MTS Systems CMS-50FF Instructions Manual

Medical Test Supply
CMS-50FF
Instructions for CMS-50FF Pulse Oximeter
Dear users, thank you very much for purchasing the Pulse Oximeter. This Manual is written and compiled in accordance with the council directive MDD93/42/EEC for
medical devices and harmonized standards. In case of modifications and software upgrades, the information contained in this document is subject to change without notice.
The Manual describes, in accordancKe with the Pulse Oximeter’s features and requirements, main structure, functions, specifications, correct methods for transportation, installation, usage, operation, repair, maintenance and storage, etc. as well as the safety procedures to protect both the user and equipment. Refer to the respective chapters for details.
Please read the User Manual carefully before using this product. The User Manual which describes the operating procedures should be followed strictly. Failure to follow the User Manual may cause measuring abnormality, equipment damage and human injury. The manufacturer is NOT responsible for the safety, reliability and performance issues and any monitoring abnormality, human injury and equipment damage due to users' negligence of the operation instructions. The manufacturer’s warranty service does not cover such faults.
Owing to the forthcoming renovation, the specific products you received may not be totally in accordance with the description of this User Manual. We would sincerely regret for that.
This product is medical device, which can be used repeatedly.
WARNING:
Uncomfortable or painful feeling may appear if using the device ceaselessly, especially for
the microcirculation barrier patients. It is recommended that the sensor should not be applied to the same finger for over 2 hours.
For the special patients, there should be a more prudent inspecting in the placing process.
The device cannot be clipped on the edema and tender tissue.
The light (the infrared is invisible) emitted from the device is harmful to the eyes, so the
user and the maintenance man should not stare at the light.
Testee cannot use enamel or other makeup.
Testees fingernail cannot be too long.
Please refer to the correlative literature about the clinical restrictions and caution.
This device is not intended for treatment.
The User Manual is published by our company. All rights reserved.
Medical Test Supply
CMS-50FF
CONTENTS
1Safety ..................................................................................................................................................... 1
1.1Instructionsforsafeoperations ........................................................................................................... 1
1.2Warning ........................................................................................................................................... 1
1.3Attention ......................................................................................................................................... 1
2Overview ............................................................................................................................................... 2
2.1Features ........................................................................................................................................... 2
2.2Majorapplicationsandscopeofapplication ......................................................................................... 3
2.3Environmentrequirements ................................................................................................................ 3
3Principle ................................................................................................................................................ 3
4Technicalspecifications ......................................................................................................................... 4
4.1Mainperformance ............................................................................................................................ 4
4.2MainParameters ............................................................................................................................... 4
5Installation ............................................................................................................................................ 5
5.1Viewofthefrontpanel ................................................................ ........................................................ 5
5.2Wristbandinstallationandprobeconnection ........................................................................................ 5
5.3Accessories ...................................................................................................................................... 6
6OperatingGuide .................................................................................................................................... 6
6.1Applicationmethod .......................................................................................................................... 6
6.2Attentionforoperation .................................................................................................................... 11
6.3Clinicalrestrictions ......................................................................................................................... 12
7Maintaintransportationandstorage ................................................................................................ 12
7.1CleaningandDisinfecting................................................................................................................ 12
7.2Maintain ........................................................................................................................................ 12
7.3Transportationandstorage ............................................................................................................... 12
8Troubleshooting .................................................................................................................................. 13
9KeyofSymbols ................................ ..................................................................................................... 13
10FunctionSpecification ....................................................................................................................... 14
Appendix1 ................................................................................................................................ ............. 15
Medical Test Supply
CMS-50FF
1 Safety
1.1 Instructions for safe operations
Check the main unit and all accessories periodically to make sure that there is no visible
damage that may affect patient’s safety and monitoring performance about cables and transducers. It is recommended that the device should be inspected once a week at least. When there is obvious damage, stop using the device.
Necessary maintenance must be performed by qualified service engineers ONLY. Users are not
permitted to maintain it by themselves.
The oximeter cannot be used together with devices not specified in User’s Manual. Only the
accessory that appointed or recommendatory by manufacture can be used with this device.
This product is calibrated before leaving factory.
1.2 Warning
Explosive hazardDO NOT use the oximeter in environment with inflammable gas such as some ignitable anesthetic agents.
