cost effective stationary biogas analyser
accurate measurements, incl. plug&play, pre-calibrated sensors
safety in use with gas flow restrictor orifice at gas inlet and ventilation
sampling from low suction up to high pressure gas
sample gas conditioning for fast and reliable measurements
no dilution of the sample gas, nor use of compressed air is required
discontinuous measurement, user settable up to 24 measurements per 24 hours
up to 2 sites monitoring (time sharing technique) with only 1 analyser
IP 54 cabinet for use in harsh environment
ready to run delivery, minimum installation work, low service downtime
DATA SUBJECT TO CHANGE WITHOUT NOTICE
* overload for short term measurements only
Calorific value: 0 – 50 MJ/m3; MJ/kg
HMI human machine interface
Backlit keyboard, password protected operation
RS485 digital interface (Modbus RTU)
Data storage and event logging on SD card
Stainless steel flow restrictor orifice
Sample gas shut-down solenoid valve
Stainless steel gas fittings with 1/8" ID threads
Condensate trap with automatic condensate draining pump
Teflon particulate filter, internal Viton hosing
Sample flow 40 ... 60 l/h
Sample inlet pressure: -40 inH2O to +120 inH2O (-100 mbar to +300 mbar)
Sample venting: atmosphere pressure
15.74” x 19.68” x 11.81” (400 x 500 x 300 mm) ( H x W x D ) for wall or rack mounting
41°F …113°F (+5°C…+45°C) or -4°F …113°F (-20°C…+45°C) with cabinet heater
Indoor or outdoor (rain and sun shade is mandatory user scope of supply)
Universal 90 - 240 Vac / 47 - 63 Hz / 90 W (300 W with cabinet heater)
I/O-module with: 4x analog output 4...20 mA, floating, max. load 500R,
2x alarm relays, potential free contacts 24 Vdc/5A, DIN rail RS 485, Profi bus converter,
LEL (CH4) monitoring inside cabinet