
- 1 -
Programmable Gain Amplifier PGA
Manual

Imprint
Information in this document is subject to change without notice. No part of this document may be
reproduced or transmitted without the express written permission of Multi Channel Systems MCS GmbH.
While every precaution has been taken in the preparation of this document, the publisher and the author
assume no responsibility for errors or omissions, or for damages resulting from the use of information
contained in this document or from the use of programs and source code that may accompany it.
In no event shall the publisher and the author be liable for any loss of profit or any other commercial damage
caused or alleged to have been caused directly or indirectly by this document.
© 2018 Multi Channel Systems MCS GmbH. All rights reserved.
Printed: 26. 09. 2018
Multi Channel Systems MCS GmbH
Aspenhaustraße 21
72770 Reutlingen Germany
Phone+49-71 21-90 92 5 - 0
Fax +49-71 21-90 92 5 -11
sales@multichannelsystems.com
www.multichannelsystems.com
Microsoft and Windows are registered trademarks of Microsoft Corporation. Products that are referred to in
this document may be either trademarks and/or registered trademarks of their respective holders and should
be noted as such. The publisher and the author make no claim to these trademark.

- 3 -
Table of Content
1. Introduction ................................................................................................................. 4
1.1 About this Manual .................................................................................................... 4
2. Important Information and Instructions .................................................................... 5
2.1 Operator's Obligations .............................................................................................. 5
2.2 Guarantee and Liability ............................................................................................. 5
2.3 Important Safety Advices .......................................................................................... 6
2.4 Terms of Use for the Program ................................................................................... 7
2.5 Limitation of Liability ................................................................................................. 7
3. First Use of the PGA .................................................................................................... 8
3.1 Welcome to the Programmable Gain Amplifier ......................................................... 8
3.2 Installing the Software .............................................................................................. 9
3.3 Setting Up and Connecting the PGA ......................................................................... 9
4. Operating the PGA .................................................................................................... 11
4.1 Adjusting the Gain Settings .................................................................................... 11
4.2 Signal Amplification and Filters ............................................................................... 12
4.3 Differential Amplifier and Noise Reduction .............................................................. 13
5. Troubleshooting ........................................................................................................ 14
5.1 About Troubleshooting ........................................................................................... 14
5.2 Technical Support ................................................................................................... 14
5.3 Insufficient Power ................................................................................................... 15
6. Appendix .................................................................................................................... 16
6.1 Pin Layout ............................................................................................................... 16
6.2 Contact Information ............................................................................................... 17
6.3 Ordering Information .............................................................................................. 18

- 4 -
1 Introduction
1.1 About this Manual
This manual comprises all important information about the first installation of the hardware and software,
and about the daily work with the instrument. It is assumed that you have already a basic understanding
of technical and software terms. No special skills are required to read this manual.
If you are using the device for the first time, please read the important safety advice before installing the
hardware and software, where you will find important information about the installation and first steps.
The printed manual and help are basically the same, so it is up to you which one you will use. The help offers
you the advantage of scrolling through the text in a non-linear fashion, picking up all information you need,
especially if you use the “Index” and the “Search” function. If you are going to read larger text passages,
however, you may prefer the printed manual.
The device and the software are part of an ongoing developmental process. Please understand that the
provided documentation is not always up to date.
The latest information can be found in the "Help".
Check also the MCS web site www.multichannelsystems.com for downloading up-to-date manuals and help
files.

