MSA Ultima X, Ultima XE, Ultima XIR, Ultima XA Maintance Manual

Page 1
Installation & Maintenance Instructions
Ultima® X Series Gas Monitors
Registered Address
ABLE Instruments & Controls Ltd Cutbush Park, Danehill, Lower Earley, Reading, Berk shire, RG6 4UT. UK.
Reading
Tel: +44 (0)118 9311188 | Email: info@able.co.uk
Aberdeen
Tel: +44 (0)1224 725999 | Email: ab@able.co.uk
Web
able.co.uk
E-commerce
247abl e.com
Page 2
Ultima®X Series Gas Monitors
Instruction Manual
In North America, to contact your nearest stocking location, dial toll-free 1-800-MSA-INST. To contact MSA International, dial (724) 776-8626. Inquiries can also be e-mailed to customer.service@msaSafety.com.
© MSA 2014 - All Rights Reserved
This manual is available on the internet at www.msaSafety.com
Manufactured by
MSA NORTH AMERICA
1000 Cranberry Woods Drive, Cranberry Township, PA 16066
(L) -Y Rev 9 10036101
Page 3
THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Like any piece of complex equipment, this instrument will perform as designed only if it is used and serv­iced in accordance with the manufacturer’s instructions. OTHERWISE, IT COULD FAIL TO PERFORM AS DESIGNED AND PERSONS WHO RELY ON THIS PRODUCT FOR THEIR SAFETY COULD SUSTAIN SEVERE PERSONAL INJURY OR LOSS OF LIFE.
The warranties made by Mine Safety Appliances Company with respect to the product are voided if the product is not used and serviced in accordance with the instructions in this manual. Please protect yourself and others by following them. We encourage our cus­tomers to write or call regarding this equipment prior to use or for any additional infor­mation relative to use or service.
MSA Instrument Warranty
1. Warranty- Seller warrants that this product will be free from
mechanical defect or faulty workmanship for the following periods:
Gas Monitor: eighteen (18) months from date of shipment or one (1) year from installation, whichever occurs first
Oxygen, Toxic or Catalytic Combustible Sensor: eighteen (18) months from date of shipment or one (1) year from installation, whichever occurs first
IR Sensor source: ten (10) years from date of shipment. All other IR components: two (2) years from date of shipment.
This warranty is applicable provided the product is maintained and used in accordance with Seller's instructions and/or recommendations. This warranty does not apply to expendable or consumable parts whose normal life expectancy is less than one (1) year. The Seller shall be released from all obligations under this warranty in the event repairs or modifications are made by persons other than its own or authorized service personnel or if the warranty claim results from physical abuse or misuse of the product. No agent, employee or representative of the Seller has any authority to bind the Seller to any affirmation, representation or warranty concerning the goods sold under this contract. Seller makes no warranty concerning components or accessories not manufactured by the Seller, but will pass on to the Purchaser all warranties of manufacturers of such components. THIS
WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR STATUTORY, AND IS STRICTLY LIMITED TO THE TERMS HEREOF. SELLER SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE.
"! WARNING
Page 4
NOTE: This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide rea­sonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequen­cy energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interfer­ence at his own expense.
This is a class A product in accordance with CISPR 22. In a domestic environment, this product may cause radio inter­ference, in which case the user may be required to take ade­quate measures.
2. Exclusive Remedy- It is expressly agreed that Purchaser's sole
and exclusive remedy for breach of the above warranty, for any tortious conduct of Seller, or for any other cause of action, shall be the repair and/or replacement at Seller's option, of any equipment or parts thereof, which after examination by Seller is proven to be defective. Replacement equipment and/or parts will be provided at no cost to Purchaser, F.O.B. Seller's Plant. Failure of Seller to successfully repair any nonconforming product shall not cause the remedy established hereby to fail of its essential purpose.
3. Exclusion of Consequential Damage- Purchaser specifically
understands and agrees that under no circumstances will seller be liable to purchaser for economic, special, incidental or consequential damages or losses of any kind whatsoever, including but not limited to, loss of anticipated profits and any other loss caused by reason of nonoperation of the goods. This exclusion is applicable to claims for breach of warranty, tortious conduct or any other cause of action against seller.
"
WARNING
i
Page 5
General Warnings and Cautions
1. The Ultima X Series Gas Monitors described in this manual must be installed, operated and maintained in strict accordance with their labels, cautions, warnings, instructions, and within limitations stated. Verify that the class, group, and temperature ratings of the equipment agree with the actual classification of the location.
2. The Ultima X Series Gas Monitor is designed to detect gases or vapors in air. It cannot measure the concentration of gases or vapors in steam or inert or oxygen-deficient atmospheres. The oxygen sensor can measure oxygen-deficient atmospheres.
3. Electrochemical sensors are sealed units which contain a corrosive electrolyte. Should a sensor develop leakage, it must be immediately removed from service; then, remove it from the sensing head and discard it properly. Caution must be exercised so that the electrolyte does not contact skin, eyes, clothing or circuitry; otherwise, serious personal injury (burns) and/or equipment damage may result.
4. Use only genuine MSA replacement parts when performing any maintenance procedures provided in this manual. Failure to do so may seriously impair instrument performance. Repair or alteration of the Ultima X Series Gas Monitor, beyond the scope of these maintenance instructions or by anyone other than an authorized MSA service personnel, could cause the product to fail to perform as designed and persons who rely on this product for their safety could sustain serious personal injury or loss of life.
5. Do not locate the general-purpose enclosure models in an area which may contain a flammable mixture of gas and air; otherwise, an explosion may occur. The general-purpose Ultima X Series Gas Monitors can be a source of ignition and must not be mounted in an area where a flammable mixture of combustible gas and air may become present; otherwise, an explosion may occur. If such a location must be monitored, use an explosion-proof Ultima X Series Gas Monitor model.
6. The Ultima XIR Infrared combustible gas monitor detects the presence of most combustible gases by identifying the difference in the amount of infrared light energy absorbed during the presence of these gases. This monitor, however, does NOT detect the presence of hydrogen gas and must never be used to monitor for hydrogen gas.
7. The standard Ultima XIR Infrared Combustible Gas Monitor does
"
WARNING
ii
Page 6
not detect the presence of acetylene gas and the presence of acetylene gas will degrade sensor performance. Custom-built acetylene sensors are available through your MSA representative.
8. Gas detectors depend on an unimpeded gas flow for proper operation. In environments where contamination is possible, ensure that the flow remains unobstructed at the sensor. Failure to follow this may prevent gas detection and generate inaccurate readings.
9. CSA performance Certification to standard C22.2 No. 152 is valid only when the instrument is calibrated on methane per the instruction manual.
10. Install Product in accordance with all markings and the regulations of the country in use.
11. Product components may have different hazardous location ratings. Ensure all components are suitable for the area of installation and protection technique.
Failure to follow the above can result in serious personal injury or loss of life.
1. As with all gas monitors of these types, high levels of, or long exposure to, certain compounds in the tested atmosphere could contaminate the sensors. In atmospheres where an Ultima X Series Gas Monitor may be exposed to such materials, calibration must be performed frequently to ensure that operation is dependable and display indications are accurate.
2. The Ultima X Series Gas Monitor must not be painted. If painting is done in an area where a Monitor is located, care must be exercised to ensure that paint is not deposited on the sintered, metal flashback arrestor in the inlet fitting of the Ultima X Series Gas Monitor, if so equipped. Such paint deposits would interfere with the diffusion process, whereby a sample of the atmosphere being monitored diffuses into the Monitor.
3. The only absolute method to ensure proper overall operation of an Ultima X Series Monitor is to check it with a known concentration of the gas for which it has been calibrated. Consequently, calibration checks must be included as part of the routine inspection of the system.
4. Protect the Ultima X Series Gas Monitor from extreme vibration. Do not mount the sensing head in direct sunlight as this may cause overheating of the sensor.
Failure to follow the above can result in injury, product damage and/or an unsafe condition.
"
CAUTION
iii
Page 7
Table of Contents
Chapter 1,
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Identifying Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Installing Your Gas Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Installing the Ultima XA Gas Monitor . . . . . . . . . . . . . .1-6
Installing the Ultima XE Gas Monitor . . . . . . . . . . . . . .1-6
Installing the Ultima XIR Gas Monitor . . . . . . . . . . . . .1-8
Electrical Connections for Ultima X Gas Monitors . . . . . . . . .1-9
Wiring for all Models . . . . . . . . . . . . . . . . . . . . . . . . .1-10
Use of External Controllers . . . . . . . . . . . . . . . . . . . . . . . . . .1-12
Identify PCB Configuration . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
Installing the Ultima X Remote Sensor Module . . . . . . . . . .1-20
Electrical Connections for Remote Sensors . . . . . . . . . . . . .1-21
Chapter 2,
Start-up and Calibration . . . . . . . . . . . . . . . . . . .2-1
Initial Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Calibration Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Ultima Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Ultima Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Calibration Output Signal . . . . . . . . . . . . . . . . . . . . . . .2-6
Calibration Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Ultima X Series Gas Monitor Calibration Procedure . . . . . . . .2-7
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Span Gas Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
INITIAL Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Standard Calibration . . . . . . . . . . . . . . . . . . . . . . . . .2-13
OXYGEN Calibration . . . . . . . . . . . . . . . . . . . . . . . . .2-18
XIR Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-18
iv
Page 8
Chapter 3,
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Chapter 4, Maintenance . . . . . . . . . . . . . . . . . . . .4-1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Ultima XIR Cleaning Procedure . . . . . . . . . . . . . . . . . . . . . . .4-4
Replacing an Ultima XE or Ultima XA Sensor . . . . . . . . . . . .4-5
Obtaining Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Appendix A,
Optional Features . . . . . . . . . . . . . . . . . . . . . . . .A-1
1) Internal Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Unpacking, Mounting and Wiring . . . . . . . . . . . . . . . .A-1
Ultima X Series Gas Monitor Internal Relays . . . . . . .A-2
Relay Specifications . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Alarm Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Fault Relay or Trouble . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Relay Connections . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
2) Optional RESET Push-button . . . . . . . . . . . . . . . . . . . . . . .A-7
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-7
RESET Button Selection . . . . . . . . . . . . . . . . . . . . . . .A-7
Optional Push-button Calibration . . . . . . . . . . . . . . . .A-8
3) Optional Horn Relay Software . . . . . . . . . . . . . . . . . . . . . .A-9
To Activate the Horn Relay . . . . . . . . . . . . . . . . . . . . .A-9
To Reset the Horn Relay . . . . . . . . . . . . . . . . . . . . . . .A-9
Appendix B, Calibration Guide for Additional XIR/XI Gases .B-1
Appendix C,
General Certification Information . . . . . . . . . . .C-1
v
Page 9
Appendix D,
HART Specific Information . . . . . . . . . . . . . . . . .D-1
HART Field Device Specification . . . . . . . . . . . . . . . . . . . . . .D-1
Host Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-2
Status Information . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-3
Extended Device Status . . . . . . . . . . . . . . . . . . . . . . .D-3
Universal Commands . . . . . . . . . . . . . . . . . . . . . . . . .D-5
Common-Practice Commands . . . . . . . . . . . . . . . . . .D-5
Burst Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-6
Catch Device Variable . . . . . . . . . . . . . . . . . . . . . . . . .D-6
Command #129: Read Sensor Gas Type . . . . . . . . . .D-7
Command #130: Read Device Real Time Clock . . . .D-7
Command #131: Read Alarm Setpoints . . . . . . . . . . .D-8
Command #132: Read Alarm Control Actions . . . . . .D-9
Command #133: Read Min, Max, Avg Values . . . . . .D-9
Command #134: Read Last Cal Date . . . . . . . . . . . .D-10
Command #135: Read Gas Table . . . . . . . . . . . . . . .D-10
Command #136: Read Input Voltage Value . . . . . . .D-10
Command #137: Read Auto Zero Comp Value . . . .D-11
Command #139: Read Sensor Status message . . . .D-11
Command #140: Read Swap Delay Status . . . . . . .D-11
Command #141: Read Cal Signal Status . . . . . . . . .D-12
Command #142: Read Alert Option Status . . . . . . . .D-12
Command #143: Read Sensor Temperature . . . . . .D-13
Command #144: Read Relay Normal State . . . . . . .D-13
Command #173: Write RTC . . . . . . . . . . . . . . . . . . .D-14
Command #174: Write Alarm Setpoints . . . . . . . . . .D-15
Command #175: Write Alarm Setpoint
Control Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-16
Command #176: Write Average Interval . . . . . . . . . .D-17
Command #177: Write Upper Trim Point . . . . . . . . .D-18
Command #178: Write Gas Table . . . . . . . . . . . . . . .D-19
Command #179: Write Sensor Data Sheet
Reset Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-20
Command #180: Write Sensor Swap Delay Enable .D-21
Command #181: Write Cal Signal Enable . . . . . . . .D-22
Command #182: Write Calibration Mode . . . . . . . . .D-23
Command #183: Write Calibration Abort . . . . . . . . .D-24
Command #184: Write Calibration Step . . . . . . . . . .D-25
vi
Page 10
Command #185: Write Alarm Acknowledge . . . . . . .D-26
Command #186: Write Protect Mode . . . . . . . . . . . .D-27
Command #187: Write Alert Option . . . . . . . . . . . . .D-28
Command #188: Write Relay Normal State . . . . . . .D-29
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-32
Power-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-32
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-32
Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-32
Busy and Delayed-Response . . . . . . . . . . . . . . . . . .D-33
Long Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-33
Non-Volatile Memory . . . . . . . . . . . . . . . . . . . . . . . . .D-33
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-34
Write Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-34
Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-34
Capability Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-34
Default Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-35
Calibration Using a HART® Communicator . . . . . . . . . . . . .D-35
Sensor Zero Selection Menu . . . . . . . . . . . . . . . . . .D-35
Standard Zero/Span Calibration Selection Menu . . .D-37
Initial Calibration Procedures . . . . . . . . . . . . . . . . . .D-40
User Calibration Selection Menu . . . . . . . . . . . . . . .D-40
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-54
Span Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-54
Zero Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-57
vii
Page 11
List of Figures
Figure 1-1. General-Purpose Ultima XA Monitor . . . . . . . . . .1-1
Figure 1-2. -Proof Ultima XE Monitor . . . . . . . . . . . . . . . . . . .1-2
Figure 1-3. Explosion-Proof Ultima XIR Monitor . . . . . . . . . . .1-2
Figure 1-4. General-Purpose XA Remote Sensor Module . . .1-3
Figure 1-5. Explosion-Proof XE Remote Sensor Module . . . .1-3
Figure 1-6. Explosion-Proof XIR Remote Sensor Module . . .1-4
Figure 1-7. Ultima XE and XIR Mounting Bracket . . . . . . . . .1-7
