Please ensure both yourself and any helpers read these instructions carefully before using your bath
lift.
This bath lift is designed and manufactured to assist individuals in getting in and out of the bath. It should not be used for any other
purpose. Do not use it in combination with other lifting or moving devices. If you are unable to get out of the bath unaided, do not
bathe without someone on call to help in an emergency.
Contents
Page
Unpacking the bath lift
1. Unpacking the bath lift 2
Before you start
2. Charging the battery 2
3. If you have a telecare-enabled bath lift 3
4. Preparing the bath for installation 3
5. Installing the bath lift 4
Using the bath lift
6. Connecting the hand control 5
7. Using the bath lift 6
Removing and storing the bath lift
8. When the bath lift is not in use 7
9. Removing the bath lift from the bath 7
10. Cleaning 8
Warranties, maintenance & servicing
11. Maintenance and servicing 8
12. Warranty 8
Symbols
13. Symbols 10
Technical data
14. Technical data 11
Troubleshooting
15. Troubleshooting
Parts list
16. Accessories 12
17. Parts list 13
1
Operating Instructions
1. Unpacking the bath lift
Open the packaging and ensure the bath lift is not damaged. If it has been damaged
in transit, contact your supplier immediately.
Inside the box you will nd:
• A bath lift, complete with two side aps and hand control attached. (see g 1)
• Side ap protectors x 2 (see g 1)
• A battery (see g 2)
• Battery charger and charger cradle attached (see g 2)
We recommend that the bath lift is always separated and carried as two lighter,
individual sections. Refer to section 5 “Installing the bath lift” for details on how to
move the bath lift.
2. Charging the battery
Side ap protectors
Hand control
Front suction feet
Charger cradle
Side ap
Figure 1
Tracks
Charger
Before rst operation of your bath lift or after long periods of non-use
(typically 7-10 days), fully charge the battery as follows. A at battery will take
up to 2 hours to charge.
i ) Slide the battery into the track on the charger cradle until it clicks into place (see
g 2).
ii ) Plug the charger into the mains socket outside the bathroom and switch on the
electricity supply.
iii) The light indicator on the charger box will initially ash in a quick sequence of
red, green, red, green, before a constant red light appears to indicate charging is in
progress (see g 2).
iv ) When the battery has been fully charged, the red light turns to a constant green light,
which indicates the battery is ready for use, and can remain continuously on charge
until you are ready to take your bath.
v ) When you are ready to use your fully charged battery (see paragraph iv), ensure it is
removed from the charger cradle and that the mains socket is switched o, or that the
charger is removed from the mains socket. (This prevents the battery discharging and
will reset the charger ready for its next use.)
Battery
Light indicator
Figure 2
The battery can be charged after each bath without the risk of overcharging.
NOTE:
•
Do not charge in the bathroom or any humid area. Always charge in a dry area, ensuring your hands are dry rst.
•
Do not cover the battery or charger whilst charging.
•
Always charge on a hard surface.
•
Ensure that no metal objects touch the contacts on the battery, as this will damage the battery.
•
Use only the Mountway charger supplied to charge the battery and for no other purpose.
2
Operating Instructions
3. If you have a telecare-enabled bath lift
The telecare-enabled bath lift has been designed to integrate with the Tunstall
Lifeline 400, 4000+ and Connect+ units. The telecare-enabled bath lift can be added
to an existing monitoring solution or as part of a new installation to suit individual
needs
If you have a telecare-enabled bath lift, you will nd a telecare-enabled battery inside
your bath lift box rather than the standard battery (see g 3).
Programming the telecare-enabled bath lift to Tunstall 400, 4000+ and
Connect+ home units
i ) Press and hold down the green cancel key located on the Lifeline unit (see g
4) until it bleeps (approx. 5 seconds), then release the green cancel key and the
LifeLine unit will bleep twice. The red alarm button will ash slowly.
ii ) Press the green cancel key again until the LifeLine unit bleeps, then releasethe
cancel key and the LifeLine unit will bleep twice. The red alarm button will now
ash rapidly.
iii) Press the red button on the end of your bath lift’s telecare-enabled battery (see
g 3) and a red light will illuminate at the end of the battery. The Lifeline unit will
bleep to conrm acceptance.
iv) Press and release the green cancel key on the Lifeline unit and the unit will bleep.
