MSA Chillgard LC Instruction Manual

Page 1
Chillgard® LC
Control Module
Instruction Manual
WARNING
"!
THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HA VE THE RESPONSIBILITY FOR INSTALLING, USING, MAINTAINING OR SERVICING THIS PRODUCT. Like any piece of complex equipment, this prod uct will perform as designed only if installed, used and serviced in accorda nce with t he manufact urer’s in structi ons. OTHE RWISE , IT COULD FAIL TO PERFORM AS DESIGNED AND PERSONS WHO RELY ON THIS PRODUCT FOR THEIR SAFETY COULD SUST AIN SEVERE PERSONAL INJUR Y O R DEATH.
The warranties made by Mine Safety Appliances Company with respect to this product are voided i f t he p r odu c t is n ot i nst a lle d , us ed an d s er vi ce d in a cc o rd ance w it h th e ins t ru cti o ns in this manual. Plea se protec t yourself and others by follow ing them. We encourage our customers to write or call regarding this equipment prior to use or for any additional information relative to use or repairs.
MSA North America 1-800-MSA-INST or FAX (724) 776-8783 MSA International (412) 967-3354 or FAX (412) 967-3451 In Cana da 1-800-267-0672 or FAX (416) 663-5908
© Mine Safety Appliances Company 2008 - All Right s Reserved This manual is available on the internet at www.msanet.com
Manufactured by
MSA NORTH AMERICA
PITTSBURGH, PENNSYLVANIA 15230
(L) REV 4 I MZ001-036 10036865
Page 2
MSA
Permanent Instrument Warranty
1. Warranty- Seller warrants that this product will be free from mechanical defect or faulty workmanship for a period of two years from date of shipment, provided it is maintained and used in accordance with Seller’s instructions and/or recommendations. This warranty does not apply to expendable or consumable parts whose normal life expectancy is less than one (1) year such as, but not limited to, non-rechargeable batteries, filament units, filter, lamps, fuses etc. The Seller shall be released from all obligations under this warranty in the event repairs or modifications are made by persons other than its own or authorized service personnel or if the warranty claim results from physical abuse or misuse of the product. No agent, employee or representative of the Seller has any authority to bind the Seller to any affirmation, representation or warranty concerning the product. Seller makes no warranty concerning components or accessories not manufactured by the Seller, but will pass on to the Purchaser all warranties of manufacturers of such components. THIS WARRANTY IS IN
LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IM PL IED O R STAT UT ORY, AND IS STRICTLY LIMITED TO THE TERMS HEREOF. SELLER SPECIFICALLY DISCLAIMS ANY
WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE.
2. Exclusive Remedy- It is expressly agreed that Purchaser’s sole and exclusive remedy for breach of the above warranty, for any tortious conduct of Seller, or for any other cause of action, shall be the repair and/or replacement at Seller’s option, of any equipment or parts thereof, which after examination by Seller is proven to be defective. Replacement equipment and/or parts will be provided at no cost to Purchaser, F.O.B. Seller’s Plant. Failure of Seller to successfully repair any nonconforming product shall not cause the remedy established hereby to fail of its essential purpose.
3. Exclusion of Consequential Damage- Purchaser specifically understands and agrees that under no circumstances will seller be liable to purchaser for economic, special, incidental or consequential damages or losses of any kind whatsoever, including but not limited to, loss of anticipated profits and any other loss caused by reason of nonoperation of the goods. This exclusion is applicable to claims for breach of warranty, tortious conduct or any other cause of action against seller.
Page 3
Chillgard LC Control Module General Warnings and Cautions
General Warnings and Cautions
WARNING
"!
1. The Control Module described in this manual must be installed, operated, and maintained in strict accordance with the labels, cautions, warnings, instructions, and within the limitations stated.
anyone other than authorized MSA service personnel, could cause the product to fail to perform as designed, and persons who rely on this product for their safety could sustain serious personal injury or death.
2. The Control Module must not be installed in outdoor areas or in locations where explosive concentrations of combustible gases or vapors might occur in the atmosphere: Class 1, Group A, B, C, and D areas as defined by the NEC. Because the Control Module is not explosion-proof, it must be located in non-hazardous areas.
3. Use only genuine MSA replacement parts when performing any maintenance procedures provided in this manual. Failure to do so may seriously impair instrument performance. Repair or alteration of the Chillgard LC System, beyond the scope of these maintenance instructions or by
Failure to comply with the above warnings can result in serious personal injury or death.
4. The Chillgard LC Control Module must be installed, located and operated in accordance to all applicable codes. These codes include, but are not limited to, the National Fire Prevention Code and National Electric Code.
5 Protect the Chillgard LC Control Module from
vibration and heating; otherwise, improper operation may result, which can result in personal injury or death.
6. Do not exceed the relay contact ratings listed in Section 1, TABLE B-1 Otherwise, relay operation may fail, which can result in personal injury or death.
Page 4
Chillgard LC Control Module Table of Contents
Table of Contents
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Table 1-1. Chillgard LC System General Operating Specifications . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Figure 1-1. Chillgard LC Control Module Standard Version
(24 Volt AC/DC or 110/220 AC Version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Figure 1-2. Chillgard LC Control Module With Internal Transformer . . . . . . . . . . . . . . . . . . . . . 1-1
Termi n o l o g y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Connection of the Chillgard Instruments Using RS-485 Communications . . . . . . . . . . . . . . . . . . . . . . . 1-3
Multiple Sensor Modules with a Chillgard LC Control Module (Daisy Chain). . . . . . . . . . . . . . . . . . 1-3
Multiple Sensor Modules with a Chillgard LC Control Module (Two Branch). . . . . . . . . . . . . . . . . . 1-3
Figure 1-3. Connection of the Sensor and LC Instruments Using RS-485 Communications . . 1-3
Figure 1-4. Multiple Sensor Modules with a Chillgard LC Control Module (Daisy Chain) . . . . . 1-4
Figure 1-5. Multiple Sensor Modules with a Chillgard LC Control Module (Two Branch) . . . . . 1-4
Section 2
Insta llation and Set-u p . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Recei vi n g . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Unpac ki n g th e S ystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
WARNIN G. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
"
Initi a l In sp e c t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Location of the Control Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
" WAR N ING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Figure 2-1. Basic LC Control Module with Door Open . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Wiring Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Opening the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
" CA U TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Unit Po w e r Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
"!CAUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
" WAR N ING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
" CA U TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Tabl e 2 -1 . Option S w i tc h Ta b l e. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2- 2
Figure 2-2. Transformer Version LC Control Module with Door Open . . . . . . . . . . . . . . . . . . . 2-2
Figure 2-3. Standard Version Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Figu re 2-4. Tra n s fo r m e r V e rsion Wiring D i a g ra m . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Wirin g to th e Ch i l l g a r d Se n s o r s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Addressing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
" WAR N ING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Tabl e 2 -2 . C h i l l g a r d D ip Swi tch Pos i t i o n s. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Figure 2-5. Chillgard LS Sensor Board Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Figure 2-6. Chillgard M-100 Jumper Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Analog Signal Output Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
"!