Figure 1-8. Ultima XIR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Figure 1-9. Ultima XE Grounding Terminals . . . . . . . . . . . . . .1-9
Figure 1-10. General-Purpose Two-Wire Operation . . . . . . .1-14
Figure 1-11. Explosion-Proof Two-Wire Operation . . . . . . . .1-14
Figure 1-12. General-Purpose Three-Wire Operation . . . . . .1-15
Figure 1-13. Explosion-Proof Three-Wire Operation . . . . . . .1-15
Figure 1-14. Two-Wire Printed Circuit Board
(no HART Protocol) . . . . . . . . . . . . . . . . . . . . . .1-16
Figure 1-15. Two-Wire Printed Circuit Board
(with HART Protocol) . . . . . . . . . . . . . . . . . . . . .1-17
Figure 1-16. Three-Wire Printed Circuit Board
(no HART Protocol) . . . . . . . . . . . . . . . . . . . . . .1-18
Figure 1-17. Three-Wire Printed Circuit Board
(with HART Protocol) . . . . . . . . . . . . . . . . . . . . .1-19
Figure 1-18. Remote Module General-Purpose
Ultima X Series Wiring . . . . . . . . . . . . . . . . . . . .1-20
Figure 1-19. Remote Module Explosion-Proof
Ultima X Series Wiring . . . . . . . . . . . . . . . . . . . .1-20
Figure 2-1. LCD Gas Concentration Display . . . . . . . . . . . . . .2-1
Figure 2-2. Ultima Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Figure 2-3. Ultima Controller . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Figure 2-4. Ultima X Optional Push-button Calibrator . . . . . . .2-5
Figure 2-5. Apply ZERO Gas Flag . . . . . . . . . . . . . . . . . . . .2-14
Figure 2-6. Apply SPAN Gas Flag . . . . . . . . . . . . . . . . . . . . .2-15
Figure 2-7. Calibration End Display . . . . . . . . . . . . . . . . . . .2-17
Figure 4-1. "Change Sensor" Scrolls Across the Display . . . .4-6
Figure 4-2. Sensor Assembly and Sensor Guard
for General-Purpose Model . . . . . . . . . . . . . . . . .4-7
Figure A-1. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Figure A-2. Relay Printed Circuit Board . . . . . . . . . . . . . . . . .A-6
Figure D-1. Zero cal step screen . . . . . . . . . . . . . . . . . . . . .D-41
viii
Page 12
Figure D-2. Span cal step screen . . . . . . . . . . . . . . . . . . . . .D-41
Figure D-3. Select Sensor Calibration
from the Sensor Trim Menu . . . . . . . . . . . . . . . .D-42
Figure D-4. First Warning screen . . . . . . . . . . . . . . . . . . . . .D-43
Figure D-5. Second Warning screen . . . . . . . . . . . . . . . . . .D-44
Figure D-6. Standard Calibration function select screen . . .D-45
Figure D-7. Calibration initiated screen . . . . . . . . . . . . . . . .D-46
Figure D-8. Selection Confirmation screen . . . . . . . . . . . . . .D-47
Figure D-9. Sensor Zero Countdown screen . . . . . . . . . . . .D-48
Figure D-10. Zero Adjustment screen . . . . . . . . . . . . . . . . . .D-49
Figure D-11. Span Countdown screen . . . . . . . . . . . . . . . . .D-50
Figure D-12. Adjusting Span screen . . . . . . . . . . . . . . . . . . .D-51
Figure D-13. Calibration Completion message . . . . . . . . . . .D-52
Figure D-14. Calibration Gas Reminder screen . . . . . . . . . .D-53
Figure D-15. Loop Control Reminder message . . . . . . . . . .D-54
Figure D-16. Calibration Status screen . . . . . . . . . . . . . . . . .D-55
Figure D-17. Sensor Trim Point screen . . . . . . . . . . . . . . . .D-56
Figure D-18. Additional Sensor Status screen . . . . . . . . . . .D-57
Figure D-19. Device Status screen . . . . . . . . . . . . . . . . . . . .D-58
ix
Page 13
List of Tables
Table 1-1. Installation Outline Drawing List . . . . . . . . . . . . . .1-10
Table 1-2. Cable Length and Wire Size for Units without
Internal Relays . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
Table 1-3. Installation Outline Drawings
for Ultima X Power Supplies . . . . . . . . . . . . . . .1-12
Table 1-4. Remote Module Wiring and Placement . . . . . . . .1-22
Table 1-5. Remote Sensor Wiring Cable . . . . . . . . . . . . . . . .1-22
Table 1-6. Low Temperature Wiring Cable . . . . . . . . . . . . . .1-22
Table 2-1. Instrument Operation . . . . . . . . . . . . . . . . . . . . . . .2-2
Table 2-2. Factory-set Span Values . . . . . . . . . . . . . . . . . . . .2-9
Table 2-3. Calibration Guide for Combustible Gas Sensor . .2-11
Table 3-1. Performance Specifications . . . . . . . . . . . . . . . . . .3-1
Table 3-2. Sensor Response to Interferants . . . . . . . . . . . . . .3-4
Table 4-1. Operational Display Messages . . . . . . . . . . . . . . . .4-1
Table 4-2. Configuration Display Messages . . . . . . . . . . . . . .4-2
Table 4-3. Troubleshooting Guidelines . . . . . . . . . . . . . . . . . .4-2
Table 4-4. Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Table A-1. Power Cable Distances for the Ultima X
Series Gas Monitor with Internal Relays . . . . . .A-2
Table A-2. Relay Specifications . . . . . . . . . . . . . . . . . . . . . . . .A-2
Table A-3. Push-button Calibration . . . . . . . . . . . . . . . . . . . . .A-8
Table D-1. Device Identification . . . . . . . . . . . . . . . . . . . . . . .D-1
Table D-2. Current Values . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-2
Table D-3. Device Variables Exposed by the Ultima Monitor .D-3 Table D-4. Dynamic Variable implemented by Ultima Monitor D-3
Table D-5. Additional Device Status (Command #48) . . . . . .D-4
Table D-6. Supported Commands . . . . . . . . . . . . . . . . . . . . .D-6
Table D-7. Device-Specific Commands . . . . . . . . . . . . . . . . .D-7
Table D-8. Gas Type Descriptions . . . . . . . . . . . . . . . . . . . .D-30
Table D-9. Alarm Control Actions . . . . . . . . . . . . . . . . . . . . .D-30
Table D-10. Gas Table Values . . . . . . . . . . . . . . . . . . . . . . .D-30
Table D-11. Calibration Modes . . . . . . . . . . . . . . . . . . . . . . .D-31
Table D-12. Sensor Status Codes . . . . . . . . . . . . . . . . . . . .D-31
Table D-13. Sampling Rates . . . . . . . . . . . . . . . . . . . . . . . . .D-32
Table D-14. Command Response Times . . . . . . . . . . . . . . .D-33
Table D-15. Capability Checklist . . . . . . . . . . . . . . . . . . . . . .D-34
Table D-16. Default Configuration . . . . . . . . . . . . . . . . . . . .D-35
x
Page 14
1-1
Chapter 1, Installation
General Description
The Ultima X Series Gas Monitor is designed to sample the environment where mounted and alert you to potentially dangerous levels of your target gas, depending on your particular model. The Ultima X Series device uses various detection methods, depending on the gas of interest. Detection methods can be electrochemical, infrared, pellement or other technologies. The Ultima XE Gas Monitor is an explosion-proof device suitable for installation in hazardous locations. The Ultima XA Gas Monitor is a general-purpose version in a plastic enclosure for use in nonexplosive atmospheres only. The Ultima X Gas Monitor can be ordered with the standard 4 to 20 mA analog output or with an optional HART (Highway Addressable Remote Transducer) protocol, which is superimposed on the 4 to 20 mA signal. The unit is factory-calibrated and shipped ready for installation.
The main sensor input is provided via a five-terminal interface that provides a digital interface for 3 VDC or 5 VDC sensor modules. Many different sensor modules are available, providing sensing capability for a large variety of gases. The operating range varies with the type of cell (e.g., electrochemical, pellistor or infrared combustible, etc.).
Identifying Your Unit
The Ultima XA Gas Monitor is housed in a rugged, plastic general­purpose enclosure (FIGURE 1-1).
Figure 1-1. General-Purpose Ultima XA Monitor
Page 15
The Ultima XE Gas Monitor is housed in a 316 stainless steel explosion-proof enclosure (FIGURE 1-2 shows the Ultima XE with the optional explosion-proof HART port).
.
The Ultima XIR Gas Monitor is housed in a 316 stainless steel explosion-proof enclosure (FIGURE 1-3).
If your application requires the sensor head to be located separately from the control unit, all models are available with Remote Sensor Modules (shown in FIGURES 1-4, 1-5 and 1-6).
Figure 1-3. Explosion-Proof Ultima XIR Monitor
Figure 1-2. Explosion-Proof Ultima XE Monitor
1-2
Page 16
Figure 1-5. Explosion-Proof XE Remote Sensor Module
Figure 1-4. General-Purpose XA Remote Sensor Module
1-3
Page 17
To determine your sensor type and options, check the shipping carton. Checked items are included in the carton. The carton label identifies:
Type of unit supplied (Gas Monitor, Gas Monitor Less Sensor, or Remote Sensor Module)
Type of gas (combustible gas, toxic gas or oxygen)
Range [% LEL, PPM (parts per million), or %]
Output (2- or 3-wire, 4 to 20 mA, or 4 to 20 mA with HART)
Any options such as internal relays and/or LEDs.
If your unit contains internal relays, see Appendix A.
Also check the sensor ID label located on the inside of the sensor. This can be viewed by unscrewing the lower portion of the sensor. If performing this while the unit is powered, see the following Warning. The sensor ID label identifies the detectable gas and the gas range.
For Ultima XE sensors marked Class I, Groups A, B, C and D and not used in Class II areas, unscrew sensor cap at least three full turns (but no more than four full turns from its tightly-closed position), wait 10 seconds, and then remove cap completely. Failure to follow this warning can result in the ignition of a hazardous atmosphere.
For the Ultima XE Sensor marked Class II Groups F and G, atmosphere must be free of dust and the power removed from the unit before the sensor cap can be removed from the housing. Failure to follow this warning can result in the igni­tion of a hazardous atmosphere.
"
WARNING
Figure 1-6. Explosion-Proof XIR Remote Sensor Module
1-4
Page 18
Your Ultima XE Series Gas Monitor may also include a separate HART Module if you ordered the following optional accessories:
1). Internal Power Supply and explosion-proof HART Port
2). External RESET Push-button and explosion-proof HART Port.
Installing Your Gas Monitor
NOTE: Reference installation outline drawings listed in TABLE 1-1.
Generally, the Ultima X Series Gas Monitors or remote sensing module should be mounted close to the area where a leak is likely to occur or where the gas is expected. Install the Ultima X Series Gas Monitors or the remote sensing module at a high level (ceiling) or low level (floor), depending on the density of the gas most likely to be found. Install the unit so that the front display of the unit is not blocked or hidden from view.
Mount the Ultima XE or XA Gas Monitor or remote sensing module with the sensor inlet fitting (FIGURE 1-1, 1-2, 1-4 or 1-5) pointed downward; otherwise, the inlet may become clogged with particulate matter or liquids.
Mount the Ultima XIR Gas Monitor or XIR Remote Sensing Module with the sensor inlet fitting extended horizontally from the main enclosure (FIGURE 1-3 and 1-6) to prevent the build-up of particulate or liquid matter on the monitor's opti­cal surfaces.
Do not paint the Ultima X Series Gas Monitors. If painting is done in an area where a sensor is located, exercise CAU­TION to ensure paint is not deposited on the sensor inlet fit­ting. Such paint deposits would interfere with the diffusion process, whereby a sample of the monitored atmosphere diffuses into the sensor. In addition, solvents in the paint may cause an alarm condition to occur.
Protect the Ultima X Series Gas Monitors from extreme vibration. Do not mount sensing head in direct sunlight as this may cause overheating of the sensor.
"
CAUTION
1-5
Page 19
Do not locate the general-purpose enclosure models in an area which may contain a flammable mixture of gas and air; otherwise, an explosion may occur. The general-purpose Ultima X Series Gas Monitors can be a source of ignition and must not be mounted in an area where a flammable mixture of combustible gas and air may become present; otherwise, an explosion may occur. If such a location must be moni­tored, use an explosion-proof gas monitor.
Installing the Ultima XA Gas Monitor
Remove lid and drill enclosure for power, signal and optional relay cable entry. Use one of the following methods to mount the general-purpose Ultima XA Gas Monitor/Less Sensor or the Ultima XA Gas Monitor.
Using customer-installed wiring holes, install the Ultima XA Gas Monitor to the end of rigid conduit.
Use mounting holes in the corners of the Ultima XA enclosure to mount directly to a wall.
Use mounting holes in the corners of the Ultima XA enclosure to mount to the optional Mounting Kit (P/N 10047561); see FIGURE 1-7.
The Ultima XA gas sensor is not shipped attached to the main enclosure. Ensure the sensor wiring harness is through the entry and the sensor is pointed downward.
Installing the Ultima XE Gas Monitor
The optional Mounting Bracket Kit (P/N 10047561) can be attached to the rear holes of the Ultima XE Gas Monitor (FIGURE 1-7).
The Ultima XE Gas Monitor main enclosure can be rotated 360° and mounted to ensure easy access to any of the four entryways. The electronics assembly inside the metal enclosure can be repositioned in any of the four self-aligning interior holes to ensure the display is properly oriented.
The Ultima XE Gas Monitor sensor is not shipped attached to the main enclosure. Ensure the sensor wiring harness is through the entry and the sensor is pointing downward. Tighten with a strap wrench.
"
WARNING
1-6
Page 20
.
Figure 1-7. Ultima XE and XIR Mounting Bracket
1-7
Page 21
Installing the Ultima XIR Gas Monitor
The Ultima XIR Gas Monitor contains no user- or field-serv­iceable parts and must be returned to the factory for repair. Any attempt to open the monitor will damage the unit and void the warranty.
Under no circumstances should a wrench or pry-bar be applied to the two legs that support the unit's reflectors dur­ing installation or removal of the sensor (FIGURE 1-8). Applying force to the legs can permanently damage the monitor.
It is recommended that the monitor's environmental guard be installed on the unit at all times. If the monitor is to be operated without the guard, frequent checks should be made to ensure particulate or liquid matter has not collected on the windows.
The optional Mounting Bracket (P/N 10047561) can be attached to the rear holes of the Ultima XE Gas Monitor (FIGURE 1-7).
The Ultima XIR Gas Sensor is factory-installed on the stainless steel gas monitor. The Ultima XIR Monitor must be installed with the XIR sensor in a horizontal position (see FIGURE 1-3) to prevent the build-up of particulate or liquid matter on the monitor optical surface.
The Ultima XIR Gas Sensor is intended for use only on metal enclosures.
.
Figure 1-8. Ultima XIR
"
CAUTION
"
WARNING
1-8
Page 22
Electrical Connections for Ultima X Gas Monitors
Before wiring the Ultima X Series Gas Monitors, disconnect power source supplying the monitor; otherwise, electrical shock or ignition of hazardous atmospheres could occur.
For Ultima XE and XIR installations, the internal grounding terminal (located on the interior bottom of the Ultima XE main enclosure) must be used for equipment grounding. The external grounding terminal is only to be used as a supple­mental bonding connection where local authorities permit or require such a connection. See FIGURE 1-9 for location of grounding terminals.
.
NOTE: For Ultima X Series units with internal relays, see Appendix A.
This assembly is marked to identify power, ground and signal connections.
A two-wire connection is possible for certain:
Toxic Gas models
Oxygen models
A three-wire connection is required for all:
Combustible Gas models
Toxic and Oxygen models with internal relays.
Figure 1-9. Ultima XE Grounding Terminals
"
WARNING
1-9
Page 23
Wiring for all Models
Install wiring in accordance with the electrical code of the country in use and UL 61010-A1 or CSA C22.2 No. 1010.1, as applicable. In these installations, twisted-pair, instrument quality cable is recommended. Shielded cable is recommended for cable runs where interferences from radio frequency interference (RFI), electromagnetic interference (EMI) or other noise sources exist (such as motors, welding equipment, heaters, etc.).
NOTE: See Installation Outline Drawings for wiring details as specified
in TABLE 1-1.
Table 1-1. Installation Outline Drawing List
MODEL TYPE DOCUMENT NO.
Ultima XA Gas Monitor SK3015-1027
Ultima XE Gas Monitor SK3015-1025
Ultima XIR Gas Monitor SK3015-1026
Conduit may also be needed in areas where large amounts of electrical noise is expected.
Use caution when selecting a cable size. TABLE 1-2 expresses the maximum cable length when only using the Ultima X Series Gas Monitors. Ultima X Series options may take additional power which requires a heavier cable or a short cable run. Cable distances for units with internal relays are specified in Appendix A, TABLE A-1.
When selecting cable size, consider future needs (i.e., addition of sensors and/or options available with the Ultima X Series Gas Monitors). See Chapter 3, TABLE 3-1, "Performance Specifications" for proper input voltage.
Ensure that water and dirt are not able to enter the unit via the wire or conduit. If the unit is installed in a location known to be wet or damp, it is good practice to loop or bend the entry into the unit that prevents water incursion.
All cable shields should be terminated to earth ground at one end only.
