MRU Instruments 61417RT, 63467HT User manual

Page 1
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
Inspect Shipment for Damage
Carefully inspect the entire shipment for damage in the presence of the shipper’s agent, removing packaging material if necessary. Note any damage to packaging and/or goods on Packing List and have it signed by the shipper’s agent prior to accepting the shipment. Submit damage claim to MRU immediately.
NOTE: Damage claims not received by MRU within 3 days of receipt of shipment will not be accepted.
Save the original box and the packing material for use if the analyzer must be shipped in the future.
The products described in this manual are subject to continuous development and improvement and it is therefore ac­knowledged that this manual may contain errors or omissions. MRU encourages customer feedback and welcomes any comments or suggestions relating to the product or documentation.
This manual is intended solely as a guide to the use of the product.
MRU shall not be liable for any loss or damage whatsoever arising from content errors or misinterpretation of informa­tion’s from this manual or any mis-use resulting from the use of this manual.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3
Page 3
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
1 Content
1
Content 4
2
INTRODUCTION 6
2.1 The company MRU GmbH........................................................................................................................ 6
2.2 OMS 420.................................................................................................................................................. 7
RETURNED GOODS 7
3
3.1 Return of analyzer according to ElektroG................................................................................................. 7
4 Safety 8
5
O2 sensor – operating theory 9
6
COe sensor – operating theory 10
7
System components, general information 11
7.1 Optimizing combustion ..........................................................................................................................11
7.2 OMS420 features ................................................................................................................................... 12
7.3 Probe models ......................................................................................................................................... 13
7.3.1 Compact model OMS420 (# 61417) ........................................................................................13
7.3.2 Remote transmitter model OMS420RT (# 63467RT)................................................................. 14
7.3.3 High temperature model OMS420HT (# 63467HT) ...................................................................15
7.4 Principle of flow guidance tube.............................................................................................................. 16
7.5 Transmitter electronics...........................................................................................................................17
7.6 Manual calibration................................................................................................................................. 18
7.7 Recommended calibration gas ............................................................................................................... 18
Software flow chart 19
8
9
Operation 20
9.1 Start-up.................................................................................................................................................. 20
9.2 Warm-up................................................................................................................................................ 20
9.3 Main measuring menu ........................................................................................................................... 21
9.4 Info menu............................................................................................................................................... 21
9.5 Service menu.......................................................................................................................................... 22
9.6 Settings menu ........................................................................................................................................ 23
9.6.1 PIN code level 1........................................................................................................................23
9.6.2 Set analog output..................................................................................................................... 24
9.6.3 Damping................................................................................................................................... 24
9.6.4 Back-purge setup...................................................................................................................... 24
9.6.5 Interval auto-calibration ...........................................................................................................25
9.6.6 Set calibration cylinder values .................................................................................................. 25
9.6.7 Set temperature measuring unit ............................................................................................... 25
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.6.8 Set gas humidity....................................................................................................................... 26
9.6.9 Solenoid test............................................................................................................................. 26
9.6.10 OMS420 settings ...................................................................................................................... 26
9.6.11 Reset to (factory) defaults ........................................................................................................ 27
9.7 Calibration ............................................................................................................................................. 28
9.7.1 OMS420HT model compact probe (# 61417)............................................................................ 29
9.7.2 OMS420RT model remote transmitter (# 63467RT).................................................................. 30
9.7.3 OMS420HT model high temperature (# 63467HT).................................................................... 31
9.7.4 Pin Code level 2........................................................................................................................32
9.7.5 O2 cell setup (by factory only) .................................................................................................. 33
9.7.6 CO cell setup (by factory only).................................................................................................. 33
9.7.7 Adjustment O2 cell amplifier (by factory only).......................................................................... 34
9.7.8 Adjustment 4 mA – 20 mA (factory only) .................................................................................35
9.7.9 Calibration AIR ......................................................................................................................... 36
9.7.10 Calibration AIR & GAS.............................................................................................................. 36
9.7.11 CO correction factor.................................................................................................................. 38
9.7.12 O2 heat resistor regulation (factory only) ................................................................................. 39
9.7.13 CO heat resistor regulation (factory only)................................................................................. 39
9.7.14 History auto-calibration............................................................................................................ 40
Troubleshooting 40
10
11
Technical specifications 41
12
Appendix 42
12.1 Error codes............................................................................................................................................. 42
12.2 Modbus Slave specification....................................................................................................................43
12.3 Declaration of confirmity........................................................................................................................ 44
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5
Page 5
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
2.2
Thank you for purchasing the MRU OMS420 In-Situ oxygen and combustibles (O
OMS 420
2 and COe monitoring probe.
Please read this instruction manual carefully before attempting to operate the analyzer. After you have become familiar
with this manual, move on to installation, operation and maintenance of the analyzer. Incorrect use of the analyzer could cause an accident or injury.
Product development and improvement are dynamic goals of MRU, and specifications of this analyzer are subject to
change without prior notice.
Modification of this analyzer is strictly prohibited unless written approval is obtained from the manufacturer. MRU will
not be responsible for any issues of any kind resulting from any modification made to the analyzer without written per­mission.
It is important that this manual remains in the custody of the actual operator of the analyzer.
After reading the manual carefully, it should be stored in a safe, but accessible place.
This instruction manual should be delivered to the end user immediately upon delivery.
