MRC MG-3002 Operation Manual

Page 1
G ( Gauss), mT ( milli Tesla ), USB/RS232
AC/DC MAGNETIC METER
Y
our purchase of this
A
f
precision measurement.
A
lthough this METER is a complex and delicate instrument, its durable structure will allow many years of use if proper operating techniques are developed. Please read the following instructions carefully and always keep this manual within easy reach.
Model : MG-3002
OPERATION MANUAL
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529
mrc@mrclab.com
MRC. 5.
18
Page 2
TABLE OF CONTENTS
1. FEATURES..................................................................1
2. SPECIFICATIONS........................................................2
3. FRONT PANEL DESCRIPTION......................................
.
4
3-1 Display................................................................. 4
3-2 Power Button........................................................4
3-3 Hold Button..........................................................
.
4
3-4 REC Button...........................................................4
3-5 ZERO Button ........................................................4
3-6 Mode Button ( DC, AC Button ), SET Button............4
3-7 Input socket.........................................................
.
4
3-8 Probe Handle........................................................4
3-9 Probe Sensing Head..............................................
.
4
3-10 Probe Plug.......................................................... 4
3-11 Probe Head Cover...............................................
.
4
3-12 RS-232 Output Terminal......................................
.
4
3-13 Reset button.......................................................4
3-14 DC 9V Power Adapter Input Socket......................
.
4
3-15 Stand.................................................................
.
4
3-16 Tripod Fix Nut.....................................................4
3-17 Battery Compartment/Cover.................................4
3-18 Battery Cover Screw............................................4
4. MEASURING PROCEDURE...........................................
.
5
4-1 AC/DC magnetic field measurement.......................
.
5
4-2 Data Hold.............................................................7
4-3 Data Record.........................................................
.
7
5. ADVANCED MEASURING PROCEDURE..........................8
5-1 Auto power ON/OFF management..........................9
5-2 Change the unit ( mT, G )......................................9
5-3 For the further calibration usage, not available
for the user..........................................................10
6. RS232 PC SERIAL INTERFACE.....................................
.
10
7. BATTERY REPLACEMENT.............................................11
8. SYSTEM RESET..........................................................
.
12
Page 3
1. FEATURES
* Wide range, general purpose magnetic measurement.
for industrial, mechanical, material, electrical, laboratory field usage.
* Unit : G ( Gauss ), mT ( milli Tesla ). * DC and AC magnetic field measurement * DC Range : 300.00 mT/3000.0 mT. AC Range : 150.00 mT/1500.0 mT. * Resolution : 0.01/0.1 mT, 0.1/1 G. * N pole/S pole indicator. * Use Hall sensor with automatic Temp. compensation. * Zero button for DC magnetic function.
* Separate probe, easy operation and convenient
for remote measurement.
* Data hold function for freezing the desired value
on display.
* Records Maximum and Minimum readings with Recall. * RS232/USB computer interface. * Microprocessor circuit assures maximum possible
accuracy, provides special functions and features.
* Heavy duty & compact housing with hard carrying case,
designed for easy carry out & operation.
* Auto shut off is available to save battery life. * Power supply from batteries or DC 9V adapter in.
1
Page 4
2. SPECIFICATIONS
Circuit Custom one-chip of microprocessor LSI
circuit.
Display LCD size : 52 mm x 38 mm
dual function LCD display. Measurement mT milli Tesla Unit G Gauss Measurement mT Range 1 300.00 mT x 0.01 mT Range, DC Range 2 3,000.0 mT x 0.1 mT
G Range 1 3,000.0 G x 0.1 G
Range 2 30,000 G x 1 G Measurement mT Range 1 150.00 mT x 0.01 mT Range, AC Range 2 1,500.0 mT x 0.1 mT
G Range 1 1,500.0 G x 0.1 G
Range 2 15,000 G x 1 G Accuracy DC ± ( 5 % rdg. + 10 digit )
AC ± ( 5 % rdg. + 20 digit ) Frequency AC measurement : Response 50 Hz / 60 Hz Sensor Hall sensor. Temp. Probe head build the Temp. sensor Compensation for using the ATC ( automatic Temp. for hall sensor compensation ). Field Direction Uniaxial. Data Hold Freeze the display reading. Memory Recall Maximum & Minimum value. Display Approx. 1 second. Sampling Time Power off Manual off by push button or
Auto shut off to save battery life. Operating 0 to 50 . Temperature Operating Less than 85% R.H. Humidity
2
Page 5
Data Output RS 232/USB PC serial interface.
*Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
*
Connect the optional USB cable USB-01 will get the USB plug.
Power Supply 006P DC 9V battery
( Alkaline or Heavy duty type ).
DC 9V adapter input.
* AC/DC power adapter is optional.
Power Current Approx. DC 20.5 mA . Weight Approx. 270 g/0.60 LB.
( include battery & probe ) Dimension Main instrument :
195 x 68 x 30 mm ( 7.6 x 2.6 x 1.2 inch )
Probe :
198 x 25 x 19 mm
( 7.8 x 1.0 x 0.7 inch ).
Accessories Instruction manual....................... 1 PC
Included Magnetic Probe............................1 PC
Carrying case, CA-06....................1 PC
Optional RS232 cable, UPCB-02 Accessories USB cable, USB-01
Data Acquisition software, SW-U801-WIN
* Above specification tests under the environment RF Field Strength less than 3
V/M & frequency less than 30 MHz only.
3
Page 6
3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display 3-2 Power Button 3-3 Hold Button 3-4 REC Button ( Enter Button ) 3-5 ZERO Button 3-6 Mode Button ( DC, AC Button ), SET Button 3-7 Input socket 3-8 Probe Handle 3-9 Probe Sensing Head 3-10 Probe Plug 3-11 Probe Head Cover 3-12 RS-232 Output Terminal
3-13 Reset button....
3-14 DC 9V Power Adapter Input Socket 3-15 Stand 3-16 Tripod Fix Nut 3-17 Battery Compartment/Cover 3-18 Battery Cover Screw
4
Page 7
4. MEASURING PROCEDURE
4-1 AC/DC magnetic field measurement
1)Power ON the meter by pressing the " Power
Button " ( 3-2, Fig. 1 ) once.
2)
Press the " Mode Button " ( 3-6, Fig. 1 ) once , the Display will show the following screen in sequence :
Under mT unit Under G unit
N(S) 0 mT a1 N(S) 0 G a2
.0 0
N(S) 0.0 mT b1 N(S) 0 G b2
0 .0
0 mT c1 0 G c2 .0
AC 0 AC
0.0 mT d1 0 G d2 0
AC .0 AC
Note :
a1 screen is ready for DC 0.1 mT resolution measurement. b1 screen is ready for DC 0.01 mT resolution measurement. c1 screen is ready for AC 0.1 mT resolution measurement. d1 screen is ready for AC 0.01 resolution measurement.
a2 screen is ready for DC 1 G resolution measurement. b2 screen is ready for DC 0.1 G resolution measurement. c2 screen is ready for AC 1 G resolution measurement. d2 screen is ready for AC 0.1 G resolution measurement.
5
Page 8
3) Display indicator
DC magnetic field measurement
* The Display will show " N " or " S " indicator. * North pole : Display will show " N " indicator * South pole : Display will show " S " indicator
AC magnetic field measurement
* Display will show " AC " indicator. * Display will not show " N (S ) indicator.
4)unit selection :
How to select the unit ( mG, uT ), the procedures please refer to chapter 5-1 " Change the unit ( mT, G ) ", page .
5)Contact and close the testing material to the probe
head ( refer to Fig. 3 ) , the Display will show the magnetic field value along with the unit ( mG, uT ).
Contact side ( measuring face )
6
Page 9
6)Zero adjustment :
When power ON the meter, due to the environment interference, the display may show certain values ( not zero value), it is normal. * Before the measurement, fix the probe at the exact
position, press the " ZERO Button " ( 3-5, Fig. 1 ) continuously at least two seconds until the display show " 0 " value, in the same time the Display will show the " 0 " indicator then release the finger from the button.
* Press the " ZERO Button " ( 3-5, Fig. 1 )
continuously at least two seconds again, the " O " indicator will be disappeared and the Display will return to the " no zero " value ( show the environment magnetic field value ).
4-2 Data Hold
During the measurement, press the " Hold Button " ( 3-3, Fig. 1 ) once will hold the measured value & the LCD will display a " HOLD " symbol. * Press the " Hold Button " once again will release the data
hold function.
4-3 Data Record ( Max., Min. reading )
* The data record function records the maximum and
minimum readings. Press the " REC Button " ( 3-4, Fig. 1 ) once to start the Data Record function and there will be a " REC " symbol on the display.
* With the " REC " symbol on the display :
a)Press the " REC Button " ( 3-4, Fig. 1 ) once, the "
REC. MAX " symbol along with the maximum value will appear on the display. If intend to delete the maximum value, just press the " Hold Button " ( 3-3, Fig. 1 ) once, then the display will show the " REC. " symbol only & execute the memory function continuously.
