Monarch 6125-010 User manual

Page 1
MONARCH INSTRUMENT
Instruction Manual
Pocket Laser Tach 200
(PLT200)
Tachometer / Rate Meter / Totalizer / Timer
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 2
SAFEGUARDS AND PRECAUTIONS
LASER RADIATION
AVOID DIRECT EYE EXPOSURE
CLASS 2 LASER PRODUCT
MAX OUTPUT POWER: 1mW
EMITTED WAVELENGTH: 650nm
CLASSIFIED TO IEC 60825-1:2001
WARNING - This product emits a visible beam of laser light. Avoid exposure to the laser radiation. The use of optical viewing aids (binoculars, for example) may increase the ocular hazard.
CAUTION - The laser beam should not be intentionally aimed at people or animals.
CAUTION - Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
Read and follow all instructions in this manual carefully, and retain this manual for future reference.
Do not use this instrument in any manner inconsistent with these operating instructions or under any conditions that exceed the environmental specifications stated.
This instrument is not user serviceable. For technical assistance, contact the sales organization from which you purchased the product.
In order to comply with EU Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE): This product may contain material
which could be hazardous to human health and the environment. DO NOT DISPOSE of this product as unsorted municipal waste. This product needs to be RECYCLED in accordance with local regulations, contact your local authorities for more information. This product may be return­able to your distributor for recycling - contact the distributor for details.
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 3
TABLE OF CONTENTS
TABLE OF CONTENTS
1.0 OVERVIEW .......................................................................... E-1
2. 0 FEATURE LOCATIONS ......................................................... E-1
3.0 LCD DISPLAY SYMBOLS .................................................... E-2
4.0 PLT200 SPECIFICATIONS ...................................................E-3
5.0 INPUT / OUTPUT ................................................................. E-7
6. 0 PREPARATION FOR MEASUREMENT ................................. E-8
6.1 Non-Contact Preparation .......................................... E-8
6.2 Direct Contact Preparation....................................... E-8
6.3 Connecting External Sensors .................................. E-9
7. 0 TAKING MEASUREMENTS ................................................E-10
7.1 Non-Contact Measurements .................................. E-10
7.2 Direct Contact Measurements ............................... E-10
8.0 TACHometer Mode........................................................... E-11
8.1 TACHometer Setup .................................................. E-11
8.2 TACHometer Operation ...........................................E-13
9.0 RATE Mode........................................................................ E-13
9.1 RATE Setup .............................................................. E-14
9.2 RATE Operation ....................................................... E-16
10.0 TOTALizer Mode ...............................................................E-17
10.1 TOTALizer Setup......................................................E-17
10.2 TOTALizer Operation ...............................................E-20
11.0 TIMER Mode ......................................................................E-21
11.1 TIMER Setup .............................................................E-21
11.2 TIMER Operation ......................................................E-22
12.0 BATTERIES ........................................................................E-23
13.0 CLEANING .........................................................................E-24
14.0 OPTIONS /ACCESSORIES ................................................E-24
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 4
1.0 OVERVIEW
The Pocket Laser Tach 200 is a multifunction Tachometer, Ratemeter, Totalizer and Timer. It is programmable to read in English or Metric units. An input socket accepts remote sensing devices and an output socket allows for pulse output to external indicating devices. The PLT200 can be tripod mounted and “Locked-On” for accurate and continuous operation. This tachometer also stores minimum, maximum and last measurement in memory.
2.0 FEATURE LOCATIONS
LCD display
Min / Scroll Down arrow
Tripod mounting bushing (underside)
Start / Reset button
Belt clip
Output socket
Max / Scroll Up arrow
Battery compartment
Menu / Select and Lock-on button
Input socket
AVOID EXPOSURE - LASER RADIATION IS EMITTED FROM THIS APERTURE
E-1
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 5
3.0 LCD DISPLAY SYMBOLS
On Target Indicator. Blinks on whenever there is an input signal. Will appear to be solid on at higher frequencies.
Low Battery icon. Indicates that the batteries are low and need to be replaced.
Times Ten icon. Indicates that the value shown is ten times that which is displayed.
Laser Indicator. Red laser is on when this indicator is illuminated.
Lock icon. Indicates that the unit is “Locked” on and making continuous measurements (Lock mode).