Telecare-enabled battery
Red light
Red button
Lifeline Connect+
Figure 3
Red alarm
button
Green
cancel key
v) Make a test call using the programmed telecare-enabled battery (see g 4) by
pressing its red button and ensure it raises an alarm call.
Making an alarm call
i ) To raise an alarm while using the bath lift, press the red button on the end of the
battery (see g 3). This will allow you to talk to the control centre operator ‘hands
free’ using the powerful speaker and microphone of the Lifeline 400, 4000+ and
Connect+.
ii ) The control centre operator will summon the appropriate assistance from your
carer, warden or emergency services.
NOTE:
An alarm call cannot be raised unless a Tunstall Lifeline 400, 4000+ or Connect+
home unit is installed in your home and the telecare-enabled battery is programmed
correctly to this home unit (see section “Programming the telecare-enabled bath lift to
Tunstall Lifeline 400, 4000+ and Connect+ home units” above).
Figure 4
3
Operating Instructions
4. Preparing the bath for installation
i ) Remove any non-slip mats from the bath.
ii ) Ensure the bottom of the bath is clean and smooth, with no oil or bath cleaner
residues that will prevent the suction feet from sticking.
iii ) The suction feet may not be suitable in baths that have been re-enamelled or
that have rippled or textured surfaces. For some baths with rippled or textured
surfaces, the front suction feet (see g 7) can be relocated to the middle of the
bath lift base frame, to clear the bath’s textured area. A centre sucker xture kit is
available, if required (see section 16).
iv ) It is essential that the side aps do not catch under any obstructions such as bath
handles, handrails or soap dishes, when rising. If your bath has handles, t the
side ap protectors to them (see g 1 and g 5) to prevent the side aps from
catching.
5. Installing the bath lift
i ) If your bath lift is supplied with optional soft or cushioned covers, t them at this
stage (see cover tting instructions).
Side ap protector
Backrest assembly
Seat unit
Figure 5
Backrest
Actuator
Figure 6
Seat
ii ) Place the backrest assembly carefully on the oor when not attached to the seat
unit. The actuator forms a useful handgrip for transporting (see g 6).
iii) Pick up the seat unit (see g 7). The rear of the base frame forms a useful
handhold for this purpose.
NOTE: Never use the sucker release handle (see g 7) to carry the seat unit.
iv ) Place the seat unit in the bottom of the bath and as far back as possible at the
opposite end to the taps. Ensure there is enough clearance at the rear of the
bath for the backrest to fully recline without resting on the bath or wall.
v ) Press rmly down on the seat unit, ensuring all four suction feet are stuck to the bath
surface. All suction feet swivel and can stick to the contours of the bath as required.
vi ) Before tting the backrest assembly to the seat unit, ensure the actuator end is
fully retracted (see g 8).
vii ) Place the backrest assembly face down onto the seat unit, whilst aligning the
hinges at the bottom of the backrest with the slots located at the rear of the seat
unit (see g 9).
Sucker
release
handle
ü
X INCORRECT
Rear of
base frame
Backrest
CORRECT
Base frame
Actuator end fully retracted
Actuator not fully retracted
Front
suction
feet
Figure 7
Actuator
Figure 8
viii) Ensure the backrest hinges locate into BOTH slots, with the bottom edge of the
backrest positioned in front of the two blocks (see g 9).
ix ) Hold the top of the backrest and lift upwards until the backrest rests in the fully
reclined position. Check the hinges are still located in both seat unit slots.
Hinge
Seat
unit
slot
Seat unit blocks
Seat
unit
slot
Hinge
Figure 9
4
Operating Instructions
NOTE:
•
If either backrest hinge has detached from the slots during assembly, the blocks
at the rear of the seat unit should guide them back into position.
•
If the backrest is forced OVER either block, it will not be assembled correctly and
the bath lift SHOULD NOT be used. Follow steps v to ix again if this occurs.
x ) Unfold the two side aps so they rest against the sides of the bath. (Side aps
ease transfer on and o the bath lift and also prevent nger entrapment.)
xi ) Fill the bath with water to the required level and check the temperature before
using the bath lift.
xii) The battery can now be connected to the hand control on your bath lift (see
section 6).
NOTE:
Do not connect the battery to the hand control until the backrest is fully
assembled to the seat unit.
