MSE UK Ltd is required under the Health and Safety at Work Act, 1974 and other UK legislation as designers,
manufacturers, suppliers and importers of articles for use at work to ensure that, as far as is reasonably practicable,
articles which we design, produce, supply or import are safe and without risk to health.
We are required to provide information on the safety and handling precautions to be observed when installing,
operating, maintaining and servicing our products. Such advice is contained in this manual.
We are also obliged to update this information should circumstances change and to operate a system to this end.
We should also like to point out, however that you as users have an important responsibility in the provision
and maintenance of safe working practices and conditions.
Accordingly, we draw the following matters to your attention:
1. This apparatus should only be used as intended and within its design parameters by
suitably qualified and trained personnel who have read and understood the relevant
sections of this manual.
2. This manual should be readily available to such personnel at all times.
3. In addition to that which is written in the manual, normal common-sense safety precautions
must be taken at all times to avoid the possibility of accidents. Particular care is required
when working with apparatus at high temperature or pressure.
4. Installation, maintenance, repairs and servicing should only be carried out by a MSE (UK)
Ltd approved engineer, and suitably trained personnel should only carry out connection to
electrical supplies.
TECHNICAL SUPPORT, WARRANTY SERVICE AND MAINTENANCE
UK customers; if you are in any doubt whatsoever regarding the correct use of this apparatus, or if you require any
technical data or assistance, please contact the MSE UK Ltd Technical Support Department at:
OVERSEAS CUSTOMERS: Should contact their local Distributor.
ELECTRICITY SUPPLIES: Voltage and frequency
MSE (UK) Ltd electrical apparatus is offered and labelled in a number of voltage ranges and, where necessary,
different frequencies of mains supply. MSE UK Ltd does not accept any responsibility for the operation of any such
apparatus should it be connected to electricity supplies which are normally outside, or vary outside, the stated
voltage and frequency values for which it is designed, nor for any consequential loss, damage or injury, howsoever
caused.
Page 3
MSE UK Ltd Warranty
Terms and Conditions
We hope that you do not have the need to use the extensive warranty cover that MSE UK Ltd.
extends to you. However should you have a problem, our prompt response is greatly helped if you have
filled in and posted the pre-paid Warranty Registration Card supplied with your new equipment.
This entitles you to a one-year warranty from the date of delivery. During this period, component
parts proven to be defective in materials or workmanship will be repaired or replaced at our expense.
Installation, commissioning and calibration are not covered by this warranty agreement. The MSE
approved service agent must be contacted for warranty determination and direction prior to any work
being carried out.
These warranties are only applicable to new products, and not second hand nor refurbished
products even if repaired by MSE UK Ltd. Any such products are covered by separate warranty terms and
conditions which will be made available on request.
Replacement or repair of component parts or equipment under this warranty shall not extend the
warranty to either the equipment or the component part beyond the original one-year warranty period
unless agreed in writing by MSE UK Ltd.
The above warranties are extended to the original purchaser upon full invoice payment. A
purchase receipt or other proof of purchase may be required before warranty service will be performed.
These warranties only cover failures due to defective workmanship which occur during the normal
operation of the product by the original purchaser, and not failures which result from accident, misuse,
abuse, neglect, mishandling, misapplication, alteration, faulty installation, electrical power fluctuations,
dust, or other environmental extremes, modification or service other than by an approved service agent or
following the written authority of the manufacturer, or damage that is attributable to acts of God.
Expendable items such as motor brushes, door seals, lid seals, "O" rings or lamps are excluded.
MSE UK Ltd, or its approved service agent, reserves the right to repair defective equipment on
the premises of the customer, or at a service station, at the sole discretion of MSE UK Ltd or their
approved agent. In the event of return to an approved service centre the customer is responsible for the
safe packaging of the instrument and notification to the service centre. Neither MSE UK Ltd nor its agents
are responsible for any damage occurring during shipment.
Specification and Material Changes: MSE UK Ltd reserves the right to supply our latest and
improved models at time of shipment.
Taxes: The prices quoted do not include any taxes imposed by the State or Country in which the
purchase was made.
Installation: Installation of all equipment shall be at the expense of the purchaser unless stated
otherwise. Access to the site, and the provision of required utilities e.g. Power, water and drainage to
suitable connections, will be the responsibility of the purchaser, and at the purchaser's expense.
Limitation of liability: In no event, whether as a result of breach of contract or warranty, shall MSE
UK Ltd be liable for any consequential or incidental damages including, but not limited to, loss of profit or
revenues, loss of use of the equipment or any associated equipment, down time costs, costs of substitute
equipment, costs of labour, costs due to delays or claims of purchaser's own customers for such
damages. The purchaser agrees to indemnify MSE Ltd and to hold them harmless from any and all
liability, claims, demands, actions, suits, expenses or costs, including attorney's fees relating to such
consequential or incidental damages.
All expressed and implied warranties, including the warranties of merchantable quality and fitness
for a particular purpose, are limited to the application period of one year.
Validity: Legal rights vary from country to country and states within countries, so some or all of
the exclusions or limitations listed above may not apply, but if any part of these conditions shall be found
to be unenforceable it shall not affect the validity or enforceability of the remainder of the conditions.
