Morris 57314 Instruction Manual

Page 1
1
Instruction Manual
2in1 LAN Tester & Multimeter
Model: 57314
Page 2
2
Contents
Introduction...............................................
Features...................................................
Safety Precautions………………………..
Meter Description.....................................
Electrical Specification.............................
Operation………………………………….. AutoRanging Mul timeter…………………. Multi-Netw ork Modular Cable Tester……
Introduction
2 in 1 LAN Tester & Multimeter is a innovative tester that allows the user to easily measure DC/AC Voltage/Current, Resistance, Continuity, diode and verify the cable continuity, open short, cross-connect. The included remote terminator allows the user to t est i nstall ed cabl e either at a wall ja ck or a patch panel add ing up t o v alu e and convenience. Proper u se and care of this meter wi ll provide years of reliable service.
Features
2in1- LAN Tester & Multimeter
Measures DC/AC Voltage, DC/AC Current, Resistance,
Continuity ,diode and LA N tester
3-1/2 digit (2000 count) LCD display for multimeter functions
LED Displays the actual pin configuration of 10BASE-T
AND 10BASE-2 Thin Ethernet, FJ45/RJ11 modular,
258A, TIA-568a/568b and Token Ring cables
Double Molded housin g
CATIII 600V; CATII 1000V
Provides easy to read continuity and fault status display
Checks for continuity, open wire, ground wire, short ed p air
and crossed pair faults
Page 3
3
Allows for remote testing of installed cables from wall jack or patch panel
auto or manual scanning for LA N tester
Autoranging with auto power off for multimeter functions
Safety International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.
Double insulatio n
Safety Precautions
1. Improper use of this meter can cause damage, shock,
injury or death. R e ad an d und er st a nd th is user’s manual before operating the meter.
2. Make sure any covers or battery doors are properly
closed and secured.
3. Always disconnect the test leads from any voltage
source before replacing the battery or fuses.
4. Do not exceed the maximum rated input limit s.
Input Limit s
Function
Maximum Input
V DC or V AC
600V DC/AC
µA AC/DC
200mA/250V fast acting Fuse
Resistance, Diode & Continuity Test
600V DC/AC
Page 4
4
5. Use great care when making measurements if the
voltages are greater than 25VAC rms or 35VDC. These voltages are considered a shock hazard.
6. Always discharge capacitors and remove power from
the device under test before performing Diode, Resistance or Continuity tests.
7. Remove the battery from the meter if the meter is to be
stored for long periods.
Meter Description
Page 5
5
LANtester part Description (1~9) :
1. Remote terminator with LED display for receiving end
2. Jack RJ45
3. LAN-connector
4. Jack RJ45
5. Jack RJ45
6. Led display for sourcing end (Jack 1)
& Led display for receiving end (Jack 2)
7. Test switch for auto scan
8. LAN tester Power switch
9. Test switch for manual scan
Digital multimeter part Description (10~17) :
10. Function switch
11. 3 1/2 Digit (2000 count)LCD display for DMM functions
12. MODE button
13. MAX Hold button
14. Data Hold button
15. COM input jack
16. V, Ω,uA,mA input jack
17. Battery Cover
Page 6
6
Specifications Electrical Specifications
Function
Range
Accuracy
DC Voltage 200mV,
±(0.5% rdg + 3d)
2.000V, 20.00V,
±(1.0% rdg + 3d)
200.0V, 600V
±(1.0% rdg + 3d)
AC Voltage 50-60Hz
2.000V, 20.00V
±(1.0% rdg + 5d)
200.0V, 600V
±(1.5% rdg + 10d)
DC Current 200.0µA,
2000µA
±(1.5% rdg + 3d)
20.00mA,
200.0mA
±(2.0% rdg + 3d)
AC Current 200.0µA,
2000µA
±(1.8% rdg + 8d)
20.00mA,
200.0mA
±(2.5% rdg + 8d)
Resistance
200.0 ±(0.8% rdg + 5d)
2.000k,
20.00k, 200.0k
±(1.2% rdg + 3d)
2.000M ±(2.0% rdg + 5d)
20.00M ±(5.0% rdg + 8d)
Page 7
7
Max input voltage 600V AC/DC Diode Test Test current 1mA max., open circuit
voltage of 1.5V typical
Continuity Check Audible signal if the resistance is <150 Display 2000 count 3 -1/2 digit LCD Over range indication
LCD displays “OL”
Polarity Minus (-) sign for negative polarity. Low Battery Indication “BAT” symbol
indicates low battery condition.
Input Impedance >7.5MΩ (VDC & VAC) AC Response Average responding
ACV Bandwidth 50Hz to60Hz Auto Power Off 15 minutes (approximately) Fuse mA, µA ranges; 0.2A/250V fast acting
Fuse
Batteries 9V battery and two “AAA” batteries Operating Temperature 32
o
F to 104oF (0oC to 40oC)
Storage Temperature 14oF to 122oF (-10oC to 50oC) Weight 308g Size 162x74.5x44.0mm Standard IEC61010-1 CAT III-600V Pollution degree II,
CE Approved
Operation
AC/DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure AC/ DC voltages if a motor
on the circuit i s being sw itched ON or OFF. Lar ge voltag e surges may occur that can damage the meter.
Page 8
8
1. Insert the bla ck te st l ead into the negative COM terminal
and the red test lead into the positive V terminal .
2. Set the function switch to VAC or VDC position.
3. Connect the test leads in parallel to the circuit under
test.
4.
Read the voltage measurement on the LCD display.
AC/DC CURRENT MEASUREMENTS
1. Set the function switch to the µA/mA position.
2. Insert the black test lead into the negative COM
terminal and the red test lead in to the po siti ve µA/mA terminal.
