Morita Root ZX II User manual

Dental root canal measurement and treatment unit
Low Speed Handpiece Module
Operation Instructions
* Low Speed Handpiece Module must be connected to Canal Measurement Module,
which is sold separately. This unit cannot be used as an independent unit.
This manual is for Low Speed Handpiece Module. To measure a root canal refer to the manual for Canal Measurement Module.
USP. 5980248 USP. 5897315 USP. 5902105 USP. 5096419 USP. 5211556 USP. 5295833 DBP. PAT. 4126753 DBP. PAT. 4139424 DBP. PAT. 4232487
KOREA PAT. 159070 JAPA N PAT. 2873722 JAPA N PAT. 2873725 JAPA N PAT. 3071245 JAPA N PAT. 3071270 JAPA N PAT. 3113109 JAPA N PAT. 3113095 JAPA N PAT. 3213484
Thank you for purchasing ROOT ZX II Low Speed Handpiece Module. For optimum safety and performance, read this manual thoroughly before using the unit and pay close attention to warnings and notes. Keep this manual in a readily accessible place for quick and easy reference. This manual contains essential safety information.
CONTENTS
Page
1. Prevent Accidents.................................................................................................................... 1
2. Parts Identification................................................................................................................. 5
3. Assembling the Unit............................................................................................................... 6
4. Before Using the Unit............................................................................................................. 8
5. Operating the Unit ................................................................................................................11
6. After Using the Unit..............................................................................................................22
7. Sterilization, Replacement Parts and Storage....................................................................24
I
(1) Sterilization ................................................................................................................24
(2) Replacement Parts......................................................................................................26
(3) Storage ........................................................................................................................27
Maintenance and Inspection.................................................................................................28
8.
9. Troubleshooting......................................................................................................................29
10. Replacement Parts List.........................................................................................................31
11. Technical Specifications ........................................................................................................32
12. Warranty.................................................................................................................................37
Operation 12. 21. ’04
Page 1
1. Prevent Accidents
Most operation and maintenance problems result from insufficient attention being paid to basic safety precautions and not being able to foresee the possibilities of accidents. Problems and accidents are best avoided by foreseeing the possibility of danger and operating the unit in accordance with the manufacturer’s recommendations.
Note the meaning of the following symbols and expressions:
WARN ING This warns that it may result serious injury of the patient or
operator if the instructions are not followed properly.
PROHIBITION
NOTE
This alerts the user to the possibility of damage to the equipment,
The user ( e.g. the hospital, clinic etc. ) is the party responsible for the maintenance and proper operation of ROOT ZX II Low Speed Handpiece Module. ROOT ZX II Low Speed Handpiece Module must only be operated by dentists and other legally licensed professionals.
The user can not use in such a way that may result in serious injury of the patient or operator.
potential injury of the patient or operator, or important points concerning operation and performance.
Do not use this equipment for anything other than its specified purpose.
Operation 12. 21. ’04
WARN ING
This unit must not be connected to or used in combination with any other apparatus or system. It must not be used as an integral component of any other apparatus or system. J. Morita Mfg. Corp. will not be responsible for accidents, equipment damage, bodily injury or any other trouble which results from ignoring this prohibition.
Do not injure your fingers when inserting or removing files.
Do not use damaged file holders; an accurate measurement can not be made with a damaged file holder.
When continuous tone is heard while the main power switch is on and without any operation, some electrical part may be malfunction. Do not use the unit and send the unit to J. Morita office for repairing.
This unit is for prescription use only.
A rubber dam should be used when performing endodontic treatment.
Caution: US Federal law restricts this unit to sale by or on the order of a dentist in U.S.A.
