MP Biomedicals FastPrep-24, 6004-500 Instruction Manual

www.mpbio.com
MP Biomedicals • 29525 Fountain Parkway • Solon, OH 44139 • tel: 1.800.854.0530 • fax: 1.800.334.6999
One Call
One Source
A World of
Biotechnology
Reagents
Precautionary Instructions
The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definition of these symbols are as follows;
NOTE :- Throughout this manual "NOTE" may be found. These Notes are helpful information to aid in the particular area of function being described.
Text with a "CAUTION" indicator will explain possible Safety infractions that could have the potential to cause minor to moderate injury or damage to equipment.
Text with a "WARNING" indicator will explain possible Safety infractions that will potentially cause serious injury and equipment damage.
Read, understand and practice the precautionary an operating instructions. Know the limitations and hazards associated with using this unit. Observe the precautionary and operational decals placed on the unit.
DO NOT operate this unit in an environment where other devices are being used that intentionally radiate electromagnetic energy in an unshielded manner.
DO NOT use sharp objects such as a pencil point or ballpoint pen to operate the buttons on the control panel as damage may result.
This unit should be operated in temperatures between 15º C to 35º C, transported and stored in temperatures between 5º C to 43º C, with Relative Humidity ranging from 30%-60%.
DO NOT disassemble, modify, or remodel the unit or accessories.
and instead
This device should be kept out of the reach of children.
WARNING
=CAUTION
=WARNING
CAUTION
®
The FastPrep -24 is CE in compliance with the European
Low Voltage and EMC Directives
www.mpbio.com
LISTED
(LABORATORY EQUIPMENT)
3RRV
U
L
Model #6004-500
www.mpbio.com
Table of Contents
Section I Introduction ..........................................................................................5
Section II Installation ...........................................................................................6
2.1 Unpacking............................................................................................... 6
2.2 Inspection............................................................................................... 6
2.3 Set-up..................................................................................................... 7
2.4 Connecting the Power............................................................................. 7
Section III Operation............................................................................................8
3.0 Controls & Functions .............................................................................. 8
3.1 Loading & Securing the Samples.............................................................. 9
®
3.2 Preparing the FastPrep -24 for Operation.............................................. 12
®
3.3 Programming the FastPrep -24.............................................................. 12
3.4 Save Programs...................................................................................... 13
Section IV Specifications....................................................................................15
Section V Warranty and Liability..........................................................................16
Section VI Appendices......................................................................................18
2.1 APPENDIX 1 Maintenance & Cleaning......................................................... 18
®
2.2 APPENDIX 2 An Explanation of FastPrep -24 Instrument Speed Settings......21
®
2.3 APPENDIX 3 FastPrep -24 Product Line.......................................................22
®
2.4 APPENDIX 4 Accessories and Spare Parts for FastPrep -24......................... 30
www.mpbio.com
5
I. INTRODUCTION
®
Different parts of FastPrep -24 are shown below: (Fig 1 & 2)
The FastPrep -24 Instrument is a high-speed benchtop homogenizer offering the ultimate in speed and performance for the lysis of biological samples. Indeed, simultaneous homogenization of 24 samples in 2ml tubes,12 samples in 15ml tubes ,48 samples in 2ml tubes or 2 samples in 50ml tubes takes place within 40 seconds.
The FastPrep -24 Instrument uses a unique, optimized motion to disrupt cells through the multidirectional, simultaneous beating of specialized Lysing Matrix beads on the sample material.
Developed for difficult and resistant samples, the FastPrep -24 Instrument lyses thoroughly and quickly any tissues and cells and thus allows easy and reproducible isolation of stable RNA, active proteins and full-length genomic DNA. Samples and buffers are added to a Lysing Matrix tube containing specialized Lysing Matrix beads. The ergonomic design of the instrument ensures an easy loading of the sample tubes that remains securely sealed during the processing. The homogenization speed and duration times are digitally controlled. After setting your speed and time with the touch of a button, just push "run", and in less than a minute your samples are completely lysed.
Dome
Sample Tubes Holder
Sample Tubes
Right Side Wings
Bowl
Securing Knob (ratchet)
Spoke Plate
Micro Switch (under Dome)
Dome Clip (Dome Fastener)
Enclosure
Front Panel
Emergency Switch
Fig. 1 (Front view)
www.mpbio.com
®
®
®
II INSTALLATION
2.1 Unpacking
Carefully remove the FastPrep -24 Instrument and accessories from the shipping carton.
The list of items included with quantity are given below.
1. AC Power Cord - 1
2. Springs - 8
3. Fuse 10 Amp - 2
4. Screw Driver - 1
5. Instruction Manual - 1
Compare the packing list to the box contents. If there is a discrepancy. call MP Biomedicals.