E-2
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 6
4.0 PLT200 SPECIFICATIONS
Laser Specifications:
Classification: Class 2 (per IEC 60825-1 Ed 1.2 2001-8)
Complies with FDA performance standards for laser products except for deviations pursuant to Laser Notice No. 50, dated July 26, 2001.
Maximum Laser Output: 1mW Pulse Duration: Continuous Laser Wavelength: 650 nm Beam Divergence: < 1.5 mrad Beam Diameter: 4 x 7 mm typical at 2 meters
Laser Diode Life: 8,000 operating hours MTBF (1 year
warranty)
Non-Contact Specifications:
Ranges: RPM 5 – 200,000
RPS 0.084 – 3,333.3
RPH 300-999,990
Resolution: Fixed: 1 (10 above 99,999)
Auto-ranging: 0.001 to 1.0 (10 above 99,999)
Accuracy: ±0.01% of reading or resolution limit
Operating Range: up to 25 feet (7.62 m) or up to 70 degrees off
perpendicular to T-5 tape target
Contact Specifications using optional Remote Contact Assembly:
Range: Contact Tips: 0.5 to 20,000 RPM
10 cm / 12-inch Wheel: 0.5 to 12,000 RPM
Resolution: Fixed: 1 (10 above 99,999)
E-3
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
Auto-ranging: 0.001 to 1.0 (10 above 99,999)
www.GlobalTestSupply.com
Page 7
Contact Specifications (continued):
Accuracy: Revs: ±0.05% of reading (RPM) or resolution limit
(with no slippage)
Linear: ±0.5% of reading or resolution limit (with no
slippage)
Contact Measurements Ranges:
TACHOMETER:
Revolutions per Minute (RPM) 0.5 to 20,000 RPM Revolutions per Second (RPS) 0.0833 to 333.33 RPS
Revolution per Hour (RPH) 30 to 999,990 RPH
RATES: Wheel Circumference:
Inches per Second 10 cm: 0.033 to 1312.3 IPS
12 in: 0.100 to 2,400.0 IPS
Inches per Minute 10 cm: 1.969 to 78,740 IPM
12 in: 6.000 to 144,000 IPM
Inches per Hour 10 cm: 118.11 to 999,990 IPH
12 in: 360.00 to 999,990 IPH
Feet per Second 10 cm: 0.003 to 109.36 FT/S
12 in: 0.009 to 200.00 FT/S
Feet per Minute 10 cm: 0.164 to 6,561.7 FT/M
12 in: 0.500 to 12,000 FT/M
Feet per Hour 10 cm: 9.843 to 393,700 FT/H
12 in: 30.000 to 720,000 FT/H
Yards per Second 10 cm: 0.001 to 36.453 YPS
12 in: 0.003 to 66.667 YPS
Yards per Minute 10 cm: 0.055 to 2,187.2 YPM
12 in: 0.167 to 4,000.0 YPM
E-4
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 8
Contact Measurements Ranges (continued):
RATES: Wheel Circumference:
Yards per Hour 10cm: 3.281 to 131,233 YPH
12 in: 10.000 to 240,000 YPH
Miles per Hour 10 cm: 0.002 to 74.564 MPH
12 in: 0.006 to 136.36 MPH
Centimeters per Second 10 cm: 0.084 to 3,333.3 CM/S
12 in: 0.21 to 3,048.0 CM/S
Centimeters per Minute 10 cm: 5.000 to 200,000 CM/M
12 in: 15.240 to 365,760 CM/M
Centimeters per Hour 10 cm: 300.00 to 999,990 CM/H
12 in: 914.40 to 999,990 CM/H
Meters per Second 10 cm: 0.001 to 33.333 M/SEC
12 in: 0.003 to 60.960 M/SEC
Meters per Minute 10 cm: 0.050 to 2,000.0 M/MIN
12 in: 0.153 to 3,657.6 M/MIN
Meters per Hour 10 cm: 3.000 to 120,000 M/H
12 in: 9.144 to 219,460 M/H
TOTALIZER:
Counts: 0 to 999,999 Scale Totals in Inches, Feet, Yards, Centimeters or Meters
Input: Internal or External optics or linear contact wheel
Timer Specifications:
Minutes:Seconds.Tenths to 99:59.9
Accuracy: ±0.2 second
Resolution: 0.1 second
E-5
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 9
Display: Dual LCD Display (5-digit upper/scrolling, 5-digit
alphanumeric lower display)
Batteries: 2 “AA” 1.5 V (DC) alkaline included
(Note: Batteries are NOT rechargeable.)