Do not use the unit when the AC adapter is connected. ( see page 7, 22 )
Never operate the unit with an external power supply. ( see page 7, 23 )
If an electrical storm occurs while the battery is being charged, do not touch the AC adapter or the charger’s power supply cord as there would be a risk receiving an electric shock. ( see page 7, 23 )
After attaching contra angle to the micromotor, give the micromotor a light tug to confirm it is securely attached. ( see page 8 )
After attaching the micromotor to handpiece cord, give the micromotor a light tug to
confirm it is securely attached. ( see page 8 ) Plastic sleeves must be replaced after each patient. ( see page 8 )
Never use stretched, deformed or damaged files. ( see page 9)
Give the file a light tug to confirm it is securely held in place. If the file is not securely placed, it could come out and injure the patient. ( see page 9)
Take care to avoid the risk of a file being swallowed by the patient. ( see page 9)
Check the unit’s operation before each patient. If the indicators in the display do not appear normally, the unit may not be able to make an accurate measurement. In this case, stop using the unit and have it repaired. ( see page 10)
When measuring the length of a root canal, make sure that the file electrode does not come into contact with an electric power source such as an electrical socket. This could result in a severe electrical shock. ( see page 12)
Before measuring length of a root canal, make sure that the rotation speed does not appear on the display. If the rotation speed appears on the display, the unit is set for root canal preparation mode, and the handpiece will operate. ( see page 12)
When shaping a root canal, make sure that the file or contrary electrode does not come into contact with an electric power source such as an electrical socket. This could result in a severe electrical shock. ( see page 13)
Check the settings on the display after selecting memories. ( see page 14 )
In some cases such as a blocked root canal, a measurement can not be made.( see page 16 )
Accurate measurement is not always possible, especially in cases of abnormal or unusual root canal morphology; make sure to take an x-ray to check the measurement results. ( see page 16 )
Stop using the unit immediately if it does not seem to be working properly. ( see page 16 )
If the motor overheats, remove the micromotor from the patient’s mouth, wait until it cools down to resume treatment. ( see page 18 )
Electric noise could cause malfunction of the unit. Do not depend entirely on the unit controlling itself; always watch the display, listen to the sound and be aware of tactile feedback. ( see page 19 )
Page 2
Operation 12. 21. ’04
WARN ING
Accurate measurement is not always possible depending on the root canal condition. Make sure to take an x-ray to check the results. Also Nickel-Titanium files can sometimes wear out rather quickly depending on the shape and the degree of curvature of the root canal. Stop using the unit immediately if it does not seem to be working properly. ( see page 19 )
When shaping and cleaning a root canal, make sure that the file or contrary electrode does not come in contact with an electrical power source such as an electrical socket. This could result in a severe electrical shock. ( see page 19 )
Nickel-Titanium files are more easily broken than stainless steel files by the amount of torque applied to them. Do not try to force the file down the root canal. Also do not use these files for the root canals that have a relatively sharp curve near the apical foramen. ( see page 19 )
Nickel-Titanium files will eventually break due to metal fatigue and should be replaced before they reach this point. ( see page 19 )
Always examine files for separation and other deformities or damage before using them. Any type of deformity could result in the file breaking. ( see page 19 )
If the file touches the oral mucosa or a tooth, it will automatically start to rotate and could injure the patient. ( see page 19 )
Do not touch the oral mucosa with the metal part at the end of the contra angle. This could result in injuring the patient. ( see page 19 )
If the contra angle’s file release button is pressed against the teeth opposite the one being treated, the file could come out and injure the patient. ( see page 19 )
Never press the file release button while the micromotor is running. This could cause the button to heat up and burn the patient or the file to come out and injure the patient. ( see page 19 )
Be careful when using the foot switch; the motor may rotate regardless of the measurement results. ( see page 20 )
If [F.02] appears in the display, noise has been detected. Turn the unit off and then back on again. If [F.02] still appears, stop using the unit and contact your local dealer or the J. Morita regional office. ( see page 22 )
Autoclave the contra angle after each patient. ( see page 24 )
Page 3
Worn Area
Worn Area
File Electrode
Replace the file electrode when it is worn to the extent indicated in the illustration. It could break if used after reaching this point. ( see page 26 )
Operation 12. 21. ’04
Page 4
PROHIBITION
Do not use this unit in conjunction with an electric scalpel or on patients who have a pacemaker.
Do not use this unit in the medical operation room.
