Multichannel Systems 60SquareMEA200-50iR-Ti User Manual

60SquareMEA200/50iR-Ti
Microelectrode Array with 60 square electrodes
21 31 41 51 61 71
53
54
55
58
62
63 73
64
65
66
67
68
72 82
74
75 85
76
77
78
83
84
86
87
REF
12
13
14
16
17
32
22
33
23
24 34
25 35
26
36
27
37 47 57
28 38
42 52
43
44
45
46 56
48
Technical Specifications 60SquareMEA200/50iR-Ti
Temperature compatibility 0 - 125 °C Dimension (W x D x H) 49 mm x 49 mm x 1 mm Base material Glass Track material Ti (Titanium) Contact pads TiN (Titanium nitride) Electrode square size 50 x 50 μm Interelectrode distance (centre to centre) 200 μm Electrode height Planar Electrode type TiN (Titanium nitride) Isolation type Silicon nitride 500 nm (PEVCD) Electrode impedance 30 - 50 k Electrode layout grid 8 x 8 Number of recording electrodes 59 Number of reference electrodes 1 internal reference electrode (iR)
MC_Rack Source layout in “Data Source Setup” 2 dim. (MEA) or Configuration Channel map Default
MEA perfusion chamber (w/o) Without ring (gr) Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm (pr) Plastic ring without thread: ID 26.5 mm, OD 30 mm, height 6 / 3 mm (pr-T) Plastic ring with thread: ID 26 mm, OD 30 mm, height 6 / 15 mm
Multi Channel Systems MCS GmbH
Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25­Fax
info@multichannelsystems.com www.multichannelsystems.com
0
© 2013 Multi Channel Systems MCS GmbH
Product information is subject to change without notice.
60SquareMEA200/50iR-Ti
Microelectrode Array with 60 square electrodes
Standard electrode layout grid 8 x 8
MEA pins
Electrode #
33 33
22 22
12 12
23 23
13 13
34 34
24 24
14 14
15 REF
25 25
35 35
16 16
26 26
17 17
21 32 31 44 43 41 42 52 51 53 54 61 62 71
21 32 31 44 43 41 42 52 51 53 54 61 62 71
21 31 41 51 61 71
12
13
14
15
15
16 26 36
17
22 32 42 52
33
23
24 34
25 35
43
44
45
46 56 66
27 37 47 57
28 38
48
62
53
63 73
64
54
55 65
67
58
68
72 82
83
74
84
75 85
76
86
77 87
78
Electrode #
MEA pins
63 63
72 72
82 82
73 73
83 83
64 64
74 74
84 84
85 85
75 75
65 65
86 86
76 76
87 87
27 27
36 36
MEA pins
28 37 38 45 46 48 47 57 58 56 55 68 67 78
28 37 38 45 46 48 47 57 58 56 55 68 67 78
Electrode #
77 77
66 66
Electrode #
MEA pins
The numbering of MEA electrodes in the 8 x 8 grid follows the standard numbering scheme for square grids: The first digit is the column number, and the second digit is the row number. For example, electrode 23 is positioned in the third row of the second column.
The specified MEA pin numbers are the channel numbers that are used in the data acquisition program. The electrode 15 is replaced by a big internal reference electrode which is connected to pin 15 of the amplifier.
Multi Channel Systems MCS GmbH
Aspenhaustrasse 21 72770 Reutlingen Germany
Fon +49-7121-9 09 25­Fax
info@multichannelsystems.com www.multichannelsystems.com
0
© 2013 Multi Channel Systems MCS GmbH
Product information is subject to change without notice.
Loading...