NOTICE:
It is prohibited to transfer part or all of this manual in written format without MRU written permission
Product development and improvement are dynamic goals of MRU, and descriptions and illustrations of the analyzer
used herein are subject to change without prior notice.
Please note:
Our warranty and guarantee obligations for OMS420 do not cover the usage of the analog signal 4 - 20 mA for regula­tion- and control-purposes. We exclude any liability for consequential damages.
3 RETURNED GOODS
Packing regulation of 12.07.1991
If your local waste facility does not except MRU packing materials for disposal, you may return it to MRU or our local sales representative. Packing materials returned to MRU must be returned prepaid.
3.1
Return of analyzer according to ElektroG
MRU GmbH is required to accept the return, for proper disposal, of all analyzers delivered after 13th of August 2005. Analyzers must be returned to MRU prepaid.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7
Page 6
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
4 Safety
The O2 probe may only be used in original, undamaged condition and in accordance with the operation manual.
All individuals dealing with the installation, commissioning, operation and maintenance of the analyzer or probe must be
qualified to do so and must strictly observe this operation manual.
Unauthorized modifications to any part of the analyzer or probe can create safety risks and are not permitted.
Power other than that specified in this manual must never be provided to the probe.
Service of transmitter electronics by non skilled personnel is not allowed.
Do not allow condensate to come into contact with the sensors.
Do not attempt to clean the probe with water.
The probe shall not be used in under-stoichiometric combustion conditions, due to the possible presence of flammable
gases, eventually over the LEL (low explosion level).
Power must always be provided to the probe, even during boiler shut-down, to prevent the formation of condensate
which can damage the sensors.
Do not use the probe for any purpose other than that specified in this manual.
Exposure to corrosive gases such as silicone vapor, alkaline and heavy metals, P, Pb, high SO2, etc. will shorten the life
time of the sensors.
It is mandatory to the user to insure that all persons operating this equipment are properly trained in
its operation and fully understand the operating principals of the equipment.
MRU GmbH, its affiliates and agents cannot be held responsible in any way for damage or injuries re-
sulting from improper use, misuse or neglect in operating this equipment.
Caution
Probes installed inside flue ducts and stacks operate at elevated temperatures (often 1.000 ate danger of serious skin burns to operators if proper handling precautions and extreme care are not taken.
o
F and higher) cre-
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8
Page 7
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
5 O
Heated zirconium oxide (ZrO2) is used as a ceramic solid electrolyte that is a good oxygen ion conductor at temperatures of ap­proximately 1.550
sensor – operating theory
2
o
F (850 °C), generated by an internal low power (20 W) heater element. The heater element is a PTC type, self­regulating device that does not require a thermocouple for temperature regulation. Constant sensor temperature is maintained by controlling the heater voltage and current to fixed resistance of the heater element.
The electro-motive force (emf ) that is generated across the solid electrolyte by the presence of oxygen ions can be measured as a sensor voltage (according to Nernst law).
where:
U
= offset voltage (for P
0
O2 ref
= P
R = universal gas constant
T = zirconium temperature
F = Faraday constant
P
= oxygen partial pressure reference side
O2ref
P
= oxygen partial pressure sample side
O2sample
O2sample
)
This voltage is measured by micro-controller based transmitter electronics and converted into a standard 4 – 20 mA signal, lin­earized for oxygen in the range of 0 – 25 %.
The expected lifetime of this sensor is about 5 years + under normal operating conditions, and is not dependent on fuel type, but:
CAUTION:
If combustible gas (CO, H
2, HC) in high concentrations are present in the sample gas, erroneous O2 readings will re-
sult due to local combustion at the sensors’ hot surface.
Exposure to corrosive gases (silicone vapor, alkaline and heavy metals, P, Pb, high SO
2, etc.) will shorten the life of
the sensor.
Condensation of flue gas moisture close to the sensor’s flange must be prevented.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
9
Page 8
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
6 COe sensor – operating theory
A solid ceramic electrolyte with thin-layer technology is used to measure the combustible gases (CxHy equivalent CO2). The ce­ramic electrolyte is a good oxygen ion conductor at temperatures of approximately 1.300 low power (10 W) self regulating heater element.
o
F (700 °C) generated by an internal,
The heated electrolyte has an initial voltage (U0 approximately zero mV) in the absence of combustible gases (H2, CO or CH4 or
3H8, etc). In presence of these gases, the output voltage increases (see chart below) as these gases are oxidized at the hot sur-
C face of sensor
The cell voltage is measured by micro-controller based transmitter electronics and converted into a standard, linearized 4 – 20 mA signal for combustibles equivalent carbon monoxide (COe) in the range of approximate 0 – 1000 ppm.
Since the sensor is reacting to the presence of any combustible gas, but calibrated with CO+H be reported.
The expected lifetime of this sensor is about 5 years + under normal operating conditions, and is not dependent on fuel type, but:
CAUTION:
Find Quality Products Online at: sales@GlobalTestSupply.com
Exposure to corrosive gases (silicone vapor, alkaline and heavy metals, P, Pb, high SO2, etc.) will shorten the life of the sensor.
www.GlobalTestSupply.com
10
2, equivalent CO measurements will
Page 9
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
7 System components, general information
The OMS420 is used to continuously measure oxygen and combustible gas concentrations in flue and stacks of industrial boilers or furnaces, and those measurements are used to finely tune the combustion process.