Blocked canals cannot be accurately measured.
This unit must not be connected to or used in combination with any other apparatus or system. It must not be used as an integral component of any other apparatus or system. J. Morita Mfg. Corp. will not be responsible for accidents, equipment damage, bodily injury or any other trouble which results from ignoring this prohibition.
Illumination devices such as fluorescent lights and the Film viewer which use an inverter can cause ROOT ZX II to operate erratically. Do not use ROOT ZX II near devices such as these.
Electromagnetic wave interference could cause this unit to operate in an abnormal, random and possibly dangerous manner. Cellular phone, transceivers, remote controls and all other devices which transmit electromagnetic waves located inside the building should be turned off.
Operation 12. 21. ’04
2. Parts Identification
Foot Switch
Low Speed Handpiece Module
Handpiece Cord
Micromotor
Page 5
Canal Measurement Module (Sold separately)
Liquid Crystal Display
Set Switch
Select Switch
Mode Switch
Power Switch
Contra Angle
Contrary Electrode (Canal Measurement Module accessory)
* Low Speed Handpiece Module must be connected to Canal Measurement Module. * Low Speed Handpiece Module cannot be used as an independent unit. * In this manual, root canal measurement unit is called Canal Measurement Module and root
canal treatment unit is called Low Speed Handpiece Module.
Accessories
Micromotor (1) Contra Angle (1) Handpiece Cord (1)
TR800-Blue Line TR400-Green Line (Option)
Handpiece Rest (1) AR Oil (1) AC Adapter (1)
Battery (1) Foot Switch (1) Plastic Sleeve (20)
(Pre-installed)
Operation 12. 21. ’04
3. Assembling the Unit
p
Canal Measurement Module
Canal Measurement Module
Cover
Stopper
Cover
Low Speed Handpiece Module
Page 6
Attaching Low Speed Handpiece Module to Canal Measurement Module
1. Hold the cover and slide the stopper on the bottom towards the liquid crystal display.
2. Slide the cover in the direction indicated by the arrow in the illustration and remove it from Canal Measurement Module .
* The cover and batteries will not be used.
3. Line up the tabs on Low Speed Handpiece Module with the notches in Canal Measurement Module and put them together.
4. Slide Low Speed Handpiece Module all the way down until it is securely attached.
Canal Measurement
Module
Low Speed
Hand
iece Module
NOTE
If the catch on the bottom is not back in its original
Notches
Tab s
place after attachment, push it in the direction shown by the arrow in the illustration.
After installation, give Low Speed Handpiece Module a light tug to confirm it is securely attached.
Operation 12. 21. ’04
Page 7
Charging Battery
The battery is built into Low Speed Handpiece Module.
Low Speed Handpiece Module
AC Adapter Plug
1. Line up the arrow ( ) on the AC adapter’s connector with the triangle ( ) above its jack on
the left side of Low Speed Handpiece Module and plug it in. Then plug the adapter into the electrical power receptacle.
WARN ING
Do not use the unit when the AC adapter is connected.
NOTE
The battery is not charged when the unit is shipped from the factory and must be charged before using the unit.
If the plug for the AC adapter does not fit the socket, it is the user’s responsibility to find a suitable plug adapter.
2. The Amber Charge LED on the back of Low Speed Handpiece Module starts flashing on and off and then, after a few seconds, it will stop flashing and stay on to show that the battery is being charged. It takes about 60 minutes to fully charge the battery.
3. Amber Charge LED goes out when the battery is fully charged.
4. Disconnect the AC adapter from Low Speed Handpiece Module and unplug it.
WARN ING
Never operate the unit with an external power supply.
If an electrical storm occurs while the battery is being charged, do not touch the AC adapter or
the charger’s power supply cord as there would be a risk receiving an electric shock.
NOTE
Do not pull or yank the cord when disconnecting the AC adapter.
Amber Charge LED On
Operation 12. 21. ’04
4. Before Using the Unit
Handpiece Cord Plug
Page 8
Assembling Micromotor
* Contra angle must be lubricated with AR Oil before
using for the first time. Refer to “Lubricating the Contra Angle” on page 25.
