Moore Industries IPH User Manual

April 2016 170-746-00D
Field-Mount
Current-to-Pressure Transmitter
IPH
Field-Mount
Current-to-Pressure Transmitter
IPHIPH
IPHIPH
IPH
IPH

Table of Contents

Introduction.....................................................................................................
Description ...................................................................................................................... 1
Specifications .................................................................................................................. 1
Ordering Information....................................................................................................... 2
11
1
11
Calibration .......................................................................................................
Necessary Equipment..................................................................................................... 2
Preparing for Calibration ................................................................................................. 2
Calibration Process ......................................................................................................... 4
Installation.......................................................................................................
Phase One: Mounting ..................................................................................................... 6
Phase Two: Electrical Connections................................................................................. 6
Phase Three: Pneumatic Connections............................................................................ 6
Operation.........................................................................................................
Maintenance ....................................................................................................
Necessary Equipment..................................................................................................... 7
22
2
22
66
6
66
77
7
77
77
7
77
Disassembling the IPH.................................................................................................... 7
Cleaning the Subassembly.............................................................................................. 7
Changing the Pneumatic Block ....................................................................................... 8
Re-assembly ................................................................................................................. 10
Troubleshooting............................................................................................
1010
10
1010
IPH
Introduction
This is the users’ manual for Moore Industries’ IPH Current-to-Pressure Transmitter. It contains all of the information that is needed to configure, install, operate, maintain, and troubleshoot the IPH. If you would like additional assistance, call one of our STAR centers.
The following guidelines are used in this manual:
WARNING
could injure the operator.
Caution
damage or destroy the unit.
Note
condition, or operation of the unit.

Description

The IPH is Moore Industries’ rugged and compact Field-Mount Current-to-Pressure (I/P) Transmitter. It is designed to meet NEMA 4X and IP55 requirements, is watertight, dust-tight, and resistant to corrosion and chemicals. These qualities make the IPH the ideal transmitter for extended use in harsh situations.
- Hazardous procedure or condition that
- Hazardous procedure or condition that could
- Information that is helpful for a procedure,
The 2-wire (loop-powered) IPH accepts a 4-20mA input from a DCS, PLC or PC-based control system. It converts the current signal to a pneumatic signal (3-15psig, 0.2-1 Bar, 20-100 kPa, etc.), providing precise proportional control of valves, actuators, and other pneumatically-controlled devices. Reverse pneumatic outputs (15-3psig, 1-0.2 Bar, 100-20 kPa, etc.) are also available.
The IPH is available with an optional coalescing filter/regulator that combines an air filter, a miniature supply line regulator, and a pressure gauge that reads in both psi (0-60) and Bars (0-4).
Model/Serial Number. Moore Industries tracks all their IPH units with a system of model numbers and serial numbers. Please be prepared to provide these numbers if you need service information or assistance.
The IPH serial and model numbers can be found on the instrument’s metal tag.

Specifications

Performance
Specifications and information subject to change without notice.
Accuracy: Less than
0.5% of span including the combined effects of linearity, hysteresis, and repeatability-defined as independent linearity per SAMA standard PMC 20.1-
1973. (For 3-27psig, error will not exceed 1% of span) Step Response: 0.3 sec­onds into 100ml (6 cu. in.) at 90% of output span
Supply Pressure Effect:
Maximum 0.3%/1.4psig (0.3%/0.1BAR)
Air Consumption:
0.1SCFM (dead-ended) Air Capacity: 1.4 SCFM maximum
Performance
(continued)
Ambient
Temperature
Maximum Input: 150%
of input span without damage RFI/EMI Effect: With field strengths of 10 volts per meter, at frequencies of 20-500MHz, it has less than ±0.1% of span change (standard)
Mounting Position Effect: Negligible, unit
mounts in any position, but should be calibrated in final orientation
Range: –40 to +60°C (–40 to +140°F) Effect: Less than ±0.1% of input span per °C over above range (0.005%/°F)
Adjustments
Connections
Weight
Zero: Multiturn screw
adjusts zero to ±3% of span Span: Multiturn potenti­ometer fully adjustable to 100% of output span
Pneumatic: 1/4-inch NPT female for both sup­ply air and output air. Air supply also has an addi­tional (normally plugged) port for optional filter/ regulator (-FR1 option) Electrical: WDNS and WDNA, 1/2-inch NPT fe­male; WDMS and WDMA, M20 x 1.5 female
993 grams (2lb, 3oz)
The Interface Solution Experts 1
IPH

