Mru AIR SWG 100 User Manual

SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
1 / 108
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
2 / 108
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
3 / 108
1. General Information .......................................................................................................... 6
1.1. Weather and environmental conditions ...................................................................... 6
1.2. Installation instructions ............................................................................................... 6
1.3. General important instructions for the plant operator ................................................. 6
2.1. Important information about the user’s operation manual .......................................... 7
11.1. General warning ......................................................................................................... 7
11.2. Safety Information ...................................................................................................... 7
11.3. Packing ...................................................................................................................... 9
11.4. Return of hazardous waste ........................................................................................ 9
11.5. Return of analyzer ...................................................................................................... 9
11.6. MRU guarantee conditions....................................................................................... 10
12. Analyzer Description ...................................................................................................... 11
12.1. Intended use ............................................................................................................ 11
12.2. Principle of operation ............................................................................................... 12
12.3. Marking and range of application ............................................................................. 13
12.4. Possible applications and analyzer options ............................................................. 13
12.5. Features ................................................................................................................... 13
13. Overview of the SWG-100BIO Ex ................................................................................... 14
13.1. Scope of supply and delivery ................................................................................... 14
13.2. Material required for the installation (not scope of supply) ....................................... 14
13.3. Dimensions and views of the analyser ..................................................................... 15
13.4. Position of the gas in and outlets ............................................................................. 16
13.5. Operating unit: Display and keypad ......................................................................... 17
14.1. Measurement menu ................................................................................................. 19
15. Explosion protection systems of SWG100BIO-EX....................................................... 20
15.1. Electrical components .............................................................................................. 20
15.2. IP 65 cabinet ............................................................................................................ 20
15.3. CH
measurement of inside cabinet ......................................................................... 21
4
15.4. Remote Power Supply Box ...................................................................................... 24
16. Sample Gas measurement ............................................................................................. 27
16.1. Cycle configuration .................................................................................................. 27
16.2. H2Slow sensor protection (optional) ........................................................................ 32
17. Analyzer mounting and installation .............................................................................. 34
17.1. Content of your order ............................................................................................... 34
17.2. Installation rules to guarantee explosion protection ................................................. 34
17.3. Prerequisites for installation ..................................................................................... 34
17.4. Steps for the mechanical mounting and installation ................................................. 35
17.5. Mounting the analyser on a wall or steel rack .......................................................... 36
17.6. Correct tubing of the SWG100BIO-Ex ..................................................................... 37
18.1. Flow restrictor orifice (#65114) ................................................................................. 38
18.2. Installation of the power supply ................................................................................ 40
18.3. Installation of the external switch-off circuit .............................................................. 41
18.4. Connection of the I/O module .................................................................................. 43
19. Power up and commissioning the analyser ................................................................. 45
19.1. Power up the analyser ............................................................................................. 46
19.2. Check country and language ................................................................................... 47
19.3. Check date and time of the instrument .................................................................... 47
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
4 / 108
19.4. Configure sampling points ....................................................................................... 47
19.5. Configuration of the alarm relays ............................................................................. 48
19.6. Configuration of the Modbus .................................................................................... 50
19.7. Configuration of the external control (Option: IO module) ........................................ 52
19.8. Configuration of the analog outputs at the I/O module ............................................. 56
19.9. Configure alarm output of I/O module ...................................................................... 57
19.10. Configure the AUX-input on software-side: ......................................................... 59
20. Operating the analyser ................................................................................................... 60
20.1. Administrator PIN code ............................................................................................ 60
20.2. Power-On of analyser .............................................................................................. 60
20.3. Self-Test .................................................................................................................. 60
20.4. Main menu measurement ........................................................................................ 61
20.5. Data Storage ............................................................................................................ 63
20.6. Power-Down of analyzer .......................................................................................... 64
20.7. Backup/restore all individual user -settings ............................................................... 65
20.8. Default configuration ................................................................................................ 65
20.9. Update the firmware ................................................................................................. 67
21. Service and maintenance ............................................................................................... 70
21.1. Preparing and information about the maintenance .................................................. 70
21.2. Regular maintenance works by the operator ........................................................... 70
21.3. Position and overview of the service-parts............................................................... 71
21.4. Cleaning the surface of SWG100BIO-Ex. ................................................................ 77
21.5. Maintenance of the switch-off relay (K2) and the solenoid valve for CH4 supervision 77
22. Exchange of electrochemical sensors.......................................................................... 79
22.1. Analyzer calibration .................................................................................................. 86
23.1. Troubleshooting ....................................................................................................... 90
24. Technical specification .................................................................................................. 91
25. Declaration of conformity .............................................................................................. 95
26. APPENDIX ....................................................................................................................... 99
27. General information ........................................................................................................ 99
27.1. Special informationen about the Modbus Slave function ......................................... 99
27.2. Special informationen about the Profibus Slave function ....................................... 100
28. Defined registers to be read by the master ................................................................ 101
28.1. Analyser Status (address 0 and some mirror addresses) ...................................... 104
28.2. Analyser System Alarm (address 2 and some mirror addresses) .......................... 105
28.3. Spare parts list ....................................................................................................... 106
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
5 / 108
Inspect Shipment for Damage
Carefully inspect the entire shipment for damage in the presence of the shipper’s agent, removing packaging material if necessary. Note any damage to packaging and/or goods on Packing List and have it signed by the shipper’s agent prior to accepting the shipment. Submit damage claim to MRU immediately.
NOTE: Damage claims not received by MRU within 3 days of receipt of shipment will not be accepted.
The products described in this manual are subject to continuous development and improvement and it is therefore acknowledged that this manual may contain errors or omissions. MRU encourages customer feedback and welcomes any comments or suggestions relating to the product or documentation.
Please forward all comments or suggestions to the Customer Feedback Department at the following address:
MRU GmbH Fuchshalde 8 + 12 74172 Neckarsulm / Obereisesheim GERMANY
Fon (+49) 71 32 99 62 0 (Reception) Fon (+49) 71 32 99 62 61 (Service) Fax (+49) 71 32 99 62 20 Email: info@mru.de Homepage: www.mru.eu
This manual is intended solely as a guide to the use of the product. MRU shall not be liable for any loss or damage whatsoever arising from content errors or
misinterpretation of information’s from this manual or any mis-use resulting from the use of this manual.
FOR MORE INFORMATIONS ABOUT COMPANY MRU PLEASE VISIT OUR WEBSITE
www.mru.eu
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
6 / 108
WARNING
Please check correct installation before powering up the analyzer!
2.
ATTENTION
In case of not intended use the guarantee will void. Regular controls,

