MRU Instruments 945142, 945132 User manual

MGA
prime
USER MANUAL
9452EN-PR
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Content
1 Introduction ................................................................................................................................. 5
1.1 Intended use .......................................................................................................................................5
1.2 About us ............................................................................................................................................... 6
2 Information for product and safety ......................................................................................... 7
2.1 Safety manual ....................................................................................................................................7
2.2 Safety precautions ............................................................................................................................ 7
3 Description ................................................................................................................................... 8
3.1 Task ........................................................................................................................................................8
3.2 Gas ow diagram .............................................................................................................................. 8
3.3 The measuring instrument ............................................................................................................ 9
3.4 Connectors ....................................................................................................................................... 10
3.5 Probes ................................................................................................................................................ 10
3.6 Gas sampling probe “TR” ............................................................................................................. 11
3.7 Gas conditioning ............................................................................................................................ 11
3.8 IR measurement ............................................................................................................................. 12
4 Operation .................................................................................................................................... 13
4.1 Commissioning ............................................................................................................................... 13
4.2 Switch on .......................................................................................................................................... 13
4.3 Switch o / Reset ........................................................................................................................... 13
4.3.1 Switch o ................................................................................................................................. 13
4.3.2 Reset .......................................................................................................................................... 13
4.4 Operating panel ............................................................................................................................. 14
5 Settings ....................................................................................................................................... 15
5.1 Analyzer settings ............................................................................................................................ 15
5.2 Setting time and date .................................................................................................................. 16
5.3 Conguration of measurement program ............................................................................. 16
5.4 Gas ow measurement ................................................................................................................ 18
6 Measurement ............................................................................................................................. 19
6.1 Preparation of each measurement .......................................................................................... 19
6.1.1 Power supply .......................................................................................................................... 19
6.1.2 Charging state of the battery ........................................................................................... 19
6.1.3 Connections to the instrument ....................................................................................... 19
6.1.4 Operating temperature ...................................................................................................... 20
6.1.5 Filter ........................................................................................................................................... 20
6.1.6 Switch-on, warm-up phase, zero point ......................................................................... 20
6.1.7 Instrument leak test ............................................................................................................. 21
6.2 How to take a Measurement...................................................................................................... 21
6.2.1 Fuel type selection and O2 reference ............................................................................. 21
6.2.2 Store the measurement results ....................................................................................... 22
7 Maintenance and cleaning ...................................................................................................... 23
7.1 Cleaning and maintenance ........................................................................................................ 23
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
8 Data memory.............................................................................................................................. 24
8.1 Organization of the Data memory .......................................................................................... 24
8.2 Information about the data memory ..................................................................................... 24
8.3 Site administration ........................................................................................................................ 24
8.4 Data transfer via USB (CSV export) .......................................................................................... 26
8.5 Export of measurements ............................................................................................................. 26
9 Extras ........................................................................................................................................... 27
9.1 Access key ......................................................................................................................................... 27
9.2 Internal Log Settings .................................................................................................................... 27
9.3 Service values .................................................................................................................................. 28
9.4 Analog output setup (4 – 20 mA) ............................................................................................. 28
9.4.1 Setting of lower limit (4 mA): ............................................................................................ 29
9.4.2 Setting of upper limit (20 mA): ........................................................................................ 29
9.4.3 Setting analog outputs during zeroing ........................................................................ 29
9.4.4 Pin assignment of the 4-20 mA interface ..................................................................... 30
9.5 Info ...................................................................................................................................................... 30
9.6 Options .............................................................................................................................................. 30
10 Information on the instrument components ....................................................................... 31
10.1 Firmware update ............................................................................................................................ 31
11 Specications ............................................................................................................................. 32
11.1 NDIR measured values ................................................................................................................. 32
11.2 Technical data ................................................................................................................................. 32
11.4 Gas sampling and conditioning ............................................................................................... 33
11.4.1 Electrochemical-, temperature- and pressure sensors ........................................... 33
11.5 Calculated values ........................................................................................................................... 34
11.5.1 Data communication........................................................................................................... 34
11.5.2 Analysis and calculations ................................................................................................... 34
11.6 Fuel types .......................................................................................................................................... 35
12 Appendix ..................................................................................................................................... 36
12.1 Error diagnosis regarding the measuring instrument ...................................................... 36
12.2 Insert a static IP-address .............................................................................................................. 37
12.2.1 Settings for the software MRU4win ............................................................................... 39
12.3 Spare parts ....................................................................................................................................... 41
13 Declaration of conformity ........................................................................................................ 42
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU

