The MSA BD96 SL (referred to hereafter as compressed air breathing apparatus)
is a self-contained breathing apparatus operating independent of the ambient air.
Breathable air is supplied to the user from compressed air cylinder(s) via a
pressure reducer, a demand controlled dosage assembly (see Instruction Manual
for the Lung Governed Demand Valve) and a face piece (see Instruction Manual
for the Face piece). The exhalation air is released directly into the ambient
atmosphere.
It is imperative that this operating manual be read and observed when using the
compressed air breathing apparatus. In particular, the safety instructions, as well
as the information for the use and operation of the apparatus, must be carefully
read and observed. Furthermore, the national regulations applicable in the user's
country must be taken into account for a safe use.
Alternative use, or use outside this specification will be considered as noncompliance. This also applies especially to unauthorised alterations to the
apparatus and to commissioning work that has not been carried out by MSA or
authorised persons.
Danger!
This product is supporting life and health. Inappropriate use, maintenance or servicing may affect the function of the device and thereby
seriously compromise the user’s life.
Before use the product operability must be verified. The product must
not be used if the function test is unsuccessful, it is damaged, a
competent servicing/maintenance has not been made, genuine MSA
1.2. Liability information
MSA accepts no liability in cases where the product has been used
inappropriately or not as intended. The selection and use of the product are the
exclusive responsibility of the individual operator.
Product liability claims, warranties also as guarantees made by MSA with respect
to the product are voided, if it is not used, serviced or maintained in accordance
with the instructions in this manual.
spare parts have not been used.
Danger!
This compressed air breathing apparatus is a pure gas protection
device. It is not suitable for underwater diving.
4 BD 96 SL
GB
MSA DESCRIPTION
61 2
8
9
2
2. Description
3
4
5
7
1
1011
Fig. 1 BD SL Compressed air breathing apparatus
1 Combo pressure gauge with couplings 7 Carrying assembly
2 Tension clamp 8 Handle
3 Cylinder retaining strap 9 Quick-Fill coupling (Option)
4 Cylinder support 10 Pressure reducer
5 Cylinder separator 11 Multi-channel line
6 Shoulder strap 12 Hip belt
2.1. Standard Model (BD 96 SL)
The carrying assembly consists of an anatomically designed backplate of
antistatic plastic with carrying slots for easy handling. On the lower part of the
carrying assembly are the supports for the pressure reducer and an elastic bump
protector. On the upper part of the assembly, a cylinder support is attached with
integrated line guide.
The carrying belts and the hip belt are adjustable in length.
One or two compressed air cylinders can be placed in the cylinder support. The
cylinder strap is freely adjustable and after inserting the compressed air
cylinder(s), it is tied down tightly and secured with the tension clamp.
BD 96 SL 5
GB
DESCRIPTION MSA
2
2
4
1
3
Fig. 2 Pressure reducer
1 Multi-channel line connection
2 Compressed air cylinder connection
3 Quick-Fill coupling (not for standard model)
4 Lead seal
The pressure reducer (see Fig. 2) is mounted in the lower area of the carrying
plate. On the pressure reducer, there is a safety valve and the multi-channel line
for connecting the combo pressure gauge. The pressure reducer reduces the
cylinder pressure to approx. 7 bar and the safety valve activates on non-permitted
pressure rise to prevent damage insuring the continued supply of breathable air.
The combo pressure gauge (see Fig. 3) is connected at the end of the multichannel line. It consists of the pressure gauge (1) itself, the coupling for the lung
governed demand valve (2) as well as an acoustic warning device (3) (signal
whistle). It triggers a continuous warning signal when the cylinder pressure drops
below 55±5 bar. The second connection (4) connects a second lung governed
demand valve (e.g. backup set).
6 BD 96 SL
GB
MSA DESCRIPTION
r
r
g
g
2.2. Model with Quick-Fill Coupling (BD 96 SL-Q)
Besides the standard unit fittings, this model also comes with a high pressure
safety coupling (see Pos. 3 in Fig. 2).It is then possible to fill 300 bar compressed
air cylinder(s) whilst the compressed air breathing apparatus is in use.
