MSA AutoMaXX-AS, AutoMaXX Series, AutoMaXX-AS-B-G, AutoMaXX-AS-B, AutoMaXX-AS-C Operating Manual

...
MSAsafety.com
Operating Manual
AutoMaXX - Series
Negative and Positive Pressure Lung Governed Demand Valve
Order No.: D4085399/05
MSA Europe GmbH Schlüsselstrasse 12 8645 Rapperswil-Jona
Switzerland
© MSA 2018. All rights reserved
MSA CONTENTS
AutoMaXX - series 3
Contents
1. Safety regulations ............................................................................................... 5
1.1. Correct use ................................................................................................. 5
1.2. Liability information ..................................................................................... 5
1.3. Series ......................................................................................................... 6
2. Description ........................................................................................................... 8
3. Operation............................................................................................................ 10
3.1. Preparing the device ................................................................................ 10
3.2. Connecting the AutoMaXX ....................................................................... 10
3.3. Putting the AutoMaXX into operation
(only Positive pressure lung governed demand valve) ............................. 10
3.4. Switching off the AutoMaXX
(only positive pressure lung governed demand valve) ............................. 11
3.5. Removing the AutoMaXX ......................................................................... 11
3.6. Bypass system (only for AutoMaXX-AS) .................................................. 11
3.7. Additional function .................................................................................... 11
4. Care and maintenance ...................................................................................... 12
4.1. Maintenance instructions .......................................................................... 12
4.2. Maintenance intervals .............................................................................. 12
4.3. Cleaning / Disinfection .............................................................................. 13
4.4. Testing and replacing the diaphragm ....................................................... 13
4.5. Visual and function check ......................................................................... 15
4.6. Tightness check ....................................................................................... 15
4.7. Checking the closing pressure
(only positive pressure lung governed demand valve) ............................. 16
4.8. Checking the opening pressure
(only negative pressure lung governed demand valve) ............................ 16
4.9. Overhaul ................................................................................................... 17
4.10. Storage ..................................................................................................... 17
4.11. Malfunction ............................................................................................... 17
CONTENTS
MSA
4 AutoMaXX - series
5. Technical data ................................................................................................... 17
6. Approved system components ........................................................................ 18
7. Ordering information ......................................................................................... 20
7.1. Positive pressure lung governed demand valves ..................................... 20
7.2. Negative pressure lung governed demand valves ................................... 21
7.3. Accessories .............................................................................................. 21
MSA SAFETY REGULATIONS
AutoMaXX - series 5
1. Safety regulations
1.1. Correct use
The AutoMaXX is a lung governed demand valve for a compressed air breathing apparatus or compressed airline breathing apparatus. It is connected between compressed air supply and full face mask and controls the inhaled and exhaled air.
The lung governed demand valve series are only permitted for specific system components (exact allocation Chap. 6).
Warning!
The AutoMaXX is not an independent breathing apparatus. It must be used exclusively mounted on a full face mask.
The AutoMaXX is not suitable for underwater diving.
It is imperative that this operating manual be read and observed when using the AutoMaXX. In particular, the safety instructions, as well as the information for the use and operation of the apparatus, must be carefully read and observed. Furthermore, the national regulations applicable in the user's country must be taken into account for a safe use.
Alternative use, or use outside these specifications will be considered as non­compliance. This also applies especially to unauthorised alterations to the apparatus and to commissioning work that has not been carried out by MSA or authorised persons.
Danger!
This product supports life and health. Inappropriate use, maintenance or servicing may affect the function of the device and thereby seriously compromise the user’s life.
Before use, the product operability must be verified. The product must not be used, if the function test is unsuccessful, it is damaged, a competent servicing/maintenance has not been made, genuine MSA spare parts have not been used.
1.2. Liability information
MSA accepts no liability in cases where the product has been used inappropriately or not as intended. The selection and use of the product are the exclusive responsibility of the individual operator.
Product liability claims, warranties also as guarantees made by MSA with respect to the product are voided, if it is not used, serviced or maintained in accordance with the instructions in this manual.
SAFETY REGULATIONS
MSA
6 AutoMaXX - series
1.3. Series
This manual applies to the following lung governed demand valves from the AutoMaXX series:
Name of lung governed demand valve
Medium pressure line
Special feature
AutoMaXX-AE, Lung Governed Demand Valve, threaded connection M45X3
Silicon line 650
AutoMaXX-AE-G, Lung Governed Demand Valve, threaded connection M45X3, Rubber line
Rubber line 650
AutoMaXX-AE-SHORT, Lung Governed Demand Valve, threaded connection M45X3
Silicon line 500
AutoMaXX-AS, Lung Governed Demand Valve, plug-in connection
Silicon line 650
AutoMaXX-AS-B, Lung Governed Demand Valve, plug-in connection BYPASS
Silicon line 650 with bypass
AutoMaXX-AS-B-G, Lung Governed Demand Valve, plug-in connection BYPASS
Rubber line 650 with bypass
AutoMaXX-AS-C, Lung Governed Demand Valve, plug-in / threaded connection M45X3
Silicon line 650
Combination plug­in/M45x3
AutoMaXX-AS-C-G, Lung Governed Demand Valve, plug-in / threaded connection M45X3, Rubber line
Rubber line 650
Combination plug­in/M45x3
AutoMaXX-AS-G, Lung Governed Demand Valve, plug-in connection, Rubber line
Rubber line 650
AutoMaXX-AS-MicroMaXX Rubber line 1100
Long line, small plug-in nipple, only for MicroMaXX
AutoMaXX-AS-SHORT Lung Governed Demand Valve, plug-in connection
Silicon line 500
AutoMaXX-N, Lung Governed Demand Valve, threaded connection RD40X1/7
Silicon line 650
AutoMaXX-N-G, Lung Governed Demand Valve, threaded connection RD40X1/7, Rubber line
Rubber line 650
AutoMaXX-N-MicroMaXX Rubber line 1100
Long line, small plug-in nipple, only for MicroMaXX
AutoMaXX-NR, Lung Governed Demand Valve, for RESCUE Set
Rubber line 1500
AutoMaXX-N-SHORT, Lung Governed Demand Valve, threaded connection RD40X1/7
Silicon line 500
MSA SAFETY REGULATIONS
AutoMaXX - series 7
Explanations on model designation
AS: Positive pressure, plug-in connection, suitable for full face masks PS-MaXX (non-PS!)
AE:
Positive pressure, thread connection M45x3 according to DIN EN 148-3, suitable for full face masks -PF
AS-C: Positive pressure, offers the options AS and AE as combination, suitable for full face masks-PF
N:
Negative pressure, round thread RD 40 according to DIN EN 148-1, suitable for round thread full face masks
G: Rubber medium pressure line (instead of silicon)
B: Bypass for constant air flow
R: Rescue set (component of rescue set)