Page 5
Table of Contents Chillgard LC Control Module
Relay Ou t p u ts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2- 7
"
Start-Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Setup Date, Time, and Logging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
CAUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2- 7
"
Initi a l Cal i b ration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
WARNIN G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
"
Section 3
Displa y Sc reens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Figure 3-1. Display Screen Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Figure 3-2. Start-up and Normal Operation Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Figure 3-3 Set-up Screens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Figure 3-4 Span Calibration Screens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Figu re 3-5 Zero C a li b r a ti o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3- 4
Figu re 3-6 Cali b ra t i o n Ch e ck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Figure 3-7 Diagnostics Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Figure 3-8 Data Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3- 6
Figu re 3-9 Alar m L e ve l Sc re e n. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Figu re 3-10 Set u p Al a r m L a tching . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Figu re 3-11 Set u p Al a r m R e l a ys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Figu re 3-12 Audi o A l a rm Scree n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Figure 3-13 Analog Outputs and Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Figu re 3-14 Set u p Ti mes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Figure 3-15 Setup Logging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3- 9
Figu re 3-16 Revi e w L o g. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Section 4
Mainte na nce. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
General Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Obtai n i n g R e p l a ce m e n t P a rts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
"!WARNING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Tabl e 4 -1 . R e p l a c e ment Pa rt s. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
" WAR N ING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4- 1
Table 4-2. Troubleshooting Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Section 5
Data L oggi ng. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Numbe r of Data Poi n ts Used. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
User Se t u p Op ti o n s. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Viewing Alarms/Events and Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Figure 5-1. Log Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Section 6
Calibr a t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Diffu si o n V e rs i o n C a l ib r a t i o n. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Calib r a t i o n Equ i p ment - (FIGUR E 6 - 1 ). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
"!WARNING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Page 6
Chillgard LC Control Module Table of Contents
Initial Calibration Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Figu re 6-1. Cal i b ra ti o n E q u i p ment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
" CA U TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
" CA U TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Calib r a t i o n Gu i d e l i n e s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Calibration Check Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
"!WARNING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Figure 6-2. Applying Calibration Gas to the Chillgard LS Refrigerant Monitor. . . . . . . . . . . . . . 6-2
" CA U TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Single or Four-Point Pumped Version Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Calib r a t i o n Equ i p ment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
" CA U TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Tabl e 6 -1 . R P C a l i b ra ti o n G a s e s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
" WAR N ING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Calibration Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
" CA U TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
" CA U TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Calib r a t i o n Gu i d e l i n e s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Figure 6-3. Kit Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Figure 6-4. Using Zero Scrubber for Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Figure 6-5. Using Zero Gas Cylinder for Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Calibration Check Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
" WAR N ING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
" CA U TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Figure 6-6. Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
" CAU T ION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Appendix A
RS-232 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A- 1
RS-232 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Table A-1. RS-232 Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Table A-2. Information Structure (Chillgard LC Control Module) . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Tabl e A- 3 . Da t a St ru c t u r e. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Tabl e A- 4 . Ala rm Stru c tu re . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Table A-5. Alarms and Events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Appendix B
Insta llation Ou t line Dra w ing. . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Figure B-1. Installation Outline Drawing for Standard Version . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Figu re B -2 . In stall a t i o n Ou tl i n e D ra w i n g fo r T r a n sformer Versi o n. . . . . . . . . . . . . . . . . . . . . . . B-3
Page 7
Chillgard LC Control Module Section 1, General Information
Section 1 General Information
Introduction
General Description
The Chillgard LC unit is a control module which can communicate with up to eight Chillgard LS or twelve Chillgard M100 Refrigerant Sens ors. The Chillgard LC Control Module can be installed remotely from the sensors, thus enabling remote control of that Chillgard LS or the Chillgard M100 Sensor.
The Chillgard LC comes in two major versions based upon the power supply in the unit:
24 Volt AC/DC or 110/220 VAC version
Internal Transformer version - 120 Volt input,
24 VAC output for powering Remote Sensor Modules.
See FIGURES 1-1 and 1-2 to identify your unit.
Table 1-1. Chillgard LC System
General Operating Specifications
OPERATING SPECIF ICATIONS
100 to 240 VAC, 50-60 Hz, 25 W
VOLTAGE RATING
POWER REQUIREMENTS
TROUBLE RELAY
ALARM 1 RELAY
ALARM 2 RELAY
ALARM 3 RELAY
24 VDC +10% -0% 24 VAC Transforme r Version: 120 V, 60 Hz, 75 VA
.15 am ps at 120 VAC; .10 am ps at 240 VAC
0.6 am ps at 24 VAC
0.1 am ps at 24 VDC Minimum wire size: #18 AWG The LC Controller can power the Chillgar d LS and seve n Chillgard M-100 s ensors
Normally energized, Form C contact: 240 VAC, 5 amp resistive SPDT
One relay, Form C contacts: 240 VAC, 5 amps resistive SPDT
One relay, Form C contacts: 240 VAC, 5 amps resistive SPDT
One relay, Form C contacts: 240 VAC, 5 amps resistive SPDT
+10%
Figure 1-1. Chillgard LC Control Module
Standard Version
(24 Volt AC/DC or 110/220 AC Version)
ANALOG OUTPUTS
MAXIMUM OUTPUT SIGNAL LOAD FOR 4-20 mA OUTPUT
4 to 20 mA sourcing, 500 ohm load , 0-6 VDC, 2 K ohm load
600 ohms (includes wiring)
Figure 1-2. Chillgard LC Control Module
With Internal Transformer
1-1
Page 8
Section 1, General Information Chillgard LC Control Module
Table 1-1. Chillgard LC System
General Operating Specifications
OPERATING SPECIF ICATIONS
AUDIO ALARM DRIVE OUTPUT
OPERATING TEMPERATURE
STANDARD VERSION DIMENSIONS
STANDARD VERSION WEIGHT
TRANSFORMER VERSION DIMENSIONS
TRANSFORMER VERSION WEIGHT
TRANSPORT AND STORAGE CONDITIONS
TEMPERATURE -40 HUMIDITY 99% RH non-condensing POLLUTION
DEGREE AND INSTALLATION CATEGORY
24 VDC 50-ohm load maximum (availa ble only with non-beacon models)
o
0 to 40
C (32 to 104oF)
7.5" high, 12.5" wide, 3.5" deep
19.05 cm high, 31.75 cm wide,
8.89 cm deep
3.5 pounds (1.59 kilograms)
14.7" high, 11.2" wid e, 5" deep 373 mm high, 284 mm wide,
127 mm deep
8.4 pounds (3.82 kilograms)
o
C to +60oC (-40oF to 140oF)
2
ZEROING- The process of placing a zero gas on
the unit during calibration.
SPAN- Full-scale or up-scale reading on
meter display.
