ADPT-NN-16
16-Electrode NeuroNexus Probe
Adapter for MPA8I Amplifiers
Connecting two MPA8Is to the adapter ADPT-NN-16
Side view
Important: Connect both MPA8Is
Important: Connect both MPA8Is
with the black side down. Otherwise,
Connect MPA8I No. 1 here
Connect MPA8I No. 2 here
MPA8I No. 1
MPA8I No.2
with the black side down. Otherwise,
the preamplifiers do not work.
the preamplifiers do not work.
Channel assignment, output pins (connected to MPA8I)
NeuroNexus Probe
NN-ADAPTER
MPA8I No. 1
MPA8I No. 2
G
G
87654321R
G
87654321R
G
Shown are the output pins of the adapter that are connected to the miniature preamplifiers MPAI when
looking directly at the pins. The labeled channels are the ground channels (G), the reference channel (R),
and the 8 recording channels (1 to 8) of the MPA8I. Please see the MPA8I manual for details.
Warning: The device may only be used together with the MPA8I from Multi Channel Systems
!
MCS GmbH and the 16-channel probe from NeuroNexus, and only for the specified purpose.
Damage of the device and even fatal injuries can result from improper use.
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25Fax
+49-7121-9 09 25-11
info@multichannelsystems.com
www.multichannelsystems.com
0
© 2012 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.
ADPT-NN-16
16-Electrode NeuroNexus Probe
Adapter for MPA8I Amplifiers
Connecting the NeuroNexus Probe to the Adapter ADPT-NN-16
Complete setup with NeuroNexus probe and two MPA8Is
Important:
Black side of MPA8I faced down !
NeuroNexus Probe
NN-Adapter
MPA8I No. 1
MPA8I No. 2
Top view without probe
MPA8I MPA8I
No. 2
1
2
3
4
5
6
7
8
R
No. 1
8
7
6
5
4
3
2
1
G
NeuroNexus Probe-Adapter
Channel assignment:
Input pins connected to the NeuroNexus probe
Pin Layout for MC_Rack:
MPA8I MC_Rack
No. 1
Pin 1 - 8 Channel 1 - 8
No. 2
Pin 1 - 8 Channel 9 - 16
Shown are the input pins of the adapter that are connected to the probe. The labeled channels are
the ground channel (G), the reference channel (R), and the 8 recording channels (1 to 8) of the MPA8I.
Please see the MPA8I manual for details. If you use MC_Rack for data acquisation, see Pin Layout above.
G and R have been connected together as factory-default settings. You can change this connection
to meet your requirements.
Important: Operation of the MPA8I is differential. The reference channel R has to be used for obtaining
a proper signal. Please see the MPA8I manual for details.
Warning: The device may only be used together with the MPA8I from Multi Channel Systems
!
MCS GmbH and the 16-channel probe from NeuroNexus, and only for the specified purpose.
Damage of the device and even fatal injuries can result from improper use.
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25Fax
+49-7121-9 09 25-11
info@multichannelsystems.com
www.multichannelsystems.com
0
© 2012 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.