MRC INE-DDBJ-350 Instruction Manual

Page 1
INE-DDBJ-350
Portable Conductivity Meter
Instruction Manual
PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com
MRC.12.16
Page 2
TIPS
Please read this instruction in detai
l before operating the
meter.
The meter must be sent to metrological servic
e or other
qualifi
ed departments to recheck after being used for one year, and only those qualified ones can be used again
The expiration date of
the electrode should see
Electrode Instructi
on Manual. After the expiry date, whether used or not, its performance will be affected, should be replaced timely.
When the meter is not used for a long time, connect the AC adapter and then switch on.
Do not use pow
er adapters that do not meet the
requirements of our meter.
The meter has a variety
of low power consumption functions,reduce to use these function will let the meter stand long time.
-- LCD backlight;
-- Use automatic backli
ght function;
-- Use automatic shutdown function during operation.
Page 3
CONTENT
1. Installation
1.1 Unpacking
1.2 Structure
1.3 Installation
1.3.1 The installation of measuring probe
1.3.2 The installation of the power charger
1.3.3 The installation of the USB cable
1.3.4 Daily use
2. Operation
2.1 Introduction
2.1.1 Terms
2.1.2 Features
2.1.3 Technical specifications
2.1.4 Operation
2.1.5 Operation diagram
2.2 Measurement parameters
2.3 Start-up, shutdown and buttons
2.4 Initial interface
2.5 meter operation
2.5.1 System setup
2.6 conductive probe calibration
2.6.1 Several methods of conductive probe calibration
2.6.2 Before calibration
2.6.3 Check calibration parameters
2.6.4 Start calibration
2.6.5 Calibration results report
2.7 Measurement
2.7.1 Preparation
2.7.2 Check measurement parameters
2.7.3 Check reading mode
2.7.4 Check calibration result
2.7.5 Check temperature options
2.7.6 Check sample, alarm setting
2.7.7 Start measurement
2.7.8 End measuring, result report
2.8 Data center
2.8.1 View conductive probe calibration information
2.8.2 View stored result
2.8.2.1 View settings
2.8.2.2 View result
2.8.2.3 Result report
3. Maintenance
3.1 Maintenance
3.2 Common troubleshooting
4. Appendix Ordering information
Page 4
1 INSTALLATION
1.1 Unpacking
In the packing box, users can find below parts:
1. INE-DDBJ-350 Portable conductivity meter 1set
2. Accessories 1set
1.2 Structure
1.2.1 Front view
Front View Of The Meter (Figure1)
The meter is composed of main unit and electrode system, both are indispensable. The meter supports conductive probe and ATC probe, allows the measurement of conductivity, TDS, resistivity, salinity value and temperature value.
1.2.2 Rear view
Rear Panel Of The Meter (Figure2)
Page 5
1.2.3 Accessories
Accessories Diagram (Figure 3)
1. The conductive probe
2. The ATC probe
3. The charger
4. The USB cable
1.3 Installation
Open the meter package, take out INE-DDBJ-350 portable conductivity meter and related accessories.
1.3.1 The installation of measuring electrode
1.3.1.1 The installation of conductive probe
Find measuring electrode socket (8 pins) above the meter (Mini type socket,
pay attention to the number of pins, don’t mistake with temperature socket), insert the conductive probe (pay attention to the direction).
1.3.1.2 The installation of temperature electrode
Find temperature electrode socket (6 pins) above the meter (Mini type
socket), insert the ATC probe (pay attention to the direction).
1.3.2 The installation of power charger
The meter uses lithium batteries(Li-on), supplied with a charger.
To connect the USB cable and the charger, and then insert the other end of the charger into the AC power. This charger is suitable for the following power supply:
Input: 100~240VAC, 0.15A, 50/60Hz; Output: 5VDC, 1A.
CAUTION:
1 This charger is only used to this meter, it is not recommended to use it on other kinds
of meters, also other kinds of charger is not recommended.
Page 6
2 When the meter is not used for a long time, the li-on may be low battery because of
standing long time and unable operation, user need to connect the AC adapter and then switch on.