1-10
Page 24
Table 1-2. Cable Length and Wire Size for Units Without Internal Relays
GAS SENSOR DC WIRE MAXIMUM MAXIMUM TYPE OUTPUT VOLTAGE SIZE CABLE LENGTH LOAD
SUPPLY (AWG) WITHOUT HART RESISTANCE
(FEET) (METERS) (OHMS)
Oxygen or Toxic 2 Wire 12 VDC 22 4000 1219 100
900(w/HART) 274 (w/HART) 50 (w/HART)
24 VDC 22 7,000 2134 500
Oxygen or Toxic 3 Wire 24 VDC 22 10,000 3048 500
Combustible 3 Wire 12 VDC 18 900 274 250
16 1,400 427 250
12 3,600 1097 250
Combustible 3 Wire 24 VDC 18 2,500 762 500
16 4,200 1280 500
12 10,000 3048 500
XIR 3 wire 12 VDC 18 300 91 250
16 500 152 250
12 900 274 250
XIR 3 wire 24 VDC 18 2,000 610 500
16 3,500 1067 500
12 5,000 1524 500
An external power supply is required to supply 8-30 VDC to the Ultima X Series Gas Monitor (For power requirements, see Chapter 3, "Specifications"). All connections should be made by following appropriate wire code procedures.
For proper installation of an AC power supply used with an Ultima X Series transmitter, refer to the following drawings for detailed information. Optional 12 VDC or 24 VDC internal and external power supplies can be ordered with the Ultima X Series Gas Monitors.
1-11
Page 25
Table 1-3. Installation Outline Drawings for Ultima X Power Supplies
MODEL POWER POWER SUPPLY POWER INSTALLATION OUTLINE
SUPPLY OUTPUT VOLTAGE SPECIFICATION DRAWING NUMBER
XA External 12 VDC 1.25 Amps, 15 W 10000020129
External 24 VDC 0.46 Amps, 11 W 10000020127
Internal 12 or 24 VDC see above SK3015-1027
XE External 12 VDC 1.25 Amps, 15 W 10000020130
External 24 VDC 0.46 Amps, 11 W 10000020128
Internal 12 or 24 VDC see above SK3015-1025 (XE) or
SK3015-1026 (XIR)
Use of External Controllers
The Ultima X Series Gas Monitors may be connected to any device capable of accepting 4-20 mA analog signals, such as:
Suprema Controller
Model 9010/9020 Controller
GasGard family controllers
Quad Gas Controller
Programmable Controllers
DCS’s, etc.
When using any of the the Ultima X Series accessories (such as relays) with the 4 to 20 mA output Ultima X Series Gas Monitor, a three-wire connection must be used. Failure to use a three-wire connection could damage the electronics within the Ultima X Series Gas Monitor which can result in serious personal injury or loss of life.
Be sure to install your Ultima X Series Gas Monitor accord­ing to National Electrical and local procedural codes. Failure to do so can result in an unsafe condition.
"
WARNING
1-12
Page 26
Identify PCB Configuration
Identify the main pc board as a two-wire or a three-wire unit:
For XA Gas Monitors:
while looking at the main pc board, locate the identifying label on the underside of the lid:
A-ULTX-PCB-A-1 is a two-wire unit, 4-20 mA output
A-ULTX-PCB-A-2 is a two-wire unit with HART protocol on the 4-20 mA output
A-ULTX-PCB-A-3 is a three-wire unit, 4-20 mA output
A-ULTX-PCB-A-4 is a three-wire unit with HART protocol on the 4-20 mA output
For XE and XIR Gas Monitors:
locate the identifying label on the side of the plastic shroud for the main pc board:
A-ULTX-PCB-E-1 is a two-wire unit, 4-20 mA output
A-ULTX-PCB-E-2 is a two-wire unit with HART protocol on the 4-20 mA output
A-ULTX-PCB-E-3 is a three-wire unit, 4-20 mA output
A-ULTX-PCB-E-4 is a three-wire unit with HART protocol on the 4-20 mA output.
1-13
Page 27
Two-wire 4 to 20 mA Ultima X Series Monitors operate in the current loop mode (FIGURE 1-10 for general-purpose) (FIGURE1-11 for explosion-proof).
.
.
Figure 1-11. Explosion-Proof Two-Wire Operation
Figure 1-10. General-Purpose Two-Wire Operation
1-14
Page 28
Three-wire Ultima X Series Monitors operate in the current source mode (see FIGURE 1-12 for general-purpose) (FIGURE 1-13 for explosion-proof).
.
.
Figure 1-13. Explosion-Proof Three-Wire Operation
Figure 1-12. General-Purpose Three-Wire Operation
1-15
Page 29
Installation of Two-Wire, 4-20 mA Output (no HART Protocol)
1. Connect 8-30 VDC power lead to J8-1 (see FIGURE 1-14)
2. Connect J8-2 to the 4-20 mA output on the remote system.
3. Connect the sensor module to main pc board connector J-1.
4. Assemble lid on the enclosure.
Figure 1-14. Two-Wire Printed Circuit Board
(no HART Protocol)
1-16
Page 30
Installation of Two-Wire, 4-20 mA Output with HART Protocol
1. Connect 12-30 VDC power lead to J8-1 (see FIGURE 1-15).
NOTE: The HART signal is not available below 12 VDC
on the two-wire pc board.
2. Connect J8-2 to the 4-20 mA input on the remote system.
3. Terminate the 4-20 mA line with 230-500 Ohms of resistance.
4. Connect sensor module to main pc board connector J-1.
5. Assemble lid on the enclosure.
Figure 1-15. Two-Wire Printed Circuit Board
(with HART Protocol)
1-17
Page 31
Installation of Three-Wire, 4-20 mA Output (no HART Protocol)
1. Connect 8-30 VDC power lead to J8-1 (see FIGURE 1-16)
2. Connect J8-2 to the 4-20 mA output on the remote system.
3. For three-wire operation, connect the signal ground to J8-3.
4. Connect the sensor module to main pc board connector J-1.
5. Wire for optional relays, if applicable (see Appendix A).
6. Assemble lid on the enclosure.
Figure 1-16. Three-Wire Printed Circuit Board
(no HART Protocol)
1-18
Page 32
Installation of Three-Wire, 4-20 mA Output with HART Protocol
1. Connect 8-30 VDC power lead to J8-1 (see FIGURE 1-17)
2. Connect J8-2 to the 4-20 mA input on the remote system.
3. Terminate the 4-20 mA line with 230-500 Ohms of resistance.
4. For three-wire operation, connect the signal ground to J8-3.
5. Connect the sensor module to main pc board connector J-1.
6. Wire for optional relays, if applicable (see Appendix A).
7. Assemble lid on the enclosure.
Figure 1-17. Three-Wire Printed Circuit Board
(with HART Protocol)
1-19
Page 33
Installing the Ultima X Remote Sensor Module
The Remote Sensor Module is used with the Ultima X Gas Monitor for installations requiring remote placement of the gas sensor.
FIGURES 1-18 and 1-19 show the general-purpose and explosion-proof configurations.
Figure 1-19. Remote Module Explosion-Proof
Ultima X Series Wiring
Figure 1-18. Remote Module General-Purpose
Ultima X Series Wiring
1-20
Page 34
The Remote Sensor Module should be mounted in a manner similar to the Ultima X (see Chapter 1, "Installing Your Gas Monitor") and at a maximum distance outlined in TABLE 1-6.
Permanently connect 1/4" ID tubing to the post on the windguard. Route this tubing to the Ultima X Gas Monitor, ensuring that there are no kinks, leaks or other obstructions. Secure this tubing near the monitor; it is used to deliver check gas to the sensor module during calibration.
Electrical Connections for Remote Sensors
Before wiring the Ultima X Series Remote Sensor Module, disconnect the power source feeding the Remote Sensor Module and the Ultima X Series Gas Monitor/Less Sensor; otherwise, electrical shock or ignition of hazardous atmos­pheres could occur.
When installing an Ultima X Series Remote Sensor Module with its mating Ultima X Series Gas Monitor/Less Sensor, fol­low National Electrical and local procedural Codes; failure to do so can result in an unsafe condition.
Five conductors are required for the Ultima XE and Ultima XA Remote Sensor Modules. Four conductors are required for the Ultima XIR Remote Sensor Module. The Ultima X Series Monitor has a five-wire terminal to accommodate up to #16 AWG conductors. For wiring details, see the applicable Installation Outline Drawing listed in TABLE 1-1.
Some installations require metal pipe or metallic conduit. In these cases, separate conductors or unshielded cable may be used.
For open wiring, shielded wire or cable should be used to minimize the possibility of noise interference and contact with other voltages. Selection of this shielded cable must comply with local requirements.
"
WARNING
"
WARNING
1-21
Page 35
Table 1-4. Remote Module Wiring and Placement
GAS TYPE MINIMUM WIRE SIZE MAXIMUM DISTANCE
Toxic and Oxygen 20 AWG 100 feet (30m)
Catalytic Combustible 18 AWG 50 feet (15 m)
16 AWG 100 feet (30 m)
*IR Combustible 16 AWG 50 feet (15 m)
12 AWG 100 fee (30 m)
TABLES 1-7 and 1-8 show suggested cables for Ultima X Series installations; other cables are available which are also adequate.
Table 1-5. Remote Sensor Wiring Cable
SUPPLIER CATALOG NUMBER DESCRIPTION
Alpha Wire Corp 5525 5 cond., shielded, 18 AWG
5535 5 cond., shielded, 16 AWG 5514 4 cond., shielded, 20 AWG
Table 1-6. Low Temperature Wiring Cable
SUPPLIER CATALOG NUMBER DESCRIPTION
Alpha Wire Corp 45525 5 cond., shielded, 18 AWG
45366 6 cond., shielded, 16 AWG 45545 5 cond., shielded, 14 AWG
At the Ultima X Series Remote Sensor Location:
1. Open the Ultima X Series Remote Sensor cover by removing lid.
2. For the Ultima XA Gas Monitor, route the power and signal cable from the Gas Monitor through a customer-created opening in the enclosure and wire it to the appropriately labeled connection on the terminal block (FIGURE 1-4).
For the Ultima XE or XIR Gas Monitor, route the cable from the Gas Monitor through a wire entry hole in the enclosure and wire it to the appropriately labeled connection on the terminal block (FIGURE 1-5).
3. Verify the sensor connector is firmly seated on the terminal board.
4. Re-install the cover of the Ultima X Series Remote Sensor.
1-22
Page 36
NOTES:
Grounding
Incoming power and signal cable shield should be earth grounded at the power source.
Connect power and remote sensor cable shields to shield terminals on main pc board.
Provide shield terminations inside the sensor housing as indicated on Installation Outline Drawings for Remote Sensor. See TABLE 1-1 for Installation Outline Drawing document numbers.
Cable Size
Cables larger than #16 AWG will require a splice of smaller cable to fit the connector.
1-23
Page 37
2-1
Chapter 2, Start-up and Calibration
Initial Start-up
The Ultima X Series Gas Monitors are factory-calibrated and ready for immediate use.
Once power is applied to the unit, the LCD shows a test of all display words. The software version number displays; then, a 30­second (self-check) countdown for sensor stability begins.
During the 30-second countdown, the output signal is the same as the calibration signal when enabled during a normal calibration. This is described later in this chapter under "Ultima X Series Gas Monitor Calibration Output Signal".
For units with LEDs, the Alert red LED will be solid ON during the 30-second countdown.
After the 30-second countdown, observe that the gas type and gas concentration (ppm, % Gas, or % LEL) alternately flash (FIGURE 2-1).
For units with LEDs, the Normal green LED will be solid ON after the 30-second countdown.
A complete listing of instrument operation features can be found in TABLE 2-1.
During normal operation, the Ultima X Monitor displays the gas concentration of the surrounding environment. The corresponding output signal can be transmitted to a controller or read directly from the optional HART port with an HCF-approved communicator (such as the Emerson 375 HART Communicator, or equivalent).
.
Figure 2-1. LCD Gas Concentration Display
Page 38
2-2
NOTE: The catalytic combustible model of the Ultima X Series Gas
Monitors is capable of detecting concentrations of certain com­bustible gases above 100% LEL. When exposed to these con­centrations, the Ultima X Series Gas Monitors will display one of two modes:
+LOC % LEL - The Ultima X Series Gas Monitor has been
exposed to a high concentration of gas (above the LEL) and it is possible that the over-range condition may still exist.
OVER % LEL - The Ultima X Series Gas Monitor has been
exposed to a high concentration of gas (above the LEL) and the over-range condition definitely still exists.
In either mode, correct the condition causing the excessive gas level and vent or purge the area before attempting the following.
In the +LOC % LEL mode, the output signal will also be locked at full-scale. If this condition occurs, the Ultima X Series Gas Monitor must be unlocked by performing a "Zero Function" with the Ultima X Series Gas Monitor Calibrator or Controller. The Ultima X Series Gas Monitor will not revert to a normal condition until a successful zero operation has been performed. This is an exclusive safety feature of the Ultima X Series Gas Monitor which pre-empts the possibility of ambiguous readings when the sensor is exposed to concentration of gas above 100% LEL
In the OVER % LEL mode, the combustible gas is over the100%LEL range. It returns to normal operation when gas concentration level falls below 100%LEL.
Table 2-1. Instrument Operation
NOTES:
1
ALERT option causes the 4-20 mA output to be set to 3.75 mA during O2sensor calibration (if the Cal Signal Option is also enabled). If the ALERT option is disabled and the Cal Signal enabled, the output is set to 21 mA during the O2sensor calibration.
2
Swap Delay timeout is 60 seconds if enabled; 0 seconds otherwise.
3
The Swap Delay feature enables a one-minute hold-off of the Sensor Missing Fault, allowing the user to "Swap" or change sensors without having the 4-20 mA set to a fault condition.
4
Alarming operation are followed if the alarms are enabled.
"
CAUTION
Page 39
2-3
OPERATION LEDs 4 to 20 mA FAULT RELAY
GREEN RED
NORMAL ON OFF Gas value Energized
steady
ALARMING OFF Flashing Gas value Energized
FAULT OFF ON 3.0 mA De-energized
steady
POWER OFF ON <3.75 mA De-energized UP (HART steady Version)
POWER OFF ON <3.1 mA De-energized UP steady (Non-HART Version)
COUNT OFF ON steady ALERT option1disabled; Energized DOWN (All 21.0 mA for O2; if ALERT option disabled Versions) 3.75 mA for others
ALERT option1enabled: De-energized
3.75 mA for all if ALERT option enabled
3
SENSOR OFF ON steady 3.0 mA if Swap Delay De-energized if Swap Delay MISSING/ timeout2expired, timeout2expired, Swap COUNT- Swap Delay3disabled Delay3 disabled or FAULT DOWN or FAULT
Previous gas value Energized if Swap Delay
3
if Swap Delay3enabled enabled and Swap Delay and Swap Delay timeout2timeout2not expired not expired
SENSOR OFF ON steady 3.75 mA if cal signal Energized if ALERT CAL enabled and ALERT option disabled
option1enabled; gas value if cal signal disabled
21.0 mA for O2if De-energized if ALERT cal signal enabled and option enabled
1
ALERT option1disabled
CAL OFF ON steady 4 mA if 4 mA calibration Energized if ALERT option 4-20 selected disabled
20 mA if 20 mA De-energized if ALERT calibration selected option enabled
1
CAL FAULT OFF ON steady Gas value De-energized two seconds
every minute
UNDER- OFF ON steady 3.0 mA if gas value De-energized RANGE 0 or less;
gas value otherwise
OVER- ON OFF
4
21.0 mA Energized
RANGE/ steady
4
LOC
Page 40
2-4
Calibration Basics
While the Ultima X Series Gas Monitor is factory-calibrated, it is good practice to calibrate the unit once it is installed in its final environmental destination.
As with any type of gas monitor, the only true check of its performance is to apply gas directly to the sensor. The frequency of the calibration gas tests depends on the operating time and chemical exposures of the sensors. New sensors should be calibrated more often until the calibration records prove sensor stability. The calibration frequency can then be reduced to the schedule set by the safety officer or plant manager.
Catalytic Combustible sensors located in areas where non-combustible chemicals may leak, particularly ones known to reduce the sensitivity (see following list) should be calibrated after such exposures.
