THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO
HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE
PRODUCT. Like any piece of c omplex equipment, the unit will perform as
designed o nly if it is ins talled, used a nd serviced in acc ordance with the
manufacturer’s instructions. OTHERWISE IT COULD FAIL TO PERFORM AS
DESIGNED AND PERSONS WHO RELY ON THIS PRODUCT FOR THEIR
SAFETY COULD SUSTAIN SEVERE PERSONAL INJURY OR DEATH.
The warranties made by Mine Safety Appliances Company with respect to the
product are voided if the product is not used and serviced in accordance with
the instructions in this manual. Please protect yourself and others by following
them. We encourage our customers to write or call regarding this equipment
prior to use or for any additional information relative to use or repairs.
In the U.S., to contact your nearest stocking location, dial toll-free 1-800-MSA-INST.
To contact M SA International, dia l 1-412-967-3354.
This manual is available on the internet at www.msanet.com
Manufactured by
MSA NORTH AMERICA
P.O. Box 42 7, Pi tt sb ur gh, P ennsylvan ia 15230
(L) Rev 7711266
Page 2
MSA Permanent Instrument Warranty
1. Warranty- Seller warrants that this product will be free from
mechanical defect or faulty workmanship for a period of
eighteen (18) months from date of shipment or one (1)
year from installation, whichever occurs first, provided it is
maintained and used in accordance with Seller’s
instructions and/or recommendations. This warranty does
not apply to expendable or consumable parts whose
normal life expectancy is less than one (1) year such as,
but not limited to, non-rechargeable batteries, sensor
elements, filter, lamps, fuses etc. The Seller shall be
released from all obligations under this warranty in the
event repairs or modifications are made by persons other
than its own or authorized service personnel or if the
warranty claim results from physical abuse or misuse of
the product. No agent, employee or representative of the
Seller has any authority to bind the Seller to any
affirmation, representation or warranty concerning the
goods sold under this contract. Seller makes no warranty
concerning components or accessories not manufactured
by the Seller, but will pass on to the Purchaser all
warranties of manufacturers of such components. THIS
WARRANTY IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESSED, IMPLIED OR
STATUTORY, AND IS STRICTLY LIMITED TO THE
TERMS HEREOF. SELLER SPECIFICALLY DISCLAIMS
ANY WARRANTY OF MERCHANTABILITY OR OF
FITNESS FOR A PARTICULAR PURPOSE.
2. Exclusive Remedy- It is expressly agreed that Purchaser’s
sole and exclusive remedy for breach of the above
warranty, for any tortious conduct of Seller, or for any
other cause of action, shall be the repair and/or
replacement at Seller’s option, of any equipment or parts
thereof, which after examination by Seller is proven to be
defective. Replacement equipment and/or parts will be
provided at no cost to Purchaser, F.O.B. Seller’s Plant.
Failure of Seller to successfully repair any nonconforming
product shall not cause the remedy established hereby to
fail of its essential purpose.
3. Exclusion of Consequential Damage- Purchaser
specifically understands and agrees that under no
circumstances will seller be liable to purchaser for
economic, special, incidental or consequential damages
or losses of any kind whatsoever, including but not limited
to, loss of anticipated profits and any other loss caused by
reason of nonoperation of the goods. This exclusion is
applicable to claims for breach of warranty, tortious
conduct or any other cause of action against seller.
Page 3
WARNINGS
"
1. Each instrument described in this manual must be
installed, operated and maintained in strict accordance
with its labels, cautions, warnings, instructions, and within
the limitations stated.
2. These monitors must not be installed in areas or locations
where explosive concentrations of combustible gases or
vapors might occur in the atmosphere: Class 1, Group A,
B, C, and D areas as defined by the NEC. Since the
monitors are not explosion-proof, they must be located in
non-hazardous areas.
3. The Toxgard II monitor is designed to detect toxic gases or
vapors in the air. It cannot measure the concentration of
toxic gases or vapors in steam or inert or oxygen-deficient
atmospheres. Neither will it indicate the presence of
explosive or combustible mists or sprays, lubrication oils,
nor explosive dusts, such as grain or coal dust.
4. As with all gas detection instruments of this type, high
levels of or long exposure to certain compounds - all of
which are not known, will contaminate the sensor.
Calibrate the monitor frequently to ensure that system
operation is dependable and its indications accurate.
5. The Toxgard II monitor must not be painted. Also, painting
should not be done near the sample line inlet. If painting is
being done in an area where the sample line inlet is
located, care must be exercised to make sure that paint is
not deposited on the sample inlet fitting of the monitor.
Such paint deposits would interfere with the sampling
process, whereby a sample of the atmosphere being
monitored is drawn into the Toxgard II unit.
6. The only absolute method to ensure proper overall
operation of a gas detection instrument is to check it with
a known concentration of the gas for which it has been
calibrated. Consequently, a calibration check should be
included as part of the routine inspection of the system.
Page 4
7. The sensor is a sealed unit which contains an electrolyte.
Should the sensor develop leakage, it should be
immediately removed from the sensing head and properly
discarded in accordance with local, state and federal
laws. To obtain a replacement sensor, contact MSA at the
address given in Chapter 5, "Obtaining Replacement
Parts." Caution must be exercised so that the electrolyte
does not contact skin eyes, clothing or circuitry. Get
immediate medical attention if electrolyte contacts skin or
eyes.
8. Perform periodic leak check on the sample line and all flow
system components and fittings of the Toxgard II monitor.
9. Use only genuine MSA replacement parts when performing
any maintenance procedures provided in this manual.
Failure to do so may seriously impair instrument
performance. Repair or alteration of the Toxgard II unit,
beyond the scope of these maintenance instructions, or
by anyone other than an authorized MSA serviceman,
could cause the product to fail to perform as designed,
and persons who rely on this product for their safety could
sustain severe personal injury or death.
10. Properly vent the exhaust of the Toxgard II monitor to a
safe area. Improper venting of the exhaust can cause
personal injury or death.
11. Do not install additional filters into the sample inlet fittings.
Adding the filters will impair instrument performance.
12. Ensure the gas sample and Toxgard II unit are at the same
temperature; otherwise, condensation will form, possibly
clogging or filling the sample line.
13. Do not mix units with different area classifications. All units
used in a classified area must have the proper area
classification. Otherwise, an explosion hazard exists,
resulting in possible explosion, injury or death.
FAILUR E TO CO MP LY WITH T HE ABOVE WARNI N GS CAN RESULT
IN SERIOUS PERSONAL INJURY OR DEATH.
Several cartons may arrive with your order; locate the carton(s)
marked as the Toxgard II Monitor. Upon receipt, ensure that your
package contains:
• Toxgard II Monitor (FIGURE 1-1)
• Toxgard II Monitor Instruction Manual (P/N 711266).
If your unit is a Toxgard II Monitor with Digital Output,
there will be an additional manual (P/N 711203)
• Toxgard II Remote Sensor Module, if ordered as a
Remote Sensor model (FIGURE 1-2)
1. Remove the Toxgard II unit from its shipping container(s),
taking care not to damage sensitive electrical components.
2. Search through the packing material and the containers to
avoid discarding valuable parts.
Initial Inspection
1. Turn the coin slot lock 1/4-turn clockwise and open the door
or use the key for the key lock, if so equipped.
2. Carefully inspect the components and assemblies inside the
case. If damage or shortage is noted, advise the carrier
promptly; make the proper claim with the carrier.
General Description
The Toxgard II Monitor is an easy -to-use, low-cost, one-sensor
monitor. It is designed to sample t he environment where the
monitor is mounted and to alert y ou to potentially dangerous
levels of one of the following gases , depending on your
particular model:
• Combust ible gas• Hydrog en Cyanide
• Oxygen • Hydrog en Chloride
• Carbon Monoxide• Nitric Oxid e
• Hydrog en Sulfide• Nitrogen Dioxide
• Chlorine• Sulfur Dioxide
• Chlorine Dioxide
1-1
Page 12
Chapter 1, Set-up
1-2
Figure 1-1.
Toxgard II Monitor
Page 13
Chapter 1, Set-up
Figure 1-2.
Remote Sensor Model
• All models are available as Remote Sensor models
(FIGURE 1-2). Combustible gas models require the remote
explosion-proof sensor module.
To determine your sensor type and options, check the shipping
carton (FIGURE 1-3) and sensor ID label. Checked items are
included in the carton.
Identifying Your Unit
To determine your model type and other inst alled options
(FIGURES 1-1 and 1-4), check the ID label on the shipping
carton (FIGURE 1-3). This label identifies :
• target gas, sensor location and range
• s ampling accessories (if applicable)
• visual and audible alarm options
• output accessories
• m ounting hardware.
Figure 1-3.
Shipping Carton Label Identifying Sensor Type
1-3
Page 14
Chapter 1, Set-up
Options and Accessories (Battery Back-up not shown)
Figure 1-4.
Mounting Your Gas Monitor
Generally, the unit or the r emote sensing module should be
mounted close to the area where a leak is likely to occur or
where the gas is expected. The Toxgar d II Monitor or Remote
Sensing Module should be installed at a high level (ceiling),
breathing level or low level (floor), depending on the density of
the gas being monitored. Mount the unit so that the front display
is not blocked or hidden from v iew.
WARNING
"
Do not locate the Toxgard II Monitor in an area which may contain
a flammable mixture of gas and air; otherwise, an explosion may
occur.
The general-purpose Toxgard II Monitor can be a source of ignition
and must not be mounted in an area where a flammable mixture of
combustible gas and air may be present; otherwise, an explosion
may occur.
1-4
Page 15
Chapter 1, Set-up
Mount the Toxgard II Monitor or the Remote Sensing Module with
its sensor inlet fitting (FIGURE 1-1 or 1-2) pointed downwards;
otherw ise, the inlet may become clogged with particulate matter
or liquids. Do not paint the Toxgard II Monitor. If painting is done
in an area where a sensor is located, exercise CAUTION to ensure
paint is not depos ited on the sensor inlet f itti ng. Suc h pain t depos its
would interfere with the diffusion process, whereby a sample of
the monitore d atmosph ere di ffuses in to the s ensor. Failu re to follow
the ab ove can result in serious per sonal injur y or death.
CAUTION
"
Protect the Toxgard II Monitor from extreme vibration. Do not mount
the sensing head in direct sunlight as this may cause the sensor
to overheat.
Ventilation patterns must be established bef ore positioning the
Toxgard II unit or sensor assembly. Smoke tubes (P/N 458481)
are useful in measuring the direction and r ate of air flow to
determine areas where the gas of int erest may tend to
accumulate.
Using suitable hardware (not supplied), m ount the Toxgard II
monitor to any flat surface v ia its four mounting lugs (see
FIGURE 1-6). Allow sufficient room t o open the unit front door.
The Toxgard II unit is designed to meet NEMA 4X requirements,
suitable for many damp, wet and/or outdoor locations.
Gas Sampling Options
Although the Toxgard II unit norm ally operates in the diffusion
mode (i.e., gases or vapors to be detected enter the sensor input
by natural molecular diffusion), a sam ple can be drawn into the
instrument if the DC pump or air aspirator option was purchased.
The Toxgard II P ump or Aspirator Model:
• dr aws a gas sample from the monitored area, through a sample
line (available separately) to the Toxgard II Gas Monitor sensor
• samples areas that are remote, inaccessible, too hot or too cold
for direct sensor monitoring, such as:
• ducts carrying combustible or toxic gas
• wet wells
• printing drying ovens
• blanking or inerting operations
• has electronics physically located in the main monitor enclosure
(FIGURE 1-1)
1-5
Page 16
Chapter 1, Set-up
• c omes equipped with a special flow cap for use on the Toxgard
II Gas Monitor to ensure proper sensor sampling.
• pr ov ides notification via the Toxgard II Monitor LED display and
fault relays if a flow fault is detected.
Table 1-1. Performance Specifications
for Toxgard II Pump or Aspirator Model
SAMPLE
TRANSPORT TIME
NOMINAL SAMPLE
FLOW RATE
MINIM UM SAMPLE
FLOW RATE
MAXIMU M SAMPLE
TUBING LENGTH
MAXIMUM
EXHAUST TUBING
LENGTH
INLET FITTINGS
EXHAUST FITTING
Only Remote Sensors with proper area ratings can be used in a
hazardous ar ea. The Toxgard II Monitor cannot be installed in a
hazardous area. Otherwise, an explosion hazard will exist resulting
is a possible explosion, injury or death.
30 seconds at 0.5 lpm with 50 feet (15.25 meters) of
.180 (4.57 mm) ID sample t ubing
25 feet (7.5 meters) maximum for Chlorine, Nitrogen
Dioxide and Chlorine Dioxide
20 feet (6 meters)
1/4" (6.35 mm) OD Tube Fitting
1/4" (6.35 mm) OD Tube Fitting
WARNING
"
Installation of DC Pump or Aspirator
Sampling
Sample Line Placement
• Remove all protective packaging plugs and/or caps from gas
INLET, CALIBRATION and EXHAUST PORTS.
• In-line filter must be used in the sample line.
• M SA tubing is available (P/N 600771).
The pump or aspirator draws a gas sam ple to the Toxgard II
Gas Monitor sensor (FIGURE 1-1):
• Can be extended up to 100 feet (30 Meters) away from the
Toxgard II Gas Monitor.
1-6
Page 17
Chapter 1, Set-up
• Us es 1/4-inch (6.35 mm) OD tubing. The tubing must be
compatible with the sampled gas. Sample tubing must also be
3/16" ID for 100-foot lines to prevent "Lo Flo."
• It takes a maximum of 30 seconds for the sample gas to reach
the Toxgard II Gas Monitor when 50 feet (15.25 meters) of
tubing is used:
• To decrease this time, shorten the sample line length.
• It is generally good practice to make the sample line as
short as possible.
• Depending on the gas characteristics, the end of the sampling
tubing and the sample inlet should be mounted to best optimize
sampling of that particular gas. Consult your architect, facility
manager or safety engineer for guidance in proper placement of
the sampling tube inlet.
• Testing for ventilation patterns is useful in establishing sample
inlet location. Smoke tubes (P/N 458481) are useful in
measuring the direction and rate of air flow to determine which
areas to monitor.
• A particulate filter (purchased separately) should be used at the
inlet end of the sample line to help prevent dirt and dust from
clogging the sample line. This filter must be compatible with the
gas being sampled.
• It is recommended that a stainless steel or Teflon* sample line
be used for monitoring gases that are highly reactive such as:
• Nitrogen Dioxide
• Sulfur Dioxide.
• Use only a Teflon sample line for the following gases:
• Chlorine
• Chlorine Dioxide
• Hydrogen Chloride
• Ammonia.
• Wit h the above gases, do not use an end-of-line filter. The
sample line should be inspected periodically and replaced if
dirty. Do not attempt to clean the sample line by applying
compressed air.
