MRC PWR-3030 Operation Manual

PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com
MRC.VER.01-2.12
Operation Manual for
Power Supply
PWR-3030
1
Packing list
3:5:
- 1x PWR-3030 Power Supply
- 1x Power Cord
- 1x PWR-3030 Power Supply Instruction Manual
MRC is liable for all missing or damaged parts / accessories within 7 days after customer receives this instrument package. Please contact MRC imm ediately reg arding th is issue. If n o respo nse is received within such a time period from consignee party, it will be the consignee party’s whole responsibility.
2
Table of contents
Packing list................................................................................................ 1
Warning .................................................................................................... 3
Product Specifications .............................................................................. 6
Product Description .................................................................................. 7
Introduction ............................................................................................... 8
Overview ............................................................................................ 8
Controls and Features ....................................................................... 9
Installation Instructions ........................................................................... 10
Operation Instructions ............................................................................. 10
Display Screen ................................................................................ 11
Constant Setup ................................................................................ 11
Programming Setup ......................................................................... 14
Limiting Parameter Setting .............................................................. 17
Troubleshooting Guide ............................................................................ 18
Maintenance ........................................................................................... 21
Ordering Information ............................................................................... 21
Warranty ................................................................................................. 22
3
Warning
MRC PWR-3030 Power Supply has b een tested and found to comply with the limits for the CE regulation. Also, PWR-3030 Power Supply series is RoHS compliant to deliver confident product which meets the environmental directive. These limits are designed to provide reasonable protection against harmful interference when the instrument series is operated in a commercial environment. This instrument series used together with power supply unit generates, uses, and can radiate radio frequency energy, and if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this instrument series in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at their expense. Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. It is strongly recommended for the user to read the following points carefully before operating this equipment.
1. Read and follow the manual instr ucti o ns care ful l y.
2. Do not alter the equipment. Failure to follow these directions could result in personal and/or laboratory hazards, as well as invalidate equipment warranty.
3. Use a properly grounded electrical outlet with correct voltage and current handling capacity.
4. Disconnect from power supply before maintenance and servicing. Refer servicing to qualified personnel.
5. Never use this instrument series without having the safety cover correctly in position.
6. Do not use the unit if there is any sign of damage to the external tank or cover. Replace damaged parts.
7. Do not use in the presence of flammable or combustible material; fire or explosion may result. This device contains components which may ignite such materials.
8. Refer maintenance and servicing to qualified personnel.
9. Ensure that the system is connected to electrical service according to local and national electrical codes. Failure to make a proper connection may create fire or shock hazard.
10. Use appropriate mat er ial s and op er ate correctly t o av oi d pos si bl e ha zards of explosion, implosi on or r elease o f tox ic or flammabl e gases aris ing from
4
overheated materials.
11. The unit shall be operated only by qualified personnel.
Safety Information
Use high level of precaution against any electrical device. Before connecting the electrical supply, check to see if the supply voltage is within the range stated at the rating label, and see to it that the device be seated firmly. Place the unit in a safe and dry location; it must NOT touch the surrounding. Follow the safety precautions for chemicals / dangerous materials. If needed, please contact qualified service representative or mrc@mrclab.com
Environmental Conditions
Ensure the instrument is installed and operated strictly in the following conditions:
1. Indoor use only
2. ≤95% RH
3. 75 kPa – 106 kPa
4. Altitude must not exceed 2000 meters
5. Ambient to 40°C operating temper at ur e
6. Pollution degree: 2
7. Mains supply voltage fluctuati ons up t o ±10% of the normal voltage
Avoiding Electrical Shock
Follow the guidelines below to ensure safe operation of the unit. PWR-3030 power supply has been designed to use with shielded wires thus minimizing any poten tial shock h az ard to t he user. MRC recommends against the use of unshielded wires.
To avoid electrical shock:
1. In the event of solution accidentally spilling into the instrument, it must be
dried out for a period of time (at least 2 hours) and restored to NORMAL CONDITION before each operation.