Page 4
ELECTRICITY SUPPLY
Before connecting this apparatus to the electricity supply, check the information given
on the apparatus rating plate and ensure that;
A) Your supply is single phase A.C. (alternating current) of the stated frequency
with neutral nominally at earth potential.
B) Your supply voltage is within the stated range.
C) The current rating is within the capacity of your outlet.
D) Your plug or electricity supply circuit is fitted with a suitable fuse.
220v - 240v 110v - 120v
Fuse rating 5 amp see note for 110V - 120V
WARNING! This apparatus must be earthed.
The wires in the mains lead are coloured in accordance with the following code;
220v - 240v 110v - 120v
Live Brown Black
Neutral Blue White
Earth Green and Yellow Green
Connect the wires to a non-reversible 3-pin plug as follows;
Green and Yellow or Green to terminal marked E (Earth), G (Ground), coloured Green
or Green and Yellow or marked with the Earth symbol.
Blue or White wires to terminal marked N (Neutral) or Common or coloured Blue.
Brown or Black wires to terminal marked L (Live) or Phase or coloured Brown.
Note 110v - 120v installations to comply with National and State Wiring Codes.
IMPORTANT Consult an electrician if in any doubt or if your supply system has any of
the following:
No earth.
A coloured code different from the above.
Reversible plugs.
Supply and return leads that are both above earth potential.
NOTE: Do not touch plug pins within two seconds of plug removal from socket.
Page 5
CONTENTS 5
1. IMPORTANT INFORMATION.................................................................................7
2. GENERAL DESCRIPTION.....................................................................................8
2.1 Controls And Indicators..................................................................................9
10. PRODUCT DISPOSAL - ISO 14001 COMPLIANCE......................................34
11. HOW TO OBTAIN SERVICE ON YOUR FALCON CENTRIFUGE ...............35
Page 7
GENERAL INFORMATION 7
1. IMPORTANT INFORMATION
The following general instructions, and the precautions regarding the
prevention of corrosion in rotors and accessories, as described under
MAINTENANCE, must be noted and carefully observed.
The maximum load should always be checked before commencing any operation. If
the sample has a relative density greater than 1.2, the maximum permissible speed of
the rotor should be calculated as given in this publication, and applied.
Any liquid spilt in the centrifuge should be removed immediately because corrosion is
a major cause of failure in centrifuges. Where particularly corrosive solutions are used,
the samples should be placed in sealed containers and all necessary precautions
observed. The following list gives some examples of corrosive liquids used, particularly
in the biochemical field:
Phenol/cresol/water, chloroform/isoamyl alcohol, salt solutions (especially ammonium
sulphate), solutions of ammonium hydroxide and solutions of acids such as
hydrochloric, trichloroacetic and perchloric acids.
It should be noted that many other substances used in different fields are often equally
corrosive. Users should always check whether or not a substance being used is
corrosive in order to take suitable precautions.
NOTE: The following materials are prohibited:
Flammable or explosive materials
Materials which chemically interact vigorously
The load must be distributed symmetrically round the rotating assembly so that it is
evenly balanced and complies with the manufacturer's instructions in this respect.
After use the sleeves and adaptors in centrifuge rotors should be removed to allow
thorough cleaning of rotors and accessories and so prevent corrosion.
Special care is necessary when using infective material in centrifuges. Containers,
when used with such materials, should be closed while being centrifuged and sterilised
immediately after use, using non-corrosive means.
NOTE: Sealed containers and related components are intended to be part of
biosafety systems such as are specified in international and national
biosafety guide lines, and cannot be relied on as the only means of safeguarding
workers and the environment when handling pathogenic
micro-organisms.
Care should be taken to preserve intact, sets of accessories which have been supplied
as sets, in which all the members possess closely similar weights.
It is advisable to have the centrifuge checked by a competent engineer, preferably the
manufacturer's representative, at least every six months.
Carbon brushes, whilst being of the highest quality obtainable are expressly excluded
from the guarantee. A spare set is included in the accessory pack. The method of
replacement is detailed in section 6.7 of this manual.
Page 8
GENERAL DESCRIPTION 8
2. GENERAL DESCRIPTION
The Centaur 2 centrifuge is CE marked in line with required European Directives
for Electromagnetic and Low Voltage Compatibility, see “Declaration of
Conformity” for EC Directive/standard used.
120v units designed for CSA approval to the following standards:
CAN/CSA C22.2 No. 1010.1-92
CAN/CSA C22.2 No. 1010.2.020-94
INDEPENDENT SPEED SENSING - 120 volt Machines only
To comply with certain USA standards the 120 volt machines are fitted with a
small window in the cover. This permits independent speed check directly from the
rotor with use of an optically coupled tachometer.
Page 9
GENERAL DESCRIPTION 9
2.1 Controls And Indicators
Figure 1.
Control Panel of the Centaur 2 Centrifuge
1. The Power On rocker switch, when on (switch illuminated) connects power
through to various circuits and components within the centrifuge.
2. The Fuse Holder carries a 20mm long, 5 Amp glass cartridge fuse. It is easily
removable.