NOTE: Do not run the sample tubing in water or across
areas of vibration, doorways, man ways or access
ways; otherwise, a sample tubing leak or tripping
hazard may result.
4. Using suitable hardware (not supplied):
1-7
Page 18
Chapter 1, Set-up
• Fasten the sample tubing to supports.
• Do not tightly pinch the tubing; otherwise, flow may
be restricted.
• Fasten the end of the sample tubing and sample inlet in
the monitoring area.
• The sample inlet should be pointed downward to
prevent dirt and water from entering the sample
tubing line.
5. Install the end-of-line filter at the end of the sampling tubing.
6. Check for leaks along the entire length of the sample
tubing line. Any leak will dilute the gas sample from the
area of interest and give a lower than actual gas reading.
Exhaust Line Placement
1. For DC pump only:
Attach 1/4-inch (6.35 mm) OD sample tubing
to the exhaust outlet fitting.
For Aspirated systems only:
Attach 3/8-inch (9.52 mm) OD sample tubing
to the exhaust outlet fitting.
2. Route the exhaust tubi ng into a safe are a where the gas
sample can be vent ed. Avoid an y sharp bends or elbows.
Route the tubing to all ow any col lected water to drain.
• The maximum exhaust tubing length is 20 feet (6 meters).
NOTE: Do not run the exhaust tubing in water or across
areas of vibration or across doorways, man ways or
access ways; otherwise, a leak in the exhaust
tubing or a tripping hazard may result.
3. Fasten the exhaust tubing to supports using suitable
hardware (not supplied).
4. Fasten the end of the exhaust tubing in the area of interest
by using suitable hardware (not supplied).
• Exhaust inlet should be pointed downward to prevent dirt
and water from entering the exhaust tubing line.
NOTE: Exhaust to a safe or non-hazardous area only.
5. Check for leaks along entire length of the exhaust tubing line.
1-8
Page 19
Chapter 1, Set-up
Blocked Flow Detection of Sample Line
Units equipped with the pump or air aspir ator options are
supplied with a flow detector. This f low detector activates the
fault relay when insufficient flow exis ts for proper gas detection.
Mounting the Air-line Toxgard II Monitor
WARNING
"
These units must not be located in areas which may contain a
flammable mixture of gas and air; otherwise, an explosi on m ay occur.
CAUTION
"
Protect the Toxgard II Monitor from extreme vibration. Do not mount the
sensing head in direct sunlight as this may cause the sensor to overheat.
1. Using suitable hardware (not supplied) mount the Toxgard II
monitor to any flat surface; allow room to open the door.
2. Connect the source of compressed air being monitored to
the pressure regulator (fittings not included). Install the
regulator as close as possible to the sample take-off.
• The maximum regulator inlet pressure is 300 PSI.
• The regulator outlet pressure should be set to a nominal
10 PSI.
NOTE: An in-line filter trap (P/N 626967 or equivalent) is
recommended to prevent clogging of the flow
system due to air line contamination, such as oil, dirt
or moisture.
CAUTION
"
Do not exceed 300 pounds maximum inlet pressure; otherwise,
damage to the unit may occur.
3. Ensure the pressure regulator is fully counter-clockwise to
reduce the initial pressure to the unit.
Flow Block
The Toxgard II Monitor equipped with the flow block option can
be used to allow a sample to be deliver ed from a pressurized
stream to the Toxgard II unit. Inlet pressure to the flow block
should not exceed 10 PSI.
Mounting the Toxgard II Rem ote Sensor Module
The Remote Sensor Module:
1-9
Page 20
Chapter 1, Set-up
• is mounted where the gas
is expected to exist
• is mounted with the sensor
pointing downward via the
3/4" NPT threads located
on the end of the
module, opposite the
sensor
• can also be mounted with
the mounting strap (P/N
655700); see FIGURE 1-5
• wiring is routed to the
Toxgard II Monitor.
Permanently connect 3/8" OD
tubing to the center post of
the sensor inlet. Route this
tubing to the Toxgard II
Monitor, ensuring there are no
kinks, leaks or other
obstructions. Secure this tubing near the monitor; it is used to
deliver check gas to the s ensor module during calibration.
• Install an external switch or circuit breaker close to the unit for
disconnection purposes.
• Install an over-current protection device (e.g., fuse) in the
supply wiring to the unit; two-amp, slow acting fuses
are recommended.
WARNING
"
Before wiring the Toxgard II Monitor, disconnect power source
suppl ying the monit or; otherwi se, electri cal shock coul d occur.
CAUTION
"
Make all wiring connections and conduit runs according to accepted
commercial wiring practices. Install the Toxgard II unit in compliance
with the applicable requirements of the National Electrical Code
and/or any other local code requirements.
NOTE:For Toxgard II with Digital Output Monitors, see
manual (P/N 711203) for wiring instructions.
1-10
Page 21
Chapter 1, Set-up
Figure 1-6.
Toxgard II Cable Entries
• Route power, ground, and (where applicable) signal wiring
through the electrical entry holes in the case (FIGURE 1-6).
• S epar ate the power wiring from any signal wiring.
• The Toxgard II unit operates on an AC power source of 120 or
220 VAC (no selection switch is necessary).
CAUTION
"
If unsure of the power supply voltage available, contact your plant
engineer or safety officer. Damage may occur to the instrument if
the wrong power supply voltage is applied to the instrument.
1-11
Page 22
Chapter 1, Set-up
Outputs
Relays
Relay Specifications
Table 1-2. Relay Specifications
3 ALARMS
RELAYS
RELAY
RATINGS
FAULT
(NORMALLY-ENERGIZED)
HORN RELAY
At 125 or 250 VOLTS AC,
NON-INDUCTIVE
AT 30 VOLTS DC,
NON-INDUCTIVE
AT 1/10 HP
Alarm Relays
There are three alarm relays, a fault relay and a horn relay in
the Toxgard II Gas Monitor.
NOTE:Alarms are automatically disabled at the following
times to avoid "nuisance" alarms:
• during first minute of power-up, while cell stabilizes
• during calibration and for two minutes
following calibration.
SPDT (S ingle pole, do uble throw)
SPDT (S ingle pole, do uble throw)
SPST (Single pole, single throw), N/O
5.0 Amps
5.0 Amps
5.0 Amps
Alarm Features
There are alarm levels in the Toxgard II unit, and a relay for
each level. There are many user-select able alarm features
enabled or disabled by selecting or de-selecting them.
• Relays are standard electromechanical devices chosen for long
life and high reliability.
• Relay contacts are brought out to the terminal block and
used to control external lights, horns, and other equipment
(FIGURE 1-6).
• Alarms generally default to 10, 20 and 30% of the full-scale
reading and are set when the gas reading is above these
values.
• The Oxygen Model is a special case where:
• Caution is set to 19.5%, negative acting
• Warning is set to 18% oxygen, negative acting
• A lar m is set to 22% oxygen, positive acting.
1-12
Page 23
Chapter 1, Set-up
• These default setpoints are factory-set. The user can change
these setpoints; see Chapter 1, "Changing Parameter Settings."
• A r e available as N/O (normally-open) or
N/C (normally-closed) contacts.
• For Normally-Open contacts:
• Upon activation, the relay contacts close to provide an
electrical path. These contacts remain closed as long as:
• The alarm condition exists within the Monitor or
• The latching mode is selected and the RESET
button has not been pushed.
• For Normally-Closed contacts :
• Upon activation, the relay contacts open to interrupt the
electrical path. These contacts remain open as long as:
• The alarm condition exists within the Toxgard II Gas
Monitor or
• The latching mode is selected and the RESET
button has not been pushed.
• S ee Chapter 1, "Dip Switch Settings" to determine the dip
switch settings for the other alarm options.
Fault Relay
• During normal operation, the Fault relay is energized.
• When a fault is detected or power is cut or turned OFF, the
relay changes state. The form C contact can be wired as either
normally-open (NO) or normally-closed (NC) for your needs.
• See the Chapter 5, TABLE 5-1, Troubleshooting Guidelines for
causes of faults.
Horn Relay/Audible Outputs
Audible Outputs
• The Toxgard II Monitor comes equipped with one of the
following audible alarms:
• Piezo Horn (standard) - 75 db at five feet
• Federal Horn - 95 db at five feet (not adjustable)
• Adjustable-Tone Banshee Horn - 95 db at five feet
(See Chapter 1, "Banshee Horn Audible Settings").
• The Audible Alarm sounds when the Toxgard II Monitor enters
a Fault or Alarm State, if that option is enabled (see TABLE 1-5)
• These options are factory- set at time of purchase.
1-13
Page 24
Chapter 1, Set-up
Horn Relay
• This Single-pole, single-throw (SPST), normally-open (N/O)
relay works in parallel with the instrument horn.
• These contac ts are open when the instrument horn is OFF or
not selected for operation [see FIGURE 1-6 for terminal block
(TB9) location].
• The contacts:
• close when the instrument horn activates
• remain closed until the horn is de-activated due to:
• elimination of the hazardous condition
• pres sing the HORN ACKNOWLEDGE button.
Relay Wiring with Inductive Loads
If you are connecting the relay s to motors, fluorescent lighting or
other inductive loads, it is nec essary to suppress any spar ks that
may occur at the relay cont act. These sparks will deteriorate or
destroy the relay contacts, rendering the relay inoperative or
useless. One way to reduce this destructive sparking is to install
a Quencharc protector (MSA P/N 630413) across the contacts.
Milliamp Output
The Toxgard II Monitor may be connected to any device capable
of accepting 4 to 20 mA analog si gnals such as the:
• M SA Model 5000 Monitor
• M SA Quad Gas unit
• P r ogrammable controllers
• DCS ’s , etc.
The standard unit is supplied with a non-isolated, 4-20 mA signal.
An isolated 4-20 mA output option is als o available.
Figure 1-7.
Standard 4-20 mA Terminal Wiring
1-14
Page 25
Chapter 1, Set-up
NOTE: The return path for this signal is connected to the
Toxgard II Monitor "analog common" circuitry, and
therefore, is not "isolated." The Toxgard II Monitor
can be ordered from MSA with an optional current
isolator module which fully isolates the instrument
from supplementary instrumentation.
Isolated 4-20 mA Output
• The Toxgard II current output [the (+) or upper terminal of TB1]
connects to the isolator input (+) terminal
• The Toxgard II signal return [the (-) or lower terminal of TB1]
connects to the isolator input (-) terminal.
For Signal Loads Less Than 50 Ohms:
• The user load connects to the isolator output (S) and
(-) terminals.
For Signal Loads Greater Than or Equal to 50 Ohms:
• The user load connects to the isolator output (+) and
(-) terminals.
Under Line-powered Conditions:
• The Toxgard II isolated current output
can drive a maximum load of 310 ohms (including wire).
Under Battery-powered Conditions:
• The Toxgard II isolated current output
is limited to a maximum load of 100 ohms.
Digital Output
The Toxgard II Monitor can be used as part of a Ultima Plus
high-speed digital network in order to report all internal
parameters to a central location. All that is additionally required if
this option is selected, is a net work communications board and a
twisted-pair connection to the networ k. See Ultima Plus
instructions for complete information conc erning this option.
WARNING
"
Since t he Toxg ard I I M onit or i s ty pi call y us ed a s a saf ety ins tru ment ,
it is rec ommen ded that the sta ndar d and op tiona l output s be use d
in the pr oper manner ; otherwise, t he user may not be not ified of
a condition that may not be considered normal.
Make w iring co nnect ions and c ondui t runs ac cordi ng to ac cepted
commercial wiring practices. Install the Toxgard II unit in compliance
1-15
Page 26
Chapter 1, Set-up
with the applicable requirements of the National Electrical Code
and/or any other local code requirements.
Visual Outputs
The unit comes equipped with the following vis ual alarms:
• front-panel LEDs for Fault, Caution and Warning (standard):
• Latched Alarms cause the front-panel LED to flash
• Unlatched Alarms cause the front-panel LED to turn ON.
Also, up to two top-mounted strobes can be added (optional):
• am ber (follows Warning settings)
• red (follows Alarm settings)
Wiring for these options is fact ory-performed at time of purchase.
Horn/Alarm Acknowledge Button
• A reset push-button is mounted on the enclosure exterior.
• A n internally-mounted terminal strip allows the user to wire an
external reset button.
• The reset button silences the audio horn and resets any latched
relays or alarms with alarm setpoints no longer exceeded.
• The reset button silences the horn but not the beacon(s) when
the alarms are latched and the gas concentration is still above
the alarm setpoint.
If installing a RESET push-button:
1. Route a two-conductor cable to terminal block TB13.
• Route this cable with DC power to avoid noise
interference from relay wiring.
2. Connect the two-conductor cable to the two positions of the
N/O momentary switch.
3. Identify the cable to enable proper connection at the button.
4. Route the cable to the push-button location; wire the button.
Remote Sensors
Electrical Connections for the Remote Sensor
Module
Before wiring the Toxgard II Remote Sensor Module, disconnect the
power source feeding the Remote Sensor Module and the Toxgard
1-16
WARNING
"
Page 27
Chapter 1, Set-up
II Monitor/Less Sensor; otherwise, electrical shock could occur.
Before removing the cover of an explosion-proof Toxgard II Remote
Sensor Module, verify that the surrounding area does not contain
a flammable mixture of combustible gas and air, since a source
of ignition is exposed; otherwise, an explosion may occur if a
metal ob ject contact s the circuit ry and produc es sparks.
Toxgard II Remote Sensor Modules (excluding t he combustible
models) are supplied with 50 feet of cable. This cable has a
connector on one end which plugs into its mate within the
Toxgard II Monitor/Less Sensor. T he other end of the cable can
be cut to the exact length needed.
The combustible Remote Sensor Module is wired with a
three-conductor cable. This cable c an be acquired locally.
Terminal strips within the Remote Sensor M odule and the
Toxgard II Monitor/Less Sensor per mit easy wiring of this cable.
CAUTION
"
When installi ng a Toxgard II Rem ote Sensor M odule with its mati ng
Toxgar d I I Mon i to r /L e s s Sensor, fo ll o w Na t ional Electrical and local
proced ural Code s; fail ure t o do so ca n re sult i n a n uns afe cond iti on .
Maximum distance between the Toxgard II Remote Sensor and
the Monitor is 50 feet. Do not splice additional cable to the
supplied cable or run more than 50 feet of user-supplied cable.
For Combustible Models
Locate/acquire the proper length of three-conductor cable. TABLE
1-3 shows recommended wire gauge for the dis tance between
the Remote Sensor and the Toxgard II Monitor/Less Sensor.
Table 1-3. Wire Gauge Vs. Distance Between Units
DISTANCEWIRE GAUGE
UP TO 15 FEET16 AWG
UP TO 25 FEET14 AWG
UP TO 50 FEET12 AWG
TABLE 1-4 lists several suppliers of 12-AWG cable. If there is a
possibility of electrical interf erence from external sources (such as
large motors, welders, electrical heaters, etc.) use shielded cable.
