyoda-ku, Tokyo, Japan, 101-0025 TEL : +81-3-3251-7016 FAX : +81-3-3253-4278 Home Page: http://www.multimic.com/ E-mail: multi@multimic. com
1. FEATURES
MIS series are 3 ranges insulation resistance tester designed for testing of electrical installations and equipment in accordance with IEC international
standard.
Battery operation and heavy duty rugged case is used for high performance
insulation testing.
The big digital and bar graph LCD display with back light enabled easy observation.
Especially, useful when working in dark place.
Hand free and continuous measurements with custom made MΩ test switch.
The voltage in the circuit or capacitor can be checked by warning lamp for safe insulation measurements.
Data hold and auto power off function.
Ultra compact, light weight.
2. SAFETY SUMMARY
The CAUTIONs and WARNlNGs which appear on the following pages must be followed to ensure operator safety and to retain the operating conditions of the instrument.
Safety Symbols:
indicates the operator must refer to an explanation in this manual.
indicates terminals at which dangerous voltages may exist.
WARNING
To avoid electrical shock, use CAUTION when working with more than
60V DC or 25V rms.AC since the danger of electric shock exist. In addition, check that the test leads are normal condition.
● POSSIBLE ELECTRICALSHOCK: Do not make measurements if the case is damaged. Make sure that the terminals and the connecting cables are in a good condition and that the proper measuring function has been selected.
POSSIBLE ELECTRICAL SHOCK or FIRE HAZARD: Do not expose this instrument to rain or moisture. Do not operate the instrument in the presence of flammable gasses or fumes.
● This instrument must be used only by professionals. Any adjustment, maintenance and repair of the opened apparatus under voltage shall be carried out only by a skilled person who is aware of the hazard involved.
CAUTION Where it is likely that the protection has been impaired, the instrument
shall be made inoperative and be secured against any unintended operation. The protection is likely to be impaired if the instrument:
shows visible damage.
● fails to perform the intended measurements.
● has been subjected to prolonged storage under unfavourable conditions.
● has been subjected to severe transport stresses.
3. SPECIFICATIONS
Function: Insulation resistance, AC voltage
Display: 3 1/2 digit LCD with bar graph, max. reading of 3200 count and annunciators Response time: Less than 5 sec. (auto ranging) Data hold indication: “DH” mark on LCD readout Infinity indication: “OL (∞)” mark on LCD readout (over 3200 count) Safety standard: IEC 61010-1, CAT. II 600V phase to earth EMC standard: EN 61326 Constructional standard: IEC 61557-2 Insulation resistance: DC 500V-50MΩ or more (MIS-2D) DC 1000V-50MΩ or more (MIS-3D, MIS-4D) Withstanding voltage: AC 3700V/1minute, between input terminal and outer case Low battery indication: “B” mark on LCD readout Temperature characteristics (0~40): ±5%rdg of specified accuracy Operation temperature: 0 to 40, less than 80% RH without condensation Storage temperature: -10 to 60, less than 80% RH without condensation Power supply: 1.5V alkaline batteries (AA size, LR-6) x 6 Dimension: 170(W) x 105(H) x 54(D)mm, approx. 365g (excluding batteries)
Accessories: Line test lead .......... 1
Earth test lead ........ 1 Line test lead
Batteries ................. 6
Test leads case ...... 1 Earth test lead
Belt ......................... 1
Instruction manual .. 1 Optional accessory: Test lead with remote switch
4. MEASURING RANGES AND TECHNICAL DATA
Insulation resistance measurement
Model
MIS-2D
MIS-3D
MIS-4D
Rated voltage &
effective measuring
range
125V-20MΩ 250V-50MΩ 500V-100MΩ
250V-50MΩ 500V-100MΩ
1000V-2000MΩ
125V-20MΩ
250V-500MΩ
1000V-2000MΩ
Min. measurable resistance at rated voltage
0.125MΩ
0.25MΩ
0.5MΩ
0.25
0.5 1MΩ
0.125
0.25MΩ 1MΩ
Rated current
1mA +20% -0%
Max. no-load voltage
Rated voltage +30% -0%
Short circuit current
<2mA