7.1 Optimizing combustion
2 x 4 … 20 mA
Optimum combustion conditions are achieved by decreasing the amount of excess air in the stack gas to the point where combus­tibles start to increase. See combustion diagram below:
The absolute value of combustibles in the stack gas of a burner depends very much on the design and construction of the burner/boiler. Combustibles (CxHy) are lower in a well-designed system than they are one that is poorly designed.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
11
Page 10
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
It is important to monitor the rising level of combustibles, labeled
in the above diagram, and to trim the air/fuel ratio of the
opt
burner to compensate for changing ambient conditions (pressure and humidity) and maintain the point of maximum heat effi­ciency.
The diagram below illustrates the burner operating under two conditions:
Conventional operation with increased safety margin (higher O gas).
Improved operation with optimized combustion (lowest O
without a corresponding increase in combustibles).
2 concentration in the stack
2 concentration in the stack gas
The difference between the yellow area and the brown area represents the degree of combustion optimization, which in turn represents savings in fuel.
7.2 OMS420 features
The main features of OMS420 system are:
Compact, reliable and rugged industrial design.
Special reference air not required (uses ambient by natural diffusion).
True wet gas analysis and calculation of dry oxygen level if humidity of gas is known.
Fast response time.
Low energy consumption for both O
Micro-controller based electronics with backlit, graphic LCD display.
Linearized, galvanic isolated 4 - 20 mA signal outputs for both O
RS485 galvanic isolated digital data transfer (Modbus protocol RTU).
Field replaceable transmitter
Fast, safe and easy servicing by a single technician without removing the probe from the stack
2 and COe sensors.
2 and COe.
Dust tight and water proof enclosure IP65 (NEMA 4)
Easy operation and maintenance
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
12
Page 11
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
7.3
Probe models
There are three distinctive models of the OMS420 probe available:
- The compact model OMS420, which uses the flow guidance tube principle. The probe tube can be made of stainless steel for stack gas temperatures up to 1.200
o
F (1.000 °C)
1.800
o
F (650 °C) or made of AISI300 steel for stack gas temperatures up to
- The remote transmitter model OMS420RT, which is similar to above model but has the transmitter electronics separated from the probe by means of a 10 m (30 ft) special cable
- The high temperature model OMS420HT has no flow guidance tube but uses a ceramic tube and an air jet pump (ejector) to extract the sample from the tip of ceramic tube. It can be used for clean flue gas temperatures up to
3.100 °F (1.700 °C).
7.3.1 Compact model OMS420 (# 61417)
This model shall be used at site with low heat radiated from the process (ambient temperature to transmitter electron­ics is less then 140 °F / 60 °C).
Compact model components are:
probe with flow guidance tube and flange (ANSI150 or DN80 or DN65)
mounting flange assembly (supplied by user)
transmitter with electronics and sensors for O
2 and COe
back purge system (option) for high dust conditions Æ only for site with flying ash type of dust
pneumatic device (option) for automatic calibration
For higher temperature, less then 2.000
o
F (1.200 oC), but higher then 1.200 ° F (650 °C), the model OMS420RT with alloy AISI300
steel is recommended.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
13
Page 12
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
7.3.2 Remote transmitter model OMS420RT (# 63467RT)
This model shall be used when radiated heat from the process (duct, stack etc) will cause rising of ambient temperature higher then 130 °F (+55 °C)
Remote transmitter model components are:
probe with flow guidance tube and flange (ANSI150 or DN80 or DN65)
- SS316Ti material for temperature below 1.200 °F (650 °C)
- AISI300 material for temperature below 1.800 °F (1.000 °C)
mounting flange assembly (supplied by user)
detector head with junction box for sensors O
transmitter with electronics and switched power supply
special cable between junction box and transmitter
back purge system (option) for high dust conditions Æ only for site with flying ash type of dust
pneumatic device (option) for automatic calibration
2 and COe
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
14
Page 13
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
7.3.3 High temperature model OMS420HT (# 63467HT)
This model shall be used at sites with clean and high temperature flue gas, less then 3.100 °F (1.700 °C)
High temperature model components are:
probe with ceramic tube, ejector and flange (ANSI150 or DN80 or DN65)
mounting flange assembly (supplied by user)
detector head with junction box for sensors O
transmitter with electronics and switched power supply
special cable between junction box and transmitter
2 and COe
pneumatic device (option) for automatic calibration
back purge is not available for this model
The main differences between the remote transmitter model and the high temperature models are:
the use of ceramic tube of sampling probe on the high temperature model
the use of an ejector (air jet pump) to draw sample to the sensors..
When the ejector is purging instrument air with some 300 l/ h, a negative pressure on the back-side of a nozzle will be created. The negative pressure draws the sample gas from the tip of ceramic tube to the sensors.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
15
Page 14
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
7.4
The construction of the sampling probe is using the flow guidance principle
Principle of flow guidance tube
The tube is divided in half by a metal plate welded into the middle of the tube.
The probe is mounted on the stack by means of an 8-hole flange (ANSI 4”, 150 lbs or DN100). The tip of the divider plate is ori­ented facing towards the flow, which directs the stack gas into and through the tube at the same velocity as the flow in the stack.