The preferred colors of conductor s within the cable are black ( K),
green (G) and white (W). If splices are necessary, solder and
insulate each conductor; then, wrap ent ire splice with suitable
electrical tape or covering. Do not use wire nuts or fasteners.
1-17
Page 28
Chapter 1, Set-up
Table 1-4. Remote Combustible Sensor Wiring Cable
THE FOLLOWING CABLES ARE THREE-CONDUCTOR, 12-AWG SHIELDED, .625" OD
CABLE SUPPLIERSCATALOG NUMBER
CAROL CABLE CO.
ALPHA WIRE CORP.
BELDEN WIRE & CABLE CO .
02725
1937/3
19206
Appropriate wiring conduit must be us ed between the Toxgard II
Monitor and the Toxgard II Remote Sensor. The cable must be
routed through conduit.
Wiring the Toxgard II Monitor/Less Sensor Using the Internal
Terminal Block
1. Open the unit.
2. Locate TB2 in the upper-right portion of the main (lower)
pc board.
3. Install wires onto positions 1, 2 and 3, noting the color used
on each position.
At the Toxgard II Remote Sensor Location (FIGURES 1-8, 1-9 and
1-10)
1. Unscrew and open the Toxgard II Remote Sensor cover.
WARNING
"
Before removing the cover of an explosion-proof Toxgard II Remote
Sensor, verify that the surrounding area does not contain a
flammable mixture of combustible gas and air since a source of
ignition is exposed; otherwise, an explosion may occur if a metal
object contacts the circuitry and produces sparks.
1-18
Figure 1-8.
Combustible Remote Sensor
Page 29
Figure 1-9.
Oxygen Remote Sensor Cable
Chapter 1, Set-up
Figure 1-10.
Oxygen and Toxic Remote Sensor
2. Route the cable from the Toxgard II Monitor through the
conduit opening and wire it to the terminal block in the
bottom of the sensor.
NOTE: Separate the power and ground wiring from any
signal wiring.
3. Verify the identity of each conductor of the cable and
connect the wire to the terminal block so that position 1 in
the sensor is wired to position 1 on the Toxgard II Monitor
TB2. Similarly, sensor positions 2 and 3 connect to
Toxgard II positions 2 and 3.
If shielded wire is used:
• Cut the shield wire short.
• Do not connect the shield at the remote sensor.
1-19
Page 30
Chapter 1, Set-up
4. Screw the conduit into the conduit opening and relieve any
excess slack in the cable.
5. Re-install the cover of the Toxgard II Remote Sensor.
WARNING
"
Do not allow the cover to remain off of an explosion-proof Toxgard
II Sens or. si nce a sourc e of ig nitio n is expose d, an explo sion may
occur if a metal object contacts the circuitry and produces sparks
in an atm osphere of co mbustible ga s.
Oxygen Remo te Sensor Wiring
Fifty feet of four-conductor c able is supplied with the unit. This
cable has a connector on one end which plugs into its mate
within the Toxgard II Monitor/Less Sensor; the other end of t he
cable can be cut to the exact length needed. Do not run more
than 50 feet of cable or splic e on any additional cable.
At the Toxgard II Monitor/Less Sensor Location
The cable connects to J5 at t he upper right portion of the (lower)
main pc board.
At the Toxgard II Remote Oxygen Sensor Location
1. Unscrew and open the Toxgard II Remote Sensor cover.
WARNING
"
Before removing the cover of an explosion-proof Toxgard II Remote
Sensor, verify that the surrounding area does not contain a
flammable mixture of combustible gas and air since a source of
ignition is exposed; otherwise, an explosion may occur if a metal
object contacts the circuitry and produces sparks.
2. Route Toxgard II Monitor cable through the conduit opening
and wire it to the terminal block in sensor bottom.
3. Verify the identity of each conductor cable and connect the
wire to the terminal block (FIGURE 1-9).
Note that there are four wires needed out of the six within the
cable. Use the markings on the conductors at the end of
the original cable or see FIGURE 1-9 for the colors of the
wires that are needed.
CONDUCTOR ID
WHITE WITH BLUE
WHITE WITH G REEN
GREEN WITH WHIT E
BLUE WITH WHITE
TERMINAL BLOCK
LOCATION
3
5
6
8
1-20
Page 31
Chapter 1, Set-up
• The other wire within the cable is not used.
• It is unnecessary to connect wire to the other terminal
positions within the remote sensor.
• Do not connect the shield at the remote sensor.
4. Screw the conduit or cable grip or gland into the conduit
opening. Relieve any excess slack in the cable.
5. Re-install the cover on the Toxgard II Remote Sensor.
WARNING
"
Do not let the cover remain off of an explosion-proof Toxgard II
Sensor. Since a source of ignition is exposed, an explosion may
occur if a metal object contacts the circuitry and produces sparks
in an atm osphere of co mbustible ga s.
Toxic Gas Remote Sensor Wiri ng
(FIGURE 1-11)
Fifty feet of six-conductor cable is supplied with the unit.
• This cable has a connector on one end which plugs into its
mate within the Toxgard II Monitor/Less Sensor.
• The other end of the cable can be cut to the exact
length needed.
• Do not run more than 50 feet of cable or splice on
additional cable.
At the Toxgard II Monitor/Less Sensor Location
The cable connects to J5 at t he upper right portion of the (lower)
main pc board.
Figure 1-11.
Toxic Gas Remote Sensor Cable
1-21
Page 32
Chapter 1, Set-up
At the Toxgard II Remote Toxic Sensor Location
1. Open the Toxgard II Remote Sensor cover by unscrewing it.
WARNING
"
Before removing the cover of an explosion-proof Toxgard II Remote
Sensor, verify that the surrounding area does not contain a
flammable mixture of combustible gas and air since a source of
ignition is exposed; otherwise, an explosion may occur if a metal
object contacts the circuitry and produces sparks.
2. Route the cable from the Toxgard II Monitor through the
conduit opening and wire it to the terminal block in the
bottom of the sensor.
3. Verify the identity of each conductor cable and connect the
wire to the terminal block (FIGURE 1-10).
CONDUCTOR ID
ORANGE WITH
WHITE
WHITE WITH BLUE
WHITE WITH
ORANGE
WHITE WITH GREEN
GREEN WITH WHITE
BLUE WITH WHITE
TERMINAL BLOCK
LOCATION
1
3
4
5
7
8
• There are no connections to positions 2 and 6 of the
terminal block within the remote sensor.
• Do not connect the shield at the remote sensor.
4. Screw the conduit or cable grip or gland into the conduit
opening; relieve any excess slack in the cable.
5. Re-install the Toxgard II Remote Sensor cover.
WARNING
"
Do not allow the cover to remain off of an explosion-proof Toxgard
II Sensor. Since a source of ignition is exposed, an explosion may
occur if a metal object contacts the circuitry and produces sparks
in an atm osphere of co mbustible ga s.
Pump Wiring
(FIGURE 1-12)
Dip Switch Settings
Many of the Toxgard II features can be set with the SW2 and
SW3 Dip Switches located on the main monitor printed circuit
1-22
Page 33
Chapter 1, Set-up
board (FIGURE 1-1). TABLES 1-5 and 1-6 show the positions for
user-selectable settings (FIGURE 1-5) :
Table 1-5. SW2 Switch Settings
NOTE: LE FT side of each SW2 Dip Swi tch indicates the ON posit ion and
*NO AUDIBLE ALARM WILL SOUND IF SWITCHES 3, 4 AND 5 ARE ALL OFF.
RIGHT side of each SW2 Dip Switch indicates the OFF position.
NOTE: LEFT side of each SW3 Dip S witch indic ates the ON p osition an d
NOTE: The delay time selected v ia switches 6 - 8
RIGHT side of each SW3 Dip Switch indicates the OFF position.
NUMBERFUNCTION
1
2
3
4
5
6
7
8
will apply to any delay sw itch(es) 3 - 5 in th e ON position.
*OFF DELAY TIM E
SPARE (LEAVE IN OFF POSITI ON)
SPARE
OFF DELAY ON/O FF - CAUTION
OFF DELAY ON/OFF - WARNING
OFF DELAY ON/OFF - AL ARM
*OFF DELAY TIME - BIT2
*OFF DELAY TIME - BIT1
*OFF DELAY TIME - BIT0
BIT2BIT1BIT0
OFFOFFOFF5
OFFOFFON15
OFFONOFF30
OFFONON60
ONOFFOFF90
ONOFFON120
ONONOFF150
ONONON180
TIME
(IN SECONDS)
Controller/Calibrator Operation
The Toxgard II Monitor is fully compatible wit h the Ultima
Controller and the Ultima Calibrator. T hese devices can be used
to set various parameters in the Toxgard II Monitor, including
alarm conditions, auto calibration conf iguration, etc.
The devices can also be used for calibrating the Toxgard II
Monitor. When using these devices, point them at the front panel
of the Toxgard II Monitor, aiming dir ectly at the triangular IR
symbol (located next to the word "WARNING" at the center of
1-24
Page 35
Chapter 1, Set-up
the front panel). For full Calibrat or and Controller operational
details, refer to instruction manual (P/N 813379).
Internal Push-buttons
A pair of internal push-buttons on the Toxgard II main pc board,
lower-right corner (FIGURE 1-1) can be us ed in place of the
Ultima Controller and Calibrator for most functions.
The right-hand button marked "CAL" is used to initiate all
instrument calibration functions, eliminating t he need to use the
Controller or Calibrator to perf orm calibration.
The left-hand button marked "MODE" allows t he user to scroll
and view the Toxgard II conf iguration parameters. This operation
is comparable with the Controller "s end display item" feature.
Viewing Configuration Parameters
At initial installation, use the Toxgard II MODE push-button
(FIGURE 1-1) to review the Toxgar d II configuration parameters.
To scroll, press and release t he MODE push-button. The Toxgard
II Monitor:
• temporarily stops displaying gas measurement readings
• displays a value and other relevant information for a parameter
(TABLE 1-7).
NOTE: Pressing and releasing the MODE push-button
within 10 seconds enables the user to scroll to the
next parameter. If the user does not press the
push-button within 10 seconds, the Toxgard II
Monitor ends the parameter viewing mode and
returns to a display of the actual gas measurement.
The Toxgard II Monitor is able to detect and notify
the user of a hazardous condition even while
scrolling the parameters. Although the LED
continues scrolling, the hazardous condition is noted
by the horn, relays and strobes (if equipped). When
the scroll list ends, the LED display will show the
alarm condition and gas measurement value.
1-25
Page 36
Chapter 1, Set-up
Table 1-7. Parameters Displayed During MODE Push-button
Operation
TEXT MESSA GEDESCRIPTIO N
MIN VALUEMinimum gas concentration of the previous time interval
MAX VALUEMaximum gas concentration of the previous time interval
AVG VALUEAverage gas concentration of the previous time interval
ZERO VALUE
SPAN VALUE
LEVEL #1 (CAUTION)
LEVEL #2
(WARNING)
LEVEL #3 (ALARM)
CURRENT TIMECurrent time, in a 24-hour format
FUTURE CAL TIME
AVG INTERVAL
CURRENT DATEDisplay toggles between current day of month and year
FUTURE CAL DATE
PREVIOUS CAL
DATE
MUX ADDR ESSInstrum ent addres s when used on a Ultima Pl us Network
CALIBRATION
INTERVAL
CAL SIGNAL STATUS ON/OFF status of the 3.75-mA signal during calibration
TEMPERATURET oxic sensor temperature
VALUE
VALUE
VALUE
Gas concentration expected to be applied at the next
calibration
Setpoint, enabled/disabled, latched/unlatched information
Time th at an Autoca l calibratio n will be perfo rmed
(applic able only for o peration with an Autocal unit )
Time in terval (in ho urs) used to det ermine min imum,
average and maximum gas concentration values
Date tha t the next Autoc al calibration will be perform ed.
Display toggles between day of month and year
(applic able only for o peration with an Autocal unit )
Date of the previous calibration
Days be tween Autocal c alibrations ( for Autocal opera tion
only)
MSA diagnostic; please ignore
1-26
Page 37
Chapter 1, Set-up
Changing Parameter Settings
Method 1 - Push-buttons
The following parameters can be set by using the MODE and
CAL push-buttons (located on t he Toxgard II Monitor main printed
circuit board):
2. While holding the MODE button, press the CAL button; then,
release both buttons.
Once in the Parameter Setting Mode:
1. Use the MODE button to change the displayed value.
2. Use the CAL button to accept and save the displayed setting.
Once a single parameter is set:
• the display advances to the next parameter.
When changing an alarm value or span value:
• each press of the MODE button increases the value
• the step value of that increment is determined by the
full-scale range
• if an entry error is made:
1.Press the CAL button repeatedly until "END" displays.
2.Restart parameter setting mode and enter correct data.
1-27
Page 38
Chapter 1, Set-up
NOTE:During the Parameter Setting Mode, normal unit
operation is suspended until this mode is exited.
No network communications, peripheral
communications, or alarm updates occur in this
mode. The current signal output will be equivalent to
a CAL signal of 3.75 mA.
When "END" displays, the Parameter Setting Mode
is complete and normal operation resumes.
• If no key is pressed during the Parameter Setting Mode:
• the unit returns to normal operation in one minute
• any values that were accepted by using the CAL button
are saved.
When the unit returns to normal operation:
Use the MODE button to display and verify the settings.
Alarm Setting:
• if the alarm is set to OFF (disabled):
• latching status and value are not shown
• once the upper limit is reached, the value rolls over to the
minimum value
• for toxic and oxygen units:
• the upper limit is the full-scale reading
• for combustible units:
• the upper limit is 60% LEL.
Method 2 - Ultima Controller
Parameters, alarm levels and their latched/ unlatched modes of
operation are set and changed using an Ultima Controller
(P/N 809086). Refer to the Ultima Cont roller Instruction Manual
(P/N 813379) for full instr uctions.
Battery Backup
The Toxgard II Monitor can be configured wit h an (internal)
rechargeable backup battery that perm its full instrument operation
if main power is interrupted or t he Toxgard II Monitor is used as
a transportable instrument for temporary gas monitoring of an
area; the carrying handle and side- mount sensor are convenient
options if the Toxgard II Monitor is used as a transportable gas
monitor.
1-28
Page 39
Chapter 1, Set-up
This battery is:
• a 12-volt , sealed lead acid type with a nominal power rating
of 2.2 amp-hours
• c ontinuously charged by the Toxgard II Monitor during
normal operation.
Transfer to the backup batt ery is automatic when main power
loss occurs. While operating on the backup battery, "BATT" lights
on the Toxgard II display. When m ain power is restored, the
"BATT" turns OFF, and the batt ery will start to recharge.