CAUTION: Do not lift the device by cover.
2.2 Inspection
Inspect the unit for any damage that may have occurred during shipment. Should there be any damage, report it to the carrier and contact MP Biomedicals immediately. Save the packaging material in the event a return is necessary.
2.3 Set-up
To assure safe operation and best result, read this manual before operating the
®
®
FastPrep -24 Instrument. The FastPrep -24 Instrument comes fully assembled, requiring very little set-up. Install the system on a clean, dry-stable surface within 4 feet (1.2m) of a compatible electrical outlet.
Ventilation
®
Allow 1-2 inches (3-5 cm) of space around he FastPrep -24 Instrument for proper ventilation. This unit is "FOR INDOOR USE ONLY". Avoid operating in areas of excessive humidity or extremes of temperature.
6
www.mpbio.com
®
2.4 Connecting the Power
®
The FastPrep 24 can work on 110 VAC/60 Hz or 220 VAC/50 Hz. Make sure the rocker switch located on the rear panel is OFF when connecting the power. Connect the power cord to the Instrument (power fixture is at the back side) and plug it in to a compatible outlet.
This symbolizes Alternating Current
III OPERATION
3.0 CONTROLS & FUNCTIONS
Functions of control panel keys are given below (Ref. Fig. 2)
1. SET Key - Press this key to select Speed & Time as displayed on LCD
Speed: Selectable from 4.0 m/s
to 6.5 m/s. In steps of 0.5 m/s (Default 4.0 m/s).
Tube holder: Selectable
Time: Selectable from I sec. to 60 sec. in steps of I second. (Default 20 seconds)
2. RUN Key - To Start or Stop the Instrument.
3. Key - To Increase the selected value of speed, tube holder & time.
4. Key - To Decrease the selected value of speed ,tube holder & time.
Fig. 2 (Control Panel)
and to choose program
(model 6004)
list of
tube holders
7
www.mpbio.com
FastPrep®24
RUN
SPEED
TIME
SET
1
2
3
4
LCD
-
3.1 LOADING & SECURING THE SAMPLES
1. Loosen the securing knob (Ratchet Nut) by rotating counter-clockwise until it come out (Ref. Fig. 3)
2. Remove the securing knob (ratchet Nut).
Fig. 3
3. Take out the assembly of spoke plate and test tube holder (
Fig. 4
Ref Fig. 6)
9
www.mpbio.com
Securing Knob (Ratchet Nut)
Spoke Plate
Three Step
Spoke Plate
Tube Holder
4. Sightly uplift the spoke plate and rotate clockwise so that retention spokes move away from holes to leave them open for loading (Ref. Fig. 5).
Fig. 5
5. Load sample tubes into the cavities of the sample holder, so that they fit snugly in the holes. It is preferred to position the tubes symmetrically.
6. When all the sample tubes are loaded, place this assembly into the
®
FastPrep -24 Instrument. Align the locking pin of Aluminium three steps with the holes under the Tube Holder, ensuring its proper placement (Ref. Fig. 6).
Fig. 6
10
Retention Spoke Plate
Sample Holder
www.mpbio.com
Locking Pin
Hole Under the Tube (for matching)
Aluminium Three Step
7. Now uplift and rotate the spoke plate counter-clockwise so that the retention spokes take the position above each sample tube cap. Place the securing knob (ratchet nut) on this assembly to tighten it completely (Ref. Fig. 7).
Fig. 7
8. Close the dome properly and secure the dome clip.
WARNING: Sample tubes must be secured properly before running the
®
FastPrep -24 Instrument.
11
Sample Tube
Retention Spoke Plate
www.mpbio.com
12
®
www.mpbio.com
3.2 PREPARING THE FastPrep 24 FOR OPERATION
1. Turn the main Rocker Switch ON (
®
2. Factory-set default values automatically program FastPrep -24 to operate at speed 4.0 m/sec, Tube holder MP:24*2 and run time 20 seconds (to override these values. See programming)
This symbolizes On and Off supply
3.3 PROGRAMMING THE FastPrep 24
Make sure Emergency Switch is pulled out).
User can take default setting or his own. Five different combination of setting can be saved as described in next section.
can be changed at this moment. Press p or
q key to change the default Speed to User Selectable Speed. Once speed is selected,
press SET again, flashing of 4.0M/S will stop to indicate that speed has been selected.
Switch ON Power Supply using Rocker Switch Provided at the Back, Make sure
Emergency Switch is pulled out.
On Power ON the LCD will Show WELCOME followed by MP Bio
Press SET key to start programming
Once SET key is pressed, LCD will show 4.0 M/S which means it's the default Speed, If Option 1 is needed than press SET key again, the 4.0 M/S message on LCD will start flashing to indicate that it
-
®
-