3 When connected to PC, the meter will automatically start charging function.
1.3.3 The installation of USB cable
Users need to connect the USB cable if use our company's data acquisition
software.
Connect the meter and PC with the USB cable, then run the data acquisition
software on
PC to realize communication.
CAUTION:
The meter uses USB communication, to install the USB driver on the PC before use.
1.3.4 Daily use
For daily use, the electrode socket should be kept dry, clean.
Relevant electrode maintenance please reference to Electrode Operation
Instruction.
2 OPERATION
2.1 Introduction
Before using the meter, make sure that you have read the meter installation guide and
operation
guide. During operation, pay attention to as following:
Non-original power adapter may cause unnecessary safety problems. Prevent corrosive gas invasion. The interface of the meter must be kept clean, dry, avoid contact with acid, alkali,
salt solution.
After test samples, the electrode should be immersed in DI water. If the operating
interval more than 6 hours or for long-term storage, wash the electrode and store it in a empty protective bottle.
2.1.1 Terms
Cell Constant: The cell constant of a pair of platinum sheets is the ratio of distance
between the two sheets to the area of each sheet. The ratio is measured in cm
-1
.
TDS factor: The Total Dissolved Solids(TDS) factor is the factor to convert
conductivity and TDS.
Temperature Coefficient: The temperature coefficient is the % change of
conductivity with temperature in %/.
Page 7
2.1.2 Features
INE-DDBJ-350 portable conductivity meter is a new design meter with following features: It adopts new microprocessor technology, 4.3 "TFT touch screen, reach 65k color
gamut;
It adopts li-on to stand long time for operation; It adopts new UI design specification, navigational operation and sliding operations; It supports the measurement of conductivity, resistivity, TDS, temperature and
salinity value;
It supports balance condition judgment, balance conditions prompt. Supports alarm
settings, support users to customize the alarm value;
It supports multiple reading methods, including continuous reading method, smart
reading method and timed reading method, for long-timing or interval timing;
It supports to store measuring results of conductivity, TDS, salinity, resistivity, each
for 1000 sets;
It supports multiple view methods that user views stored data according to storage
number and storage time, the results will display in curve;
It supports conductive probe calibration function, calibrate Cell constant or TDS
factor;
It supports multi-points conductivity calibration, up to 2 points; It supports automatic identification of conductivity standard solution, default to four
kinds of GB standard solution;
It supports a variety of conductivity compensation mode, including pure water
compensation;
it supports automatic temperature compensation, automatic calibration, automatic
range and automatic frequency switch functions in the full range;
It supports GLP norms, to set the user ID, sample ID; It supports firmware upgrade function; It supports USB can be connected with PC; It supports the function of power-off protection. When the meter is to be turned off or
automatic shutoff, the stored measuring data, calibration data and setting parameters will not be lost;
It supports power consumption management, including automatic backlight
adjustment, automatic power off. In the set time, the meter will automatically reduce the backlight if there is no operation; If there is no further operation, the meter will automatically power off to save power consumption.
2.1.3 Technical Specifications
1. Measuring range
Conductivity: 0.010 μS/cm~1000 mS/cm; Resistivity: 5.00.cm~100.0 M.cm; TDS: 0.01 mg/L~300 g/L; Salinity: (0.00~8.00)%;
Temperature: (-5.0~105.0).
2. Resolution
Conductivity: automatic grade;
Page 8
Resistivity: automatic grade; TDS: automatic grade; Salinity: 0.01%; Temperature: 0.1.
3. Error of the meter
Conductivity: ±1.0%FS Resistivity: ±1.0%FS TDS: ±1.0%FS Salinity: ±0.2% Temperature: ±0.2.
4. Normal operating conditions:
Ambient temperature: (0.0~40.0); Relative humidity: No more than 85%; Power supply: li-on battery; No vibration that could affect performance exists; Non-corrosive gas in the surrounding air; No strong electromagnetic interference to surrounding except the geomagnetic
field.