Ordering Information

Unit Input Output Supply Pressure Options Housing
IPH
Current-to­Pressure Transmitter
When ordering, specify: Unit / Input / Output / Supply Pressure / Option [Housing] Model number example: IPH / 4-20MA / 3-15PSIG / 20PSI / -FR1 [WDNS]
4-20MA
into 260 ohms, nominal
3-15PSIG 3-27PSIG 15-3PSIG 20-100KPA .2-1BAR .2-1KGCM2
*The unit’s output must match the supply pressure to its right.
20PSI 35PSI 20PSI 140KPA
1.4BAR
1.4KGCM2
-FR1 Combines a
coalescing filter, a miniature supply line regulator, and a pressure gauge that reads in both psi (0-
60) and Bars (0-4)
-ISF FM approved Intrinsically Safe
-ISE PTB/CENELEC Intrinsically Safe
-NE CENELEC Type N
WDNS Aluminum body with PBT polyester cover; NPT pneumatic ports and NPT electrical entry port WDNA Aluminum body with aluminum cover; NPT pneumatic ports and NPT electrical entry port WDMS Aluminum body with PBT polyester cover; NPT pneumatic ports and metric electrical entry port WDMA Aluminum body with aluminum cover; NPT pneumatic ports and metric electrical entry port
NOTE: Add “P” suffix to the model housing designator for 2-inch pipe mounting hardware (e.g., WDNSP, WDMSP)

Calibration

Every IPH is fully tested and calibrated at the factory prior to shipment. Before installation, however, your IPH’s should be bench-checked to verify proper operating levels and to set the desired unit zero and unit span. Calibration should be conducted in an appropriate testing environment.

Necessary Equipment

Table 1 lists the equipment required to calibrate the IPH. This equipment is not supplied by Moore Industries, but should be available in most labs or maintenance areas.
Table 1. IPH Calibration Equipment
Equipment Specifications
Current So urce
Cali brated, adj ustable. Must be capabl e of di scret e ou t p ut levels withi n t h e appropri ate, rated range for the unit un der test; 4-20 or 10-50 mA. Accuracy of ±0.1%

