EcoFlexMEA36
Flexible Microelectrode Array with 36 electrodes for use with 32-Channel Miniature Preamplifier MPA32I for in vivo or in vitro applications.
37 mm
Electrode field: 1800 x 1800 m, 32 recording electrodes, 2 reference electrodes, 2 ground electrodes
Connect the EcoFlexMEA36 directly to a 32-Channel Miniature Preamplifier. Insert the EcoFlexMEA36 into the MPA32I with the electrode field up. The additional connector can be used for connecting a silver pellet or a silver wire for grounding the bath.
Technical Specifications EcoFlexMEA36
Temperature compartibility 10 - 125 °C Dimension (W x D) 37 mm x 30 mm Thickness of the electrode field 50 m Weight < 10 g
Base material Polyimide (Kapton) Contact pads and track material Gold Electrode diameter 50 m Interelectrode distance (centre to centre) 300 m
28 mm
23 mm
30 mm
6 mm
9 mm
14 mm
Electrode height Planar Electrode type Gold electrodes Isolation type Polyimide (Kapton) Electrode impedance Approximately 50 k Electrode layout grid 6 x 6 Number of recording electrodes 32 Number of reference electrodes 2 internal reference electrodes Number of ground electrodes 2 ground electrodes
MC_Rack Version 4.1.1 and higher Data Source Setup Source Layout: Configuration, Total Number of Channels: 64, Amplifier: FA32I/S or FA64I/S, MEA: EcoFlexMEA36. Channel map Display: Layout tab > Default Map (EcoFlexMEA36.cmp).
Cleaning Rinse with distilled water. EcoFlexMEAs made from Polyimide (Kapton) are heat stable and autoclavable.
Multi Channel Systems MCS GmbH
Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25­Fax
info@multichannelsystems.com www.multichannelsystems.com
0
© 2013 Multi Channel Systems MCS GmbH
Product information is subject to change without notice.
EcoFlexMEA36
Electrode layout
300 m
GND 2
5
B1 B2 B3 B4 B5 B6
7
C1 C2 C3 C4 C5 C6
3
A2 A3 A4 A5
4
6
9
D1 D2 D3 D4 D5 D6
11
E1 E2 E3 E4 E5 E6
REF
F2 F3 F4 F5
GND 1 is a large ground electrode connected to pin 1 of the MPA32I input connector. GND 2 is a second ground electrode connected to pin 36. The REF electrodes are reference electrodes connected to pin 2 and 35, respectively. Both ground inputs and both reference electrode inputs are equal, that is, they are connected to each other inside the standard MPA32I. Please see the MPA32I manual for details.
12
13
1
2
8
1610
14
15
31
30
24
18
17
29
28
2632
22
20
19
GND 1
27
25
23
21
REF
300 m
50 m
The numbers in the electrodes are the recording channel numbers that refer to the channel numbers of the data acquisition program. Please make sure that you have selected “Configuration” in the “Channel Layout” under “Data Source Setup” with a total number of 64 channels. In “Amplifier”, please choose FA32I/S or FA64I/S and in “MEA” EcoFlexMEA36. In Layout tab of the display, please click ”Default Map”.
The letter digit code below is the electrode identifier and refers to the position of the electrode in the grid.
Important: The channel map is constructed by looking on the back side of the electrodes, because the FlexMEA electrodes are placed on the preparation upside down!
If you use a MEA64-System with more than one MPA32I, the signal collector SC2x32 leads the output channels of the second amplifier to channel numbers 33 to 64. Please read the SC2x32 data sheet.
Warning: The device may only be used together with the MPA32I from Multi Channel Systems
!
MCS GmbH, and only for the specified purpose. Damage of the device and even injuries can result from improper use.
Multi Channel Systems MCS GmbH
Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25­Fax
info@multichannelsystems.com www.multichannelsystems.com
0
© 2013 Multi Channel Systems MCS GmbH
Product information is subject to change without notice.
Loading...