SPANNING- The process of placing a full-scale or
span gas on the unit during calibrati on.
SPAN GAS VALUE- The gas concentration that
gives the instrument a full-scale or up-scale value. This value is printed on the calibration gas cylinder containing the gas.
FLOW RATE- Volume of gas drawn through the
sample line per minute.
EXHAUST GAS- Sample gas after it passes
through the sensor.
PUMP- The electric motor driven dev ice that
moves the gas sample to the refriger ant monitor.
ALARMS- The Chillgard LC System has three
alarms to alert the user at specific, user-adjustable refrigerant gas concentrat ions. (Level 1 = Caution, Level 2 = Warning and Level 3 = Alarm.)
RELATIVE HUMIDITY- The percent of water vapor
saturation in air at a given t emperature.
POINT NUMBER- The location or ar ea from which
a gas sample is drawn. Up to four , eight, or 12 areas, sequentially numbered from 1 to 12, can be sampled.
Terminology
Become familiar with the following terminology.
ZERO- A zero (0) indication on the meter
display usually indicates fresh air ( no refrigerant gas present).
FRESH AIR- Air that has no possibility of
containing refrigerant gas.
TEMPERATURE EFFECT- The gas response
displayed by the instrument (PPM) can change +3 ppm for each degree (C) that the instrument is operating above/below the temperature at which the instrument was last calibrated.
1-2
Page 9
Chillgard LC Control Module Section 1, General Information
Connection of the Chillgard Instruments Using RS-485 Communications
The Chillgard LC instrument acts as t he Master device in the network. The LC inst rument initiates and controls all communications on the network. The LC instrument sends out commands and requests for data from individual sens ors. When the instruments are powered up, the LC unit automatically detects the number and gas type of all instruments on the network.
In a simple system using a single sens or module and an LC Control module, the two m odules are connected together; both units hav e the RS-485 terminators "IN" (FIGURE 1-3). Ref er to Section 2, Installation and Setup, for s etting the address switches and FIGURES 2-2 and 2-3 f or the location of the termination jumpers.
NOTE:Each sensor must have a different address.
The total length of the RS-485 cable cannot exceed 1000 feet (304 meters). MSA recommends Belden #9841 low-capacitance cable.
Multiple Sensor Modules with a Chillgard LC Contro l Mo du le (Daisy Chain)
See FIGURE 1-4. NOTE:Up to eight Chillgard LS Sensors can be
connected to a Chillgard LC Controller. (They must be addressed as Points 1 through 8.)
LS1 devices can be on points 1 through 8.
LS4 devices must be on point 1 and/or 5
and will also use the next three physical addresses.
LS4 on point 1 takes addresses 1, 2, 3, and 4.
LS4 on point 5 takes addresses 5, 6, 7, and 8.
LS4 cannot be on addresses 9, 10, 11, or 12.
Multiple Sensor Modules with a Chillgard LC Contro l Mo du le (Two Branch)
See FIGURE 1-5. Each end of the net work must have the terminator IN; modules in t he middle must have terminators OUT. RS-485 connec tions from the Chillgard LC Control unit mus t have no more than two terminations.
Figure 1-3. Connection of the Sensor and LC
Instruments Using RS-485 Communications
1-3
Page 10
Section 1, General Information Chillgard LC Control Module
Figure 1-4. Multiple Sensor Modules
with a Chillgard LC Control Module (Daisy Chain)
with a Chillgard LC Control Module (Two Branch)
1-4
Figure 1-5. Multiple Sensor Modules
Page 11
Chillgard LC Control Module Section 2, Installation and Set-Up
Section 2 Installation and Set-up
Receiving Unpacking the System
To unpack the equipment:
1. Carefully remove the Chillgard LC Control Module from its shipping container(s) in order to prevent damage to sensitive electrical components. If any damage is found, report it to the shipper immediately.
WARNING
"
Do not install or operate a damaged unit. It may not function properly and may not alert you to any gas conditions.
2. Search through all packing material and containers to avoid inadvertently discarding usable or valuable parts. Report any shortages immediately to MSA.
Initial Inspection
With the front door open, carefully inspect components and assemblies inside the enclosure. If damage or shortage is evident , advise and promptly file the proper claim with t he carrier.
Location of the Control Module
WARNING
"
Explosion Hazard! Unit must not be located in areas that may
contain a flammable mixture of gas and air; failure to follow this requirement can result in death, serious injury, or equipment or property damage.
Figure 2-1. Basic LC Control Module with Door Open
2-1
Page 12
Section 2, Installation and Set-Up Chillgard LC Control Module
Wiring Connections
Opening the Uni t
All wiring to the Chillgard LC unit is made via the bottom entries. Open the unit to provide complete access to all wiring connections.
Ensure that all wiring codes are followed. These codes include, but are not limited t o, the National Electrical Code.
CAUTION
"
Component Damage! Monitor components must be protected from
splashing, spraying, or dripping water. Failure to do so may cause damage to internal components.
Unit Power W ir ing
!CAUTION
"
Instr ument Damage! Correct powe r voltage mus t be connected to
the instrument. Failure to use cor rect voltage may res ult in instrume nt damage.
A separate, dedicated power sourc e is recommended for the refrigerant contr ol module to ensure that the unit remains power ed when other circuits are shut down for s ervicing, routine maintenance or shift changes. Supply ear thground is installed on lug first and all com ponent earthgrounds are connected afterwards .
Standard Version
The control module uses a wide range power supply which can accept AC power f rom 100 to 240 volts, 50 or 60 Hz. I f 24 Volt AC or DC is available, power can be supplied directly to the control board (FIGURE 2-3). The 24 V olt power source used with this equipment must be separated from mains by double or reinforc ed insulation. If the 110/220 Volt power opt ion is available, power can be supplied to terminal block as shown in FIGURE 2-1.
Transformer Version
If the Chillgard LC unit contains t he internal transformer, supply power as shown in FIGURE 2-4. The resulting 24 VAC output from the transformer can be used to power up to five diffusion LS units or two pumped f our-point LS units. However, voltage drop will occ ur over extended line lengths with multiple sensors. Ensure minimum required voltage is present at the sensor terminals for each sensor module.
Figure 2-2. Transformer Version LC Control Module
with Door Open
WARNING
"
If insufficient voltage is supplied, unit could display incorrect readings. check each sensor module for minimum required voltage at each unit.
With 18 AWG wire, a Chillgard LS Sensor Module can be located a maximum distance of 450 feet from the Chillgard LC Control Module. Two Chillgard LS four-point Sensor Modules daisy-chained from the Chillgard LC Contr ol Module can be connected a maximum of 150 f eet from the Chillgard LC Control Module. Wit h 18 AWG wire, seven Chillgard M-100 sensors can be daisy-chained from the LC Control Module to a distance of 1500 feet.
CAUTION
"
If the Chillgard LC transformer or other AC power is used to power the Chillgard M-100 Sensor, the Chillgard M-100 AC power must be isolated from the LC power for proper operation.