Preparing for Calibration

To prepare the IPH for calibration, remove the screw-on cap and connect the unit to your calibration equipment as shown in Figure 1 on page 3, which shows the calibration setup.
Unit Connections and Controls. The IPH has two labeled terminals on its faceplate, located under the top protective cover of the housing. The terminal labeled “+I” is for connection of the positive current input, and the terminal labeled “–I” is for negative connections.
DC Multimeter
Air Supply
Air Pressure Gauges
Pneumati c Test Load
Screwdriver
2 The Interface Solution Experts
Calibrated. Accuracy of ±0.1% (optional equipment)
Filtered, regulated, instrument-quality.
Two calibra ted gauges; #1 with accuracy of ±2%, #2 with accur acy of ±0.1%.
Calibrated. Volume of appr oximately 120 ml (75 in3), per IE C spec #770.
Slotted-tip. Head width of 5 mm (0.1875 in), maximum.
IPH
The two IPH controls are also located inside the unit housing under the screw-on cap. They consist of a mechanical adjustment screw and an electronic potentiometer, each accessed through the front panel. They are labeled “zero” and “span”. The zero adjust­ment screw mechanically provides a control range for zero offsets of ±3 percent of rated unit span.
The span pot electronically adjusts unit full-scale output to 100 percent of rated span. It requires ap­proximately 22 turns to move its wiper from one extreme to the other, clockwise for maximum or counterclockwise for minimum span. It is equipped with a slip clutch to prevent damage if the adjustment is turned beyond the wiper stop.
Note: Always use clean, dry, instrument air when
calibrating or operating the IPH.
All pneumatic lines used in calibration and operation must be “blown down” (purged) prior to connection to the IPH. Any condensation or oil residue in the lines, if introduced into the pneumatic chambers of the IPH, may result in poor unit performance.
Calibration Steps:
(See Figure 1)
1. Orient IPH vertically or horizontally based on how it will be positioned in the application.
2. With zero air input (supply off), connect 1/4-inch pneumatic tubing between the appropriate output port of the regulated instrument air supply and the calibrated pressure gauge #1 (accuracy of ±2% of span). Connect another hose from the pressure gauge to the port labeled “IN” on IPH.
3. Connect 1/4-inch pneumatic tubing between the IPH port labeled “OUT” and the appropriate port of pressure gauge #2 (accuracy of ±0.1% of span), then from gauge #2 to the appropriate pneumatic load.
4. Run current source wiring through conduit opening in housing, and to front panel of IPH.
5. Connect positive lead of adjustable current source to +I terminal of IPH. Connect negative source lead to –I IPH terminal. A dc multimeter may also be connected to verify level of current input.
Figure 1. IPH Calibration Setup
PRESSURE GAUGE #2
(ACCURACY ±0.1%)
PNEUMATIC
TEST LOAD
OUTPUT
PORT
IPH
SPAN
ZERO
CURRENT TO PRESSURE TRANSMITTER
INPUT
PORT
+I
DC
MULTIMETER
+
-I
_
_
ADJUSTABLE
CURRENT
SOURCE
+
PRESSURE GAUGE #1
(ACCURACY ±0.2%)
REGULATED
INSTRUMENT
AIR SUPPLY
The Interface Solution Experts 3
IPH
6. When connections are complete, apply 4mA of input current. (0% of total current)
7. Apply appropriate filtered, instrument-quality air to supply line: 20 or 35 psi (40 kPa, or 1.4 bar). Verify appropriate supply pressure by checking Supply Pressure field of unit model number.
