The mask/helmet combination series Ultra Elite-H/F1 with standard thread connection according to
EN 148-1 are not complete respiratory protective devices by themselves, but serve as facepieces in
accordance with DIN 58610 [EN 443 and EN 136 CL3+ with the exception of selection 7.11] for use
with respiratory filters, compressed air breathing apparatus, fresh air hose breathing apparatus and
compressed air line breathing apparatus.
The mask as part of a respiratory protective device ensures an appropriately tight fit of the user’s face
against the ambient atmosphere.
The helmet as part of the whole protective device ensures the protection of the upper half of the head
mainly against the effects of impact, penetration, heat and flames during firefighting.
The masks can only be used in conjunction with following MSA GALLET helmets:
Safety Regulations
•F1SF [EN 443:2008]
•F1E [EN 443:2008]
The instructions for use for the helmets have to be regarded.
The mask/helmet combination series Ultra Elite-H/F2 X-TREM with standard thread connection
according to EN 148-1 are not complete respiratory protective devices by themselves, but serve as
facepieces in accordance with DIN 58610 [EN 136 CL3+ with the exception of selection 7.11].
WARNING!
The helmet type F2 X-TREM is not certificated according to EN 443. Therefore this helmet in combination with the mask must not be used for firefighting in buildings and structures.
The mask/helmet combination provides the function of head protection and respiratory protection in
one system.
WARNING!
According to the European directive 89/656/EWG it has to be verified before first use of the mask that
a correct size has been chosen [proper tight fit], that the mask can be worn in combination with other
protective equipment [for example a protective jacket], that it is a correct choice for the conditions at
the particular place of use and that it fulfils the ergonomic requirements.
It is imperative that this operating manual be read and observed when using the device. In particular,
the safety instructions, as well as the information for the use and operation of the device, must be carefully read and observed. Furthermore, the national regulations applicable in the user's country must be
taken into account for a safe use.
WARNING!
This product is supporting life and health. Inappropriate use, maintenance or servicing may affect the
function of the device and thereby seriously compromise the user's life.
Before use the product operability must be verified. The product must not be used if the function test
is unsuccessful, it is damaged, a competent servicing/maintenance has not been made, genuine MSA
spare parts have not been used.
•Gallet F1 XF [EN 443:2008]
GB
Ultra Elite-H Mask – Negative Pressure
5
Alternative use, or use outside this specification will be considered as non-compliance. This also
applies especially to unauthorized alterations to the product and to commissioning work that has not
been carried out by MSA or authorized persons.
1.2Liability Information
MSA accepts no liability in cases where the product has been used inappropriately or not as intended.
The selection and use of the product are the exclusive responsibility of the individual operator.
Product liability claims, warranties also as guarantees made by MSA with respect to the product are
voided if it is not used, serviced, or maintained in accordance with the instructions in this manual.
Changes and modifications not expressly approved by the manufacturer will void the user’s authority
to operate the equipment.
Safety Regulations
Ultra Elite-H Mask – Negative Pressure
GB
6
2Description
1
2
3
4
5
The face blank is made of a special soft rubber compound and assures a snug, comfortable fit and a
tight seal. Some versions of the mask are also available in small size.
The inhalation air flows from the connector of the mask past the inhalation valve to the inside of the
lens [thus keeping the lens largely fog-free] and then through the check valves into the nose cup.
The exhalation air passes through the exhalation valve directly to the ambient air.
Description
Fig. 1 Overview mask/helmet combination
1Helmet Visor4Adapter with tightening strap
2Helmet5Connector with inhalation and exhalation valves
3Lens
Ultra Elite-H Mask – Negative Pressure
GB
7
2.1Mask Versions
1
2
3
4
5
6
Ultra Elite-H•Standard version
Ultra Elite-H small•as standard version, small size
2.2Marking/Certification
Marking
The mask is marked on the outer face blank as shown in Fig. 2:
The Declaration of Conformity can be found under the following link: https://MSAsafety.com/DoC
Ultra Elite-H Mask – Negative Pressure
9
3Use
The mask is either carried using the harness in front of the chest or in the mask container.
WARNING!
Ensure that the top of the mask seal only lies on the user's forehead. Hair should not be between the
mask's seal and the user’s skin.
The mask could otherwise leak. This danger also exists for instance for mask wearers with beards or
deep scars in the sealing area.