5. Outside dimensions: L×W×H, mm: 201* 96 * 55
6. Weight(kg)About 0.3
2.1.4 Operation
The meter using the touch screen as the operation to control equipment, the meter is
equipped with a stylus.
By default, blue highlighted item means selected option, light color means invalid
option. Click highlighted option can perform the corresponding operations.
There are two input modes, character input and digital input. The title of the input
window is input message prompt, above is the input window, below is the button.
Press "Cancel" key to cancel the input; Press the "Enter" key to confirm the input; Press "Bsp" means backspace, that is to remove an input character.
Meter Button Input Diagram (Figure 4)
Page 9
Total Operation Diagram (Sheet 5)
2.2 Measurement parameters
The meter supports GLP norms, each result contains a complete set of measurement
parameters. Including several parts, such as prompt parameters, conductive probe calibration parameters, measure control parameters, sample parameters, measuring results parameters, etc. Following is a detailed description. Measurement parameters: it is the current measurement parameter types that
is the main measurement parameters. The meter displays in the window, take the main measurement parameters, auxiliary parameters, temperature data as a group of display unit. Main measurement parameters including the conductivity, TDS, salinity, resistivity, their corresponding auxiliary parameters respectively are resistivity, conductivity, conductivity and conductivity.
Setting user ID
Basic information Display
current time
User ID
Measuring signal monitoring Shows current
result
Shows last
calibration result
Power on, self-test
Enter into the initial interface. The initial interface has three
parts: Basic information, measuring signal monitoring, system
System menu System
settings
View stored
results
Click the measure windows to start measurement
Set system time, backlight, auto power off, touch screen, etc
View saved measurement results in different ways
Click the calibration result to calibrate directly
2.1.5 Operation chart
Page 10
Reading method: it is the reading method of current measurement parameters.
The meter supports three reading methods, continuous reading method, smart reading methods and timed reading method. Continuous reading method: This reading method is suitable for continuous
monitoring of the sample data, observing the change trend of sample. It requires user to manually terminate the measurement.
Smart reading method: Set the balance condition first. When the measurement
started, the meter will automatic measure, calculate and display the measured value, once meet the balance condition, this measurement is finised finished. Balance conditions includes balance time and balance value. When all
the measured values meets the demand of balance value at the set balance time. It means satisfies balance conditions, this measurement is finished.
The meter provide four balance conditions to select, "Fast, Medium,
Strict, Custom", different parameters have different balance conditions. Fast, medium, and strict are three kinds of balance conditions provided by the meter, the user cannot modify. User is only allowed to modify custom type balance conditions.
Note: User needs to manually save the measurement result.
Timed reading method: The meter supports two timing reading methods, one is
fixed time reading method, another is fixed interval time reading method. Fixed time reading method: When the measurement started, the meter works until the set time, for example set the time as 300 seconds, the meter will continuously measure for 300 seconds.
Note: Fixed time method does not automatically save measuring results.
Fixed interval time reading method: User needs to set measurement interval time and measurement quantity, then the meter will automatically collect and record data according to the set interval, the minimum time interval is 1 second.
Calibration result, it is the last calibration results of used electrode, is also the
electrode parameter of current measurement parameters to be used.
Corrector, it is the calibration operator last time.  Correction time, it is the last calibration time.  Temperature compensation coefficient, this parameter is corresponding to the
conductivity, TDS and salinity measurement parameters. When user needs to measure in high accuracy, the influence of the temperature will lead to inaccurate conductivity measurement, set temperature compensation coefficient, the default value is 0.020, that is 2.00% / .
Page 11
Set Cell constant, this parameter is corresponding to conductivity and salinity
measurement parameters. Usually there are two ways to get Cell constant: Calibrate the conductive probe with standard conductivity solution, or set the Cell constant value directly.
Set TDS factor, this parameter is corresponding to the TDS measurements.
Usually there are two ways to get TDS : Calibrate the conductive probe with standard conductivity solution, or set the TDS factor directly .
Conductivity compensation, this parameter is corresponding to conductivity
measurement, the meter supports three kinds of compensation mode, including no compensation mode, linear compensation mode, pure water compensation mode.