8. Set span pot to approximately mid-scale (22 turns in either direction, then approximately 11 turns in the opposite direction).
9. Set zero adjustment screw fully clockwise, then five turns counterclockwise.
10. Allow approximately 30 seconds for calibration setup to stabilize.
Calibration Process
To perform the recommended bench-check for the IPH, first perform the setup as described in the preced­ing section. The calibration procedure consists of a basic check and adjustment of unit zero and span, based on the reading of pressure gauge #2 (accuracy of ±0.1%).
Table 2 lists the values to be used during calibration. Refer to the table when performing the following:
1. Check unit zero setting. Monitor reading of pressure gauge #2 (output), and turn zero adjust screw counterclockwise to lower output, clockwise to raise output. Set zero adjust screw so that pressure output is at 0% of appropriate range (±3%) when 0% input is applied.
2. Check unit span setting. Increase input to 100% of rated span (20 mA).
3. Monitor reading of pressure gauge #2 (output), and adjust span pot so that reading is at 100% of appropriate pressure range for your unit.
Check the values listed in Table 2.
4. Repeat steps 1 through 3 until IPH outputs 0% of rated pressure range at 0% current input, and 100% of output pressure range at 100% current input.
5. Verify the accuracy of your adjustments by input­ting the appropriate percent of span levels listed in table 2.
Table 2. IPH Calibration Values
At
0% of
Rated Inpu t
IPH Output
Configuration
3-15 psig 3 psi 15 psi 6 psi 9 psi 12 psi
15-3 psig 15 psi 3 psi 12 psi 9 psi 6 psi
0.2- 1 ba r 0.2 bar 1 bar 0.4 bar 0.6 bar 0.8 bar
1-0. 2 ba r 1 bar 0.2 bar 0.8 bar 0.6 bar 0.4 bar
3-27 psig 3 psi 27 psi 9 psi 15 psi 21 psi
3-16.6 psig 3 psi 16.6 psi 6.4 psi 9.8 psi 13.2 psi
20-100 kPa 20 kPa 100 kPa 40 kPa 60 kPa 80 kPa
Range
(4mA)
Adjust Zero until Gauge
#2 Reads:
At
100% of
Rated Inpu t
Range (20mA)
Adjust Sp an
until Gauge #2
Reads:
Verif y by Applying
25% of Rated Input
Range (8mA)
Gauge #2 will
Read:
Verif y by Applying
50% of Rated Input
Range (12mA)
Gauge #2 will
Read:
Verif y by Applying
75% of Rated Input
Range (16mA)
Gauge #2 will
Read:
100-20 kPa 100 kPa 20 kPa 80 kPa 60 kPa 40 kPa
4 The Interface Solution Experts
Figure 2. IPH Current-to-Pressure Transmitter Housing Dimensions
CONDUIT WIRE ENTRY
METRIC (M20 x 1.5)
OR NPT (1/2-14)
143 mm
(5.6 in)
85 mm (3.4 in)
THREADS
-FR1 OPTION
(COALESCING
FILTER/REGULATOR
AND PRESSURE GAUGE)
IPH
SCREW-ON CAP
21 mm (0.8 in)
51 mm (2.0 in)
1/4 NPT FEMALE PNEUMATIC OUTPUT PORT
119 mm
OUT
(4.7 in)
33 mm
(1.3 in)
53 mm (2.1 in)
IN
1/4 NPT FEMALE
PNEUMATIC
INPUT PORT
(UNUSED PORT IS
PLUGGED)
140 mm
(5.5 in)
WITH
-FR1 OPTION SLIP-ON DRAIN
CONNECTION FOR
1/4-INCH FLEXIBLE TUBING
0.36 DIA. (9.15mm)
MOUNTING HOLES (2)
91 mm (3.6 in)
157 mm
(6.2 in)
1/4 NPT PNEUMATIC
INPUT PORT FOR OPTIONAL
-FR1 FILTER/REGULATOR/GAUGE
36 mm (1.4 in)
23 mm (0.9 in)
INSTRUMENT
TAG
UNIVERSAL MOUNTING BRACKET
(SURFACE OR PIPE MOUNT; VERTICAL OR HORIZONTAL)
72 mm
(2.9 in DIA.)
9.7 mm
(0.38 in DIA.)
69 mm (2.7 in)
69 mm (2.7 in)
89 mm
(3.5 in)
NOTE:
For NEMA4X and IP55 Installations, the unit must be mounted
vertically or within 45 degrees of vertical.
The Interface Solution Experts 5
IPH