3. For current measurements up to 2000µA DC/AC, s et
the function switch to the mA position
4. Pres s the MODE b utt on t o ind ica te “D C” / “A C” on the
display.
5. Remove power from the circuit under test, then open
up the circuit at the point where you wish to measure current.
6. Touch the black test probe tip to the negative side of
the circuit. Touch the red test probe tip to the pos itive side of the circuit.
7. Apply power to the circuit.
8. Read the current in the display
Page 9
9
RESISTANCE MEASUREMENT
WARNING: To avoid electric shock, disconnect power to
the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the lin e cords.
1. Set the function switch to the Ω position.
2. Insert the black test lead into the negative COM terminal
and the red test lead into the positive Ω terminal.
3. Touch the test probe tips across the circuit or part under
test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the resistance reading.
4. Read the resistance in the display
CONTINUITY CHECK
WARNING: To avoid electric shock, never m easure
continuity on c ircuits or wires that have voltage on them.
1. Set the function switch to the position.
2. Insert the black test lead into the negative COM terminal
and the red test lead into the positive Ω terminal.
3. Press the MODE button to indicate  on the display.
4. Touch the test probe tips to the circuit or wire you wish to
check.
5. If the resistance is less than approximately 150Ω, the
audible signal will sound. If the circuit is open, the display will indicate “OL”.
Page 10
10
DIODE TEST
1. Set the function switch to the position.
2. Press the MODE button to touch the test probes to the
diode to indicate on the display under test. Forward voltage will typically indicate 0.400 to 0.700V. Reverse voltage will indicate “OL”. Shorted devices will indicate near 0V and an open device will indicate “OL” in both polarities
MAX Hold button
To hold the highest reading on the LCD
1. Press the MAX hold button. The meter reading will not
change as readings change.
2. Press the MAX hold button again to return to normal
operation.
Hold Button
The Data Hold function allows the meter to “freeze” a measurement for later reference
1. Press the “DATA HOLD” button to “freeze” t he di spl ay , t h e “HOLD” indicator will appear.
2. Press the “DATA HOLD” button to return to normal operation.
AUTO POWER OFF The auto off feature will turn the meter off after 15 minutes.
Page 11
11
REPLACING THE BATTERY
1. Remove the bottom cover and s ecure the screw.
2. Replace old battery with fresh two 1.5V AAA & 9V type battery.
3. Replace the bottom cover and secure the screw.
REPLACING THE FUSES
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the fuse cover.
1. Disconnect the test leads from the meter.
2. Remove the protective rubber holster.
3. Remove the battery cover (two “B” screws) and the
battery.
4. Remove the four “A” screws securing the rear cover.
5. Lift the center circuit board straight up from the
connectors to gain access to the fuse holders.
6. Gently remove the old fuse and install the new fuse into
the holder.
7. Always use a fuse of the proper size and value
(0.2A/250V fast blow for the 200mA range).
8. Align the center board with the connectors and gently
press into place.
9. Replace and secure the rear cover, battery and battery
cover.
Multi-Network Modular Cable Tester Operation Note: Make sure the battery power is sufficient. Insufficient
battery power will lead to dimmed LEDs and incorrect results.
Page 12
12
10 Base-T Test
1. Plug one en d of the test ed c able into the transmitting RJ45 jack on the master unit marked with a '' and the other
end of the cable into the remaining receiving RJ45 jack.
2. Slide power switch on. The upper row of LEDs will start to scan in sequence if the Auto/Manual button is set on "Auto" mode. The LED for pin 1 will light up if the button is in "Manual" mode.
3. Switch back and f orth fr om Auto or Manual sc anning mode by pressing the Auto/Manual button on the side of the master-testing unit.
4. Once both ends of the cable are plugged in properly, the second row of LEDs will illuminate according to the corresponding LEDs in the top row.
5. Read the results of the LED display for the pin configuration status of the tested cable. If you fail to read the result s the f irs t ti me in Auto mode, y ou may wait for the second LED scan, or si mply s witch to M anual mode for pin by pin testing. In Manual m ode, pressing the square "Test" button will advance testing to the next pin.
RJ11 Modular Cable Test
Please follow directions for the UTP/STP Cable Test & use the operations manual for the correct LED pin out display
Coaxial Cable Test
1. Plug the two attached BNC adapter cables on both RJ45 jacks. Then connect the tested cable to each end of the BNC adapter cables
2. For the remaining testing procedures, please refer to 10Base-T Test steps 2 to 5.
Page 13
13
Note:
1. The center pin of BNC should be read on LED 2.
2. As C oaxial cable has only two wires, we suggest y ou r ea d the result of the LED scan using Manual mode.
Remote Test
1. Plug one end of the tested cable to the transmitting RJ45 jack on the master unit marked with a ' ' and plug the other end into the remote terminator. If the tested cable is installed in a patch panel or wall plate, you may use the included patch cable to solve the connector gender problem.
2. Now, set the Auto/Manual switch to Auto mode for one-person testing.
3. Read the test results from the LED display on remote terminator.
Note: The LED display on the remote unit will scan in
sequence correspondi ng to the transmitt ing en d of the master unit.
Sample Test Results 1 2 3 4 5 6 7 8 G
1. Continuity: Pin 2 has continuity
1 2 3 4 5 6 7 8 G
2. Open: Pin 2 is opened
1 2 3 4 5 6 7 8 G
3. Short: Pin 2 and Pin 3 are s horted
1 2 3 4 5 6 7 8 G
4. Miswire: Pin 3 and Pin 6 are miswired
Page 14
14
Caution:
1. Operating the tester in live circuits may damage the tester
2. Leaving the battery in the tester for long periods of time without use could drain power from the battery