1. General Information

1.1. Weather and environmental conditions

The SWG100BIOGAS analyser is designed for operating ambient temperatures of +5°C to +45°C (without cabinet heater) resp. -20°C to +45°C (with cabinet heater). The SWG100BIO-Ex. is for designed for use in hazardous Zone 2, see the conformity marking in the corresponding section.
In case of outdoor mounting it is important that the analyser is protected against rain and sun. For extreme environmental conditions like high humidity, salty sea air, etc. further protective measures are necessary. These should be clarified with the manufacturer (MRU).
Any additional protective measures for outdoor mounting have to be provided by the plant operator. The manufacturer (MRU) is supporting the plant operator in choosing appropriate protective measures.

1.2. Installation instructions

Installation instructions, which are described in chapter 17 of the operation manual, have to be strictly carried out.
Please operate the analyzer only in upright position!

1.3. General important instructions for the pl ant operator

To guarantee continuous operation of the SWG100 BIOGAS analyser, the functions, processes and operation of the analyser have to be monitored regularly by the plant operator – especially in case of any initial installation. Thus, it will be possible to take suitable measures to improve the availability and life time of the analyser.
As the plant operator gains more experience concerning the maintenance requirements of the analyser, the monitoring frequency may be reduced to more extended periods of time.
It is important that the filter-units, listed in chapter 10 of the operation manual, are checked regularly and, if necessary, service parts are changed.
inspections and the exchange from polluted and exhausted filters by the operator are also an important part of the determination “not determined use”. See chapter 9 regular maintenance work.
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
7 / 108
DANGER
Identifies an immediate, impending hazard that, if ignored, will
WARNING
Identifies an immediate, impending hazard that, if ignored, may
CAUTION
Identifies a possibly dangerous situation that, if ignored, may result
9.
ATTENTION
Identifies a possibly harmful situation that, if ignored, may result in
11.
NOTE
Identifies user tips and other especially important information.
DANGER
Any maintenance action must be done only, when the atmosphere is