6. Connecting the hand control
NOTE:
If the battery has not been used before or has not been used for a long period
(typically 7-10 days), ensure you place it on charge before using the bath lift.
Slide the fully charged battery into the tracks on the hand control (see g 10) until it
Battery
Battery status
indicator
Tracks
Battery
Hand control
Figure 10
Hand control
Figure 11
clicks into place (see g 11).
Always check the battery status indicator on the hand control before using the bath
lift (see g 12). Press the UP button and observe the following guidelines:
•
If GREEN light, proceed with your bath, provided the battery has been charged
within the last 7-10 days.
•
If FLASHING GREEN LIGHT, proceed with your bath, provided the battery has
been charged within the last 7-10 days. Then charge immediately afterwards
(see section 2).
•
If RED light, do NOT use the bath lift. Charge IMMEDIATELY (see section 2).
•
If in any doubt, recharge the battery before use (see section 2).
NOTE:
The battery is for use solely with Mountway bath lifts and cannot be used to
operate any other product.
5
Hand control
Up
Down
Figure 12
Operating Instructions
7. Using the bath lift
i ) To raise the bath lift, press the UP button on the hand control (see g 12). If the side aps are caught, immediately lower the
bath lift by pressing the DOWN button on the hand control. Pull each side ap free from the obstruction, then resume the lifting
operation.
ii ) As the bath lift reaches the top of the bath, the side aps will fold out to form a transfer platform.
iii) When the bath lift’s seat and side ap are level with and supported by the top of the bath, sit on the side ap and move onto the
seat. Ensure no part of the body can become trapped when transferring onto the seat.
iv) Turn as you lift each leg over the side of the bath then make yourself comfortable on the seat. You may nd it easier to transfer if
you wet the seat and the side aps rst. A slide and swivel aid is available, should you nd transferring dicult (see section 16).
v) Before lowering the bath lift, extend your feet out in front of you and keep hands above the seat to avoid any risk of entrapment.
vi) To lower yourself to the bottom of the bath, press the DOWN button on the hand control (see g 12). When the button is
released, the bath lift will stop and remain at that position until the button is pressed again.
vii) When you have reached the bottom of the bath, continue to press the DOWN button to begin the backrest reclining. Release
the DOWN button when you nd your desired recline position. Maintain pressure on the DOWN button to stop the backrest at
its fully reclined position. To reduce the backrest angle, press the UP button.
viii) Whilst bathing, use the suckers on the battery to attach the hand control and battery to the edge of the bath or to a tiled wall.
If the hand control and battery fall into the water, there is NO danger as both are waterproof. Ensure you wipe the
battery dry before charging.
ix) When nished bathing, press the UP button to raise the bath lift. The backrest will move into the upright position rst, then
the bath lift will begin to rise. When the button is released, the bath lift will stop and remain at that position until the button is
pressed again.
x) Once the side aps are level with the top of the bath, remove your nger from the UP button and move sideways onto the side
ap.
xi) Turn whilst lifting each leg out of the bath.
xii) Drain the bath.
NOTE:
• If the RED light appears on the hand control whilst lowering, the bath lift will not go down any further. If this happens, press
the UP button and the bath lift will raise you to the top of the bath. Charge the battery immediately.
• In the event of an emergency, sliding the battery out of the hand control can instantly terminate all bath lift movement.
• When the bath lift reaches its maximum limit of travel (up or down), there will be a short delay before it will operate again.
• The bath lift is not designed for continuous operation. Allow at least 15 minutes between baths.
• The bath lift should not be used under a constant ow of water (e.g. running tap) and ensure the actuator is not totally
submerged in water.
6
Operating Instructions
8. When the bath lift is not in use
i ) The bath lift may be stored in its lowest position at the bottom of the bath with
the backrest fully reclined. If the bath lift is stored in the bath, leave the battery
connected to the hand control unless it is being charged.
ii ) The bath lift may be removed from the bath for storage (see section 9 below).
9. Removing the bath lift from the bath
i ) Ensure the bath lift is in its lowest position with the backrest fully reclined.
ii ) Remove the battery from the hand control.
iii) Fold both side aps onto the seat.
iv) Fold the backrest forward until it lies at on the seat.
v ) Wind the hand control cable around the cable winders on the rear of the backrest,
then attach to the retainer clip (see g 13). This will prevent potential damage to
the hand control caused by the trailing cable.