Silanes, Silicates, Silicones and Halides (compounds containing Fluorine, Chlorine, Iodine or Bromine)
TABLE 3-2 in Chapter 3 lists interferants for electrochemical sensors.
Before calibrating, the Ultima X Series Gas Monitor must be powered for a minimum of one hour to allow the sensor to settle into its new environment.
Before attempting a calibration, power the unit at least one full hour.
To ensure a fully functional sensor, perform a calibration check and adjustments at initial start-up and at regular intervals.
When it is determined that calibration adjustments are required, the Ultima X Series Gas Monitor provides a one-man, non-intrusive method of adjustment at the unit.
To calibrate the unit, one of the following accessories is necessary:
Ultima Calibrator P/N 809997 (FIGURE 2-2)
Ultima Controller P/N 809086(FIGURE 2-3)
Optional Push-button Calibration (FIGURE 2-4). Instructions for use of the optional push-button are given in Appendix A.
"
CAUTION
Page 41
2-5
HART®-compatible communications interface with Device Description Language capability (DDL) or generic HART interface with Manufacturer Specific Command capability. This hand-held HART Communicator must be HART revision 7 compliant and can be obtained from a HART-authorized supplier. See Appendix D for command definitions.
.
.
.
Figure 2-4. Ultima X Optional Push-button Calibrator
Figure 2-3. Ultima Controller
Figure 2-2. Ultima Calibrator
Page 42
Ultima Calibrator
The Ultima Calibrator allows the following functions:
Zero
Calibration (zero and span)
Changing address for some models.
Ultima Controller
The Ultima Controller also provides the above functions, plus access to the following features:
Three levels of alarm and relays
Date of last successful calibration
Maximum gas readings over selected time periods
Average gas readings over selected time periods
Changing span gas value from factory-set value
Access to real-time clock for time and date
Changing of full scale value.
NOTE: See Ultima Controller/Calibrator manual (P/N 813379) for
full functionality.
NOTE: When an Ultima X Series Gas Monitor has an active latched
alarm (indicated by a flashing alarm display):
An infrared (IR) remote device (such as the Ultima Calibrator or Controller) may be used to reset this alarm.
The next IR command it receives from a calibration device will reset the latched alarm (if it is not beyond the alarm threshold). The intended IR command will be ignored and interpreted as an 'alarm reset.' When the latching alarm function is inactive, other valid IR commands may be used.
Calibration Output Signal
The Ultima X Series Gas Monitor is shipped with the calibration output signal DISABLED so the output signal will track the gas concentration value during the calibration process. In some applications, it may be desirable to disable or lock the output to a pre-determined output value to prevent activation of alarm devices. The calibration signal can be ENABLED using the Ultima Controller or a HART controller with DDL­or manufacturer-specific command capability. When the calibration
2-6
Page 43
2-7
signal is enabled, the output signal is 3.75 milliamps for the 4 to 20 milliamp output models.
NOTE: For oxygen sensors, the calibration signal will be 21 mA.
Oxygen can be set to a 3.75 mA calibration signal by turning ON the ALERT option as described in the Ultima Controller manual.
Calibration Kit
Calibration Kits are available for the Ultima X Gas Monitors. For the recommended calibration kit, see Ultima Controller/Calibrator manual (P/N 813379).
Ultima X Series Gas Monitor Calibration Procedure
Read all calibration instructions before attempting an actual calibration. Also, identify and become familiar with all of the calibration components. During the calibration, it is necessary to quickly apply the span gas to the unit. Prior connection of the calibration components will aid in the ease of unit calibration.
The only true check of any gas monitor's performance is to apply gas directly to the sensor. The calibration procedure must be performed regularly.
NOTES:
If this is the first calibration or, if the sensor element has been changed or replaced, see Chapter 2, "Initial Calibration."
If this is an oxygen sensor, see subsequent section, "Oxygen Calibration."
If this is an XIR sensor, see subsequent section, "XIR Calibration."
Apply power to the unit at least 1 hour before calibrating.
Due to the unstable nature of Chlorine Dioxide (ClO2), Chlorine gas is used as a calibration simulant. If using the MSA calibration system and gas cylinder (P/N 710331), the response ratio is 2:1. In other words, the 2 ppm sample of Chlorine should be set to read 1 ppm of ClO2. The default value for the calibration gas on the ClO
2
Ultima X Series Gas Monitor is 1 ppm.
For Cl2and ClO2calibration, do not mix regulators. Use only one regulator for each of these gases. They will not work properly if one regulator is used for multiple gases.
Page 44
2-8
Due to the reactivity of HCL with flow system components, the flow control regulator must only be used for HCL gas. HCL gas must be run through the flow control regulator and tubing for five minutes before attempting a calibration. After a successful calibration, flush the flow control regulator and tubing with 100% Nitrogen for five minutes. Store the flow control regulator in the desiccated bag included in Calibration Kit 54 or equivalent dry container.
Equipment Required
Three calibration kits (numbered 40, 41, and 54) are available from MSA for diffusion Ultima X Series Gas Monitors. Kit 40, 41, and 54 are housed in a convenient carrying case and contain all items necessary (less gas) for a complete and accurate calibration.
These Kits do not calibrate Ultima X Series units equipped with a flow cap.
NOTE: The calibration procedure for the sample draw Ultima XE/XA
Monitor is the same as the procedure for the diffusion version, except calibration gas is applied to the calibration entry port of the inlet flow block and the cal kit for pumped units provides a flow matching regulator.
The check or calibration gases can also be carried in the case. See TABLE 2-2 for the appropriate zero and span gas cylinders for your Ultima X Series Gas Monitor.
TABLE 2-2 shows the recommended calibration kit for Ultima X Series Gas Monitors. Typically, Cal Kit 41 uses 0.25 LPM regulator and a calibration cap to contain the calibration gas. Cal Kits 40 and 54 use a
1.5 LPM regulator and no calibration cap. If Cal Kit 41 is recommended and the application is such that the calibration cap cannot be used (such as for a remote sensor application), Cal Kit 40 may be used. However, any time Cal Kit 40 is used, ambient wind conditions must be minimized to avoid a calibration with increased sensitivity.
NOTE: The Ultima XIR uses Cal Kit 40 and does require a calibration
cap. This calibration cap (P/N 10041533) is shipped with the product.
Page 45
2-9
These calibration kits contain zero caps to use in place of zero calibration gas. These caps can only be used when the ambient air does not contain the gas the monitor is detect­ing. If there is any doubt, use zero gas when zeroing the Ultima X Monitor; otherwise, improper calibration could occur.
Span Gas Values
The Ultima X Series Gas Monitor is factory-shipped with a preset span gas value (TABLE 2-2). This span gas value can be changed via the Ultima Controller or a HART controller; otherwise, the span gas must correspond to preset concentrations. See Section 3 of the Controller/Calibrator Manual (P/N 813379) to change the span gas value. See Appendix D for the equivalent HART command
The span gas value of Ultima X Gas Monitor catalytic combustible models are pre-set to one of the broad categories shown in TABLE 2-2. Specific span gas values for all combustible models are listed under each category given in TABLE 2-3.
Always calibrate for the least sensitive gas or vapor (higher number category) expected to be measured (TABLE 2-3); otherwise, instrument readings may be incorrect.
Table 2-2. Factory-set Span Values
GAS TYPE RANGE SPAN GAS RP CAL WARM-UP
PRESET CYLINDER KIT TIME VALUES P/N
CARBON 0-100 PPM 60 PPM 710882 40 15 minutes MONOXIDE 0-500 PPM 300 PPM 10027938
0-1000 PPM 400 PPM 10028048
SULFUR 0-25 PPM 10 PPM 10028070 40 15 minutes DIOXIDE 0-100 PPM 10 PPM 808978
HYDROGEN 0-10 PPM 5 PPM 710414 40 15 minutes SULFIDE 0-50 PPM 40 PPM 10028062
0-100 PPM 40 PPM 10028062 0-500 PPM 250 PPM 10089547
NITRIC OXIDE 0-100 PPM 50 PPM 10028074 40 15 minutes
NITROGEN 0-10 PPM 5 PPM 710332 41 41 30 minutes DIOXIDE
"
WARNING
"
WARNING
Page 46
2-10
GAS TYPE RANGE SPAN GAS RP CAL WARM-UP
PRESET CYLINDER KIT TIME VALUES P/N
CHLORINE 0-5 PPM 2 PPM 710331 41 30 minutes
0-10 PPM 2 PPM 30 minutes 0-20 PPM 10 PPM 10028066 30 minutes
HYDROGEN 0-50 PPM 10 PPM 10028072 41 30 minutes CYANIDE
HYDROGEN 0-10 PPM 8 PPM 10028070 41 30 minutes FLUORIDE
(7)
CHLORINE 0-3 PPM 1 PPM 710331 41 30 minutes DIOXIDE
(4)
OXYGEN 0-10% 5% 493580 40 15 minutes
0-25% 20.8% 10028028
(2)
15 minutes
NATURAL GAS
(3)
0-100% LEL 25% LEL
(1)
10028034 40 15 minutes
PETROLEUM 0-100% LEL 40% LEL
(1)
10028034 40 15 minutes
VAPORS
(3)
(GASOLINE)
GENERAL 0-100% LEL 55% LEL
(1)
10028034 40 15 minutes
SOLVENTS
(3)
NON- 0-100% LEL 29% LEL
(1)
10028034 40 - - -
METHANE IR
METHANE IR 0-100% LEL 50% LEL
(5)
10028032 40 - - -
PHOSPHINE 2.0 PPM 0.5 PPM 710533 41 24 hours
ARSINE 2.0 PPM 1.0 PPM 710533 41 24 hours
SILANE 25 PPM 5 PPM 10014897 41 4 hours
DIBORANE 50 PPM 15 PPM 10014897 41 30 minutes
FLUORINE 5.0 PPM 4.0 PPM 710331 41 30 minutes
BROMINE 5.0 PPM 2.5 PPM 710331 41 30 minutes
AMMONIA 100 PPM 25 PPM 10028076 40 30 minutes
0-1000 PPM 300 PPM 10044014 40 30 minutes
HYDROGEN 0-1000 PPM 500 PPM 10022386 40 30 minutes
ETHYLENE
(6)
0-10 PPM 4.0 PPM 10028070 40 24 hours
OXIDE
CARBON 0-5000 PPM 2000 PPM 479266 40 - - ­DIOXIDE IR 0-2% 1.5% 807386
0-5% 2.5% 479265
HYDROGEN CHLORIDE 0-50 PPM 40 PPM 10028078 41 30 minutes
NOTES:
1
Calibrated with Propane (.6% gas by volume)
2
Not required for standard calibration procedure
3
For combustible gas, it is good practice to calibrate unit with gas to be detected
4
ClO2is calibrated with Cl2or use ClO2Calibrator Kit (P/N 710420)
5
Methane IR is calibrated with 50% LEL Methane
6
ETO is calibrated with SO2.
7
Hydrogen Fluoride (HF) is calibrated with Sulfur Dioxide (SO2). 10 ppm SO2equals 8 ppm HF.
Page 47
2-11
Table 2-3. Calibration Guide for Combustible Gas Sensor
CATEGORY 31: FOR CATALYTIC TYPE 1S NATURAL GAS
To detect the following gases, recalibrate with 0.6% propane & set span gas value accordingly Acetaldehyde 23 Hydrogen 16 Acetylene 24 MAPP Gas 20 Butadiene, 1, 3 25 Methane 20 Carbon Monoxide 20 Methanol 20 Ethane 24 Methylene Chloride 24 Ethylene 25 Monomethyl Amine 22 Ethylene Dichloride 22 Trigonox B 22
CATEGORY 32: FOR CATALYTIC TYPE 1S PETROLEUM VAPORS
To detect the following gases, recalibrate with 0.6% propane & set span gas value accordingly 1, 1, 1-Trichloroethane 32 Ethylene Oxide 36 Acetic Acid 28 Freon 152A 28 Acetone 37 Gasoline 35 Acrolein 28 Hexane 40 Acrylonitrile 26 Isoprene 33 Allyl chloride 30 Methyl Acetate 34 Benzene 37 Methyl Chloride 32 Butane (n) 36 Methyl Propene (2) 29 Butane (iso) 32 Methyl t-Butyl Ether 35 Butanol (iso) 38 Pentane (n) 36 Butene-1 34 Pentane (iso) 36 Butene-2 37 Pentene 35 Butyl Acetate (n) 28 Propane 29 Butylene 33 Propanol (n) 36 Butyraldehyde 30 Propanol (iso) 37 Chlorobenzene 38 Propylene 33 Cyclohexane 37 Propylene Oxide 33 Dimethoxyethane 26 Tetrahydrofuran 30 Dioxane, 1, 4 39 Toluene 39 Epichlorhydrin 33 Trichloroethylene 35 Ethanol 30 Triethylamine 38 Ether, Diethyl 37 Vinyl Acetate 34 Ether, Dimethyl 30 Vinyl Chloride 32
CATEGORY 33: FOR CATALYTIC TYPE 1S GENERAL SOLVENTS
To detect the following gases, recalibrate with 0.6% propane & set span gas value accordingly Amyl alcohol 43 JP-4 41
Butanol (n) 48 Methyl Cellosolve 49 Butyl Acrylate 46 Methyl Ethyl Ketone 52 Cellosolve 42 Methyl Isobutyl Ketone 53 Di isopropylamine 42 Methyl Methacrylate 40 Diethylamine 41 Naphtha, VM&P 53 Ethyl Acetate 43 Octane (iso) 52 Ethyl Acrylate 52 Propyl Acetate 45 Ethyl Benzene 41 Styrene 42 Heptane 42 Xylene 50 Hexene 42
Page 48
2-12
CATEGORY 38: ULTIMA XIR METHANE
To detect the following gases, recalibrate with 2.5% methane & set span gas value accordingly
Methane 50
CATEGORY 39: ULTIMA XIR NON-METHANE
To detect non-methane gases, recalibrate with stated % propane & set span gas value as given in Appendix B
For additional gases for the Ultima XIR, see Appendix B.
INITIAL Calibration
When the unit is powered up for the first time, or when a new sensor module is placed in the unit, an INITIAL Calibration is recommended. This procedure enables the unit to gather data about the sensor to make accurate decisions for the CHANGE SENSOR function and the CAL FAULT function to work properly. During normal use, INITIAL calibration should only be used when a standard calibration will not clear a fault condition due to use of incorrect calibration gas or another similar situation.
The INITIAL calibration is accomplished by:
simultaneously pressing the ZERO and CALIBRATE buttons of the Ultima Calibrator or
pressing and holding SPAN button on the Ultima Controller or
using the optional push-button calibration as outlined in Appendix A, "Optional Push-button Calibration"
using the HART Communicator as described in Appendix D.
After starting the INITIAL calibration:
The display should show "APPLY ZERO GAS"
The word "ICAL" on the display distinguishes an INITIAL Calibration from a regular calibration. If "ICAL" does not appear, abort the calibration; then, retry the above procedure.
NOTE: The zero or calibration process can be aborted at any time sim-
ply by pressing any button during the 30-second countdown on the Calibrator while aiming at the unit or by pressing and releas­ing the push-button if push-button calibration is available.
The remainder of the procedure is now the same as that for a regular calibration, as described in the following procedure.
Page 49
2-13
Standard Calibration
A standard calibration includes a "zero" and "span" procedure as described in the following procedures. If the user chooses to only perform a "zero" procedure, they may do so by pressing the ZERO button on the Calibrator or Controller instead of the CALIBRATE button as described as follows, or by using the optional push-button calibration as outlined in Appendix A, "Optional Push-button Calibration". Both the "zero" and the "span" functions are available on the HART Controller and are described in Appendix D.
Zeroing
1. If Using the zero cap:
If the ambient air is suitable, with no traces of the gas of interest, place the appropriate Calibration Kit zero cap over the SensorGard inlet and wait two minutes; otherwise, use zero gas.