- 5 -
2 Important Information and Instructions
2.1 Operator's Obligations
The operator is obliged to allow only persons to work on the device, who
are familiar with the safety at work and accident prevention regulations and have been instructed
how to use the device;
are professionally qualified or have specialist knowledge and training and have received instruction
in the use of the device;
have read and understood the chapter on safety and the warning instructions in this manual and
confirmed this with their signature.
It must be monitored at regular intervals that the operating personnel are working safely.
Personnel still undergoing training may only work on the device under the supervision of an experienced
person.
2.2 Guarantee and Liability
The general conditions of sale and delivery of Multi Channel Systems MCS GmbH always apply. The operator
will receive these no later than on conclusion of the contract.
Multi Channel Systems MCS GmbH makes no Guarantee as to the accuracy of any and all tests and data
generated by the use of the device or the software. It is up to the user to use good laboratory practice to
establish the validity of his findings.
Guarantee and liability claims in the event of injury or material damage are excluded when they are the result
of one of the following:
Improper use of the device.
Improper installation, commissioning, operation or maintenance of the device.
Operating the device when the safety and protective devices are defective and/or inoperable.
Non-observance of the instructions in the manual with regard to transport, storage, installation,
commissioning, operation or maintenance of the device.
Unauthorized structural alterations to the device.
Unauthorized modifications to the system settings.
Inadequate monitoring of device components subject to wear.
Improperly executed and unauthorized repairs.
Unauthorized opening of the device or its components.
Catastrophic events due to the effect of foreign bodies or acts of God.

- 6 -
2.3 Important Safety Advices
Warning: Make sure to read the following advice prior to install or to use the device and the
software. If you do not fulfill all requirements stated below, this may lead to malfunctions or
breakage of connected hardware, or even fatal injuries.
Warning: Obey always the rules of local regulations and laws. Only qualified personnel should
be allowed to perform laboratory work. Work according to good laboratory practice to obtain
best results and to minimize risks.
The product has been built to the state of the art and in accordance with recognized safety engineering rules.
The device may only
be used for its intended purpose;
be used when in a perfect condition.
Improper use could lead to serious, even fatal injuries to the user or third parties and damage to the
device itself or other material damage.
Warning: The device and the software are not intended for medical uses and must not be used
on humans.
High Voltage
Electrical cords must be properly laid and installed. The length and quality of the cords must be in accordance
with local provisions.
Only qualified technicians may work on the electrical system. It is essential that the accident prevention
regulations and those of the employers' liability associations are observed.
Each time before starting up, make sure that the power supply agrees with the specifications of
the product.
Check the power cord for damage each time the site is changed. Damaged power cords should
be replaced immediately and may never be reused.
Check the leads for damage. Damaged leads should be replaced immediately and may never be
reused.
Do not try to insert anything sharp or metallic into the vents or the case.
Liquids may cause short circuits or other damage. Keep the device and the power cords always dry.
Do not handle it with wet hands.
Requirements for the Installation
Make sure that the device is not exposed to direct sunlight. Do not place anything on top of the device,
and do not place it on top of another heat producing device, so that the air can circulate freely.

- 7 -
2.4 Terms of Use for the Program
You are free to use the program for its intended purpose. You agree that you will not decompile,
reverse engineer, or otherwise attempt to discover the source code of the software.
2.5 Limitation of Liability
Multi Channel Systems MCS GmbH makes no guarantee as to the accuracy of any and all tests and data
generated by the use the software. It is up to the user to use good laboratory practice to establish the validity
of his findings.
To the maximum extent permitted by applicable law, in no event shall Multi Channel Systems MCS GmbH
or its suppliers be liable for any special, incidental, indirect, or consequential damages whatsoever (including,
without limitation, injuries, damages for data loss, loss of business profits, business interruption, loss of
business information, or any other pecuniary loss) arising out of the use of or inability to use the program
or the provision of or failure to provide Support Services, even if Multi Channel Systems MCS GmbH has
been advised of the possibility of such damages.