Installation

The installation of the IPH is carried out in three phases. The first phase is the physical mounting of the unit. Next is the electrical connections phase, and finally, pneumatic connections can be made. It is strongly recommended that IPH’s be installed in this order. It is also strongly suggested that each unit be calibrated according to the instructions in this manual before being placed into service.
The IPH may be installed at any angle: either surface­mounted or attached to pipe or round conduit. When deciding on the placement of the IPH unit, make sure to consider the possibility of needing to: access the front panel, check the fittings, read the -FR1 Option gauge, or drain the filter.
Closed Loop/Open Loop. The IPH should be installed in a closed loop. A closed loop is the best way to measure a control variable. It also helps determine if a deviation from a desired value exists and provides feedback for actuator loading pressure.
An open loop has limitations that are not consistent with precise control. Long term drift of the loop dynam­ics, load fluctuations that require constant adjustments of the actuator loading pressure, and performance quality variations due to inconsistencies between operating personnel are all problems commonly associated with open loops. A controlled variable cannot be directly measured in an open loop; this prevents compensating adjustments to the system input.

Phase Two: Electrical Connections

To complete the electrical connections, route the wiring through the conduit port to the terminal block, then use a slotted-tip screwdriver with a maximum head width of 3 mm (0.125 inch) to loosen the terminal screws.
The terminals are comprised of compression-screw sockets that accommodate 22-14 AWG wiring.
Connect positive lead (+) to terminal labeled “+ negative lead (–) to terminal labeled “–I”. Tighten the
terminal screws until snug. Use shielded, twisted-pair wiring for low-level input.
Ground the shielding wire as close as possible to the installed IPH.
I”, and

Phase Three: Pneumatic Connections

To complete the final phase of installation, the pneu­matic connections, connect the supply line to the 1/4-inch NPT female port labeled “IN”. Connect the output line to the 1/4-inch NPT female port labeled “OUT”.
Seal all fittings with Teflon® tape, or equivalent. Always “blow down” (purge) all tubing and the controlled device before connecting the IPH.
Filters. The IPH requires filtered, dry, regulated, instrument-quality air to prevent clogging and to ensure extended periods of maintenance-free operation. Moore Industries suggests the following levels of filtering protection:

Phase One: Mounting

Figure 2 gives the IPH’s outline dimensions. The illustration also gives the dimensions of the available
-FR1 Option hardware, which is recommended for most installations.
After placing the IPH in the desired location and orientation, secure the housing with the optional pipe mounting hardware or other appropriate fasteners.
If you have further questions on how to mount the IPH, contact your Moore Industries Sales Representative for assistance.
NOTE:
For NEMA4X and IP55 Installations, the unit must be
mounted vertically or within 45 degrees of vertical.
6 The Interface Solution Experts
Pre-filter – A general purpose “rough” filter, used to reduce particulate matter to 5 microns in size. Also removes bulk liquids. Although not required, this filter is especially recommended to protect the
0.01 micron final filter when used.
Final Filter – A second, final filter is recom­mended, to remove particulate matter in sizes down to 0.01 micron. This filter removes virtually all condensable liquids from the air stream as well.
Filter/Regulator Module Option – A combined filter/regulator assembly, the -FR1 Option, offered as an accessory for the IPH, removes particles down to 0.01 micron, supplying regulated, instru­ment-quality air to the unit. This space-saving module is affixed to the IPH supply port, and comes with a pressure gauge scaled in both psi (0-60) and bars (0-4).
IPH
Operation
Once the IPH unit’s calibration has been checked or adjusted and the unit has been installed properly, the IPH operates unattended with the exception of the minor maintenence procedures that are described in the next section.
Remember that if an IPH is installed in an open loop, it may appear to drift over extended periods of time because of the lack of corrective feedback.
If the unit is determined to be the cause of a loop irregularity, carry out the maintenance procedure in the next section of this manual. If problems persist, refer to the Troubleshooting Section.
Instrument-quality Air. Air from the application continuously flows through the IPH during operation. Depending upon the purity of the air supply, the unit’s internal assembly may have to be removed and cleaned to ensure continued optimum performance.
Initially, random checks can help establish a satisfac­tory maintenance interval geared to the user’s air supply cleanliness. Refer to the next section for instruction on the disassembly and cleaning of the IPH.
this equipment is not available, your facility may not be qualified to perform the operations described in this section. Contact Moore Industries’ Customer Service Department for more information.

Disassembling the IPH

To disassemble the IPH for maintenance/cleaning, unscrew the protective top cover, then unscrew the two screws from the base and remove the unit.
Take the internal assembly out of the housing by pulling straight out, away from housing base. Apply steady pull, disengaging the pneumatic fittings from their ports in the housing base. Do not disassemble the pneumatic block at this time.
Separate the faceplate from the transducer by remov­ing the screws on both sides of the plate assembly.

Cleaning the Subassembly

1. Place subassembly on work surface with zero
screw and span pot down.
2. Use syringe to fill Orifice #1 with trichloroethane
(TCE). See Figure 3.

Maintenance

Before beginning IPH maintenance, the unit must be removed from its application.
Warning: It is strongly recommended that the
maintenance procedures be performed in a clean,
controlled environment, such as a technicians’
bench, a laboratory, etc. Several internal parts are
small and precision-machined and can be easily lost
or damaged if an attempt is made to perform
maintenance in the field.
After maintenance, each IPH should be recalibrated before it is returned to service. Refer to the Calibration Section of this manual for instructions.