DO NOT use the oximeter while the teste measured by MRI and CT. Please do not break the wristband, for fear it becomes out of use, or the unexpected drop of
the device which is due to the looseness of the wristband in the process of using. Users who are allergic to the wristband is not recommended to use it.
The person who is allergic to rubber cannot use this device. The disposal of scrap instrument and its accessories and packing’s (including battery, plastic
bags, foams and paper boxes) should follow the local laws and regulations. Please check the packing before use to make sure the device and accessories are totally in accordance with the packing list, or else the device may have the possibility of working abnormally. Please choose the accessories and probe which are approved or manufactured by the manufacturer, or else it may damage the device. Please choose the battery chargers which should be ensured compliance with the requirements of IEC 601-1, or else it may damage the device.
Please don't use the device in the course of charging. The device can only be matched with the compatible probe. Please don't measure this device with functional tester for the device's related information.
1.3 Attention
Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture.
If the oximeter gets wet, please stop operating it. When it is carried from cold environment to warm or humid environment, please do not use it
immediately.
DO NOT operate keys on front panel with sharp materials. High temperature or high pressure steam disinfection of the oximeter is not permitted. Refer to
User Manual in the relative chapter (7.1) for instructions of cleaning and disinfection. Do not have the oximeter immerged in liquid. When it needs cleaning, please wipe its surface with medical alcohol by soft material. Do not spray any liquid on the device directly. When cleaning the device with water, the temperature should be lower than 60℃.
Medical Test Supply
CMS-50FF
As to the fingers which are too thin or too cold, it would probably affect the normal measure of the patients' SpO2 and pulse rate, please clip the thick finger such as thumb and middle finger deeply enough into the probe. The pulse oximeter can be used to adult or infant. Whether the device is used to adult or infant, it depends on the probe selected.
Whether the device is used to adult or infant, it depends on the probe selected. The update period of data is less than 5 seconds, which is changeable according to different
individual pulse rate. Please read the measured value when the waveform on screen is equably and steady-going, this measured value is optimal value. And the waveform at the moment is the standard one. If some abnormal conditions appear on the screen during test process, pull out the finger and reinsert to restore normal use.
The device has normal useful life for three years since the first electrified use. This device has the function of alarming; users can check on this function according to chapter
6.1 as a reference. The device has the function of limits alarming, when the measured data is beyond the highest or lowest limit, the device would start alarming automatically on the premise of the alarming function is on. The device has the function of alarming, this function can either be paused, or closed (default setting) for good. This function could be turned on through menu operation if you need. Please check the chapter 6.1 as a reference. The device may not work for all patients. If you are unable to achieve stable readings, discontinue use.
2 Overview
The pulse oxygen saturation is the percentage of HbO
2
in the total Hb in the blood, so-called the O
2
concentration in the blood. It is an important bio-parameter for the respiration. A number of diseases relating to respiratory system may cause the decrease of SpO2 in the blood, furthermore, some other causes such as the malfunction of human body's self-adjustment, damages during surgery, and the injuries caused by some medical checkup would also lead to the difficulty of oxygen supply in human body, and the corresponding symptoms would appear as a consequence, such as vertigo, impotence, vomit etc. Serious symptoms might bring danger to human's life. Therefore, prompt information of patients' SpO2 is of great help for the doctor to discover the potential danger, and is of great importance in the clinical medical field. The Pulse Oximeter features in small volume, low power consumption, convenient operation and being portable. It is only necessary for patients to put one of his fingers into a probe for diagnosis, and a display screen will directly show the measured value of pulse oxygen saturation with the high veracity and repetition.
2.1 Features
A. Operation of the product is simple and convenient. B. The product is small in volume, light in weight and convenient in carrying. C. Low power consumption
Medical Test Supply
CMS-50FF
2.2 Major applications and scope of application
The Pulse Oximeter can be used in measuring the pulse oxygen saturation and pulse rate through finger. The product is suitable for being used in family, hospital, oxygen bar, community healthcare, physical care in sports (It can be used before or after doing sports and it is not recommended to use the device during the process of having sport) and etc.
The problem of overrating would emerge when the patient is suffering from toxicosis
which caused by carbon monoxide, the device is not recommended to be used under this circumstance.