- 8 -
3 First Use of the PGA
3.1 Welcome to the Programmable Gain Amplifier
The Programmable Gain Amplifier (PGA) is used for amplifying and filtering raw data from a miniature
preamplifier, for example, a MPA32I. The PGA is a differential (I-type) filter amplifier. It operates similar to
the standard I-type filter amplifiers (FA) from Multi Channel Systems MCS GmbH, but was designed for users
who demand a higher flexibility. Whereas the gain in the standard FA is fixed, the gain of a PGA is softwareselectable, in the range of 10 to 5000, for each individual channel. This is useful when signals from different
sources with either different preamplification stages or different signal amplitudes, for example, spikes and
local field potentials, are recorded.
Gain settings for each channel are selected in the user-friendly program PGA-Control. Simply pick the desired
gain for each channel from the drop-down list and download the settings onto the amplifier via the USB port.
You can print a report sheet and save the gain configuration for later use.
Standard versions of the PGA are available with 16, 32, and 64 inputs (PGA16, PGA32, PGA64, respectively).
Standard bandwidth is 1 to 5000 Hz, which is useful for recording all signal types, such as cardiac signals,
spikes, and local field potential recordings. Signals can be separated by digital filtering in MC_Rack, the data
acquisition and analysis software from MCS for use with the ME- / MEA-System. Special versions feature an
integrated splitter, with two different pass bands for recording signals with different frequencies, for
example, for separating local field potentials from spike signals: Input signals are split and the two pass bands
(1 to 300 Hz, 300 to 5000 Hz) are sent to two separate outputs. Splitter versions are available with 16 input
and 32 output channels, or 32 input and 64 output channels (PGA1632 and PGA3264, respectively). Please
refer to chapter "Signal Amplification and Filters" for more information.

- 9 -
3.2 Installing the Software
System Requirements
Software: One of the following Windows operating systems is required: Windows 10, 8.1 and 7, Vista or
Windows XP (English and German versions supported). Other language versions may lead to software errors.
Hardware: USB Port 2.0 or USB Port 1.1
Installing the Software
Please check the system requirements before you install the software. MCS cannot guarantee that the
software works properly if these requirements are not fulfilled.
Important: Please make sure that you have full control over your computer as an administrator. Otherwise,
is possible that the installed software does not work properly.
Double-click “Setup.exe” on the installation volume. The installation assistant will show up and guide you
through the installation procedure.
Follow the instructions of the installation assistant.
3.3 Setting Up and Connecting the PGA
Note: You can use an USB hub for connecting the PGA to the computer, for example, if you have no free
USB port or if you need to extend the USB cable. Please note that other devices that are connected to the
same USB port and that send or receive continuous data streams, for example, a web cam or USB speakers,
can hinder the transmission of the gain settings from the computer.
The following instructions refer only to the operation of the PGA. For general recommendations on the
grounding and shielding of the setup, please refer to the Miniature Preamplifier Manual.
Provide a power supply, for example PS40W, and a computer with USB 1.1 port in the immediate vicinity
of the installation site.
The recommended value of supply voltage should be US = ± 6.0 V. The limits for proper operation ranges
from ± 5.7 V to ± 12.0 V. Please use the lowest possible supply voltage. The higher the supply voltage,
the more heat will be internally generated by the voltage regulators!
Note: The PGA should only be operated if the brightness of the power LEDs is stable (without flickering),
otherwise the DC shows instabilities.

4 mm black plug connector
Warning: Do not mismatch the polarity of the power supply. False connection may
damage the device.
Important: Make sure that the reference input is always connected, because signals applied to the reference
input are subtracted from the recording signal input channels. If the reference input is unconnected or
connected to an electrode in an active area (for example, it sees signals of interest), this will either lead to
signal loss or to a very high noise level. Connect the reference input to the ground of the setup when you
are not using a reference. For more information, please see Differential Amplifier and Noise Reduction.
1. Place the PGA on a stable surface, where the air can circulate freely and the PGA is not exposed
to direct sunlight.
2. Connect the input signal source, for example, a SC8x8 or SC2x32 signal collector connected
to miniature preamplifiers, to the 68-pin MCS standard input connector of the PGA.
3. Connect the 68-pin MCS standard output connector to the MC_Card input of the data acquisition
computer by a 68-pin MCS standard cable. Make sure the IPS10W as a power supply source is
properly installed to the data acquisition computer. Check the output power if you are in doubt.
For more information, please see the ME-Systems Manual.
4. If the data acquisition computer has no IPS10W, connect a suitable power supply, for example
a PS40W, to the 4 mm power input sockets. An internal power supply transforms the supply voltage
into the internal supply voltage.
5. Connect the USB connector to a free USB 1.1 port of the computer with a USB cable Type A - B.
The connector on the PGA side is type B; the connector on the computer side is type A.
The computer connection is necessary for programming the PGA, but not for operating it.
6. Switch on the PGA. The LEDs on the rear panel must light up, otherwise the power is insufficient.
7. Install the PGA-Control program provided together with the PGA to the connected computer.
The PGA driver is installed automatically as well.

