MRC AM-4214SD Operating Manual

TABLE OF CONTENTS
1. FEATURES.................................................................................
2. SPECIFICATIONS........................................................................
3. FRONT PANEL DESCRIPTION..................................................................
3-1 Display........................................................................................
3-2 Power Button ( ESC, Backlight Button ).........................................................................................
3-3 Hold Button ( Function Button, Next Button ).........................................................................................
3-4 REC Button ( Enter Button, Unit Button ).........................................................................................
3-5 SET Button (
Button, Time check Button )..............................................................................
3-6 Logger Button (
Button, Sampling time check Button )........................................................
3-7 Probe input socket..................................................................
3-8 Type K/J Probe Input Socket................................................
3-9 SD card socket....................................................................................
3-10 RS-232 Output Terminal......................................................
3-11 Reset Button........................................................................
3-12 DC 9V Power Adapter Input Socket.............................................
3-13 Battery Compartment/Cover.................................................
3-14 Battery Cover's Screws.........................................................
3-15 Stand........................................................................................
3-16 Tripod Fix Nut............................................................................
3-17 Probe Plug....................................................................................
3-18 Telescope Probe.......................................................................
3-19 Sensing Head...............................................................................
4. MEASURING PROCEDURE....................................................................
4-1 Function selection ....................................................................
4-2 Air velocity/Temp. measurement..............................................
4-3 Type K/J thermometer ...................................................................
10
4-4 Data Hold...............................................................................................
11
4-5 Data Record ( Max./ Min. reading )....................................................................
11
4-6 LCD Backlight ON/OFF....................................................................
12
5. DATALOGGER................................................................................
12
5-1 Preparation before execute datalogger function...............................................................................
12
5-2 Auto Datalogger ( Set sampling time
1 second ).......................
13
5-3 Manual Datalogger ( Set sampling time = 0 second ).....................
14
5-4 Check time information...............................................................................
14
5-5 Check sampling time information...............................................................................
15
5-6 SD Card Data structure.......................................................................
15
6. Saving data from the SD card to the computer....................................
16
7. ADVANCED SETTING...............................................................................
17
7-1 Set clock time ( Year/Month/Date, Hour/Minute/ Second )..........
18
7-2 Decimal point of SD card setting.....................................................................
19
7-3 Auto power OFF management ...........................................................
19
7-4 Set beep Sound ON/OFF.................................................................
20
7-5 Select the thermometer type to Type K or Type J.................................................................
20
7-6 Select the temperature unit to or ...............................................................
21
7-7 Set sampling time .........................................................................
21
7-8 SD memory card format....................................................................................................
21
8. POWER SUPPLY from DC ADAPTER.....................................................
22
9. BATTERY REPLACEMENT.............................................................
22
10. SYSTEM RESET.................................................................................
23
11. RS232 PC serial interface.................................................................................
23
12. Optional Type K Temp. probe.................................................................................
25
13. PATENT.....................................................................................
26
1. FEATURES
* Combination of hot wire and standard thermistor,
deliver rapid and precise measurements even at low
air velocity value. * Slim probe, ideal for grilles & diffusers. * Air velocity : m/s, Ft/min, Km/h, Knot, Mile/h, * Air temperature ( , )℃ ℉ * Air Temp. used thermistor sensor, fast response time. * Type K, Type J thermocouple thermometer. * Real time SD memory card Datalogger, it Built-in Clock
and Calendar, real time data recorder , sampling time set
from 1 second to 3600 seconds. * Manual datalogger is available ( set the sampling
time to 0 ), during execute the manual datalogger
function, it can set the different position ( location ) No.
( position 1 to position 99 ). * Innovation and easy operation, computer is not need
to setup extra software, after execute datalogger, just
take away the SD card from the meter and plug in the
SD card into the computer, it can down load the all the
measured value with the time information (
year/month/date/ hour/minute/second ) to the Excel
directly, then user can make the further data or graphic
analysis by themselves. * SD card capacity : 1 GB to 16 GB. * LCD with green light backlight, easy reading. * Can default auto power off or manual power off. * Data hold, record max. and min. reading. * Microcomputer circuit, high accuracy. * Power by UM3/AA ( 1.5 V ) x 6 batteries or DC 9V adapter. * RS232/USB PC COMPUTER interface. * Separate probe, easy for operation of different
measurement environment.
1
* Applications : Environmental testing, Air conveyors,
Flow hoods, Clean rooms, Air velocity, Air balancing,
Fans/motors/blowers, Furnace velocity, Refrigerated case,
Paint spray booths.
2. SPECIFICATIONS
2-1 General Specifications
Circuit Custom one-chip of microprocessor LSI
circuit.
Display LCD size : 52 mm x 38 mm
LCD with green backlight ( ON/OFF ).
Measurement
Air velocity:
Unit m/s (meters per second)
Km/h ( kilometers per hour ) Ft/min ( FPM, feet per minute ) Knots ( nautical miles per hour ) Mile/h ( mph, miles per hour )
Air temperature:
, ℃ ℉
Type K/ Type J thermometer.
, ℃ ℉
Sensor
Air velocity & Air flow :
Structure Tiny glass bead thermistor.
Air temperature :
Thermistor.
Type K, Type J thermometer :
Type K/J thermocouple probe.
* Probes are optional.
2
Datalogger Auto 1 second to 3600 seconds Sampling Time
@ Sampling time can set to 1 second,
Setting range
but memory data may loss.
Manual Push the data logger button
once will save data one time.
