Multichannel Systems ADPT-NNCP-16 User Manual

ADPT-NNCP-16
16-Electrode Chronic Probe Adapter for MPA8I Amplifiers
Connecting the MPA8I to the NeuroNexus Chronic Probe Adapter
Side view
Connect MPA8I No. 1 here Connect MPA8I No. 2 here
MPA8I No. 1
MPA8I No.1
MPA8I No.2
MPA8I No.2
Important: Connect the
Important: Connect the
MPA8I with the black side
MPA8I with the black side
down. Otherwise the device
down. Otherwise the device
does not work.
does not work.
16-Electrode Chronic Probe from NeuroNexus
NNCP-Adapter
MPA8I No. 1
MPA8I No. 2
G
G
87654321R
G
87654321R
G
Shown are the output pins of the adapter that are connected to the miniature preamplifiers MPAI when looking directly at the pins. The labeled channels are the ground channels (G), the reference channel (R), and the 8 recording channels (1 to 8) of the MPA8I. Please see the MPA8I manual for details.
Warning: The device may only be used together with the MPA8I from Multi Channel
!
Systems MCS GmbH and the 16-channel chronic probe from NeuroNexus, and only for the specified purpose. Damage of the device and even injuries can result from improper use.
Multi Channel Systems MCS GmbH
Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25­Fax
+49-7121-9 09 25-11
info@multichannelsystems.com www.multichannelsystems.com
0
© 2012 Multi Channel Systems MCS GmbH
Product information is subject to change without notice.
ADPT-NNCP-16
16-Electrode Chronic Probe Adapter for MPA8I Amplifiers
Connecting the NeuroNexus Chronic Probe CP16 to the NNCP-Adapter
Top view
Important: Black side of MPA8I faced down !
Omnetics connector A9279-001 (NSD-18-W-GS-w/2GP), male, fixed on the adapter.
The red arrow shows the direction of the probe. Because of the guide posts it is just possible to plug the male connector of the adapter into the female connector of the probe in one required position!
Pin Layout of MPA8I No.1 and No. 2 and NeuroNexus Chronic Probe CP16
Pin
MPA8I No. 1 Pin NNCP16
Reference 10 CH 01 9 SITE 8 CH 02 8 SITE 7 CH 03 7 SITE 6 CH 04 6 SITE 5 CH 05 5 SITE 4 CH 06 4 SITE 3 CH 07 3 SITE 2 CH 08 2 SITE 1 Ground 1
Reference
10
9
8
7
6
5
4
3
20
19
18
17
16
15
14
13
Guide post
MPA8I No. 2 Pin NNCP16
Guide post 11 CH 01 12 SITE 9 CH 02 13 SITE 10 CH 03 14 SITE 11 CH 04 15 SITE 12 CH 05 16 SITE 13 CH 06 17 SITE 14 CH 07 18 SITE 15 CH 08 19 SITE 16 Guide post 20
2
Ground
The photo shows the top of the adapter NNCP from Multi Channel Systems MCS GmbH. To insert the chronic probe from NeuroNexus Technologies C16 into the adapter you have to turn the chronic probe upside down. It is only one way possible to plug in the probe.
Multi Channel Systems MCS GmbH
Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25­Fax
+49-7121-9 09 25-11
info@multichannelsystems.com www.multichannelsystems.com
1
12
Guide post
11
0
© 2012 Multi Channel Systems MCS GmbH