2. If Using zero gas cylinder:
a. Locate the zero gas cylinder and the Calibration Kit Flow
Controller.
b. Screw the Flow Controller onto the top of the zero gas cylinder.
c. Locate the Tube Assembly from the cal kit.
d. Push the smaller end of the Tube Assembly over the Flow
Controller gas outlet and ensure tubing completely covers the gas outlet.
e. When using Cal Kit 40, connect the other end of the tubing
over the SensorGard inlet.
When using Cal Kit 41, locate the cal cap (with hole for tubing) and push the tubing through the hole in the bottom of the cap. Then, connect the end of the tubing over the sensor inlet and push the calibration cap over the entire sensor inlet.
f. Turn on zero gas flow by turning knob on the flow controller.
3. Point the Calibrator or Controller at the Ultima X Series Monitor display; press the CALIBRATE button.
NOTE: The zero or calibration process can be aborted at anytime
during the 30-second countdown interval; simply press any button on the Calibrator or Controller while aiming it at the unit or by pressing and releasing the push-button if push­button calibration is available.
NOTE: The 30-second countdown interval is omitted for oxygen
units; it is electronically zeroed.
Page 50
2-14
The display shows:
A countdown from 30 to 0 seconds
APPLY ZERO GAS (FIGURE 2-5).
.
4. After the 30 second countdown:
The display alternates between "CAL" and a value. This value is the actual reading of the gas concentration the sensor is detecting.
Once the gas value on the display is stable, the alternating display stops. If the calibration is successful, the display will show END.
a. If using the zero cap: remove it.
b. If using a zero gas cylinder:
1) Turn OFF the gas flow by turning the flow controller knob.
2) Remove the tubing from the SensorGard.
If the calibration output signal is enabled during calibration, it will be held at the lockout value for an additional two minutes or until after the span routine if performing a full calibration.
c. If CAL FAULT appears on the display, this indicates:
An unsuccessful attempt to zero or calibrate the Ultima X Series Monitor
The Ultima X Series Monitor is operating with the calibration parameters defined before the calibration was attempted.
Figure 2-5. Apply ZERO Gas Flag
Page 51
2-15
• See Troubleshooting Guidelines found in Chapter 4.
To extinguish the CAL FAULT, a complete, successful calibration procedure must be performed.
The Ultima X Series Gas Monitor allows automatic zero adjustment only within a pre-defined range. It cannot make corrections outside this range, such as when an empty or wrong cylinder of gas is applied or failure to begin gas flow within the allotted 30-second countdown occurs.
If only a ZERO was performed, the procedure is complete and the user should return the calibration equipment to the cal kit. If a CAL was performed, the gas monitor will continue to the "span" sequence as described in the following section.
Spanning
5. During a regular calibration, the Ultima X Series Gas Monitor automatically begins the span countdown after a successful zeroing of the unit. The span countdown is 30 seconds (FIGURE 2-6).
NOTE: The span process can be aborted at any time during the
countdown by simply pressing any button on the Calibrator while aiming it at the unit or by pressing and releasing the push-button if push-button calibration is available.
.
Figure 2-6. Apply SPAN Gas Flag
Page 52
2-16
6. Locate the span gas cylinder and the Calibration Kit Flow Controller.
7. Screw the Flow Controller onto the top of the span gas cylinder.
8. Locate the Tube Assembly from the cal kit.
9. Push the smaller end of the Tube Assembly over the gas outlet of the Flow Controller and ensure that the tubing completely covers the gas outlet.
10.When using Cal Kit 40, connect the other end of the tubing over the SensorGard inlet.
When using Cal Kit 41, locate the cal cap (with hole for tubing) and push the tubing through the hole in the bottom of the cap. Then, connect the end of the tubing over the sensor inlet and push the calibration cap over the entire sensor inlet.
11. Turn ON the gas flow by turning the flow controller knob.
It is good practice to have all calibration components previously assembled.
Ensure that any calibration gases are applied during the 30­second count down period.
If a CAL FAULT indication is on the Ultima X Series Gas Monitor display before the user is able to apply the gas, a steady state gas condition was reached, causing the unit to use a wrong reading as a span indication.
It is necessary to restart the calibration process to clear this condition.
12. After the 30 second countdown:
The display alternates between "CAL" and a value. This value is the actual reading of the gas concentration the sensor is detecting.
Once the gas value on the display is stable, the alternating display stops. If the calibration is successful, the display will show END for approximately two seconds. (FIGURE 2-7).
No user adjustments are necessary.
The display will show the span gas value while the span gas is flowing to the unit.
13.Turn OFF the gas flow by turning the knob on the flow controller.
Page 53
2-17
If the calibration output signal is enabled during calibration, it will be held at the lockout value for two additional minutes after END is displayed.
When the span gas is removed from the sensor, the sensor reading should change to show an ambient condition.
If a CAL FAULT appears on the display, this indicates:
An unsuccessful attempt to calibrate the Ultima X Series Gas Monitor
The Ultima X Series Gas Monitor is operating with the calibration parameters defined before the calibration was attempted.
To extinguish the CAL FAULT flag, a complete calibration procedure must be performed.
The Ultima X Series Gas Monitor allows automatic zero and span adjustments within a pre-defined range. It cannot make corrections outside this range, such as when an empty or wrong cylinder of gas is applied or failure to begin gas flow within the allotted 30­second countdown occurs.
14. After a successful calibration, remove the tubing from the Flow Controller and remove the Flow Controller from the cylinder; return all items to their appropriate location in the Calibration Kit.
Figure 2-7. Calibration End Display
Page 54
2-18
OXYGEN Calibration
NOTE: If this is the first calibration after the sensor element is replaced,
perform an "Initial Calibration."
Oxygen calibration is slightly different from other gases. When the ZERO function is performed, the 30-second countdown is omitted because the Ultima/Ultima X Series unit performs the zero electronically. No calibration cap or zero gas is necessary.
To meet the specification stated, it is necessary to span the oxygen Ultima/Ultima X Series Gas Monitor with the Calibration Kit and an oxygen cylinder. The concentration of oxygen in air varies slightly due to changing relative humidity and pressure levels. These variations in oxygen levels are detected by the oxygen Ultima/Ultima X Series Gas Monitor. To meet the reproducibility specification, it is necessary to use a calibration gas cylinder. This ensures the same concentration of oxygen for every calibration.
25% Oxygen Ultima/Ultima X Series Gas Monitor
For the SPAN function, ambient air is generally adequate for the 25% oxygen Ultima/Ultima X Series Gas Monitor as the expected default span value is 20.8%. Therefore, when the display prompts "APPLY SPAN GAS" it would be adequate to simply allow the countdown to occur without applying gas.
NOTE: If the sensor is located in an area of normally low or enriched
oxygen, then a 20.8% oxygen sample must be applied when the display prompts: "APPLY SPAN GAS".
XIR Calibration
Although a full calibration (zero and span) can be performed on the Ultima XIR Gas Monitor, a no-gas calibration is sufficient to properly calibrate the monitor. Typically, a zero adjustment is all that is required for a full calibration. Normally, any degradation of the sensor's performance is associated with slight drifts in its zero response which, in turn, will adversely affect its span performance. Restoring the sensor's zero is typically sufficient to restore its span performance.
A zero adjustment is performed by one of the following methods:
pressing the ZERO button on the Calibrator or Controller
using the optional push-button calibration as outlined in Appendix A, "Optional Push-button Calibration"
using the HART Controller or DCS, as described in Appendix D.
Page 55
2-19
Follow the "Zeroing" instructions given earlier in this chapter. After completing the zeroing function, perform a span check to ensure proper operation. If the span check is unsuccessful, perform a full calibration.
NOTE: For calibration of an XIR sensor operating with a Flow Cap,
temporarily replace the Flow Cap with the Environmental Guard (packaged with the instrument) and perform the calibration procedure.
The Calibration Cap must be removed from the XIR environ­mental guard after completing the Zeroing and/or Spanning procedure; otherwise, the sensor cannot perform properly.
Calibration Documentation
The Ultima X Series Monitor records the date of the last successful calibration. This date can then be displayed on the front-panel LCD (with the use of the Controller or via the HART Controller).
"
WARNING
Page 56
3-1
Chapter 3, Specifications
Table 3-1. Performance Specifications
GAS TYPES Combustibles, Oxygen & Toxics
TEMPERATURE TOXICS & OPERATING 0 to 40°C (32 to +104°F) RANGE OXYGEN RANGE
*EXTENDED -20 to +50°C (-4 to +122°F) RANGE
OPERATING 0 to +30°C (32 to +86°F) RANGE NH
3
*EXTENDED -10 to +40°C (+14 to +104°F) RANGE NH3, Cl2, ClO
2
Calibrate within operating range
CATALYTIC SINGLE & -40 to +60°C (-40 to +140°F) COMBUSTIBLES DUAL
MODULES
IR SINGLE & -40 to +60°C (-40 to +140°F) COMBUSTIBLES DUAL
MODULES
STORAGE -40 to +70°C (-40 to +158°F) TEMPERATURE or limits of the sensor RANGE
DRIFT ZERO DRIFT Less than 5%/year, typically
SPAN DRIFT Less than 10%/year, typically
NOISE Less than 1% FS
*Extended Range = The sensor may not meet all of the accuracy parameters listed.
Page 57
3-2
ACCURACY
GAS LINEARITY REPEATABILITY
CARBON MONOXIDE +2% FS +1% FS or 2 ppm
OXYGEN +2% FS +1% FS
HYDROGEN SULFIDE +10% FS or 2 ppm +1% FS or 2 ppm
CHLORINE +10% FS or 2 ppm +5% FS or 1 ppm
SULFUR DIOXIDE +10% FS or 2 ppm +1% FS or 2 ppm
NITRIC OXIDE +10% FS or 2 ppm +1% FS or 2 ppm
NITROGEN DIOXIDE +10% FS or 2 ppm +4% FS or 1 ppm
HYDROGEN CYANIDE +10% FS or 2 ppm +4% FS or 2 ppm
HYDROGEN CHLORIDE +10% FS or 2 ppm +10% FS or 2 ppm
CATALYTIC <50% LEL +3% FS +1% FS COMBUSTIBLE GAS >50% LEL +5% FS +1% FS
IR COMBUSTIBLE <50% LEL - +2% +2% FS GAS: METHANE, PROPANE >50% LEL - +5% +2% FS
CHLORINE DIOXIDE +10% FS or 2 ppm +5% FS or 1 ppm
ETHYLENE OXIDE +10% FS +5% FS
AMMONIA +10% FS +5% FS
HYDROGEN +5% FS +5% FS
PHOSPHINE +10% FS +10% FS
ARSINE +10% FS +10% FS
SILANE +10% FS or 2 ppm +1% FS or 2 ppm
DIBORANE +10% FS or 2 ppm +1% FS or 2 ppm
FLUORINE +10% FS or 2 ppm +5% FS or 1 ppm
HYDROGEN FLUORIDE +10% FS +10% FS
BROMINE +10% FS or 2 ppm +5% FS or 1 ppm
Page 58
3-3
STEP TIME TO REACH Less than 12 seconds (typically 6 seconds) CHANGE 20% OF SCALE- Less than 20 seconds (ETO) RESPONSE OXYGEN & TOXICS
TIME TO REACH Less than 30 seconds (typically 12 seconds) 50% OF SCALE- Less than 45 seconds (ETO) OXYGEN & TOXICS
TIME TO REACH Less than 10 seconds 50% OF SCALE­COMBUSTIBLES
TIME TO REACH Less than 30 seconds 90% OF SCALE­COMBUSTIBLES
HUMIDITY 15 to 95% RH, non-condensing,
24 hours or less
15 to 60% RH (SO2**)
35 to 95% RH, long term
SENSOR CATALYTIC 3 years, typically LIFE COMBUSTIBLES
OXYGEN & TOXICS 2 years, typically
AMMONIA ***
FULL REPLACEMENT 1 year from installation; 10 years for IR
WARRANTY Sensor source (see "MSA Instrument
Warranty" in this manual for complete details)
WIRING OXYGEN 2-wire or 3-wire REQUIRE- & TOXICS MENTS
COMBUSTIBLES 3-wire
RELAYS 3-wire
**SO2 sensor should not be used in dirty or humid environments.
***0-100 ppm NH3sensor is consumable at a rate of 10% for every 200 ppm/hours of
exposure. 0-1000 ppm NH3sensor is consumable at a rate of 10% for every 1500 ppm/hours of exposure.
Page 59
3-4
POWER CON- OXYGEN & TOXICS 8 VDC 250 mA max SUMPTION * 12 VDC 175 mA max (TOTAL UNIT 24 VDC 100 mA max WITH RELAYS)
CATALYTIC 8 VDC 600 mA max COMBUSTIBLES 12 VDC 400 mA max
24 VDC 210 mA max
IR COMBUSTIBLES 8 VDC 870 mA max
12 VDC 550 mA max 24 VDC 290 mA max
SIGNAL COMBUSTIBLES 3-wire current source OUTPUT
OXYGEN 2-wire current sink & TOXICS 3-wire current source
XA SIZE 9.423" H x 5.125" W x 3" D PHYSICAL (239.34 mm x 130 mm x 76 mm)
WEIGHT 1.5 pounds (0.7 kilograms)
XE SIZE 10.280" H x 6.312" W x 3.911" D inches PHYSICAL (261.11 mm x 160.33 mm x 99.34 mm)
WEIGHT 10.4 pounds (4.72 kilograms)
* The HART output signal is not available below 12 VDC on the two-wire pc board.
Table 3-2. Sensor Response to Interferants
If your readings are higher or lower than expected, it could be due to the presence of an interferant gas.
• The gas listed in column 1 is presented to the sensor.
• Column 2 indicates the concentration of that gas presented to the sensor.
• The remaining columns indicate the respective responses by the sensors to each particular gas.
For Example:
Scan column 1 until you locate "hydrogen". Column 2 shows that 500 ppm of hydrogen was presented to the sensor. Column 3 shows that a CO (filtered) sensor gave an equivalent response of 200 ppm. Column 4 shows that an H2S sensor gave an equivalent response of 0.5 ppm, etc.