2. Never connect or disconnect wires loading from the power jacks when the
red indicator light of power switch is on.
3. WAIT at least 5 seconds after stopping a run before handling output leads
or any connected apparatus.
4. ALWAYS mak e sur e that your hands, work area, and ins t r uments are clean
and dry before making any connections or operating the power supply.
5. ONLY connect the power cord to a properly grounded AC outlet.
5
Avoiding Damage to the Instrument
1. Do not attempt to operate the device if damage is suspected.
2. Protect this unit from physical damage, corrosive agents and extreme temperatures (direct sunlight, etc.).
3. For proper ventilation and safety concerns, keep at least 10 cm of space behind the instrument, and at least 5 cm of space on each side.
4. Use high level of precaution against the damages on the unit.
5. Do not operate the unit out of environmental conditions addressed above.
6. Prior to applying any cleaning or decontamination methods other than manufacturer’s recommendation, users should check with the manufacturer’s instruct i on to see if the proposed method wil l damag e th e equipment.
7. When the PWR-3030 power supply is taken into a cold room environment, the unit can be operated immediately. However, when the PWR-3030 power supply is removed from the cold room environment, let the unit equilibrate to room temperature for a minimum of 2 hours before using it
.
Equipment Operation
Follow the guidelines below to ensure safe operation of the unit:
1. Check the displayed figures to see if the unit is functioning correctly before using this unit.
2. NEVER access dang erous ch emic als or ot her ma teri als to prev ent pos sible hazard of explosion and damage.
3. Do not operate the unit without lids or covers to prevent possible hazards.
4. A tem por ar y c onductivity caused by condensation mig ht occ ur even though this series is rated Pollution Degree 2 in accordance with IEC 664.
Symbols
The symbols used on PWR-3030 Power Supply are explained below.
Indicates an area where a potential shock hazard may exist. Consult the manual to avoid possible personal injury or instrument damage.
Indicates disposal instruction. DO NOT
throw this unit into a municipal trash bin when this unit has
reached the end of its lifetime. To ensure utmost prote
ction of the
global environment and minimize pollution, please recycle this unit.
Max. voltage: 300V Max. Current: 3000mA
Max. Watt: 300W
6
Product Specifications
Output Voltage / Inc. 5 - 300V / 1V Output Current / Inc. 10 - 3,000mA / 10mA Max. Watt / Inc. 300W / 1W Operating Mode Voltage or current or power Control Microprocessor controller Program Storage 30 programmed files Program Multi-Step Up to 6 steps Terminal Pairs 4 pairs Display 2.6” LCD Timer Constant mode: 1-9999 min. with alarm, continuous
Programmable mode: 1-999 mi n with alarm,
continuous Safety Device No Load detection
Leakage detection Sudden load change detection Over temperature protection Over current detection Over voltage detection
Shrouded plugs and sockets Crossover Yes Operating Temperature
Ambient - 40°C Unit Dimension 190 x 305 x 95 mm (W x L x H) Construction Material Flame retardant ABS faceplate and aluminu m Rated Voltages 100 - 240V; 47/60 Hz Input Rating 360W Weight approx. 2.5 kg
7
Product Description
The MRC PWR-3030 Power Supply series is a microprocessor controlled and designed to meet most electrophoresis needs. This manual describes the setu p and operation of the PWR-3030 Power Supply including important information on safety and maintenance of the unit. The PWR-3030 power supply is capable of running horizontal & vertical electrophoresis, SDS-PAGE, native PAGE application, two-dimensional electrophoresis, and electro-blotting. It comes equipped w ith a timer and al arm function as w el l as a pause function. Furthermore, the powerful specific ati ons plus four terminal pairs can be used for multiple electrophoresis units simultaneously.