2.1. Important information about the user ’s operation manual

The users/operation manual is an important part of this delivery. It will explain how to use this analyzer properly and sets forth safety and environmentally friendly procedures. It is the responsibility of all users to read and familiarize themselves with this manual, paying particular attention to the safety instructions. The most important safety details are clearly marked with an attention sign.
Safety declarations
In this document the safety hints will be declared as follow:
result in severe bodily injuries or death.
result in severe bodily injuries, material damage or death.
in minor injuries.
damages to the device or its surroundings.

11.1. General warning

verified to be not explosive.

11.2. Safety Information

The following safety procedures must be followed at all times. They are significant and essential part of this manual. Failure to follow safety procedures can result in the loss of your warranty claims.
Biogas or other similar gases (landfill gas, bio-methane, coal seam gas etc) is containing flammable gas component CH4 and toxic gas component as well (H2S and CO2). Analyzers is continuously sampling a certain volume (approx 50l/h) of the sample gas, and is venting it to ambient air. For this reason, there are two aspects which must be considered:
1) toxicity danger of sample gas
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
8 / 108
2) flammable (explosion) danger of sample gas
1) Inhaling toxic gases is harmful to health and can even cause death in some cases.
It is the responsibility of analyzer user to ensure that person is skilled and trained in safety aspects of gases being analyzed and procedures to follow while using this instrument.
Local regulations for possible exposition to toxic gases have to be known and obeyed by the user of the analyzer
Using a personal gas detector inside the biogas plant is highly recommended since H2S in higher (very dangerous) concentration cannot be detected by human nose. Only small concentrations around few ppm can be detected by human nose
CO2 gas is heavier than air and therefore operator shall avoid working at underground or confined spaces. Beside of that CO2 is also odorless!
It is not allowed using the biogas analyzer in confined space or rooms
without forced ventilation.
Sample gas exiting the analyzer will flow in the ambient air and only outdoor use or forced ventilation rooms are suitable for using the biogas analyzer
2) Regarding flammable gases (e.g. CH4 methane) and hazardous area of operating the
instruments, user must also be able to recognize the area classification and be aware of using the instrument there. This area classification is country specific, please observe and notice it.
Stationary analyzers are allowed to be mounted in hazardous area zone 2 only if they have the certificate of compliance. These instruments shall never be located in confined places or rooms without forced ventilation.
Only trained personnel should carry out installation of stationary instrument and/or maintenance, service and repair. Opening the stationary analyzer cabinet can expose personnel to injuries and shocks from mains voltage!
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
9 / 108
CAUTION
Moisture or condensate, being pumped out of the condensate outlet
Please carefully clean all parts that come into contact with the
Safety regulations
The analyzer may only be used as indicated in this manual. Our analyzers are checked according to the following regulations: VDE 0411 (EN61010) and DIN VDE 0701 before they leave the MRU GmbH factory. MRU technical products are designed and manufactured according to DIN 31000/ VDE 1000 and UVV = VBG 4 of the professional guilds for fine mechanics and electrical engineering. MRU GmbH assures that the analyzer complies to the essential requirements of the legal regulations of the member states of the electro-magnetic compatibility (89/336/EWG) and to the low-voltage regulations (3/23/EWG).
Specific safety regulations
No part of the analyzer, or any other metal parts & accessories shall be used as electric conductors. The analyzer shall not be used in or under water. The analyzer shall not be placed near or directly exposed to open fire or heat. The analyzer shall avoid dropping.
port can be slightly acidic.
In case of skin contact IMMEDIATELY: clean affected parts of the body.
Avoid getting liquid in eyes.
condensate.

11.3. Packing

Packing regulation of 12.07.1991 If your local waste facility does not accept MRU packing materials for disposal, you may return it to MRU or our local sales representative. Packing materials returned to MRU must be returned prepaid.

11.4. Return of hazardous waste

Waste Disposal/Returns/Warranty ­MRU GmbH is required to accept the return of hazardous waste such as electro-chemical sensors that cannot be disposed of locally. Hazardous waste must be returned to MRU prepaid.