Table 2-1. Option Switch Table
SWITCH 1 CUSTOMER RS232 PORT SPEED SET
SWITCH 2 N/A SWITCH 3 N/A
19,200 BPS 8, N, 1 UP
9,600 BPS 8, N, 1 DN
2-2
Page 13
Chillgard LC Control Module Section 2, Installation and Set-Up
Figure 2-3. Standard Version Wiring Diagram
Figure 2-4. Transformer Version Wiring Diagram
2-3
Page 14
Section 2, Installation and Set-Up Chillgard LC Control Module
Wiring to t he Chil lgard Sens ors
The Chillgard LC Control Module can communicate to up to 12 sample points. Communication between these devices is achieved through RS-485 modbus protocol [see the Chillgard LS Instruc tion Manual (P/N 10035164) and/or the Chillgard M-100 Instruction Sheet (P/N 10073591) for definition of protocol]. This protocol allows for dais y-chaining of the sensor modules to the c ontrol module. See FIGURE 2-3 for wiring to the control m odule.
Addressing
Each sensor module must have a dist inct address. The sensor module address swit ch is used to specify the addresses of each sensor module (FIGURE 2-5 and 2-6). The addresses also determine the point number on the dis play.
WARNING
"
If the addresses of each sensor module are not unique to each unit, the control modul e may not recognize one or more of the sensor modules.
The dip switch positions are shown in TABLE 2-2.
Table 2-2. Chillgard LS Dip Switch Positions
POINT ID
1100 0 0 0 -­2101 1 0 0 -­3102 0 1 0 -­4103 1 1 0 -­5104 0 0 1 -­6105 1 0 1 -­7106 0 1 1 -­8107 1 1 1 --
1234
NOTE: 1 indicate s closed.
DIP SWITCH
For four-point sensor modules, there ar e two valid dip switch settings: 100 and 104.
If dip switch setting 100 is chosen, addresses 101 to 103 will be assigned to the four-point unit and will not be available for use on other units.
Figure 2-5. Chillgard LS Sensor Board Connections
2-4
Page 15
Chillgard LC Control Module Section 2, Installation and Set-Up
Figure 2-6. Chillgard M-100 Jumper Locations
2-5
Page 16
Section 2, Installation and Set-Up Chillgard LC Control Module
If dip switch setting 104 is chosen, addresses 105 to 107 will be assigned to the four-point unit and will not be available for use on other units.
Analog Signal Output Wi r i ng
The unit has two availa ble analog outputs. The 4 to 20 mA analog output is software-se lectable between 10% or 100% of full scale:
4-20 mA, isolated, current sourcing
0-6 VDC outputs, 0.5-Volt steps to indicate
point being monitored. For example, 1.5 V represents point 3 at the milliamp output.
Analog output wiring should enter the unit through the hole provided on the left side of the unit. The maximum wire size that these connectors can accept is Listed #12 AWG; t he maximum cable length is 500 feet (166 meters) .
It is suggested that Listed #18 AWG, twisted-pair wire be used. If shielded wire is nec essary, ground the shields of all cables at t he receiving end of the signal. Do not ground or connect the shields at the Chillgard LC Control Module. See FIGURE 2-2 for connecting the analog outputs.
CAUTION
"!
Bundle low voltage wiring together (lower than 30 volts), separate from high voltage wiring (higher than 30 volts).
Optional Alarm Beacon
Your unit may have an optional alarm kit installed. This kit is made up of a beac on on the top of the unit. It is factory-wired so no additional wiring is necessary; it lights when an Alar m 3 indication is given by the instrument. Connection is made to the strobe connector (FIGURE 2- 2). IP rating shown on the strobe light applies t o the strobe only.
Rela y Outputs
Alarm Relays
There are three alarm relay outputs:
Alarm 1 (factory-set to trip at 50 ppm)
Alarm 2 (factory-set to trip at 150 ppm)
Alarm 3 (factory-set to trip at 1000 ppm;
300 ppm for R 123)
Each relay can be set up as latching/non-latching and/or normally-energized/normally de-energized. Contacts are Form C at 240 V olts AC 5 amps resistive. User can adjust alarm trip points via the front pane l (see Secti on 3, FIGURE 3-11).
Alarm Relay Connections Wiring
Three refrigerant level alarm relay out puts are provided. All alarm relays are Form C, SPDT relays which can be wired to either closed or opened contacts in an alarm condition. Use copper conductors only.
Each relay has contacts for:
NORMALLY OPEN (NO)
COMMON (COM)
NORMALLY CLOSED (NC)
The function of each relay connec tor terminal is indicated on FIGURE 2-2.
NOTE:The maximum wire size that these
connectors can accept is #12 AWG.
Fault Relay
There is one relay (the Fault r elay) within the unit that indicates that a trouble or start-up condition exists. This relay is configured from the factory and operates differently than the alarm relays. It operates in a normally-energized mode.
This relay is energ ized when the instrument is:
normally operating,
in the calibration mode, or
in the setup condition.
The relay is de-energized when:
a fault is detected,
the unit is in the start-up state, or
the main power is lost.
This means that power is no longer prov ided to the relay coil. It is not pos sible to change the configuration of the Trouble relay.
The relay connector function or ident ification:
NORMALLY OPEN (NO),
COMMON (COM),
NORMALLY CLOSED (NC)
as marked in FIGURE 2-2 refers to the relay contacts as if the relay is de-energized or in the trouble condition. A relay contact is provided between the Normally Closed (NC) and Common (COM) position. This contact will be made in t he event that main power to the unit is lost or any other trouble condition exists.
2-6
Page 17
Chillgard LC Control Module Section 2, Installation and Set-Up
Remote Alarm Silence Switch
The audible alar m and latched ref rigerant level alarms can be remotely reset throug h a switch t hat has a momentary c ontact opening (normally clo sed set of contac ts). Connections are made as shown in FIGURE 2-3 with t he wiring entering the enclosure through the hole on t he left side of the instrument. The switch must have signal-level contacts , typically gold plated. Th e maximum distan ce from the reset switch to the co ntrol module, using 18 AWG wire, is 250 feet. The maximum wire size that these connectors can accept is #12 AWG.
Audible Alarm Output
An outpu t is provided to drive the horn on the bottom of the uni t (FIGURE 2-2).
CAUTION
"
All field wiring must be done in accordance with national and local electrical codes.
Start-Up
The following steps outline the procedures t o power-ON the Chillgard LC Control Module:
1. Before applying power to the unit, verify proper power will be applied to the unit.
2. Turn the instrument ON at the circuit breaker or fuse that supplies power to the instrument. (The instrument does not have a power switch.)
A green LED indicates that power is ON
and communication is established to a sensor module.
NOTE:The display will begin to indicate the gas
concentration of each point connected to the Chillgard LC Control Module.
After power-ON, allow for unit s tabilization (about 30 minutes) before checking calibration of instrument.