1 Introduction

y This manual enables you to understand and safely operate this MRU
Analyzer MGAprime.
y Please read this manual with great vigilant and get familiar with the
product before using it.
y This analyzer may only be operated by competent personnel and for its
intended use.
y Please pay special attention to all safety directions and warnings to
prevent personal injuries and damaging of the product.
y We can’t be held responsible for any injuries and/or damages that oc-
cur by not following the instructions in this manual.
y Always keep the manual near you when working with the analyzer, to
be able to read instructions as needed. Please ensure to hand over all documents to when handing the analyz­er over to others.

1.1 Intended use

The Analyzer MGAprime is designed for the gas analysis of ue gases, as they are emitted from gas/oil burners, engines, or heating and power appliances. The instrument is intended to support the user in control and indicative measurements in an ecient, accurate and reliable way
Introduction
The analyzer is specically not intended as a safety device or personal protective equipment; it should not be used as a warning device to warn people against the presence of harmful gases.
The instrument was manufactured according relevant normatives and regulations. It has to used within it’s intended use.
The Instrument must not be modied from the design or safety engineer­ing. Modications of any kind by the user will render the declaration of conformity.
This instrument meets the requirements of the valid European and national regulations. You can ind the declaration of conformity in the ap­pendix.
Syntax
Please note that this manual makes use of the scientic notation of gases (NO2), while the instrument itself and it’s screen shots display the gases in upper case letter only, i.e. (NO2).
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Information for product and safety

2 Information for product and safety

2.1 Safety manual

All general information and safety precautions of MRU products are listed in the supplied separate safety manual. Therefore this manual must be read and observed before the rst use of the instrument.
Instrument-specic safety and warning requirements in this manual are prexed before dangerous actions.

2.2 Safety precautions

The used category’s of safety precautions are here explained once more.
DANGER
Identies an immediate, impending hazard that, if ignored, will result in severe bodily injuries or death.
WARNING
Identies an immediate, impending hazard that, if ignored, may result in severe bodily injuries, material damage or death.
CAUTION
Identies a possibly dangerous situation that, if ignored, may result in minor injuries.
ATTENTION
Identies a possibly harmful situation that, if ignored, may result in dam­ages to the device or its surroundings.
NOTE
Identies user tips and other especially important information.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Description

3 Description

3.1 Task

The instrument is designed for the gas analysis of ue gases, as they are emitted from gas/oil burners, engines, or heating and power appliances. The instrument is intended to support the user in control and indicative measurements in an ecient, accurate and reliable way The instrument provides a full set of all equipment and sensors required for a emission control measurement:
y heated probe incl heated lter y heated sample line y gas conditioning unit including lters and gas cooler y gas pump and ow control y gas sensors
Available accessories include sensors for temperature or ow measurement. The user interface allows for a modern and intuitive way to operate the instrument. Running a commercial LINUX operating system, it allows as well for lot of options for data transfer and storage. For an overview on all available options please refer to the company’s home page or sales representatives.
3.2 Gas ow diagram
The analyzer draws a sample of the ue gases from the duct using a built­in gas pump through the probe is cleaned and dried using a gas cooler and built-in lter and analyses the extracted gas with electrochemical and NDIR sensors. Draft and temperature are measured at the tip of the sampling probe.
2 3 5 2 6 7
1
A
B
C
E
F
D
11 9 24810
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
A Fresh air inlet B Sample gas inlet
C Condensate outlet D Di. Pressure connector
E Vent collection box F Vent outlet
1 Sample gas lter (PTFE) 2 Dust lter
3 Auto-zero solenoid valve 4 Sample gas pump
5 Double stage gas cooler 6 Sample ow sensor
Oxygen sensor
7
O2-ECS or O2 paramagnetic
9 Infrared (NDIR) bank 10 Acrodisc PTFE lter
11 Di. pressure sensor

3.3 The measuring instrument

The measuring instrument consists of a compact and robust metal hous­ing with shock-absorbing rubber corners. All electrical and pneumatic connections are located on the both front sides of the instrument. It is operated exclusively via the touch-sensitive touch screen.
Description
8 Vent pump
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Description