- 11 -
4 Operating the PGA
4.1 Adjusting the Gain Settings
1. Switch the power supply on. The PGA’s power LED should be lighting.
2. Start the PGA-Control program.
3. Specify the gain settings for all channels by selecting the appropriate gain factor from the drop-down
menu, or load a previously saved gain configuration ("*.amp" file). A different gain level can be
selected individually for each channel. The option "Set All" can be used to define a general gain level
for all channels.
4. Always click the "Download" button to apply the settings to the PGA.
5. It is recommended to print out the gain settings for later reference by clicking the "Print" button.
6. If you want to keep the pattern for future use, save the file as an "*.amp" file.
7. Switch off the power supply after use. If the PGA is switched off, the last gain settings are stored.
When starting the PGA later, the stored settings will automatically be restored.
PGA Gain Settings Example
In the screenshot above, the gain of the PGA channel 1 is set to 10, channel 2 to 50, 3 to 100, 4 to 1000,
channel 5 to 5000, and the rest of the available channels to a gain factor of 10. The pass band of the
amplifier is displayed on top of the channel numbers.

- 12 -
4.2 Signal Amplification and Filters
The PGA combines a band pass filter and the signal amplification in one instrument.
Different filter settings are used for different applications to enhance the signal-to-noise ratio. The pass band
of the filter amplifier should be chosen according to the signal type. It is generally useful to set the upper limit
of the amplifier near the highest expected signal frequency, but also at a safe distance to make sure that the
full gain is used for signal amplification. The PGA is available with a standard bandwidth of 1 to 5000 Hz,
which is suitable for most applications.
A bandwidth of 10 to 5000 Hz is optional, which has the advantage that the amplifier is more stable towards
low-frequency noise and oscillations, but has the disadvantage that slow signal components such as plateaus
are removed from the input signal. Other bandwidth configurations are available on request. The PGA1632
and the PGA3264 are special versions that feature two different pass bands. Signals are split and the two
pass bands are send to two separate output channels. For 16 input and 32 output channels, or 32 input and
64 output channels, respectively. These amplifiers are perfect for recording signals with different frequencies:
You can record slow signals like field potentials with a bandwidth of 1 to 300 Hz on one half of the channels,
and fast signals like spikes with a pass band of 300 Hz to 5 kHz on the other half. Please refer to the label on
the amplifier to check the bandwidth of the amplifier version that you have in use.
If you have a complete ME-System with MC_Card, you can use the digital filter of the MC_Rack program
to change the pass band and filter the raw data. This way, you are much more flexible in designing your
experiments. As a further advantage, you can see the original (not filtered) data as well. This is especially
important because all filters are known to distort signals. On the other hand, you may need a higher
sampling rate to avoid aliasing, and you will have a lower signal to noise ratio.
The gain of a PGA can be adjusted from 10 to 5000. Usually, a total gain (of preamplifier and PGA)
of about 1000 to 2000 is used for extracellular recording of biological signals. For large signals, for example,
from whole-heart preparations, you need a lower gain to prevent a saturation of the amplifier.
Please note that the ratio of the output signal to the input signal, that is, the gain, is not a fixed parameter
for the complete bandwidth. The gain that was specified for the amplifier, for example, 1200 is not fully
reached at the borders of the amplifier's pass band. The general rule is, that at the lower and upper limit
of the frequency band, the gain is "Root of 2/2", that is approximately 70 %, of the full gain. Therefore,
you should use a bandwidth that is at a safe distance of the signals of interest. Outside the pass band,
the gain decreases with the frequency and finally approaches zero.
For information on the gain and filters of the MC_Card data acquisition system, please see the ME-System
Manual. For more information on gain and filters in general, please refer to standard literature or contact
your local retailer.