Setup Date, Time, and Logging
The data log on the LC unit must be initialized by setting the date and time under the Setup - Setup Time and Date Menu.
Setup logging next on the following menu:
1. Turn ON logging.
2. Set the log frequency.
3. Clear the log file to reset any readings.
4. Reset the LC unit to re-initialize the log to a clean start.
" CAUTION
If this is not done, data recorded will not have a time stamp and data review will not be possible.
Initial Calibration
WARNING
"
Calibrate after installation; otherwise, false or erroneous readings can result.
See Chillgard LS Instruction Manual (P/N 10035164) for calibration or c alibration check instructions. See Chillgard M100 Instruc tion Sheet (P/N 10073591) for calibration or c alibration check instructions.
2-7
Page 18
Chillgard LC Control Module Section 3, Display Screens
Section 3 Display Screens
All instrument operation is performed via the front panel which consists of four keys and a two-line by 20-character LCD display. There is no reason to open the unit for set-up, calibration or diagnostic testing of the instr ument. The most commonly used, self-explanatory screens appear on the following pages. Simply follow the on-screen menus. The step-by step approac h guides you through each operation.
The Display Screen Overview (FIGURE 3-1) shows a general system function flow. See
the following FIGURES for specific Display Screen details:
Start-up and Normal Operation Screens
(FIGURE 3-2)
Calibration Screens
(FIGURES 3-4 through 3-5)
Information Screens
(FIGURES 3-7 through 3-8 and FIGURE 3-16)
Set-up Screens
(FIGURES 3-3 and 3-10 through 3-15).
3-1
Page 19
Section 3, Display Screens Chillgard LC Control Module
3-2
Figure 3-1. Display Screen Overview
Page 20
Chillgard LC Control Module Section 3, Display Screens
Figure 3-2.
Start-up and Normal Operation Screens
Figure 3-3.
Set-up Screens
3-3
Page 21
Section 3, Display Screens Chillgard LC Control Module
Figure 3-4.
Span Calibration Screens
3-4
Figure 3-5.
Zero Calibration
Page 22
Chillgard LC Control Module Section 3, Display Screens
Figure 3-6.
Calibration Check
Figure 3-7.
Diagnostics Menu
3-5
Page 23
Section 3, Display Screens Chillgard LC Control Module
Figure 3-8.
Data Screens
Figure 3-9.
Alarm Level Screen
3-6
Page 24
Chillgard LC Control Module Section 3, Display Screens
Figure 3-10.
Setup Alarm Latching
Figure 3-11.
Setup Alarm Relays
3-7
Page 25
Section 3, Display Screens Chillgard LC Control Module
Figure 3-12.
Audio Alarm Screen
3-8
Figure 3-13.
Analog Outputs and Password
Page 26
Chillgard LC Control Module Section 3, Display Screens
Figure 3-14.
Setup Times
Figure 3-15.
Setup Logging
3-9
Page 27
Section 3, Display Screens Chillgard LC Control Module
3-10
Figure 3-16. Review Log
Page 28
Chillgard LC Control Module Section 4, Maintenance
Section 4 Maintenance
General Maintenance
Under normal operation conditions, the Chillgard LC Control Module requires minimal maintenance.
Obtaining Replacement Parts
To obtain replacement parts, address t he order or inquiry to:
MSA North America P.O. Box 427, Pittsburgh, PA 15230
or call, toll-free, 1-800-MSA-INST.
WARNING
"!
Use only genuine MSA replacement parts when performing any maintenance procedures. Failure to do so may seriously impair unit performance. Repair or alteration of the Chillgard LC Control Module, beyond the scope of these instructions or by anyone other than authorized MSA service personnel, could cause the product to fail to perform as designed and persons who rely on this product for their safety could sustain serious personal injury or death.
Table 4-1. Replacement Parts
PART PART NO.
Mounting Panel, Chillgard LC 10033888 RFI Filter 10034402
Display Assembly, Chillgard LC 10039866 Transfo rmer, Chillg ard LC 10039 002 Printe d Circuit Boar d Assembly, Co ntrol 10033 233 Printed Circuit Board, Membrane Switch 10034274
Strobe Light, red, 2 4 VDC 63467 4 Buzzer 637123 Horn 10034190
WARNING
"
Hazardous Voltage! Disconnect all electric power, including
remote disconnects before servicing. follow proper lockout/tagout procedures to ensure the power cannot be inadvertently energized. Failure to disconnect power before servicing could result in death or serious injury.
4-1
Page 29
Section 4, Maintenance Chillgard LC Control Module
Table 4-2. Troubleshooting Guidelines
TROUBLE DESCRIPTION SOLUTION
1. Check AC power to unit.
2. Verify AC power to unit is wired properly.
3. Check for loose wires on terminal barrier input.
Unit will not turn ON No power
4. Check wiring to the unit power supply. Remove power supply cover and chec k fuse; replace if necessary.
5. Check input cable to mai n board on power supply.
6. Check for 24 VDC power supply output.
7. Check communications
1. Check that plu g is co nn ected to circuit board
Beacon will not light Beacon alarm
Under-range failed
Calibrati on failed
Zero limi t is minus 20 ppm
Coefficients out of range
2. Beacon is controlled by Al arm 3; ch eck that the unit has exceeded Alarm 3.
3. Replace beacon asse mbly.
1. Adjust display zero to 0.0 via the keypad with zero air or zero scrubber applied.
2. Check the zero air cylinder; replace if necessary.
3. Check the zero scrubber; replace if necessary.
1. Replace zero scrubber or change zero air supply.
2. Check span gas supply.
3. Check all tubing, filters and fittings for leaks.
4. For sequencer units, make sure the unit is locked on the selected sample port used for calibration.
5. Leak test the flow system.
6. Return to MSA for service.
4-2
Page 30
Chillgard LC Control Module Section 4, Maintenance
Table 4-2. Troubleshooting Guidelines
TROUBLE DESCRIPTION SOLUTION
1. Remove all input lines to the sensor module. Attach one line at a time to check for sample input. Check all end of line filters. Sample flow failure is always the present point being sampled .
Sensor fl ow failed
Leaky o r blocked sample lin e
2. Che ck opera t io n of all manifold samp le valv es .
3. Leak test the flow system.
4. Return the unit to MSA for service.
Sensor te mperature range fa iled
Infrared source failed in sens or module
Memory protect
External reset failed
Display failure
Checks for tempera ture rang e of
o
over 67
Checks the power of source assembly
Checks checksum of setup a nd cal valu es
Checks the external reset butto n
Display communications
C
1. Sensor temperature exceeded; mount unit in a lower temperature location.
1. Check the co nnection of the IR source into the motherboard.
2. Replac e op t ic al be nc h.
3. Return to MSA for service.
1. Replace the control board
2. Return to MSA for service.
1. If not used, check for jumper.
2. If used, verify switch is wired norma lly closed.
1. Repower unit.
2. Check for broken or cracked display.
3. Replace display.
Audio al arm failure Audio ala rm
No data in th e Data Log
Unit fails startup
Loggin g not set
Password is set inadvertently
1. Check output terminal s.
2. Check f or faulty horn bu zzer.
1. See Chapte r 2, "Initial Calibration"
1. Call MSA for service
4-3
Page 31
Chillgard LC Control Module Section 5, Data Logging
Section 5 Data Logging
The Chillgard LC unit allows 798 sets of data and events to be logged. This data is eit her the maximum or average of the concentration calculated. The structure of t he data stored and transmitted is shown in Appendix C. The download can be initiated by sending the ASCII character "L" or "0x4C." Datalog information can be downloaded by connecting to the RS232 por t (see FIGURE 2-3).