Necessary Equipment

To perform the basic maintenance procedure for the IPH, refer to Table 3, which lists the required equipment. These materials are not supplied by Moore Industries, but should be available in maintenance areas prepared to perform this type of procedure. If
3. Clean nozzle air passage by gently passing
cleaning wire back and forth through opening in subassembly base.
4. Soak several small strips of clean paper in TCE.
5. Set the subassembly on its side and slide one strip
between the nozzle and the flapper. Carefully apply slight pressure to flapper until it rests against nozzle with soaked paper in between.
6. Maintain pressure while pulling paper out. Repeat
with other strips of paper until no residue is trans­ferred to paper.
7. Inspect air passage O-ring. If damaged, contact
Moore Industries’ Customer Service Department for replacement.
8. Use instrument air supply to dry and generally
“blow out” subassembly. Put small amount of alcohol in air passage and set subassembly aside.
The Interface Solution Experts 7
IPH

Cleaning the Pneumatic Block

Refer to Figure to disassemble the IPH pneumatic block, and when performing the following:
1. Place block on work surface with its two socket­head screws facing up. Use a screwdriver to mark one side of the block with several small scratches for use when re-assembling parts.
2. Dip cleaning wire in TCE and use it to clean orifice #1, the air passage between block and mechanical subassembly.
3. Remove two screws that hold the pneumatic block together. Separate section #1 from block.
4. Use swabs dipped in TCE to clean internal sur­faces and to remove any dirt particles or oil. Flush all parts with alcohol after cleaning.
5. Push cleaning wire dipped in TCE through the small orifices on underside of section #1, in the

Table 3. IPH Maintenance Equipment

center opening, and along the outside edge. Use a syringe filled with alcohol to flush the openings. Set section #1 aside.
6. Remove spring, disk, and diaphragm from top of section #2. Inspect each for deterioration and dirt. Use compressed air to blow off all parts. Flush orifice with alcohol after cleaning.
7. If necessary, remove diaphragm from bottom surface of section.
8. Push cleaning wire dipped in TCE through brass fitting in center of section.
9. Clean small orifices along edge of section (top and bottom) with cleaning wire and flush with alcohol syringe.
10. Use swabs dipped in TCE to clean all surfaces.
11. Repeat cleaning/flushing procedure for orifices and surfaces of section #3.
Equipment Specifications/Notes
Instrum ent-Qua lity Air
or Nitrogen Supply
White, B ond Paper Clean, undyed, and unlaminated.
Cotton Swabs Clean, general-utility swabs for use in cleaning surfa ces and absorbin g excess
Trichloroethan e
(TCE)
Isopropyl Alcohol Clean, general-purpose flushing solution.
Syringe For injecting alcohol into small orifices.
Screwdriver Slotted-tip. Head width of 5 mm (0.1875 in), maximum.
Probe T echnic i a n’s tool f o r ma nipulat i on of very s m all parts .
Cleaning Wire 0.005-inch diameter, maximum.
Reduced to between 20 and 30 psig, and fitted with a hose and fine tip or nozzle.
solvent and alcohol.
Rho-T r on - TPC-400 or equivalent.
Hex Keys One 3/32 in, minimu m length 5.5 in. Ball-t ipped head re com m ended;
Removeable
Thread-locking Compound
8 The Interface Solution Experts
One 5/ 32 in, stan da r d; One 4 mm, s t a ndard
Loctite® #242 or equivalent removable threadlocker.

Figure 3. IPH Internal Assemblies

IPH
PNEUMATIC
BLOCK
SECTION #1
HOLE
ROLL IS UP
ROLL IS DOWN
HOLE
MECHANICAL
SUB-ASSEMBLY
NOZZLE
FLAPPER
ORIFICE #1
UPPER SPRING (HEAVY GAUGE WIRE)
PLASTIC DISK UPPER RUBBER
DIAPHRAGM BRASS
FITTING
SECTION #2
MIDDLE RUBBER DIAPHRAGM
HOLE
RESTRICTOR
HOLE
SUPPLY
SCREEN
SECTION #3
LOWER RUBBER DIAPHRAGM/DISK
SECTION #4
The Interface Solution Experts 9
IPH
12. Remove integrated diaphragm/fitting from section #4.
13. Repeat cleaning/flushing procedure for orifices and surfaces.
14. Check to make sure that the spring is in good condition.
15. Locate the supply port air filter screen on the pneumatic block.
16. Remove filter screen with needle, and flush with TCE. Rinse with alcohol and set aside to dry.
Re-assembly
1. Dry all parts of the pneumatic block and the mechanical subassembly with the specified air or nitrogen supply. Make sure all parts are clean and dry.
2. Use the scratch markings made earlier to make sure that each section of pneumatic block is oriented properly, then use small socket-head screws to secure re-assembled block. Tighten screws to between 35 and 45 Nm of torque (26 to 33 ft/lbs).
3. Line up pin on section #1 of pneumatic block with hole on base plate of mechanical subassembly.
4. Place a small amount of thread-locker on the threads of the head screw used to hold the pneu­matic block and mechanical subassembly together, and secure two subassemblies to each other with the screw. Make sure screw is as tight as pos­sible.
Drain Check. System filters have automatic drains that depend on the fluctuation of system pressure to induce drainage. A stable system may not drain efficiently.
Check periodically for clogs and drain system’s filters by pushing the drainage valve with a small probe or wire.