The detector body is mounted on the flanged side of the tube through a hole in the flange cut for that purpose. In the body, behind a filter screen (the measuring side of detector), the two sensors for oxygen and combustibles are exposed to the stack gas flowing through the filter screen. In the detector behind the sensors, ambient air for reference gas diffuses through another filter screen (for dust protection) and flushes the backside of sensors (the reference side of the detector).
blowback in the stack
back-purge solenoid valve
Illustration for compact probe only
The flange feeds compressed air to several holes placed strategically around the detector and through a small tube, with air­releasing orifices spaced along its entire length, which is mounted along the metal plate in the center of the probe. The timing, duration and number of pulses of compressed air is controlled by user-settable electronic parameters and released by a solenoid valve. During purging, compressed air blows across the filter screen protecting the sensors, and from the holes in the tiny blow­back tube in the center of the probe, dislodging any accumulation of particulates so they will flow freely out of the probe and back into the stack.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
16
Page 15
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
7.5
The transmitter is housed in an aluminum enclosure (IP65, NEMA4X) that contains:
Transmitter electronics
Printed circuit board with u-processor (PCB “MAIN”)
Backlit, graphic display and dust proof keypad
Printed circuit board for the sensor connection (PCB “OMS”)
Electrical connector for power supply and data transfer
„OMS“ pcb (1)
MAIN“ pcb (2)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
17
Page 16
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
7.6
To perform calibration of the instrument, the user can use following schematic diagram (or similar).
It is required to use moistured calibration gas (see § 4.7) and two hand-ball valves to select between instrument air (zero gas) and calibration gas (span gas) supply to calibration inlet port of unit.
Manual calibration
Manual calibration
The procedure of calibration is described in chapter § 6.7.
7.7 Recommended calibration gas
For automatic calibration or manual calibration (see procedure in chapter 6.7.3) customer shall use following calibra­tion gases:
1) zero gas = instrument air (21 %O
2) span gas = 2…3 %O
2 ; 400…600 ppm CO ; 40…60 ppm H2 ; rest N2
2 , 0 ppm HC)
Note: It is highly recommended to use water bubblier to moisture the calibration gas!!!
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
18
Page 17
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
8 Software flow chart
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
19
Page 18
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9 Operation
9.1 Start-up
Prior to start-up, use the following check list to verify that all conditions are set for proper start-up:
Checklist transmitter
Transmitter removed from the probe? (Note: always power up with transmitter removed from the probe/stack!!)
Cast cover closed and screwed on?
Transmitter easily accessible and visible?
Ambient temperature around transmitter electronics in operating range of 0 °F to 140 °F (-20°C to 60 °C)?
Correct location of transmission cable (not in close proximity to high power supply cables or engines)?
Connection for power supply connected properly?
Signal connection connected properly?
Power supply (factory provided line power fuse) switched on?
Start up considerations:
Wiring
: Most problems are due to incorrect wiring. Please double check the wiring.
Shield should be grounded only at one side of the cable.
Leaks:
Check the calibration inlet port plug for correct fit.
Insulation:
Temperature:
Check that the mounting flange has been properly insulated to prevent gas condensation.
Check mounting flange temperature: min. 160 °F (70 °C) and max. 300 °F (150 °C).
Check ambient temperature of transmitter for max. 100 °F (+60 °C).
Note: If flange temperature at site with mounted probe and transmitter is below
(70 °C)
it is necessary to use a flange heater (ask MRU) to prevent condensation!!!
After power is switched on, the MRU logo and model of the unit will be displayed!
9.2 Warm-up
Warm-up time: minimum 30 minutes
160 °F
LCD will display a time count down for 30 minutes.
During warm-up, some inside measured values (heaters current and voltages) will be compared with credible thresholds and in case of “out-of-range”, an error message will be displayed. After countdown, if everything is OK, the message will change from “please wait” to the main measuring menu.
The “SKIP” function is used only for service purposes, to allow the operator to access other unit functions without waiting until the end of warm-up interval time. During this time, measurement values are not credible.
During warm-up, all other menus (info, service and setting) are accessible.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
20
Page 19
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
H
9.3
After warm-up, the unit will start automatically by displaying the main measurement menu.
O2 real time value with 0.01% resolution CO
Main measuring menu
real time value with ppm resolution
e
NOTE:
If between 30 min no button is pressed during the measurement, the level will be set on "0". I.e. the
Press “INFO” key for “info” menu (see chapter 9.4) Press “SERV” key for “service” menu (see chapter 9.5) Press “SET” key for “settings” menu (see chapter 9.6) Press “CALIB” key for “calibration” menu (see chapter 9.7)
9.4 Info menu
In the “INFO” menu the following parameters can be called up from a scroll up/down list:
menus SET and CALIB are only by renewed PIN input usable.
Next back-purge in hours/minutes
Next automatic set to zero in days/hours
If the display shows „----„ the automatic set to zero is not activated
Gas humidity display in % 1 - 25 or disabled
O
2 value wet: The O2 display is calculated to wet“ (*)
dry: The O The gas humidity can be set by “SET” to
4 - 20 mA 0
4 - 20 mA CO setting range of analog COe output
Firmware installed firmware version
Hardware installed hardware version
Serial number display of the serial number
O
2 sensor installed
CO sensor installed/not installed
(*) Formula for O
2 setting range of analog O2 output
2 calculation dry/wet
2 display is calculated to dry“(*).