WARNING!
The tightness of the mask/helmet combination depends on the correct size of the helmet, the adjustment of the adapter with tightening strap [if possible] and the helmet´s inner shell.
It is necessary to adjust the helmet’s inner shell accurately to your head size.
If the helmet or the inner shell do not fit your head properly, or if the wrong mask size is chosen, there
is a risk of leakage.
3.1Putting on the Helmet
(1)Select a helmet suitable for the user’s head size.
(2)Adjust the helmet according to the helmet’s operating
manual.
(3)Grip the helmet with both hands with the chinstrap open.
(4)Pull the helmet from the forehead over the head.
(5)Close and slightly tighten the chin strap.
Use
Ultra Elite-H Mask – Negative Pressure
GB
10
3.2Donning the Mask
WARNING!
In order to guarantee a proper fit for those wearing glasses, the Ultra Elite-H Mask spectacle kit must
be worn since ordinary glasses cannot be worn under the mask.
When fitting the mask in position the chin strap of the helmet must be loose. Remove
chin piece if necessary.
(1)Push helmet backwards onto neck so that the forehead
area is not covered.
(2)Open buckles and pull tightening straps on the belt
tongues until the grip tabs of the tightening straps reach
the stop [longest setting].
(3)Attach the helmet adapter to the slots on both sides of
the helmet.
(4)Draw pulling tabs of tightening straps on both sides of
the helmet until the mask fits onto the face.
(5)Push the helmet forward again until fitting comfortably.
(6)Tighten chin straps and fix their end with Velcro fastener.
(7)Ensure that the fit of the respiratory mask is checked by
a second person.
Use
WARNING!
The chin strap must fit tightly under the chin.
If necessary, remove chin protector.
Ultra Elite-H Mask – Negative Pressure
GB
11
3.3Leak Check
In order to check the mask-to-face tightness a leak test must be performed before each use.
(1)Seal the mask connector or the inlet opening side of an
attached filter with the ball of your hand.
(2)Test tightness by inhaling and exhaling. Whilst doing so:
(3)If necessary retighten the straps, or
when inhaling there must be negative pressure, no
inflowing air should be noticeable.
try a mask of a different size.
If the leak test fails the mask must not be used.
3.4Removing the Mask
(1)Open buckles on the adapters.
(2)Pull mask until stopped by the straps.
(3)Lift both hooks of the helmet adapter to the side and pull
them back out of the helmet slots.
(4)Remove the mask to the front.
Use
Do not grip the mask by the exhalation valve!
(5)Remove the helmet.
Ultra Elite-H Mask – Negative Pressure
GB
12
Cleaning, Disinfection
4Cleaning, Disinfection
The cleaning and disinfection of the masks is performed in accordance with the cleaning intervals
chapter 5.2.
There are three possible kinds of cleaning:
•Using a washing machine
•disk washing type (Meiko)
•tumbling type (Miele)
•Using an ultrasonic cleaner
•Cleaning by hand.
WARNING!
Do not use cleaning products containing hydrocarbons or solvents [e.g. nitro-thinner].
Cleaned parts must not be dried in radiant heat [sun, radiators]. When using a drying cabinet, the
temperature must not exceed +60°C.
Perform a tightness test after every cleaning, disinfection and maintenance or after every exchange of
parts.
4.1Cleaning/Disinfection of the Helmet
(1)Clean and disinfect the helmet according the helmet operating manual.
4.2Cleaning/Disinfection with a Washing Machine
Follow the washing agent’s user instructions on this CD/DVD.
(1)After cleaning check the two screws for the lens ring. Retighten the screws hand-tight if neces-
sary.
4.3Cleaning/Disinfection with an Ultrasonic Cleaner
When cleaning/disinfecting with an ultrasonic cleaner please follow the cleaning agent’s
user instructions on this CD/DVD.
Ultra Elite-H Mask – Negative Pressure
GB
13
Cleaning, Disinfection
4.4Cleaning by Hand
(1)Remove inhalation and exhalation valve discs.
(2)Unbutton the nose cup.
(3)Unscrew the speaking diaphragm with the special tool.
(4)Clean mask using lukewarm water and mild cleaning agent EW 80 and thoroughly rinse in
running water.
(5)Disinfect the mask after cleaning. Removed components must be separately cleaned and disin-
fected.