11.5. Return of analyzer

MRU GmbH is required to accept the return, for proper disposal, of all analyzers delivered after 13th of August 2005. Analyzers must be returned to MRU prepaid.
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
10 / 108

11.6. MRU guarantee conditions

MRU granted of the analyzer SWG100 BIOGAS a guarantee of 12 month.
1. 6 month on MRU spare parts.
2. The term of the guarantee conditions starts as of the invoice date.
3. Excluded from the guarantee conditions are damages, which occurred by:
Improper use.
Improper application.
Improper mounting.
Deliberate or negligent destructions.
External influence for example fall, hits, solvents, acids, gases, by normal wear or
transport damages. This applies also for defects, which are caused from high pollution and/or moisture (condensate) in the gas route or on the sensors.
4. As well excluded from the guarantee conditions are typical consumable- and spare parts.
5. The guarantee condition denatured immediately, if no original consumable parts are installed.
The guarantee is only for original MRU consumable parts and sensors.
6. With the replacement of the type plate or the serial- number of the devices all guarantee
conditions will be invalid.
7. The service of a guarantee conditions will not enlarge the guarantee time. Demands because
of consequential damages are excluded.
8. The transport costs for round trip takes the consignor or the warrantee.
9. MRU reserves the right, to determine individual conditions or exceptions. These will be
separated informed.
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
11 / 108

12. Analyzer Description

12.1. Intended use

The instrument is intended for analyzing the composition of biogases / biomethane / landfill gases and determine the concentration of several components like CH4, CO2, O2 and H2S. The instrument may optionally be equipped to monitor several sites in time sharing technique (cyclical one by one sampling). In particular, the instrument is not foreseen to serve as a gas detector or as a safety device.
In case of not intended use the guarantee will void. Regular controls, inspections and the exchange from polluted and exhausted filters by the operator are also an important part of the determinations “not determined use”- see chapter 9 regular maintena nce work.
The SWG-100BIO Ex. is an extractive measuring system. It has maximal 4 sample gas inlets, to measure the gas concentration from 4 different sample points, one by one. It has an internal gas pump to suck the sample gas from the sampling points. After the measurement, the sample gas is vented through the sample gas outlet (Vent). All sample points are equipped with a flow restrictor orifice, which allows also sampling the gas from the higher-pressure gas pipe of the engine, while limiting the maximum amount of gas in case of internal cabinet piping leakage.
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
12 / 108

12.2. Principle of operation

Sample gas from one or more sampling points is fed into the analyzer by dedicated ports. Internally mounted electric valves select one point at a time to feed sample gas to the analysis unit.
The instrument is equipped with a non-dispersive infrared (NDIR) bench for analysis of CO2 and CH4. Two separate infrared detectors for each CO2 and CH4 are included, each operating with a different optical path length and stabilized by referring to a reference detector. The IR source is a highly efficient and stable IR emitter, pulsed at a frequency of several Hertz. By design NDIR technique offers good stability and selectivity together with long life time of sensor (only limited by corrosion or dust, which can be prevented or removed by regular servicing the instrument).
The instrument is optionally equipped with a number of electro-chemical sensors ECS in order to detect gas components like oxygen O2 or H2S. Those sensors offer a reliable and effective way to detect the target gases. They are typically of limited life time (several years) but may be easily replaced once the end-of-life is reached.
In regular time intervals (user settable), the instrument automatically switches to purge the sensors with fresh (ambient) air for re-adjust the zero point.
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
13 / 108

12.3. Marking and range of application

The SWG100BIO-EX was designed for use in hazardous area with potentially explosive atmosphere and was certificated by ATEX directive:
EX-Protection: [EX] II 3G Ex nA nC IIC T3 Gc Certification no.: EPS 16 ATEX 1xxx X
The SWG100BIO-EX is operating in a temperature range of -20°C … +45°C with intern cabinet heater.
The specifications of the analyzer are:
Main power supply: 230 VAC / 50 Hz Power consumption: 36 VA Protection class: IP65 Fuses: 230 VAC/ 50Hz / 6 A Data output interface: 4…20 mA / analog outputs
RS485 Modbus RTU