Page 60
3-5
ND = No Data
INTER- CONCEN- CO H2SCl2SO
2
NO NO2HCN HCL
FERANT TRATION filtered filtered
(PPM)
Acetone 1000 0 0 0 0 ND 0 ND ND
Acetylene 12000 0 0 0 0 ND ND ND ND
Ammonia 25 0 0 0 0 ND 0 0 0
Arsine 1 0 0 0 0 0 ND ND 1
Benzene 20 0 0 0 0 ND ND 0 ND
Bromine 2 0 0 2.5 ND 0 0 0 ND
Carbon Dioxide 5000 0 0 0 0 0 0 0 0
Carbon Disulfide 15 0 0 0 0 0 ND 0.1 0
Carbon Monoxide 100 100 0.3 0 0.2 ND 0 0 0
Chlorine 5 0 -3 5 0 0 0 -0.2 0
Diborane 20 0 0 0 0 ND ND ND 0
Ethylene 50 100 0.1 0 0 ND 0 -0.3 ND
Ethyl Alcohol 100 115 0 0 0 ND ND 0 ND
Ethylene Oxide 10 ND ND ND 0 ND ND ND ND
Ether 400 3 0 0 0 ND 0 ND ND
Fluorine 5 0 0 2.5 0 0 ND 0 0
Freon 12 1000 0 0 0 0 0 0 0 0
Germane 1 0 0 0 0 0 ND ND 1
Hexane 500 0 0 0 0 ND 0 0 ND
Hydrogen 500 200 0.5 0 15 ND -10 0 0
Hydrogen Chloride 50 0 0 0 0 4 0 ND 50
Hydrogen Cyanide 10 0 0 0 0 0 0 10 0
Hydrogen Fluoride 10 0 0 0 0 ND ND ND 6.5
Hydrogen Sulfide 10 1 10 -0.1 0 1 -8 50 40
MEK 200 0 0 0 0 0 0 ND ND
Mercaptan (Methyl) 5 0 4.5 -0.1 0 1 ND 6 ND
Methane 5000 0 0 0 0 0 0 0 0
Nitric Oxide 100 0 2 0 2 100 ND -3 40
Nitrogen Dioxide 5 -1 -4 0.5 -5 1.5 5 ND 0
Phosphine 0.5 ND 0 0 ND 0 ND ND 2
Silane 5 0 0 0 0 0 ND ND 7
Sulfur Dioxide 10 0 0.3 0 10 0.5 ND -0.3 0
Tichloro­ethylene 1000 0 0 0 0 0 ND ND ND
Page 61
3-6
ND = No Data
INTER- CONCEN- ClO2HF PH
3
ASH4SiH4GeH3B2H6Br
2
FERANT TRATION
(PPM)
Acetone 1000 0 0 ND ND ND ND ND 0
Acetylene 12000 0 0.1 ND ND ND ND ND 0
Ammonia 25 0 0 ND ND ND ND ND 0
Arsine 1 0 ND 0.7 1 1 1 5 0
Benzene 20 0 ND ND ND ND ND ND 0
Bromine 2 1 ND ND ND ND ND ND 2
Carbon Dioxide 5000 0 0 ND ND ND ND ND 0
Carbon Disulfide 15 0 ND 0 0 0 0 0 0
Carbon Monoxide 100 0 0 0 1 0 0 0 0
Chlorine 5 2.5 5 ND ND ND ND ND 4
Diborane 20 0 ND 3.5 5 4 5 20 0
Ethylene 50 0 0 0.5 1 1 1 2 0
Ethyl Alcohol 100 0 0 ND ND ND ND ND 0
Ethylene Oxide 10 0 ND ND ND ND ND ND ND
Ether 400 0 0 ND ND ND ND ND 0
Fluorine 5 1 ND ND ND ND ND ND 2
Freon 12 1000 0 0 0 0 0 0 0 0
Germane 1 0 ND 0.7 1 1 1 5 0
Hexane 500 0 0 ND ND ND ND ND 0
Hydrogen 500 0 0 0 0 0 0 0 0
Hydrogen Chloride 50 0 30 ND ND ND ND ND 0
Hydrogen Cyanide 10 0 0 ND ND ND ND ND 0
Hydrogen Fluoride 10 0 10 ND ND ND ND ND 0
Hydrogen Sulfide 10 0 3 ND ND ND ND ND 0
MEK 200 0 ND ND ND ND ND ND 0
Mercaptan (Methyl) 5 0 ND ND ND ND ND ND 0
Methane 5000 0 0 ND ND ND ND ND 0
Nitric Oxide 100 0 2 ND ND ND ND ND 0
Nitrogen Dioxide 5 0.2 2.5 ND ND ND 0.5 ND 0.4
Phosphine 0.5 0 0 0.5 1 0.7 1 3 0
Silane 5 0 ND 0.1 0.2 5 0.2 15 0
Sulfur Dioxide 10 0 8 ND 1 2 3 6 0
Tichloro­ethylene 1000 0 0 ND ND ND ND ND 0
Page 62
ND = No Data
INTER- CONCEN- F
2
NH3H
2
EtO
FERANT TRATION
(PPM)
Acetone 1000 0 ND ND ND
Acetylene 12000 0 ND ND ND
Ammonia 25 0 25 ND 0
Arsine 1 0 ND ND ND
Benzene 20 0 ND ND ND
Bromine 2 12 ND ND ND
Carbon Dioxide 5000 0 0 0 ND
Carbon Disulfide 15 0 ND ND ND
Carbon Monoxide 100 0 0 2 ND
Chlorine 5 10 0 0 0
Diborane 20 0 ND ND ND
Ethylene 50 0 0 40 ND
Ethyl Alcohol 100 0 ND ND 10
Ethylene Oxide 10 ND ND ND 10
Ether 400 0 ND ND ND
Fluorine 5 5 ND ND ND
Freon 12 1000 0 0 0 0
Germane 1 0 ND ND ND
Hexane 500 0 ND ND ND
Hydrogen 500 0 ND 500 ND
Hydrogen Chloride 50 0 0 0 ND
Hydrogen Cyanide 10 0 0 3 0
Hydrogen Fluoride 10 0 ND ND ND
Hydrogen Sulfide 10 -0.2 0.5 1 ND
MEK 200 0 0 ND 3
Mercaptan (Methyl) 5 -0.2 ND ND ND
Methane 5000 0 ND ND ND
Nitric Oxide 100 0 0 3 ND
Nitrogen Dioxide 5 1 ND ND 0
Phosphine 0.5 0 0 0 0
Silane 5 0 ND ND ND
Sulfur Dioxide 10 0 0 0 ND
Tichloro­ethylene 1000 0 ND ND ND
3-7
Page 63
4-1
Chapter 4, Maintenance
General
The Ultima X Gas Monitor is constantly performing a self-check. When a problem is found, it displays the appropriate error message (TABLE 4-3, "Troubleshooting Guidelines"). When a critical error is detected within the unit, the 4-20 mA output signal goes to a fault condition of 3.0 mA
The "Sensor Warning" indication is not an error and does not affect the output. TABLES 4-1 and 4-2 describe the messages that users may see.
Table 4-1. Operational Display Messages
MESSAGE INDICATES
MM/DD/YY Format for date scrolling
VER Software version level will display next
TIME Time will display next
DATE Date will display next
MIN MIN value for this interval will display next
MAX MAX value for this interval will display next
AVG AVG value for this interval will display next
Adr Instrument's address will display next
End End of calibration cycle
Err An Error code will display next
HR Special case indicates hours (two characters or less)
OVER Gas value is greater than the set range
Page 64
4-2
Table 4-2. Configuration Display Messages
MESSAGE INDICATES
CAL SIG ON Instrument will output the calibration signal during calibration
CAL SIG OFF Instrument will output gas value during calibration
LTCH/ Latching relay operations
UNLTCH/ Non-latching relay operations
INCR/ Increasing Alarm relay operations
DECR/ Decreasing Alarm relay operations
ENER Energized relay operations
DENER De-energized relay operations
CAL Normal calibration or 4-20 calibration cycle
iCAL Initial calibration cycle
OFF Alarm is OFF
ON Alarm is ON
RNGE Instrument's operational full-scale will display next
PCAL Instrument's previous calibration date will display next
TBLE Instrument gas table selection (if applicable)
ALERT OP ON Instrument output will follow ALERT mode
ALERT OP OFF Instrument output will not follow ALERT mode
SWAP DELAY ON 60-second delay after sensor missing before fault
SWAP DELAY OFF Fault occurs at sensor missing condition
Table 4-3. Troubleshooting Guidelines (In priority order)
MESSAGE INDICATES ACTION
MN FLASH Program memory on the Replace main pc board FAULT main PCBA is invalid
MN RAM Defective RAM memory location Replace main pc board FAULT was found on the main PCBA
MN EEPROM EEPROM on the main PCBA Replace main pc board FAULT is invalid
SENSOR Instrument has lost communica- Connect or replace sensor MISSING tion with the sensor module
SNSR FLASH Sensor module program memory Replace sensor module FAULT is invalid
SNSR RAM Sensor module has a defective Replace sensor module FAULT RAM location
SNSR DATA Sensor module datasheet is Send reset data sheet command FAULT invalid from the controller; if error persists,
replace sensor
INVALID Attached sensor module is not Replace with correct sensor type SENSOR compatible with main instrument
Page 65
4-3
MN SUPPLY Power supply on main PCBA Check sensor wiring FAULT is out of range or replace main pc board
RELAY Error with the internal relays Cycle power to the unit FAULT has occurred or replace main pc board
SNSR POWER Power at the sensor module Correct wiring error, replace main FAULT is out of range pc board, or replace sensor module
IR SOURCE IR source failure Replace or consult factory FAULT
FIXED 4-20 mA is at a set level Exit Fixed Current Mode CURRENT and will not change when using the HART MODE Controller gas is applied or
under fault conditions
- SUPPLY The negative supply sensor Check wiring or
FAULT module is out of range replace sensor module
REF SIG IR reference detector failure Replace or consult factory FAULT
ANA SIG IR analytical detector failure Replace or consult factory FAULT
LOW SIGNAL Low IR signal Clean optics or replace
sensor module. If in cleaning mode, no action required
PARAM FAULT An operational parameter is Restart; replace, if necessary
out of range or sensor failed internal check
CONFIG Main EEPROM memory Use Controller to reset all RESET was reset configurations (e.g., alarm levels,
calibration signals ON or OFF, etc)
CHANGE Sensor is at its end of life Replace sensor SENSOR
ZERO CAL Instrument did not Repeat calibration; check for proper FAULT calibrate successfully calibration gas; check for blockage OR SPAN in the flow system CAL FAULT
SENSOR Sensor approaching end of life Prepare to replace sensor module WARNING
CHECK CAL Calibration should be verified Perform bump test or calibration
+LOC Instrument is locked in Recalibrate or reset sensor
over-range condition
OVER % LEL Sensor is exposed to a The instrument will return to normal
gas concentration above the LEL operation when the gas concentra-
tion drops below 100% LEL
und Under-range condition - quick Recalibrate or replace sensor
Und Under-range condition - slow Recalibrate or replace sensor
The highest priority message is displayed first. Lower priority messages are output only after the highest priority message is cleared. A manual selected scanning of low-priority messages is not possible.
Page 66
Ultima XIR Cleaning Procedure
The presence of particulate matter, oil films, liquid water, or the residue from water drops on the two monitor windows can adversely affect its performance. The environmental guard is designed to prevent foreign solids or liquids from reaching the monitor's optical system. Additionally, heating elements are incorporated into the unit to prevent water condensation. Under severe conditions, however, some material may collect on these surfaces and it may be necessary to occasionally check and clean the windows.
1. Remove the environmental or flow cap.
2. Place an opaque object (piece of paper, end of wrench handle, etc.) between the light source window and the mirror to completely obscure the light path for two to three seconds.
The Ultima XIR/Ultima XI Monitor enters the Cleaning Mode for two minutes.
NOTE: While in the Cleaning Mode, the sensor will not
respond to the presence of gas.
The analog current output is 3.0 mA during this time.
The display indicates 'low signal'’.
3. While both windows are made of a highly durable material that is not easily scratched, avoid excessive pressure when cleaning them. Clean, cotton-tipped applicators are the most convenient tool to remove material collected on the windows.
Use a dry applicator or one moistened with distilled water to wipe the window and remove dust.
Use an additional clean, dry applicator to remove any residual water.
Use an applicator moistened with isopropyl alcohol to remove heavy deposits of solids, liquids or oil films. Clean the window again with a second applicator moistened with distilled water; then, dry the window with a final applicator.
Avoid using excessive amounts of water or alcohol in the cleaning procedure, and inspect the window to ensure that the entire surface is clean.
The unit remains in the Cleaning Mode for a minimum of two minutes. If active cleaning is still in progress at the end of this period, the sensor detects the motion of this object in its light path and automatically extends the Cleaning
4-4
Page 67
4-5
Mode for 15 seconds. Further 15-second Cleaning Mode extensions continue until no motion is detected.
NOTE: When the cleaning process is complete, be sure to
remove all objects from the light path.
4. When exiting the Cleaning Mode, the unit returns to normal operation. If water or isopropyl alcohol was used, allow the unit to operate for 15 minutes to completely dry before replacing the environmental guard and continuing to monitor for combustible gas.
5. Replace the environmental or flow cap.
6. After cleaning the windows, it is advisable to check the sensor’s response to both zero and calibration gas.
Do not place foreign objects in the sensor's analytical region (except per the "Ultima XIR Cleaning Procedure" as described above); otherwise, the infrared beam can be par­tially blocked, causing the sensor to generate false read­ings. All objects must be removed from the sensor's analyt­ical region for it to function properly. Similarly, if water or isopropyl alcohol is used to clean the sensor's windows, any residue from the cleaning procedure must be complete­ly dissipated before returning the unit to service. Checking the sensor's response to zero gas is the best way to purge residual cleaning materials from the sensor and to make sure that sensor's reading is stable before zeroing or cali­brating the sensor (see Chapter 2, "Start-up and Calibration").
Replacing an Ultima XE or Ultima XA Sensor
The only routine maintenance item is the sensing element itself, which has a limited lifetime. When the Ultima X Series Gas Monitor's read-out indicates that the sensor must be changed, there is very little sensor lifetime remaining. It is good practice to obtain a replacement sensing element before the sensing element within your unit becomes inoperative. Typically, the Ultima X Series Monitor LCD display shows a maintenance message when the sensor is due for replacement (FIGURE 4-1).
"
CAUTION
Page 68
4-6
Electrochemical sensors are sealed units which contain a corrosive electrolyte. If electrolyte is leaking from the sen­sor, exercise CAUTION to ensure the electrolyte does not contact skin, eyes or clothing, thus avoiding burns. If con­tact occurs, rinse the area immediately with a large quantity of water. In case of contact with eyes, immediately flush eyes with plenty of water for at least 15 minutes. Call a physician.
Do not install a leaking sensor in the sensing head assem­bly. The leaking sensor must be disposed of in accordance with local, state and federal laws. To obtain a replacement sensor, contact MSA at the address given under "Obtaining Replacement Parts."
1. There is no need to open the main enclosure; simply unscrew the sensor assembly located on the bottom of the Ultima X Series Gas Monitor main assembly (FIGURE 4-2).
For Ultima XE sensors marked Class I, Groups A, B, C and D and not used in Class II areas, unscrew sensor cap at least three full turns (but no more than four full turns from its tightly-closed position), wait 10 seconds, and then remove cap completely. Failure to follow this warning can result in the ignition of a hazardous atmosphere.
"
WARNING
"
CAUTION
"
WARNING
Figure 4-1. "Change Sensor" Scrolls Across the Display
Page 69
4-7
For the Ultima XE Sensor marked Class II Groups F and G, atmosphere must be free of dust and the power removed from the unit before the sensor cap can be removed from the housing. Failure to follow this warning can result in the igni­tion of a hazardous atmosphere.
2. Identify the sensor assembly needed via the A-ULTX-SENS code on the interior sensor label and obtain the appropriate sensor assembly; replace sensor assembly.
NOTE: Alarm setpoints and relay functions (energized/de-
energized, latching/unlatching, and upscale/downscale) will not change when changing a sensor module from its current gas type to the same gas type (e.g., carbon monoxide to carbon monoxide). Alarm setpoints and the upscale/downscale relay function will change to the new sensor's default settings when changing a sensor module from its current gas type to a different gas type (e.g., carbon monoxide to oxygen).
3. The Ultima X Series Gas Monitor is shipped with the Sensor Swap Delay enabled. This means that the 4-20 mA output signal and the FAULT relay will hold off a fault indication for 60 seconds after the sensor missing indication is displayed on the instrument. This setting allows the operator to exchange sensor modules without a FAULT indication.
4. Refer to Chapter 2, "Calibration" to calibrate with the new sensor.
It is recommended that all other maintenance be performed at an MSA factory-authorized service center.
Figure 4-2. Sensor Assembly and Sensor Guard
for General-Purpose Model
Page 70
4-8
Obtaining Replacement Parts
See TABLE 4-4 for replacement sensor kits. To obtain a replacement sensor, address the order or inquiry to:
Mine Safety Appliances Company 1000 Cranberry Woods Drive Cranberry Township, PA 16066
or call, toll-free, 1-800-MSA-INST. Inquiries can also be e-mailed to
customer.service@msanet.com.
Use only genuine MSA replacement parts when performing any maintenance procedures provided in this manual. Failure to do so may seriously impair sensor and gas moni­toring performance. Repair or alteration of the Ultima X Series Gas Monitor, beyond the scope of these maintenance instructions or by anyone other than authorized MSA serv­ice personnel, could cause the product to fail to perform as designed and persons who rely on this product for their safety could sustain serious personal injury or loss of life.
Table 4-4. Replacement Parts
GAS SELECTION SENSOR KIT PART NO.
GENERAL- EXPLOSION- GENERAL­PURPOSE PROOF PURPOSE PLASTIC MODEL E STAINLESS MODEL A MODEL G
Carbon Monoxide, 100 ppm A-ULTX-SENS-11-0... A-ULTX-SENS-11-1... A-ULTX-SENS-11-6...
Carbon Monoxide, 500 ppm A-ULTX-SENS-12-0... A-ULTX-SENS-12-1... A-ULTX-SENS-12-6...