MRC PWR-3030 power supply provides Constant Voltage or Constant Current or Constant Power to instruments used in electrophoresis. 4 pairs of terminals and the powerful specification equipment enable more capabilit y of the PWR-3030 power supply than other existing similar product on the market. PWR-3030 also has a 2.6” LCD screen where many parameters are displayed at the same time, which provides a better aspect of the user friendl y concept to the user.
Features of 3:5 power supply:
- Compact size
- Advanced capacity: 300W, 3000mA, 300V
- Microprocessor controller
- Constant voltage, constant current, and constant power
- Four outlet terminal pairs
- LCD display
- Timer with alarm function
- Advanced safety devices
- Stackable
- Wide applications for DN A, RN A and protein electrophoresis
10
Insta llation Instructions
The MRC PWR-3030 Pow er Suppl y is actually a pre-installed instrument. As long as it is placed on a sturdy and level surface in a safe and dry place, and is further connected with a well-prepared electrophoresis system, it is ready for operation.
Operation Instructions
*Note: To operate under constant voltage or constant current modes, adjust the other parameter to the maximum value. For example, to operate under constant voltage, adjust current to max before running using constant voltage, and vice versa.
The PWR-3030 Power Supply is designed to operate under two modes: Constant Mode or Programming Mode, depending on your electrophoresis conditions.
Use the Constant Voltage / Current / Power Operation for applications that require only one specific voltage limit, current limit, and power limit continuously during the entire durat ion of electrophoresis.
11
Display Screen
Switch on the power, the screen will first show the version and model name. After the welcome page you will be able to select the oper ation mode: Constant or Program mode (Screen shown below)
Constant Setup
Program Setup
Constant Setup
This section gives an instruction for constant mode operation. Constant Voltage / Current / Power Mode allow you to specify a voltage limit or a
current limit to be used continuously during the entire duration of electrophoresis. Review the guidelines provided in this manual before starting electrophoresis using PWR-3030 Power Supply. We recommend reading the guidelines provided in this manual for best results before starting an operation.
1. Place the PWR-3030 Power Supply on a s tur d y and level sur face in a safe and
dry place, away from laboratory traffic.
2. Ensure that the AC power switch is OFF, and then plug the three-pronged
power cord into a grounde d thr ee-pro ng AC outlet of the appropriate volt age (110V to 240V as indicated on the rating sticker near the AC cord on the back of the unit).
3. Connect the DC output jacks from the electrophoresis unit; insert the red
lead (+) into the red output jack, and the black lead (-) into the black output jack.
12
4. Turn on the power switch. The version and model name will appear on the screen.
5. Use Key and Key to select , and then press
the Key or to enter the next screen: .
6. Use , , , or to move cursor to the p ar am ete r s such
as voltage (V) or current (mA) or power (W) or time (Min.). Press the key to set the specified parameter.
7. Use or to set the appropriate value, and then press .
Move to the next parameter until all the parameters are set in the same operation method.
8. Press Key to start electrophoresis. the LED will light up,
and the screen will show the real time parameter values, .
Press or to see the following screen, . Press
or to go back to
9. Press Key to temporarily interrupt the power to the ongoing
13
electrophoresis without terminating the operation. , the LED is
flashing. Press Key to restart the run.
10. Press t he Key again to stop electrophoresis.
11. When the run is completed, operation stops with alarm and
is shown on the screen. Press Key to terminate a timed run, and turn the AC power OFF by the switch on the rear.
12. T o change limits of electrophoresis in progress
If you need to make changes to the current running limits, you must stop
electrophoresis by pressing the K ey. Press Key to enter the
setting screen, . Enter the changes and then press Key once again to restart your operation.
Note: After stopping and restarting an operation, the timer resets to selected time and does not take into account the time that electrophoresis was in progress before it was stopped.