Warning!
In compressed air breathing apparatus with Quick-Fill couplings the use
of 200 bar compressed air cylinders is not permitted.
2.3. Technical Data
High pressure connection: 200 bar resp. 300 ba
Medium pressure : 5 bar to 9 ba
Operating temperature: -30°C to +60°C
ht (approx.) : 3.8 k
Wei
Dimensions (approx.) : Length 585 mm
Width 290 mm
Height 145 mm
Approvals
: The compressed air breathing apparatus
conforms to the Directives 89/686/EEC and
94/9/EG. It is a container unit with compressed air in accordance with EN 137.
ATEX
BVS 03 ATEX H 010 X
IM1c
II 1 G c IIC T6 -30°C ≤ Ta ≤ +60°C
II 1 D c
0158
BD 96 SL 7
GB
USING THE COMPRESSED AIR BREATHING APPARATUS MSA
3. Using the Compressed Air Breathing Apparatus
Warning!
The compressed air breathing apparatus may only be put into use in a
fully maintained and tested condition. If malfunctions or defects are
noticed prior to use, do not use the compressed air breathing apparatus
under any circumstances.
Get the apparatus checked and repaired by an authorised service
3.1. Connecting one Compressed Air Cylinder
Fig. 4 Compressed air breathing apparatus with one compressed air cylinder
(1) Place compressed air breathing apparatus horizontally so that the back face
(2) Bring cylinder separator (5) into a horizontal position opposite the tension
(3) Check gasket on pressure reducer (10) for proper condition.
(4) Open tension clamp on cylinder retaining strap (3) and extend strap
(5) Push compressed air cylinder through the cylinder strap (3) with the cylinder
(6) Screw cylinder valve onto pressure reducer, if necessary, bring the
(7) Fasten cylinder strap, do not over tighten.
(8) Check position of compressed air cylinder, retighten if necessary.
centre.
is uppermost (see Fig. 1).
clamp (2) until it catches.
(see Fig. 7).
valve toward the pressure reducer, so that it lies on the central support (4).
compressed air breathing apparatus with valve up into a vertical position.
8 BD 96 SL
GB
MSA USING THE COMPRESSED AIR BREATHING APPARATUS
Attention!
Tighten cylinder retaining strap only when clamping lever is in clamped
position, otherwise, it can not be closed.
(9) Hinge tension clamp down until it catches.
(10) Fasten end of the cylinder retaining strap onto Velcro strip.
(11) Briefly open cylinder valve and check for escaping air, retighten if
necessary.
3.2. Connecting two Compressed Air Cylinders
Fig. 5 SCBA apparatus with twin compressed air cylinders
(1) Place breathing apparatus down in a horizontal position with the back face
uppermost (see Fig. 1).
(2) Hinge cylinder separator (5) into a vertical position until it catches.
(3) If there is no T-piece connected to the pressure reducer (10), check gaskets
on pressure reducer and T-piece and screw in T-piece loosely.
(4) Open tension clamp on cylinder retaining strap (3) and extend strap
(see Fig. 7).
(5) Push one compressed air cylinder through the cylinder retaining strap so
that the cylinder valve points to the T-piece, and lies on one of the outer
supports.
(6) Loosely screw the cylinder valve onto the T-piece.
(7) Push a second compressed air cylinder through the cylinder retaining strap
so that the cylinder valve points to the T-piece, and lies on the other outer
support.
(8) Loosely screw the second cylinder valve onto the T-piece.
BD 96 SL 9
GB
USING THE COMPRESSED AIR BREATHING APPARATUS MSA
y
r
r
With right-angled T-piece, align the compressed air cylinders by hinging
the pressure reducer and pulling the cylinders together (see Instruction
Manual for T-pieces).
(9) Tighten all three high pressure connection handwheels.