12.4. Possible applications and analyzer options

The biogas analyzer SWG100 is the industrial measuring solution to be used at all hazardous zone 2 areas of::
biogas (anaerobic digestion) plants
cogeneration heat and power engines (CHP)
municipal or industrial waste water treatment sites
coal seam gas sites (coal bed methane)
food and animal waste processing plants
biomethane (gas to grid) plants
landfill sites

12.5. Features

This analyzer can be equipped with additional options and/or additional accessories to full fill multiple other measuring tasks such as
H2S purge pump
H2S
H2 measurement with EC cell
CO measurement with EC cell
Multiple sampling point switchover from 2x up to maximum 4x sites
Module with 4 channel analog outputs/inputs 4-20mA, with 2x “fail safe” alarm
Converter module of RS485 into Profibus
Cabinet heater for freeze protection Ask our sales representatives for available options and accessories or check out our MRU
website.
low measurement with protected EC cell using cutoff solenoid valve and air high measurement with EC cell
relays
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
14 / 108

13. Overview of the SWG-100BIO Ex

13.1. Scope of supply and delivery

The delivery contains the follow components:
SWG-100BIO Ex analyser, with a stainless steel case.
A stainless-steel thread filter for the zero gas inlet, with a G1/8 inch male thread.
Flow restrictor orifices for the sample gas inlets, with a G1/8 inch female th read. The
number of flow restrictors is depended from the number of gas sample inlets.
A fitting for the gas outlet (vent), with a G1/4 inch male thread.
Mounting material.
Optional: Flame arresters for the sample gas inlets, with a G1/4 inch male thread.

13.2. Material required for the installation (not scope of supply)

Material for tubing or hosing the sample gas inlets and outlet.
For the sample gas inlets: Flexible PTFE tube or stainless steel tubing with a diameter of 4/6 mm. It could be necessary to install a heated sample line on the tubes, to protect the sample gas from freezing.
For the sample gas outlet (Vent): Flexible PTFE tube with a diameter of 8/10 mm. It could be necessary to install a heated sample line on the flexible tube, to protect the vent gas pipe for freezing.
For the condensate outlet: Flexible tube with a diameter of 4/6 mm. It could be necessary to heat the condensate tube to prevent that the condensate for freezing.
Power supply lines and signal lines
Cable for power supply: Cable type: H 07 RN-F 5 G 0,75 mm²
Signal lines for 4…20 mA analog outputs.
Signal lines for Modbus RTU digital transfer.
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
15 / 108
642 mm
630 mm
223 mm
1 2 1 2 3 4 1
2 3 4
Reference:
Reference:
Reference:

13.3. Dimensions and views of the analyser

Top view:
1. Sample gas inlets G1/8 inch female thread.
2. Sample gas outlet G1/4 inch female thread
Front view / left view / right view
1. Heat sink of gas cooler unit
2. Calibration gas inlet G1/8 inch female thread
3. Zero gas inlet G1/8 inch female thread
4. Condensate outlet G1/8 inch female thread
Bottom view
1. Cable gland M16 for power supply
2. PE bolt for earthing cable
3. Cable gland M16 for I/O modules
4. Breather drain
MRU GmbH, D-74172 Neckarsulm
16 / 108

13.4. Position of the gas in and outlets

1
Display (color, with backlight)
2
Keypad
3
Sample biogas inlet ports (1 to max 4 ), female G1/8” threads
4
Cabinet lock
5
Inlet cable glands M16 for I/O modules
6
Inlet cable gland M16 for mains power supply
7
Heat sink for gas cooler unit
8
Condensate outlet port, female G1/8” thread
9
Zero gas inlet port, female G1/8” thread
10
Calibration gas inlet port, female G1/8” thread
11
Vent outlet port, female G1/4” thread
1
2
3
4
4 5 6
7
11
8 9 10
SWG100 BIOGAS User Manual
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
17 / 108
14.
NOTE
Operation of the keypad is intrinsically safe.
DANGER
Explosion hazard
off itself, the
1 2 3 4 5
6 7 8
9
Control unit

