
60MEA-SG
60MEA-Signal Generator
Control button
Grounding Socket
LED
DIP Switch
Battery
Gold Contact Pads
60MEA-Signal Generator is a very convenient tool for MEA-System and MEA2100-HS(2x)60 users.
Use the 60MEA-SG instead of setting up a complete experiment for training, controlling, and
troubleshooting purposes.
Note: When using a MEA-SG which is not adapted to MEA2100-System, the contact to the internal
reference must be disconnected. Otherwise the delivered signals are not acceptable. Please insert
the MEA-SG in correct orientation as shown on the picture. Cover the designated contact pad
connected to the reference electrode (Channel 15 in MC_Rack), for example with isolation tape.
Switch on : Press control button. Switch off : Press control button longer than two seconds.
Important : To change DIP switch position, please switch off the device!
!
Table : DIP switch position, number of control button presses, and corresponding signals
Switch 1 Switch 2 Control button Signal
presses n times
OFF OFF 60MEA-SG ON Sinus 0.005 Hz
1 Sinus 0.01 Hz
2 Sinus 0.03 Hz
3 Sinus 1.25 Hz
4 Sinus 12.5 Hz
ON OFF 60MEA-SG ON EPSP
1 Population Spike
2 Spikes
OFF ON 60MEA-SG ON ECG Atrium
1 ECG Ventricle
2 Ventricle FP
ON ON 60MEA-SG ON ERG with Spikes
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25Fax
+49-7121-9 09 25-11
info@multichannelsystems.com
www.multichannelsystems.com
0
© 2012 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.

60MEA-SG
60MEA-Signal Generator
Switch
1
OFF
ON
Switch
2
OFF
press
Button
n times
1
60MEA-SG
ON
Signal Source
Artificial Sine Wave (1.75 mV) 0.005 Hz
Note: sine waves < 1 Hz might not be
visible because of the hardware filter
bandwidth.
2
3
4
5
OFF
Sine Wave
1
60MEA-SG
Sine Wave
Sine Wave
Sine Wave
Hippocampal Slice
EPSP
ON
1
Hippocampal Slice
Population Spike
0.01 Hz
0.03 Hz
1.25 Hz
12.5 Hz
Signal Type
2
Hippocampal Neurons
Spikes
OFF
ON
1
60MEA-SG
Heart
ECG Atrium
ON
1
Heart
ECG Ventricle
2
Cardiomyocytes
Ventricle FP
ON
1
ON
60MEA-SG
Retina
ERG with Spike
ON
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25Fax
+49-7121-9 09 25-11
info@multichannelsystems.com
www.multichannelsystems.com
0
© 2012 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.