![](/html/51/51f4/51f47f7dfa29365fa63f6f3582ca5fb8c8d101a4aa8131b8e313ba282483fbf6/bg1.png)
OPTION
Gas ow
measurement
VARIOluxx
MGAluxx
MGAprime
USER MANUAL
9452EN-OPT
![](/html/51/51f4/51f47f7dfa29365fa63f6f3582ca5fb8c8d101a4aa8131b8e313ba282483fbf6/bg2.png)
Legal notices / Intellectual property rights comments
Original user manual
© 2018 by MRU
No part of this manual my be published in any form (print, photocopy, electronic media or
any other publication form) without a written approval by the publisher.
All user trade marks and name mark descriptions, even those which are not marked as such,
are properties of the respective owners.
Edition: 2018-08-14, V01
MRU
Content
1 Gas ow measurement ............................................................................................................... 3
1.1 Setting the parameter ..................................................................................................................... 3
1.1.1 Assign a measured value ......................................................................................................3
1.1.2 Set parameters ........................................................................................................................4
1.1.3 Store the measurement results ..........................................................................................5
1.1.4 Continuous data logging ......................................................................................................5
1.1.5 Export der Messung auf SD-Karte .....................................................................................5
2 Appendix ....................................................................................................................................... 6
2.1 Calculating the molar mass ........................................................................................................... 6
2.2 Calculating the gas density .......................................................................................................... 6
2.3 Calculating the current Flow velocity ........................................................................................6
2.4 Calculating the current volume ow .........................................................................................7
2 / 8 Version 01
OPTION Gas ow measurement
![](/html/51/51f4/51f47f7dfa29365fa63f6f3582ca5fb8c8d101a4aa8131b8e313ba282483fbf6/bg3.png)
MRU
1 Gas ow measurement
This option allows the user to determine the gas velocity at a single poin
by means of a pitot tube (order separately). Based on the dierential
pressure between the ports of the pitot tube and the absolute pressure,
the gas velocity is calculated. The gas density as measured by the instrument is taken into account.
Assuming the measured gas velocity is represenative for the gas ow in a
duct, the ow rate is calculated as the product of the measured velocity
and the duct's cross section area which can be inserted accordingly.
1.1 Setting the parameter
Before measuring you must enter all parameters for the gas ow measurement in the menu Measure.
1.1.1 Assign a measured value
Gas ow measurement
f Double touch the value eld.
Ö A list with of all available measured values is displayed.
f Choose the wanted value and “replace” or "insert".
OPTION Gas ow measurement
3 / 8Version 01