No compensation: At this point, the measurement reading is absolutely
conductivity value.
Linear compensation mode: The meter will automatically compensate to
the conductivity corresponding to the reference temperature.
Pure water compensation mode: this compensation method is
corresponding to the measurement of pure water.
Conductivity reference temperature, this parameter is corresponding to the
conductivity measurement. The meter supports two reference temperature: 20.0  and 25.0 , the default is 25.0 .
Salinity compensation mode, this parameter is corresponding to the salinity
measurement.
Note: This meter does not support seawater mode!
Balance condition, the meter provides four balance conditions for, as "Fast,
medium, strict, custom", which are corresponding to different parameters and the measurement environment.
Balance time, this parameter is corresponding to balance condition, its a
parameter of balance condition, it means the minimum waiting time to meet the data stability condition. During this time, measurements data should all meet the balance value, the meter will wait all the time once do not meet. The longer the time is, the more strict, and vice versa.
Balance value, this parameter is corresponding to the balance condition, its a
parameter of balance condition. It is the degree of data can be accepted, the greater the number is, the more easier to balance, and vice versa. Different measurement parameters have different values.
Note: users can only modify the custom of Balance time, Balance values.
Temperature display, the meter supports () Celsius or Fahrenheit ()
temperature display.
Temperature compensation mode, each measurement parameter supports
independent temperature compensation, when user selects a manual temperature compensation, the meter calculate
with the manual temperature value set by the user; When selects automatic compensation, the meter calculate with the temperature collected from ATC probe. If no ATC probe is connected with the meter, the meter will still use the manual setting temperature.
Manual temperature, use manual temperature when automatic temperature
compensation is invalid.
Page 12
Alarm switch, the meter supports alarm function. When the measured value
exceeds the alarm value, the meter will alert to remind users, the default display numbers turn to yellow. If user doesn’t need this function, close the alarm switch.
High alarm limit, this parameter is corresponding to the alarm function, to set the
upper alarm limit.When the alarm function is enabled, if the measurement results exceeding the maximum limit alarm, will alert to remind the user.
Low alarm limit, this parameter is corresponding to alarm function, to set the
alarm limit.
2.3 Start-up, shutdown and buttons
The meter built-in power supply, press the round Button(Figure 5)under the screen to turn on the meter.
After start-up, wait after complete self-inspection, enters into the initial interface.
If the battery is too low, may not be able to start-up, at this time, connect the charger
to start-up, and recharge in time.
After operation, user can long press On-Off key more than 5 seconds to close the
meter. If user enables automatic shutdown function, the meter will automatically power off if no operation in a set time.
During operation, user can open or close the LCD display at any time by short
pressing On-Off key.
There is a power light inside On-Off key. Flashing in the state of power on or charging,
the flashing number represents the electric quantity; maximum four times, means the battery is fully charged. When in shutdown state, the indicator light should be off in a few second. Charging in shutdown state, the light will normally on if fully charged.
2.4 Initial state
After start up and the meter enters into initial state after self-test. There are three parts:
basic information, measuring signal monitoring and the system menu.
Basic information interface displays the current time and the login user ID;
Round Button (Figure 5)
Sample ID, its the name of the sample, maximum 10 characters.
Page 13
Measuring signal monitoring interface displays the current measurement signal
and the last conductive probe calibration results; Click measurement window can start formal measurement, click the ">" button on the right to switch to view other parameters; Click ">" button in the last correction result window to calibrate electrode directly .
The system menu interface includes system settings and view stored results.
All the operation began from this interface, and will return to here after the completion of corresponding operations. We referred to as the initial state, click on the arrow keys or sliding around can switch interface, as shown in figure 6:
Initial State Of The Meter (Figure 6)
Page 14
2.5 Operation
2.5.1 System settings
System settings includes function modules, such as system time, buzzer, automatic backlight, automatic shutdown, user ID management, touch screen, language selection, about, etc. In the initial state, select menu item "System Settings" to enter, as shown in figure 7.