Number of Data Points Used
If the unit is only a single point monitor or all but one point is inactive, the value is recorded continuously. With four points active, 12 values per point are evaluated every hour . The number of values can be considerably lower if the monitor enters extended dwell because of high gas concentrations.
User Setup Options
Concentrations are recorded either every 15 minutes or hourly (FIGURE 3-15). If the logging function is turned OFF, the cl ock/calendar chip is stopped to save battery ener gy.
Before logging can be used, the user must:
1. Turn logging ON.
2. Select an interval of 15 minutes or one hour for readings to be logged.
3. Clear the log.
4. Restart the hardware.
Setting any value starts the c lock/calendar. Hours or days may be skipped or repeat ed. The log can be cleared in the Setup Menu.
Viewing Alarms/Events and Data
From the Review Events Screen, the user can view just alarms, just events, or both. The latest event is presented first. When the memory is filled, the oldest entries are ov erwritten.
From the Review Data Screen, the us er can view the recorded concentrations. The particular point must be selected. Again, the lates t data is presented first; when the memory is f illed, the oldest entries are overwritten.
The log entry number displays to giv e the user a sense of where in the log they ar e located (FIGURE 5-1).
Figure 5-1. Log Description
5-1
Page 32
Chillgard LC Control Module Section 6, Calibration
Section 6 Calibration
Two calibration procedures are available:
1. Diffusion for non-pumped sensors like a single-point, non-pumped Chillgard LS Module
2. Pumped for sensors which incorporate a pump to draw air through tubing to the sensor.
NOTE:The Chillgard M-100 sensor can be
cal-checked, but does not require calibration.
Diffusion Version Calibration
Calibra ti on Equipment - (FIGURE 6-1)
Calibration of the monitor requires a supply of:
ZERO GAS (air or nitrogen) It may be possible to use ambient air if you are sure it does not contain any possible interferant gases or contaminants.
SPAN GAS (A known refrigerant concentration) that measures approximately 10% of the full-scale calibration of the unit. See TABLE 5-3 for available refrigerant gas cylinders.
Relative humidity may have a small effec t on the output of the unit. If dry gas is used, Nafion Tubing (P/N 813628) can be used t o humidify the sample stream going to the monitor.
Both ZERO gas and SPAN gas must be carefully applied to the unit to avoid press urizing the internally mounted optical bench.
See Chillgard LS manual (P/N 10035164) TABLE 5-2, Calibration Accessories Parts List, for the appropriate calibration parts and calibr ation gases available for the Chillgard LS Refrigerant M onitor.
WARNING
"!
Exercise care during the span calibration to ensure that the unit can accurately detect refrigerant gas. Improper calibration can cause improper readings across the full-scale range of the monitor.
0.25 LPM Flow Controller P/N 478359
Calibratio n Adapter P/N 10034395
Figure 6-1. Calibration Equipment
Initial Calibration Procedures
Refer to FIGURES 3-4 and 3-5 for cal ibration menus.
During the initial calibration procedures, alarm relays of any connected contr ol instrumentation may activate. Disconnect or dis able any equipment or alarms.
The following equipment is required for initial calibration:
Tubing Assembly with calibration adapter (FIGURE 6-1)
Calibration Gas
Flow Controller (0.25 LPM) (FIGURE 6-1)
Preparation for Calibration
To verify the instrument is operating properly and to make initial calibration adjustments, perform the following:
1. Remove the light gray cover to open the enclosure on the Chillgard LS and LC.
6-1
Page 33
Section 6, Calibration Chillgard LC Control Module
CAUTION
"
Component Damage! Monitor components must be protected from
splashing, spraying, or dripping water. Failure to do so may cause damage to internal components.
2. Deactivate the equipment connected to the outputs, or disconnect the wiring to the outputs.
3. Replace lid on the enclosures.
CAUTION
"
If any control instruments connected to the Chillgard LS or Chillgard LC Refrigerant Monitor are wired to external devices (e.g., horns, exhaust fans, and fire suppression systems), these devices may activate while adjustments or repairs are performed during the following procedures.
To prevent activating these devices while adjusting the Chillgard LC Monitor, disconnect the wiring from the relay. Return all wiring to the relay when the calibration procedure is completed.
Calibrat io n Gu id e lin e s
Once the Chillgard LS Refrigerant Monitor is operating, perform periodic calibration c hecks to ensure proper instrument operation.
Perform calibration to monitor long-term changes (drift) in both the ZERO and SP AN readings. If there is an unacceptable change in either of these readings, make adjustments to obt ain proper readings.
When routine calibration does not res tore proper readings, perform the procedures out lined under "Initial Calibration."
If following Calibration procedures fails t o restore proper readings of the instrument, see Chillgard LS manual (P/N 10035164) Section 4, "Troubleshooting Guidelines" for guidelines to correct the instrument.
Keep written records of the c alibration readings obtained and any adjustments made. Analysis of these records enables review and c ontrol of the time between checks.
Check a new Chillgard LS Refrigerant M onitor installation at least once a week by performing the steps outlined in the following section.
Calibration Check Procedure
The calibration check procedure involves chec king the SPAN and ZERO readings on t he instrument.
Figure 6-2. Applying Calibration Gas
to the Chillgard LS Refrigerant Monitor
During the calibration check procedure, any control instrumentation connected to t he Chillgard LS Refrigerant Monitor may activate. Disconnect or disable any equipment or alarms connect ed to the monitor during the calibration pr ocedure.
Applying Calibration Gases to the Instrument
Arrange Span and Zero gas cylinders wit h regulator, tubing, and cal cap as shown in FIGURE 6-2. Refer to FIGURE 3-6 for t he calibration menus.
WARNING
"!
Refrigerant Hazard! During calibration, the Chillgard LS Monitor
is not sampling and monitoring the intended area. Exercise caution in the area as appropriate. Failure to do so may result in death or serious personal injury.
1. Lock the Chillgard LC Control Module to the point being calibrated.
Only the point being calibrated displays.
2. Perform the Zero Calibration first.
Use Nitrogen, not room air, to calibrate
the zero point and avoid any contamination.
6-2
Page 34
Chillgard LC Control Module Section 6, Calibration
Be sure sure to apply gas to the sensor
for three minutes.