3.4 Connectors

15
14
13
12
11
10
9
16 1 2
17 18
8
3
4
5
6
7
1 Loudspeaker 2 Ethernet (LAN)
3 USB socket 4 Second USB socket (Option)
5 RS485 (Option) 6 Analog outputs 4 ... 20 mA
7 Mains power supply 8 Sample gas lter
Condensate outlet port
9
Hose connection DN 4/6
11 Fresh air inlet port 12 Sample gas inlet port
13 Outlet fan of gas cooler 14 Outlet fan of gas cooler
15 Pressure-/di. pressure 16
17 Combustion air temperature 18 AUX socket

3.5 Probes

The Analyzer is available with dierent probes, both with xed and ex­changeable probe tubes.
y for high and less dust content y for fuel temperatures up to 800 °C (stainless-steel probe tube), y for fuel temperatures up to 1.200 °C (Inconel steel probe tube), y for fuel temperatures up to 1.700 °C (ceramic probe tube) y with and without heated pre-lter y with and without heated gas sampling line y probe tubes in dierent lengths, from 300mm to 2000mm
A complete list of available probes can be found in the current price list of this analyzer.
Sample gas outlet port (VENT)
10
Hose connection DN 4/6
Pressure-/di. pressure (Absolute pressure)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Description

3.6 Gas sampling probe “TR”

Heated probe with heated and exchangeable glass lter. The probe tube includes a gas temperature sensor and is available in dierent tube lenghts.
1
9
8
7
2 3 4
5
6
1 Probe handle 2 Probe tube
3 Fast locking coupling 4 Probe cone
5 Cable plug (14-pin) 6 Heated hose line
7 Cable coupler (5-pin) 8 Fast locking coupling
9 Filter lock

3.7 Gas conditioning

The sucked sample gas is dried and ltered before it is fed to the sensors. A double stage sample gas cooler with Peltier element is used for dry­ing. The condensate liquid appearing in the gas cooler is pumped to the condensate outlet by means of a peristaltic pump. The condensate forms drops at the outlet of the instrument. Optionally, connect a hose (DN 4/6) to the condensate drain.
The VENT output delivers the sample gas after the analysis stage. If
For subsequent ltering, a round lter is used on the front of the measur­ing instrument.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Description

3.8 IR measurement

The instruments NDIR gas sensor is able to detect up to 8 dierent gases. It is most advanced in terms of its long-term stability due to a dedicated stabilization technology including a permanent zeroing by operating the bench at two dierent gas pressure values.
Due to its low noise and being drift free, it is perfectly suited to long­term measurements.
Principle of the IR-bench (NDIR)
21 3 4 5 6
1 IR source 2 Gas entry
3 Sample gas cell 4 Gas exit
5 Band pass lter 6 IR detector
An infrared source delivers IR radiation in the wavelength range between 1 and 10 um, which is relevant for the absorption of gas components to be measured.
The target gas absorbs a portion of the IR radiation, which is detected by a wavelength selective detectors and the end of the sample gas cell. The absorption value is correlated with the gas concentration, while all eects of cross sensitivity to other gases are corrected by an internal software algorithm.
As the IR bench is operated successively at two dierent gas pressure values, it is possible to eliminate all drift eects, which would otherwise contribute to the absorption signal.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU

4 Operation

4.1 Commissioning

Operation
The instrument is delivered as a complete assembly ready for use.
f Check the instrument regarding condition and integrity after delivery. f Connect the instrument to the power grid.
Ö The instrument switches on and start the operating system. Ö Blue LEDs for ON and power supply are switched on. Ö The instrument runs a start procedure which includes:
y self test y warm-up of the NDIR bench y cool down the double stage gas cooler, indicated by the symbol y Zeroing, indicated by symbol
f Charge battery for more than 8h is recommended after rst start to
allow the battery to charge completely. Operation of heated probe and sample line is only supported when connected to power grid.
Ö The battery is charged as soon as teh connection to power grid is
established.
Ö The blue LED will be blinking slowly.
Heating of probe and probe tube are unsupported in battery mode.