(10) Briefly open cylinder valves and check for escaping air, retighten if
necessary.
(11) Check tight position of compressed air cylinders, retighten if necessary.
(12) Tighten cylinder strap on loose end.
Attention!
Tighten cylinder retaining strap only when clamping lever is in clamped
position, otherwise, it can not be closed.
(13) Hinge tension clamp down until it catches.
(14) Fasten end of the cylinder strap onto Velcro strip.
(15) Briefly open cylinder valve and check for escaping air, retighten if
necessary.
3.3. Donning the Compressed Air Breathing Apparatus
(1) Check all components of the compressed air breathing apparatus for
defects and malfunctions.
(2) Don breathing apparatus with shoulder straps fully extended.
(3) Close hip belt and tighten loose ends.
(4) Tighten shoulder straps until the back plate fits comfortably.
(5) Adjust shoulder straps to achieve a comfortable weight distribution between
shoulder straps and hip belt.
(6) If required, connect lung governed demand valve to medium pressure
coupling (see Instruction Manual for Lung Governed Demand Valve).
3.4. Condensed Prior to Use Check
(1) Open cylinder valve(s) and check pressure on the combo pressure gauge.
The pressure values must read:
for 300 bar c
for 200 bar cylindersminimum 180 ba
lindersminimum 270 ba
(2) Close cylinder valve(s) and observe pressure gauge.
•The pressure must not drop more than 10 bar in 60 s.
(3) Carefully activate flushing mode of the lung governed demand valve, closing
exit port as much as possible.
(4) Observe pressure gauge.
•The warning signal must sound at 55±5 bar.
10 BD 96 SL
GB
MSA USING THE COMPRESSED AIR BREATHING APPARATUS
3.5. Donning the Face piece (Full Face Mask)
(1) Don full face mask and check face piece-to-face fit (palm test) (see
Instruction Manual of Full face mask)
(2) Open cylinder valve(s) fully.
Warning!
When using two compressed air cylinders, always open the valves of
both cylinders. Only then will both cylinders be empty uniformly.
(3) Connect lung governed demand valve to face piece (see Instruction Manual
for Lung Governed Demand Valve)
(4) The compressed air breathing apparatus is ready for use.
3.6. During Use
(1) Regularly check tight fit of face piece and lung governed demand valve and
retighten if necessary, as well as the air supply shown on the pressure
gauge.
(2) Leave area immediately if the warning signal sounds.
Independent from the warning signal, an earlier retreat may be required;
the retreat start is based on the reading shown on the pressure gauge.
Danger!
The warning signal sounds when the air supply in the compressed air
cylinders is reduced.
In such cases, immediately leave the area, there is danger of air
deficiency.
3.7. Use of the Second Connection
(1) Remove safety cap from medium pressure coupling of Second Connection
on the combo pressure gauge.
(2) Connect medium pressure line of lung governed demand valve of second
user pushing until the coupling audibly catches.
Warning!
When the Second Connection is in use, more air is consumed and
service time is considerably reduced. Always keep this in mind when
using your apparatus.
BD 96 SL 11
GB
USING THE COMPRESSED AIR BREATHING APPARATUS MSA
3.8. Filling with Quick-Fill
1
Fig. 6 Fill with Quick-Fill (optional)
1 Quick-Fill coupling
With the Quick-Fill function, the compressed air cylinders of the breathing
apparatus can be filled during use (see Instruction Manual for Quick-Fill).
3.9. Removing the Compressed Air Breathing Apparatus
(1) Remove lung governed demand valve.
(2) Close cylinder valve(s).
(3) Activate flushing mode of lung governed demand valve, releasing all air
pressure.
(4) Open hip belt.
(5) Extend shoulder straps by lifting the slides.
Danger!
Do not throw off compressed air breathing apparatus. This could
damage the valve and any remaining compressed air could escape
suddenly.
(6) Remove compressed air breathing apparatus.
This could cause fatal injury to you or to any bystanders.
12 BD 96 SL
GB
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.