Cable
winder
Cable winder
with retainer clip
Hand
control
Figure13
vi) Hold the actuator and lift the backrest assembly upwards and clear of the seat
unit and place the backrest carefully on the oor out of the way.
vii) Pull the sucker release handle attached to the rear suction feet at the back of the
seat unit and the rear feet will be released (see g 14).
viii) Now hold the rear of the base frame, NOT the sucker release handle, and tilt the
seat unit forward until the front suction feet automatically release.
ix ) Firmly grip the base frame by the rear, and lift out of the bath.
x ) The bath lift may be stored with the backrest folded at onto the seat unit (see g
15). The backrest will not be locked to the seat in this position and care should be
taken whilst handling.
NOTE:
•
Do not attach the battery to the hand control when the backrest is folded at.
•
Only operate the bath lift when the backrest is assembled and locked to the
seat unit.
Rear of base frame
Sucker release
handle
Figure14
Flat-folded bath lift
Figure15
7
Operating Instructions
10. Cleaning
Clean the bath lift as frequently as you clean your bath, using non-abrasive cleaners.* If used communally, we recommend
disinfecting the bath lift after each use. SOLVENT-based cleaners, such as white spirit, should NOT be used.
i ) To disinfect the bath lift, use a soft brush or cloth and wipe down with a sterilising solution or anti-bacterial cleaner. Rinse the
product with clean water, then wipe and dry using a clean cloth.
ii ) Bath/medicinal oils and other solutions are easily removed with washing-up liquid. Do not allow oils or other additives to build
up. They should be removed frequently otherwise, over the course of time, they may lead to degradation of the suction feet.
iii) The bath lift can be cleaned in an automated washing machine (to a maximum temperature of 73° C for 3 minutes), however the
battery MUST be removed and washed by hand. The battery must not be subjected to high temperature cleaning.
iv) Do not clean or rinse the actuator under a constant ow of water, e.g. running tap, or submerge it in water.
iv) If your bath lift is tted with comfort covers, ensure you remove them on a regular basis to clean the bath lift surfaces.
It is also advisable to clean the bath surface under the suction feet from time to time.
* Cleaning agents should be used in accordance with the manufacturer’s guidelines.
11. Maintenance and servicing
There are no user serviceable parts on the bath lift. Repairs and servicing must be entrusted to a Mountway authorised dealer.
The bath lift should be inspected annually by a Mountway authorised dealer, who will check that there are no loose ttings or any
other evidence of mechanical damage.
12. Warranty
This product is sold with the benet that, in the event of any defect of manufacture or material appearing on the frame within 5
years of the date of receipt of the product, the defect will be rectied free of charge, provided that:
i ) Reasonable evidence is provided (e.g. purchase invoice or delivery note) that the product was purchased not more than 5 years
prior to the date of claim; and
ii ) The defect is not attributable to accidental damage (either in transit or otherwise), fair wear and tear, willful damage, misuse or
unauthorised repair.
The battery is guaranteed for 2 years to power (when fully charged) one bathing cycle* at full load (22 stone/140kg/308lbs).
The actuator and consumables (e.g. suction feet and the charger) are also guaranteed for 2 years
Consumers’ statutory rights are not aected.
* The bathing cycle:
For one bath, it is assumed that the bath lift will be raised from the bottom to the height of the bath and then lowered into the
water supporting a person of up to 22 stone/140kg/308lbs. After bathing, the bath lift will raise the person to the height of the bath
then be lowered to the bottom of the bath for storage.
8
Operating Instructions
13. Symbols
To comply with international standards for medical equipment, the following symbols are found on the Aquila labels:
SYMBOL EXPLANATION
Only direct current from a small battery is used to power your bath lift. There is NO connection to mains
electricity when in use, so there is no risk of electric shock.
The bath lift is categorised as type B medical equipment and meets international standards for electrical
safety.
The bath lift is sealed to the international standard against the eects of water spray.
IP66
The battery stick is sealed to the international standard against the eects of immersion in water to a
IP67
depth of 1 metre.
The hand control is protected against long periods of immersion in water under pressure.
IP68
This symbol is found on the hand control and means the light represents the battery status.
PLEASE READ THE INSTRUCTIONS BEFORE INSTALLATION OR USE.
Do not bin/dispose.