Oxygen, 10% - compensated A-ULTX-SENS-13-0... A-ULTX-SENS-13-1... A-ULTX-SENS-13-6...
Oxygen, 25% - compensated A-ULTX-SENS-14-0... A-ULTX-SENS-14-1... A-ULTX-SENS-14-6...
Hydrogen Sulfide, 10 ppm A-ULTX-SENS-15-0... A-ULTX-SENS-15-1... A-ULTX-SENS-15-6...
Hydrogen Sulfide, 50 ppm A-ULTX-SENS-16-0... A-ULTX-SENS-16-1... A-ULTX-SENS-16-6...
Hydrogen Sulfide, 100 ppm A-ULTX-SENS-17-0... A-ULTX-SENS-17-1... A-ULTX-SENS-17-6...
Chlorine, 5 ppm A-ULTX-SENS-18-0 not available A-ULTX-SENS-18-6...
Sulfur Dioxide, 25 ppm A-ULTX-SENS-70-0... A-ULTX-SENS-70-1... A-ULTX-SENS-70-6...
Nitric Oxide, 100 ppm A-ULTX-SENS-20-0... A-ULTX-SENS-20-1... A-ULTX-SENS-20-6...
Nitrogen Dioxide, 10 ppm A-ULTX-SENS-72-0... A-ULTX-SENS-72-1... A-ULTX-SENS-72-6...
Hydrogen Cyanide, 50 ppm A-ULTX-SENS-22-0... A-ULTX-SENS-22-1... A-ULTX-SENS-22-6...
Sulfur Dioxide, 100 ppm A-ULTX-SENS-71-0... A-ULTX-SENS-71-1... A-ULTX-SENS-71-6...
"
WARNING
Page 71
4-9
GAS SELECTION SENSOR KIT PART NO.
GENERAL- EXPLOSION- GENERAL­PURPOSE PROOF PURPOSE PLASTIC MODEL E STAINLESS MODEL A MODEL G
Hydrogen Chloride, 50 ppm A-ULTX-SENS-23-0... not available A-ULTX-SENS-23-6...
Chlorine Dioxide, 3 ppm A-ULTX-SENS-24-0... not available A-ULTX-SENS-24-6...
Combustible Gas, 100% LEL A-ULTX-SENS-31-0... A-ULTX-SENS-31-1... A-ULTX-SENS-31-6... Natural Gas and H2, 5% CH
4
Combustible Gas, 100% LEL A-ULTX-SENS-32-0... A-ULTX-SENS-32-1... A-ULTX-SENS-32-6... Petroleum Vapors,
2.1% Propane
Combustible Gas, 100% LEL A-ULTX-SENS-33-0... A-ULTX-SENS-33-1... A-ULTX-SENS-33-6... Solvents, 2.1% Propane
Comb Gas IR - Methane, not available A-ULTX-SENS-38-1... not available 5% CH
4
Comb Gas IR - Non Methane, not available A-ULTX-SENS-39-1... not available
2.1% Propane
Phosphine, 2 ppm A-ULTX-SENS-41-0... A-ULTX-SENS-41-1... A-ULTX-SENS-41-6...
Arsine, 2 ppm A-ULTX-SENS-42-0... A-ULTX-SENS-42-1... A-ULTX-SENS-42-6...
Silane, 25 ppm A-ULTX-SENS-43-0... A-ULTX-SENS-43-1... A-ULTX-SENS-43-6...
Germane, 3 ppm A-ULTX-SENS-44-0... A-ULTX-SENS-44-1... A-ULTX-SENS-44-6...
Diborane, 50 ppm A-ULTX-SENS-45-0... not available A-ULTX-SENS-45-6...
Bromine, 5 ppm A-ULTX-SENS-46-0... not available A-ULTX-SENS-46-6...
Fluorine, 5 ppm A-ULTX-SENS-47-0... not available A-ULTX-SENS-47-6...
Ammonia, 100 ppm A-ULTX-SENS-48-0... not available A-ULTX-SENS-48-6...
Hydrogen, 1000 ppm A-ULTX-SENS-49-0... A-ULTX-SENS-49-1... A-ULTX-SENS-49-6...
ETO, 10 ppm A-ULTX-SENS-50-0... not available A-ULTX-SENS-50-6...
Ammonia, 0-1000 PPM A-ULTX-SENS-54-0... not available A-ULTX-SENS-54-6...
Oxygen-Solvent Tolerant, not available A-ULTX-SENS-55-1... not available 0-25%
Carbon Monoxide, A-ULTX-SENS-57-0... A-ULTX-SENS-57-1... A-ULTX-SENS-57-6... 0-1000 PPM
Chlorine, 0-20 PPM A-ULTX-SENS-61-0... not available A-ULTX-SENS-61-6...
Oxygen-Solv & CO2Tolerant, not available A-ULTX-SENS-62-1... not available 0-25%
Oxygen-Low, 0-25% not available A-ULTX-SENS-63-1... not available
Oxygen-Low, Solv Tolerant, not available A-ULTX-SENS-64-1... not available 0-25%
SENSOR REPLACEMENT PARTS
PART PART NO.
XE Sensor Guard 10028904
XIR Sensor Guard 10041265
XIR Flow Cap 10042600
Page 72
A-1
Appendix A, Optional Features
1) Internal Relays
General Information
The internal relays are designed to enable Ultima X Series Gas Monitors to control other equipment. There are four relays within the Ultima X Series Gas Monitor's module:
three alarm relays
one fault relay.
Once configured, the relays activate when the Ultima X Gas Monitor detects an alarm condition. Similarly, the fault relay de-energizes when a fault condition is detected.
The alarm relays are enabled in the non-latching, de-energized mode at the factory.
To disable or configure the alarms, you need the Ultima Controller (P/N 809086) or a HART Communicator.
The fault relay is normally-energized so the relay de-activates into a fail-safe condition if a fault or power outage occurs. See "Fault Relay" later in this Appendix.
To prevent false alarms in the following instances, alarms/relays are temporarily disabled:
1) During the first minute from power-up
2) During calibration
3) For two minutes after calibration.
4) For one minute after the sensor missing indication displays (if the Delay feature is enabled).
Unpacking, Mounting and Wiring
Unpack, mount and wire the Ultima X Series Gas Monitor according to Chapter 1, "Installation". All electrical connections to the Ultima X Series Gas Monitor can be made via the clearly marked board-mounted connections.
"
CAUTION
Page 73
A-2
NOTE: To avoid electrical noise problems, do not run AC lines from
relays in the same conduit or cable tray as the DC Signal lines.
See Ultima Controller and Calibrator Manual (P/N 813379) for complete relay configuration information.
Power cable wiring lengths for the Ultima X Series Gas Monitor with internal relays differ from models without internal relays (TABLE A-1).
Table A-1. Cable Length and Wire Size for Units With Internal Relays
GAS SENSOR DC WIRE MAXIMUM MAXIMUM TYPE OUTPUT VOLTAGE SIZE CABLE LENGTH LOAD
SUPPLY (AWG) WITHOUT HART RESISTANCE
(FEET) (METERS) (OHMS)
Oxygen or Toxic 3 Wire 12 VDC 16 2,500 762 250
24 VDC 16 8,000 2,438 500
Combustible 3 Wire 12 VDC 16 900 274 250
24 VDC 16 3,000 914 500
XIR 3 wire 12 VDC 16 400 152 250
24 VDC 16 2,500 762 500
In all installations, twisted instrument-quality cable is recommended.
Shielded cable is recommended in situations where radio frequency interference (RFI), electro-magnetic interference (EMI) or other electrical noise sources exist or are anticipated.
Ultima X Series Gas Monitor Internal Relays
Relay Specifications
Table A-2. Relay Specifications
TEMPERATURE -40 to +60°C (-40 to +140°F) RANGE
HUMIDITY 15 to 95% RH, non-condensing
RELAYS 3 ALARMS SPDT FAULT (Single pole, double throw)
(NORMALLY-ENERGIZED) SPDT (Single pole, double throw)
RELAY At 125 or 250 VOLTS AC, 5.0 Amps or 5 Amps @ 1/10 Hp RATINGS NON-INDUCTIVE
At 30 Volts DC, 5.0 Amps or 5 Amps @ 1/10 Hp NON-INDUCTIVE
Page 74
A-3
Alarm Relays
There are three alarm relays and one fault relay in the Ultima X Series Gas Monitors. The three alarm relays:
Activate when the Monitor detects a gas concentration level that exceeds setpoints
Alarms 1, 2 and 3 generally default to 10%, 20% and 30% of the full-scale reading and are set when the gas reading is above these values.
The Oxygen Model is a special case where:
Alarm 1 is set to 19% oxygen and activates below this setpoint
Alarm 2 is set to 18% oxygen and activates below this setpoint
Alarm 3 is set to 22% oxygen and activates above this setpoint.
These default setpoints can be changed or verified via the Ultima X Controller.
See Ultima/Ultima X Controller and Calibrator manual (P/N813379).
The Controller can also enable the latching alarm function.
Are factory-set to a de-energized position.
All relay connections have a normally-open set of contacts and a normally-closed set of contacts. These contacts are labeled as NCD (normally-closed, de-energized) or NCE (normally­closed, energized).
The units are shipped with alarm relays factory-set to the de­energized (non-alarm) position and the trouble relay set to the energized (non-fault) position.
Upon activation, the relay contacts change state and remain changed for as long as:
The alarm condition exists within the Ultima X Series Gas Monitor or
The latching mode is selected (see "Note on Resetting Latched alarms with Controller or Calibrator" in Chapter 2).
These defaults can be changed or verified via the Ultima X Controller.
Page 75
A-4
Fault Relay or Trouble
It is a normally-energized, single-pole, double-throw (SPDT) relay.
During normal operation, the relay contacts are normally closed (NC) and normally open (NO) as shown in FIGURE A-1.
NOTE: FIGURE A-1 depicts the version of the printed circuit board
assembly without HART components. FIGURE 1-17 is the equivalent HART version.
When a fault is detected or power is cut or turned OFF, these contacts change as follows:
normally-closed contacts open
normally-open contacts close.
Provides an electrical path for fail-safe relay operation. In the event of any failure, including power loss, the relay will change to a fault condition.
The Fault Relay can remain STEADY ON or PULSED. These two different modes can communicate different information to any PLC or DCS connected to the fault relay:
Fault Relay STEADY ON indicates:
Ultima X Series sensor is not connected properly or
Ultima X Series Gas Monitor internal fault or
An inoperative relay.
Figure A-1. Relay Contacts
Page 76
Fault Relay PULSED (once per minute) indicates:
Improper calibration of the Ultima X Series Gas Monitor or
Ultima X Series Gas Monitor CHECK CAL or CAL FAULT displayed.
Relay Connections
All electrical connections to internal relays can be made directly on the pc board (see FIGURE A-1).
If you are connecting the relays to motors, fluorescent lighting or other inductive loads, it is necessary to suppress any sparks or inductive feedback that may occur at the relay contact. These effects may render the unit inoperative. One way to reduce these effects is to install a *Quencharc®across the load being switched. This device is available from MSA as (P/N 630413).
Before wiring the Ultima X Series Gas Monitors, disconnect power source supplying the monitor; otherwise, electrical shock or ignition of hazardous gases could occur.
The Ultima X Series Gas Monitor must be disassembled for relay wiring. The following procedure must be performed:
1. Remove the Ultima X Series Gas Monitor cover.
2. Pull on the wiring plugs to disconnect the connectors on the exposed board.
NOTE: Observe connector locations for later re-insertion.
3. Route customer-supplied cable into the enclosure and connect to the appropriate wiring plugs.
4. Identify each conductor of the cable to enable proper connection at the control equipment.
5. If installing a RESET push-button:
a. Route a two-conductor cable to terminal block J10
(FIGURE A-2).
"
WARNING
A-5
Page 77
Route this cable with DC power to avoid noise interference from relay wiring.
b. Connect the two-conductor cable to the two positions of
terminal block J10.
c. Identify the cable to enable proper connection at the
button.
d. Route cable to the push-button location; wire the button.
6. Re-install the wiring plugs.
Ensure that wiring plugs are firmly seated into their mates.
7. Pull the cable away from the unit to relieve any excess slack.
It is important not to have excess wire or cable within the module to avoid unwanted AC noise.
8. Re-install the cover of the Ultima X Series Gas Monitor.
Figure A-2. Relay Printed Circuit Board
A-6
Page 78
2) Optional RESET Push-button
If you are going to specify a switch to use with the Ultima X series, it should have, as a minimum, an equivalent (or better) approvals classification. The MSA provided pushbutton (P/N 10046923) is explosion-proof for Class I, Groups B, C, and D only. When relied upon for an explosion-proof capacity, do not install on equipment mounted or intended to be mounted in any other hazardous location.
This pushbutton has been approved for use in Group A with the Ultima X Series when the protection technique is Nonincendive and the hazardous location is Division 2. Check product labels for protection technique.
When the pushbutton is used in conjunction with the Ultima XE Gas Monitor, the highest classification rating for the sys­tem is reduced to Class I, Div. 1, Groups B, C, & D. See MSA P/N 10048833 for installation instructions. Misuse can result in ignition of hazardous gases.
General
A RESET button is an optional feature to allow latching relays to be reset at the sensor location.
This may silence any alarm horns or turn OFF any equipment connected to the relays.
Latching relays can be configured on the Ultima X Series Gas Monitor via the Ultima Controller.
In a latching configuration: when the RESET button is pushed and any alarm is latched and not in its active alarm state, the alarm will reset.
NOTE: An IR command can mimic the RESET button per
Controller and Calibrator manual (P/N 813379).
In a non-latching configuration: the RESET button has no affect on the alarms.
RESET Button Selection
The RESET push-button can be acquired locally and wired to the Ultima X Series Gas Monitor during unit installation.
"
WARNING
A-7
Page 79
The RESET push-button must be a normally-open type with a momentary contact when pushed.
The electrical ratings must be at least 1 amp at 250 volts AC.
Specific push-buttons may be sourced by manufacturers such as Appleton Electric and Crouse Hindes, Inc.
Optional Push-button Calibration
The following procedure is used to enter the calibration by using the push-button.
1. Press and hold the push-button until the heart is displayed.
2. Release the push-button.
At this time, any recoverable alarms will be acknowledged.
3. Press and hold the push-button within three seconds of the push-button release.
4. Release the push-button when the desired calibration is displayed. See TABLE A-3.
Table A-3. Push-button Calibration
CALIBRATION DISPLAYED PUSH-BUTTON TYPE DATA HOLD TIME
Zero Calibration CAL ZERO 5 seconds
Span Calibration CAL SPAN 10 seconds
ICAL iCAL 20 seconds
Refer to Chapter 2, "Startup and Calibration" for more information on calibration.
5. The calibration can be aborted during the 30-second countdown by pressing the push-button until the is displayed.
When the push-button is released, the calibration will be aborted.
A-8
Page 80
3) Optional Horn Relay Software
The Ultima X Series Gas Monitor is available with optional Horn Relay Software to allow an audible horn to be used with Relay 1. The following describes the use and functionality of this optional feature.
Relay 1 is considered the Horn Relay. It does not function directly with Alarm 1 as in the standard software. To configure, note that:
• Alarm 1 function is still active on the display
• Alarm 1 display has its own latching/non-latching configuration setting
• Horn Relay is configured as normally energized/de-energized via the Alarm 1 configuration setting; this is the only Alarm 1 configuration setting that uniquely controls the Horn Relay.
• Alarm/Relay 2 and Alarm/Relay 3 action remains unchanged.
NOTE: All relays, including Horn Relay hardware, have NO (normally
open) and NC (normally closed) terminals.
To Activate the Horn Relay
The Horn Relay is initially active when any alarm (1, 2, or 3) condition is active. Once Horn Relay is reset, it is set again if the gas level:
• clears below or above the active setpoint and then moves back through the setpoint or
• continues to move through the next alarm set point.
To Reset the Horn Relay
The Horn Relay is reset by momentary contact closure using the pushbutton or by IR communications, regardless of whether or not the alarm has cleared.
• The Horn Relay automatically resets if all alarms are clear (such as when all alarms are unlatched and fall below or rise above their setpoints for negative acting alarms).