System Setup Diagram (Figure 7)
S
ystem time: to record the meter execution time, such as operation time,
storage time, conductive probe calibration time and so on. If the deviation is larger, user need to adjust according to the local time, time format is year/month/days hour: minute: second;
Buzzer: Set the prompt way of buzzer, including whether buzz or not when
the button is valid, and prompt the user whether buzz or not when perform to the special circumstances.
Automatic backlight: Automatic backlight includes two parts, automatic
backlight control and brightness.
Automatic backlight control: The meter has automatic backlight adjustment function that reduce power consumption. If no any operation in the set time (just click on the touch screen), the meter will automatically reduce the backlight brightness, show black screen state, click on the touch screen at any time to wake the screen. Automatic backlight control including 6 options: off, 10sec, 20sec, 30sec, 60 sec, 120 sec, respectively means backlight always on, does not close; no any key operation within 20 seconds, 30 seconds, 10 seconds, 60 seconds and 120 seconds, the backlight will automatically shut down (enter power saving mode, there is still a fainter display. In power saving mode, press the screen any place, will be back to show to the set backlight value);
Page 15
Brightness control is for the purpose of saving power consumption,
the meter also allows user to set the brightness of the backlight. The higher brightness, the greater the power consumption, but showed the best effect.
Automatic shutdown: The meter has automatic shutdown function. There
are ten options: off, 1 min, 2 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 60 min; If open the automatic shutdown function, the meter will directly turn off within the setting time.
User ID management: this function only sets the user ID.Touch screen: Including two functions, touch screen calibration and touch
screen test.
Touch screen calibration: to calibrate touch screen if the touch
screen buttons position deviation is larger. The meter adopts five point calibration method, to continuously click on more than five times at five locations on the touch screen to finish the calibration. Since the calibration results will directly affect your operation, once enter into the calibration state, follow steps carefully. Store the calibration data after calibration.
Touch screen test: To verify the results of calibration, the meter also
provides a test module. After entering test mode, user clicks on the screen at any position according to your usage habits to check the accuracy, including the intensity of the click, size of the location and the click effect of using stylus or finger.
Restore defaults: The meter supports restore factory defaults function.
Click restore factory default to restore if the meter parameters is confusion. Restore the factory default as follows:
1. User ID is set to Admini;
2. Buzzer prompt is all on;
3. Automatic backlight is off;
4. Automatic shutdown is off;
5. The storage of measurement results are all cleared;
6. Reset all measured parameters to factory default. Cell constant is 1.000.
About: Displays the version of the meter information and company
information, etc. The meter supports firmware upgrade function.
2.6 conductive probe calibration
2.6.1 Several methods of conductive probe calibration
The meter provides a variety of ways to calibrate Cell constant:
In the measurement signal monitoring interface, click the last correction
result to re-calibrate;
Before measurement start, re-calibrate when checking the calibration results
;
The meter supports edit and select function of calibration results, allows users to save last correction data to current calibration and reedit, so the meter will prompt the user whether to keep or delete the last calibration results or not before starting the calibration, as shown
Page 16
in figure. If user wants to restart correction, choose to delete, or else choose to keep. For example, the last calibration use 146μS/cm and 1408μS/cm to complete two point correction, but user only wants to re-calibrate point 1408μS/cm, then can choose to keep current calibration data.
Keep Last Calib Data Diagram (Figure 8)
The meter
adopts the navigational direction calibration, users complete the
correction in accordance with the instructions step by step.
2.6.2 Before calibration
The conductive probe will have some drift under different service environment or unused for a long time, and need to re-calibrate with the conductivity standard solution, the meter is up to 2 point calibration. The meter supports automatic identification of conductivity standard solution,
the default four standard solution. Standard data obtained by potassium chloride solution, the nominal concentration respectively is 1 mol/L , 0.1 mol/L, 0.01 mol/L, 0.001 mol/L. There are usually two methods to get conductivity standard solution: Self­preparation, purchase directly.