4.2 Switch on

f Touch the
Ö LED lights blue
f Release the ⏻ button
Ö LED lights red, analyzer runs up
button for 3 sec. minimum
4.3 Switch o / Reset
4.3.1 Switch o
f Touch the “Context menu” on the display f “Turn instrument o”
Ö Do you wish to turn instrument o?
f “YES”
Ö “The system will shut down”
or
f Touch the
Ö Do you wish to turn instrument o? Ö “YES” Ö “The system will shut down”
button

4.3.2 Reset

f Touch the ⏻ button during ashing LED for 30 sec. minimum
f After change to continuous lighting, release the ⏻ button
Ö The instrument will switching o with reset
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Operation

4.4 Operating panel

All functions are controlled via the touch surface of the instrument. Dif­ferent gestures are available in the individual menus and windows.
1412 1311 18
1 Power-on and reset
2 Reserve
3 Reserve
4 LED display mains operation/battery charging mode
16 17
8910 6
7
1915
1
2
3
4
5
5 Reserve
6 Current ow rate
7 Current temperatures heated hose
8 Selected measuring program, e.g. Test or measurement program
9 Current temperatures of NDIR bench
10 Current temperatures of gas cooler
Access to detailed information on the instrument components.
11
Especially for service or inquiry
12 Menu info
13 Menu contacts
Status bar: display of zero point, alarms, executed measuring pro-
14
gram, selected fuel, heat-up-, cool-down phase
15 Menu measure
16 Menu settings
17 Menu extras
18 Battery Charge indicator
19 Context menu with window-dependent additional functions
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU

5 Settings

5.1 Analyzer settings

Settings
After the analyzer has been inspected and is ready for start-up it can be switched on and personalized settings can be entered. These settings can be changed at any time.
After the analyzer has been inspected and is ready for start-up it can be switched on and personalized settings can be entered. These settings can be changed at any time.:
Country Option By changing the country the
O2 reference values settings are lost. The fuel list is reset. Similarly, country-specic
defaults and methods of measurement are selected as a result. Ensure the correct setting of the country in which you are performing the measure­ment to ensure that all relevant ones are set up
Language Option Select instrument languages
Modbus slave Modbus address of the instrument for the
remote control via Modbus
Temperature heating hose
Combustion analysis
130°C ­+180 °C
on/o
Setting combustion analysis to ON allows for several calculation performed by the instrument, which are fuel type dependent, e.g. Lambda, heat loss.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Settings
Negative gas readings
VNC on/o VNC Viewer for remote control via LAN
Reference tem­perature
Interval auto-zero
on/o Negative gas readings Caused by tem-

5.2 Setting time and date

In the Extras menu, you can check the date and time and, if necessary, set it. When connected to the Internet, the time automatically synchronizes itself.
perature drift of a sensor are suppressed (shown as zero) or displayed
Calculation of the values for a standard state at reference temperature
Interval time after which the instrument performs a zero-point
f The time and date can be set in the middle black number series by
pushing up and down the gray number row.
f Accept the changes with ACCEPT.
5.3 Conguration of measurement program
The Analyzer provides in the Measure menu various measurement programs, which can be selected in the main window. Each measuring program denes the properties of the measuring window:
y Measurement program
The measured value window can be freely set with regard to the dis­played measured values, of the underlying fuel. The purge limit of the CO sensor (if available) can be set
y Test program
The measured value window shows predened values and cannot be changed. Can be used in instrument testing to obtain standardized dis­plays, e.g. can easily be checked with test gases.
y If so, further measuring programs
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Settings
The measurement window for gas analysis can be congured and adapt­ed to your needs. The measuring window initially displays 12 measured values, by a scrolling gesture it will display more values.
Moving a value eld
f Touch and hold the value eld.
Ö Value eld will be framed.
f Move the value eld to a dierent position.
Ö The other value elds move automatically.
Assign a measured value
f Double touch the value eld.
Ö A list with of all available measured values is displayed.
f Choose the wanted value and “replace”.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Settings
5.4 Gas ow measurement
Set up measurement window
With the ow measurement option are further measurements available:
y v-ow y Flow rate y Mass ow carbon
Parameter for gas ow measurement
f Choose the right context button - menu item “gas ow measurement”.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU

6 Measurement

6.1 Preparation of each measurement

6.1.1 Power supply

The analyzer can be operated with an internal battery to warm up the instrument or to use internal instrument functions. A mains connection is required heating hose.

6.1.2 Charging state of the battery

The battery symbol in the display indicates approximately the remaining capacity of the battery. From 2% remaining capacity, the charging indicator starts to ash red. If the instrument is not connected to the mains power supply within one minute, the analyzer switches o to prevent battery discharge. Even when the battery is discharged, the instrument can be operated completely on the power supply.