%97.20max2 =O oxygen content in air
=wetO2 measured O2-value in wet stack gas
=dryO2 calculated (%) O2-value dry
Set gas humidity“
=
water content (%) in stack gas, (value entered manually),
100
therefore calculation of
Find Quality Products Online at: sales@GlobalTestSupply.com
22
www.GlobalTestSupply.com
wetOdryO
100
−⋅=H
⎞ ⎟
21
H
=
1100
⎜ ⎝
wetO
2
dryO
2
Page 20
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.5
Service menu
Displays actual 6 lines from a scroll up/down list with measured (A/D converter) components information.
Press the “UP” or “DOWN” key and return back to main measuring menu by pressing “BACK”.
standard value allowed deviation
O
2 cell - 10 mV (with operation temperature and 21 % O2) ± 5 mV
O2 heater 12 V (depending on flange temperature) ± 2 V
O2 heater 1,3 A (heater current) ± 0,3 A
CO cell 5 mV (after warm-up and with fresh air ) ± 10 mV
CO heater 7,5 V (depending on flange temperature) -3,0 V…. + 1 V
CO heater 0,4 A (heater current) ± 0,1 A
Unit temperature less then 160 °F (70 °C)
Typical mV-values for COe sensor: O
2 = 3 %, CO = 200 ppm Æ 98,0 mV 2 = 3 %, CO = 500 ppm Æ 190,0 mV
O
2 = 10 %, CO = 200 ppm Æ 32,0 mV
O
2 = 10 %, CO = 500 ppm Æ 93,0 mV
O
If the system detects an error, the plain text meaning will be displayed after pressing “ERROR” key
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
22
Page 21
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.6
This menu is protected by a user level 1 PIN code. Changing the pin code is described in following chapter.
9.6.1 PIN code level 1
Settings menu
Entering the factory default PIN code 1111 (user level 1), gains access to the SET main menu and the above settings.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
23
Page 22
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.6.2 Set analog output
PREV or NEXT move cursor to values which can be changed EDIT change value by means of UP/DOWN and POS OK save values BACK return to main menu
Factory defaults are: 20mA: O
20mA: CO 0 – 1.000 ppm Hold after purge typical 20 sec. (from 5 to max 200 seconds)
Notice: The last measured value on analog output 4-20mA after back-purge is hold even after back-purge cycle is finished.
9.6.3 Damping
This function allows user to set a damping (averaging the measurement over a period of time) of measurement, in the display as well in the analog output.
2 0 – 21.00 %
PREV or NEXT move cursor to values which can be changed EDIT change value by means of UP/DOWN and POS OK save values BACK return to main menu
Integration times from 0 to 30 seconds are settable, where 0 seconds means no damping.
9.6.4 Back-purge setup
This function allows setup of the optional back-purge solenoid valve to control:
interval time between back purges (hours and minutes)
pulse duration or interval time for energizing the solenoid valve (opening the valve)
number of pulses during one back purge cycle
PREV or NEXT move cursor to values which can be changed EDIT change value by means of UP/DOWN and POS OK save values BACK return to main menu
If hours and minutes are 0, the back purge function is deactivated. The delivery state of this function is off.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
24
Page 23
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.6.5 Interval auto-calibration
This function allows setting the interval of auto calibration
PREV or NEXT move cursor to values which can be changed EDIT change value by means of UP/DOWN and NEXT BACK return to main menu START Starts the auto cal immediately
Interval: 0 – 99 days .Note: Purge time: 1 – 10 minutes Auto-cal: AIR or AIR and Gas AIR: 1 point calibration with AIR (zero gas) AIR/GAS: 2 points calibration with combined AIR an GAS (zero gas and span gas) First calibration must always be the zero calibration with AIR”. The delivery state of this function is off.
9.6.6 Set calibration cylinder values
This function allows setting the values of span gas cylinders used for calibration.
If “interval” is 0, the auto cal. function is deactivated!
PREV or NEXT move cursor to values which can be changed EDIT change value by means of UP/DOWN and POS OK save values BACK return to main menu
The set values for “zero” are fixed values, programmed into the unit’s firmware.
The values for the calibration gas cylinder setting have to be set to the correct cylinder values for both O
9.6.7 Set temperature measuring unit
2 and CO.
EDIT change value by means of UP/DOWN OK save values BACK return to main menu
Set the temperature units to °C or °F
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
25
Page 24
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.6.8 Set gas humidity
This function allows setting the value of gas humidity (if known) and selection of either wet oxygen measurement or dry oxygen calculation.
NEXT move cursor to values which can be changed EDIT change value by means of UP/DOWN OK save values BACK return to main menu
Change value of gas humidity (XX.X %) if this is known.
Using the formula:
Default value for humidity = disabled
In the main menu the readings are w Æ wet oxygen measurement and d Æ dry oxygen calculation
Humidity H % = (1 - O2wet/O2dry) x 100 the dry oxygen is calculated and displayed.
9.6.9 Solenoid test
With the menu option "Solenoid test" allows to test each individual solenoid valve (if installed)
A manual switch ON or OFF of all 3 individual solenoid valves is possible.
By means of the key F1 the solenoid valve SV1 is switched ON or OFF, key F2 the solenoid valve SV2 is switched ON or OFF and with the key F3 the single solenoid valve SV3 is switched ON or OFF.
The display shows the current O2 and/or CO value(s).
