Motic DSK, DSK500 Series Manual

Contents
1. Important 3
1.1 Getting Ready 3
1.2 Maintenance and Storage 4
1.3 Caution 4
3. Controls 6
3.1 Observation bridge mount 6
3.2 Power supply unit 6
3.3 Large Base 7 4 Operation 8
4.1 Focus Adjustment 8
4.1.1 Focusing at the primary observer Position 8
4.1.2 Focusing at the secondary observer Position 8
4.2 Using the Pointer 9
4.2.1 Pointer Brightness Adjustment 9
4.2.2 Pointer Displacement 9
4.3 Photomicrography Cautions 10
5. Specification 11
5.1 Operating environment 12
6.1 Assembly Diagram 13
6.2 Detailed Assembly Procedure 14
6.2.1 Mounting the Pillar 14
6.2.2 Mounting the Focusing Assembly 14
6.2.3 Mounting the Side by Side attachment 15
7. Mounting 16
7.1.1 Mounting the Microscope Body 16
7.1.2 Mounting the Observation Tubes 16
7.1.3 Mounting the Pointer Illumination 16
7.1.4 Connecting the Power Cord 17
Correct assembly and adjustments are very important for the microscope to manifest its full performance. If you want to assemble the microscope by yourself, see Chapter, “ASSEMBLY” first.
1. Important
The Motic DSK is a side-by-side dual viewing stereomicroscope system. As it allows two observers to sit side by side, it is ideal for education and training purposes. Note that there are few restrictions on the installation location of this attachment. The orientation of the images observed by the primary observer and the secondary observer is the same in both the vertical and horizontal directions.
1.1 Getting Ready
1. This manual pertains only to the DSK attachment. Before using this attachment in conjunction with the SK500/SK700 Stereo microscope and associated options, make sure that you have carefully read and understood the corresponding manuals, and that you understand how the various components of the microscopic system are used together.
2. The DSK attachment is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact.
3. Do not use the attachment anywhere where it may be exposed to direct sunlight, high temperature and humidity, dust, or vibrations. (For operating environment conditions, see Section, “SPECIFICATIONS”
4. Before replacing the pointer illumination bulb, be sure to set the main switch to “(OFF)”, unplug the power supply unit and other cords and wait until the bulb and its surroundings have fully cooled down.
5. be sure to use only the specified tungsten bulb when replacing the pointer illumination bulb.
Applicable bulb 6V1OWGE (mfd. by Hosobuchi Electric Lamp)
6. Do not plug the pointer illuminator cord into any unit except the Motic deliverd power supply unit.
7. Make sure this attachment is installed in a room where there is as little vibration as possible and that the work surface on which this attachment is installed is sturdy and level (with inclination within 5°). If vibration is still noticeable, use a anti-vibration pad.
8. Before placing a specimen which is sensitive to static electricity (such as a packaged circuit board) on the stage of the large base, place a conductive mat or similar object on the stage.
1.2 Maintenance and Storage
To clean the lenses and other glass components, simply blow dirty away using a commercially available blower and wipe gently using a piece of cleaning paper (or clean gauze).
If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute alcohol.  Since the absolute alcohol is highly flammable, it must be handled carefully.  Be sure to keep it away from open flames or potential sources of electrical sparks for
example, electrical equipment that is being switched on or off.
 Also remember to always use it only in a well-ventilated room.  The equipment uses plastic resins extensively in its external finish. Do not attempt to
use organic solvents to clean the non-optical components of the microscope. To clean these components, use a lint-free, soft cloth lightly moistened with a diluted neutral detergent.  Never disassemble any part of the microscope as this could result in malfunctions or reduced performance.  This equipment should be disposed of by following the rules and regulations of your national or local government.
1.3 Caution
 If the attachment is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the attachment may also be damaged. Always operate the equipment as outlined in this instruction manual.
2. Nomenclature
If you have not yet completed the assembly of the microscope yet, see Chapter, “ASSEMBLY” first.
DSK500
Fig.1
3. Controls
3.1 Observation bridge mount
Fig.2
3.2 Power supply unit
Fig.3a Fig.3b