6.1.3 Connections to the instrument

Measurement
for the measurement including heated gas sampling probe and
CAUTION
Acid from the condensate
Acid burns may result from weakly acidic liquids from the condensate.
f If you come into contact with acid, wash the area immediately using a
lot of water.
DANGER
Risk due to toxic gases
There is a risk of poisoning. Noxious gases are sucked in by the measuring device and released into the ambient air.
f Only use the measuring device in well ventilated spaces.
f Connect the gas sampling probe to the instrument (gas plug and elec-
trical plug).
f Consider to connect a hose or collecting container if necessary to the
condensate outlet when appropriate.
f The hose and any connected reservoir connected to the outlet must
not be closed against ambient or include an air outlet to avoid over­pressure.
f Please note that measuring gas may leak at the instrument side or at
the VENT outlet. Connect a pump to the VENT output to collect the sample gas. If an “active VENT” option is installed, an internal gas pump feeds the sample gas completely to the VENT outlet, where an evacuat­ing gas hose should be connected.
f In the case the option active vent is installed, note that the sample gas
at the exit may be diluted with ambient air and is therefore not ade­quate to be reused in the process.
f Ensure that ambient fresh air can be sucked in at the fresh air inlet.
Consider to connect a hose leading to fresh ambient air, when the instrument’s environment contains toxic gases or a high CO2 level.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Measurement
f The ow rate of the gas should be within the specied range. Other-
wise, please check probe and lter for clogging
f Temperatures of NDIR and heating hose should be within the specied
range in order to guarantee a sucient measuring accuracy.
In the Extras menu under “Connections” the connection drawings are displayed:

6.1.4 Operating temperature

The internal gas cooler operates at 5°C, which is the dew point of the sample gas to the sensors. Components along the gas line may be dam­aged if they are colder than 5°C and condensation appears internally. Therefore, if the Analyzer has been stored very cold below 0°C, it is essen­tial to wait for the Analyzer to warm up in a warm environment in order to avoid such condensation! In such cases, take a typical warm-up time for the instrument of one hour into account, especially when wet ue gases are to be measured. If the operating temperature is not within the permissible range, a corre­sponding message is displayed.

6.1.5 Filter

The probe lter and the round lter must be checked before and after each measurement

6.1.6 Switch-on, warm-up phase, zero point

After switching on, the instrument can always be operated, even if no gas analysis can take place during the warm-up phase. The instrument inde­pendently performs the following actions during the warm-up phase:
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Measurement
y Heating the probe and the heating hose y Warm up the NDIR bench (if available) y After the operating temperatures have been reached, the gas pump is
switched on and the analyzer takes the zero point with fresh air.
y After the zeroing, the analyzer is ready for operation
If a new zero point is required by further heating the instrument, it can be started via the context menu.

6.1.7 Instrument leak test

f Check all connections for correct t. f Check all hoses and hose connections (from the tip of the probe to the
gas inlet of the measuring instrument) for leaks.
The Analyzer has a built-in “leak test” in menu Extras for checking the tightness of the gas paths. This is done by measuring the remaining gas ow when the gas path is closed.
f Seal the sample gas inlet f If the complete gas path is to be tested for leakage - at the probe tip.
If the system is tight, the trac light is green and the l / h pointer points
to 0.

6.2 How to take a Measurement

6.2.1 Fuel type selection and O2 reference

The fuel selection can be selected in the menu Measure with the O2 reference / fuel in the “context menu”. For this, the combustion analysis have to be switched on, also see 5.1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Measurement

6.2.2 Store the measurement results

The measured values can be stored in a site via the context menu / Store menu entry. The measurement itself continues until switch o of the analyzer.
Continuous data logging
A continuous measurement logging is possible via the context menu / “Start logging”.
With “Start Logging”, the currently measured values are stored in a site every 10 seconds.
End the data logging with “Stop logging”.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU

7 Maintenance and cleaning

7.1 Cleaning and maintenance

The Analyzer needs only low maintenance eort for long value preserva­tion
Acid from the condensate
Acid burns may result from weakly acidic liquids from the condensate.
f If you come into contact with acid, wash the area immediately using a
lot of water.
After every measurement:
f Remove the gas sampling tube from the analyzer, so that the hose can
dry.
Occasionally:
f Cleaning of the probe and the probe tube. f after longer disuse load battery rst and afterwards approx. all 4 weeks. f Check the lter in the probe head (if available), replace if necessary. f Check the round lter at the front of the instrument and replace if
necessary.
Maintenance and cleaning
CAUTION
Maintenance
An annual service check and if necessary adjustment of the sensors at an MRU service department (www.mru.eu) are recommended for the pres­ervation of value.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Data memory

