MRC STE-TIN, STE-TAN Operation Manual

Operation Manual
STE-TIN/TAN
Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529
mrc@mrclab.com
Thank you for choosing this Steam Sterilizer.
Please read the instructions manual carefully in order to install the equipment and operate the equipment in an efficient manner.
TABLE OF CONTENTS
1.GENERAL ------------------------------------------------------------ 1
2.TECHNICAL PARAMETER ---------------------------------------- 2
3.PACKING CONTENT ----------------------------------------------- 3
4.INSTALLATION ---------------------------------------------------- 4
5.OPERATION ---------------------------------------------------------- 5
5.1 READY ------------------------------------------------------------- 5
5.2 PREPARE THE MATERIAL TO BE STERILIZED -- 6
5.3 SELECTING THE STERILIZATION PROGRAM -------------- 7
5.4 RUNNING THE STERILIZATION PROGRAM ----------------- 8
5.5 TEST PROGRAMS -------------------------------------------------11
6. ADVANCED SETTING -------------------------------------------- 12
7. MAINTENANCE ---------------------------------------------------- 17
8. TRANSPORTAND STORAGE ------------------------------------ 20
9. ERROR CODES ----------------------------------------------------- 21
10. SAFETY DEVICES ------------------------------------------------ 22
APPENDIX
1. CHARACTERISTICS OF THE FEEDING WATER -------------- 23
2. DIAGRAMS OF THE STERILIZATION PROGRAMS --- 24
3. ELECTRICAL DRAWING ----------------------------------------- 31
4. HYDRAULIC DRAWING ----------------------------------------- 33
--------
------
Instructions manual
Instructions manual
This sterilizer described in this manual is intended for the sterilization of
re-useable surgical instruments and material.
It operates automatically with 134 and 121 sterilization temperatures.
It has been produced in accordance with the EN 13060.
℃℃
1 General
4
3
2
1
1 Distilledwater tank 2 LCDscreen 3 Controlpanel 4 Mainswitch 5 Drainconnector of distilledwater tank 6 Doorhandle 7 Drainconnector of usedwater tank 8 USBport (optional)
9 Printer port 10 Printer power 11 Bacteriological filter 12 Safety valve 13 Condenser ventilation 14 Circuit breaker 15 Power supplycord 16 Rating plate
In order to proper use the sterilizer, please be sure to read the warning and attention carefully for safety.
Security Notice
This symbol is grounding protection inside the machine.
HOT SURFACE. This symbol is visible on the front of the panel after open the door.
Important safety information. This symbol is used to draw the attention of the reader to particularly important notions for operator safety.
5
6
7
11
13
15
12
16
8
9
10
1
14
2 Technical Parameter
Chamber
Rated Voltage
Main Fuses
Nominal power (CLASS B)
Sterilization Temperature
Capacity of the distilled
water tank
Operation temperature
Outside size
Net weight CLASS B)
Noise
Relative Humidity
Atmospheric pressure
Item
12
16
Φ230mmX360mm
Φ200mmX360mm
AC220V-240V(AC110V-130V), 50-60Hz
F16A/250V(F25A/250V for 110V)
2000VA
121 /134
℃℃
Approx 2.5L (water at level MAX)
Approx 0.5L (water at level MIN)
5-40
445mm(width)X 410mm(height)X 605mm(depth)
76kPa -106kPa
max. 80%, non condensing
<70dB
41kg
45kg
2
Nominal power (CLASS N)
1900VA
Net weight CLASS N)
39kg
43kg
Instructions manual
Instructions manual
3 Packing Content
Quantity
Item
No.
2
3 4
5
6
7
Instrument tray
Instrument tray rack
Instrument tray handle
Door adjustment tool
Draining hose
Instructions manual
2
1
1
1
2 1
1
1
Steam sterilizer
1
3
Door seal
8
4 Installation
* There must leaves 10cm gap around sterilizer, and 20cm on top side. the
clearance required for the movement of the door leave an at least
350mm fan-shaped space in front of the door.
