Use and Operation
Electrodes wear out after repeated use or over a longer time of use. The noise produced by electrodes in a bad condition can interfere with the signals of electrodes nearby.
OUT
to amplifier
To silence the defective electrodes, you can remove them from your input signal with the 64-channel grounding unit. The channels are lead through the GND64; single channels can be picked and switched off. The pin assignment is not affected. The amplifier inputs not in use are grounded.
Do not mismatch the input and output sockets, otherwise the GND64 does not work.
In the figure above, all channels are in a switched on state. To switch off a channel, move the respective slider up with your finger tip or with a pin.