3. Enter the CAL - USER - ZERO - NEXT ­CHANGE Menu to select the point to be calibrated (see FIGURE 3-5).
4. Press OK on the correct point; then, press NEXT.
5. After a reading displays, press ADJ; then, press UP or DN to return the displayed gas value to 0 ppm.
6. Perform Span Calibration using the appropriate sensor gas at 100 ppm.
Be sure to apply gas for three minutes.
7. Enter the CAL - USER - SPAN - NEXT ­CHANGE Menu to select the point to be calibrated (see FIGURE 3-4).
8. Press OK on the correct point; then, press NEXT.
9. Press ADJ; then, press UP or DN to set the span to the gas concentration of the type being used (100 ppm).
10. Close the regulator valve and remove the span gas cylinder from the sampling tubing.
11. Remove the calibration cap and return to normal operation.
Calibration Equipment
Equipment needed:
Calibration Kit (MSA ATO #50; FIGURE 6-3)
A SPAN gas cylinder
Optional ZERO gas cylinder.
A ZERO gas cylinder may not be needed.
The Calibration Kit contains a ZERO gas scrubber which can be used in place of a ZERO gas cylinder if the ambient air around t he Chillgard LS contains little or no refrigerant.
Relative humidity may have a small effec t on the output of the unit. If dry gas is used, Nafion Tubing (P/N 813628) can be used t o humidify the sample stream going to the monitor.
CAUTION
"
The zero gas scrubber must be replaced periodically. The frequency of replacement depends on the concentration of the ambient refrigerant vapors.
The SPAN or ZERO cylinders (if needed) m ay be included with the Calibration Kit; cy linders shown in TABLE 6-1 are available from MSA.
CAUTION
"
Equipment Failure! Do not leave any alarm device or equipment
disabled or disconnected during normal operation of the instrument; otherwise, the equipment will not function as intended when the instrument detects an alarm situation. Failure to do so may result in serious personal injury.
Single or Four-Point Pumped Version Calibration
Introduction
As with any type of gas monit or, the only true check of its performance is t o apply gas directly to the sensor. The frequency of the calibration gas test depends on the operating time and exposures of the sensors. New monitors should be calibrated more often until the c alibration records prove stability. The c alibration frequency can then be reduced to the sc hedule set by the safety officer or plant manager.
Perform the calibration procedure regularly and maintain a log of calibration adjustments. Calibration frequency may increase for a variety of reasons. If calibration cannot be performed at any step, STOP; consult MSA at 1-800-MSA-INST.
Table 6-1. RP Calibration Gases
DESCRIPTION CONCENTRATION PART NO.
R-11 in Nitrogen 100 ppm 803499 R-12 in Nitrogen 100 ppm 804866 R-123 in Nitrogen 100 ppm 803498 R-134A in
Nitrogen R-22 in Nitrogen 100 ppm 804868
Become familiar with the Calibration Kit components (FIGURE 6-3).
Exercise care during the span calibration to ensure that the unit can accurately detect refrigerant gas. Improper calibration can cause improper readings across the full-scale range of the monitor.
100 ppm 803500
WARNING
"
Calibration Procedures
Preparation for Calibration
To verify the instrument is operating properly and to make initial calibration adjustments, perform the following:
1. Remove the light gray cover to open the enclosures on the Chillgard LS and LC Monitors.
6-3
Page 35
Section 6, Calibration Chillgard LC Control Module
Tube and Tee Asse mbly
(P/Ns 6038 06 and 636866 )
Connector Assembly (P/N 711533)
Zero Gas Scrubber (P/N 803873)
(Replace protective caps after use)
Figure 6-3. Kit Components
CAUTION
"
Component Damage! Monitor components must be protected from
splashing, spraying, or dripping water. Failure to do so may cause damage to internal components.
2. Deactivate the equipment connected to the outputs, or disconnect the wiring to the outputs.
CAUTION
"
1.5 LPM Flow Controller (P/N 4783 58 )
Span Ga s Scrubb er (P/N 803874)
(Replace protective caps after use)
1. Temporarily block the sample inlet at the end-of-line filter(s) and verify that the monitor gives a Fault alarm.
2. After checking for leaks, remove the sampling line for the Chillgard LS inlet.
3. Attach the Calibration Kit connector assembly to the in l e t.
Calibrat io n Gu id e lin e s
Once the Chillgard LS Refrigerant Monitor is operating, perform periodic calibration c hecks to ensure proper instrument operation.
If any control instruments connected to the Chillgard LS or LC Refrigerant Monitor are wired to external devices (e.g., horns, exhaust fans, and fire suppression systems), these devices may activate while adjustments or repairs are performed during the following procedures.
To prevent activating these devices while adjusting the Chillgard LC Refrigerant Monitor, disconnect the wiring from the relay. Return all wiring to the relay when the calibration procedure is completed.
Before calibrating the Chillgard LS Monitor, leak-check the sample line(s) connected t o the monitor:
6-4
Perform calibration to monitor long-term changes (drift) in both the ZERO and SP AN readings. If there is an unacceptable change in either of these readings, make adjustments to obt ain proper readings.
When routine calibration does not res tore proper readings, perform the procedures out lined under "Initial Calibration."
If following Calibration procedures fails t o restore proper readings of the instrument, see Section 5, "Troubleshooting Guidelines" for guidelines to correct the instrument.
Keep written records of the c alibration readings obtained and any adjustments made. Analysis of
Page 36
Chillgard LC Control Module Section 6, Calibration
these records enables review and c ontrol of the time between checks.
Calibrat ion Equipment
Calibration of the monitor requires a supply of:
ZERO GAS (nitrogen) It may be possible to use ambient air if you are sure it does not contain any possible interferant gases or contaminants.
SPAN GAS (A known gas concentration that measures approximately 10% of the full-scale calibration of the unit.
Carefully apply both ZERO gas and SPAN gas t o the unit to avoid pressurizing t he internally mounted sensing cell. See TABLE 5-2, "Calibration Accessories Parts List" for the appropriate calibration parts and calibr ation gases available for the Chillgard LS Refrigerant M onitor.
Figure 6-4. Using Zero Scrubber for Zero Calibration
Figure 6-5. Using Zero Gas Cylinder
for Zero Calibration
6-5
Page 37
Section 6, Calibration Chillgard LC Control Module
Calibration Check Procedure
The calibration check procedure involves chec king the SPAN and ZERO readings on t he instrument. During the calibration check procedure, any control instrumentation connected to t he Chillgard LS Refrigerant Monitor may activate. Disconnect or disable any equipment or alarms connect ed to the monitor during the calibration pr ocedure.
Zeroi ng the M o nitor
When zero gas is required, att ach a zero gas scrubber or zero gas cylinder to the connector as shown in FIGURES 6-4 and 6-5.
WARNING
"
If the sampling line is not re-attached, the monitor cannot sample from the remote location.
CAUTION
"
During calibration, the Chillgard LS Refrigerant Monitor is not sampling and monitoring the intended area. Exercise caution in the area as appropriate.