Warning: All valves can be switched at the same time. It does not take place monitoring on a meaningful position of the single solenoid valves. This operation is only for authorized technical personal!
9.6.10 OMS420 settings
This function allows change of display contrast and set of transmitter address for multiple transmitter connection on the same RS485 data bus.
to get back to main settings menu
PREV or NEXT move cursor to values which can be changed EDIT change value by means of UP/DOWN and POS OK save values BACK return to main menu
LCD – Contrast 0-20 (10 for default)
Device-ID (Modbus) Slave address (RS 485-Modbus RTU)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
26
Page 25
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.6.11 Reset to (factory) defaults
YES System parameters will be reset to factory defaults NO Function will be aborted
Factory defaults
Auto-calibration OFF
Back-purge OFF
4 – 20 mA Range O2 20,96 %
4 – 20 mA range CO 1.000 ppm
O2 cylinder1 21,00 %
O2 cylinder2 2,00 %
CO cylinder1 0 ppm
CO cylinder2 1.000 ppm (mixture gas with additional 50 ppm H2 gas is expected!)
Pulse duration 2 sec.
Numb. of pulses 6
Hold after purge 20 sec.
Unit temp. °F
Display O
Humidity --.- % (without)
Interval (days) 0 (autocalibration = OFF)
Purge time (min.) 0
Auto-Cal. AIR (only 1 point calibration)
2 wet
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
27
Page 26
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7 Calibration
The transmitter should be powered up for at least 2 … 3 hours before calibration. A test (calibration) gas cylinder is connected to the calibration gas inlet port for calibration.
Adjustment can be done manually or automatically using the solenoid valves of the optional pneumatic unit PU420.
The test gas (clean fresh air/instrument air or calibration gas) flows through the hole inside the small flange (sensor mani­fold), purges the sensors and exits through the filter screen to the inside probe tube mounted on the stack.
On site calibration is done without removing the probe from the stack.
After calibration, tighten the calibration inlet port plug with a 14 mm wrench for 1/6 of a turn.
Perform adjustment:
1 Purge sensor with ambient air (40 l/h using the calibration gas inlet port). 2 When being purged with fresh air, the analog output signal should be 17.37 mA (for 0 – 25 % O 3 With calibration gas (2% O
in N2), the analog output should be 5.28mA (for 0 - 25% O2 measuring range).
2
4 The transmitter is now ready for operation.
Use calibration gas cylinder at least every 3 - 6 months.
Calibration with ambient clean air can be done automatically on a daily basis using the pneumatic option PU420.
measuring range).
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
28
Page 27
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7.1 OMS420HT model compact probe (# 61417)
Pos. Description
1 Electrical connector 2 Hose connection cal gas 3 Probe tube (stainless steel) 4 Flange 4”ANSI 150 or DN65PN6 5 Probe head (transmitter) 6 Calibration gas cylinder 7 Pressure regulator with flow volume control
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
29
Page 28
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7.2 OMS420RT model remote transmitter (# 63467RT)
Pos. Description
1 Hose connection back purge air 2 Hose connection cal gas and back purge air 3 Probe tube (stainless or Inconel steel) 4 Flange DN65 PN6 or 4” ANSI 150 5 Probe head 6 Calibration gas cylinder 7 Pressure regulator with flow volume control 8 Transmitter 9 Special connection cable (10 m or 30 m) 10 Compressed air or Nitrogen N2 (6 – 8 bar) supply
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
30
Page 29
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7.3 OMS420HT model high temperature (# 63467HT)
Pos. Description
1 Hose connection injector 2 Hose connection cal gas and back purge air 3 Probe tube (Inconel or ceramic) 4 Flange 4”AINSI 150 5 Probe head 6 Calibration gas cylinder 7 Pressure regulator with flow volume control 8 Transmitter 9 Special connection cable (10 m or 30 m) 10 Compressed air or Nitrogen N2 (6 – 8 bar) supply
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
31
Page 30
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
IMPORTANT:
Only authorized persons or manufacturer’s trained technicians are authorized to perform calibration
adjustments on the OMS420.
9.7.4 Pin Code level 2
Calibration with AIR (zero gas) or GAS (span gas) requires the use of level 2 pin code
see chapter 9.6.1
POS position UP / DOWN to change the numbers
NOTE:
Below displayed values e.g. O2 factor and offset values are examples!
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
32
Page 31
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7.5 O2 cell setup (by factory only)
If there is no calibration gas available and the cell must be replaced, this function permits changing a cell without calibration. This can only be done if the cell factors have been measured at MRU and delivered together with the spare sensor.
PREV or NEXT move cursor EDIT change value by means of UP/DOWN and POS OK save values BACK return to calibration menu
1) O2 cell setup requires input (edit function) of O2 factor and O2 offset values delivered with the spare sensor
2) Temperature compensation of 20.9 % O flange temperature change. The difference of air oxygen measurement at lower temperature to measurement at higher tempera­ture will be edit as “Temp.offs” value.
NOTE:
This procedure should be followed by new calibration using calibration gas cylinder.
Only authorized persons or manufacturer’s trained technicians are authorized to perform adjustments on the OMS420.
9.7.6 CO cell setup (by factory only)
If there is no calibration gas available and the cell must be replaced, this function permits changing a cell without calibration. This can only be done if the cell factors have been measured at MRU and delivered together with the spare sensor.