2.3 Environment requirements
Storage Environment a) Temperature:-40℃~+60
b) Relative humidity :5%95% c) Atmospheric pressure :500hPa~1060hPa
Operating Environment a) Temperature:10℃~40 b) Relative Humidity :30%75% c) Atmospheric pressure:700hPa1060hPa
3 Principle
Principle of the Oximeter is as follows: An experience formula of data process is established taking use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive Hemoglobin (Hb) and Oxyhemoglobin (HbO2) in glow & near-infrared zones. Operation principle of the device is: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning & Recording Technology, so that two beams of different wavelength of lights can be focused onto human nail tip through perspective clamp finger-type sensor. Then measured signal can be obtained by a photosensitive element, information acquired through which will be shown on screen through treatment in electronic circuits and microprocessor.
Figure 1.
Medical Test Supply
CMS-50FF
4 Technical specifications
4.1 Main performance
A. SpO2 value display B. Pulse rate value display, bar graph display C. Pulse waveform display D. Low-voltage indication: low-voltage indicator appears before working abnormally which is due
to low-voltage
E. The display mode can be changed F. Screen brightness can be changed G. A pulse sound indication H. With alarm function I. With SpO2 value and pulse rate value of storage, the stored data can be uploaded to computers J. It can be connected with an external oximeter probe K. Real-time data can be transmitted to computers L. With clock function
4.2 Main Parameters A. Measurement of SpO2
Measuring range: 0%~100% Accuracy: When the SpO2 measuring range is 70%100%, the permission of absolute error is ±2%; Below 70% unspecified B. Measurement of pulse rate Measuring range:30bpm250bpm Accuracy: ±2 bpm or ±2% (select larger) C. Resolution SpO2 : 1%, Pulse rate: 1bpm. D. Measurement Performance in Weak Filling Condition SpO2 and pulse rate can be shown correctly when pulse-filling ratio is 0.4%. SpO2 error is ±4%, pulse rate error is ±2 bpm or ±2% (select larger). E. Resistance to surrounding light The deviation between the value measured in the condition of man-made light or indoor natural light and that of darkroom is less than ±1%.
F. Power supply requirement: : 3.6 V DC ~ 4.2V DC. G. Optical Sensor
Red light (wavelength is 660nm6.65mW) Infrared (wavelength is 880nm, 6.75mW)
H. Adjustable alarm range:
SpO2 0%~100% Pulse Rate: 0bpm~254bpm
Medical Test Supply
CMS-50FF
5 Installation
5.1 View of the front panel
Figure 2. Front view
Figure 3. Left and right view
1. Probe jack: It is used to connect a SpO2 sensor to measure the oxygen saturation and pulse rate.
2. USB port: It is used to connect a personal computer to export the trend data or charge the lithium battery via a data line.
3. Charging indication light. When the device is in the state of battering charging, the indication light appears to be orange, and when the battery status is full, the light turns to green.
5.2 Wristband installation and probe connection
A. Put one side of the wristband on which there is no ring through the belt hole above the device,
then put it through the other belt hole under the device.
B. Make the alignment signs on the probe in accordance with the alignment signs on the device, then plug the probe into the jack. C. It appears as the following figure after installation.
Medical Test Supply
CMS-50FF
Figure 3.
5.3 Accessories
A. a wristband B. a User Manual C. a power adapter D. a data line E. a disk (PC software) F. An adult-oximeter probeModel:CMS04
An infant-oximeter probe (May purchase selectively)
6 Operating Guide
6.1 Application method
A. Install the wristband according to the instructions of chapter 4.2. a) Put the suitable probe into the jack on the right side of the oximeter. (The probe is limited to be
produced by our company; never replace it with the similar ones by other manufacturers).
b) Put the finger into the probe. c) Turn on the device by long pressing the button on the panel. d) Do not shake the finger and keep the patient in a stable state during the process. e) The data can be read directly from the screen on the measuring interface.
Fingernails and the luminescent tube should be on the same side.
If the alarm function is on, the device will provide medium-priority alarm signal when
probe or finger is out. Intermittent alarm will occur and the user interface presents "FINGER OUT".
Medium priority indicating that prompt operator response is required.
Medical Test Supply
CMS-50FF
Figure 4.
Actual probe may be different with the probe as figure 4,please accept the actual probe with the
device
Attention:
CLICK = short press of power button and PRESS = prolonged push of power button (1sec) B. Change display direction
On the measuring interface, enter the clock interface by Click the button, and then Click the button to change the display direction within 30 seconds.