7
Page 10
b)Press the " REC Button " ( 3-4, Fig. 1 ) again, the
" REC. MIN " symbol along with the minimum value will appear on the display. If intend to delete the minimum value, just press the " Hold Button " ( 3-3, Fig. 1 ) once, then the display will show the " REC. " symbol only & execute the memory function continuously.
c)
To exit the memory record function, just press the " REC Button " for 2 seconds at least. The display will revert to the current reading.
Remark : When make the DC magnetic field measurement
and execute the " RECORD " function, it can not change from the " North pole " to " South pole " or from the " South pole " to " North pole " during the measurement. It can make the measurement under the same pole ( S pole or N pole ) only, otherwise the Max., Min. value will be no sense.
5. ADVANCED SETTING
Press the " SET Button " ( 3-6, Fig. 1 ) continuously at least two seconds will enter the " Advanced Setting " mode, then press the " SET Button " ( 3-6, Fig. 1 ) once a while in sequence to select the three function, the lower Display will show :
PoFF.....Auto power OFF management
Unit......Select the measurement unit to mT or G
CLr....... For the further calibration usuage, not available
for the end user
8
Page 11
5-1 Auto power OFF management
When the lower display show " PoFF "
1)Use the " REC Button " ( 3-4, Fig. 1 ) to enter
will enter into the setting of " Auto power off management ". Press the " Mode Button " ( 3-6, Fig. 1 ) once in sequence to select upper value to " yES " or " no ".
yES - Auto Power Off management will enable. no - Auto Power Off management will disable.
2)
After select the desiring function ( YES or no ), press the " REC Button " ( 3-4, Fig. 1 ) to save the function with default.
5-2 Select the measurement unit to mT or G
When the lower display show " unit "
1)Use the " REC Button " ( 3-4, Fig. 1 ) to enter
will enter into the setting of " Select the measurement unit of mT or G " Press the " Mode Button " ( 3-6, Fig. 1 ) once in sequence to select the Display text to " t " or " g ".
t - milli Tesla ( mT ) G - Gauss ( G )
2)
After Display unit is selected to " mT " or " G ", press the " REC Button " ( 3-4, Fig. 1 ) will save the setting function with default.
9
Page 12
5-3 For the further calibration usage, not available for the user
When the lower display show " CLr " This setting just for the further calibration usage, not
available for the end user.
6. RS232 PC SERIAL INTERFACE
The instrument has RS232 PC serial interface via a 3.5 mm terminal ( 3-12, Fig. 1 ). The data output is a 16 digit stream which can be utilized for user's specific application. A RS232 lead with the following connection will be required to link the instrument with the PC serial port.
Meter PC
(3.5 mm jack plug) (9W 'D" Connector)
Center Pin..........................Pin 4
Ground/shield......................
.
Pin 2
2.2 K resister
Pin 5
The 16 digits data stream will be displayed in the following format :
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
10
Page 13
Each digit indicates the following status :
D15 Start Word = 02 D14 4 D13 1 D12 & Annunciator for Display D11 mT = E3 G = B5 D10 Polarity
0 = Positive 1 = Negative
D9 Decimal Point(DP), position from right to the
left, 0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D8 to D1 Display reading, D8 = MSD, D1 = LSD
For example :
If the display reading is 1234, then D8 to D1 is : 00001234
D0 End Word = 0D
RS232 setting
Baud rate 9600 Parity No parity Data bit no. 8 Data bits Stop bit 1 Stop bit
7. BATTERY REPLACEMENT
1)When the left corner of LCD display show " ", it
is necessary to replace the battery. However, in-spec. measurement may still be made for several hours after low battery indicator appears before the instrument become inaccurate.
2)Loose the " Battery Cover Screw " ( 3-18, Fig. 1 ), slide
the " Battery Cover " ( 3-17, Fig. 1 ) away from the instrument and remove the battery. Replace with DC 9 V battery ( 006P Alkaline/heavy duty ), and reinstate the cover.
3)Make sure the battery cover is secured after changing
the battery.
11
Page 14
8. SYSTEM RESET
If the meter happen the troubles such as :
CPU system is hold ( for example, the key button can not be operated... ).
Then make the system RESET will fix the problem. The system RESET procedures will be either following method :
During the power on, use a pin to press the " Reset Button " ( 3-13, Fig. 1 ) once a while will rest the circuit system, After execute the " System reset " the setting value of : Advanced Setting " will be cleared and return to default value.
12