Battery Life: 30 hours continuous typical with batteries provided
External Input:
Absolute max: -0.3 V to 5 V (DC)
Minimum: low below 1.2 V and high above 2 V (TTL compatible)
Edge: Triggers on Positive edge
Power Out: 3.0 V nominal, approx. 2.8 V @ 20 mA max
Pulse Output: 0 V to 3.3 V (DC) pulse
Same shape as External Input signal or high when internal optics sees a reflection
Dimensions: 6.92” (17.58 cm) H x 2.4” (6.10 cm) W x 1.6” (4.06 cm) D
Weight: Approx. 7 oz. (210 g)
This product is designed to be safe for indoor use under the following conditions (per IEC61010-1).
Installation Category II per IEC 664
Pollution Degree Level II per IEC 664
Temperature: 40 °F to 105 °F (5 °C to 40 °C)
Humidity: Maximum relative humidity of 80% for temperatures up
to 88 °F (31 °C) decreasing linearly to 50% relative humidity at 100 °F (40 °C). Humidity non-condensing.
Specifications subject to change without notice.
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
E-6S-17
Page 10
5.0 INPUT / OUTPUT
Input socket (
Output socket (
Input: Accepts remote sensor or
Remote Contact Assembly (RCA). 1/8” (3.5mm) stereo phone plug.
Output: 1 pulse per revolution TTL
output on internal operation. Pulse repeater with external sensors. 1/8” (3.5mm) mono phone plug.
TT
T)
TT
SS
S)
SS
Signal Input
+3V Out to
Sensor
Common
(GND)
Common (GND)
+3V Out to Sensor
Signal Input
Input Connector Detail (Stereo plug)
Pulse Output
Common
(GND)
Common (GND)
Pulse Output
Output Connector Detail (Mono plug)
E-7
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 11
6.0 PREPARATION FOR MEASUREMENT
6.1 Non-Contact Preparation
For Internal operation (Red laser) or External operation using optional Remote Optical Sensor (ROS-Red LED).
1. Clean Shaft
2. Apply 1/2” square T-5 Reflective tape
For Small Shafts:
As small as 1/8” wide on side or radius edge
6.2 Direct Contact Preparation
For External operation ONLY using optional Remote Contact Assembly (RCA).
Select and install contact option:
1. Contact Tip (Convex tip shown.
Use Concave tip for small shafts.)
Align flats
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
E-8
Page 12
2. 10 cm Wheel 3. 12 inch Wheel
OR
Install with pin in shaft
Tighten screw securely into flat on shaft.
fully seated in slot. Tighten screw.
6.3 Connecting External Sensors
Plug sensor into
Input socket
SS
S
SS
In
(shown with optional 12 inch wheel)
Remote Contact
Assembly (RCA)
Remote Optical
Sensor (ROS-P)
E-9
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Infrared Sensor
(IRS-P)
Magnetic Sensor
with Amplifier
(MT-190P)
Page 13
7.0 TAKING MEASUREMENTS
7.1 Non-Contact Measurements
Hand-held
Laser
External Sensor (ROS shown)OR
Light
7.2 Direct Contact Measurements
Rotational Linear
(Use concave tip for small shafts)
to RCA
from PLT200 from PLT200
ONLY USE MODERATE PRESSURE
WARNING: Making measurements in direct contact with rotating equipment can be dangerous. Keep all loose clothing and hair away from exposed moving machinery. Keep the hand holding the instrument well behind the back end of the Remote Contact Assembly. Properly replace all machinery guards after completing measurement. Do not use for rotation greater than 20,000 RPM.
E-10
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 14
8.0 TACHometer Mode
Tachometer measures speed or linear rate with respect to time. Time intervals are seconds, minutes, or hours. Rotational speed can be measured in Revolutions (Revs) per second, per minute, or per hour. The most common measurement is RPM or Revs per minute using the optical tachometer mode.