Troubleshooting the IPH

Many components of the IPH have been thermally aged, tested, and selected using a computer-aided design program. This usually makes field repair unnecessary.
It is therefore recommended that any properly main­tained unit found to be performing below specifications be returned to the factory in accordance with the instructions found on the back cover of this manual.
If a problem is suspected with the IPH, review the following steps:
1. Verify that bench instruments used to take mea­surements have the proper range and accuracy and are within current certification period limits.
2. If a change in the relationship between the input and output is detected, attempt a re-calibration of the IPH.
3. If the response time lengthens, or if the span drops, this may indicate a blockage due to air supply contamination. Follow the instructions in the Maintenance Section of this manual.
5. Place mounting bracket over subassembly and pneumatic block and screw the bracket onto the subassembly.
6. Slide internal assembly into housing, being careful not to crimp or pinch the current wiring, and mate fittings in housing with ports on pneumatic block of internal assembly. Press down until locked into place.
7. Reseat the IPH and screw it in.
8. Calibrate unit. Refer to Calibration Section of this manual for detailed instructions.
10 The Interface Solution Experts
The Zero Adjust Screw. This screw has a flange to prevent its being turned too far counterclockwise during adjustment. Forcing the screw past the point at which its flange comes into contact with the unit front panel could disengage it from the internal assembly.
If this happens, disassemble the IPH as described in the Maintenance Section of this manual.
Visually inspect the screw, and if serviceable, hold down the flapper arm and turn the screw clockwise until its pin is in a position to hold the arm.
Refer to the Calibration Section of this manual, and recalibrate unit.
IPH
IPH:
Current-to-Pressure Transmitter

Appendix A: Intrinsic Safety

This page contains the installation diagram for the IPH carrying the intrinsically safe option. It also includes guidelines for setting up zener barriers necessary in these types of applications.
REVISION
DRAWING NUMBER
CB
G.E. 09/00
100-100-53 B
ECO 13163
REVISED BY DATE BY APPROVAL
NOTICE RE PROPRIETARY INFORMATION: This drawing and the inform-
ation contained herein are the proprietary property of Moore Industries-
International, Inc. (MII) and should not be reproduced or disclosed to any
third party without the written consent of an authorized officer of MII.
= 30 VDC
i
U
or
Entity Parameters
max
V
(Power/Loop, +PS & -PS):
ITTER
SM
AN
E TR
R
T TO
ESSU
EN
R
PR
R U
C
= 85 mA
= 0.6375 W
= 0 µF
i
i
i
I
P
or
or
max
I
max
C
P
ELEC IN
cable
cable
+ C
+ L
i
i
= 5 mH
³ C
³ L
i
o
o
L
C
or
or
a
a
L
C
L
These diagrams must be used to augment the installa­tion instructions earlier in this manual for units that are to operate in areas requiring intrinsically safe instru­mentation.
t
t
or V
or I
oc
sc
³ V
³ I
i
i
I
U
or
or
max
max
I
V
PNEU IN
PNEU OUT
CSA International
Class I, Div. 2, Groups A-D.
Class II, Div. 1, Groups E, F & G.
Hazardous 'Classified' Locations/Areas:
Intrinsically Safe: Class I, Div. 1, Groups A-D.
II 1G EEx ia IIC.
CENELEC/ATEX
Class II, Div. 2, Groups F & G. Class III.
Intrinsically Safe:
£ +60˚C
amb.
Environmental Protection: IP65 & NEMA 4X
-I
+I
No modifications are permitted
SPAN
ZERO
GROUND
CURRENT TO
PRESSURE TRANSMITTER
Non-Hazardous Safe Area
(3)
-PS
+PS
(2)
Apparatus
Associated
Barrier or other
(1)
Apparatus
Unspecified
Hazardous (Classified) Locations - FM (US NEC 500):
Intrinsically Safe System
Current-to-Pressure T ransmitter
CONTROL DRAWING
Field Installation Diagram:
For Hazardous 'Classified' Locations.
TITLE
09/00
IPH,
09/00
09/00
CATEGORY
Current-to-Pressure Transmitter
Current-to-Pressure Transmitter
IPH:
IPH:
Gus H. Elias
DRAWN
CHECKED
/mm
= ±inch
(UNLESS NOTED)
TOLERANCES
DO NOT SCALE DRAWING
DECIMALS.X.XX
W.Ho
ENGINEER
/2.54
/0.76
= ±.1
= ±.03
NONE
Gus H. Elias
SCALE
/0.25
= ±.010
= ± 30'
=+.003-.002/+.08-.05
ANGLES:
.XXX
HOLES:
CERTIFIED PRODUCT
Non-Incendive: Class I, Div. 2, Groups A-D.
T Code: T6 @ 60˚C Maximum Operating Ambient
Class II, Div. 2, Groups F & G and Class III, Div. 2.
Intrinsically Safe: Class I,II,III; Div. 1; Groups A-G.
Exceed 1 W
Ground/Earth Path
Resistance Must Not
Temperature. Range: -40˚C £ T
exceed 30 VDC & the output current (Isc, It or Io) must not exceed 85 mA. Also, it must be installed per the manufacturer's guidelines.
be approved by the "specific" (CSA/EECS/FM/LCIE/SAA/SIRA/TUV, etc..) certifying agency for I.S. connections in: "Class I-III, Division 1, Groups
be in accordance to 'ANSI-P12.6' (Installation of I.S. Systems for Hazardous 'Classified' Locations) and the National Electric Code 'ANSI/NFPA 70' . Also, a
be used when installed in Class II and Class III environments. For CSA applications, adhere to the 'Canadian Electric Code C22.1' most current publication on I.S. installation
Substitution of components may impair the Intrinsic Safety of the unit. DO NOT open the unit when either energized or when an explosive gas atmoshpere is present. Disconnect power
drawing.
This is a controlled 'Related' or 'Schedule'
without the notification and final approval of
or the Certifying Agency (schedule dwgs.).
the Q.A. Certification Engineer (related dwgs.)
Notes:
(1) Apparatus which is unspecified except that it must not be supplied from, or contain under normal or abnormal conditions a source of potential with respect to earth in excess of 250 VRMS or 250 VDC
which is considered to be the Safe Area's maximum voltage.
(2) The Barrier or other Associated Apparatus must
A Shunt Zener Barrier is NOT required for Non-Incendive (or Class I, Division 2 or Type N) installations.
A-G" locations. The output voltage (Voc, Vt or Vo) must not
4- For FM applications, installation must
(3) The combined Capacitance and Inductance of the inter-connecting cables and the PC Prog. Transmitters must not exceed the values indicated on the Associated Apparatus.
dust-tight conduit seal must
before servicing. Also read, understand and adhere to the manufacturer's installation and operating procedures.
guidelines. For CENELEC/ATEX applications, adhere to 'BS5345 or EN 60079-14:1997' or any equivalent, most current and pertaining publication on I.S. installation guidelines.
5- Warning:
The Interface Solution Experts 11
GUIDELINES AND CERTIFICATIONS

