SAFETY REGULATIONS MSA
1. Safety Regulations
1.1. Correct Use
The ALTAIR PRO is a portable instrument for monitoring toxic gases and oxygen
deficiency or excess in the workplace. It is used exclusively for monitoring and not
for measuring gas concentrations in the ambient air. It must be serviced and
repaired by qualified and authorised personnel.
It is imperative that this operating manual be read and observed when using this
instrument. In particular, the safety instructions, as well as the information for the
use and operation of the apparatus, must be carefully read and observed.
Furthermore, the national regulations applicable in the user's country must be
taken into account for a safe use.
Attention!
This product is supporting life and health. Inappropriate use,
maintenance or servicing may affect the function of the device and
thereby seriously compromise the user’s life.
Before use, the product operability must be verified. The product must
not be used if the function test is unsuccessful, it is damaged, a
competent servicing/maintenance has not been made, genuine MSA
spare parts have not been used.
Alternative use, or use outside this specification, will be considered as noncompliance. This also applies especially to unauthorised alterations to the
apparatus and to commissioning work that has not been carried out by MSA or
authorised persons.
1.2. Liability Information
MSA accepts no liability in cases where the product has been used
inappropriately or not as intended. The selection and use of the product are the
exclusive responsibility of the individual operator.
Product liability claims, warranties also as guarantees made by MSA with respect
to the product are voided, if it is not used, serviced or maintained in accordance
with the instructions in this manual.
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
MSA SAFETY REGULATIONS
1.3. Safety and Precautionary Measures to be Adopted
Attention!
The following safety instructions must be observed implicitly. Only in
this way can the safety and health of the individual operators, and the
correct functioning of the instrument, be guaranteed.
Use instrument only for stipulated gases
The ALTAIR PRO is used for monitoring and detecting gases in the ambient air.
Thereby, it must be used only for gases for which a sensor is installed in the
instrument.
The sensor must not be blocked during use.
Observe oxygen concentration
Although the ALTAIR PRO unit will detect up to 25% oxygen, all ALTAIR PRO
versions are not certified or approfed for use in atmospheres containing more
than 21% oxygen.
Check function and alarm before each use
In case of failed function tests, the instrument must be removed from use.
Conduct additional tests after any significant impact
The instrument function should also be checked after severe mechanical stress
(dropping, impact etc.) and when the instrument or the sensors have been
exposed to silicon, silicates, lead-containing substances, hydrogen sulphide or
contaminated compounds.
Defective vibration alarm possible
At ambient temperatures < 0 °C, the vibration alarm can give a false alarm or can
fail completely. In such conditions, do no rely solely on the vibration alarm.
Observe BG information
In Germany, observe the BG information BGI 836 (5).
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
DESCRIPTION MSA
2. Description
2.1. Instrument Overview
1
2
3
4
5
6
7
Fig. 1 Instrument view
1 Alarm LED 5 Button
2 Infrared port 6 Display with backlight
3 Sensor cover 7 Stipulated gas type
4 Acoustic alarm transmitter
The instrument is a portable gas meter for monitoring gases in the ambient air
and in the workplace.
However, all instrument models are calibrated and permitted only up to
21% oxygen-content.
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
MSA DESCRIPTION
The response levels for the individual gases are set at the factory
and cannot be
changed during operation. After commissioning the instrument, changes are
possible only with the MSA FiveStar® Link™ Software version 4.4 or higher
(available as option; Æ Description of software).
If the ambient air contains higher gas concentrations, the instrument
ignores the Fresh Air Setup and goes directly to measuring mode after
the self-test.
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
DESCRIPTION MSA
2.2. On-Screen Indicators
1
2
Fig. 2 Display
1 Self-test in progress
2 Bump check indicator
3 Hour glass
4 Confidence indicator
5 Alarm indicator
6 Battery life indicator
7 Numeric indicator
8 Gas concentration units
4
3
5
6
7
8
Battery Life Indicator
Attention!
If an alarm is triggered while using the instrument as a monitor, leave
the area immediately.
Remaining on site under such circumstances can cause serious
damage to health or can even lead to death.
The battery condition icon continuously displays in the upper right-hand corner of
the display. As the battery is depleted, battery icon segments go blank until only
the outline of the battery icon remains.
Each battery indicator segment represents approximately 1/4 of the total battery
capacity.
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
MSA DESCRIPTION
Battery Warning
A battery warning indicates that a nominal two days of operation remain before
the instrument's battery is depleted.
The duration of remaining instrument operation during battery warning
depends on ambient temperature (colder temperatures may reduce
battery life) and the number and duration of alarms during battery
warning.
When the ALTAIR PRO goes into battery warning the:
battery outline indicator flashes
audible alarm sounds every 30 seconds
alarm lights flash every 30 seconds
Battery Shut Down
Attention!
If a Battery Shutdown condition occurs, stop using the instrument and
leave the area immediately. The instrument can no longer alert you of
potential hazards since it does not have enough power to operate
properly.
Failure to follow this warning can result in serious personal injury or
death.
When the battery can no longer operate the instrument, the instrument goes into
Battery Shutdown mode:
battery outline indicator flashes
alarm icon turns ON
alarm sounds every 30 seconds
alarm lights flash
no gas reading display
display alternates between "
BAT" and "ERR"
no other instrument pages can be viewed
The instrument remains in this state until it is turned OFF or the battery is
completely depleted. The alarm lights and the audible alarm can be silenced by
pushing the button.
When the instrument battery is running low, remove it from service and
replace the battery.
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
DESCRIPTION MSA
Confidence indicator
The Confidence indicator flashes once every 60 seconds to notify the user that
the instrument is ON and operating normally. In addition, both LEDs will flash
briefly every 60 seconds.