The amount of time that the bac kup battery can operate the
Toxgard II Monitor depends on the:
• type of sensor
• ins talled options
• pr es ence of alarm conditions
• am bient conditions.
TABLE 1-7 can be used as an approx imation of expected battery
operating time. In this table:
• " Minimum Time" represents a Toxgard II Monitor
configured with:
• str obes flashing
• alarm conditions asserted
• Federal horn operational
• internal pump
• " Maxim um Time" represents a Toxgard II Monitor
configured with:
• no strobes
• no alarms
• no horn
• no pump.
• A c tual battery lifetime falls somewhere between the indicated
Minimum Time and Maximum Time values.
1-29
Page 40
Chapter 1, Set-up
Table 1-8. Expected Battery Operating Time at 77o F
SENSOR TYPEMINIMUM TIMEMAXIM UM TIME
Combustible1 hour10 hours
Toxic o r oxygen0.5 hours24 hou rs
NOTE:As ambient temperature drops, the expected battery
operating time period decreases. At -4
o
F (-20oC),
the times shown in TABLE 1-8 are reduced to 60%.
During battery operation, the Toxgard II unit monitors backup
battery voltage. When voltage is r educed to a level where
continued operation yields unreliable instrument oper ation and
possible battery damage the instrument LED dis plays Err6 for 3
seconds and then turns itself OFF until main operating power
is restored.
WARNING
"
The "Faul t" relay mus t be connec ted to an alar ming device to alert the
user the instr ument is no longer operational when th e above condi tion
occurs.
NOTE: The Toxgard II battery pack requires up to 24-30
hours of uninterrupted main power to fully charge.
NOTE:Normally, the unit automatically transfers power to the
optional backup battery when the main power is
lost. However, if the user wants to start the unit from
an unpowered state using the battery, press SW6
on the main printed circuit board (FIGURE 1-12).
Banshee Horn Audible Settings
The Banshee Horn can produce several distinctly different
sounds. These sounds and the volume of these sounds can be
adjusted as follows:
1. Disconnect the main operating power from the
Toxgard II Monitor.
2. If the Toxgard II Monitor is equipped with an internal
back-up battery, carefully disconnect the WHITE wire from
the battery (+) terminal.
NOTE: Record the position of the RED and BLACK
1-30
wires connecting the horn to TB12 on the
Toxgard II Monitor main pc board. When this
Page 41
Chapter 1, Set-up
Figure 1-13.
Banshee Horn Audible Settings
procedure is complete, return these wires to their
original positions.
3. Remove the four screws from the face of the horn.
4. Carefully separate the body halves of the horn.
5. To change the sound of the Banshee Horn, move the
push-on jumper to the position to provide the desired
sound (FIGURE 1-13).
NOTE: The internal horn label shows which sound
corresponds with each position.
6. To change the volume of the Banshee Horn, use a small,
flat screwdriver to adjust the volume control shown on
FIGURE 1-13.
NOTE: The loudest volume is obtained in the fully
clockwise position.
7. When settings are complete, verify that the internal rubber
gasket lines up with the screw holes.
8. Carefully mate the two halves of the horn and place the
excess slack of the RED and BLACK wires inside the T.
9. Replace the four screws and tighten to ensure a
weatherproof seal.
10. Move the RED and BLACK wires connecting the horn and
the Toxgard II Monitor main pc board into their
original positions.
1-31
Page 42
Chapter 1, Set-up
11. Reconnect the main operating power, and reconnect the
WHITE wire to the Toxgard II Monitor internal backup
battery (if so equipped).
WARNING
"
Verify the proper o peration, sou nd and loud ness of the Ba nshee
Horn; otherwise, the horn cannot properly alert the user.
Autocalibration Option
The Toxgard II Monitor is available with an autocalibration option
This user-selectable option allows periodic, unmanned gas
calibration of the system.
• This option is not available for units with sampling options, nor
is it available for monitoring of the following target gases:
• chlorine
• chlorine dioxide
• hydrogen chloride
• nitrogen dioxide.
For installation details, refer to t he autocalibration Instruction
Manual (P/N 815379).
Duct-Mounted Remote Sensor Installation
Refer to Appendix A for this opt ion.
1-32
Page 43
Chapter 2, Start-up
Chapter 2
Start-up
Initial Start-up
The Toxgard II Monitor is fact ory-calibrated and ready for
immediate use. Its range and output are preset. Once power is
applied to the unit, the display shows a test of all display words
and segments. The software version number displays; then, a
25-second warm-up countdown for sensor st ability begins:
• a blank screen appears between each second of countdown
time.
• There is no signal output during this warm-up time.
The Monitor shows the software v ersion number again; then, it
begins to display the gas concentrat ion in the corresponding unit
of measurement (FIGURE 2-1):
• ppm (for toxic gases)
• % Gas (for oxygen only)
• % LEL (for combustible gases).
Figure 2-1.
Gas Concentration Display
2-1
Page 44
Chapter 2, Start-u p
After initial start-up, the Toxgard II Monitor also displays a
calibration flag as a reminder to perf orm a field calibration
(FIGURE 2-2). While this message is display ed, the Toxgard II
Monitor is functioning within factory- set calibration parameters.
Once the first successful fi eld calibration is performed, the
calibration flag turns OFF.
CAUTION
"
To ensure a fully functional sensor, perform a calibration check
and adjustments at initial start-up.
During normal operation, the Toxgard II Monitor displays the gas
concentration of the surrounding environment. The corresponding
output signal can be transmitted t o a controller. For a complete
listing of display and output conditi ons other than normal
operation, see TABLE 5-1.
NOTE:The combustible gas model of the Toxgard II Monitor is
capable of detecting concentrations of certain
combustible gases above 100% LEL. When
exposed to these concentrations, the Toxgard II
Monitor will display one of two modes:
• LOC % LEL - The Toxgard II Monitor has be en
exposed to a high con cen trat ion of ga s (abov e th e
LEL) and there is a possibility that the over-r ange
condition may s till ex ist.
2-2
• 100 % LEL - The Toxgard II Monitor has been
exposed to a high concentration of gas (above the
LEL) and the over-range con dition definitely still exists.
Figure 2-2.
Calibration Flag Display
Page 45
Chapter 2, Start-up
CAUTION
"
In either mode, co rrect th e condit ion ca using th e exces sive g as leve l
and vent or purge the area before attempting the following.
In the LOC % LEL mode, the output signal is also locked
at full-scale. If this condition occurs, the Toxgard II
Monitor must be unlocked by performing a "Zero
Function," either with the Ultima Calibrator or
Controller or by using the internal "CAL"
push-button.
NOTE:The Ultima Calibrator and Controller are compatible
with both the Ultima Gas Monitor and the Toxgard II
monitor.
In the LOC mode, The Toxgard II Monitor will not
revert to a normal condition until a successful zero
operation has been performed. This is an exclusive
safety feature of the Toxgard II Monitor which
pre-empts the possibility of ambiguous readings
when the sensor is exposed to concentration of gas
above 100% LEL.
In the 100% LEL mode, the combustible gas is over the
100% LEL range. It will return to normal operation
when the gas concentration level falls below 100%
LEL.
Calibration
As with any type of gas monit or, the only true check of its
performance is to apply gas directly t o the sensor. The frequency
of the calibration gas test s depends on the operating time and
chemical exposures of the sensors. New sensors should be
calibrated more often until the calibrat ion records prove sensor
stability. The calibration frequency c an then be reduced to the
schedule set by the safety off icer or plant manager. Before
calibrating the sensors, it is good practice to apply power to the
unit to allow the sensor to adapt to the new environment. This is
especially important for the following unit s:
• chlorine
• nitric oxide
• hydrogen cyanide
• hydrogen chloride
• chlorine dioxide.
2-3
Page 46
Chapter 2, Start-u p
Sensors should be powered at least one half hour before any
calibration attempt is made.
Combustible sensors located in areas wher e non-combustible
chemicals may leak, particularly ones
known to reduce the sensitivity (see
following list) should be calibrated
after such exposures.
Non-combustible Chemicals that
Reduce Sensor Sensitivity
• Silanes, Silicates, Silicones and
Halides (compounds
containing Fluorine, Chlorine,
Iodine or Bromine)
• TABLE 2-1 lists interferants for
electrochemical sensors.
When it is determined that calibrat ion
adjustments are required, the
Toxgard II Monitor provides a
one-man, non-intrusive method of
adjustment at the unit.
Figure 2-3.
Ultima
®
Calibrator
To calibra te the unit, the internal
push-buttons or one of the
following accessories can be used:
• Calibr ator P/N 809997
(FIGURE 2-3)
• Controller P/N 809086
(FIGURE 2-4)
Calibrator
The Calibrator allows the following
functions:
• Zero
• Calibr ation (zero and span)
• Display address for digital
output mod els .
Toxgard II Monitor Controller
The Controller also provides the
above functions and access to t he
additional following features:
2-4
Figure 2-4.
Ultima
®
Controller
Page 47
INTER-
FERANT
Acetone
Acetylene
Ammonia
Benzene
Carbon
Dioxide
Carbon
Disulfide
Carbon
Monoxide
Chlorine
Dimethyl
Sulfide
Ethanol
Ethylene
Ethyl Ether
Freon 12
Freon 113
Freon 12Bl
Freon 13Bl
Hexane
Hydrogen
Hydrogen
Chloride
Hydrogen
Cyanide
Hydrogen
Sulfide
Isopropanol
Mercaptan
(Ethyl)
Mercaptan
(Methyl)
Methane
Methanol
Methyl Amine
Methyl
Bromide
Methyl Ethyl
Ketone
Nitric Oxide
Nitrogen
Dioxide
Sulfur Dioxid e
Tichloroethylene
Chapter 2, Start-up
Table 2-1. Sensor Response to Interferants
CONCENTRATION
(PPM)
100000000
12000
10000
50000002.500
10010070.21003-30
1001150000
4003
1000-20
100010
100000
100000
500000000
500280.5150-150
10000
50000000100
1000
5000
2000
100102.22-33201001
15014.7NA40000357.50
10000
COH2SSO2HCNNO
No Data
No Data No Data
0
No Data
000
20000000
1520
No Data
0.1
No Data
CL2HCLNOCLO
2
No Data No Data No Data No Data
No Data No Data No Data No Data
0
No Data No Data
No Data No Data
No Data No Data
No Data No Data
0
0
0
0
0
500-6-0.245-102.5
520.43-0.200
No Data No Data No Data No Data
50990.10-0.300
No Data No Data No Data
No Data
No Data
No Data No Data No Data No Data No Data No Data No Data
No Data No Data No Data No Data No Data No Data No Data
No Data No Data
200
2.4
No Data No Data No Data No Data
0
No Data No Data No Data No Data No Data
No Data
0
No Data No Data
No Data No Data
No Data No Data
No Data No Data
No Data No Data
1004
0
0
0
0
0
No Data
1050310002.500
10010050-8-0.54.54-0.25
504000000
572.7
574.5
No Data
No Data
No Data
6.5
No Data No Data No Data No Data No Data
6
50130000.100
No Data No Data No Data
No Data
No Data
No Data
0
No Data
0
No Data No Data No Data No Data
0
No Data No Data No Data No Data
0
No Data No Data No Data No Data
0
No Data No Data
No Data No Data
0
No Data No Data
No Data No Data
0
0
0
0
10-2-8.4-10-201012-126
No Data
No Data No Data No Data No Data No Data No Data
0
2
2-5
Page 48
Chapter 2, Start-u p
• Automatic calibration at timed intervals when used with MSA
autocalibration module.
NOTE: If the autocalibration module was pur c hased for use
with the Toxgard II Monitor, consult the MSA Ultima
Controller Instruction Manual (P/N 813379) for
set-up and operation.
• Three levels of alarm
• Date of last successful calib ratio n
• M aximum and minimum gas readings over selected time periods
• Average gas readings over selected time periods
• Changing of span gas value from factory-set value
• Access to real-time clock for a time and date stamp
See Calibrator and Controller instruction m anual (P/N 813379) for
proper calibration procedures.
Toxgard II Monitor Calibration Output Si gna l
The Toxgard II Monitor is shipped with the calibration output
signal disabled so the output signal will t rack the gas
concentration value during the calibration process. In some
applications, it may be desirable to dis able or lock the output
during calibration to a pre-determined output value to prevent
activation of alarm devices. The calibration signal can be enabled
using the Controller. When the calibrati on signal is enabled, the
output signal is 3.75 milliamps during calibr ation.
NOTE:The Controller is compatible with both the Ultima Gas
Monitor and the Toxgard II Monitor.
2-6
Page 49
Chapter 3, Calibration
Chapter 3
Calibration
The Toxgard II Gas Monitor provides non-intrusive calibration
through the use of the Calibrator or Controller. The unit can also
be calibrated using the internally-mounted "CA L" push-button.
NOTE:The Calibrator and Controller are compatible with both
the Ultima Gas Monitor and the Toxgard II monitor.
When calibrating any Toxgard II Gas Monitor, refer to the
Calibration Kit instructions.
When first placing the Toxgard I I Gas Monitor into operation it is
good practice to perform a calibr ation. When replacing the
sensor, it is necessary to do an initial calibration which allows the
unit to store the initial sensor parameters for proper end-of-life
indication. Refer to Chapter 3, "Initial Calibration."
Read the appropriate calibration instr uctions before attempting an
actual calibration. Also, identify and become familiar with all of
the calibration components. During the c alibration, it is necessary
to quickly apply the span gas t o the unit. Prior connection of the
calibration components will aid in the ease of unit calibration.
Equipmen t R equ ir ed
Three calibration kits (numbered 40, 41, and 54) are available
from MSA for diffusion calibr ation of Toxgard II Gas Monitors.
Kit 40 is also used to calibrat e the Toxgard II Air Line Monitor.
Calibration Kits 42 and 43 are used to calibrate Toxgard II
monitors with pumped samples. All Calibration Kits are housed in
a convenient carrying case and contain all items necessary (less
gas) for a complete and accurat e calibration.
The check or calibration gases can als o be carried in the case.
See TABLE 3-1 for the appropr iate zero and span gas cylinders
for your Toxgard II Gas Monitor .
WARNING
"
These calibration kits contain zero caps in place of zero calibration
gas. Th ese cap s can o nly be used wh en the am bient a ir doe s not
contai n the gas the mo nit or is de tecti ng. If there i s any doub t, use
zero gas when zeroing the Toxgard II Monitor; otherwise, improper
calibr ation could o ccur.