- 13 -
4.3 Differential Amplifier and Noise Reduction
If the reference electrode is properly placed into an inactive region near the recording site, the noise voltage
is usually a common component of the input voltages from the reference electrode and the recording
electrodes. Thus, the noise will be canceled out when the difference of the amplifier inputs is taken as
the recording signal.
As a differential amplifier, the PGA has a reference input in addition to the ground input. The power ground
and the reference input are connected to the power ground and reference output of the connected
miniature preamplifier.
The reference input of the miniature preamplifier (in case that a differential preamplifier is used) should
be connected to a reference electrode that is ideally identical to the recording electrodes and placed into
a comparable but inactive area or tissue. This way, background or noise signals that are picked up by both
the reference electrode and the recording electrodes are removed from the recording signal already in the
miniature preamplifier. (For more details on this issue, please see the MPA8I or MPA32I Manual.)
So what is the benefit of the differential operation of the PGA, if the noise is already canceled out in the
preamplifier? The differential operation of the PGA compensates the potential difference between the
miniature preamplifier output and the PGA input that results from the supply voltage drop in the line from
the preamplifier to the PGA. If multiple preamplifiers are connected to a PGA, the amplifier offsets are
averaged. Also, it is possible to connect a non-differential preamplifier or the electrodes themselves directly
to the PGA and cancel out the noise in the PGA. However, connecting the electrodes directly to the PGA
is not recommended, as the distance from the electrodes to the first amplification stage should be kept as
short as possible to optimize the signal-to-noise ratio.

- 14 -
5 Troubleshooting
5.1 About Troubleshooting
The following hints are provided to solve special problems that have been reported by users. Most problems
occur seldom and only under specific circumstances. Please check the mentioned possible causes carefully
when you have any trouble with the product. In most cases, it is only a minor problem that can be easily
avoided or solved.
If the problem persists, please contact your local retailer. The highly qualified staff will be glad to help you.
Please inform your local retailer as well, if other problems that are not mentioned in this documentation
occur, even if you have solved the problem on your own. This helps other users, and it helps MCS to optimize
the instrument and the documentation.
Please pay attention to the safety and service information (chapter "Important Safety Advice" in the manual
or help). Multi Channel Systems has put all effort into making the product fully stable and reliable, but like all
high-performance products, it has to be handled with care.
5.2 Technical Support
Please read the Troubleshooting part of the manual or help first. Most problems are caused by minor
handling errors. Contact your local retailer immediately if the cause of trouble remains unclear. Please
understand that information on your hardware and software configuration is necessary to analyze and finally
solve the problem you encounter. If you have any questions or if any problem occurs that is not mentioned in
this document, please contact your local retailer. The highly qualified staff will be glad to help you.
Please keep information on the following at hand:
Description of the error (the error message text or any other useful information) and of the context
in which the error occurred. Try to remember all steps you had performed immediately before the
error occurred. The more information on the actual situation you can provide, the easier it is to track
the problem.
The serial number of the device. You will find it on the device.
The software and hardware version you are currently using. In the PGA-Control window, click
"About".
The operating system and service pack number on the connected computer.
The hardware configuration (microprocessor, frequency, main memory, hard disk) of the connected
computer. This information is especially important if you have modified the computer or installed new
hard- or software recently.

- 15 -
5.3 Insufficient Power
You see no signals or strange artifacts on all channels. This behavior can be caused by an insufficient supply
power, indicated by a flickering or dark power LED. Please see the PGA data sheet for more information on
the supply power requirements. If the voltage drops beyond a critical level, the amplifier cannot operate
properly, resulting in artifacts or in a saturation of the amplifier. MCS standard cables with a total length
of up to 3 m and the isolated power supply IPS10W are recommended for operating the PGA.
Possible causes:
? Loose cable or bad contact. The power distribution from the IPS10W in the connected computer
is obstructed.
Unplug and reconnect the 68-pin MCS standard cable connecting the PGA with the data acquisition
computer. Make sure the connectors are properly plugged in and not tilted.
? The output power of the power supply is too low for operating the amplifier. This is especially likely
if you use a custom power supply, or the power supply has a technical defect.
— Please check the output power of the power supply and compare it with the PGA data sheet.
Contact your local retailer, describe the problem and your hardware configuration, and ask for a power
supply that is suitable for your amplifier.

