![](/html/62/6223/6223f26a72508a621f8fde684ab020c8a669535df7b3891dcfb07ee0a59e2059/bg1.png)
TCW1 Warming Plate
Warming Plate for use with Temperature Controller TC01 / TC02
Warming plate for in vivo experiments with narcotized rats or mouses, and for in vitro experiments,
for example, for keeping slices or cell cultures on microelectrode arrays (MEA) warm.
Technical Specifications
Dimensions (W x D x H) 220 x 120 x 10 mm
Weight 600 g
Material Aluminium, black anodized
Integrated sensor:
Type PT 100 1/3 DIN 8
Accuracy +/- 0.1 °C
Measuring method Four wire measuring bridge
Resistance 10 -15
Heating temperature range Ambient temperature to 50 °C
Settling time +/- 2 min
Maximal temperature 105 °C
Connection to temperature controller TC01 / TC02 D-SUD 9 female connector
Cleaning: Please wipe the warming plate with a soft tissue and detergent or alcohol.
Do not immerse the plate into water!
Do not autoclave the warming plate because of the connected cable!