(6)Dry mask and components and reassemble mask in reverse order.
(7)Perform a tightness test [ chapter 5.5].
4.5Disinfection by Hand
Clean the mask before disinfecting.
Use a disinfectant according to the user instructions on this CD/DVD.
(1)Disassemble and clean the mask (see chapter 4.4).
(2)Disinfect the mask using recommended disinfectant. Please observe the user instruction for the
disinfectant.
(3)Thoroughly rinse mask in running water.
(4)Dry mask and components and reassemble mask in reverse order.
(5)Perform a tightness test (see chapter 5.5).
Ultra Elite-H Mask – Negative Pressure
GB
14
5Maintenance
5.1Maintenance Instructions
This device should be regularly checked and serviced by trained specialists. Inspection and service
records must be maintained. Always use original parts from MSA.
Repairs and maintenance must be carried out only by authorised service centres or by MSA. Changes
to devices or components are not permitted and will result in loss of approval.
MSA is liable only for maintenance and repairs carried out by MSA.
MSA recommends the following maintenance intervals. If necessary considering the
usage, tasks may be at even shorter intervals than indicated.
Observe national laws and regulations!
If in any doubt, ask your local MSA contact person.
5.2Maintenance Intervals
Maximal Intervals
DevicesWork to be Performed
helmetSee helmet operating manual.
masks
mask/
helmet
combination
*)
For a 2-year interval cleaned and disinfected masks have to be stored airtight. Otherwise masks
should be cleaned and disinfected at least semi-annually. After each cleaning and disinfection the
facepiece must be checked.
**)
For airtightly packed facepieces, which are not exposed to increased climatic and mechanical stress
[for example transport on vehicles], this interval can be extended to 2 years.
Cleaning and Disinfec-
*)
tion
Visual, Functional and
Tightness Check
Replacement of the exhalation valve disc
Replacement of the
Speech Diaphragm
User checkX
**)
Before
use
After
6-monthly
Use
XX
XX
Two
years
*)
**)
Four
years
X
Maintenance
Six
years
X
GB
Ultra Elite-H Mask – Negative Pressure
15
5.3Maintenance of the Exhalation Valve
The year of manufacture is located on the valve disc.
In case of a leak remove the exhalation valve disc and replace it with a new one as follows:
(6)Open the pivot cap of the connector.
(7)Pull the old exhalation valve disc off the connector.
(8)Replace the defective components.
(9)Close the pivot cap of the connector.
(10) Perform a tightness test [ chapter 5.5].
5.4Replacing the Speech Diaphragm
The year of manufacture is located on the speech diaphragm.
(1)Unbutton the nose cup.
(2)Unscrew the threaded socket from the mask inside with the special tool.
(3)Remove the speech diaphragm.
(4)Assemble the new speech diaphragm with the gasket towards the connector.
(5)Perform a tightness test [ chapter 5.5].
5.5Tightness Test of the Mask
The testing of the masks for tightness is performed using an applicable MSA test device
in accordance with the relevant operating manual.
Maintenance
(1)Fit mask tight onto the test device.
Use the test plate or test band to fix the mask.
(2)Generate a negative pressure of 10 mbar.
(3)Measure the pressure change after 1 min.
The mask including the exhalation valve meets the requirements if for a moistened exhalation valve and a negative pressure generated inside the mask the pressure change does not
exceed 1 mbar in a minute.
Leaking masks must not be used.
Ultra Elite-H Mask – Negative Pressure
GB
16
Safekeeping and Storage
5.6Visual Test and Function Test
Visual Test
(1)Inspect the mask for possible damages like for example deformations, stickings or cracks. Valve
discs, especially exhalation valve discs, are crucial functional elements of the mask.
(2)Defective or damaged parts have to be replaced immediately.
(3)For helmets: refer to the helmet operating manual.
Functional Test
After assembling the mask the mobile parts, especially the valve discs, have to be tested for unrestricted mobility.
5.7Helmet Maintenance
For detailed information about helmet maintenance refer to the helmet operating
manual.
6Safekeeping and Storage
WARNING!
In order to avoid damage to or the deformation of the masks keep no additional loose objects in the
mask container.
For the safekeeping of the mask the mask container should be used. This also applies to masks with
installed filters, except CO and NO filters.
