MSA 697444 Instructions Manual

Bitrex Qualitative Fit Test Kit
instructions
YOUR MSA BITREX QUALITATIVE FIT TEST KIT INCLUDES:
1. Test hood
2. Nebulizer #1 (sensitivity)
3. Nebulizer #2 (fit test)
4. Sensitivity Solution
6. Two sets of replacement nebulizer inserts
PREPARATION
1. Connect snaps at the top of the hood,
so that the connections help to support itself.
2. Remove an applicator tube of Sensitivity
Test solution from its storage box. With the fill tip pointed upward, pinch along the full length of the applicator tube using your thumb and forefinger until glass ampoule inside of the applicator tube is thoroughly crushed.
It is important to keep the applicator tip pointed upward when pinching the tube. This is to prevent the solution from squirt­ing out during the pinching process.
3. Remove the stopper from the nebulizer
labeled “No. 1 - Sensitivity Test Solution”. Insert the tip of the applica­tor tube into the opening of the nebuliz­er. Squeeze along the applicator tube to transfer the solution from the appli­cator down into the nebulizer until the applicator is completely drained of the solution.
Do not attempt to remove the tip of the applicator. It contains a porous filter which is designed to allow the solution to pass through it while filtering out the
glass ampoule particles. It is also not necessary to puncture a hole or cut the tip of the applicator. The applicator tip already has the proper hole size in it to allow the solution to squirt into the neb­ulizer. See diagram on back.
NOTE It is recommended that the protective stopper be replaced back on the nebulizer until the test is ready to be performed.
4. Repeat steps 2 and 3 using an applica-
tor tube of Fit Test Solution and the nebulizer labeled “No. 2 - Fit Test Solution.”
5. Once the applicator solutions have
been drained into the nebulizers, the applicator tubes can be safely dis­posed.
SENSITIVITY TEST
This test is conducted to assure that the person being fit tested can detect the taste of the Bitrex solution at very low levels. The sensitivity test solution is a 100 to 1 dilution of the fit test solution.
The test subject should not eat, drink, or chew gum for at least 15 minutes before the test.
1. Explain the fit test procedures to the
subject.
2. Have the subject don the hood without
a respirator.
3. Position the hood forward so there is
about six inches between the subjects face and the hood. This is especially important for the fit test, it helps to ensure even dispersion of the aerosol and allows free movement of the sub­jects head while wearing a respirator.
4. Instruct the subject to breath through
their mouth.
5. Using the sensitivity test solution nebu-
lizer #1, inject the aerosol into the hood through the hole in the hood. Inject ten squeezes of the bulb, fully collapsing and allowing the bulb to expand fully on each squeeze.
6. Ask the subject if they can detect the
taste of the Bitrex aerosol. If tasted note the number of squeezes and pro­ceed to the fit test.
7. If the subject does not taste the sensi-
tivity solution, inject an additional 10 full squeezes of the aerosol into the hood. Repeat with 10 more squeezes of the aerosol into the hood if still not tasted.
8. If 30 squeezes of the nebulizer were
inadequate to produce a response from the subject, the test should be ended and another type of fit test must be used.
9. Remove the hood and give the subject
a few minutes to clear the taste from their mouth.
FIT TEST
1. Have the test subject don and fit check
the respirator per the instructions pro­vided with the respirator.
NOTE The fit test is to be performed with the test subject wearing a respirator with a particu­late filter or prefilter.
2. Have the test subject don and position
the hood as before, and to breath through their mouth.
3. Using the fit test nebulizer #2, inject the
fit test aerosol through the hole in the hood using the same number of full bulb squeezes as required in the sensi­tivity test, ie. 10, 20, or 30 squeezes.
4. To maintain an adequate concentration
of aerosol during this test, inject one­half of the number of squeezes used in step #3 above, every 30 seconds.
part no.
697444
TAL 109 (L) Rev. 3 © MSA 2001 Prnt. Spec. 10000005196 (C) Mat. 816975
Doc. 816975
WARNING
WARNING
5. Instruct the subject to indicate if
they detect the taste of Bitrex aerosol at any time during the test.
6. After the initial aerosol is injected (step 3),
instruct the test subject to perform the following exercises for 60 seconds each:
a. Normal breathing b. Deep breathing - breaths should be
deep and regular.
c. Turning head from side to side-
movement should be complete with one turn every second.
d. Nodding head up and down-move-
ment should be complete and made about once per second.
e. Talking, reciting the alphabet or
reading aloud the “Rainbow Passage” (see below).
f. Jog in place. g. Normal breathing.
7. If the entire test is completed without the
subject detecting the taste of the Bitrex aerosol, the test is successful and the respirator fit is deemed adequate.
8. If the test subject does detect the taste
of the Bitrex aerosol, terminate the test, (this indicates an inadequate fit). Wait 15 minutes and perform the tests over, starting with the sensitivity test. Fit test protocol meets 29 CFR Part 1910.134.
CLEANING/REFILLING
Immediately after completing the test, safely dispose of unused solution. Rinse the nebulizers with warm water to prevent clogging. Wipe out the inside of the hood with a damp cloth or paper towel to remove any deposited Test Solution. The Nebulizers must be thoroughly rinsed in water, shaken dry and refilled at least each morning and afternoon or at least every (4) four hours.
OSHA
The MSA qualitative fit test kit meets the performance criteria for fit testing respira­tors under the current OSHA standards.
OPERATING INSTRUCTIONS FOR NEBULIZER
1. Remove stopper (A) from vapor tube (B)
holding in horizontal position. Add solu­tion through vapor tube (B). Do not fill above O-Ring (D). Overfilling will cause nebulizer to leak.
2. Remove stopper (C) for maximum
aerosol delivery. Do not close stopper during test. You must hold nebulizer in upright position (as shown in picture) for it to function properly.
Replacement Parts
Part Number Description
697445 Sensitivity Solution 697446 Test Solution
697447
Nebulizer, Sensitivity
Test Solution 697448 Test Hood 697449 Training Video
697459
Nebulizer, Fit Test
Solution
Hold nebulizer in upright position as shown.
THE RAINBOW PASSAGE”
WHEN THE SUNLIGHT STRIKES RAINDROPS IN THE AIR, THEY ACT LIKE A PRISM AND FORM A RAINBOW. THE RAINBOW IS A DIVISION OF WHITE LIGHT INTO MANY BEAUTIFUL COLORS. THESE TAKE THE SHAPE OF A LONG ROUND ARCH, WITH ITS PATH HIGH ABOVE, AND ITS TWO ENDS APPARENTLY BEYOND THE HORIZON. THERE IS, ACCORDING TO LEGEND, A BOILING POT OF GOLD AT ONE END. PEOPLE LOOK, BUT NO ONE EVER FINDS IT. WHEN A MAN LOOKS FOR SOMETHING BEYOND HIS REACH, HIS FRIENDS SAY HE IS LOOKING FOR THE POT OF GOLD AT THE END OF THE RAINBOW.