Track-It TM USB Blind Data Logger Instructions
The Track-It
data logger that can record up to 2 channels of data. It has a wide range of
triggering and alarm options as well as differing input options. It has a built in
multicolor LED to view status in the field. It works in conjunction with the Track-It
TM
PC Software. The unit plugs directly into a USB port on the PC to allow
programming, upload of data and to display data on the PC in real time (certai n
modes only).
Push Button/LED functionality: The push button can be used to check status or to change the state of the data logger. For instance,
it can be used to start and stop recording if desired. This feature can also be disabled to ensure the data logger state remains
unchanged until connected to a PC.
Press and Hold, release when LED turns steady green: Changes record state on/off
Press and Hold, release when LED turns steady red: Reset alarm indication
Press and Hold, release after LED goes off: No change
TM
Data Logger is an extremely versatile compact battery powered
TO CHECK STATUS
Short press and release:
LED will flash in various patterns to indicate data logger status:
The flash sequence is: Battery status > Record status > Alarm status
One long red flash at the beginning of the flash sequence indicates low batter y. (no red flash = battery ok)
One long red flash at the end of the flash sequence indicates an alarm. (no red flash = no alarm)
Slow red/green blink: Connect to PC. Unit is NOT set up or memory is full.
TO CHANGE STATE
The table below summarizes the push-button / LED functionality.
STATUS—Press & Release Button
Lo Batt. >>>> Record >>>>> Alarm
CHANGE STATE—Press & Hold Button
Release
Release
Release after LED’s go off
LED Pattern
LED Sequence
Low battery - one RED blink (No red if battery OK)
Recording or triggered to record - fast GREEN/RED blink
Idle/Not recording - one long GREEN blink
Alarm occurred - one long RED blink (No red if no alarm)
Not set up or memory full - slow RED/GREEN blink, connect to PC
LED pattern
Release button while LED GREEN - Turn record mode on/off*
Release button while LED RED - Reset alarm indication*
Release button when LED goes off - No change
*NOTE: Feature must be enabled in software
RESULT OF BUTTON RELEASE
STATUS
Analog Input
Most Track-It TM Data Loggers have the ability to measure up to 2 channels of analog inputs (4-20mA, 0-5Vdc etc) using external
analog modules that plug onto the USB connector. These are programmed using the Track-It
NOTE: Analog recordings are made via the USB connector. If possibl e (button programmed) ST OP RECORDING before plugging the
recorder into the PC so as not to record the USB signals, other wise use the Track-It
Device Setup tab as soon as the device is plugged into the PC.
TM
PC software.
TM
software "Stop Recording" button under the
Rev: 0.92 06/22/11

Protection
The data Logger comes with a rubber boot t hat can be installed to protect the USB
connector and provide some degree of bump and splash protection. The unit shou ld
have the boot installed when traveling. The button can be o perated when the b oot is
in place. The boot cannot be fitted when analog modules are attached. The unit
should not be immersed in liquid.
TM
Track-It
Software
The Track-It TM PC Software allows the data logger to be programmed, allows the data to be retrieved, displa yed, saved to disk or
exported to Excel
TM
. Data can also be displayed in real time (on device s using internal sensors). The software has both simple and
advanced modes for programming data loggers. Some of the features are:
• Delayed recordings, fixed duration recordings by time or numb er of samples, multiple record times, manual record by button
press, record on alarms.
• Sample rates from 2 seconds to 24 ho urs. Instantaneous, average, maximum or minimum values.
• Two Alarms, high or low, latched or momentary and record under these alarm conditions
• Input scaling and offset for analog modules. Engineering unit selection for internal sensors.
• Button function - record off/on, status indication, reset alarm indication.
• Display of data graphically, digitally or tabular formats with alarm indication.
• Export user selected data in Excel
TM
or CSV formats. Filter data to be exported.
• Simple setup (single screen) or advanced mode.
Replacing the Battery
To replace the battery - loosen the two screws on the back of the Logger and remov e the bottom case half. Cauti on - the screws have
small plastic washers. There are two battery options. Replace the relevant battery then replace the cover and the screws.
Option 1 - Coin Cell CR2032 3.0V Option 2 - Lithium Thionyl Chloride EF651625 3.6V (LTC-7PN)
Rev: 0.92 06/22/11