First, prepare conductivity standard solution by themselves. To prepare standard
solution refer to the following data:
KCl Standard solution for testing Cell constant
Cell constant(cm-1)
0.01 0.1 1 10
KCL solution concentration(mol/L)
0.001 0.01 0.01 or 0.1 0.1 or 1
Page 17
Relation of KCl solution concentration and conductivity value (unitμS/cm)
The composition of standard solution
Second, choose and buy conductivity standard solution. One is to purchase
directly standard conductivity standard solution produced by professional manufacturers. The other way is to buy standard solution from our company. There are four kinds of standard conductivity solution produced by ourselves.
Prepare 1~2 kinds of conductivity standard solution, DI water according to
requirements, prepare get ready, click "Next".
Prepare Before Cell Constant Calibration Diagram (Figure 9)
Temp
Concentration (mol/L)
15.0 18.0 20.0 25.0 35.0
1 92120 97800 101700 111310 131100
0.1 10455 11163 11644 12852 15353
0.01 1141.4 1220.0 1273.7 1408.3 1687.6
0.001 118.5 126.7 132.2 146.5 176.5
Concentration (mol/L)
Volume concentration KCl(g/L)solution (20in air)
1 74.2457
0.1 7.4365
0.01 0.7440
0.001 Dissolve 100mL 0.01mol/L of solution to 1L
Page 18
2.6.3 Check calibration parameters
The meter supports automatic identification function of the conductivity solution, the default four JJG standard solution, standard data using potassium chloride solution, nominal concentration respectively is 1 mol/L , 0.1 mol/L, 0.01 mol/L,
0.001 mol/L. If user adopts non-standard solution for the meter, set the recognition type as "Manual identification", and manually enter the nominal value.
Check Measurement Parameter Diagram (Figure 10)
2.6.4 Start calibration
Specific calibration process is as follows:
Prepare 1~2 kinds of standard solution according to the need (Options);
Insert the conductive probe and ATC probe into corresponding measuring
electrode socket; clean the probes with DI water and put in the standard solution;
If it is automatic recognition, the meter will automatically identify the solution,
and display the standard value; If the recognition is failure, check whether the ATC probe is properly connected, whether the conductive probe is damaged, or whether the shelf life of standard solution is over.
If it is manual identification, to manually input the nominal value.
If the solution cannot be identified or no input nominal value, this calibration
can not continue.
After reading is stable, press the "Calib" key, the meter store calibration
results and switch to report page, shows the calibration results;
If user need to continue to calibrate other standard solution, click "Continue
Calibration" button to return to calibration status. Clean the probes and put in another standard buffer. After successful recognition by the meter, and the reading is stable, press
Page 19
"Calibrate" button to complete the correction.
The meter supports repeating calibrate of the same standard solution, that is
support the function of matching the same solution, if the current correction solution differ 50uS/cm with previous standard solution, the meter consider to be the same solution, will automatically cover the calibration data of previous standard solution. This function can prevent repeated correction, or allow user to calibrate at will. If necessary, user will need to reset some correction parameters to complete correction. Click "Setup" button or sliding around to switch pages.
End correction, press the "<" button to exit the correction, shown in Figure
11.
Calibration Diagram (Figure 11)
2.6.5 Calibration results report
After each end of the calibration, the meter switch to the calibration results pages,
report page indicates the current calibration points, calibration result and the detailed data of calibration solution.
The meter allows at any time to switch to the calibration results page to view the calibration results. If there is a previous calibration data, the meter allows to delete certain calibration results. At this time, click the "<" button to end correction, shown in Figure 12.
Page 20
Calibration Result Diagram (Figure 12)
2.7 Measurement
2.7.1 Preparation
Before measurement, user should understand better the nature and character of
measured sample; the basic operation and application of the meter; understand the use and maintenance of conventional electrodes.
To prepare conductive probe before measurement, if need calibration, then purchase
or self prepare. Following are some preparation and matters need attention for conductive probe.
Clean:Thoroughly clean conductive probe with DI water, and then clean
according to the conductive probe maintenance parts. If continues error or unstable, refer to the conductive probe maintenance section.