6-6
Figure 6-6. Span Calibration
Page 38
Chillgard LC Control Module Section 6, Calibration
Assemble this setup when calibrating a Chillgar d LS-4 sensor. Be sure that the pressure well outlet is at least six inches long t o minimize dilution by the outside air. This pressure well allows for a flow outlet when the flow regulator has a higher flow than the Chillgard LS-4 sensor. Tr ying to limit the flow from the regulator is quite difficult and may lead to under-pressurization of the Photo acoustic Bench. If this under -pressurization occurs, concentration readings f rom that sensor will be low.
1. Lock the Chillgard LC sensor onto the first sampler address of the Chillgard LS-4 sensor (point 1 or point 5).
2. Remove any sampling tubes from the first point; calibration gases will be applied directly to the sampler.
3. Perform Z ero Calib r a tion firs t.
Use Nitrogen, not room air, to calibrate
the zero point and avoid any contamination.
Be sure sure to apply gas to the sensor
for three minutes.
4. Enter the CAL - USER - ZERO - NEXT ­CHANGE Menu to select the point to be calibrated (point 1 or point 5). (See FIGURE 3-5.)
5. Press OK on the correct point; then, press NEXT.
6. After a reading displays, press ADJ; then, press UP or DN to return the displayed gas value to 0 ppm.
7. Perform Span Calibration using the appropriate sensor gas at 100 ppm.
Be sure to apply gas for three minutes.
8. Enter the CAL - USER - SPAN - NEXT ­CHANGE Menu to select the point to be calibrated. (See FIGURE 3-4.)
9. Press OK on the correct point; then, press NEXT.
10. Press ADJ then UP or DN to set the span to the gas concentration of the type being used (100 ppm).
11. Close the regulator valve and remove the SPAN gas cylinder from the sample tubing.
12. Remove tubing from the sensing cell on the unit.
13. Re-connect or enable all equipment and alarm devices connected to any control equipment monitoring the Chillgard LS Refrigerant Monitor.
CAUTION
"
Do not leave any alarm device or equipment disabled or disconnected during normal operation of the instrument; otherwise, the instrument will not function as intended when it detects an alarm situation.
6-7
Page 39
Chillgard LC Control Module Appendix A, RS-232 Output
Appendix A RS-232 Output
Introduction
The instrument is ready to monitor; however, it is necessary to configure the unit to your specific requirements. Your Chillgard LC Control Module may be connected to Multi-Point Sequenc er(s) and/or multi-sensor modules, enabling your unit to sample from up to 12 individual sampling locat ions.
Your unit also contains several ot her features described in this appendix:
RS-232 output
Password protection
Sampling point identification.
RS-232 Output
The RS-232 output broadcasts certain information about the Chillgard LC Control Module (TABLE A-3). This output conforms t o the specification for RS-232 signal levels and is capable of driving its signal up to 200 feet when using low c apacitance RS-232 cable.
Connection to the RS-232 output is v ia a 9-pin sub "D" female connector.
Table A-1. RS-232 Parameters
9600 Bau d or 19,200
COMMUNICATION PARAMETERS
The log data is an array of s tructures. The array size is 792 and the structure lengt h is 22 bytes. The first byte is a sync character "R" followed by 23,760 bytes. The first s tructure is the Log Information.
(See TABLE 2-1) 8 bits No parity One stop bit
Table A-2. Information Structure
(Chill ga r d LC C ontrol Module)
ITEM
Total 30
DESCRIPTION VALUE OR TYPE
NO.
0 Monitor type ASCII "R" 1 1 Monitor Type ASCII "L" 1 2 Monitor Type ASCII "C" 1 3 Scale char 0x0 3 1 4 ID Number unsigne d integer 2 5 Year (0 - 99)* bcd number 1 6 Month (1 - 12) bcd number 1 7 Date (1 - 31) bcd number 1 8 Hour (0 - 23) bcd number 1 9 Minute (0 - 59) bcd number 1
10 Average or Max char 1
Point 1 Gas
11
Type Point 2 Gas
12
Type Point 3 Gas
13
Type Point 4 Gas
14
Type Point 5 Gas
15
Type Point 6 Gas
16
Type Point 7 Gas
17
Type Point 8 Gas
18
Type Point 9 Gas
19
Type Point 10 Gas
20
Type Point 11 Gas
21
Type Point 12 Gas
22
Type 23 Sequencer Points char 1 24 Log Index unsigned inte ger 2
*The base/year is 2000.
The last stru cture is filled with 0xff .
char 1 char 1 char 1
char 1 char 1 char 1 char 1
char 1 char 1 char 1 char 1 char 1
SIZE
(BYTES)
The information structure is followed by data and event structures. Data s tructures are always identified by Alarm/Event value of 0x01. Empty structures have Alarm/Event v alue of 0x00.
A-1
Page 40
Appendix CA RS-232 Output Chillgard LC Control Module
Table A-3. Data Structure
ITEM
NO.
1 Alarm or Event char = 0x00 1
2 Year* bcd 1
3 Month bcd 1
4Date bcd 1
5Hour bcd 1
6Minute bcd 1
7 Conce ntration P oint 1 integer 2
8 Conce ntration P oint 2 integer 2
9 Conce ntration P oint 3 integer 2
10 Concentration Point 4 integer 2
11 Concentration Point 5 integer 2
12 Concentration Point 6 integer 2
DESCRIPTION
VALUE OR
TYPE
SIZE
(BYTES)
Table A-4. Alarm Structure
ITEM
Total 30
DESCRIPTION VALUE OR TYPE SIZE (B YTES)
NO.
Alarm or
1
Event 2Year bcd 1 3Month bcd 1 4Date bcd 1 5Hour bcd 1 6Minute bcd 1 7 Point (1 - 12) c har 1 8spaces 0x00 15
*The base/year is 2000.
char 1
Table A-5. Alarms and Events
DATA CONCENTRATION 0x01
LOG_ALARM 1 0x02
ALARMS
LOG_ALARM 2 0x04
13 Concentration Point 7 integer 2
14 Concentration Point 8 integer 2
15 Concentration Point 9 integer 2
16 Concentration Point 10 integer 2
17 Concentration Point 11 integer 2
18 Concentration Point 12 integer 2
Total integer 30
*The base/ye ar is 2000.
EVENTS
LOG_ALARM 3 0x08 LOG_WARMUP 0x10 LOG_READY 0x20 LOG_FAULT 0x40 LOG_CAL_SETUP 0x80
A-2
Page 41
Chillgard LC Control Module Appendix B, Installation Outline Drawing
Appendix B Installation Outline Drawing
B-1
Page 42
Appendix B, Installation Outline Drawing Chillgard LC Control Module
B-2
Figure B-1. Installation Outline Drawing
for Standard Version
Page 43
Chillgard LC Control Module Appendix B, Installation Outline Drawing
Figure B-2. Installation Outline Drawing
for Transformer Version
B-3
Loading...