8 Data memory

8.1 Organization of the Data memory

Basis of the data memory of the analyzer is a set of sites stored in the instrument. Every site exists of a unique site number and 12 freely usable text lines which can have, e.g., the address, customer name etc. The instrument can store up to 1.000 dierent sites. Sites can be created in the instrument and be changed. Measurements are stored by assigning them to a site. Measurements can be, on this occasion, singles ue gas measurements or other measuring programs available in the instrument.

8.2 Information about the data memory

In the menu Contacts, item “storage” you select “memory info “to get information about the actual memory volume. The part of free memory, the total number of the stored sites and the number of the measure­ments stored all together, split in the kind of the measurement is listed.

8.3 Site administration

Available sites are listed in this menu
The “ADD SITE” button creates a new site number. Further changes may be introduced after selecting a specic line.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Data memory
y Changes in the site description may be entered and stored y A site may be deleted by the “delete” button. y Measurements assigned to the selected site are displayed when the
“Measurement” button is activated.
In the menu Measure you can see stored measurements
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Data memory

8.4 Data transfer via USB (CSV export)

The data exchange format is CSV. A character-separated values (CSV) le is a simple text format for a database table. Each record in the table is one line of the text le. Each eld value of a record is separated from the next by a character. The analyzer uses a semi-colon ‘;’ as value separator (other implementations use sometimes a comma). CSV is a simple le format that is widely supported, so it is often used to move tabular data between dierent computer programs, for example Microsoft Excel™ or Access™, that support the format. Also other computer programs oer this type of interface because it is widely spread and easy to use.
The following functions are available
y Export of sites y Export of ue gas measurements
f Insert USB Stick f Select “site” and “measurement” in Contacts menu f Now press “EXPORT CSV”
After this… you nd in the USB stick directory “1113Export” the export les. The le name is such as those “09_04_2018_15_02_02__Measurement__ Natural_gas.csv”….
f Open this le (*.csv) with Excel…. (don’t use the dat les… don’t use
“Internal log settings” are only for internal use or diagnostics)

8.5 Export of measurements

This function is used to export the measurements from the analyzer to a computer program. Attention, This function is not suitable for back-up or for the transfer to another analyzer because the exported le cannot be imported again! The created le has the name ‚EMIxxxxx.csv’, in which the xxxxx are con­tinuing 5 digit numbers with leading zeros. The created le has a column header with the following information: Site number, Date/Time, Measuring program name, Fuel type, CO2max, O2reference, and all mea­sured values that the analyzer can measure.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU

9 Extras

Extras
IIn the Extras menu, the time / date setting (chapter 5.2) and the connec­tion possibilities of the instrument (chapter 3.4) are further menu items for service purposes available.
The rst menu items have already been explained during “settings”.

9.1 Access key

The entry of an access key (password) is for maintenance action and allows experienced users to operate on the operating system level. Not required for regular use.

9.2 Internal Log Settings

The analyzer stores internal parameters at regular intervals to allow for an optimized support from experienced service sta. These les can be copied to a connected USB stick and sent via email if required to do so.
f Copy internal log to USB stick.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Extras

9.3 Service values

This screen displays a number of internal parameters and their values.
In case of unexpected behavior of the instrument it might be helpful to communicate those values to our worldwide service sta: www.mru.eu

9.4 Analog output setup (4 – 20 mA)

There are 8 analog outputs available (4-20 mA). Each analog output (channel) is assigned a measured variable and an output range.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Extras

9.4.1 Setting of lower limit (4 mA):

This setting determines the lower end value, corresponding to 4 mA. If the measured value falls below the set value, the analog output stops at 4 mA.

9.4.2 Setting of upper limit (20 mA):

This setting determines the upper end value, corresponding to 20 mA. If the measured value rises above the set value, the value stops at 20 mA.

9.4.3 Setting analog outputs during zeroing

The following settings are possible
y Hold - The outputs kept the last values from before zeroing y 2mA - The outputs change to 2 mA to indicate the zero point
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Extras

9.4.4 Pin assignment of the 4-20 mA interface

9.5 Info

In the Info menu, version information and installed options can be viewed.