• If any alarm remains latched after the gas value moves out of the alarm range, the Horn Relay remains active until reset by the user.
Clearing latched alarms is done by using the optional pushbutton or the IR Controller.
A-9
Page 81
B-1
Appendix B, Calibration Guide for Additional XIR/XI Gases
Page 82
B-2
Page 83
C-1
Appendix C, General Certification Information
Page 84
C-2
Page 85
Appendix D, HART Specific Information
HART Field Device Specification
The Ultima X Series Gas Monitor is available with an optional HART (Highway Addressable Remote Transducer) output communications protocol. With this option, the Ultima X Series Gas Monitor complies with HART Protocol Revision 7 and uses the 16-bit manufacturer and device codes. This document specifies all the device specific features and documents HART Protocol implementation details (e.g., the Engineering Unit Codes supported). These specifications assume the reader is somewhat familiar with HART Protocol requirements and terminology.
This specification is a technical reference for HART-capable HOST Application Developers, System Integrators and knowledgeable End Users. It also provides functional specifications (e.g., commands, enumerations and performance requirements) used during Field Device deployment, maintenance, testing, and operations. It is recommended that the 4-20 mA output be the primary gas monitoring signal. The HART signal can be the secondary method.
NOTE: The two-wire units’ HART protocol does not comply fully
with Immunity Standards EN61000-4-3 (2006) and EN61000-4-6 (2007).
Table D-1. Device Identification
MANUFACTURER NAME MSA MODEL NAME(S) ULTIMA
Manufacture ID Code 0x6008 Device Type Code 0xE09F
HART Protocol Revision 7.0 Device Revision 1
Number of Device Variables 1 Notes:
Physical Layers Supported FSK,
4-20 mA
Physical Device Category Current Output
D-1
Page 86
Host Interface
Analog Output
The three-wire 4-20 mA current loop is connected on terminals marked 8-30 VDC(1), 4-20 mA OUT(2), and GND (3-wire)(3). The two-wire 4-20 mA current loop is connected on the 8-30 VDC(1) and 4-20 mA OUT(2) terminals. Refer to the installation outline drawings shown in Chapter 1, TABLE 1-1 for details.
This is the main output from this transmitter, representing the process gas measurement, linearized and scaled according to the configured instrument range. This output corresponds to the Primary Variable (PV). HART communications are supported on this loop. This device has a Capacitance Number (CN) of 1.
An inoperative device can be indicated by down-scale or up-scale current, depending on the sensor type. Current values are shown in TABLE D-2.
Table D-2. Current Values
DIRECTION VALUES VALUES
(% OF RANGE) (MA OR V)
Linear over-range Down 0% 4.00 mA
Up +105.0% +1.0% 20.64 to 20.96 mA
Device malfunction indication Down: less than 3.5 mA
Up: greater than 20.96 mA
Maximum current 22.0 mA
Multi-drop Current draw 3.5 mA
Lift-off voltage, 3-wire PCBA 8 VDC
Lift-off voltage, 2-wire PCBA 13 VDC @ 250 Ohms
D-2
Page 87
Table D-3. Device Variables Exposed by the Ultima Monitor
VARIABLE DESCRIPTION VARIABLE DESCRIPTION
Gas Type Sensor gas type description Last Cal Date Date sensor was
last calibrated
Alarm Gas value at which an Auto Zero comp Amount of compensated Setpoints alarm status bit is set below zero drift
Alarm Action increasing or decreasing Alert option status See next section
alarm type, latching or non latching
Alarm Status Indication of alarm setpoint Swap Delay status See next section
exceeded
Input Voltage Device input voltage level
Min/Max/Avg Minimum, maximum and Sensor Temp
avg. value of PV over time
Avg Interval Time interval for min, max, Sensor Status Status returned by
avg (1,8 or 24 hr) sensor
Gas Table Linerization table selection Relay EN/De-eng Relay EN/De-eng
RTC Date Device real time clock date Cal Sig Cal Signal Status
RTC Min Device real time clock minutes
RTC Hrs Device real time clock hours
Table D-4. Dynamic Variable implemented by Ultima Monitor
MEANING UNITS
PV Gas Value %, %LEL, PPM
Status Information
Device Status
Bit 4 ("More Status Available") is set when any failure is detected. Command #48 gives further details.
Extended Device Status
The Ultima Monitor can predict when certain maintenance will be required. This bit is set if a sensor fault or maintenance warning is detected. "Device Variable Alert" is set if the PV is out of limit.
D-3
Page 88
Table D-5. Additional Device Status (Command #48)
Command #48 returns 5 bytes of data, with the following status information: BYTE BIT MEANING CLASS DEVICE
STATUS BITS SET
0 0 Configuration Reset Error 4,7
1 Main ram fault Error 4,7
2 Main flash fault Error 4,7
3 EEprom write error Error 4,7
4 Incompatible sensor Error 4,7
5 Sensor quick under range Error 4,7
6 Sensor UNDer range Error 4,7
7 Calibration fault Error 4,7
1 0 Sensor Missing Error 4,7
1 Sensor Overrange Warning
2 Overrange Lock Warning
3 Parameter Fault Error 4,7
4 Sensor Warm up Warning
5 Sensor Config Reset Warning
6 Sensor Power Fault Error
7 5V Power Fault Error
2 0 Zero Countdown Info
1 Apply Zero Gas Info
2 Span Countdown Info
3 Apply Span Gas Info
4 Cal Aborted Info
5 Zero Fault Info
6 Span Fault Info
7 Cal OK Info
D-4
Page 89
BYTE BIT MEANING CLASS DEVICE
STATUS BITS SET
3 0 End of Life Warning Warning 4,7
1 Sensor Swap Delay Info
2 Change Sensor Fault Error
3 Sensor Power Fault Error 4,7
4 Internal Comm Fault Error
5 Cal Sig Enable Info
6 Alert Option Enable Info
7 Relay Fault Error
4 0 Alarm 1 Set Warning
1 Alarm 2 Set Warning
2 Alarm 3 Set Warning
3
4
5
6
7
"Not used" bits are always set to 0.
Some bits used in this transmitter indicate device or sensor failure and, therefore, also set bit 7 and bit 4 of the Device Status byte.
These bits are set or cleared by the self-test executed at power up, or following a reset. They are also set (but not cleared) by any failure detected during continuous background self-testing.
D-5
Page 90
Universal Commands
All Universal commands have been implemented in the Ultima Gas Monitor. The Ultima Gas Monitor returns a 7 in the Universal rev to indicate the device is using the expanded 16-bit manufacturer and device codes.
Common-Practice Commands
The following Common Practice commands have been implemented in the Ultima X device:
Table D-6. Supported Commands
COMMAND # DESCRIPTION
35 Write Range Values
38 Reset “Configuration Changed” flag
40 Enter/Exit Fixed Current Mode (See the following Warning)
42 Perform Master Reset
45 Trim DAC Zero
46 Trim DAC Gain
48 Read Additional Device Status
59 Write Number of Response Preambles
71 Lock Device
72 Squawk
80 Read Device Variable Trim Point
The gas monitor will NOT register gas concentration changes on the 4-20 mA signal line if the operator places the unit in Fixed Current Mode. Implement alternate protection measures when the unit is placed in this mode. Make sure the unit is returned to Standard Operational Mode prior to use for gas detection. Failure to follow this warning can result in serious personal injury or loss of life.
Burst Mode
This device supports burst mode.
Catch Device Variable
This Field Device does not support Catch Device Variable.
"
WARNING
D-6
Page 91
Table D-7. Device-Specific Commands
The following device-specific commands are implemented in the Ultima Monitor:
COMMAND # DESCRIPTION
129 Read Sensor Gas Type
130 Read Device RTC
131 Read Alarm Setpoints
132 Read Alarm Control Actions
133 Read Min/Max/Average Values
134 Read Last Cal Date
135 Read Gas Table
136 Read Input Voltage
137 Read Auto Zero Comp
139 Read Sensor Status
140 Read Swap Delay Status
141 Read Cal Signal Status
142 Read Alert Option Status
143 Read Sensor Temperature
144 Read Relay Normal State
173 Write Device RTC
174 Write Alarm Setpoints
175 Write Alarm Control Actions
176 Write Average Interval
177 Write Upper Trim Point
178 Write Gas Table
179 Write Sensor Data sheet Reset
180 Write Sensor Swap Delay Enable
181 Write Cal Signal Enable
182 Write Calibration Mode
183 Write Calibration Abort
184 Write Calibration Step
185 Write Alarm Acknowledge
186 Write Protect Mode
187 Write Alert Option
188 Write Relay Normal State
Command #129: Read Sensor Gas Type
Reads the Gas Type of the sensor currently connected to the Ultima Gas Monitor.
D-7
Page 92
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0-3 ASCII Sensor gas type description
(see TABLE D-8)
Command #130: Read Device Real Time Clock
Reads the Real Time clock hours and minutes from the Ultima X Gas Monitor.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned RTC Hours
1 Unsigned RTC Minutes
Command #131: Read Alarm Setpoints
Reads the Ultima X Alarm Setpoint values.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0-3 Float Alarm 1 Setpoint Value
4-7 Float Alarm 2 Setpoint Value
8-11 Float Alarm 3 Setpoint Value
D-8
Page 93
Command #132: Read Alarm Control Actions
Reads the Ultima X Alarm Control Actions.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Bit Enum Alarm 1 Control Actions (see TABLE D-9)
1 Bit Enum Alarm 2 Control Actions (see TABLE D-9)
2 Bit Enum Alarm 3 Control Actions (see TABLE D-9)
Command #133: Read Min, Max, Avg Values
Returns the Ultima minimum, maximum and average values recorded over an average interval. The average interval can be a value of 1 , 8, or 24 hours. For a one-hour interval, the value is updated at the top of each hour. For an eight-hour interval, the values are updated at 800, 1600 and 2400 hours.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0-3 Float Minimum Value
4-7 Float Maximum Value
8-11 Float Average Value
12 Unsigned Average interval (1, 8, or 24)
D-9
Page 94
Command #134: Read Last Cal Date
Returns the Ultima last calibration date of the currently connected sensor.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0-2 Unsigned Last sensor calibration date
Command #135: Read Gas Table
This command returns the Ultima sensor Gas Table currently in use. The Gas Tables are linearization reference tables used with certain sensors to provide accurate response for different gases from the same sensor.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned Gas Table Number (see TABLE D-10)
Command #136: Read Input Voltage Value
Returns the Ultima input supply voltage value. This number should be in the range of 8-30 volts DC.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0-3 Float Input Voltage Value
D-10
Page 95
Command #137: Read Auto Zero Comp Value
Returns the Ultima Automatic Zero Compensation value. This value is accumulated by the device when the sensor reading attempts to drift below zero. This value is used to compensate the actual Zero calibration. The device will attempt to compensate up to 10 counts (display units) before setting the under-range bit.
Command #139: Read Sensor Status message
Returns the Ultima X sensor status message. This is a single byte containing hex codes. This byte is sent from the sensor module to the main processor and passed to the HART communications processor.
Request Data Bytes
BYTE FORMAT DESCRIPTION
None
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Enum Sensor Status message (See TABLE D-12)
Command #140: Read Swap Delay Status
This command returns the Ultima X sensor swap delay status. This is a single byte containing a 0 if disabled or 1 if enabled. If enabled, the swap delay will hold off a sensor missing error for 1 minute. This hold­off allows a sensor module to be swapped out with a calibrated sensor module without triggering a "sensor missing" alarm and dropping the 4­20 mA to the trouble level.
Request Data Bytes
BYTE FORMAT DESCRIPTION
None
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Enum Sensor Swap Delay Status (0 –disabled, 1 - enabled)
D-11
Page 96
Command #141: Read Cal Signal Status
This command returns the Ultima X Cal Signal status. This is a single byte containing a 0 if disabled or 1 if enabled. If enabled, the output will be set to 3.75 mA during calibration (21 mA for oxygen). If disabled, the output will track the gas concentration.
Request Data Bytes
BYTE FORMAT DESCRIPTION
None
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Enum Sensor Cal Signal Status (0 –disabled, 1 - enabled)
Command #142: Read Alert Option Status
This command returns the Ultima X Alert Option Status status. This is a single byte containing a 0 if disabled or 1 if enabled. If enabled, the Alert Option will cause the 4-20 mA to be set to 3.75 mA during calibration of an Oxygen sensor (if the Cal Signal Option is also enabled). If the Alert Option is disabled and the Cal Signal enabled, the output will be set to 21 mA during Oxygen sensor calibration.
Request Data Bytes
BYTE FORMAT DESCRIPTION
None
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Enum Alert Option Status (0 - disabled, 1 - enabled)
D-12
Page 97
Command #143: Read Sensor Temperature
This command returns the Ultima X Sensor Temperature. This is a single byte containing an integer value representing the temperature returned by the gas sensor. Not all gas sensors have an on-board temperature.
Request Data Bytes
BYTE FORMAT DESCRIPTION
None
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned Sensor temperature (°C)
Command #144: Read Relay Normal State
This command returns the Ultima X to the Normal relay state. This is a single byte containing a bit map of the three alarm relays’ non alarm states. Not all gas sensors have on-board relays.
Request Data Bytes
BYTE FORMAT DESCRIPTION
None
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Bit 0 Alarm #1 0 = normally de-energized, 1 = normally energized
0 Bit 1 Alarm #2 0 = normally de-energized, 1 = normally energized
0 Bit 2 Alarm #3 0 = normally de-energized, 1 = normally energized
D-13
Page 98
Command #173: Write RTC
Writes the Ultima Real Time Clock hours and minutes values.The real time clock is used to compute the minimum, maximum and average values and to date stamp the last sensor calibration date.
Request Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned RTC hours (0-23)
1 Unsigned RTC minutes (0-59)
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned RTC hours (0-23)
1 Unsigned RTC minutes (0-59)
Command-Specific Response Codes
CODE CLASS DESCRIPTION
0 Success No Command-Specific Errors
1-2 Undefined
3 Error Parameter too large
4 Undefined
5 Error Too few data bytes
6 Undefined
7 Error In write protect mode
8-15 Undefined
16 Error Access Restricted
17-31 Undefined
32 Error Busy
33-127 Undefined
D-14
Page 99
Command #174: Write Alarm Setpoints
Writes the Ultima Alarm Setpoint values. The Ultima Gas Monitor uses alarm setpoint values to set alarm status bits in the device. The alarms can be enabled or disabled, set to increasing or decreasing and can be set to latching (see "Command 175: Write Alarm Setpoint Control Actions"). The adjustment range is greater than zero and less than full ­scale.
Request Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned Alarm Number (1, 2, or 3)
1-4 Float Alarm Setpoint Value
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned Alarm Number (1, 2 or 3)
1-4 Float Alarm Setpoint Value
Command-Specific Response Codes
CODE CLASS DESCRIPTION
0 Success No Command-Specific Errors
1-2 Undefined
3 Error Parameter too large
4 Error Parameter too small
5 Error Too few data bytes
6 Undefined
7 Error In write protect mode
8-15 Undefined
16 Error Access Restricted
17-18 Undefined
19 Error Invalid device variable index
20-31 Undefined
32 Error Busy
33-127 Undefined
D-15
Page 100
Command #175: Write Alarm Setpoint Control Actions
Writes the Ultima X Alarm Setpoint Control Actions. The Ultima X Gas Monitor uses alarm setpoint Control Actions to enabled or disabled, set to increasing or decreasing and to set the alarm to latching or non latching.
Request Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned Alarm Number (1, 2, or 3)
1 Bit Enum Alarm Control Action Value (see TABLE D-9)
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned Alarm Number (1, 2, or 3)
1 Bit Enum Alarm Control Action Value (see TABLE D-9)
Command-Specific Response Codes
CODE CLASS DESCRIPTION
0 Success No Command-Specific Errors
1-4 Undefined
5 Error Too few data bytes
6 Undefined
7 Error In write protect mode
8-15 Undefined
16 Error Access Restricted
17-18 Undefined
19 Error Invalid device variable index
20-31 Undefined
32 Error Busy
33-127 Undefined
D-16
Loading...