@ Set the sampling time to
0 second.
@ Manual mode, can also select the
1 to 99 position ( Location ) no.
Memory Card SD memory card. 1 GB to 16 GB.
*
It recommend use memory card 4 GB.
Advanced * Set clock time ( Year/Month/Date, setting Hour/Minute/ Second )
* Decimal point of SD card setting * Auto power OFF management * Set beep Sound ON/OFF * Set thermometer type to Type K or Type J * Set temperature unit to or * Set sampling time * SD memory card Format
Temperature Automatic temp. compensation for the Compensation Anemometer function and the type K/J
thermometer. Data Hold Freeze the display reading. Memory Recall Maximum & Minimum value. Sampling Time Approx. 1 second. of Display Data Output RS 232/USB PC computer interface.
* Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
* Connect the optional USB cable
USB-01 will get the USB plug.
Operating 0 to 50 . Temperature Operating Less than 85% R.H. Humidity
3
Power Supply
* AC/DC power adapter is optional.
Alkaline or heavy duty DC 1.5 V battery ( UM3, AA ) x 6 PCs, or equivalent.
* AC/DC power adapter is optional.
DC 9V adapter input. ( AC/DC power adapter is optional ).
Power Current Normal operation ( w/o SD card save
data and LCD Backlight is OFF) :
Approx. DC 30 mA.
When SD card save the data and LCD
Backlight is OFF) :
Approx. DC 50 mA.
Weight 515 g/ 1.13 LB. Dimension
Main instrument :
203 x 76 x 38 mm
Telescope Probe :
Round, 12 mm Dia x 280 mm ( min. length ). Round, 12 mm Dia x 940 mm ( max. length ).
Accessories *
Instruction manual..............................
1 PC
Included *
Telescope Probe...................................................
1 PC
*
Hard carrying case ( CA-06 )..................................
1 PC Optional SD Card ( 1 G ) Accessories SD Card ( 2 G )
Type K thermocouple probe. AC to DC 9V adapter. USB cable, USB-01. RS232 cable, UPCB-02. Data Acquisition software, SW-U801-WIN.
4
2-2 Electrical Specifications (23± 5 )
Air velocity
Measurement
Range Resolution Accuracy
m/s 0.2 to 5.0 m/s 0.01 m/s ± ( 5% + a )
5.1 to 25.0 m/s 0.1 m/s reading
Km/h
0.70 to 18.00 km/h
0.01 Km/h
18.0 to 72.0 km/h 0.1 Km/h or
Mile/h 0.50 to 11.20 mph 0.01 mph ± ( 1% + a )
( mph ) 11.2 to 44.7 mph 0.1 mph full scale
Knot 0.40 to 9.70 knot 0.01 Knot
9.7 to 38.8 knot 0.1 Knot Ft/min 40-3940 ft/min 1 Ft/min @ a = 0.1 m/s, 0.3 km/h, 0.2 mile/h, 0.2 knot, 20 ft/min
Note:
m/s - meters per second km/h - kilometers per hour ft/min - feet per minute knot - nautical miles per hour mile/h - miles per hour (international knot)
Air temperature
Measuring Range 0 to 50 /32 to 122 Resolution 0.1 /0.1 Accuracy ± 0.8 /1.5
Type K/J thermometer
Sensor Resolution Range Accuracy Type Type K 0.1 -50.0 to 1300.0 ± ( 0.4 % + 0.5 )
-50.1 to -100.0 ± ( 0.4 % + 1 )
0.1 -58.0 to 2372.0 ± ( 0.4 % + 1 )
-58.1 to -148.0 ± ( 0.4 % + 1.8 )
Type J 0.1 -50.0 to 1100.0 ± ( 0.4 % + 0.5 )
-50.1 to -100.0 ± ( 0.4 % + 1 )
0.1 -58.0 to 2012.0 ± ( 0.4 % + 1 )
-58.1 to -148.0 ± ( 0.4 % + 1.8 )
@
Above specification tests under the environment RF Field Strength
less than 3 V/M & frequency less than 30 MHz only.
5
3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display
3-2 Power Button ( ESC, Backlight Button )
3-3 Hold Button ( Function Button, Next Button )
3-4 REC Button ( Enter Button, Unit Button )
3-5 SET Button (
Button, Time check Button )
3-6 Logger Button (
Button, Sampling time check Button )
3-7 Probe input socket
3-8 Type K/J thermometer socket
3-9 SD card socket
3-10 RS-232 Output Terminal
3-11 Reset Button
3-12 DC 9V Power Adapter Input Socket
3-13 Battery Compartment/Cover
3-14 Battery Cover Screws
3-15 Stand
3-16 Tripod Fix Nut
3-17 Probe Plug
3-18 Telescope Probe
3-19 Sensing Head
6
4. MEASURING PROCEDURE
4-1 Function selection
1)Turn on the meter by pressing the " Power Button "
( 3-2, Fig. 1 ) momentarily.
* Pressing the " Power Button " ( 3-2, Fig. 1 )
continuously and > 2 seconds again will turn off the
meter.
2)The meter can select 2 kind function as :
a. Air velocity/Temp. d. Type K/J Thermometer
Pressing the " Function Button " ( 3-3, Fig. 1 ) continuously ( not release the button ), the Display will show the following text in sequence :
Air Air velocity/Temp. measurement tP Type K/J Thermometer measurement
Until the Display show the desired mode ( Function ), just release the " Function Button " ( 3-3, Fig. 1 ), the meter will execute this function with default.
7