C. Enter and exit the clock interface a) On the measuring interface, Click the button in order to enter the clock interface, and it will
automatically return to the measuring interface if there are no more operations within 30 seconds. b) On the measuring interface, Press the button for about 10 seconds in order to enter the clock interface, and the device would return to the measuring interface by Press the button for about 10 seconds again.
D. Pause alarm a) Alarm including the alarm of measure data's going beyond the limits, the alarm of low-voltage,
the alarm of probe or finger's out of position. b) On the measuring interface, if the alarm function is on, during the period of alarming, you can
pause it by Click the button, but the function will be renewed in about 60 seconds.
c) If you want to turn off the alarm for good, you should enter the menu for operation. E. Menu operations
When the device is under the measuring interface, Press the button for about 1 second in order to enter the menu interface shown as figure 5. Users can adjust the setting through the main menu, such as backlight, alarm, clock, data transmission (with the data line), data storage, and power off. The specific operation methods are as the following:
Medical Test Supply
CMS-50FF
Figure 5. Main Menu Interface
a) Backlight adjustment On the main menu interface, Click the button to select Brightness, Press the power button and hold
to adjust the backlight brightness. b) Alarm setting On the main menu interface, click the button to select “Alarm, Press the power button (1sec) to enter the alarm setting interface as shown in Figure 6:
a. Adjusting the high and low limits of alarms Click the button to select Direction, then Press the button to choose Up or Down. (this will be the
direction the value of the high-low limits of SpO2 and pulse rate will be adjusted) To raise the SpO2 and pulse rate limit, choose Direction as up, then Click the button to highlight
the parameter to be adjusted: SpO2 high limit (SpO2 ALM HI), SpO2 low limit (SpO2 ALM LO), Pulse rate high limit (PR ALM HI), Pulse rate low limit (PR ALM LO), Press the button and hold to adjust the selected limit to the desired higher value and release the button once the higher limit has been reached.
To lower the SpO2 and pulse rate limit, choose Direction as down, then Click the button to choose the parameter to be adjusted. Press the button and hold to adjust the selected limit to the desired lower value and release the button once the lower limit has been reached.
If the alarm function is on, the device will provide medium-priority alarm signal when
the data of SpO2 or pulse rate is beyond the limit. Intermittent alarm will occur and the measurement shows in yellow. Medium priority indicating that prompt operator response is required.
b. The alarm state setting
Click the button to select Alarm, then Press the button to choose alarm on or off, press “on” to turn on the alarms and “off” to turn off the alarms.
Medical Test Supply
CMS-50FF
Figure 6.Alarm Setting Menu
c. Pulse sound indication setting Click the button to select Pulse Sound, and then Press the button to choose to have the Pulse
Sound (heart beat) alarm on or off”. d. Exit the Alarm settings Click the button to select EXIT, then Press the button to exit the Alarm Settings Menu. cClock setting On the main menu interface, Click the button to select Clock, then enter the clock setting interface
by press the button.
Figure 7. Clock Setting Menu
a. When entering the clock setting menu, the menu choice bar would be on the item of set time, and the state would always be no whenever it enters the clock setting menu on the purpose of avoiding unexpected changes of time due to improper operation. You can change the state by press the button, choose yes to reset the time, choose no to forbid time resetting. b. Click the button to select the parameter that you want to change, then adjust the data by press the button. c. Click the button to select Exit, then exit the clock setting menu by press the button. If you have reset the time or date, when exiting the clock setting menu, firstly the renewed time and date
Medical Test Supply
CMS-50FF
would be displayed on the screen, then it returns to the main menu; if you didn’t reset the time and date, when exiting the clock setting menu, the device would return to the main menu directly.
dReal-time data transmission setting Firstly, please install the affiliated software into the computer, and two icons would appear on the
desktop after installation. The icon of "SpO2 "is a program for receiving real-time data which is shown as figure 8; the icon of "SpO2 Review" is a program for receiving stored data which is shown as figure 9. a. Please connect the device to computer with the affiliated data line, then double click the "SpO2 "icon to start the program. b. On the main menu interface, Click the button to select “Usb”, then Press the button to choose whether transmit the real-time data to computer which displays the data synchronously or not, choose on to permit transmission, choose off to forbid transmission c. When you unplug the data line from computer, there is a dialog box "Save data at view" appearing on the desktop, in which you can input some patient's basic information.