3-1
Page 50
Chapter 3, Calibrati on
Table 3-1. Factory-set Span Values
GAS TYPERANGE
CARBON
MONOXIDE
SULFUR DIOXIDE
HYDROGEN
SULFIDE
NITRIC OXIDE
NITROGEN DIOXIDE
CHLORINE
HYDROGEN
CYANIDE
HYDROGEN
CHLORIDE
CHLORINE
DIOXIDE
(4)
OXYGEN
NATURAL GAS
PETROLEUM
VAPORS
(3)
(3)
SPAN GAS
PRESET
VALUES
0-100 PPM60 PPM
0-500 PP M30 0 PPM
0-25 PPM10 PPM
0-10 PPM5 PPM
0-50 PPM40 PPM
0-100 PPM40 PPM
0-100 PPM50 PPM
0-10 PPM5 PPM
0-5 PPM2 PPM
0-50 PPM10 PPM
0-100 PPM40 PPM
0-3 PPM1 PPM
0-25%20.8%
0-100% LEL25% LEL(1 )
0-100% LEL40% LEL(1 )
MSA RP
CYLINDER P/N
CALIBRATION
71088240
1002793840
1002807040
71041440
10028062
1002807440
71033241
71033141
1002807241
1002807854
71033141
10028028
(2)40
1002803440
KIT
40
40
(GASOLINE)
GENERAL
SOLVENTS
(3)
0-100% LEL55% LEL(1 )
NOTES:
1Cali brated with Propane (.6% ga s by volume)
2Not requ ired for standard cal ibration procedure
3For co mbustible gas, it is goo d practice to calibrate unit with the gas
to be detected
4CLO
5 Change span gas valu e with Controller to us e low range calibration
is calibrated with CL2
2
cylinder.
3-2
Page 51
Chapter 3, Calibration
Calibration Kits 40 and 42
Calibration Kit 40 will calibrate all dif fusion Toxgard II Gas
Monitors and Calibration Kit 42 will calibrat e all sample draw
Toxgard II Gas Monitors for the following gases (FIGURE 3-1):
• Carbon Monoxide
• Oxygen
• Hydrogen Sulfide
• S ulfur Dioxide
• Nitric Oxide
• All Combustible gases.
Calibration Kits 41 and 43
Calibration Kit 41 will calibrate all dif fusion Toxgard II Gas
Monitors and Calibration Kit 43 will calibrat e all sample draw
Toxgard II Gas Monitors for the following gases (FIGURE 3-2):
• Chlorine
• Nitrogen Dioxide
• Chlorine Dioxide
• Hydrogen Cyanide.
NOTE:The calibration procedure for the sample draw Toxgard
II Monitor is the same as the procedure for the
diffusion version, except calibration gas is applied to
the calibration entry port of the inlet flow block.
Calibration Kit 54
Calibration Kit 54 will calibrate diffus ion Hydrogen Chloride
Toxgard II Monitors (FIGURE 3-5).
Span Gas Values
The Toxgard II Monitor is fact ory-shipped with a preset span gas
value that can be changed via the Cont roller (P/N 809086);
otherwise, the span gas must correspond to preset
concentrations. See TABLE 3-1 for factory-set span values.
CAUTION
"
To ensure a fully functional sensor, perform calibration checks
and adjustments at initial start-up and at regular intervals.
3-3
Page 52
Chapter 3, Calibrati on
Item 1 - Tubing (P/N 711112)
- 3/16" I D side connects to Item 3;
- 1/4" ID side connects to sensor
Item 2 - Z ero Cap (P/N 710535)
Figure 3-1. Calibration Kit 40 Contents
(Your Kit may also include one or two gas cylinders)
Item 1 - Tubing (P/N 711112)
- 3/16" side co nnects to Ite m 2;
- 1/4" side connects to sensor
Item 2 - .25 LPM Flow Controller (P/N 478359)
Item 3 - 1 .5 LPM Flow Controller (P/ N 478358)
Item 3 - Calibration Ca p (P/N 7104 11)
(Your Kit may also include one or two gas cylinders)
3-4
Item 4 - Ze ro Cap (P/N 813774)
Figure 3-2. Calibration Kit 41 Contents
Page 53
Chapter 3, Calibration
Item 1 - Tubing (P/N 711113)
-
- 1/8" ID side connects to Ultima or Tox gard II Sensor
Item 2 - F low Controll er, Match Flo w (P/N 69734 4)
3/16" ID side connects to Item 2
Figure 3-3. Calibration Kit 42 Contents
(Your Kit may also include one or two gas cylinders)
Item 1 - Tubing (P/N 711113)
- 3/16" ID side connects to Item 2
- 1/8" ID side connects to Ultima or Toxgard II se nsor
Item 2 - Fl ow Controlle r, Match Flo w (P/N 697344 )
Figure 3-4. Calibration Kit 43 Contents
(Your Kit may also include one or two gas cylinders)
3-5
Page 54
Chapter 3, Calibrati on
Item 1 - Tubing (P/N 711112)
- 3/16" I D side connects to Item 3;
- 1/4" ID side connects to sensor
Item 2 - Zero Cap
(P/N 710535)
Item 4 - Desiccant
(P/N 10 064306)
Item 3 - 1.5 LPM
Flow Controller
(P/N 4783 58)
Item 5 - C alibration Cover Assem bly
(P/N 10066581)
Figure 3-5. Calibration Kit 54 Contents
(Your Kit may also include one or two gas cylinders)
NOTE: Item 5 (P/N 10066581) is used only for SAFEMTX Calibration
Toxgard II Gas Monitor Calibration
The only true check of any gas m onitor’s performance is to apply
gas directly to the sensor. The calibration procedure must be
performed regularly.
Calibration Documentation
The Toxgard II Monitor records t he date of the last successful
calibration. This date can then be displayed on the front-panel
LED (with the use of the Controller ) or transmitted to a
3-6
Page 55
Chapter 3, Calibration
Control Data Center if the Ultima Plus Digital Network option
was purchased.
More frequent calibration may be required when the unit is new.
Calibration frequency depends on operating time and chemical
exposures of the instrument. Also perform calibration when
installing or changing the power source of the control
instrumentation. In some cases, it may be necessary to perform
only a zero function of the Gas Monitor in lieu of a full zer o and
span procedure. Check with your saf ety officer or safety engineer
to determine if only a zero function is necessary.
NOTES:
• If the sensor element has been changed or r eplac ed, see
Chapter 3, "Initial Calibration."
• If this is an oxygen sensor, see Chapter 3, "Oxygen Calibration."
• B efore calibrat ing Chlorine, Nitrogen Dioxide, Hydrogen
Chloride, Chlorine Dioxide, or Hydrogen Cyanide, apply power
to the unit at least 1/2-hour before calibrating.
• Due to the unstable nature of Chlorine Dioxide (ClO
), Chlorine
2
gas is used as a calibration simulant. If using the MSA
calibration system and gas cylinder (P/N 710331), the response
ratio is 2:1 (the 2-ppm sample of Chlorine should be set to read
1 ppm of ClO
ClO
Toxgard II Gas Monitor is 1 ppm.
2
). The default value for the calibration gas on the
2
To Zero the Toxgard II Gas Monitor Only
NOTE:Combustible sensor configurations are only available
with a remote sensor; side-mount and bottommount sensors are not available in a combustible
sensor configuration.
Identify your Toxgard II Gas Monitor and select the proper one of
the following two zeroing calibration proc edures.
• Use Procedure Number 1 to zero your unit for the foll owing gases:
• Carbon Monoxide
• Oxygen
• Hydrogen Sulfide
• Sulfur Dioxide
• Hydrogen Chloride
• Nitric Oxide
• All Combustible gases.
3-7
Page 56
Chapter 3, Calibrati on
• Use Procedure Number 2 to zero your unit for the following
gases:
• Chlorine
• Nitrogen Dioxide
• Chlorine Dioxide
• Hydrogen Cyanide
Procedure Number 1
• This procedure uses items found in Calibration Kit 40, 42, and
54.
• For oxygen units, skip to Step 3.
1. If Using the zero cap (FIGURE 3-6):
If the ambient air is suitable, with no traces of the gas of
interest, place the zero cap (P/N 710535), from the
Calibration Kit, over the SensorGard inlet and wait two
minutes; otherwise, use zero gas.
2. If Using zero gas cylinder:
a.Locate the zero gas cylinder and the 1.5 LPM Flow
Controller (P/N 478358).
3-8
Figure 3-6.
Zero Cap Placement
Page 57
Chapter 3, Calibration
b.Screw the Flow
Controller onto
the top of the
zero gas
cylinder.
c.Locate the Tube
Assembly
(P/N 711112).
d.Push the smaller
end of the Tube
Assembly over
the Flow
Controller
gas outlet.
• Ensure
tubing
completely
Calibration Tubing Set-up
Figure 3-7.
covers the gas outlet (FIGURE 3-7).
e.Connect the other end of the tubing
over the SensorGard inlet (FIGURE 3-8).
f.Turn on the gas flow by turning the knob on the
flow controller.
• Not ice the open arrow on the flow controller knob.
Figure 3-8.
Zero Gas Connection
3-9
Page 58
Chapter 3, Calibrati on
3. Point the Calibrator or Controller at the Toxgard II Gas
Monitor display (FIGURE 1-2); press the ZERO button or
push the right-hand internal CAL push-button (FIGURE
1-1) until the word "ZERO" appears; then, release.
NOTE: The zero or calibration process can be aborted at
any time during the 30-second countdown interval;
simply press any button on the Calibrator or
Controller while aiming it at the unit or by pushing
the internal push-button.
The 30-second countdown interval is omitted
for oxygen units. It is electronically zeroed.
The display shows:
• A countdown from 30 to 0 seconds
• APPLY ZERO GAS(FIGURE 3-9)
4. After the 30 second countdown:
• The display alternates between "CAL" and a value (for
example: 0 PPM). This value is the actual reading of the
gas concentration the sensor is detecting.
• Once the gas value on the display is stable for 15
seconds, the alternating display stops. If the calibration is
successful, the display will show END and 0
(FIGURE 3-10).
a.If using the zero cap: remove it.
b.If using a zero ga s c y linder:
1) Turn OFF the gas flow by turning the flow
controller knob.
2) Remove the tubing from the SensorGard.
• If the calibration signal is enabled during
3-10
Figure 3-9.
Apply Zero Gas Flag
Page 59
Chapter 3, Calibration
Figure 3-10.
Calibration End Display
calibration, it will be held at the lockout value
for an additional two minutes.
c.If a CAL FAULT flag appears on the unit (FIGURE 3-11),
this indicates:
• An unsuccessful attempt to zero or calibrate the
Toxgard II Gas Monitor
• The Toxgard II Gas Monitor is operating with the
calibration parameters defined before the calibration
was attempted.
To extinguish the CAL FAULT flag, a complete calibration
procedure must be performed.
• See Troubleshooting Guidelines found in Chapter 5.
The Toxgard II Gas Monitor allows automatic zero adjustment
only within a pre-defined range. It cannot make any
corrections outside this range, such as when an empty or
wrong cylinder of gas is applied or failure to begin gas
flow within the allotted 30-second countdown occurs.
Figure 3-11.
Calibration Fault Flag
3-11
Page 60
Chapter 3, Calibrati on
5. If Using Zer o Ga s :
After a successful zeroing, remove the tubing from the
Flow Controller and remove the Flow Controller from the
cylinder; return all items to their appropriate location in the
Calibration Kit.
If Using Zer o C a p:
Remove cap and return it to the Calibration Kit.
Procedure Number 2
This procedure uses the items found in Calibr ation Kit 41 and 43.
1. If Using the Zero Cap (FIGURE 3-12):
If the ambient air is suitable, with no traces of the gas of
interest, place the zero cap (P/N 813774) over the sensor
inlet and wait two minutes; otherwise, use zero gas.
2. If Using Zero Gas Cylinder:
a.Locate the zero gas cylinder and the .25 LPM Flow
Controller (P/N 478359).
b.Screw the Flow Controller onto the top of the zero
gas cylinder.
c.Locate the Tube Assembly (P/N 711112).
3-12
Figure 3-12.
Zero Cap Placement
Page 61
Chapter 3, Calibration
Figure 3-13.
Zero Gas Connection
d.Push the smaller end of the Tube Assembly over the Flow
Controller gas outlet.
• Ens ure tubing completely covers the gas outlet
(see FIGURE 3-6).
e.Connect the other end of the tubing over the SensorGard
inlet (FIGURE 3-13).
f.Turn ON the gas flow by turning the knob on the
flow controller.
• Not ice the open arrow on the flow controller knob.
3. Point the Calibrator or Controller at the Toxgard II Gas
Monitor display (FIGURE 1-2) and press the ZERO button
or push the right-hand internal push-button (FIGURE 1-1)
until the word "ZERO" appears; then, release..
The display shows:
• A countdown from 30 to 0 seconds
• APPLY ZERO GAS(FIGURE 3-9)
NOTE: The zero or calibration process can be aborted at
any time during the 30-second countdown interval;
simply press any button on the Calibrator while
aiming it at the unit or by pushing the internal
push-button.
3-13
Page 62
Chapter 3, Calibrati on
4. After the 30 second countdown:
• The display alternates between "CAL" and a value
(example: 0 PPM). This value is the actual reading of the
gas concentration the sensor is detecting.
• Once the gas value on the display is stable for 15
seconds, the alternating display stops. If the calibration is
successful, the display will show END and 0
(FIGURE 3-10)
a.If Using the Zero Cap: remove it.
b.If Using a Zero Gas Cylinder:
1) Turn OFF the gas flow by turning the flow
controller knob.
2) Remove the tubing from the Sensor inlet.
• If the calibration signal is enabled during calibration,
it will be held at the lockout value for an additional
two minutes.
c.If a CAL FAULT flag appears on the unit (FIGURE 3-11),
this indicates:
• An unsuccessful attempt to zero or calibrate the
Toxgard II Gas Monitor
• The Toxgard II Gas Monitor is operating with the
calibration parameters defined before the calibration
was attempted.
To extinguish the CAL FAULT flag, a complete calibration
procedure must be performed.
The Toxgard II Gas Monitor allows automatic zero
adjustment only within a pre-defined range. It cannot
make any corrections outside this range, such as when
an empty or wrong cylinder of gas is applied or failure to
begin gas flow within the allotted 30-second countdown
occurs.
5. If Using Zer o Ga s :
After a successful zeroing, remove the tubing from the
Flow Controller and remove the Flow Controller from the
cylinder; return all items to their appropriate location in the
Calibration Kit.
3-14
Page 63
Chapter 3, Calibration
Perform a Complete Calibration
• If this is the first calibration after the sensor element is replaced,
refer to "Initial Calibration."
• If this is an oxygen sensor, refer to "Oxygen Calibration."
• Fo r CL
one regulator (P/N 478359) for each of these gases. They will
not work properly if one regulator is used for multiple gases.
Identify your particular Toxgard I I Gas Monitor and select the
proper one of the following two c alibration procedures.
• Use Procedure Number 1 to calibrate the Toxgard II Gas
Monitor for the following gases:
• Use Procedure Number 2 to calibrate the Toxgard II Gas
Monitor for the following gases:
and CLO2 calibration, do not mix regulators. Use only
2
• Carbon Monoxide
• Oxygen
• Hydrogen Sulfide
• Sulfur Dioxide
• Hydrogen Chloride
• Nitric Oxide
• All Combustible gases.