14
Programming Setup
Instructions for operating PWR-3030 Power Supply in the Programming Operation are provided in this section. The Programmable Mode allows you
to vary levels in vol tage ( V) , c ur r ent (mA), and power (W) during specified time periods as discrete changes (STEP) or as gradients (RAMP) for up to 6 steps, depending upon your electrophoresis needs. The PWR-3030 power supply is capable of having storage room for 30 di ffere nt prog r am files for the user’s convenience. We recommend reading the guidelines provided in this manual for best results before star ti ng an opera ti on.
1. Place the PWR-3030 Power Supply on a sturdy and level surface in a safe
and dry place, away from laboratory traffic.
2. Ensure that the AC power switch is OFF, and then plug the three-pronged
power cord into a grounded three-prong AC o utl et of the appropriate voltage (110V to 240V as indicated on the rating sticker near the AC cord on the back of the unit).
3. Connect the DC output jacks from the electrophoresis unit; insert the red
lead (+) into the red output jack, and the black lead (-) into the black output jack.
4. Turn on the power switch from the back panel. TheYHUVLRQDQGPRGHOQDPH will appear on the screen.
5. Use Key and Key to select , and then press
Key or to enter the next screen, .
6. Press Key first and then use or to select appropriate
15
file number, and then press Key to enter the following screen,
.
7. Use , , , or to move cursor to the p ar am ete r s such
as voltage (V), cur rent ( mA), pow er (W), or time (min.). Press Key to set the specified parameter.
8. Use or to set the appropriate value, and then press , and move to the next parameter until all the p aram eters ar e set in th e sam e operation method.
9. Use Key to move down to for setting Step 4 – 6.
Press Key back to Step 1 – 3 screen, .
10. Press Key to start electrophoresis. , the LED is lit, and the
screen will show the real time parameter values, . Press
or to see the following screen, . Press
or Key back to .
11. Press Key to temporarily interrupt power to ongoing run without
16
terminating the operation. , the LED is flashing. Press Key to restart the run.
12. W hen electrophoresis is completed, is shown on the screen.
Press the Key again to stop electrophoresis.
13. To change limits of electrophoresis in progress.
If you need to make changes to the current running limits, you must stop
electrophoresis by pressing the Key. Press Key to enter the
setting screen, . Enter the changes and then press Key once again to restart your operation.
Note: After stopping and restarting an operation, the timer resets to selected time and does not take into account the time that electrophoresis was in progress before it was stopped.
17
Limiting Parameter Setting
The PWR-3030 Power Supply is capable of operating under limiting voltage, limiting current, or limiting power in both Constant Setup mode or Programming Setup mode. We use Programming Setup mode here as an example.
Voltage Limiting
1. Use Key, Key, Key, Key, and Key to set
Maximum Current (3A) and Maximum Power (300W), and on the real time
screen “Volt” will be shaded, for instance, .
Current Limiting
1. Use Key, Key, Key, Key, and Key to set
Maximum Voltage (300 V) and Max imum Pow er (300 W), and on t he real tim e
screen “mA” will be shaded, for insta nc e, .
Power Limiting
1. Use Key, Key, Key, Key, and Key to set
Maximum Voltage (300V) and Maximum Current (3A), and on the real time
screen “Wat” will be shaded, for insta nce, .
18
T roubleshooting Guide
Many operating pr oblems may be solv ed by carefully reading a nd following the instructions in this manual accordingly. Some suggestions for troubleshooting are given below. Should these suggestions not resolve the problem, contac t our SERVICE DEPARTMENT or a distributor in your region for assistance. If troubleshooting service is required, please include a full description of the problem.
Problem Cause Solution
No Display / lights
No AC power
Check if PWR po
wer supply is unplugged, or
AC
power source problem AC power cord is not connected
Check AC power cord connections at both ends. Use the correct cords.
The fuse has blown Replace the fuse
Repeated fuse broken
Hardware failure Contact MRC s
ervice department
Operation stops with alarm: The screen
displays
Electrophoresis leads are not connected to the power supply or to the electrophoresis unit(s), or
there is a broken circuit in
the electrophoresis cell
Check the connections to the power
supply
and on your electrophoresis
cell to make
sure the connection is
intact; check
condition of wires in electrophoresis unit. Close the circuit by reconnecting the
cables.