Figure 8. SpO2 program
Figure 9. SpO2 Review program
If the users choose to turn on the synchronizing display function on computer, it would
probably take several seconds for the data to appear on the computer screen.
eData storage setting This instrument has the ability to store 24 hours’ worth of data. It can store the measured pulse rate
and SpO2 value accurately, transfer the data to the computer, display the data and print reports (with the included SpO2 Software - Green Heart) a. On the main menu interface, Click the button to select Record, then Press the button to choose whether store the data or not, choose Yes to permit storing, choose No to forbid storing.
b. The device can storage starting time automatically. c. If the data storage function is being turned on, when return to the measuring interface, a red
"REC"sign and a flashing red dot would appear on screen, which means the device is in a state of storing. d. In the state of storing, whatever interfaces the device is on (measuring interface, menu interface, and clock interface), the sign "Recording" would appear on the screen in 30 seconds, then the clock interface would appear in succession after several seconds. And then the screen will be automatically shut down. If short press the button at this moment, the sign "Recording" would appear on the screen,
Medical Test Supply
CMS-50FF
and then the screen will be automatically shut down again; if long press the button, the device would return to the former interface.
e. If turning on the data storage function, the former data storage will be automatically removed. f. In the state of data storing, after the screen is automatically shut down, the pulse sound indication
would be off for saving power. g. When the storage space is full, it displays Memory is full on the screen, and then shut down in a few seconds. But it will still display Memory is full by the next time you turn on the device on the purpose of warning the user, if Click the button again, it will enter the measuring interface.
fUploading the data to the PC after recording a. Please connect the device with computer by the data line which is equipped with the device,
then double click "SpO2 Review" icon to open"SpO2 Review" program, click the ‘New Session’ Icon in the software, enter the patient data and then click ‘ok’. The Software will then display
“device connected, waiting for data”. b. At this time, Press the button to enter the “Main Menu” and then Click the button to select Upload. Press the button to select “on” then the data will be transferred to your computer.
c. In the state of storing, it is not applicable for the users to upload the stored date to computer. d. When the upload of stored data is finished, the menu choice bar will move to "Exit" automatically.
g Power off
On the main menu interface, Click the button to select "Power off", then Press the button to shut down the device.
h Exit the main menu
On the main menu interface, Click the button to select "Exit", hen Press the button to exit the main menu.
F. Charge There are two kinds of charging methods:
a Connect the device with computer by data line, then the device should be under charging state. b) Connect the device with power supply by power adaptor, then the device should be under charging state. c) When the device is in the state of battering charging, the indication light appears to be orange, and when the battery status is full, the light turns to green.
If the alarm function is on, the device will provide high-priority alarm signal when the
battery is in low power status. Intermittent alarm will occur and the battery icon turns red in the state of flashing. High priority indicating that immediate operator response is required.
6.2 Attention for operation
A. Please check the device before using, and confirm that it can work normally.
B. The finger should be in a proper position (see the attached illustration of figure 4 for reference), or else it may result in inaccurate measure.
Medical Test Supply
CMS-50FF
C. The SpO2 sensor and photoelectric receiving tube should be arranged in a way with the subject’s arteriole in a position there between. D. The SpO2 sensor should not be used at a location or limb tied with arterial canal or blood
pressure cuff or receiving intravenous injection.
E. Do not fix the SpO2 sensor with adhesive or else it may result in venous pulsation and inaccurate
measure of SpO2 and pulse rate.
F. Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby
light, infrared heater, direct sunlight and etc.
G. Strenuous action of the subject or extreme electrosurgical interference may also affect the
accuracy.
H. Testee cannot use enamel or other makeup. I. Please clean and disinfect the device after operating according to the User Manual (7.1).
6.3 Clinical restrictions
A. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of subject is required. For a subject with weak pulse due to shock, low ambient/body temperature, major bleeding, or use of vascular contracting drug, the SpO2 waveform (PLETH) will decrease. In this case, the measurement will be more sensitive to interference. B. For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green and acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and some with icterus problem, the SpO2 determination by this monitor may be inaccurate. C. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor blamed for serious error of SpO2 measure. D. As the SpO2 value serves as a reference value for judgment of anemic anoxia and toxic anoxia, some patients with serious anemia may also report good SpO2 measurement.