3. The Timer control allows a run period up to 30 minutes-selected by turning the
knob clockwise. When the time selected has run out, the rotor will automatically
run down to a stop. If required, the knob may be returned (anti-clockwise) to a
lesser time or to the Off position at will. It cannot be turned anti-clockwise
between the 'Hold' and '30' 'positions. The Hold position switches the centrifuge
for continuous running until manually reset to Off.
4. The Rev/min indicator is a digital display giving continuous read-out of the
rotational speed of the rotor, rounded off to the nearest hundred rev/min, e.g.
2500, 3200.
5.
The Speed control allows selection of any speed between 300 rev/min and the
maximum allowed for the rotor in use (see section 4.5 Rotors And Accessories
Tables) as indicated on the Rev/min digital display.
6. The Imbalance indicator lamp (Red) when on, gives warning that the rotor is
misloaded to a marked degree. In this event, drive to the rotor will cease and
the rotor will run down to a stop. Loading and balance must be checked and
corrected before re-starting the centrifuge.
7. The Lid Unlocked indicator lamp (Amber) when on, indicates that the rotor is
at a standstill and the lid may be opened safely. When the rotor is spinning the
lamp remains off and the lid is locked.
Page 10
GENERAL DESCRIPTION 10
2.2 Centrifuge Bowl
The centrifuge bowl is easily removable for cleaning or disinfection purposes see
section 6 (Maintenance).
2.3 Opening Lid Manually
An electrical interlock prevents the lid being opened while the rotor is turning, or
when the power is disconnected. If access to the rotor is required during a power
failure, insert the Lid Lock Release Key (supplied with accessories) to its fullest
extent, into the small hole on the top of the lid, near the front; move the point
towards the rear against a slight spring pressure to actuate the lid lock
mechanism. Press the lid actuator to the right and open the lid.
The lid is friction-controlled, to remain open at any intermediate position desired.
Front
Figure 2.
Manual Over-ride of Lid Lock
2.4 Specifications
Dimensions
Height 336mm (with lid open - 470mm)
Width 436mm
Depth 487mm
Weight 36kg (without rotor)
Power Supply
Input voltage and frequency.
Cat. No. MSB020.CX1.5 220/240V 50/60 Hz.
Cat. No. MSB020.CX1.1 110/120V 50/60 Hz.
Operating
Humidity MRH 80% up to 31°C decreasing linearly to 50% RH at 40°C
Altitude Up to 2000M
NOTE: Motor, as fitted, contains overtemperature cut-out device.
3. INSTALLATION
IMPORTANT: The rotor must not be fitted when moving or during transit.
3.1 Preparation
Remove the centrifuge from its packing and place it on a rigid, level surface.
Check that the available power supply corresponds to that stated on the voltage
plate fixed to the rear panel of the centrifuge.
NOTE: (i) An electrical interlock prevents the lid being opened unless the
centrifuge is connected to the power supply.
(ii) It is important that, at least, 100mm clearance is allowed at the
sides and rear of the centrifuge when placed in the final operating
position.
NOTE: Laboratory management procedures should require that no person or
any hazardous materials are within a 300mm boundary while the
centrifuge is operating.
3.2 Connecting To 3 Pin Plug
Connect the 3-core cable to a 3-pin plug fitted with a 5 Amp fuse as follows:
WARNING! This equipment must be earthed.
Brown wire to Live (L) Terminal
Blue wire to Neutral (N) Terminal
Yellow/Green wire to Earth (E) Terminal
Connect the centrifuge to the power supply and press the Power switch on. The
Power On and the Lid Unlocked indicators will come on. Push the lid lock actuator
to the right and open the lid.
NOTE: BS EN 61010-2-020 ‘Particular Requirements for Centrifuges’ states that
a remote switch, preferably adjacent to the room exit is a requirement
in case of unit malfunction.
Page 13
INSTALLATION 13
3.3 Recommendation For Securing Centrifuge
A Fixing Kit for securing the Centaur 2 centrifuge to a suitably sound worktop is for
securing centrifuge available from MSE as an optional accessory. It consists of an
anchor bracket which is secured to the bench, or table, by two screws with nuts
and washers; and two screws which screw into the body of the centrifuge to
secure it to the bracket. See Figure 3.
Figure 3
Securing Centrifuge to work surface using Bench fixing Kit
(Part No. 41159.188)
If the restraint kit is not used to secure the centrifuge on the bench then, it is the
recommendation of BS EN61010-2-020 that a clearance of 300mm is allowed
around the base of the centrifuge when in the final operating position.
NOTE: Laboratory management procedures should require that no person or any
hazardous materials are within a 300mm boundary while the centrifuge is
operating.
Figure 3a Recommended Clearances
Note: 75mm if solid surface (i.e. wall) otherwise 300mm
Page 14
ROTORS AND ACCESSORIES 14
4. ROTORS AND ACCESSORIES
4.1 Installation Of Rotor
Before fitting a rotor to the centrifuge ensure that the drive shaft is lightly smeared
with petroleum jelly. Check that the rotor is clean - particularly pockets in angle
rotors.
Locate the rotor on the drive shaft and, ensuring that the correct spindle nut for the
rotor in use is employed (see sections 0 & 4.5), engage and tighten the spindle nut
finger tight.