Check samples: If conductive probe work normal in standard solution but
not operation properly in the sample, check the presence of interference substances and whether the
conductance cell is damaged or not.
Platinum black series conductive probe were adhered loose platinum black
layer on the surface of the conductive probe platinum that avoid touching any object. Only use DI water to wash, otherwise it will damage the platinum black layer, lead to inaccurate electrode measurement.
If find performance of platinum black series conductive probe reducing, to
rinse the platinum in anhydrous ethanol and DI water successively.
The loose platinum black layer on the surface of platinum conductive probe
may be
Page 21
adsorbed sample components, must be washed after measurement.
Placed in a period of time or use after period of time, the constant of
conductive probe may change and need to be re-calibrated.
The meter provides two ways to start measurement. One is to Start measurement
according to the last configuration, the other is to view or modify before measurement. Usually, if user is familiar with the whole measuring process and sample condition, it is suggested to Start measurement according to last configuration; On the contrary, according to clew, step by step to check the relevant measurement parameters and then start measuring, shown in Figure 13.
Prepare Before Measurement Diagram (Figure 13)
1.7.2 Check measurement parameters
The meter supports the measurement of conductivity, TDS, salinity, resistivity and
temperature.
The meter is always displayed in window form, the window also shows three
parameters: one primary parameters, one auxiliary parameter and one
temperature value. User set measurement parameters, that is to select the
main parameter types, when the main parameters respectively are the
conductivity, TDS, salinity, resistivity, the auxiliary parameter respectively is
resistivity, conductivity, conductivity, conductivity. Temperature value will
always be displayed, as shown in Figure 14.
Page 22
Check Measurement Parameter Diagram (Figure 14)
2.7.3 Check reading mode
Check Reading Mode Diagram (Figure 15)
The navigation page includes the current measuring parameters, reading method,
conductivity compensation method, conductivity reference temperature, data
stability condition, etc., related parameters varies with reading method, shown
in Figure 15.
Page 23
2.7.4 Check calibration result
This navigation page shows the last conductive probe calibration result, this data is
also the electrode parameters to be used in current measurement.
If user needs to re-calibrate, then double-click the last calibration results data
area, confirm to the re-calibrate probes, the meter guide the user to complete
calibration, detailed calibration please refer to electrode adjustment section, as
shown in Figure 16.
Users can also set Cell constant value directly.
Check Calibration Data Diagram (Figure 16)
Page 24
2.7.5 Check temperature options
This navigation page sets temperature display type, temperature compensation method.
Check Temp Option diagram (Figure 17)
2.7.6 Check sample, alarm setting
This navigation page sets the measuring sample ID, alarm option.
If necessary, set the alarm switch, high limit alarm and low limit alarm, when
measuring results exceeded the set alarm value, the meter will express with
visible mark, yellow digital display in order to prompt the user by default, as
shown in Figure 18.
Check Sample ID Diagram (Figure 18)
Page 25
2.7.7 Start measurement
After all of the above operation is completed, click on "Next" to start formal
measurement, display as shown in figure 19.
Measurement Display Diagram (Figure 19)
In the figure, above is the measurement window, also shows the results of the measuring parameters and measuring results of related auxiliary parameters.
In the figure, below is the last time conductive probe calibration results, when the
measurement is finished, it shows the result.
The meter automatically measures values, calculates and displays the results according to the set measurement parameters and reading methods.
The meter supports data stability judgment, synchronous prompt measurement condition, convenient for user to view.
When the reading method is smart measurement mode, the meter monitors the balance condition, once balanced, it will automatically store measurement results and end the measurement.
When the reading method is fixed interval timing measurement mode, the meter monitors the timing measurement conditions, once satisfied, automatically store the measurement results; and then repeated until reach the set number of measurement, the measurement end.
When the reading method is a fixed time timing measurement mode, the meter continuously monitoring, when reach to set time, the measurement is finished (do not store measuring results).
Page 26
When the reading mode is continuous measurement mode, the meter always monitoring the whole measurement process, until user manually end the measurement. The meter allows users to store measurement results manually. During the measurement process, press the "End" key to terminate at any time.
