mru VARIOluxx, MGAprime, MGAluxx User Manual

OPTION
Gas ow measurement
VARIOluxx MGAluxx MGAprime
USER MANUAL
9452EN-OPT
Legal notices / Intellectual property rights comments
Original user manual
© 2018 by MRU
No part of this manual my be published in any form (print, photocopy, electronic media or any other publication form) without a written approval by the publisher.
All user trade marks and name mark descriptions, even those which are not marked as such, are properties of the respective owners.
Edition: 2018-08-14, V01
Content
1 Gas ow measurement ............................................................................................................... 3
1.1 Setting the parameter ..................................................................................................................... 3
1.1.1 Assign a measured value ......................................................................................................3
1.1.2 Set parameters ........................................................................................................................4
1.1.3 Store the measurement results ..........................................................................................5
1.1.4 Continuous data logging ......................................................................................................5
1.1.5 Export der Messung auf SD-Karte .....................................................................................5
2 Appendix ....................................................................................................................................... 6
2.1 Calculating the molar mass ........................................................................................................... 6
2.2 Calculating the gas density .......................................................................................................... 6
2.3 Calculating the current Flow velocity ........................................................................................6
2.4 Calculating the current volume ow .........................................................................................7
2 / 8 Version 01
OPTION Gas ow measurement
1 Gas ow measurement
This option allows the user to determine the gas velocity at a single poin by means of a pitot tube (order separately). Based on the dierential pressure between the ports of the pitot tube and the absolute pressure, the gas velocity is calculated. The gas density as measured by the instru­ment is taken into account.
Assuming the measured gas velocity is represenative for the gas ow in a duct, the ow rate is calculated as the product of the measured velocity and the duct's cross section area which can be inserted accordingly.

1.1 Setting the parameter

Before measuring you must enter all parameters for the gas ow mea­surement in the menu Measure.

1.1.1 Assign a measured value

Gas ow measurement
f Double touch the value eld.
Ö A list with of all available measured values is displayed.
f Choose the wanted value and “replace” or "insert".
OPTION Gas ow measurement
3 / 8Version 01
Loading...