13.5. Operating unit: Display and keypad

All commissioning can be carried out with the control-unit. The operating-unit can only be operated when the cabinet door is open.
Keypad for operation of control unit.
In case of system alarm, after analyser has auto-shut­cabinet might contain an explosive atmosphere. Any ignition source must be avoided when opening the door.
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
18 / 108
Number
Symbol
Description
1
ESC : abort or return to the menu above
2
Prepare Power-Down: Press this key before you
sensors
3
, , ,
Arrow keys: context dependent functions, e.g. scroll in
4
OK : confirmation key, select a marked menu point.
5
Screen shot : press this key in order to store a screen
6
Menu key: Will show all available functions in the
three function keys.
7-9
Function Keys : Activates the functions seen on the
display (2 function key bar)
Number
Function
1
Menu bar
2
Function key bar
3
Display panel
- Measurement value,
4
SD-Card in the slot
protected)
1 2 3
4
All information required to operate the analyser is displayed as shown below.
disconnect mains. The analyser will store changed user settings and other operational data and will purge the
between lines, change values, change view.
All information required to operate the analyser is displayed as shown below.
shot of the current display contents onto the SD card.
window that is currently in use – also those which have an individual key on the key pad like the printer and the
- Menu
- Indication green read­and write access
- Indication yellow only read access (SD-Card write
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
19 / 108
Please read chapter 5 for
Please read chapter 5 for
Please read chapter 5 & 6
SWG 100 BIOGAS

14.1. Measurement menu

The measurement menu is the main root of the menus. Here you can see the current measurement values, the current sampling point number and the status of the zeroing process while it's active. Via pressing the function keys F1-F2-F3 you can reach other menus:
- F1: Menu Sample Point Switching
- F2: Menu Storage
- F3: Menu Extras
Menu Memory all tasks for the management of the data memory available. Menu Extras all the other available tasks – for management and customizing your
analyzer.
details.
details.
for details.
No matter which menu is currently active you will return to the measurement menu by pressing the ESC key several times.
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
20 / 108
WARNING
The following chapter explains with which protection systems
Ex. is equipped. If one of the explained
The entire cabinet must be checked for absence of
Exchange disconnected flexible tubes. Never use

15. Explosion protection systems of SWG100BIO-EX

The explosion protection of SWG100 Bio-EX is based on the nA and nC type of protection as indicated in the ATEX declaration. Following aspects have to be followed.
Any action not in line with the guidelines below may violate the explosion protection of the SWG100 BIO-Ex and may results in danger and explosion hazard
the SWG100BIO­protection systems is triggered, the user must do the following actions to prevent an ignition inside the analyser cabinet.
Only staffs with a “hot permit” are allowed to restart the
analyser in service.
The staff must have a certified flammable gas detector to
monitor the ambient atmosphere.
The cabinet door shall not be opened by force in order to
prevent ignition sparks.
flammable gas with a certified combustible gas detector.
It is not allowed to restart the analyser in service, until the
cause of system alarm is remedied.
exhausted tubes again.

15.1. Electrical components

All components in the SWG100 BIO-EX are selected and qualified to meet the requirements of the given type of protection (nA or nC). Make sure that any change to the instrument during it’s life time are in line with the requirements of this type of protection. No change to the wiring or change of components are allowed without prior information about the impact on the type of protection of these changes. Always ask feedback of MRU for the intended changes.

15.2. IP 65 cabinet

The SWG100BIO-Ex. has an IP 65 cabinet. The cabinet prevents explosive gases and dust to penetrate inside. Prevention from dust is a basic requirement to the type of protection. Therefore, it is required to always operate the instrument with the door being closed. Perform a visual inspection in regular time intervals to verify the housing and the seals of the door and window are not damaged.
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
21 / 108

15.3. CH4 measurement of inside cabinet

While the components internal to the instrument are verified to meet the type of protection, it is nevertheless required to monitor the cabinet to prevent explosive atmosphere being present. Therefore SWG100 BIO-EX implements a regular monitoring of the CH4 content inside the cabinet with the help of the CH4-NDIR bench used for biogas measurement.
In case an explosive atmosphere is detected inside the cabinet, the SWG100BIO-EX outputs an alarm signal to the remote power supply box, which interrupts the power supply of the full instrument.
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
23 / 108
Solenoid valve for CH4 monitoring
CH4 / CO2 NDIR-bench
Menu screen in the „CH4