7 Maintaintransportation and storage
7.1 Cleaning and Disinfecting
Using medical alcohol to disinfect the device, nature dry or clean it with clean soft cloth.
7.2 Maintain
A. Please clean and disinfect the device before using according to the User Manual (7.1).
B. Please recharge the battery when the screen shows .
C. Recharge the battery soon after the over-discharge. The device should be recharged every six months when it is no regular used. It can extend the battery life following this guidance. D. The device needs to be calibrated once a year (or according to the calibrating program of hospital). It also can be performed at the state-appointed agent or just contact us for calibration.
7.3 Transportation and storage
A. The packed device can be transported by ordinary conveyance or according to transport contract. The device cannot be transported mixed with toxic, harmful, corrosive material. B. The packed device should be stored in room with no corrosive gases and good ventilation. Temperature: -40°C~60°C; Humidity: 95%
Medical Test Supply
CMS-50FF
8 Troubleshooting
Trouble
Possible Reason
Solution
The SpO2 and Pulse Rate cannot be displayed normally
1. The finger is not properly positioned.
2. The patient’s SpO
2
is too low to be
detected.
1. Place the finger properly and try again.
2. Try again; Go to a hospital for a diagnosis if you are sure the device works all right.
The SpO2 and Pulse Rate are not displayed stably
1. The finger is not placed inside
deep enough.
2. The finger is shaking or the patient
is moving.
1. Place the finger properly and try
again.
2. Let the patient keep calm
The device cannot be turned on
1. The batteries are drained or almost drained.
2. The device’s malfunction
1.Please recharge the battery
2. Please contact the local service
center.
The display is off suddenly
1. The device’s malfunction.
2. The battery is drained away or almost drained away.
1. Please contact the local service
center.
2.Please recharge the battery
The device cannot be used for full time after charge
1The battery is not full charged. 2The battery is broken
1Please recharge the battery 2Please contact the local service center.
The battery cannot be full charged even after 10 hours charging time.
The battery is broken
Please contact the local service center.
9 Key of Symbols
Signal
Description
Warning – See User Manual
The pulse oxygen saturation (%)
Pulse rate (bpm)
Full-voltage
Low-voltage
Close the alarm sound indication
Medical Test Supply
CMS-50FF
Pause the alarm sound indication
Open the alarm sound indication
Close the pulse sound indication
Open the pulse sound indication
menu button/power button/function button Type BF
SN
Serial number
1. the finger clip falls off ( no finger inserted)]
2. Probe error
3. Signal inadequacy indicator
IPX1
Ingress of liquids rank
WEEE (2002/96/EC)
10 Function Specification
Information
Display Mode
The Pulse Oxygen SaturationSpO2)
2-digit digital OLED display
Pulse RatePR
3-digit digital OLED display
Pulse Intensity (bar-graph)
bar-graph OLED display
SpO2 Parameter Specification
Measuring range
0%100%, (the resolution is 1%).
Accuracy
70%100%±2% ,Below 70% unspecified.
Average value
Calculate the Average value in every 4 measure value. The deviation between average value and true value does not
exceed 1%.
Pulse Parameter Specification
Measuring range
30bpm250bpm, (the resolution is 1bpm)
Accuracy
±2bpm or±2% (select larger)
Medical Test Supply
CMS-50FF
Average pulse rate
Moving calculate the Average pulse rate every 4 cardio-beat’s cycle.
The deviation between average value and true value does not exceed 1%
Safety Type
Interior Battery,BFType
Pulse Intensity
Range
Continuous bar-graph display, the higher display indicates the stronger pulse.
Battery Requirement
Voltage 3.7 rechargeable lithium battery × 1
Battery working life
Charge and discharge no less than 500 times.
Power Adapter
Input Voltage
100 to 240 VAC, 50/60 Hz
Output voltage
5 VDC
Output current
250mA
Output power
1.25 W
Oximeter Probe
Wavelength:660nm 880nm
Dimensions and Weight
Dimensions
61(L) × 56(W) × 24 (H) mm
Weight
About 50g (with the lithium battery*1)
Appendix 1
State
Alarm condition delay
Alarm signal generation delay
Low voltage alarm
0.6s
20ms
Spo2 alarm
400ms
20ms
Pulse rate alarm
400ms
20ms
Probe error alarm
400ms
20ms
Medical Test Supply
CMS-50FF