• Chlorine
• Nitrogen Dioxide
• Chlorine Dioxide
• Hydrogen Cyanide.
Procedure Number 1
This procedure uses the items found in Calibr ation Kit 40, 42,
and 54.
• For oxygen units, skip to Step 3.
• For Calibration Kit 54 diffusion Hydrogen Chloride, run HCL gas
through flow control regulator and tubing for five minutes before
starting a calibration. Dedicate this flow control regulator and
tubing for use with HCL gas only.
Zeroing
1. If Using the Zero Cap:
If the ambient air is suitable, with no traces of the gas of
interest, place the zero cap (P/N 710535), from the
3-15
Page 64
Chapter 3, Calibrati on
Calibration Kit, over the SensorGard inlet and wait two
minutes; otherwise, use zero gas (FIGURE 3-6).
2. If Using Zero Gas Cylinder:
a.Locate the zero gas cylinder and the 1.5 LPM Flow
Controller (P/N 478358).
b.Screw the Flow Controller onto the top of the zero
gas cylinder.
c.Locate the Tube Assembly (P/N 711112).
d.Push the smaller end of the Tube Assembly over the Flow
Controller gas outlet.
• Ens ure tubing completely covers the gas outlet
(FIGURE 3-7).
e.Connect the other end of the tubing over the SensorGard
inlet (FIGURE 3-8).
f.Turn on the gas flow by turning the knob on the
flow controller.
• Not ice the open arrow on the flow controller knob.
3. Point the Calibrator or Controller at the Toxgard II Gas
Monitor display (FIGURE 1-2); press the CALIBRATE
button or press and hold the right-hand internal
push-button for about five seconds until the word SPAN
appears on the display. Release the button when SPAN
is showing.
NOTE: The zero or calibration process can be aborted at
any time during the 30-second countdown interval;
simply press any button on the Calibrator or
Controller while aiming it at the unit or press the
internal push-button.
The 30-second countdown interval is omitted for oxygen
units. It is electronically zeroed.
The display shows:
• A countdown from 30 to 0 seconds
• APPLY ZERO GAS(FIGURE 3-9)
4. After the 30 second countdown:
• The display alternates between "CAL" and a value
(example: 0 PPM). This value is the actual reading of the
gas concentration the sensor is detecting.
3-16
Page 65
Chapter 3, Calibration
• Once the gas value on the display is stable for 15
seconds, the alternating display stops. If the calibration is
successful, the display will show END and 0
(FIGURE 3-10).
a.If using the zero cap: remove it.
b.If using a zero ga s c y linder:
1) Turn OFF the gas flow by turning the flow
controller knob.
2) Remove the tubing from the Flow Controller.
• If the calibration signal is enabled during
calibration, it will be held at the lockout value
for an additional two minutes.
c.If a CAL FAULT flag appears on the unit (FIGURE 3-11),
this indicates:
• An unsuccessful attempt to zero or calibrate the
Toxgard II Gas Monitor
• The Toxgard II Gas Monitor is operating with the
calibration parameters defined before the calibration
was attempted.
To extinguish the CAL FAULT flag, a complete calibration
procedure must be performed.
• See Troubleshooting Guidelines found in Chapter 5.
The Toxgard II Gas Monitor allows automatic zero adjustment
only within a pre-defined range. It cannot make any
corrections outside this range, such as when an empty or
wrong cylinder of gas is applied or failure to begin gas
flow within the allotted 30-second countdown occurs.
Spanning
• If this is the first calibration after the sensor element is replaced,
refer to "Initial Calibration."
• If this is an oxygen sensor, refer to "Oxygen Calibration" later in
this chapter.
5. The Toxgard II Gas Monitor automatically begins the span
countdown after a successful zeroing of the unit. The
span countdown is also 30 seconds (FIGURE 3-14).
• The span process can be aborted at any time during the
countdown by simply pressing any button on the
Calibrator while aiming it at the unit or by pressing the
internal push-button.
3-17
Page 66
Chapter 3, Calibrati on
Figure 3-14.
Apply SPAN Gas Flag
6. Turn OFF the zero gas and remove the flow controller, if
used previously.
7. Locate the span gas cylinder and the 1.5 LPM Flow
Controller (P/N 478358).
8. Screw the Flow Controller onto the top of the span
gas cylinder.
9. Locate the Tube Assembly (P/N 711112).
10. Push the smaller end of the Tube Assembly over the gas
outlet of the Flow Controller.
• Ensure that the tubing completely covers the gas outlet
(FIGURE 3-7).
11. Connect the other end of the tubing over the SensorGard
inlet (FIGURE 3-15).
12. Turn ON the gas flow by turning the flow controller knob.
• Notice the open arrow on the knob of the flow controller.
• If a CAL FAULT indication is on the Toxgard II Gas
Monitor display before the user is able to apply the gas
(FIGURE 3-11), a steady state gas condition existed for
longer than 15 seconds; this caused the unit to use a
wrong reading as a span indication.
• I t is necessary to restart the calibration process to
clear this condition.
• Ensure that any calibration gases are applied during
the 30-second count down period.
3-18
Page 67
Chapter 3, Calibration
Figure 3-15.
Span Set-up
• I t is good practice to have all calibration components
previously assembled.
13. After the 30 second countdown:
• The display alternates between "CAL" and a value.
(for example: 60 PPM for 0 to 100 ppm carbon
monoxide). This value is the actual reading of the gas
concentration the sensor is detecting.
• Once the gas value on the display is stable for 15
seconds, the alternating display stops. If the calibration is
successful, the display will show END for approximately
two seconds. (FIGURE 3-16).
• There are no adjustments necessary.
• The display will show the
span gas value while the span
gas is flowing to the unit. (For
example, it may read 60 PPM
or 25 % or 60 %LEL; FIGURE
3-17.)
14. Turn OFF the gas flow by turning
the knob on the flow controller.
Figure 3-16.
Calibration End Display
3-19
Page 68
Chapter 3, Calibrati on
• If the calibration signal is enabled during calibration, it will
be held at the lockout value for two additional minutes.
• When the span gas is removed from the sensor, the
sensor reading may drift for a few minutes; this is normal
sensor operation.
Figure 3-17.
Span Gas Connection
3-20
• If a CAL FAULT flag appears on the unit (FIGURE 3-11),
this indicates:
• An unsuccessful attempt to calibrate the Toxgard II
Gas Monitor
• The Toxgard II Gas Monitor is operating with the
calibration parameters defined before the calibration
was attempted.
To extinguish the CAL FAULT flag, a complete calibration
procedure must be performed.
The Toxgard II Gas Monitor allows automatic zero and span
adjustments only within a pre-defined range. It cannot
make any corrections outside this range, such as when an
empty or wrong cylinder of gas is applied or failure to
begin gas flow within the allotted 30-second
countdown occurs.
Page 69
Chapter 3, Calibration
15. After a successful calibration, remove the tubing from the
Flow Controller and remove the Flow Controller from the
cylinder; return all items to their appropriate location in the
Calibration Kit.
16. For Calibration Kit 54 diffusion Hydrogen Chloride:
a.After a successful calibration, flush the flow control
regulator and tubing with 100% Nitrogen for five
minutes. Use this flow control regulator with Hydrogen
Chloride gas only.
b.Store the flow control regulator and tubing in the
desiccated sealed bag included in Calibration Kit 54 or
equivalent dry container.
Procedure Number 2
This procedure uses the items found in Calibr ation Kit 41 and 43.
Zeroing
1. If Using the Zero Cap (FIGURE 3-12):
If the ambient air is suitable, with no traces of the gas of
interest, place the zero cap (P/N 813774) over the sensor
inlet and wait two minutes; otherwise, use zero gas.
2. If Using Zero Gas Cylinder:
a.Locate the zero gas cylinder and the .25 LPM Flow
Controller (P/N 478359).
b.Screw the Flow Controller onto the top of the zero
gas cylinder.
c.Locate the Tube Assembly (P/N 711112).
d.Push the smaller end of the Tube Assembly over the Flow
Controller gas outlet.
• Ens ure tubing completely covers the gas outlet
(FIGURE 3-7).
e.Connect the other end of the tubing over the SensorGard
inlet (FIGURE 3-13).
f.Turn on the gas flow by turning the knob on the
flow controller.
• Not ice the open arrow on the flow controller knob.
3. Point the Calibrator at the Toxgard II Gas Monitor display
(FIGURE 1-2) and press the Calibrator or Controller
3-21
Page 70
Chapter 3, Calibrati on
CALIBRATE button or press and hold the internal
push-button for about five seconds until the word SPAN
appears on the display. Release the button when SPAN
is showing.
NOTE: The zero or calibration process can be aborted at
any time during the 30-second countdown interval;
simply press any button on the Calibrator or
Controller while aiming it at the unit or pushing the
internal push-button.
The display shows:
• A countdown from 30 to 0 seconds
• APPLY ZERO GAS(FIGURE 3-9)
4. After the 30 second countdown:
• The display alternates between "CAL" and a value
(example: 0 PPM). This value is the actual reading of the
gas concentration the sensor is detecting.
• Once the gas value on the display is stable for 15
seconds, the alternating display stops. If the calibration is
successful, the display will show END and 0
(FIGURE 3-10).
a.If Using the Zero Cap: remove it.
3-22
b.If Using a Zero Gas Cylinder:
1) Turn OFF the gas flow by turning the flow
controller knob.
2) Remove the tubing from the SensorGard.
• If the calibration signal is enabled during
calibration, it will be held at the lockout value
for an additional two minutes.
c.If a CAL FAULT flag appears on the unit (FIGURE 3-11),
this indicates:
• An unsuccessful attempt to zero or calibrate the
Toxgard II Gas Monitor
• The Toxgard II Gas Monitor is operating with the
calibration parameters defined before the calibration
was attempted.
To extinguish the CAL FAULT flag, a complete calibration
procedure must be performed.
• See Troubleshooting Guidelines found in Chapter 5.
Page 71
Chapter 3, Calibration
The Toxgard II Gas Monitor allows automatic zero and span
adjustments only within a pre-defined range. It cannot
make any corrections outside this range, such as when an
empty or wrong cylinder of gas is applied or failure to
begin gas flow within the allotted 30-second
countdown occurs.
Spanning
• If this is the first calibration after the sensor element is replaced,
refer to "Initial Calibration."
• If this is an oxygen sensor, refer to "Oxygen Calibration."
5. The Toxgard II Gas Monitor automatically begins the span
countdown after a successful zeroing of the unit. The
span countdown is also 30 seconds (FIGURE 3-13).
• The span process can be aborted at any time during the
countdown by simply pressing any button on the
Calibrator while aiming it at the unit or by pressing the
internal push-button.
6. Turn OFF the zero gas and remove the flow controller if
used previously.
7. Locate the span gas cylinder and the .25 LPM Flow
Controller (P/N 478359).
8. Screw the Flow Controller onto the top of the span
gas cylinder.
9. Locate the Tube Assembly (P/N 711112).
10. Push the smaller end of the Tube Assembly over the gas
outlet of the Flow Controller.
• Ensure that the tubing completely covers the gas outlet
(FIGURE 3-7).
11. Locate the calibration cap (P/N 710411) and push the
tubing through the hole in the bottom of the cap.
12. Connect the other end of the tubing over the Sensor inlet
and push the calibration cap over the entire sensor inlet
(FIGURE 3-17).
13. Turn ON the gas flow by turning the flow controller knob.
• Notice the open arrow on the knob of the flow controller.
• If a CAL FAULT indication is on the Toxgard II Gas
Monitor display before the user is able to apply the gas
(FIGURE 3-11), a steady state gas condition existed for
longer than 15 seconds; this caused the unit to use a
wrong reading as a span indication.
3-23
Page 72
Chapter 3, Calibrati on
• I t is necessary to restart the calibration process to
clear this condition.
• Ensure that any calibration gases are applied during
the 30 second count down period.
• I t is good practice to have all calibration components
previously assembled.
14. After the 30 second countdown:
• The display alternates between "CAL" and a value
(for example: 60 PPM for 0 to 100 ppm carbon
monoxide). This value is the actual reading of the gas
concentration the sensor is detecting.
• Once the gas value on the display is stable for 15
seconds, the alternating display stops. If the calibration is
successful, the display will show END for approximately
two seconds (FIGURE 3-16).
• There are no adjustments necessary.
The display will show the span gas value while the span gas is
flowing to the unit. (For example, it may read 60 PPM or
25 % or 60 %LEL; FIGURE 3-17).
15. Turn OFF the gas flow by turning the knob on the
flow controller.
• If the calibration signal is enabled during calibration, it will
be held at the lockout value for two additional minutes.
• When the span gas is removed from the sensor, the
sensor reading may drift for a few minutes; this is normal
sensor operation.
• If a CAL FAULT flag appears on the unit (FIGURE 3-11),
this indicates:
• An unsuccessful attempt to calibrate the Toxgard II
Gas Monitor
• The Toxgard II Gas Monitor is operating with the
calibration parameters defined before the calibration
was attempted.
To extinguish the CAL FAULT flag, a complete calibration
procedure must be performed.
The Toxgard II Gas Monitor allows automatic zero and span
adjustments only within a pre-defined range. It cannot
make any corrections outside this range, such as when an
empty or wrong cylinder of gas is applied or failure to
3-24
Page 73
Chapter 3, Calibration
begin gas flow within the allotted 30-second
countdown occurs.
16. After a successful calibration, remove the tubing from the
Flow Controller and remove the Flow Controller from the
cylinder; return all items to their appropriate location in the
Calibration Kit.
OXYGEN Calibration
NOTE:If this is the first calibration after the sensor element is
replaced, perform an "Initial Calibration."
Oxygen calibration is slightly different from other gases. When
the ZERO function is performed, the 30-second countdown is
omitted because the Toxgard II unit performs the zero
electronically. No calibration cap or zero gas is necessary.
To meet the stated specification, it is necessary to span the
Oxygen Toxgard II Gas Monitor with t he Calibration Kit and an
oxygen cylinder. The concentration of oxygen-in-air varies slightly
due to changing relative humidity levels. These variations in
oxygen levels are detected by the oxygen Toxgard II Gas
Monitor. To meet the reproducibility s pecification, a calibration
gas cylinder must be used to ensure the same oxygen
concentration for every calibration.
25% Oxygen Toxgard II Gas Monitor
For the SPAN function, ambient air is generally adequate for
the 25% oxygen Toxgard II Gas Monitor as the expected default
span value is 20.8%. Therefore, when the display prompts
"APPLY SPAN GAS" it would be adequate to simply allow the
countdown to occur without applying gas.
NOTE:If the sensor is located in an area of normally low or
enriched oxygen, then a 20.8% oxygen sample
must be applied when the display prompts: "APPL Y
SPAN GAS".