Desciption of the CH4 cabinet monitoring

The main parts of the system are follow ing:
The CH4/CO2 NDIR- bench.
A solenoid valve, which feeds the atmosphere inside the case to th e CH4/CO2 NDIR
bench.
The figure below shows the position of the two components.
The CH4 cabinet measurement is inforced after each air purging phase / zero point. This phase takes 2 minutes and is called 'CH4 case measurement'. During this phase the menu indication bar will indicate 'CH4 case meas.' and a count-down starts.
During this phase, all solenoid valves are closed and the solenoid valve, which feeds the gas from inside the cabinet, will open. The first 90 seconds of the CH4 monitoring is used to feed the inside cabinet atmosphere to the NDIR bench. Only in the last 30 seconds the CH4/CO2 NDIR- bench detects the CH4 concentration. During this time the measured CH4 values will be shown also as life values.
The CH4 concentration is displayed as CH4 % concentration as well as CH4 concentration in % LEL in the measurement menu (see screenshots below).
If the CH4 concentration rises over an user defined value, an alarm will be triggered for 2 minutes. After the 2 minutes the analyser will power down to prevent an explosion.
The CH4 alarm threshold can be set in the general menu (find on the path: EXTRAS /
GENERAL MENU).
If the user must perform maintenance works on the analyser and the alarm appears, it is possible to delay the power down for 1 hour to max 2 hours. The alarm menu can be closed with the key F1 or ESC. The alarm is still active, the user will see a flashing red 'OFF' on the menu indication bar.
case meas.” mode
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
24 / 108
Screen after a delay was added
Right / left arrow keys General menu: Menu-
Switch off alarm. The user has 2 minutes
point to set the LEL Level
before the SWG100BIO-Ex power down.

Self test of the CH4/CO2 NDIR- bench

The CH4/CO 2 NDIR-bench being a part of the SWG100BIO-EX protection system, is requ ired to be monitoring itself. In case the bench is supposed to be not working correctly, the alarm output is activated as well, which will lead to a power down of the instrument.
This self test is performed regularly and intrinsically by the biogas sample gas measurement itself. The analyser detects the CH4 concentration on all installed sample gas inlets. If the CH4 concentrations of all inlets are below 5 Vol.%, the alarm is triggered. The user has the possibility to manually delay the power-down in 1 hour intervals (max. 24 hours), in case the missing CH4 is known to be due to the sample gas composition. Do not delay the alarm if there is any doubt on the NDIR bench’s performance. The alarm is still active, the user will see a flashing red 'OFF' on the menu indication bar.
In case the user needs to pause the self test supervision for a limited time interval, this can be done for a couple of hours (min.2:00:00 max.24:00:00) in order to prevent a switch-off alarm. To pause the function, select to the path: EXTRAS / CH4 SENSOR SUPERVISION. The following menu will appear. With F3 the supervision function can be switched off and on. In the screen a countdown appears. This countdown shows when the paused CH4 sensor supervision will restart. With the right/ left arrow keys the countdown can be set up until 24 hours.
to set the countdown up or down.
After power-on the CH4 sensor supervision will start as soon as all sample gas inlets have been measured once and not earlier than 1 hour operation.

15.4. Remote Power Supply Box

When the SWG100BIO-EX outputs an alarm (due to internal CH4 detected or due to failed self test) it is required to switch off the instrument completely by means of an external power supply box, which shall be located in safe area, in the power supply distribution cabinet.
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
25 / 108
The Remote Power Supply Box as optionally supplied by MRU, has a switch-off contactor (K1), with 3 make contacts (normally open). The power contactor K1 is activated after the analyser is powered on by pushing the power-on button until the green LED is light-on. In case of an SWG100BIO-EX alarm, the power contactor K1 is deenergized. The circuit diagram shows the correct wiring of the power co nta ctor insta lled in the remote power supply box. For more information see chapter 20.3
The circuit diagram of above pictured device must be implemented by user as part of the safety concept. User can either use the remote power supply box as offered by MRU, or install inside the power distribution cabinet a similar circuit (see diagram in chapter 20.3) using a power contactor and push-start button.
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
27 / 108
Screenshot shows default setting, when
The two auto-configurations, which can be selected.

