MRC L3201LED User Manual

11
1) While the push rod to transmitted or reflected is to the working
position “R”, fluorescent microscopy can be used.
2) Reduce the field diaphragm, If it is not centered, move it to the
center using the special tool. Adjustment can be done by opening the
3) The push rod to direct image to binocular eyepieces or camera port
to the upper left of the trinocular head will send 100% of light to the
binocular eyepieces or to the top port. It satisfies high-grape
photography, see figure 7.
Preventative Maintenance
1. Cleaning frame and stage
Disconnect the plug from mains socket before cleaning. Clean the frame
and stage with a soft cloth moistened with a mild detergent solution. Be
sure the instrument is dry before using.
2. Cleaning optical parts
Microscope eyepieces and objectives are coated. They should not be
wiped while dry as dirt or dust may scratch the coating. It is best to
remove parts from the frame prior to cleaning. Always blow loose dust
away first. Use cotton swabs or lens tissue moistened with a lens cleaner
or a small amount of alcohol, then wipe the surface clean with a good
quality lens tissue. Solvents such as Xylene should NOT be used as
cleaner.
3. Cleaning 100X oil immersion lens
The immersion oil should be removed from the lens at the end of each
workday using cotton swabs or lens tissue moistened with a lens cleaner
or a small amount of alcohol.
LED Epifluorescence Microscope
L3201LED
User Guide
HAGAVISH 3,HOLON, ISRAEL
TEL: 972-3-5595252, FAX. 972-3-5594529
mrc@mrclab.com
1
Congratulations and Thank You for your purchase of
The Microscope
This precision instrument has been designed to require a minimum of optical and mechanical maintenance. Its excellent design assures years of high quality, reliable service. We recommend you read this entire manual carefully before beginning to use the instrument.
WARNING
PLEASE DISCONNECT PLUG FROM MAINS SOCKET BEFORE
REMOVING THE BOTTOM PLATE, OPENING THE LAMP
REPLACEMENT DOOR, OR REMOVING THE LAMP HOUSE
WARNING
THE POWER CORD PROVIDED WITH THE EQUIPMENT HAS A
GROUNDED PLUG. ALWAYS USE THE POWER CORD WITH A
PROPERLY GROUNDED WALL OUTLET.
DO NOT EXPOSE THE INSTRUMENT TO HIGH TEMPERATURES OR
HUMIDITY. AVOID USING THE INSTRUMENT IN EXTREMELY DUSTY
LOCATIONS
OPERATING TEMPERATURE 5 DEGREE C. TO 35 DEGREE C.
OPERATING HUMIDITY: 20% TO 80% @ 25 DEGREE C.
CAUTION: NEVER IMMERSE THE INSTRUMENT IN WATER OR
SOLVENT
CAUTION: DO NOT PUT ANY FOREIGN OBJECTS IN THE FRAME OR
INTO ANY MOVING MECHANICAL PARTS
10
swing the attachment lens to optical center.
6. Condenser Alignment – to center the Abbe condenser into the optical
light path, see figure 6:
1) Close the field iris diaphragm.
2) Observe the image using the 5X
objective and you will see an
image similar to figure 6 A. The
image may be out of focus and out
of center.
3) To center the field iris diaphragm in the field of view, use the two
centering screws at each side of the Abbe condenser mounting ring, see
figure 6 B.
4) Once the image is in focus and centered in the field of view, open the
field iris diaphragm until the image is almost as large as the field of
view, see figure 6 C.
7. Fluorescent Illumination Alignment, see figure 7.
A B C
Figure 6
4
1 2
3
1. Field Diaphragm Adjustment Knob 2. Camera Port 3. Push Rod
To Transmitted or Reflected 4. Push Rod To Direct Image to
Binocular Eyepieces or Camera Port Compensator 5. Field
Dia
p
hragm Centering Screw
5
Loading...
+ 4 hidden pages