8.1 TACHometer Setup
1. Turn Power
Last Units selected are displayed
ON
1a. To toggle Lock
On/Off
Press and Hold
Locked On
2. Enter Setup
3. Enter selection
Last Mode selected is displayed
of Mode
4. Select TACH
OR
Repeat until TACH displayed
Mode
5. Save and advance
E-11
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 15
6. Enter selection of Units
RPS, RPM or RPH
7. Select Units
OR Repeat until
desired Units displayed
8. Save and advance
9. Enter selection
NONE, 1, 2 or 3
of number of decimal places
10. Select decimal places
OR Repeat until
desired decimal places displayed
11. Save and advance
12. Exit Setup – Ready to
DONE, then Units selected
measure
Unit will remember these settings (including lock on/off) even if turned off and back on.
E-12
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 16
8.2 TACHometer Operation
Measure
Press and hold
Recall Max
Recall Min
If unit Locked
OR
Lock on
Max Speed
Min Speed
Resets Max/Min
on:
Power OFF
Automatic after 90 seconds
OR
if unit not Locked on
9.0 RATE Mode
Measurement of units in addition to Revs requires the attachment of the Remote Contact Assembly and tips/wheels. With this attachment, the unit can measure RATE inputs-revs, inches, feet, yards, centimeters and meters either per second, per minute or per hour, as well as miles per hour.
NOTE: External Remote Contact Assembly (RCA) must be inserted into
input socket.
E-13
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 17
9.1 RATE Setup
1. Turn Power ON
1a. To toggle Lock
On/Off
2. Enter Setup
3. Enter selection of Mode
4. Select RATE Mode
5. Save and advance
6. Enter selection of Units
Press and Hold
Linear: IPS, IPM, IPH, FT/S, FT/M, FT/H,
YPS, YPM, YPH, MPH, CM/S, CM/M, CM/H, M/SEC, M/MIN, M/H
EXTRN, then scrolling
message, then last Units selected
Locked On
Last Mode selected is displayed
OR Toggles between
RATE and TOTAL. Select RATE.
Rotational: C RPS, C RPM or C RPH
E-14
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 18
RATE Setup (continued):
7. Select Units
8. Save and advance
Only for Linear Units: 8a. Enter
selection of Wheel
8b. Select
Wheel
8c. Save and
Advance
9. Enter selection of number of decimal places
10. Select decimal places
OR Repeat until
desired Units displayed
OR
Rotational Units
Linear Units
Last Wheel selected is displayed
OR Toggles
between 10CM and 12IN
NONE, 1, 2 or 3
OR
Repeat until desired decimal places displayed
11. Save and advance
E-15
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 19
12. Exit Setup – Ready to measure
DONE, USE CONTACT TIP or
[wheel selected], then Units selected
Unit will remember these settings (including lock on/off) even if turned off and back on.
9.2 RATE Operation
Measure
OR
Press and hold Lock on
Recall Max
Recall Min
If unit Locked
Resets Max/Min
Max Speed
Min Speed
on:
Power Off
Automatic after 90 seconds
OR
if unit not Locked on
E-16
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 20
Despliegue: 5 x 0.5” (12.7mm) dígitos numéricos más 5 alfanuméricos LCD
Baterías: 2 “AA” 1.5 V (CC) alcalinos incluidas
(Notese: Baterías NO se pueden volver a cargar.)
Vida de Batería: típicamentes 30 horas continuas con las baterías
suministrados
Entada externa:
Máximo absoluto: -0.3 V a 5 V (CC) Mínimo: bajo a menos de 1.2 V y alto por encima de 2 V
(compatible con TTL)
Borde: Cambia en borde positivo
Potencia Salida: 3.0 V nominal, aprox. 2.8 V @ 20 mA max
Salida Pulso: 0 V a 3.3 V (CC) pulso, Misma forma que la señal de
entrada externa o alta cuanda la óptica interna ve una réfleccion
Dimensiones:6.92” (17.58 cm) de alto x 2.4” (6.10 cm) de ancho x 1.6”
(4.06 cm) de profundo
Peso: Aprox. 7 onzas. (210 g)
Este producto fue concebido para ser segura para uso interno bajo las siguientes condiciones (según IEC61010-1).
Categoría de instalación II según IEC 664
Grado de nivel de contaminación II según IEC 664
Temperatura: 40 °F a 105 °F (5 °C a 40 °C)
Humedad: Humedad relativa maxima de 80% para temperatures hasta
88 °F (31 °C) disminuyendo linearmente hasta un 50% de humedad relativa a una temperatura de 100 °F (40 °C). Humedad sin condensar.
Especificaciones sujetos a cambios sin aviso previo.
