PROGRAMMING AND RUNNING THE PUMP
THESE INSTRUCTIONS ARE NOT A SUBSTITUTE FOR THE OPERATOR’S MANUAL
Be sure to read the operator’s manual thoroughly prior to operating the
pump. See other side for delivery set priming and loading instructions.
STEP 1
Hang feeding container so that the bottom of the bag
is at or above the level of the pump door.
STEP 2
Press and hold ON/OFF key for 1.5 seconds to turn pump on. After
running the self test, display will show last programmed rate.
STEP 3
Program the prescribed rate using the ( + ) and ( – ) keys.
STEP 4
Press RATE/DOSE key to display dose. Program the
prescribed dose using the ( + ) and ( – ) keys.
STEP 5
Press the FOOD TYPE key to display type of food
to be delivered. Press the ( + ) and ( – ) keys to
change between formula and human milk.
STEP 6
Insert stepped connector into patient’s feeding tube.
4 For a feeding tube without the ENFit® connection system,
use pre-attached transitional connector to connect set to
the feeding tube. Note: If transitional connector has become
detached from the set or is loose, screw it clockwise onto
the set firmly by hand. Be careful not to overtighten.
4 For a feeding tube with the ENFit® connection
system, remove pre-attached transitional connector
from set and screw set clockwise into feeding tube
firmly by hand. Be careful not to overtighten.
STEP 7
Press RUN/PAUSE key to start the pump.
STEP 8
When feeding is complete, press ON/OFF key
for 1.5 seconds to turn the pump o.
NOTE: User must confirm each parameter (rate, dose, and food type) before pump
will start. If dose and food type are viewed prior to pressing the RUN/ PAUSE
key, pump will alarm and display CON FIRM DOSE AND FOOD TYPE.
NOTE: Rate, dose, and food type settings can be locked so that changes cannot be made. See the User
Preference Settings section of the Infinity Orange Operator’s Manual for additional information.