2.7.8 End measuring, result report
When the measuring is finished or terminated, the meter provides a simple report, including the measurement parameters, the way of reading and balance condition and the quantity of storage results, as shown in Figure 20.
At this time, choose to remeasure, or end the measurement etc.
Measurement Result Diagram (Figure 20)
2.8 Data center
The meter supports the view of conductive probe calibration information and stored
measuring result.
2.8.1 View conductive probe
calibration information
The meter supports the view of detailed conductive probe calibration information,
including the number of the last calibration, calibration results and detailed
records of each point, as shown in Figure 21.
These information is available in the initial state, measuring state and check the
calibration results, etc.
Page 27
Check Calibration Data Diagram (Figure 21)
2.8.2 View stored result
In the initial state, click the menu item "View stored results" to enter.
2.8.2.1 View settings
The meter supports multiple view modes, view stored data according to store
numbers and storage time. User sets view modes according to the actual
needs, then press "Start search"; the meter will search results according to
search conditions and displays in curve.
Check View Option Diagram (Figure 22)
Page 28
2.8.2.2 View result
The meter shows the matching results that meet the consulting conditions in curve,
as shown in figure. The figure includes view content, current view method,
view scope, the number of matching results, the beginning number of current
curve, the end number of current curve, etc. Buttons are show on the right side
of the curve.
For clean display, the meter displays up to 25 data at a time, user click on the
"Page up", "Page down" flip to see the rest of the data. Click "<<" and ">>" to
move indicative lines and view the corresponding results, as shown in Figure
23.
View Result Diagram (Figure 23)
2.8.2.3 Result report
The meter detail records measured parameters related to each measurement
results, the report including information such as the operator, measurement
results, balance condition, temperature compensation method, conductivity
compensation method, conductivity reference temperature, sample ID and
reading method, etc.
Under this interface, allowing users to directly press "<" and ">" button to view
contiguous data, as shown in Figure 24.
Page 29
Result Report Diagram (Figure 24)
Page 30
3.1 Maintenance
Correct use and maintenance can ensure the normal and stable operation of the meter
for a long time. In order to ensure insulation, the meter sockets must be kept clean, dry, avoid
contact with acid, alkali, salt solution and prevent moisture.
After used in environmental humidity is higher, wipe dry the electrode plug a
clean cloth.
To save the electrode in dry place.
3.2 Troubleshooting
Phenomenon Cause of the problem Troubleshooting
Boot does not display
1. The power is too low;
2.The meter damage.
3. The LCD backlight is off
1. Connect the charger and turn on.
2.Replace or repair according to regulation.
3. Close the automatic backlight function, click on the touch screen or press the key to open the LCD
backlight The conductivity measurement is not correct
1. The electrode performance is bad
2. Use incorrect way of compensation
3.The conductive probe calibration error
4. The calibration of the conductivity standard solution has problem.
1. Replace the electrodes
2. Set the correct compensation
mode
3. Re-calibrate electrode
4. To buy or prepare standard
solution again
4 APPENDIX
Product ordering information
Product name and type Specifications
INE-DDBJ-350 Portable Conductivity Meter
A new generation of conductivity meter, meet the measurement requirements of portable conductivity, TDS, salinity, resistivity, temperature, measuring range: 1.0 grade, measurement range:
0.000 mu S/cm ~ 1000 ms/cm; Accuracy: ± 1.0% (FS);
DJS-1-L conductive probe Suitable for measuring the conductivity, TDS, salinity, resistivity
T-818-L ATC probe
Suitable for measuring temperature
Conductivity standard solution Four kinds of GB conductivity standard solution
3 MAINTENANCE
Page 31
Packing List
INE-DDBJ-350
Portable Conductivity Meter
Number Description Quantity
1 INE-DDBJ-350 Portable Conductivity Meter 1
2 DJS-1-L Conductive probe 1
3 T-818-L ATC probe 1
4
USB Power Supply(5V/1A
1
5
Mini USB Line
1
6
Stylus
1
7
Manual
1
Loading...