S-6
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 21
5. Save and advance
6. Enter selection of
Different options displayed for Internal or External operation.
Units
Internal or External ROS: External Remote Contact Assembly:
COUNT
Only
Rotational: REV Linear: INCH, FEET, YARDS, CM, METER
7. Select Units
OR Repeat until
8. Save and advance
OR
COUNT or REV Linear Units
Only for Linear Units: 8a. Enter selection
of Wheel
8b. Select Wheel
OR
8c. Save and
Advance
desired Units displayed
Last Wheel selected is displayed
Toggles between
10CM and 12IN
E-18
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 22
TOTALizer Setup (continued):
9. Enter selection
NONE, 1, 2 or 3 of number of decimal places
10. Select decimal places
OR
Repeat until desired decimal places displayed
11. Save and advance
12. Exit Setup – Ready to measure
Units = COUNT: Rotational/Linear Units: DONE,
then COUNT
DONE, USE CONTACT TIP or
[wheel selected], then Units selected
Unit will remember these settings (including lock on/off) even if turned off and back on.
E-19
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 23
10.2 TOTALizer Operation
Measure
Recall Max or Min
Recall Time in seconds
If unit is Locked on:
Power Off
NOTE:
Pressing
OR
Press and hold Lock on
Resets Max/Min, Total and Measure­ment Time
Automatic after 90 seconds
OR
if unit not Locked on
OR
Max or Min Speed (in last selected Tach or Rate mode units)
Shows time in seconds from when the Start / Reset button is pressed until the last input signal measured.
once before 90 seconds will keep measurements in memory and the display turned on longer.
E-20
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 24
11.0 TIMER Mode
Accumulates time in minutes, seconds, and tenths of a second. There are two modes of operation. The Manual mode operates like a stopwatch, the timing period being started and stopped by the user. The Auto mode can be stopped and started by the user or a piece of reflective tape on objects. The user can freeze the display-and view/ record a LAP time-at any time without affecting the count.
11.1 TIMER Setup
1. Turn Power ON
Last Units selected are displayed
1a. To toggle Lock
On/Off
Press and Hold
Locked On
2. Enter Setup Mode
3. Enter selection of Mode
4. Select TIMER Mode
OR
Last Mode selected is displayed
Repeat until TIMER displayed
5. Save and advance
E-21
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 25
6. Enter selection of Timer function
MAN or AUTO
7. Select Timer function
OR Toggles between
Manual and Auto
8. Save and advance
9. Exit Setup – Ready to
DONE, then Units selected
measure
Unit will remember these settings (including lock on/off) even if turned off and back on.
11.2 TIMER Operation
Measure:
Manual
Each press toggles Start and Stop
Auto
Start and Stop triggered by
OR
external remote optical sensor (ROS) or internal optics
Reset
With Timer stopped ­Resets time to 00:00.0
E-22
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 26
TIMER Operation (continued):
Lap
Power Off
12.0 BATTERIES
When displayed, replace batteries.
Remove battery cover
With Timer running ­Stops at elapsed time to date. To continue, press again.
OR
OR
Install two 1.5V “AA” alkaline batteries
If Timer stopped ­Automatic after 90 seconds (if unit not Locked on) Automatic after 99:59.9
NOTE: Both batteries face the same direction.
E-23
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Page 27
13.0 CLEANING
To clean the instrument, wipe with a damp cloth using mild soapy solution.
14.0 OPTIONS /ACCESSORIES
T- 5 Reflective Tape, 5 foot [1.5 m] roll, ½ inch [13 mm] wide RCA Remote Contact Assembly with 10 cm wheel, concave
and convex tips
CTE Concave/convex contact tips and 10 cm linear contact
wheel
12 inch Wheel 12 inch circumference wheel for use with RCA CA-4044-6 6 foot Input/Output cable, 1/8” mono phone plug to
BNC connector
ROS-P Remote Optical Sensor ROS-P-25 Remote Optical Sensor with 25 foot cable ROSM-5P Remote Optical Sensor, modulated MT-190-P Amplified Magnetic Sensor IRS-P Infrared Sensor EC-25P 25 foot extension cable for all sensors CC-10 Padded Nylon Carrying Case CC-11 Latching Carrying Case for Pocket Tach and accessories CAL-N.I.S.T. N.I.S.T. Traceable Certificate of Calibration
E-24
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Loading...