- 16 -
6 Appendix
6.1 Pin Layout
PGA Input: 68-Pin MCS Standard Connector
Channel 17 - 32 (only used in PGA32, PGA64, PGA1632)
Channel 33 - 48 (only used in PGA64, PGA3264)
Channel 49 - 64 (only used in PGA64, PGA3264)
PGA Output: 68-Pin MCS Standard Connector
Channel 17 - 32 (only used in PGA32, PGA64, PGA1632)
Channel 33 - 48 (only used in PGA64, PGA3264)
Channel 49 - 64 (only used in PGA64, PGA3264)

- 17 -
6.2 Contact Information
Local retailer
Please see the list of official MCS distributors on the MCS web site.
User forum
The Multi Channel Systems User Forum provides the opportunity for you to exchange your experience
or thoughts with other users worldwide.
Mailing list
If you have subscribed to the Mailing List, you will be automatically informed about new software releases,
upcoming events, and other news on the product line. You can subscribe to the list on the contact form of
the MCS web site.
www.multichannelsystems.com

- 18 -
6.3 Ordering Information
Please contact your local retailer for pricing and ordering information.
Amplifiers
Miniature preamplifier with 2 electrode inputs
Small sized and light weight headstage with
common ground and additional indifferent
reference electrode input, input type I,
gain = 10
Miniature preamplifier with 8 electrode inputs
Miniature preamplifier with 32 electrode
inputs
Filter amplifiers with 4, 8, 16, 32, 48, or 64
channels and input type S or I
NN is the total number of channels,
X is the input type (S or I), with custom gain
and bandwidth
Amplifier with programmable gain, 16
channels
Gain programmable from 10 to 5000
Amplifier with programmable gain, 32
channels
Amplifier with programmable gain, 64
channels
Amplifier with programmable gain, 16 input
and 32 output channels
Gain programmable from 10 to 5000, with
two different pass bands
Amplifier with programmable gain, 32 input
and 64 output channels
0.5 m 68-pin MCS standard cable
1 m 68-pin MCS standard cable
2 m 68-pin MCS standard cable
3 m 68-pin MCS standard cable

Data acquisition system with 16 analog
channels
Complete with data acquisition computer with
MC_Card and IPS10W, and software package
Data acquisition system with 32 analog
channels
Data acquisition system with 64 analog
channels
Data acquisition system with 128
analog channels
ME recording system with 16 analog
channels and filter amplifier with fixed
gain
Stand-alone system for extracellular recordings,
complete with 2 x MPA8I, integrated FA16I,
integrated 16-channel data acquisition, USB 2.0
data transfer to computer, and software package
ME recording system with 16 analog
channels and filter amplifier with fixed
gain
Complete with 2 x MPA8I, SC8x8, FA16I, data
acquisition computer with MC_Card and IPS10W,
and software package
ME recording system with 32 analog
channels and filter amplifier with fixed
gain
Complete with 2 x MPA32I, SC2x32, FA32I, data
acquisition computer with MC_Card and IPS10W,
and software package
ME recording system with 64 analog
channels and filter amplifier with fixed
gain
Complete with 2 x MPA32I, SC2x32, FA64I, data
acquisition computer with MC_Card and IPS10W,
and software package
ME recording system with 128 analog
channels and filter amplifier with fixed
gain
Complete with 4 x MPA32I, 2 x SC2x32, 2 x
FA64I, data acquisition computer with MC_Card
and IPS10W, and software package
ME recording system with 16 analog
channels and filter amplifier with
programmable gain
Complete with 2 x MPA8I, SC8x8, PGA16, data
acquisition computer with MC_Card and IPS10W,
and software package
ME recording system with 32 analog
channels and filter amplifier with
programmable gain
Complete with 2 x MPA32I, SC2x32, PGA32, data
acquisition computer with MC_Card and IPS10W,
and software package
ME recording system with 64 analog
channels and filter amplifier with
programmable gain
Complete with 2 x MPA32I, SC2x32, PGA64, data
acquisition computer with MC_Card and IPS10W,
and software package
ME recording system with 128 analog
channels and filter amplifier with
programmable gain
Complete with 4 x MPA32I, 2 x SC2x32, 2 x
PGA64, data acquisition computer with MC_Card
and IPS10W, and software package