MSA rubber products are protected by an anti-aging agent that can become visible as a light coating.
This coating is harmless and can be removed during cleaning.
To ensure a long life for rubber components, keep them in a cool, dry place that is protected from ultraviolet radiation, according to ISO 2230:2002, Rubber Products – Guidelines for Storage.
Ultra Elite-H Mask – Negative Pressure
GB
17
7Ordering Information
1
2
3
16151413
12
4
7
8
9
11
10
5
6
7.1Exploded View
Ordering Information
Fig. 3 Mask exploded view
1Nose cup valve seat9Cover Elite, assembly
2Nose cup valve disc10Inhalation valve seat Elite
3Nose cup [without valves]11Inhalation valve disc
4Carrying strap12Lens ring
5Speech diaphragm ring13Adapter with tightening strap
6Speech diaphragm14Screw for lens ring
7Exhalation valve disc Elite15Lens
8Clamp Elite16Face blank
Ultra Elite-H Mask – Negative Pressure
GB
18
7.2Mask
DescriptionArticle No.
Ultra Elite-H10045891
Ultra Elite-H – small10045892
7.3Accessories
DescriptionArticle No.
Elite spectacles [plastic frame]D2056730
Elite spectacles [metal frame]D2056733
Special tool [for replacing exhalation valve and speech diaphragm]D2055038
Exhalation valve closureD2056703
Mask hanger [package of 4]D2055753
Mask tightness test kit [for more MSA testing instruments refer to
www.MSAsafety.com]
Advantage container10026179
Test plateD6125721
Test bandD6125715
7.4Replacement Parts
DescriptionArticle No.
Service kit Ultra Elite [exhalation valve disc, inhalation valve disc, 2 nose cup
valve discs, inhalation valve seat]
Nose cup Elite, assemblyD2056915
Lens [with silicate coating]D2056802
Lens ring, assemblyD2056705
ClampD2056717
Speech diaphragm [package of 2]D2055708
Nose cup valve seat [package of 2D2056708
Nose cup valve disc [package of 20]D2055731
Threaded ring [speech diaphragm]D2056709
Inhalation valve disc [package of 20]D2056714
Inhalation valve seat [package of 5]D2056715
Exhalation valve disc [package of 10]D2056716
Carrying strap, assemblyD2055707-SP
Connector, assemblyD2056710
Connector EliteD2056713
Ordering Information
D6063705
D2056704
GB
Ultra Elite-H Mask – Negative Pressure
19
8Applicable Helmets
8.1MSA Gallet F1 Helmets
Helmets
F1 SF* [EN 443:2008]
F1 E* [EN 443:2008]
F1 E [EN 443:2008]
F1 SF [EN 443:2008]
* The mask/helmet combination with neck curtain for which all helmet components are marked with * comply with directive: vfdb-Richtlinie 0802:04
Helmets
Gallet F1 XF [EN 443:2008]
size M, L
Neck curtain:
aluminium
wool (integral)
nomex
Lamp:
Lamp holder (mounted, spare)
Integrated light module
Retro-reflective stickers:
Applicable Helmets
Colour of the shell
red*
yellow*
blue*
black*
white*
orange*
green luminescing*
nickel
Colour of the shell
all colours
GB
Grey
Ultra Elite-H Mask – Negative Pressure
20
8.2MSA Gallet F2 X-TREM Helmets
Helmets
INDUSTRY [EN397]
-unvented shell
-no chinstrap
GA3200-BA00
Optional:
2-point chinstrap
WILDLAND RESCUE [EN12492]
-Vented shell
-3-point chinstrap
GA33xx
Optional:
Retro-reflective stickers (GA3230):
TRAFFIC RESCUE [directive 89/686/EC]
-unvented shell
-3 point chinstrap
GA35xx
Optional:
Wool neck curtain (GA3240)
Applicable Helmets
Colour of the shell
White
White
Blue
Yellow
Black
Grey
Orange
Red
Green
Photoluminescent
silver, yellow, red, blue or green
GB
White
Blue
Yellow
Black
Orange
Red
Photoluminescent
Retro-reflective stickers (GA3230):
Ultra Elite-H Mask – Negative Pressure
silver, yellow, red, blue or green
21
For local MSA contacts, please visit us at MSAsafety.com
Because every life has a purpose...
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.