4.2 Removal Of Rotor
Press Power switch off. Open the centrifuge lid. Remove all rotor accessories, i.e.
buckets, trunnions etc. Loosen the spindle nut a few turns, grip the rotor as
illustrated in Figure 4 and pull upwards with the fingers whilst gently rocking the
rotor and applying downward pressure with the thumbs, to release it from the drive
shaft. Remove the nut and lift the rotor free.
Note: Do not attempt to loosen the rotor by tapping on the centrifuge motor
spindle.
Figure 4.
Removing Rotor from Drive Shaft
Page 15
ROTORS AND ACCESSORIES 15
4.3 Loading The Rotor
The rotors must carry a full complement of cups/buckets and/or trunnion rings, but
all cups/buckets need not be loaded. Loads must be placed symmetrically about
the rotor and opposing loads should balance to within 2 grams. Where possible,
balancing by sight is usually sufficient.
Before loading centrifuge tubes ensure that the appropriate colour coded-adapter
or rubber cushion is correctly positioned in the cup/bucket. Check that all tubes
project equally from the upper surface of the rotor.
Finally, if using a swing-out rotor, check that all cups of trunnion rings swing freely.
If necessary, apply a light smear of petroleum jelly to the rotor pivot slots.
Failure to load the rotor correctly can result in serious imbalance and possible
damage to the centrifuge.
Biologically hazardous material should only be centrifuged in sealed
buckets or cups.
4.4 Rotor Speed
Maximum Speeds of rotors (See section 4.5) are calculated on the basis of tubes
filled with a sample of specific gravity 1.2. If liquids of higher specific gravity are
used, the maximum speed of the rotor must be reduced according to the formula:
Where M = New maximum speed
N = Maximum rotor speed at 1.2 specific gravity
S = Specific gravity of sample used
Page 16
ROTORS AND ACCESSORIES 16
4.5 Rotors And Accessories Tables
The centrifuge complies with the requirements of BS4402 only when fitted with one of the following rotors.
* Max. rev/min at 240V or 110V ** Supplied as standard with buckets # Available as spares
34411-912 34134-604 34142-105** 4 x 100ml 3500 2246 42 125
Sealing cap 34144-308 4 x 50ml 3500 2246 29 120
43561-156 can be 34151-302 8 x 15ml 3500 2246 17 118
used if required 34153-302 20 x 7ml 3500 2246 13 118
34134-605 34142-104** 4 x 50ml 4200 3130 30 118
34411-909 34148-705 4 x 50ml 4200 3130 conical 29 112
Sealing cap assy 34141-114 4 x 1oz 4200 3130 Glass
43561-606 can be 34141-115 4 x 1oz 4200 3130 Sterilin
used if required 34144-307 4 x 15ml 4200 3130 17.5 110
34153-304 24 x Haem 4200 3130 7.7 120
34151-301 8 x Wass 4200 3130 12.7 120
34411-904 34136-102 34142-101** 8 x 15ml 4000 2612 17 120
34136-103 34142-101** 16 x 15ml 3400 2000 17 120
43161-103 16 x 15ml 3400 2000 conical 17 115
34148-101 16 x Haem. 3400 2000 7.7 120
34146-104 16 x 5 or 3ml 3400 2000 13 120
34146-103 16 x 2ml 3400 2000 12 100
34146-102 16 x 0.5 or 1ml 3400 2000 9 100
Rotor
Capacity
Max.*
Speed
Max.
RCF
Max. Tube Size
(mm)
(mm)
Ht
Page 18
ROTORS AND ACCESSORIES 18
4.5.3 8 Place Swing out rotor
43121-111
(includes buckets and spindle nut)
Bucket Trunnion Adaptor/
Cushion
Dia
34411-905# 34142-101** 8 x 15ml 3500 2200 17 115
Rotor
Capacity
Max.*
Speed
Max.
RCF
Max. Tube Size
(mm)
Ht
(mm)
Page 19
ROTORS AND ACCESSORIES 19
4.5.4 8 Place Angle rotor
34117-603
Bucket Trunnion Adaptor/
Cushion
Dia
34411-909 34142-104** 8 x 50ml 4600 3120 29 110
34411-909 34148-705 8 x 50ml 4600 3120 conical 29 110
34411-909 34141-114 8 x 1oz 4600 3120 Glass
34411-909 34141-115 8 x 1oz 4600 3120 Sterilin
34411-909 34144-307 8 x 15ml 4600 3120 17.5 110
34411-909 34153-304 48 x Haem. 4600 3120 7.7 120
34411-909 34151-301 16 x Wass. 4600 3120 12.7 120
Rotor
Capacity
Max.*
Speed
Max.
RCF
Max. Tube Size
(mm)
(mm)
Ht
Page 20
ROTORS AND ACCESSORIES 20
4.5.5 12 Place Angle rotor
34117-604
Bucket Trunnion Adaptor/
Cushion
Dia
34411-904 34142-101** 12 x 15ml 4600 3000 17 110
34411-904 43161-103 12 x 15ml 4600 3000 conical 17 110
34411-904 34148-101 12 x Haem. 4600 3000 7.7 110
34411-904 34146-104 12 x 5 or 3ml 4600 3000 13 120
34411-904 34146-103 12 x 2ml 4600 3000 12 100
34411-904 34146-102 12 x 0.5 or 1ml 4600 3000 9 100
Rotor
Capacity
Max.*
Speed
Max.
