
Plexiglass of MEA1060-UP-BC-PGP
PG-MEA1060-UP-BC
Plexiglass of Perfusion Ground Plate for MEA1060-UP-BC Amplifier
Instructions for Changing the Plexiglass Part of the Perfusion Ground Plate
of the MEA1060-UP-BC Amplifier
Perfusion Ground Plate in original state.
O-Ring-PGP
PGP-O-Ring
Disconnect the four screws fixing the plexiglass. Disconnect the plexiglass.
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25Fax
+49-7121-9 09 25-11
info@multichannelsystems.com
www.multichannelsystems.com
0
© 2012 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.

Plexiglass of MEA1060-UP-BC-PGP
PG-MEA1060-UP-BC
Plexiglass of Perfusion Ground Plate for MEA1060-UP-BC Amplifier
Instructions for Changing the Plexiglass Part of the Perfusion Ground Plate
of the MEA1060-UP-BC Amplifier
O-Ring-PGP
PGP-O-Ring
Place the new plexiglass into the open area. Make sure the orientation of the plexiglass
is correct. Insert the tubes for the perfusion IN and the perfusion OUT into the holes.
Turn the perfusion ground plate and fix the plexiglass with the four provided screws.
Mount the tubes onto the metalic pipes for the perfusion IN and the perfusion OUT.
Scope of delivery:
1 Plexiglass Part PG-MEA1060-UP-BC
4 screws (M3x8)
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25Fax
+49-7121-9 09 25-11
info@multichannelsystems.com
www.multichannelsystems.com
0
© 2012 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.