Press START/STOP to restart the run. High resistance due to tape left on a pre-cast gel, incorrect buffer concentration, or incorrect buffer volumes in the electrophoresis cell
Correct the condition by making
sure the
tape is removed from the pre-
cast gel,
buffers are prepared
correctly, and the
recommended volum e of buffer is
added to
the electrophoresis unit.
Operation stops with
alarm: Display shows
Bad connections for terminal connectors or damaged wires or damaged platinum wires
Check all the connections to terminators,
cables, wires, and gel tanks
19
Operation stops with alarm: Display shows
Circuit is interrupted
- Verify that the running buffer is correct.
- Verify the all cables are attached correctly
- Turn the Power switch off and
on again;
restart application.
- If you cannot restart the instrument, turn off the power, disconnect the power cord from the outlet, and contact Technical Service.
Operation stops with
alarm: Display shows
Circuit is interrupted
- Verify that the running buffer is correct.
- Verify the all cables are attached correctly
- Turn the Power switch off and
on again;
restart application.
- If you cannot restart the instrument, turn off the power, disconnect the power cord from the outlet, and contact Technical Service.
Operation stops with alarm: Display shows
Ground leak detected during run
Check the electrophoresis system for improper grounding. Restart the power supply by turning the Power switch off and on.
Operation stops with alarm: Display shows
Power suppl y is overheating
- Turn off power supply. Check for sufficient
airflow around the power supply fan. After cooling down, restart the power supply by turning the Power switch to the on position.
- If you cannot restart the instrument, turn off
the power, disconnect the power cord from the outlet, and contact Technical
Service. Operation stops with alarm: Display shows
and countdown
Power supply is getting restart ON
- Verify that the electrophoresis condition and system is correct.
- Pay attention to the own safety.
20
Encountering Problems
1. Check the troubleshooting section.
2. Call Technical Service or e-mail to mrc@mrclab.com
3. If the unit must be shipped back for repair, contact MRC or the distributor for a Return Authorization Number and shipping instructions. The unit will be repaired as quickly as possible and returned to you.
Replacing the Fuse
For additional fuses, contact MRC co. ltd.
To replace the fuse:
1. Turn off the main power switch at the rear of PWR-3030 Power Supply and
detach the power cord from the rear of PWR-3030 Power Supply.
2. Open the fuse compartment located inside the Power Entry Module by
inserting a small flathead screwdriver into the slot below the ON/OFF switch. Turn the screw driv er to gent ly pry open the fuse comp artmen t. Note: the fuse compartment will not open with the power cord in place.
3. Pull the fuse h older out of the compartment and inspect the fuse. I f t h e fuse
is burned or there is a break in the fuse element, replace the fuse with an identical type of fuse (4A/250V~) as provided in the fuse holder (see figure below)
4. Place the fuse holder back into the compartment.
5. Snap the cover close.
21
Maintenance
PWR-3030 Power Supply uses all solid-state components and should require no maintenance or recalibration under normal use. The housing may be cleaned with a dry cloth. If the unit must be ret urned for repair, contact our SERVICE
DEPARTMENT or your local distributor for shipping instruction.
Ordering Information
PWR-3030 Power Supply
Cat. No. Description
PWR-3030 Power Supply
22
Warranty
MRC warrants apparatus of its manufacture against defects in materials and workmanship, under normal service, for one year from the shipping date to purchaser. This warranty excludes damages resulting from shipping, misuse, carelessness, or neglect. MRC’s liability under the warranty is limited to the receipt of reasonable proof by the customer that the defect is embraced within the terms o f th e warranty. All claims made und er t his warranty must be presented to MRC within one year following the date of delivery of the product to the customer.