PREV or NEXT move cursor EDIT change value by means of UP/DOWN and POS OK save values BACK return to calibration menu
CO cell setup factor and CO offset setup factor will be edit using the values delivered together with the spare sensor.
is required when oxygen value at air is changed too much because of gas or
2
Set point of heat resistor [Ohm] must be also carried out according to § 6.7.12 NOTE:
This procedure should be followed by new calibration using calibration gas cylinder.
Only authorized persons or manufacturers trained staff are allowed to perform adjustments on the OMS420!
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
33
Page 32
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7.7 Adjustment O2 cell amplifier (by factory only)
PREV or NEXT move cursor EDIT change value by means of UP/DOWN and POS OK save values BACK return to calibration menu
This procedure is required only when a new O
Source a precision voltage device to the input connectors of the O
cell is installed.
2
cell signal (see picture below) and select
2
-10mV for offset trim and +100mV for span trim.
If actual values displayed are other than -10mV to +100mV setting can be adjusted by selecting “Set offset” or “Set span” and editing and replacing the displayed numbers with the actual numbers.
Factory default = factory adjusted values
ATTENTION:
Changes of these factors require new gas calibration. Only authorized persons or manufacturer’s trained technicians are authorized to perform adjustments on the OMS420.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
34
Page 33
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7.8 Adjustment 4 mA – 20 mA (factory only)
Connect precision 100 Ohm resistor (0.1% tolerance) to both analog outputs of O
Measure the voltage across the resistors using a precision digital multi-meter.
and CO measurements.
2
PREV or NEXT move cursor EDIT change value by means of UP/DOWN, until you read 400mV, up to2000mV on the multi-meter OK save values BACK return to calibration menu
Note:
The 4 – 20mA analog output factory adjustment can deteriorate over time, so measured values must be verified occasionally.
Only authorized persons or manufacturers trained staff are allowed to perform adjustments on the OMS420.
COe measurement
analog output
O2 measurement
analog output
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
35
Page 34
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
g
y
9.7.9 Calibration AIR
This menu performs the 1 point calibration with AIR (zero gas).
Wait approximately 5 minutes while instrument air is being sup­plied to sensors to allow signals to stabilize.
Press “CALIB” key
Wait for approximately 3 seconds.
Zero gas (AIR) calibration was successful.
If calibration fails, an error messa
Return to setting menu.
e will be displa
O2 OK. CO OK or O2 failed! CO failed!
If calibration fails, an error message will be displayed.
9.7.10 Calibration AIR & GAS
This menu performs the 2 points calibration with combined AIR and GAS (zero gas and span gas).
For the calibration, the use of moistured mixture gas (O Moisturing the span gas requires the use of water bubblier!
2, CO and H2 ) is mandatory.
ed.
With single component gas CO the calibration is not possible.
If no mixture gas is available, then see chapter 9.7.11.
Calibration of O2 only: set CO cylinder (value) = 0 Æ no calibration Calibration of CO only: set O2 cylinder (value) = 0 Æ no calibration
First calibration must always be the zero calibration with “AIR”
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
36
Page 35
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
g
g
Wait approximately 5 minutes while instrument air is being supplied to sensors to allow signals to stabilize.
Press “CALIB” key
Second calibration shall be the span calibration with “GAS”
Wait for approximately 3 seconds!.
Zero gas (AIR) calibration was successful.
If calibration fails, an error messa
e will be displayed.
Read calibration gas cylinder values and if any change is required (other cylinder values) then edit O display by pressing “O
” then the “CO” key.
2
and CO in the
2
Wait approximately 5 minutes while calibration gas is being supplied to sensors to allow signals to stabilize, then press “CALIB” key.
Wait for approximately 3 seconds.
Span gas (GAS) calibration was successful.
If calibration fails, an error messa
e will be displayed.
Return to previous setting menu.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
37
Page 36
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7.11 CO correction factor
This function allows user to perform site calibration of combustibles measurement with a comparative selective CO measurement, using portable CO analyzer (e.g. Delta65). This will make the equivalency to CO and is recommended to be carried out at sta­ble, constant combustion process.
Because sensor is measuring all combustibles in the flue gas, a difference to selective CO measurement will occur.
By using the correction factor, this difference will be minimized, the instrument will display same value as the comparing CO selec­tive measuring instrument.
1) The measured value by means of Delta65 can be set (“desired value”) by pressing “EDIT” key.
2) Using the“UP” and “DOWN” keys will change this value until it corresponds to Delta65 measurement of CO.
3) Then press “OK” key to correct the combustibles measurement to selective CO measured value.
NOTE: by pressing the key “FACT.” the CO correction factor will be set to 1000 (amplification is 1)!!!!
In that case, adjustment of COe reading according to measurement by means of Delta65 is out of order!
Use a portable analyzer to measure CO in the flue gas and to adjust accordingly the “desired value”
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
38
.