RCF
Max. Tube Size
(mm)
(mm)
Ht
Page 21
OPERATION 21
5. OPERATION
5.1 Run Procedure
Ensure that the Rotor is correctly loaded and secured and the Timer set at Off.
Close the lid and press down firmly. Press the Power switch on. The Power On
and the Lid Unlocked indicators will come on.
Select Timer to Hold, for continuous running, or to the time required. The lid will
now be locked - Lid Unlocked indicator off.
Select required speed (between 300 rev/min and the maximum speed of the rotor
in use, see section 4.5) by adjusting the Speed control until the required speed is
indicated on the Rev/min digital display. For repeat runs the Speed control need
not be moved from the position originally selected.
5.2 Stop Procedure
To terminate a run in the Hold mode, set the Timer switch to off and the rotor will
run down smoothly to a stop. In the Time mode the centrifuge will run down
automatically when the selected time has run out. In both modes automatic
braking is applied, and when the rotor has stopped, the Lid unlocked indicator will
come on. Open the lid; the Lid unlocked indicator will go off.
5.3 Imbalance
Should the imbalance indicator lamp come on, the centrifuge will run down. When
the digital panel meter indicates four zero’s, turn the Power switch and Timer to off
and wait thirty seconds, then turn the Power switch on. The lid unlocked indicator
will come on; open the lid; check the balance of the rotor; close the lid and restart
the run.
5.4 Motor Thermal Cut-out
The motor is fitted with a ‘self re-setting’ thermal cut-out to protect the motor from
over-heating. If the thermal cut-out has operated during a run the rotor will stop
and ‘0000’ will be displayed. The centrifuge will return to normal operation once
the motor has cooled.
IMPORTANT: ENSURE THAT THE VENTILATION SLOTS AT THE BACK ARE
NOT OBSTRUCTED IN ANY WAY.
5.5 Power Failure
If the mains power to the centrifuge is interrupted during a run, the rotor will coast
down to a stop and the lid will remain locked. Turn the Timer control to off.
When the power is restored, re-start the run as normal. If it is required to gain
access to the rotor during the power failure, use the Lid Lock Release Key (see
section 2.3).
If power failure occurs and the centrifuge is left unattended, subsequent
restoration of power will cause the centrifuge, if in the Hold mode, to run up to the
originally selected speed. If in the Time mode, the centrifuge will run up to speed
and continue for the time left on the Timer. If the selected time runs out during the
power failure, the lid will remain locked for an inherent delay period of
approximately 3½ minutes from the time of power failure, even if the power is
restored within this period.
Page 22
OPERATION 22
5.6 Lid Safety Circuit
The Lid Safety Circuit has a dual function:
i. It prevents power being switched to the motor until the lid is properly closed.
ii. It imposes a delay of approximately three and a half minutes in the event of a
tacho failure before the Lid Unlocked indicator comes on and the lid can be
opened.
Two micro-switches in the lid form part of the lid safety circuit. These can be
checked for correct operation as follows:
i. Press Power switch on and open the lid. The Lid Unlocked indicator lamp
should be off.
ii. Close the lid. The Lid Unlocked indicator lamp should come on. If the lamp
does not come on, check that the lid is firmly closed as this is the most likely
cause of malfunction .
iii. Switch the Timer to Hold or select a Time. The Lid Unlocked indicator should
go off. If this does not occur, then the Lid Locked microswitch has not
operated. Again, the most likely cause is improper closing of the lid.
Page 23
MAINTENANCE 23
6. MAINTENANCE
DISCONNECT THE CENTRIFUGE FROM THE POWER SUPPLY
User maintenance is minimal, being confined mainly to cleanliness and care of
rotors and accessories. Basic equipment maintenance must be carried out by a
qualified/competent electrician.
6.1 Cleanliness
Ensure that the body of the centrifuge is cleaned periodically to maintain a good
appearance. It is particularly essential to clean out the centrifuge bowl thoroughly,
especially when corrosive materials have been used. A soft cloth dampened in a
solution of mild detergent and warm water will serve for both purposes.
6.2 Drive Shaft
The centrifuge drive shaft should be cleaned periodically with a solvent to remove
excessive grease. When clean, the shaft should be re-lubricated lightly with
petroleum jelly.
6.3 Prevention Of Corrosion
Simple corrosion can usually be detected by eye and appears in the form of
pitting, or white, fluffy deposits on the surface of aluminium rotors. To prevent
corrosion, take care to avoid scratching the surface of rotors and/or accessories.
After use wash rotors, metal buckets, cups, carriers and adaptors in warm soapy
water, dry and store, inverted, in a dry place. Rotors should be protected with MSE
Rotor Spray (Cat. No.17341.151), which is an anti-corrosion and
de-waterising agent.
In addition, rotors should be periodically inspected by a responsible person so that
corrosion can be detected at the onset. Contact MSE (UK) Ltd Technical Support
for further advice on this subject.
6.4 O-Rings
Ensure O-Rings are lightly coated with silicone grease. Check O-Rings regularly
for cuts and abrasions. Seals should be replaced as necessary, or at least once a
year.