9.6 Options

Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Information on the instrument components

10 Information on the instrument components

Especially for service or inquiry This menu can be selected from the main menu with the adjacent but­ton. In addition to special instrument information and the possibility of instru­ment matching for service stations, a rmware update is also possible.

10.1 Firmware update

Performing an update:
f Please extract the obtained le 1113Update.zip. f Copy the unzipped le ‘1106.fwb’ to a USB stick in the root directory. f Switch on the analyzer... f Insert the prepared USB stick into a USB socket of the analyzer. f Press
f Select the menu item “Software update”
f “Load from USB Stick” f After copying to the analyzer press the button “Perform rmware up-
date”
Ö The rmware update starts
f After updating, turn the analyzer o
Ö After rebooting, the new rmware is installed.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Specications
11 Specications

11.1 NDIR measured values

Gas Range Resolution Accuracy
NO 250 - 4000 ppm
NO
SO
CO
2
2
2
200 - 1000 ppm
200 - 4000 ppm
40 Vol% 0,01 Vol%
CO 200 - 10000 ppm
N2O 200 - 1000 ppm
CH
4
500 - 10000 ppm
HC (C3H8) 200 - 10000 ppm

11.2 Technical data

Operating temperature +5°C - +45 °C
Rel. Humidity, non-condensing 90 %
Storage temperature -20 °C - +50 °C
Internal Battery Pack, capacity, operating hours w/o gas cooler and heated sample line
1 ppm
1 ppm
1 ppm
1 ppm
1 ppm
1 ppm
1 ppm
Repeatability: 1% v.MB. (CH4: 2% v. MB) 8h-Drift: 1% v.MB. Linearity: 2% v. MB.
Li-Ion, 96 Wh, 6 h
Display 7” touch, < 750 cd/m2, 800*480 px
Power supply (w/o heated sample line)
Weight instrument w/ 2 EC sen­sors.
Weight instrument w/ 2 EC sen­sors plus case
86 - 265 V / 47 - 63 Hz / 105 W
7,5 kg
16 kg
Size incl Case (WxHxD) 43 cm x 15 cm x 29 cm
Size incl Case (WxHxD) 51 cm x 29 cm x 51 cm
Housing Material Aluminium / TPU
IP degree of protection IP20
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Specications

11.4 Gas sampling and conditioning

11.4.1 Electrochemical-, temperature- and pressure sensors

Electrochemical sensor O2 Long Life
Measuring range 0 - 25 Vol%
Resolution 0,01 Vol%
Abs. accuracy ± 0,2 Vol%
Response time T90 < 20s
Years expected lifetime (@air) 5 Jy
Paramagnetic sensor O2
Measuring range 0 - 25 Vol%
Resolution 0,01 Vol%
Accuracy 0,1 Vol%
Flue gas temperature TA
Measuring Range with high grade steel probe pipe
Measuring range with Inconel probe pipe Short time only (up to 20 mins
0 - 800° C
0 - 1100° C 0 - 1350° C
±2° C /
Accuracy abs. / reading
1%
Air temperature Tl
Measuring range 0 - 100°C
Abs. accuracy 1 °C
Draft
Measuring range ± 120 hPa
0,02 hPa
Accuracy abs. / reading
1 %
Differential Pressure
Measuring range ± 120 hPa
Accuracy abs. / reading
0,02 hPa
1 %
Barometric Pressure P
abs
Measuring range 300 - 1200 hPa
Accuracy ± 3 hPa
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Specications

11.5 Calculated values

Velocity
v
based on dierential pressure mea­surement with Pitot tube
Typical measuring range 3 m/s - 100 m/s
Accuracy at 3 m/s 1 m/s
Accuracy > 12 m/s (reading)
± 1%
Resolution 0,1 m/s
Absolute pressure measurement

11.5.1 Data communication

USB interface master only (for connection to USB stick or accessories)
Support of external SD card reader Ο
Ethernet, RJ45 Ο
WLAN Ο
Bluetooth Ο
Ο
RS485 (AUX socket, for connection of external sensor mod­ules)
Ο
RS485 (isolated, for connection to PC) Ο
Analog I/O: 4x input, 8x output, 4... 20 mA Ο