Power supply with 40 W power and
± 7 V output voltage
For supplying power to programmable gain
amplifiers, for custom setups that lack an internal
power supply, or for other custom applications.
32-Channel Cactus Needle Adapter
Allows a direct connection of electrodes with cactus
needles to a 32-channel miniature preamplifier
MPA32I.
NeuroNexus Probe adapters for 16,
32, or 54 channels
ADPT-NN-16
ADPT-NN-32
ADPT-NN-54
The 16-Electrode NeuroNexus Probe Adapter allows
a direct connection of 16-Electrode acute probes
from NeuroNexus Technologies to two 8-channel
miniature preamplifiers MPA8I.
The 32-Electrode NeuroNexus Probe Adapter allows
a direct connection of 32-Electrode acute probes
from NeuroNexus Technologies to a 32-channel
miniature preamplifier MPA32I.
The 54-Electrode NeuroNexus Probe Adapter allows
a direct connection of 54-Electrode acute probes
from NeuroNexus Technologies to two 32-channel
miniature preamplifiers MPA32I.
Signal collector for 8 x MPA8I
Collects the 8 channels from each of the up to
8 miniature preamplifiers MPA8I and leads them
to the amplifier via a 68-pin MCS standard cable.
Signal collector for 2 x MPA32I
Collects the 32 channels from each of the up to
2 miniature preamplifiers MPA32I and leads them
to the amplifier via a 68-pin MCS standard cable.
Signal collector with blanking circuit
for 8 x MPA8I
Collects the 8 channels from each of the up to
8 miniature preamplifiers MPA8I and leads them
to the amplifier via a 68-pin MCS standard cable,
the voltage outputs are held constant during the
TTL blanking signal, stimulus artifacts are avoided.
Signal dividers with 16, 32, or 64
channels
Placed between amplifier and MC_Card, permits
to select any channel, does not interfere with the
data acquisition.
For silencing defective electrodes. Single channels
can be switched off with small toggle switches.
Inputs not in use are grounded.

- 21 -
Microelectrode Arrays
A broad range of microelectrode probes are available from several providers. The following are compatible
with the miniature preamplifiers from Multi Channel Systems MCS GmbH. If you are interested in a particular
probe that is not mentioned here, please ask Multi Channel Systems or your local retailer for compatibility.
Flexible
microelectrode
array
(FlexMEA36)
With 36 electrodes in total 32 electrodes arranged in a 6 x 6 grid
+ 2 reference electrodes + 2 ground electrodes. 30 μm electrode
diameter, 300 μm interelectrode distance. TiN electrodes and
gold tracks embedded in flexible polyimide.
Multi Channel Systems
MCS GmbH
www.multichannel
systems.com
Flexible
microelectrode
array
(FlexMEA72)
With 72 electrodes in total 72 electrodes arranged in an 8 x 9
grid + 4 reference electrodes + 4 ground electrodes. 100 μm
electrode diameter, 625 to 750 μm interelectrode distance. TiN
electrodes and gold tracks embedded in flexible polyimide.
Flexible
microelectrode
array
(EcoFlexMEA36)
With 36 electrodes in total 32 electrodes arranged in a 6 x 6 grid
+ 2 reference electrodes + 2 ground electrodes. 50 μm electrode
diameter, 300 μm interelectrode distance. Gold electrodes and
gold tracks embedded in flexible polyimide (Kapton).
Flexible
microelectrode
array
(EcoFlexMEA24)
With 24 electrodes in total 24 electrodes arranged in a 10 x 2
grid + 2 reference electrodes + 2 ground electrodes. 80 μm
electrode diameter, 300 μm interelectrode distance. Gold
electrodes and gold tracks embedded in flexible polyimide
(Kapton).
NeuroNexus
acute or
chronic probes
16-, 32-, 64-channel silicon probes
NeuroNexus
Technologies
http://neuronexus.com/