6.5 Sterilisation
Except for the 8-place swing-out rotor Cat. No. 43121.111, all rotors, buckets,
sealing caps (O-Rings removed) and the bowl may be sterilised by autoclaving at
120°C. See section 6.9 for bowl removal procedure. If the means available for
disinfection of certain microbiological agents are inadequate, the safety officer
should be consulted and appropriate steps taken
6.6 Disinfecting
The following cleaning fluids may be used
TERMINEX 2 (Available from Arrow Chemicals Ltd.)
VIRKON (Available from Antec International)
These cleaning agents if used as instructed by the manufacturer should not be
harmful to this product, or accessories supplied for use with this product.
WARNING: SOLVENTS OR GRITTY CLEANERS SHOULD NEVER BE USED
EXCESSIVE USE OF WATER SHOULD ALSO BE AVOIDED
6.7 Inspection Or Renewal Of Motor Brushes
Page 24
MAINTENANCE 24
The motor brushes should last for a minimum of 1000 hours of operation.
To renew brushes proceed as follows:
Open the lid and remove the rotor, close the lid, and disconnect the centrifuge
from the power supply.
Turn the centrifuge on to its back Unscrew and remove the two screws securing
the front rubber feet, and the other two screws that locate and secure the
instrument chassis to the base frame. Place corrugated cardboard or other
protection on the upper part of the centrifuge, then withdraw and lay the chassis
on top
Removing The Brushes
Unscrew the brush-retaining cap. The brush is held within a carrier against a
tensator roll spring. If the carrier does not spring free on removal of the cap, a thin,
stiff wire, hooked at one end may be used to assist. Do not use force. Gently
press the arms of the carrier together and withdraw the brush. On withdrawal, note
the position of the roll spring in the carrier in relation to the brush and motor as the
new brush must be fitted in the same manner. It will be seen that a small pin, fitted
near the rear end of the brush locates in a guide slot along one arm of the carrier.
Note that the free end (tongue) of the roll spring is held, near the end of one arm
of the carrier, facing in the direction of fitting.
Fitting New Brushes
Locate the new brush in the carrier with the pin located in the slot and the brush
pushed fully into the carrier against the tension of the roll spring Hold the
assembly together with the roll spring in the same relationship to brush and motor
as when the old brush was removed, and insert the front end into the brush
housing of the motor Take care to ensure that the 'tongue' of the roll spring lies,
facing forward, flat with the carrier (if not, gently press it down with a small
screwdriver) before pushing the assembly home into the brush housing. Note: If
the 'tongue is bent backwards against the carrier upon insertion, it may break off
and the spring will be ineffective Check that the carrier moves freely in and out
against the tension of the spring. Refit the brush-retaining cap and tighten fingertight. Replace and secure the instrument chassis and the rubber feet and turn the
centrifuge the right way up. Run the centrifuge with an unloaded rotor on the drive
shaft to a speed of 1000 rev/min for half-an-hour, to bed in the brushes.
6.8 Renewing The Fuse
To renew the fuse: disconnect the centrifuge from the power supply at the mains
plug. Unscrew the fuse holder cap and remove the fuse. Inspect and replace as
necessary with the correct rating. For 220V/240V model, the plug fuse may also
need to be checked. Fit the new fuse and refit the holder back into the receptacle.
If the fuse blows again during use, contact the local distributor, or MSE (UK) Ltd.
NB - If the fuse blows immediately or blows again during normal use, the power
supply should be disconnected and the service engineer called.
Page 25
MAINTENANCE 25
6.9 Removal And Replacement Of Centrifuge Bowl
To remove the bowl for cleaning or disinfection purposes proceed as follows:
Open the lid, disconnect the centrifuge from the power supply and remove the
rotor (if fitted).
Peel back the lip of the seal above the bowl vent (inside the bowl at right) note
position of the vent then lift the bowl by the vent hole above the seal at this point.
Now lift the bowl by the centre boss and work the seal lip away from around the
entire circumference of the bowl.
To refit the bowl:
Bend the lip of the seal to point upwards at any convenient position. Lower the
base of the bowl on to the seal and work the seal around the side of the bowl so
that the whole seal edge is pointing upwards and embracing the lower part of the
bowl. Lower the bowl, applying slight downward pressure with a twisting motion
and ease downwards until the seal lip is just under the bowl rim, with the vent in
the bowl in original (noted) position.
Insert the lid seal tool (Figure 5.) between the seal and the lip of the bowl and
ease the seal over the lip of the bowl while applying downward pressure on the
bowl centre boss. Work the tool round the rim of the bowl gently easing the seal
over the bowl. Exert further downward pressure on the boss to locate the bowl rim
in the recess of the seal.
Figure 5.
Use of lid seal tool
6.10 Lid Friction Adjustment
If the lid friction collar needs adjustment, e.g., because lid falls slowly downwards,
turn the socket screw, which is visible through the small hole located centrally near
the lower edge of the inside of the lid, using a 4mm socket wrench. Turning
clockwise will tighten, and anti-clockwise will loosen, friction resistance on the lid.
See Figure 6 (upper inset, annotation 20).
Page 26
SERVICING 26
7. SERVICING
This section has been included for the guidance of competent service
engineers only - preferably those trained by MSE (UK) Ltd.
