Handheld analyzer with up to 5 electrochemical sensors and a dual gas NDIR bench
± 10 ppm or 10 % reading up to 2,000 ppm
10 % reading up to 5,000 ppm
± 5 ppm or 5 % reading up to 500 ppm
10 % reading from 500 to 2,000 ppm
± 10 ppm or 5 % reading up to 1,000 ppm
10 % reading up to 2,000 ppm
5 % reading < 4,000 ppm / 10 % reading > 4,000 ppm
± 2.0 ppm or** 5 % reading
with 0.1 ppm resolution **
5 % reading < 1,000 ppm / 10 % reading > 1,000 ppm
± 2.0 ppm or** 5 % reading
with 0.1 ppm resolution **
5 % reading < 200 ppm / 10 % reading > 200 ppm
± 2.0 ppm or** 5 % reading
with 0.1 ppm resolution **
*overload range recommend only for short time measurements
**are not separate sensors; selected sensors are used with special calibration
± 4°F ... < 392°F / 1 % reading > 392°F
(with stainless steel / Inconel steel tube)
± 4°F ... < 392°F / 1 % reading > 392°F
(with suitable material of sampling tube)
Stack / Differential pressure
± 0.01 inH2O or 1% reading
Gas flow velocity measurement
1 ... 40 m/s (using Pitot tube)
41°F .... 113°F, max. 95 % RH, non condensing
not in aggressive, corrosive or high dust environments, not for use in hazardous areas
Lithium-Ion battery, 6 to 8 hours operation
100 - 240 V AC / 50 … 60 Hz 1A
approx. 2.2 lbs. (with 7 sensors)
( W x H x D) 4.3” x 8.8” x 2.04”
Data subject to change without notice
MRU Instruments, Inc.
Humble, Texas 77338
Tel.: (832) 230 - 0155
Fax: (832) 230 - 1553
info@mru-instruments.com
www.mru-instruments.com