16. Sample Gas measurement

16.1. Cycle configuration

Path and default setting

EXTRA/ MEASUREMENT CYCLE CONFIG.
When the menu “MEASUREMENT CYCLE CONFIG.” is selected the user definable setting for the measurement cycle will appear (see screenshot below).

General information

the „MEASUREMENT CYCLE CONFIG.” will be started the first time.
The menu point “CYCLE CONFIGURATION” allows the user to configure an individual measurement cycle. Every installed sample point can be configured. For the configuration the user has the following phases, which can be selected:
Zeroing.
Purging.
Stand-by.
Measurement SPx (SPx stands for Sample point 1, 2…).
The configuration is performed with the three function keys F1, F2 and F3.
F1: Delete a phase.
F2: Make an Auto-config.
F3: Insert a new phase.
OK: View/change phase details
Left/right: Change the phase type.

Auto configuration

With F2 the “Auto-config.” can be selected. The user can select one of two default cycle configurations.
One zeroing / cycle.
One zeroing / sample point.
The first program is for applications where the different measurement points have almost the same gas concentrations. The zeroing is not necessary at every change of the measurement SPX. The second program is for applications where the different measurement points have different gas concentrations. A zeroing is recommended after every measurement point change. The screenshots below show the “One zeroing / cycle” and “One zeroing / sample point” in comparison.
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
ZEROING
Measuring site valves
Valves closed
Zeroing valve
Valve open
Duration
2min to 1 h
Recommendation
5min., in general not to be changed by user as depending only on analyzer internal setup
+
=
Screenshot shows how a
Depending on the analyser type, the first or the first and second phase cannot be deleted, deactivated or moved to another position.

Delete a phase

With F1 a phase can be deleted. To do this, select the phase, which should be deleted and press F1.

Insert a phase

With F3 a new phase is inserted in the measurement cycle. With the right/left arrow keys the different phase types can be selected.
In the title-bar the entire cycle time is shown. It is called “Measurement cycle config.” With OK the “Cycle phase details” can be shown and changed.
phase can be deleted. In this example the last phase “Measurement SP3” is deleted.

Configuration of the phase details

In this chapter the different cycle phase details will be explained. Zeroing (Cycle phase details): In the cycle phase detail “ZEROING” the zeroing time can be configured.
SWG100BIO Ex. User Manual
MRU GmbH, D-74172 Neckarsulm
29 / 108
MEASUREMENT SPx
Measuring site valves
Valve of selected site is open, others closed
Zeroing valve
Valve closed
Duration
Phase duration: 2 min. to 24 h
H2S-low: Activated/protect (Optional)
STAND-BY
Measuring site valves
Valves closed
Zeroing valve
Valve closed
Duration
Phase duration: 2 min to 24h Quiet time: calculated
Measurement SPX (Cycle phase details): In the cycle phase details of “Sample point X” the measurement time and the suction delay can be configured. Each sample point can be configured individually. In the cycle phase details the following times can be set:
Suction/response time: 30 sec. to 1h Pure measurement: calculated
Stand-by (cycle phase details): In the cycle phase details “Stand-by” the sleep mode time can be configured. In the cycle phase details the following times can be set:
Phase duration: Entire Stand-by time (Purging time + Quiet time = Phase time).
Purging time: The time, to purge the analyser with ambient air, through the zero gas inlet.
Quiet time: The time, where the analyser is in the pure stand-by mode.
Purging time: 30 sec. to 1h
Purging (cycle phase details): The purging is a separate configuration point to purge the analyser with ambient air through the zero gas inlet. It can be helpful, if the analyser must switch between a sample point with different sample gas concentrations.
SWG100 BIOGAS User Manual
MRU GmbH, D-74172 Neckarsulm
PURGING
Measuring site valves
Valves closed
Zeroing valve
Valve open
Duration
30 sec. to 1 h
=

Activated/deactivated a phase

The user has the opportunity to deactivate a phase in the measurement configuration cycle. This could be necessary for example if a sample point is temporarily not in use. The activation and deactivation of a phase can be done in the cycle phase details of the concerning phase.

Example for the deactivation of a phase

In this example the “Measurement SP2” will be deactivated. The deactivated phase is grey out.
Loading...
+ 78 hidden pages