INITIAL Calibration
When a new sensor element is placed in t he sensor, an INITIAL
Calibration must be performed. This procedur e enables the unit
to gather data about the sensor t o make accurate decisions for
the CHANGE SENSOR function and the CAL FAULT function to
work properly.
The INITIAL calibration is accomplished by aiming and
pressing the ZERO and CALIBRATE buttons of the calibrator
3-25
Page 74
Chapter 3, Calibrati on
simulta neously or by pressing and h olding the inte rnal, right-ha nd
button f or ab out 10 s eco nds unt il the di spla y show s "ICAL";
release the button when "ICAL" is showing.
• The display should show "SET APPLY ZERO GAS"
• The remainder of the procedure is now the same as that for a
regular calibration.
• The presence of the words "SET" and "ICAL" on the display
distinguish INITIAL Calibration from a regular calibration. If
these words do not appear, abort the calibration; then, retry the
above procedure.
NOTE: The zero or calibration process can be aborted at any
time simply by pressing any button during the
30-second countdown on the Calibrator while aiming
at the unit or by pressing the internal push-button.
• The display leads the user through the zero and span routines
as in a regular calibration.
NOTE:This procedure must be initiated only when a new
sensor element is installed.
Toxgard II Airline Monitor, Pumped Version and
Aspirated Version Calibration (FIGURES 3-18
through 3-20)
Calibrat ion Procedure
1. Install the Regulator/Flow Controller onto the
Zero Gas Cylinder.
2. Connect the larger end of the Calibration Tube Assembly
(3/16" ID) to the Cylinder Regulator.
3. Insert the smaller end of the tube assembly (1/8" ID) into the
calibration fitting and tighten the nut.
4. Initiate the calibration procedure on the Toxgard II Air
Line Monitor, Pumped or Aspirated versions.
5. Press and hold the calibration valve button.
6. Open the Regulator/Flow Controller valve on the
Gas Cylinder.
7. At the end of this sequence, the Toxgard unit displays END.
Close the Regulator/Flow Controller valve and release the
calibration valve button.
8. Remove the Zero gas and install the Regulator/Flow
Controller onto the Span Gas Cylinder.
3-26
Page 75
Chapter 3, Calibration
Figure 3-18.
Typical DC Pump Version Installation
Figure 3-19.
Typical Air Aspirator Version Installation
3-27
Page 76
Chapter 3, Calibrati on
Figure 3-20.
Typical Air Line Monitor Installation
9. Press and hold the calibration valve button.
10. Open the Regulator/Flow Controller valve on the Span gas
cylinder and allow the gas to flow until the reading
stabilizes and the unit displays END.
11. Close the Regulator/Flow Controller valve; then, release
the calibration valve button.
12. Return the unit to normal operation.
Zero Procedure
1. Install the Regulator/Flow Controller onto the
Zero Gas Cylinder.
2. Connect the larger end of the Calibration Tube Assembly
(3/16" ID) to the Cylinder Regulator.
3. Insert the smaller end of the tube assembly (1/8" ID) into the
calibration fitting and tighten the nut.
4. Initiate the Zero procedure on the Toxgard II Air
Line Monitor, Pumped or Aspirated versions.
5. Press and hold the calibration valve button.
3-28
Page 77
Chapter 3, Calibration
6. Open the Regulator/Flow Controller valve on the
gas cylinder.
7. At the end of this sequence, the Toxgard unit displays END.
Close the Regulator/Flow Controller valve and release the
calibration valve button.
8. Return the unit to normal operation.
3-29
Page 78
Chapter 4
Specifications
Table 4-1. Specifications
GAS TYPES
TEMPERATURE
RANGE
DRIFT
NOISE
ACCURACY
TOXICS & OXYGEN
COMBUSTIBLESDUAL MODULE:
ZERO DRIFT
SPAN DRIFT
CARBON MONOXIDE
OXYGEN
HYDROGEN SULFIDE
CHLORINE
SULFUR DIOXIDE
NITRIC OXID E
NITROGEN DIOXIDE
HYDROGEN
CYANIDE
HYDROGEN
CHLORIDE
COMBUSTIBLE GAS,
H
2
COMBUSTIBLE GAS,
PETROLEUM
VAPORS
COMBUSTIBLE GAS,
SOLVENTS
CHLORINE DIOXIDE
Chapter 4, Spec ificati ons
Combustibles, Oxygen & Toxics
OPERATING
RANGE:
*EXTENDED
RANGE:
Calibr ate within op erating rang e
Less than 1% FS
GASLINEARITYREPEATABILITY
the greater of
full scale (FS)
+
2% FS
+
10% FS or 2 ppm
+
10% FS or 2 ppm
+
10% FS or 2 ppm
+
10% FS or 2 ppm
+
10% FS or 2 ppm
+
10% FS or 2 ppm
+
10% FS or 2 ppm
+2% of scale
+
2% of sca le
+
2% of sca le
+
10% FS or 2 ppm
+
2%
o
C
0 to 40
(32 to +1 04oF)
-20 to +50
(-4 to +122oF)
-40 to +90
(-40 to
+194
Less than
5%/year,
typically
Less than
10%/year,
typically
+
1% FS or 2
ppm
+
1% FS
+
1% FS or 2
ppm
+
5% FS or 1
ppm
+
1% FS or 2
ppm
+
1% FS or 2
ppm
+
4% FS or 1
ppm
+
4% FS or 2
ppm
+
10% FS or
2 ppm
+
1% FS
+
1% FS
+
1% FS
+
5% FS or 1
ppm
o
C
o
C
o
F)
4-1
Page 79
Chapter 4, Specifi cat ions
Table 4-1. Specifications
*Extended Range = The sensor may not meet all of the accuracy parameters listed.
STEP CHANGE
RESPONSE
HUMIDITY
ALTITUDE
SENSOR LIFE
HAZARDOUS
AREA RATING
WIRING
REQUIREMENTS
POWER INPUT
SIGNAL OUTPUT
RELAY OUTPUTS
PHYSICAL
BEACON
TIME TO REACH
20% OF SCALE
- OXYGEN & TOXICS
TIME TO REACH
50% OF SCALE
- OXYGEN & TOXICS
TIME TO REACH
50% OF SCALE COMBUSTIBLES
TIME TO REACH
90% OF SCALE COMBUSTIBLES
COMBUSTIBLES
OXYGEN &TOXICS
FULL
REPLACEMENT
WARRANTY
mA OUTPUT
OXYGEN & TOXICS
COMBUSTIBLES
REMOTE SENSOR
4-20 mA
DIGITAL
5 AMP WITH
120-240 VAC;
5 AMPS WITH 30
VDC
SIZE
WEIGHT
CHARACTERISTICS
INTENSITY
Less than 12 seconds (typically 6
seconds)
Less than 30 seconds (typically 12
seconds)
Less than 10 seconds
Less than 30 seconds
15 to 95% RH, non-condensing
+10,000 feet (3,000 meters)
3 years , Typically
2 years , Typically
1 year
Gener al-purpose o nly. Check Are a
Rating of Remote Sensor for
allowable applications
OXYGEN &
TOXICS
COMBUSTIBLES
120 - 220 VAC
Supplie d by Toxgar d II Monitor
COMBUSTIBLES
OXYGEN &
TOXICS
See Ul tima Plus spec ifications
Three le vels of Alarm
One Fa ult Relay
16 x 23.3 x 5 inches
(40.6 x 59.2 x 12.7 cm) max.
13 pounds (5.9 kilograms)
Strobe type
70 to 90 flashes per minute
RED: 55 candella eff.
AMBER: 230 cand ella eff.
2-wire
2-wire
+15%
current source
4-2
Page 80
Chapter 4, Spec ificati ons
Table 4-1. Specifications
HORNS
HAZARDOUS
RATING
APPROVAL
PARAMETERS
PIEZO
FEDERAL
BANSHEE
NORMAL
CONFIGURATION
CLASS I, DIVISIO N
2 CONFIGURATION
WARNING
"!
Explosion Hazard: Substitution of components may impair
suitab ility for Cl ass I, Divisio n 2.
Explosion Hazard: Do not disconnect equipment unless power has
been switched OFF or the area is known to be non-hazardous.
75 db at five feet
95 db at five feet
100 db at three fee t
Gener al-purpose uni t
Unit can be configu red to meet
Class I , Division 2 Areas; thi s
configuration will not have
relays, pu mps, batteries,
autoc alibration, o r horns as optio ns
(it will have the st andard
Piezo a larm buzzer)
Installation Category II
Pollution Degree 2
4-3
Page 81
Chapter 5, Mainten ance
Chapter 5
Maintenance
General
The Toxgard II Monitor may be cleaned with a dry soft cloth and
requires minimal periodic maintenance to assure trouble-free
operation. However, the response to gas must be checked
regularly to assure the instr ument is functioning.
NOTE:The only way to verify an instrument is functioning
properly is to perform a calibration test. This test
does not necessarily mean the unit must be
adjusted, but that it must be checked with ZERO
and SPAN gas, and observed for
acceptable readings
The more frequent the calibration t ests, the greater the chanc es
for early detection of instrum ent failure resulting from instrument
component failure, damage or power loss.
Maintain a written log of instrument readings obtained during
calibration. Increase the frequenc y of calibration tests when the
sensor nears end-of-service life, or when abnormally high
variations with respect to the instrument log are observed.
The only routine maintenance item is the s ensing element itself,
which has a limited lifetime. When the Toxgar d II Monitor’s
read-out indicates that the sensor must be changed, there is ver y
little sensor lifetime remaining. It is good practice to obtain a
replacement sensing element before the sens ing element within
your unit becomes inoperative.
The Toxgard II Monitor is constantly per forming a self-check.
When a problem is found, it displays the appropriate error code.
(Table 5-1, "Troubleshooting Guidelines"). When an er ror is
detected within the unit, the mA output signal falls to 0 mA.
• The "Change Sensor" indication is not an error and does not
affect the output.
Corrective Maintenance
See TABLE 5-1, "Troubleshooting" to ident ify an inoperative
assembly in the Toxgard II monitor; then, replace it according to
the following procedures.
5-1
Page 82
Chapter 5, Mainte nanc e
Sensor Replacement
Typically, the Toxgard II Monitor LED display shows a
maintenance message when the sensor is due f or replacement
(FIGURE 5-1).
1. To replace the sensor, remove power from the
Toxgard II Monitor.
WARNING
"
Before removing the cover of an explosion-proof Toxgard II Monitor,
sampl e th e are a ar ound t h e enc l os ure wi th a p or tab le c om bus t ibl e
gas indicator (such as the MSA Passport
sure the surrounding area does not contain a flammable mixture of
combustible gas and air. When the enclosure is open and connected
to power, a source of ignition exists which could cause the explosion
of combustible gas, resulting in damage, serious personal injury
and death.
Handle the sensor carefully; the electrochemical version is a sealed
unit which contains a corrosive electrolyte. If electrolyte is leaking
from the sens or, exe rci se CA UTION to ens ure t he elec trol yte d oes
not co ntact sk in, eyes or clot hing, th us avoid ing bur ns. If co ntact
occur s, rinse th e area imme diately w ith a larg e quantit y of water.
In case of contact with eyes, immediately flush eyes with plenty
of water for at least 15 minutes. Call a physician.
CAUTION
"
Do not install a lea king sensor in the sensing head assembly. The
leaking sensor should be disposed of in accordance with local,
state and federal laws. To obtain a replacement sensor, contact
MSA at the address given under "Obtaining Replacement Parts."
2. There is no need to open the main or remote enclosure;
simply unscrew the cap located on the bottom of the
sensor assembly (FIGURES 5-2 through 5-4). For
combustible versions, unscrew the gas-tight union
(see FIGURE 1-6).
®
Personal Alarm) to make
5-2
Figure 5-1.
Change Sensor Flag
Page 83
Chapter 5, Mainten ance
Figure 5-2.
Cap Location for Remote Combustible Gas Explosion-proof Sensor
Cap Location for Toxic Gas or Oxygen General-purpose Model
Figure 5-3.
5-3
Page 84
Chapter 5, Mainte nanc e
Figure 5-4.
Cap Location for Remote Toxic Gas or Oxygen Explosion-proof Sensor
3. Identify the sensor element needed and obtain the same
type of sensor element.
4. Toxic or Oxygen Sensors:
Unplug the sensor by pulling it straight down; properly
dispose of sensor.
Combustible Sensors:
• Combustible units have a three-pin connector; pry out the
locking clip and pull apart the connector (FIGURE 5-5).
5. Toxic or Oxygen Sensors:
Before installing the sensor, check the pins of the sensor
for a shorting clip; remove the shorting clip and the label
beneath the clip before installing the sensor. Install the
new sensor by aligning the pins; do not force onto pins.
Combustible Sensors:
Connect the three-pin connector on the new sensor
(FIGURE 5-5). Carefully reconnect the union, taking care
not to pinch the wires.
CAUTION
"!
Do not over-tighten the sensor on the inlet fitting; otherwise, damage
to the sensor may result.
8. Re-apply power to the Toxgard II Monitor.
9. Perform the INITIAL Calibr ation Procedure as outlined in the
Toxgard II Monitor Calibrator or Toxgard II Monitor
Controller instruction manual. This procedure differs
slightly from a standard calibration procedure. Use of the
INITIAL calibration procedure upon installation of a new
sensor ensures the Toxgard II Monitor will perform a
successful calibration. Failure to do so may result in a
CAL FAULT reading (FIGURE 5-6).
5-4
Page 85
Chapter 5, Mainten ance
Figure 5-5.
Combustible Three-Pin Connector
Figure 5-6.
CAL FAULT Reading
It is recommended that all other m aintenance be performed at an
MSA factory-authorized service c enter. See TABLE 5-1 for
troubleshooting guidelines.
5-5
Page 86
Chapter 5, Mainte nanc e
Table 5-1. Troubleshooting Guidelines
SYMPTOMPOSSIBLE CAUSECORRECTIVE ACTION
Unit will
not
power
up
properly.
Display
shows
CAL
FAULT.
Display
shows
CAL.
Display
flashes
Err 1*.
Display
flashes
Err 2.
Display
flashes
Err 3.
Display
shows
Err 4.
*Hydrogen Cyani de, Hydrogen Chloride and Sulfur Dioxide Sensor s
NOTE: This NOTE and the following CAUTION apply to the Toxgard II Series
sensors for Hydrogen Cyanide (HCN), Hydrogen Chloride (HCL) and Sulfur
Dioxide (S O
detection function to alert the user of a sensor disconnection problem by
displaying "ERR 1" on th e LCD display .
The sensor detection function is NOT present in the HCN, HCL, SO2 or
100 ppm H
SO2 sensor disc onnection will NOT be i ndicated by an "ERr 1 " display. A
periodic check with a known conce ntration of cali bration ga s must be
performed regula rly to ensure proper sens or function. See "Cali bration"
portion of Cha pter 2 in this manual for c alibration instructions .