Page 37
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7.12 O2 heat resistor regulation (factory only)
PREV or NEXT move cursor EDIT change value by means of UP/DOWN and POS OK save values BACK return to calibration menu
First line: regulation point actual (O
Second line: new set point (Heat resistor [Ohm]}
ATTENTION:
Changes of these parameters require new calibration. Only authorized persons or manufacturer’s trained technicians are authorized to perform adjustments on the OMS420
9.7.13 CO heat resistor regulation (factory only)
)
2
PREV or NEXT move cursor EDIT change value by means of UP/DOWN and POS OK save values BACK return to calibration menu
First line: regulation point actual (CO)
Second line: new set point (Heat resistor [Ohm]}
Heat resistor [Ohm] CO: R 25 °C * (2.35 up to 2.6)
R higher Æ higher temperature Æ lower offset signal, faster response time, low CO response signal R lower Æ lower temperature Æ larger CO response signal, and lower response time
ATTENTION:
Changes of these parameters require new calibration. Only authorized persons or manufacturer’s trained technicians are authorized to perform adjustments on the OMS420
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
39
Page 38
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
9.7.14 History auto-calibration
This function allows the user to see the last auto calibration values.
10 Troubleshooting
If the system detects an error, the plain text will be displayed after pressing the “Error” key in the service menu.
Malfunction: Corrective action:
No display
Display:
Heating current<0.5A
Sensor element defect
Display:
Configuration not ok
Checksum wrong
Display:
Flash error!
Display:
Unit to warm
Hot environment
Transmitter does not react with sample gas from stack
Calibration error Calibration factor cannot be set.
Check power supply, check cables and connectors
O
sensor element defect
2
Check heater resistance of sensor
Replace sensor
New adjustment required
New adjustment required Please contact your dealer (manufacturer).
Ambient temperature too high!
Use heat shield or attach a compressed air based Vortec cooler
Clean the filter screen.
Test gas cylinder is empty, filter screen dirty or clogged
Calibration offset value cannot be set.
Instrument air supply is not available,
Transmitter does not react with test gas Please contact your dealer (manufacturer).
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
40
Page 39
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
11 Technical specifications
Measurements Oxygen and combustibles (equivalent CO
Measurement principle Heated zirconium oxide cell for O
,
2
Heated solid electrolyte cell for CO
) contained in flue gas
e
e
Lifetime of cells more than 5 years under normal conditions (*)
Warm-up time minimum 30 minutes
Measurement range 0 – 25.0 % for O
0 – 1,000 ppm for CO
Resolution 0.01 % for O
Repeatability within ± 1 % of full scale for O
Linearity better than ± 1 % of full scale for O
Accuracy O
2: ± 0.2 % O
2
e
and 1 ppm for COe.
2
2
2
or ± 5 % of reading (whichever is larger)
2
COe: not selective measurement, accuracy is not an issue
Response time < 10 seconds
Electronics micro-controlled based, on board graphic backlit LCD
Output signals 2 x 4 – 20 mA, galvanic isolated, max. 500 R (for direct transfer to process PLC)
RS 485, galvanic isolated digital data transfer (Modbus protocol RTU)
Power supply 18 Vdc to max. 24 Vdc, 100 W for compact probe model OMS420
100…240Vac / 50-60Hz, 100W for all other models
Ambient temperature - 4
o
F … 140o F ( -20°C …+ 65°C)
Ambient humidity 5 – 95 %, non condensing
Protection IP 65 (NEMA 4)
(*) in the absence of heavy metals, silicones, silicates, aggressive and/or corrosive gases
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
41
Page 40
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
12 Appendix
12.1 Error codes
02 Configuration Error, new adjustment required
04 Flash Error, new adjustment required Please contact your dealer (manufacturer).
08 Unit to warm, ambient temperature too high!
01 Heating current<0.5A, O2 sensor not o.k.
32 Heating current<0.1A, CO sensor not o.k.
64 Heating current<0.1A, CO sensor not o.k.
128 Calibration error, calibration factor cannot be set.
Test gas cylinder is empty, filter dirty or clogged.
Calibration offset value cannot be set.
Instrument air supply is not available.
At the same time if several errors line up, then the error codes are to be added.
Her one example:
33 Heating current<0.5A, O2 sensor not o.k. and
Heating current<0.1A, CO sensor not o.k.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
42
Page 41
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
12.2
Modbus Slave specification
General information
• supports only the binary Modbus protocol (RTU)
• the slave modbus address is user definable from 1 to 49
• communication parameter are user definable as follows:
- 9600 baud
- 8-Bit
- even parity and 1 stop bit
• data types (used in table below): U 16 bit unsigned integer value (0…65535) I 16 bit signed integer value (-32768...32767) UL 32 bit unsigned integer value (0…4.294.967.295) L 32 bit signed integer value (-2.147.483.648…2.147.483.647) F 3 32 bit floating point value (reads -1E38, when not available)
Defined registers Available data with modbus command 4 READ INPUT REGISTERS:
PLC address
Protocol address
Data type
Number of registers
Register content
Device info / status
40001 0 U 1 Error-Flags
40002 1 U 1 Reserved for further flags
40003 2 U 1 Status
40004 3 I 1 O2 [%] 1152 -> 11,52 %
40005 4 I 1 CO [ppm] 123 -> 123 ppm
Error Flags Bit 0 Heating current<0.5A, O2 sensor not o.k. Bit 1 Config error Bit 2 Flash error Bit 3 Unit to warm Bit 4 Bit 5 Heating current<0.1A, CO sensor not o.k. Bit 6 Bit 7 Calibration error
Status Status 0 Busy Status 2 Measurement active Status 5 Backpurge active Status 6 Calibration active
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
43
Page 42
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
12.3
Declaration of confirmity
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
44
Page 43
OPERATING MANUAL OXYGEN MONITORING SYSTEM OMS 420
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
45
Loading...