Multichannel Systems HE-Up User Manual

HE-UP
Heating element for MEA1060-UP Amplifier
Installation
1.) You need the heating element (HE-Up), the connector board (HE-Up-Con) to the temperature controller TCX, and four provided screws each.
HE-Up
HE-Up-Con
2.) Mount the heating element onto the amplifier’s bottom. Make sure the input pins go through the hole.
Multi Channel Systems MCS GmbH
Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25­Fax
info@multichannelsystems.com www.multichannelsystems.com
0
© 2011 Multi Channel Systems MCS GmbH
Product information is subject to change without notice.
HE-UP
Input pins
Heating element for MEA1060-UP Amplifier
Installation
3.) Screw the heating element (HE-Up) tightly from the other side with the provided four bigger screws (M3x6).
Input pins
4) Mount the connector board (HE-Up-Con) so that all six pins are connected to the heating element. Tighten it with the provided four smaller screws (M2x4).
Scope of delivery: 1 HE-Up + 4 screws (M3x6) 1 HE-Up-Con + 4 screws (M2x4)
Multi Channel Systems MCS GmbH
Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25­Fax
info@multichannelsystems.com www.multichannelsystems.com
0
© 2011 Multi Channel Systems MCS GmbH
Product information is subject to change without notice.