* The place which sterilizer located must be ventilated,make sure that the
radiator not being jammed.
* The sterilizer should be placed on a levelworktable.
* Don't cover or block the door,ventilation or radiation openings on the
sterilizer.
* Don't place the sterilizer near a sink or in alocation where it is likely to
be splashed.
* Keep away from all sources of heat.
4
Instructions manual
Instructions manual
605
Distilled water tank is lacking of water.
Distilled water tank is full
Used water tank is full.
Door locked
information output to USB port
Printer connect
Notice: Before using the sterilizer at the first start-up or any time the
icon blink, it is necessary to fill the distilled water tank with distilled water.
5.1 Ready
5.1.1 Open the door and take out all the instrument trayand other accessories inside, unpack and clean them.
5.1.2 Connect the power, and connect theprinter (See 6.5)
5.1.3 Power on The switch is located underneath the control panel on the front side of the machine . After switching on, the machine turns on the LCD .
Then it will show the door position, water level, working program, date, time and etc. . The machine will alert by beep sounds if there are
problems.
5 Operation
Program
Alarm code
process curve
Temperature
Pressure
Setting
Time
Date
5
5.1.4 Filling the distilled water Open the top lid, and fill the tank with distilled water by cup or tank. When you hear a beep signal, it means the water level exceeds the max. level. The will be displayed. Please stop filling immediately.
5.2 Preparing the material to be sterilized
To get thebetter effectiveness of the sterilization process andto preserve
the material in time, follow the indications below reported.
* Arrange the tools of different metal (stainless steel, moderate steel,
aluminum, etc.) on different trays or however well separate between them;
* In case of not stainless steel tools, interpose a sterilization paper napkin or
muslin cloth between tray and tool, avoiding direct contacts between the
two different materials;
* Verify all the tools aresterilized in open position;
* Arrange the containers (glasses, cups, test-tubes, etc.) on one side or
inverted position, avoiding possible water stagnation;
* Don't overload the trays over the stated limit (see Appendix 1).
* Don't stack the trays one above the other or put them in direct contact with
the walls of the sterilization chamber.
* Always use the instrument tray handle.
* Wrap the tools one by one or, if more tools have to be set in the same wrap,
verify that they are of the same metal;
*Seal the wrap with sterilization adhesive ribbon or by a thermal sealer.
* Don't use metallic clips, pins or other, as this jeopardizes the maintenance
of the sterility;
6
The waterlevel should not exceeded thisport.
Instructions manual
Instructions manual
* Turn the sterilization paper in order to set the plastic part downward
(tray side) and the paper part upward.
Always wrap the tools in case of prolonged store.
5.3 Selecting the sterilization program
5.3.1 LCD It displays the cycle temperature,
pressure, error code, sterilization state and program.
5.3.2 temperature button
Select temperature of sterilization.
5.3.3 Program button
Select program of sterilization.
5.3.4 START/STOP button
Press this button to start the sterilization cycle, holding this button above
3 seconds to stop the cycle.
5.3.5 Select the program
Press TEMP button to select the temperature. Andpress PROGRAM button to select the program.
TEXTILE OR DRYING
Notice: The button will be locked for 10 seconds after you switch on.
It initializes its system and check the states during that time.
7
8
SOLID
WRAPPED
(CLASS B)
(CLASS N)
5.4 Running the sterilization program.
After selecting program, put the
instruments into the chamber by tray
handle.
5.4.1 After the instruments are loaded, you may close and lock the door by turning the door handle clockwise until it stops.
The icon will be lightened.
Caution You must turn the door handle to the maximum position,
otherwise the machine will alarm and stop working during the cycle.
Instructions manual
Instructions manual
SOLID
WRAPPED
PRION
LIQUID
B&D TEST
HELIX TEST
VACUUM TEST
DRYING
LIQUID
TEXTILE OR DRYING