11.5.2 Analysis and calculations

Continuous measured values
O
2
Unit
[%]
Temp. ambient air (thermocouple) [°C]
Temp. ue (thermocouple) [°C]
CO [ppm]
Draft [hPa]
Further continuously calculated values
Unit
ETA [%]
ETA condensed [%]
Losses [%]
Losses condensed [%]
Lambda -
Dew point [°C]
CO/CO2 ratio [%]
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Specications

11.6 Fuel types

This list is for Germany only. Fuel types from other countries can be obtained from MRU GmbH: Web page: www.mru.eu
Germany O2max 20,96
Fuel CO2max A1 A2 B
Test gas 0,0 0,00 0,00 0,000
Natural gas (LL) 11,8 0,37 0,66 0,009
Natural gas (E) (*) 12,1 0,37 0,64 0,009
EL heating oil 15,4 0,50 0,68 0,007
S heating oil 15,9 0,50 0,66 0,007
P/B liquid gas 13,7 0,42 0,63 0,008
Propane 13,7 0,43 0,66 0,007
Butane 14,1 0,45 0,67 0,007
Biodiesel 15,7 0,46 0,62 0,005
Dry wood 20,3 0,60 0,62 0,009
Pellets 20,3 0,74 0,77 0,000
Coal 19,1 0,59 0,65 0,009
Lignite 19,4 0,39 0,42 0,009
Peat 19,8 0,66 0,70 0,010
Coke oven gas 10,8 0,29 0,60 0,011
Coal gas 11,7 0,35 0,63 0,011
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Appendix

12 Appendix

12.1 Error diagnosis regarding the measuring instrument

Fault indication Possible causes Repair
Gas cooler is faulty! The system will shut down…
Undervoltage!!! The System will shut down…
Power consumption too high” Please check heated hose!
Hose temperature is set to 160°C
Please wait – pump is o Warm up has not been completed yet.
Flow monitoring! Flow rate too low! Please check lter
Gas cooler faulty Contact MRU service
department
Battery is discharged Connect instrument
to power grid.
Please check heated sample line.
Sample line temperature set to
160 °C.
Heated sample line is
only intended to op-
erate at higher tem­peratures than 160
°C for short time. !
Measurement not started as instrument
is in warm-up phase..
The sample gas ow has fallen below the required limit.
Filter or sample line
may be clogged. Gas pump may be faulty Flow sensor may be defect.
Heated sample line
may be damaged.
Disconnect the
instrument from
power grid and
visually inspect the sample line. Take
care to unroll the heated sample line
before operating it. During next pow-
er-up the tempera-
ture is reduced to
160 °C again..
Wait until warm up phase has
terminated.
Check sample line
and all lters on dust or water. Replace lters.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Appendix
Fault indication
Possible causes Repair
Condensate or water has been detected behind the gas cooler
stage. To protect the sensors
the pump has been switched
o..

12.2 Insert a static IP-address

f To continue operation it has
to be ensured that no water
will be fed to the sensors.
f Remove sample line. Check
external lter on possible wa­ter content. Remove water
f Use the peristaltic pump to
drain water. Therefore press the corresponding display button.
f Repeat if necessary. f In a second step, if the pro-
ceeding steps removed all
water possibly present in the gas cooler and hoses, the gas pump may be used to dry out remaining water droplets.
f Choose “network” in menu Extras.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Appendix
f Put the switch on “DHVF is active”” f Choose the desired IP address, subnet mask, standard gateway and
preferred DNS server.
f Conrm with “OK”.
Ö The modication will be active after restart.
The input of subnet mask occurs to following principle:
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Appendix

12.2.1 Settings for the software MRU4win

The PC program MRU4win can be used for a LAN connection of the ana­lyzer.
f Connect the analyzer with the network.
Read the IP address
f Select the “display setting” in the main menu using the context key.
Ö The IP address is shown at the top line.
f Modbus slave ID set to 1.
Adjust MRU 4 Win to PC
f The Modbus must be activated in the Setting menu
f „Create a Modbus Device“
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Appendix
f The IP address of the analyzer must be inserted.
Ö After these settings, the analyzer connect to the PC.
Info about the network
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Appendix

12.3 Spare parts

Part number Spare part
56879A PTFE Round lter
61158 Probe lter sintered metal 2 µm
61157 Probe lter sintered metal 20 µm
10825 Mineral wool lter element
59799 O-ring 16 x 1,5
61066 O-ring 12 x 2
61333
60074 O-ring 8 x 2
O-ring 10 x 2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MRU
Declaration of conformity

13 Declaration of conformity

Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...