Insufficient voltage or
curre nt from powe r supply.
Incorr ect wiring connections .
Wire str ands touchi ng at
input terminal block.
Cell sig nal too low or t oo
high during calibration.
Unstab le calibratio n signal.
Check power supply and
interconnecting wires.
Check c alibration g as and flow sys tem;
then, re calibrate (valid calibration or
Initial Ca libration will clear display).
Replace cell and perform Initial
Old or d amaged cell .
Failed autocal calib ration
Sensor cell replacement
was det ected or dis play
was factory-set.
Sensor c ell detecte d as
missing.
Damaged sensor body.
Calibration.
(If orig inal cell is ne w, an Initial
Calibration alone may solve the
problem.)
If the unit i s incorrectly set u p for
autocal, disable the autocal mode by
setting th e future calibratio n date to
12-31-1993 or check the autocal flow
system.
Perform Initial Cal ibration if un it is
new or cell was replaced.
Perform regular cal ibration if sa me cell
was re-installed.
Inspect cell for proper placement.
After recovery, see abov e SYMPTOM:
"Display shows CAL".
Replace sensor body and perform
Initial Calibration.
Invali d cell type.Replace or repair mai n board.
Non-volatile memory not
detected .
Non-volatile memory reset
or not in itialized.
Cycle power or repair main board.
Perform Initial Calibrat ion.
(Cell type changed.)
Damaged sensor or
sensor body
) only. T he Toxgard I I Series produ ct line contai ns a sensor
2
S version of the Toxgard II series. Therefore, an HCN, HCL or
2
Replace sensor or sensor body and
perfor m initial calib ration.
CAUTION
"!
5-6
Page 87
Chapter 5, Mainten ance
Table 5-1. Troubleshooting Guidelines
SYMPTOMPOSSIBLE CAUSECORRECTIVE ACTION
Display
flashes
Err 6.
Output
signal
not
present.
Display
shows
SENSOR.
No
response
to
manual
ZERO or
CAL
command.
Display
shows
UND.
Low gas
reading.
Display
shows
LO FLO
Battery power exhaustedConnect to line power for recharg e.
Wiring errorCheck connections.
Initial Calibration was not
performed.
Sensor c ell missing.
Sensor s ignal is
under-range.
Error detected.
Sensor c ell output i s below
acceptable signal level.
Unit is set up for
AUTOCAL use, and a utocal
unit or gases are not
present.
Cell signal is under-range.
Clogge d sensor wat er stop
disc.
Misal igned stop dis c.Re-align disc a nd gasket.
Poor gas flow or damaged
cell.
Rest ricted air flow .Clear tubing.
Pump in operative.Repla ce pump.
Perform Initial Cal ibration.
See SYMPTOM: "Displa y flashes Err
1".
See SYMPTOM: "Displa y shows UND".
See "Sym ptom" co lumn for sp ecific
error.
Replace sensing cell and
perform an Initial Calibration.
De-ac tivate AUTOCA L mode by se tting
Future Ca libration date to 1 2-31-1993.
Perform regular zero or full calibration.
If calibration fails, see above
SYMPTOM: "Display shows CALFAULT". If UND persists, replace cell
and perform Initial Calibra tion. NOTE :
Although sensor output is not present
during UND condition with no gas
present, output will correctly respond to
the presence of gas. When gas is no
longer present, output will no longer be
present. This function serves to alert the
user that an abnormal condition exists
and calibration or other maintenance is
required.
Clean or replace disc .
See SYMPTOM "Display shows CALFAULT".
5-7
Page 88
Chapter 5, Mainte nanc e
Obtaining R epl acem en t Part s
See TABLE 5-2 for replacement sens or kits. To obtain a
replacement sensor, address the order or inquiry to:
Mine Safety A ppliances Company
Instrument Division
P.O. B ox 427, Pittsbu rgh, PA 1523 0-0427
or call, toll-free, 1-800-MSA-INST.
WARNING
"
Use only genuine MSA replacement parts when performing any
maintenance procedures provided in this manual. Failure to do so
may seriously impair sensor performance. Repair or alteration of
the Toxgard II Monitor, beyond the scope of these maintenance
instructions or by anyone other than authorized MSA service
personnel, could cause the product to fail to perform as designed
and persons who rely on this product for their safety could sustain
seriou s personal inj ury or death.
Contat ct 1-800-MSA- INST for fac tory support in determin ing the appr opriate part
710225
812075
710436
812152
numbers for explosi on-proof or du ct mount S ensorgard s
5-9
Page 90
Chapter 5, Mainte nanc e
ITEMPART NO.
Main printed circuit board
Table 5-3. Spare Parts List
NOTE:
Order this replacement board according
to info rmation prov ided on the la bel of
the ex isting board . This label sh ould
read:
"A-TOX-PCB-X-Y-Z"
(X-Y-Z corresponds to the
first, sixth and seventh option code
positi ons found on the namepl ate of the
main unit)
Contac t 1-800-MS A-INST for
factory -support in d etermining the
appropriate part number.
Display printed circ uit board
Plastic Support, display printed circuit
board
Strobe, red
Strobe, gasket
Battery
Power supply
Piezo Bu zzer
Horn, federal
Pump kit, DC
Pump kit, asp irated
Air line f low assembl y
Auto cal assembly
Valve assembly, general-purpose
Valve assembly, explosion-proof
The Remote Sensor Duct Mount Kit (FIGURE A-1) provides
convenient sampling of gas within an air duct.
• This Kit enables duct gas sampling even when flow rate within
the duct is at a high velocity.
• The user can calibrate the sensor via a calibration port without
disassembling or removing the sensor from the duct.
• The sensor and the Kit are sealed to the duct to prevent gas
within the duct from escaping.
• All Toxgard II remote sensors fit into the remote sensor model
of the Duct Mount Kit.
The Remote Sensor Module Duct Mount Kit:
• A llows the sensor portion of the unit to be mounted on the duct
(perhaps in an inaccessible location), while the read-out and
calibration portion are placed in an easily accessible location to:
Figure A-1.
Ultima Duct Mount Kit
A-1
Page 92
Appendix A, Optional Toxgard II Duct Mount Kit
• save valuable maintenance time and money
• enhance calibrator safety.
Mounting
General
Mounting the Remote Sensor on a duct is easily done with the
Duct Mount Kit. The proper location on the duct will permit your
sensor to operate up to its pot ential.
Choose a location on the duct:
• Where the duct is flat or level:
• The Duct Mount Kit will not mount on a diameter of less
than eight inches (203 mm) of rounded or
curved duct work.
• Additional gasketing may be necessary to properly seal
the unit.
• A s far away from any bends or tees as possible, as those
locations may have additional turbulence
• Where reasonable access is possible to enable
sensor replacement
• That keeps the sensor pointing downward:
• It is good practice to mount the sensor on horizontal run
of the duct.
• Mount the sensor on top of the duct to allow water and
dirt to naturally drip off the sensor within the duct and
prevent clogging.
• That allows the dual module sensor calibration gas tubing to be
routed to the read-out module. Any gas-compatible tubing can
be used (not supplied in Kit).
• Keep this tubing as straight and short as possible.
• Beyond any filter or cleaning unit.
• Away from the lowest point on the duct, which tends to collect
additional dirt and dust.
• Away from any fans or blowers.
The Duct Mount Kit requires some as sembly before mounting. It
is necessary to temporarily remove the sensor from the module
housing and install the sealing gaskets. A ll necessary hardware is
provided in the Kit.
A-2
Page 93
Appendix A, Optional Toxgard II Duct Mount Kit
Figure A-2.
Duct Template
Mounting
1. Locate the duct to be sampled.
2. Cut a rectangular hole [8.875 by 2.7 inches (225 by 68.5
mm)] in th e d u c t (FIGURE A-2) .
• The gasket on the bottom of Duct Mount Kit must sit flat
on the duct.
• If necessary, pre-drill the fastener holes.
• Do not remove the gasket paper lining on the Duct
Mount Kit at this time.
• On rounded ducts, additional gasketing may be required
for proper sealing.
3. Locate and identify all components in the Duct Mount Kit
(TABLE A-1).
A-3
Page 94
Appendix A, Optional Toxgard II Duct Mount Kit
Figure A-3.
Remote Sensor Spacers
Table A-1 . D ua l m odule Duct Mount K it Parts List
DESCRIPTIONPART NUMBERFIGURE
Two Tox gard II Gas Mo nitor Spacer s
Toxgard I I Gas Moni tor Clamp
Toxgard I I Gas Moni tor Enclosur e Gasket
Toxgard I I Gas Moni tor Plate
Toxgar d II Gas Moni tor Duct Mo unt Kit Bas e
Plate
Two 1/4-inch Hex Nuts
Two 1/4-inch Lock Washers
Six 10-32 Screws
Six #10 Lock Washer s
710779A-12
710778A-13
710770A-6
710769A-7
814862A-14
631357
655245
634054
627852
4. Locate the Remote sensor of the Toxgard II Gas Monitor to
be installed in the duct.
5. Unscrew the cover of the Remote sensor and set aside.
Figure A-4.
Remote Sensor Clamp
A-4
Page 95
Appendix A, Optional Toxgard II Duct Mount Kit
WARNING
"
Before removi ng the co ver of an explo sion-pro of Remote Sensor
Toxgard II Gas Monitor, verify that the surrounding area does not
contain a flammable mixture of combustible gas and air, since a
source of ignition is exposed; otherwise, an explosion may occur
if a metal object contacts the circuitry and produces sparks.
6. For Toxic or Oxygen Remote Sensor Toxgard II Gas
Monitors:
To disconnect the sensor wiring on the termination printed
circuit board in the module, pull up on the connector
inserted into J1 (FIGURE 1-8).
For Combust ible Remote Sensor Toxgard II Gas Monitor:
To disconnect the sensor wiring on the termination printed
circuit board in the module, unscrew or loosen the screws
on the terminal wiring connector.
• Befor e removing the wires, mark the identity of each wire
to permit proper replacement (FIGURE 1-7).
7. To avoid damage, completely remove the black SensorGard
by unscrewing the black plastic ring.
8. Using a strap or pipe wrench, carefully unscrew the sensor
body from the module.
• Carefully remove sensor assembly from the sensor body.
• Ensure that wires and connector do not break when
removing the sensor body.
9. Place the enclosure plate and gasket from the Duct Mount
Kit on the sensor body
threads (FIGURE A-5).
10. Carefully feed the sensor
wires from the sensor
body back into the
module opening.
11. Using a strap or pipe
wrench, carefully screw
the sensor body back
into the module.
• Ensure that gasket
is centered and not
folded on the
sensor threads.
Figure A-5.
Sensor
A-5
Page 96
Appendix A, Optional Toxgard II Duct Mount Kit
12. For Toxic or Oxygen Remote Sensor Toxgard II Gas
Monitors:
Re-connect the sensor wiring on the termination printed circuit
board in the module by aligning the connector and
inserting it into J1 (FIGURE 1-8).
For Combust ible Remote Sensor Toxgard II Gas Monitor:
Re-connect the sensor wiring onto the wiring termination
printed circuit board in the module (FIGURE 1-7).
13. Re-install the cover of the Remote Sensor Toxgard II
Gas Monitor.
WARNING
"
Do not let the cover remain off of an explosion-proof Remote Sensor
Toxgard II Gas Monitor. Since a source of ignition is exposed, an
explosion may oc cur if a metal object contacts the circuitry and
produces sparks in an atmosphere of combustible gas.
14. Locate the Remote sensor Toxgard II Gas Monitor base
plate and two #10-32 screws and lock washers.
15. Remove the four screws holding the module mounting
bracket (FIGURE A-6).
A-6
Figure A-6.
Remote Sensor Base Plate
Page 97
Appendix A, Optional Toxgard II Duct Mount Kit
Figure A-7.
Duct Mount and Sensor
16. Fasten the Remote Sensor module to the bracket using
two 10-32 screws and lock washers.
• These screws pass through the bracket and into two
tapped holes in the module.
17. Push the Remote sensor Toxgard II Gas Monitor
through the slit gasket on the Duct Mount base plate
(FIGU R E A -7 ).
18. Fasten the sensor module bracket to the base plate using
the 10-32 screws removed in step 15.
• Do not tighten completely.
19. Fasten the calibration tubing fitting into the center post of
the SensorGard (FIGURE A-7).
A-7
Page 98
Appendix A, Optional Toxgard II Duct Mount Kit
20. Locate the two Remote sensor Toxgard II Gas Monitor
spacers. Screw one end of each spacer into the Duct
Mount Kit base plate.
21. Place the Remote sensor Toxgard II Gas Monitor clamp on
the top of the spacers.
• Ensure the folded part of the clamp is mounted on the
back side of the Remote Sensor Toxgard II Gas Monitor.
22. With the 1/4-inch hex nut and lock washers, secure the
clamp onto the spacers (FIGURE A-7).
• Tighten the #10-32 screws that secure the module
mounting bracket to the base plate.
23. Clean the area around the hole made in step 2.
• Ensure that no grease, metal shavings or other material
is in the area where the gasket will be seated.
24. Remove the protective paper backing from the gasket
found on the bottom of the Duct Mounting flange.
25. Install the entire assembly on the duct. Use proper
fasteners to secure the Toxgard II Gas Monitor Duct
Mount to the duct.
26. If necessary, tape or seal around the sensor gasket to
prevent leaks.
27. Remove the top half of the calibration fitting and connect it
to your calibration tubing.
• When calibrating, push the calibration fitting into its mate.
28. Install the read-out module by following the mounting
instructions in the instructions manual for the single
module model.
29. Refer to the instruction manual for wiring details.
A-8
Page 99
Appendix B, Definit i on of Ter ms
Appendix B
Definitions of Terms
The following defines terminology used in this m anual that may
be unfamiliar to some users:
Zero: A zero (0) indication on the meter display indicates an
ambient air present, or no hazardous gases present.
Zeroing: The process of def ining a zero condition on the display.
Span: An up-scale reading on the meter display
Spanning: The process of defining an up-scale reading on
the display.
SPAN Gas Value: The gas concentrati on which gives the
instrument an up-scale or full-scale value. This value is printed
on the cylinder containing the gas.
Exhaust G as: The gas after it passes the sensor.
Flow Rate: T he speed of the gas in the sample line.
Pump: An electromechanical device which mov es the gas
sample to the Toxgard II inst rument.
Upscale: Instrument alarms on gas concentrations greater than
or equal to the alarm value.
Downscale: I nstrument alarms on gas concentrations less than
or equal to the alarm value.
Non-latching: A larms reset automatically when the alarm
condition clears.
Latching: Alarm relays and indicator lamps remain activ e (after
alarm condition clears) until they are manually de-activated by
pressing the HORN/ALARM ACKNOWLEDGE button.
B-1
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.