Low Voltage Directive

When installing any Moore Industries product, al­ways follow all local regulations and standards for grounding, shielding, and safety. The following grounding and wiring practices must be followed in order for the unit(s) to meet the requirements set fourth in the EMC directives EN50082-2 and EN55011.
Grounding
If the unit has a metal case it is to be grounded. (DIN rail mounted units should be mounted on a grounded rail).
Wiring
Twisted shielded wire should be used for all input and output signals. The shields are to be grounded at the units, to earth ground (safety ground). The un-shielded part of the in/output wires should be no longer than 2 inches.
CE Certification-related Guidelines
The following guidelines must be followed in order to comply with EN61010-1 (Low Voltage Directive). These items affect the AC version of the IPH. If these products are to be used in a non-CE environ­ment, this directive may be disregarded.
WARNING:
If this unit is used in a manner not specified by
Moore Industries, the protection provided by the
equipment may be impaired.
Switches and Circuit Breakers
A switch or circuit breaker must be wired in series with the AC power conductors. The switch or cir­cuit breaker used must be located within three meters of the unit.
WARNING:
Terminals on this unit may be connected to
hazardous voltages. Before making ANY
connections to this unit, ALL hazardous voltages
must be de-energized.
The circuit breaker or switch will only remove power to the unit, hazardous voltages may still be connected to other terminals on the unit.
Installation Category
All of Moore Industries’ terminals are rated CAT II, except those with the -RF option. These terminals are rated CAT I.
Equipment Ratings
The IPH does not generate hazardous voltages, rather, it accommodates a current input, and gen­erates a pressure output. Products connected to the IPH should be designed to receive these in­puts.
Supply Wiring
All power connections should be made with the proper wire.
The end of each conductor should be stripped no more than 8mm. The end of the stripped wire should be tinned with solder, or inserted into a fer­rule and crimped before being placed into a termi­nal block.
Conductors connected to screw-type connections should have a ring-lug or spade-lug crimped onto the wire end.
RETURN PROCEDURES
To return equipment to Moore Industries for repair, follow these four steps:
1. Call Moore Industries and request a Returned Material Authorization (RMA) number. Warranty Repair –
If you are unsure if your unit is still under warranty, we can use the unit’s serial number to verify the warranty status for you over the phone. Be sure to include the RMA number on all documentation.
Non-Warranty Repair – If your unit is out of warranty , be prepared to giv e us a Purchase Order n umber when you call. In most cases, we will be able to quote you the repair costs at that time. The repair price you are quoted will be a “Not To Exceed” price, which means that the actual repair costs may be less than the quote. Be sure to include the RMA number on all documentation.
2. Provide us with the following documentation:
a ) A note listing the symptoms that indicate the unit needs repair b ) Complete shipping information for return of the equipment after repair c) The name and phone number of the person to contact if questions arise at the factory
3. Use sufficient packing material and carefully pack the equipment in a sturdy shipping
container.
4. Ship the equipment to the Moore Industries location nearest you.
The returned equipment will be inspected and tested at the factory. A Moore Industries representative will contact the person designated on your documentation if more information is needed. The repaired equipment, or its replacement, will be returned to you in accordance with the shipping instructions furnished in your documentation.
WARRANTY DISCLAIMER
THE COMPANY MAKES NO EXPRESS, IMPLIED OR STATUTORY WARRAN­TIES (INCLUDING ANY WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE) WITH RESPECT TO ANY GOODS OR SER­VICES SOLD BY THE COMPANY. THE COMPANY DISCLAIMS ALL WARRAN­TIES ARISING FROM ANY COURSE OF DEALING OR TRADE USAGE, AND ANY BUYER OF GOODS OR SERVICES FROM THE COMPANY ACKNOWL­EDGES THAT THERE ARE NO WARRANTIES IMPLIED BY CUSTOM OR USAGE IN THE TRADE OF THE BUYER AND OF THE COMPANY, AND THAT ANY PRIOR DEALINGS OF THE BUYER WITH THE COMPANY DO NOT IM­PLY THAT THE COMPANY WARRANTS THE GOODS OR SERVICES IN ANY WAY.
ANY BUYER OF GOODS OR SERVICES FROM THE COMPANY AGREES WITH THE COMPANY THAT THE SOLE AND EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY CONCERNING THE GOODS OR SERVICES SHALL BE FOR THE COMPANY, AT ITS OPTION, TO REPAIR OR REPLACE THE GOODS OR SERVICES OR REFUND THE PURCHASE PRICE. THE COMPANY SHALL IN NO EVENT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES EVEN IF THE COMPANY FAILS IN ANY ATTEMPT TO REMEDY DEFECTS IN THE GOODS OR SERVICES , BUT IN SUCH CASE THE BUYER SHALL BE ENTITLED TO NO MORE THAN A REFUND OF ALL MONIES PAID TO THE COMPANY BY THE BUYER FOR PURCHASE OF THE GOODS OR SERVICES.
ANY CAUSE OF ACTION FOR BREACH OF ANY WARRANTY BY THE COMPANY SHALL BE BARRED UNLESS THE COMPANY RECEIVES FROM THE BUYER A WRITTEN NOTICE OF THE ALLEGED DEFECT OR BREACH WITHIN TEN DAYS FROM THE EARLIEST DATE ON WHICH THE BUYER COULD REASONABLY HAVE DISCOVERED THE ALLEGED DE­FECT OR BREACH, AND NO ACTION FOR THE BREACH OF ANY WAR­RANTY SHALL BE COMMENCED BY THE BUYER ANY LATER THAN TWELVE MONTHS FROM THE EARLIEST DATE ON WHICH THE BUYER COULD REASONABLY HAVE DISCOVERED THE ALLEGED DEFECT OR BREACH.
RETURN POLICY
For a period of thirty-six (36) months from the date of shipment, and under normal conditions of use and service, Moore Industries ("The Company") will at its option replace, repair or refund the purchase price for any of its manu­factured products found, upon return to the Company (transportation charges prepaid and otherwise in accordance with the return procedures established by The Company), to be defective in material or workmanship. This policy extends to the original Buyer only and not to Buyer's customers or the users of Buyer's products, unless Buyer is an engineering contractor in which case the policy shall extend to Buyer's immediate customer only. This policy shall not apply if the product has been subject to alteration, misuse, accident, ne­glect or improper application, installation, or operation. THE COMPANY SHALL IN NO EVENT BE LIABLE FOR ANY INCIDENTAL OR CONSE­QUENTIAL DAMAGES.
Tel: (818) 894-7111 • FAX: (818) 891-2816 Tel: (02) 8536-7200 • FAX: (02) 9525-7296
© 2006 Moore Industries-International, Inc.
United States • info@miinet.com
Australia • sales@mooreind.com.au
Belgium • info@mooreind.be
Tel: 03/448.10.18 • FAX: 03/440.17.97
Tel: 86-21-62491499 • FAX: 86-21-62490635
The Netherlands • sales@mooreind.nl
Tel: (0)344-617971 • FAX: (0)344-615920
Specifications and Information subject to change without notice.
China • sales@mooreind.sh.cn
United Kingdom • sales@mooreind.com
Tel: 01293 514488 • FAX: 01293 536852
Loading...