7.1 Calibration Of Speed Control
Disconnect the centrifuge from the power supply Remove the instrument chassis
(see section 6.7 for removal instructions). Refit the two rubber feet temporarily and
set the centrifuge upright with the instrument panel at the front.
Connect 2 frequency metre to test points TP1 (ground) and TP2 (signal) at the
right hand side of the circuit board.
Place a correctly assembled and balanced rotor (see section 4) on the drive shaft,
close the lid and set the speed control and RV2 - PCB 'D' fully anti-clockwise, set
RV3 - PCB 'A' fully clockwise.
Switch on, turn the timer to ‘Hold’ and check that the rotor rotates slowly and the
centrifuge display indicates 100 rpm, if necessary, adjust RV2 - PCB 'A' (min
speed set) to achieve this condition.
Set the speed control fully clockwise, adjust RV1 - PCB 'A' until the frequency
meter reading matches half (1/2) that of the centrifuge display reading, adjustment
clockwise will increase the display reading whilst decreasing the meter reading
and vice-versa for anti-clockwise.
Adjust RV2 - PCB 'D' (max. speed set) so that the centrifuge runs at the highest
speed indicated on both the centrifuge and frequency meter displays.
7.2 Removal Of The Motor
Ensure that there is no rotor on the drive shaft. With the centrifuge turned on to its
back and the instrument chassis removed (section 6.7), remove the remaining
screws around the edges of the base plate. Making a note of terminations and
connections for reference, disconnect plugs and wires and remove the base-platewith-motor. The upper part of the motor is partially secured by the rubber support
disc in the body and can be pulled out.
Remove the three nuts and spring washers which secure the three rubber mounts
to the base plate. Remove the motor assembly from the base plate.
Remove the two hexagon screws, washers and spacers (note arrangement) and
remove the optical detector printed circuit board. Remove the screw in the lower
end of the motor and carefully remove the plastic tacho disc. The motor can now
be detached from the tacho box by removing the three small screws around the
centre of the tacho box.
Carefully note location and position of the motor support bush around the body of
the motor before removal. Remove the motor support bush and fit it to the new
motor as noted, before re-assembly.
Re-assembly of the motor to the tacho box and base plate is a reversal of removal
procedures. Ensure that the tacho disc is centred within the slot of the optical
detector. Ensure also that the three rubber mounts are not twisted during reassembly. Adjust balance detector as described in section 7.3.
Page 27
SERVICING 27
7.3 Adjustment Of Balance Detector
With the centrifuge upright and after operation with a loaded rotor for a few
minutes, to allow the motor to settle in its operating position, insert a 1mm thick
gauge between the microswitch actuator and the side of the tacho-box
(Item 6. Figure 6).
By turning the Balance detector adjusting screw, Item 7, Figure 6 adjust the
microswitch towards the side of the tacho-box until the microswitch just operates
(breaks).
Ensure that the microswitch operates (re-makes) when the gauge is removed.
Page 28
SERVICING 28
Figure 6
Partially Exploded View Of Centaur 2
Page 29
rev
Min
/
2
CIRCUIT DESCRIPTION 29
8. CIRCUIT DESCRIPTION
8.1 Power Input
A.C. mains power is fed to the centrifuge via a double-pole switch SW1, a 5 Amp
fuse FS1 and a filter. The Power switch SW1, when switched on, is illuminated by
a green neon lamp.
8.2 Speed Control
The motor current is phase-controlled by a series triac MR1 which is triggered by
an opto-isolator IC1. The L.E.D. of the opto-oscillator is driven by a section of the
transistor driver IC2 which in turn is fed from two cascaded Schmitt triggers IC4.
The input terminal of the first Schmitt trigger is connected to a 68nF capacitor,
which is discharged to approximately 1 Volt during each mains cross-over by IC2
pins 2, 15 and 3, 14. the Potentiometer RV2 (PCB A) sets the low end of the
speed range and RV2 (PCB D) set the maximum speed. RV4 is for speed control.
The pulse width generated by this circuit is linearly proportional to the speed
control resistance. There is, however, some non-linearity in the speed versus
speed-control angle characteristic because the pulse is used to "chop" a sine
wave in order to vary the speed.
The feedback network R18, MR10 discharges the capacitor to give a narrow
trigger pulse, thus reducing the power requirement on the 12 Volt supply to IC1.
Voltage is fed to the speed control network via the Lid Closed and the Lid Locked
micro-switches SW2 and SW3 and one contact of the Timer switch, SW4/1, so
that the motor is driven only when the lid is closed and locked and the Timer is set
to Hold; or a time is selected. The voltage is stabilised at about 9 Volts by R17,
MR7 and the 8.2 Volts supply.
Relay A is also fed through the two micro-switches and the Timer switch contact.
When Relay A is energised, the motor field and armature are connected in series
with each other and the triac MR1, across the 240 Volts mains input. With Relay A
released (which occurs when the Timer switches off), the armature is shortcircuited and a D.C. voltage is applied to the field. This applies a braking force to
the motor.
8.3 Tacho Circuit
The tacho input is taken from an opto-electronic chopper unit, which gives an
output of
Hz. This input switches VT1 which provides a fixed voltage (≈8V)
square wave to the Speed Indication and Lid Safety Circuits.