Backlight
The backlight can be manually activated with a quick push of the button. The
backlight remains ON for 10 seconds.
2.3. Monitoring Toxic Gases
With the ALTAIR PRO, you can monitor the concentration of the following toxic
gases in the ambient air:
The ALTAIR PRO Detector displays the gas concentration in parts per million
(PPM) on the measuring p
page is selected or the instrument is turned OFF.
S)
2
)
2
)
2
)
).
2
age. The instrument remains on this page until another
Attention!
If an alarm is triggered while using the instrument as a monitor, leave
the area immediately.
Remaining on site under such circumstances can cause serious
damage to health or can even lead to death.
There are four alarm setpoints in the instrument:
High Alarm
Low Alarm
STEL Alarm
TWA Alarm
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
MSA DESCRIPTION
The backlight turns ON for 20 seconds during an alarm condition.
If the gas concentratio
display and flash "
Fo
r more and detailed information Æ Section 3.3.
n reaches or exceeds the alarm setpoint the instrument will
LO" or "HI" and "ALARM" and enter an alarm sequence.
2.4. Monitoring the Oxygen Concentration
With the ALTAIR PRO the oxygen concentration is monitored in the atmosphere.
The alarm setpoints can be set to enrichment (concentration > 20.8%) or
depletion (concentration < 20.8%).
Attention!
If an alarm is triggered while using the instrument as a monitor, leave
the area immediately.
Remaining on site under such circumstances can cause serious
damage to health or can even lead to death.
When a set value is reached, an alarm is triggered:
an acoustic signal sounds,
alarm LEDs flash,
a vibration alarm is triggered and
in the display, "
and the corresponding oxygen concentration.
ALARM" is displayed in combination with "LO" or "HI"
Fo
r more and detailed information Æ Section 3.4.
The LOW alarm (lower of the two %O2 alarm settings) is latching and
will not reset when O
concentration rises above the LOW setpoint. To
2
reset alarm push the button
False oxygen alarms can occur due to barometric pressure (altitude)
changes or extreme changes in ambient temperature.
It is recommended that an oxygen calibration be performed at the
temperature and pressure of use. Be sure that the instrument is in
known fresh air before performing a calibration.
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
DESCRIPTION MSA
2.5. Duration of Use of the Instrument
Normal Operation
The instrument is ready for operation
as long as the battery warning indicator (Pos. 6 in Fig. 2) does not
display and
the self-test is successful.
Weakening Battery
A weakening battery is indicated by
the flashing battery warning indicator and
the remaining operation time indicator going off.
Attention!
You are not allowed to use the instrument even if it still shows a gas
concentration.
Flat Battery
A flat battery triggers the following alarm:
an acoustic signal sounds,
alarm LEDs flash,
battery warning indicator flashes and
the display shows "
ERR".
The alarm can be turne
Attention!
It is not permitted to continue to use the instrument!
The instrument must be taken out of use.
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
d off by pressing the button.
MSA OPERATION
3. Operation
3.1. Changing Alarm Setpoints
Alarm setpoints can only be changed manually during instrument setup,
which can be entered after battery attachment or reattachment (Æ Fig. 3).
Alarm setpoints can also be changed electronically using the
MSA FiveStar
(1) Ensure instrument is turned OFF before proceeding.
(2) Remove and re-attach battery.
(3) Turn instrument ON. Press and hold the button for 3 seconds.
ON displays during this time period.
®
Link™ Software.
The Steel Mill version cannot be turned OFF without removing the battery
or using MSA FiveStar
®
Link™ Software.
"ALARM", "SET" and "?" are displayed.
(4) To change the alarm setpoints manually, press the button once while
"ALARM", "SET" and "?" are displayed.
If the button is not pressed, the instrument turns ON, as usual, after three
seconds.
The Software Version displays for three seconds.
The gas type displays for three seconds.
["CO", "H
S", "O
2
", or "SO2" etc.].
2
The following alarm Setpoints are displayed:
Low Alarm Setpoint for three seconds, "
LO" and "ALARM" icons turn ON.
(5) To change the low alarm (LO), press the button when "LO" and "ALARM"
displays.
LO", "ALARM", "SET" and "?" are displayed.
"
(6) By pressing the button, increase the value for the low alarm, until the desired
value is reached.
On obtaining the maximum value, the display jumps and begins to count
upwards again with minimum value.
(7) Release the button and wait for 3 seconds.
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it
OPERATION MSA
High Alarm Setpoint for three seconds, "
(8) To change the high alarm (HI), press the button when "HI" and "ALARM"
displays.
HI", "ALARM", "SET" and "?" are displayed.
"
(9) By pressing the button, increase the value for the high alarm, until the
desired value is reached.
On obtaining the maximum value, the display jumps back and begins to
count upwards again from the minimum value.
(10) Release the button and wait 3 seconds.
Short Term Exposure Limits "
for three seconds.
(11) To change the STEL setpoint, press the button when the setpoint displays.
ALARM", "SET" and "?" are displayed.
"
(12) By pressing the button, increase the value for the STEL alarm, until the
desired value is reached.
On obtaining the maximum value, the display jumps back and begins to
count upwards again from the minimum value.
(13) Release the button and wait 3 seconds.
Time Weighted Average, "TWA" displays; then, the TWA setpoint displays for
three seconds.
(14) To change the TWA setpoint, press the button when the setpoint displays.
ALARM", "SET" and "?" are displayed.
"
(15) By pressing the button, increase the value for the TWA alarm, until the
desired value is reached.
On obtaining the maximum value, the display jumps back and begins to
count upwards again from the minimum value.
(16) Release the button and wait 3 seconds.
STL" displays; then, the STEL setpoint displays
HI" and "ALARM" icons turn ON.