Dots
Low Speed Handpiece Module
1. Line up the arrow ( ) on the handpiece cord connector with the triangle () above its jack on the left side of Low Speed Handpiece Module and plug it in.
NOTE
Handle Low Speed Handpiece Module carefully; do not drop, bump or expose it to other kinds of impacts or
Plug in until the tip of the arrow reaches the surface marked with the small triangle (▼).
shocks. Rough handling could cause damage.
Make sure the plug is securely plugged into the jack. A poor connection can cause malfunction.
Do not drop anything on or bang the plug after it has been inserted into the jack.
Contra Angle
2. Line up the dots on the micromotor and contra angle and slide the contra angle straight onto the micromotor until it clicks securely into place. The contra angle has a simple snap-on connection.
WARNING
After attaching contra angle to the micromotor, give the micromotor a light tug to confirm it is securely attached.
Micromotor
Micromotor
3. Line up the triangle () to connect the micromotor to handpiece cord.
Handpiece Cord
WARNING
After attaching the micromotor to handpiece cord, give the micromotor a light tug to confirm it is securely attached.
Triangle Marks
4. Put a plastic sleeve on the micromotor portion.
WARNING
Plastic sleeves must be replaced after each patient.
Operation 12. 21. ’04
Page 9
Assembling File and File Electrode
Use only nickel-titanium files for root canal preparation.
WARNING
Never use stretched, deformed or damaged files.
File Release Button
File Electrode
1. Gently pull back the file electrode.
2. Hold down the file release button on the contra angle and insert the file. Turn the file back and forth until it is lined up with interior latch groove and slips into place. Release the button to lock the file into the contra angle.
WARNING
Give the file a light tug to confirm it is securely held in place. If the file is not securely placed, it could come out and injure the patient.
Take care to avoid the risk of a file being swallowed by the patient.
NOTE
Use caution when inserting and removing files to
avoid injury to fingers.
Inserting and removing files without holding the file
release button may damage the chuck.
Do not attach the file electrode to files with a shaft diameter of 1.2mm or greater, or large shafted tools such as Largo bur. The file electrode cannot be attached to files and Gates-Glidden reamers that do not have circular shafts. Set the unit for manual mode when using these types of instruments.
3. Gently clip the file electrode securely onto the file.
Contrary Electrode
NOTE
Do not clip the file electrode to the cutting part of the
file.
The file electrode cannot be attached onto some files.
Connector
Attaching Contrary Electrode
Insert the contrary electrode (lip clip) into the connector of the handpiece cord. (The contrary electrode is an accessory provided with Canal Measurement Module.)
NOTE
Always hold the connector to connect or disconnect
cords.
Operation 12. 21. ’04
Page 10
Foot Switch Plug
Power Switch
Attaching Foot Switch
Insert the foot switch plug all the way into its jack on the left side of Low Speed Handpiece Module. [This jack is marked with a triangle (▲) pointing up.]
NOTE
Always hold the connector to connect or disconnect cords.
Checking the Function
1. Press the Power switch to turn the unit on. Display used for root canal preparation will appear.
* The unit will automatically turn off after 5 minutes of
non-use.
2. Check that the handpiece cord is properly plugged into the jack.
3. Check that the contra angle is securely attached to the micromotor.
4. Check that the file is properly installed. Give it a light tug to confirm.
5. Check that the file electrode is properly clipped onto the file.
Contrary Eelectrode
File Holder Contact
6. Check that the contrary electrode is attached to the connector of the handpiece cord.
7. Contact the file with contrary electrode and check that all the canal length indicator bars on the display are lit, the word “APEX” flashes and audible beep becomes continuous. Use caution when contacting the file with contrary electrode as the file starts to rotate as soon as the file touches the contrary electrode.
WARNING
Check the unit’s operation before each patient. If the indicators in the display do not appear normally, the unit may not be able to make an accurate measurement. In this case, stop using the unit and have it repaired.
Operation 12. 21. ’04
Loading...
+ 28 hidden pages