Page 30
CIRCUIT DESCRIPTION 30
8.4 Speed Indication
The output of VT1 is fed, via a Schmitt trigger, to a monostable circuit (IC3 pins 4,
5, 6, 8, 9 and 10). The pulse width is fixed by a time constant of 22 nf and a 50k
variable resistor RV1. The output pulses are fed to the digital display unit (PCB D)
where they are integrated to give a D.C. voltage proportional to speed.
The digital display unit consists of a 7107 analogue-to-digital converter/driver I.C.
and its associated components, driving two LED's. The two most significant digits
are driven by the 7107 to indicate the speed. The two least significant digits are
connected through twelve 1k resistors (RN1) to give a fixed 00 display.
8.5 Lid Safety Circuit
When the Timer switch is made, contact SW4/2 applies a voltage to charge up
C8. This sets one input of the NOR gate IC3 pins 11 to 13, to ‘1’, thus deenergising the Lid Solenoid and locking the lid.
When the Timer switches to ‘Off’ the charging current to C8 ceases, but the
discharge time constant is long enough to delay the energising of the solenoid for
about three-and-a-half minutes. This ensures that the lid cannot be opened before
the rotor has stopped.
To allow the lid to be opened as soon as the rotor stops, the circuit comprising:
IC4 pins 1, 2, 3, 4, 5 and 6; IC3 pins 1, 2 and 3, together with VT2, monitors that
the run-down is normal and if so, discharges the capacitor C8. The method
employed uses the outputs of two frequency discriminators: IC4 pins 5 and 6
(directly) and IC4 pins 1, 2, 3 and 4 (after a delay) as inputs to a NOR gate.
If the 60 rev/min discriminator operates sufficiently in advance of the 6 rev/min
circuit, then a pulse is generated by the NOR gate which switches on VT2, to
discharge the capacitor. Under failure conditions no pulse will be generated and
the lid will remain locked for about three-and-a-half minutes.
8.6 Balance Circuit
IC2 pins 4/13 and 5/12 form a bi-stable circuit, which is forced into the required
stable state at switch-on by the capacitor C5. In this stable state pin 13 is near
earth so that Relay A can be energised; and pin 12 is high so that the balance
lamp is off.
If the rotor is unbalanced sufficiently to cause the Balance microswitch to open
momentarily, the bi-stable circuit will switch over to its other stable state. IC2 pin
13 will then be high, de-energising the relay. Pin 12 will be low, switching the
Balance lamp LP2 on and removing the charging current to C8, so that the lid is
unlocked after about three-and-a-half minutes.
The bi-stable can be reset by switching the mains supply off for a few seconds
and then on again.
8.12 Instrument Chassis Assembly
Part No. Description Remarks
32600.019 Fuse 5A (20mm)
32900.037 Fuse Holder
38700.046 Timer Switch, 30 min, 16A, 250V
44050.035 Potentiometer, Speed Control RV4
76070.099 Knob 28mm Black
76070.095 Knob 28mm Black
76070.096 Cap 28mm Black Speed/Timer Control
76070.097 Pointer 28mm Grey
36040.067 Rocker Switch 240V Power On
36040.068 Rocker Switch 110V
34060.013 Indicator Lamp, Red Imbalance
34020.087 Indicator Lamp, Amber Lid Unlocked
35080.046 Relay RLA
46100.003 Filter
38900.042 Transformer TR1
9. Product Disposal - ISO 14001 Compliance
This product should be treated as industrial waste and disposed of accordingly.
There are no toxic material used in the manufacture of this product. The
majority of materials used in this product are recyclable, and all can be disposed
of safely. Where the product has refrigeration, it is important that prior to
disposal the refrigerant gas is recovered by a qualified person. The insulation
material is non-toxic but could be an irritant. If removed from the product it
should be bagged and disposed of at an authorised site.
Page 35
HOW TO OBTAIN SERVICE 35
10. HOW TO OBTAIN SERVICE ON YOUR CENTRIFUGE
MSE UK Ltd is committed to giving our customers the best possible service. If
your centrifuge should require service at any time please follow these procedures:
All countries except UK, USA and CANADA - Contact your local MSE (UK) Ltd
distributor
UK only For all technical and service enquires contact: -
MSE (UK) Ltd
Worsley Bridge Road
Lower Sydenham
London
SE26 5AZ
Telephone +44 (0) 870 609 4097
Fax +44 (0) 208 650 8408
E-mail: sales@mseuk.co.uk
1. Contact the repairs centre - have the model, serial number, and date of purchase and
fault description available.
2. You will be given a return goods authorisation number and directions for shipping.
3. Remove all rotors, buckets and adapters. Do not ship these items - only the centrifuge.
4. Thoroughly clean and disinfect the centrifuge.
5. Fill out the attached service request form and place inside the centrifuge.
6. Pack in a protective box (preferably that in which the centrifuge was originally supplied).
7. MSE (UK) Ltd will specify the carrier to be used and will give details of how the freight is
to be charged.
Page 36
SERVICE REQUEST FORM
Should it become necessary to have your MSE centrifuge repaired, please take a few
moments to fill out this form, which will help us to ensure you receive the best and fastest
service possible.