9
14. Technical data
Operating Instructions
The bath lift seat is made of glass-lled polypropylene /
179mm (7in)
360mm (14.2in)
179mm (7in)860mm (33.9in)
nylon. The hand control and battery are made from ABS.
The hand control button cover is made of Thermoplastic
Elastomer.
The bath lift is in the ocial category of “internally
powered equipment” and is not suitable for use in the
presence of ammable anaesthetic mixture with air or
nitrous oxide.
The equipment is tested in accordance with:
• BS EN 60601-1 : 1990
• BS EN 10535
• The European Council Directive 89/336/EEC relating to Electromagnetic Compatibility.
• The European Council Directive 93/42/EEC relating to Medical Devices.
• The European Council Directive 73/23/EEC relating to Electrical Safety.
The bath lift complies with the WEEE directive.
475mm (18.7in)
676mm (26.6in)
DISPOSAL: The plastic, metal parts, packaging materials and suction feet are recyclable but the Nickel Metal Hydride (NiMH)
battery, should be disposed of only by returning it to your supplier, the manufacturer or an authorised disposal point.
Lifting capacity (person) 22 stone/140 kg/308lbs
Net weight
9.6 kg/21.16lbs
(excluding battery)
Seat base weight6kg/13.22lbs
Operating and charging
Temperatures
Max40°C (104°F)
Min0°C (32°F)
Backrest weight3.6kg/7.94lbs
Seat height UP440mm (17.3 in)
Seat height DOWN 75mm (2.9 in)
Backrest height680mm (26.77 in)
Power input14.4V dc nominal 7A max
Duty cycleOne bath per 15 minutes
Degree of protectionIP66
Angle of recline35°
Hand control (battery)
Input14.4V dc nominal 1.0A
Output14.4V dc nominal 7A max
Degree of protectionIP68
Battery stickIP67
Shipping and storage
Weight (packaged) 11.5 kg (25.4 lbs)
Shipping volume0.07 m³ (2.48ft³)
Temperatures
Max40°C (104 °F)
Min-10°C (14 °F)
Humidity
Max 90% RH
Min 10% RH
Pressure
Max125 kPa (18.4 psi)
Min 73 k Pa (10.7 psi)
Battery charger
Input100-240V ac 50-60Hz 1.0A
Output14.4V dc nominal 1.0A
This product is suitable for transport in
pressurised aircraft holds.
10
Operating Instructions
15. Troubleshooting
SYMPTOMPOSSIBLE CAUSESOLUTIONSECTION/PAGE
Bath lift does not move.
No sound from the motor.
• Battery is at.• Charge battery.
• Battery is not connected to
hand control.
• Not pressing correct area
on hand control buttons.
• Faulty electronics in hand
control.
• Insert battery correctly into both tracks on hand
control cradle and slide all the way to the end until
it clicks.
• Press arrowhead on either UP or DOWN buttons.
Continue to press.
• Contact supplier or authorised service/repair
centre.
• Battery fault.• Contact supplier or authorised service/repair
centre. Replace battery.
• Actuator fault.• Contact supplier or authorised service/repair
centre.
Section 2/page 2
Section 6/page 5
Section 6/page 5
n/a
n/a
n/a
Red light on hand control is
on and bath lift only moves
up.
Red light on hand control
is on but bath lift does not
move.
Bath lift only moves
intermittently.
• Battery is at or has not
been fully charged and
Stop Descent™ has been
activated to prevent the
bath lift lowering.
• Battery is at and has not
been charged for long
period (typically 7-10 days)
• Charge battery immediately.
• Charge battery immediately.
• Charge battery before use if bath lift has not been
used for long periods (typically 7-10 days).
• Do NOT start bath with RED light on hand control.
Charge immediately.
• Ensure battery is placed on charge as soon as you
nish your bath if FLASHING GREEN light appears
on hand control.
• When bath lift is not in use, leave battery
connected to hand control. Disconnect only for
charging or when moving bath lift.
• Battery fault• Contact supplier or authorised service/repair
centre.
• Battery is not fully charged.• Charge battery.
• Excessive user load.• Do not use bath lift if you weigh over
22st/140kg/308lbs.
• Bath lift has encountered
obstruction during lifting,
e.g. bath handles, soap
dish.
• Lower bath lift, pull aps free from obstruction,